the lyos™ control system
TRANSCRIPT
![Page 1: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/1.jpg)
The LyoS™ Control System
FOR LYOSTAR™ 3 LYOPHILIZERS
![Page 2: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/2.jpg)
Welcome to LyoS™
![Page 3: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/3.jpg)
Welcome to LyoS™
• Pronounced \‘lī-,ōs\. • Windows®-based platform. • Leverages GE Proficy HMI/SCADA iFIX. • Provides visualization, data acquisition
and supervisory capabilities directly to an Allen-Bradley CompactLogix PLC.
![Page 4: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/4.jpg)
Key Features
• Automated Leak and Function Testing • Automatic Backfill / Automatic Stoppering • Barometric Endpoint Testing (PRISE) • Pirani vs. Capacitance Manometer
differential testing (PRCM)
![Page 5: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/5.jpg)
Sophisticated Instrumentation Set
• Sixteen (16), Type T product temperature-monitoring thermocouples.
• Two (2) Capacitance Manometers (product and condenser chambers).
• Pirani vacuum transducer (product chamber).
![Page 6: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/6.jpg)
Optional Cycle Optimization Tools
• SMART™ Freeze Dryer Technology (optional).
• ControLyo™ Nucleation On-Demand Technology (optional).
![Page 7: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/7.jpg)
Fully Validateable
• Fully configurable and fully validatable for regulated pharmaceutical and biotechnology applications.
• SP Scientific offers onsite control system IQ/OQ validation support and a more comprehensive factory-initiated System Integration Testing (SIT) program.
![Page 8: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/8.jpg)
Documentation and Support
• All original software is provided with the LyoS™ control system.
• A backup copy of the LyoS™ control system is provided.
• PLC source code is available. Requires a signed Confidentiality Disclosure Agreement (CDA).
![Page 9: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/9.jpg)
Control System Hardware / Software Comparison
Control System Feature LyoStar™ 3 LyoStar™ II
Praxair ControLyo™ Nucleation Technology
Optional Not Available
SMART™ Freeze Dryer Technology
Optional (Available June 2011)
Optional
Programmable Logic Controller (PLC)
Allen-Bradley CompactLogix
Opto 22
Vacuum Sensor Configuration (Standard)
2 Capacitance Manometers 1 Pirani Vacuum Gauge
1 Capacitance Manometers 2 Pirani Vacuum Gauge
![Page 10: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/10.jpg)
Control System Software Feature Comparison
Control System Feature LyoStar™ 3 LyoStar™ II
PVG/CM Differential Feedback Testing
Standard Offering. Not Available.
Barometric Endpoint / Pressure Rise Testing
Standard Offering. Not Available.
Vacuum Ramping During Primary Drying
Standard Offering. Not Available.
Vacuum Preseal Standard Offering. Not Available.
Product Driven Mode Standard Offering. Not Available.
Vacuum Control Using Secondary Gauge
Standard Offering. Not Available.
![Page 11: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/11.jpg)
Basic Navigation (Buttons)
The circular icons located at the top middle and top right corner of the screen provide access to additional screens and functions.
Stop
Back
Play
Alarm Summary
Maintenance
Data Export
Trend
![Page 12: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/12.jpg)
Basic Navigation (Buttons)
• Play. Start or continue an automatic cycle. • Stop. Terminate an automatic cycle. • Back. Return to the previous screen.
![Page 13: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/13.jpg)
Basic Navigation (Alarm Summary Button)
• The Alarm Summary button provides access to a detailed list of all active alarms.
• The button displays the current number of unacknowledged alarms.
• Alarms are automatically saved to the computer’s hard drive.
![Page 14: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/14.jpg)
The Alarm Summary Page
![Page 15: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/15.jpg)
The Alarm Summary Page (Alarm Print)
• Select the Alarm Print button to print
saved alarms. • The Alarm Print dialog appears with a list
of dates. • Alarms are printed by date. Dates are
formatted as yy/mm/dd.
![Page 16: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/16.jpg)
The Alarm Print Dialog
• To print the Alarm Summary, select a date and click Print Alarm.
• To close the Alarm Print dialog, click Cancel.
![Page 17: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/17.jpg)
Basic Navigation (Maintenance Button)
The Maintenance button opens the Maintenance menu dialog. • Enable Component Mode. • Open the Configuration page. • Open the Calibration page.
![Page 18: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/18.jpg)
The Maintenance Menu (Component Mode)
• Component Mode allows you to manually control system components listed on the Synoptic page such as the heater, fluid pump, vacuum pump, compressors, and associated valves.
• Click the Component button to enable. The Synoptic page will open.
![Page 19: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/19.jpg)
The Maintenance Menu (Configuration)
• Select the Configuration button to open the Configuration page.
• Configure settings used by the software to determine how the system will operate, as well react to certain conditions.
• PLC version and system serial number are displayed at the top of the page.
![Page 20: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/20.jpg)
The Configuration Page
![Page 21: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/21.jpg)
The Configuration Page (ControLyo™)
• Systems configured with ControLyo™ Nucleation Technology will include additional settings.
![Page 22: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/22.jpg)
The Maintenance Menu (Calibration)
• Select the Calibration button to open the Calibration page.
