the physics of medical therapy acceleratorshpschapters.org/.../pds/8aphysmedther_coutrakon.pdf ·...
TRANSCRIPT
The Physics of Medical Therapy Accelerators
A three part lecture series byGeorge Coutrakon, PhD
Asst. Professor in School of MedicineLoma Linda University Medical Center
Dept. of Radiation MedicineLoma Linda, California
Course Outline Feb 1st , 2008
• Introduction to Therapy Machines• What are the technical goals for hitting the target
with high doses of radiation?• What are the beam delivery systems that give the
most conformal dose to the target? • How do we shape the beam to hit the target with
good conformity both for protons and for X-rays?• How do these clinical requirements translate into
accelerator requirements?
Course Outline Feb. 2nd ( in 2 parts)
• Physics of electron and proton medical accelerators
• The physics of 3D intensity modulated (x-ray) radiation therapy (IMRT)
• Theory of operation of cyclotrons and synchrotrons for therapy
• Examples of commercial proton systems• Heavy Ion Accelerators and the push for Heavy
Ion Therapy and the Heidelberg Ion Therapy project
Electron linear accelerator mounted on a gantry
X-ray and electron treatment head
Requirements for Radiation Therapy
• Beams must penetrate the body deep enough for shallow and deep tumors ( 0 to 35 cm in water equivalent material
• Field sizes should be adjustable from 1 cm diameter up to 30 cm x 30 cm
• Dose should be uniform to +/- 3% over the tumor volume• Dose fall off at boundary of tumor volume should be
sharp, typically 90% of maximum dose to 10% of maximum dose in less than 1 cm.
• Need to treat 1 to 2 Gy/minute for a typical daily dose of 2 Gy.
• The national average for full course of treatments is about 30 daily fractions with 5 fractions per week.
Patient Treatment set-up for cranial tumors
Dose vs. Depth for typical clinical x-ray and electron beams
Dose vs. Depth in Tissue for P+, e-
and X-Rays
Dose = (Np/A) (1/ρ) dE/dx
dE/dx = 4π z2 e4 NA ρ (Z/A) (Log f(β) - β2)/mec2 β2
Energy Loss of Charged Particles in matter vs. particle momentum
Energy Loss of charged particles in matter vs. particle momentum
∫=0
0
)//(E
dxdEdER
50 to 250 50 to 250 MeVMeV is clinical range of is clinical range of interest interest
Eye Tumors at 50 Eye Tumors at 50 MeVMeV; Prostate Tumors at 250 ; Prostate Tumors at 250 Proton Energy vs Range in Tissue
05
10152025303540
0 100 200 300
Energy (MeV)
Ran
ge (c
m)
Series1
Proton intensity requirements for therapy (derived from energy loss tables, Janni, 1982)
Result: need about 1 x 109 protons/cm for 1 Gy
Requirements for Therapy Medical Accelerators
• Create beams that give the highest dose conformity to tumor and spare nearby healthy tissue. Maximize ratio of tumor control/side effects
• Reliability for high throughput--- 95% uptime required• a) Linear accelerators can treat up to 35 patients per day at 20
minute intervals• b) Proton accelerators do not fit on gantries and feed beams to 4 or 5
rooms that each can treat 40 patients per day• Energy range – Need variable energy depending on tumor depth • a) Electrons 6 – 25 MeV• b) Protons 70 – 250 MeV• c) Carbon ions 100 to 430 MeV/amu• Beam Current for 1 to 2 Gray/minute @ 1 – 2 min./treatment• a) Electrons: 300 μA to make Xrays, 1 μA for e- treatments• b) protons: 3-5 nAmps or 1 x 1012 protons/ minute• c) carbon: 3 x 109 C6+ per minute
Accelerator Requirements (con’t)
• Beam energies must be changed frequently throughout the day
• Good intensity uniformity over the treatment duration• Compact to fit in hospital environment
Accelerator requirements (con’t)
• Beam energies must be changed frequently throughout the day with good stability
• Good intensity uniformity over the treatment duration• Compact to fit in hospital environment• To achieve these goals, patients should have rotatable beam
lines called gantries that can shoot beams from any angle to a common intersection point ( called isocenter)
• The beam energy should be variable to achieve targeting at a variety of depths in the patient
• The beam delivery system will dictate the accelerator requirements, to some extent
Beam Delivery Systems (X-rays)Two modes of beam delivery in X-ray treatment are generally used
1. Conventional X-ray treatment (CRT) which uses 2 to 6 portals with fixed collimators
2. Intensity Modulated Radiation Treatment (IMRT)a) use 10 to 30 portals in 50 gantry increments called “step and shoot”b) within each step, use multiple collimator openings to modulate the beam intensity across the field. Beam energy and intensity from the accelerator are held constant.
Beam delivery systems for Hadron TherapyBoth need fewer portals per target due to the increased dose sparing of
the Bragg Peak. 1. Deliver uniform or equal SOBP along any ray trajectory throughout the
target. This fixed modulation usually accomplished by passive scattering and passive range modulation techniques
Hadron Beam Delivery Systems (con’t)
2. Conform the SOBP to the shape of the target both on the distal side and on the entry ( proximal side). This is called variable modulation since the width of the SOBP is not the same for all “rays” through the targetThis is usually accomplished using beam scanning techniques and “painting” each layer of the target with a unique energy. The energy variation is produced at the acceleraorrather than in the nozzle
Superposition of Bragg Peaks
Passive beam spreading vs. dynamic beam spreading techniques
Lateral Beam Spreading using Pb
Beam Spreading using Magnetic Deflection of Pencil Beams
Fast and Slow Scanning Magnets at LLUMC
Patient in Treatment Position in Proton Gantry
Dose distributions to prostate gland using 6 field x-rays (CRT) and 2 field protons
Propellor for Range Modulation
LLUMC Beam Delivery System
Example of spot scanning technique for Intensity Modulated Proton Therapy (IMPT)
Additional Accelerator Requirements for Hadron Beam Scanning
• Programmed sequence of energies in small (1 – 2 MeV steps) to deliver SOBP.
• Ability to change energies at the accelerator in several seconds
• Increased accuracy of accelerated energy (+/- 0.2 mm of range corresponding to about +/- 0.2 MeV.
• Programmed sequence of intensities from the accelerator with dynamic range approx. 10:1