• Calibrate the lyophilizer’s temperature sensors (i.e., shelf, condenser, ambient, interstage and product probes).
![Page 23: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/23.jpg)
The Calibration Page
![Page 24: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/24.jpg)
Basic Navigation (Trend Button)
Use the Trend button to launch the Historical Trend screen. • Monitor real-time and historical cycle data. • Customize active charts. • Create new charts.
![Page 25: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/25.jpg)
The Historical Trend Screen
![Page 26: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/26.jpg)
Basic Navigation (Data Export Button)
Use the Data Export button to open the Historical Data Export page. • Export historical data into a comma-
separated value (.CSV) file that can be viewed in any spreadsheet application.
![Page 27: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/27.jpg)
The Historical Export Page
![Page 28: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/28.jpg)
Basic Navigation (Print Button)
• Print. Prints the current full-screen view to
the available printer.
![Page 29: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/29.jpg)
Basic Navigation (Cycle Status)
The Cycle Status pull-down is accessible from the top left corner of the screen.
Cycle Status Pull-Down
![Page 30: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/30.jpg)
Basic Navigation (Cycle Status)
The Cycle Status pull-down displays the real-time status of critical system parameters.
![Page 31: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/31.jpg)
Basic Navigation (Cycle Status)
Select the Probe Avg button to open the Product Probes pop-up window.
![Page 32: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/32.jpg)
Basic Navigation (Tabs)
The LyoS™ control system includes a series of navigation tabs located across the bottom of the screen. To navigate to a particular page, simply click the tab of the page you wish to access.
![Page 33: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/33.jpg)
The Synoptic Page
• A process flow diagram of the complete lyophilization system.
• Individual components may be monitored and/or controlled from this screen.
• Critical system parameters are displayed. • Activate or deactivate individual
components with a click of the mouse.
![Page 34: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/34.jpg)
The Synoptic Page
![Page 35: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/35.jpg)
The Synoptic Page (ControLyo™)
• Systems configured with ControLyo™
Technology include additional process flow items.
![Page 36: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/36.jpg)
The Freeze-Dry Page
• The Freeze-Dry Recipe page allows you to view and edit current recipe parameters, and load a previously saved recipe.
• This page provides access to the Recipe Manager function, which allows you to create and save recipes.
![Page 37: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/37.jpg)
The Freeze-Dry Page
![Page 38: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/38.jpg)
The Freeze-Dry Page (ControLyo™)
Systems configured with ControLyo™ Technology include additional fields below the Freezing/Thermal Treatment steps.
![Page 39: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/39.jpg)
The SMART™ Page
• SMART™ technology utilizes manometric temperature measurement (MTM).
• Automated measurements of the product temperature at the sublimation interface are used to determine ideal parameters for preventing product collapse or meltback during primary drying.
![Page 40: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/40.jpg)
The SMART™ Page
Coming June 2011.
![Page 41: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/41.jpg)
The Leak Test Page
• The Leak Test page is intended to be used when verifying the vacuum integrity of the system.
• Leak Test results are stated in terms of the system’s Rate of Rise, as well as its Volume–Based Leak Rate (VBLR).
![Page 42: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/42.jpg)
The Leak Test Page
![Page 43: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/43.jpg)
The Function Test Page
• Verify the system’s performance against specifications or previous test data.
• Test results include cold and warm probe test data; shelf, condenser and vacuum pull-down rates; lowest shelf and condenser temps, lowest vacuum temps., and shelf heat-up rate.
![Page 44: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/44.jpg)
The Function Test Page
![Page 45: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/45.jpg)
The Semi-Auto Page
• The Semi-Auto page may be accessed to activate or deactivate the semi-auto functions, including Shelf (Shelf Control to setpoint or Heat Only to setpoint), Condenser, Vacuum Control to setpoint, Isolation Valve, Backfill, Vacuum Release and Defrost.
![Page 46: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/46.jpg)
The Semi-Auto Page
![Page 47: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/47.jpg)
The Semi-Auto Page (ControLyo™)
• Systems configured with ControLyo™ Nucleation Technology will include the following Semi-Auto functions.
![Page 48: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/48.jpg)
Additional Layout Items
From left to right, the top of the page layout also includes the following: • SP Scientific logo. • Current security access level. • Question Mark and Information icons. • Date and time.
![Page 49: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/49.jpg)
Security Access Levels
• The control system contains a multi-level, password-protected security system.
• The standard security access levels are identified as Guest, Operator and Supervisor.
![Page 50: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/50.jpg)
Question Mark Icon
• The Question Mark icon indicates a Help entry point.
• When selected with the Synoptic page active, illustrated components are tagged with descriptive callouts. This feature may be particularly useful when operating your lyophilizer in Component Mode.
![Page 51: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/51.jpg)
Question Mark Icon
Descriptive Callout Tags
Question Mark Icon
![Page 52: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/52.jpg)
Information Icon
Selecting the Information icon opens a window displaying a Vapor Pressure over Ice table.
![Page 53: The LyoS™ Control System](https://reader034.vdocuments.mx/reader034/viewer/2022051710/58491b321a28ab002f8b7691/html5/thumbnails/53.jpg)
Date and Time
The current date and time, according to the Windows® operating system clock, is displayed at the top right corner of the control system layout.