the audiometric threshold: measured in-situ, automated, and by the hearing aid gitte keidser, harvey...
TRANSCRIPT
![Page 1: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/1.jpg)
The audiometric threshold: measured in-situ, automated, and by the hearing aid
Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid Yeend, and Lisa Hartley
National Acoustic Laboratories
AHS June 2010Lake Como
![Page 2: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/2.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Background
Conventional audiometry
Most modern hearing aids have on-board generators that produce pure tones in-situ audiometry
Low-cost tests of hearing over the telephone or internet in demand automated audiometry
![Page 3: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/3.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Self-fitting hearing aid (concept)
Audiometer
Prescription Formula
Adjust Hearing Aid
Real-ear to coupler difference
Automatic
Plus trainability
meet high demands in developing countries where services are scarce
![Page 4: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/4.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Questions Is in-situ audiometry valid and reliable?
What is the effect of using different couplings (open vs closed domes) on in-situ threshold measurements?
Is automated audiometry valid and reliable?
![Page 5: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/5.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
In-situ audiometry 24 participants with
known sensorineural hearing loss
Measurements (twice) Conventional audiometry
(insert) Manual in-situ audiometry
(open and closed instant tips)
REDD (insert, open, closed)
![Page 6: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/6.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Thresholds in dB HL
(Test-retest, p = 0.69)
![Page 7: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/7.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
REDD values
p < 0.0000001
(Test-retest, p = 0.58)
![Page 8: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/8.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Thresholds in dB SPL
dB SPL = dB HL + REDD
![Page 9: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/9.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Reliability
![Page 10: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/10.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Reliability
Supra-aural
Valente et al
Hawkins et al
![Page 11: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/11.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Reliability
Supra-aural
Valente et al
Hawkins et al
Insert
Saunders & Morgan
Valente et al
Hawkins et al
![Page 12: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/12.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Automated audiometry
23 participants (46 ears)
Threshold measurements at 0.25, 1, and 4 kHz Manually Automated, adaptive
procedure
AA01 AA02 AA03 AA04 AA05 AA06 AA07 AA08 AA09 AA11 AA12 AA13 AA14 AA15 AA16 AA17 AA18 AA19 AA20 AA21 AA22 AA23 AA24 AA01 AA02 AA03 AA04 AA05 AA06 AA07 AA08 AA09 AA11 AA12 AA13 AA14 AA15 AA16 AA17 AA18 AA19 AA20 AA21 AA22 AA23 AA24
250 1000 4000
Frequency (H z)
-10
0
10
20
30
40
50
60
70
80
Threshold (dB HL)
![Page 13: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/13.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Validity
![Page 14: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/14.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
ReliabilityManual audiometry (5 dB Hughson-Westlake)
Automatic audiometry (2 dB final step size)
Automated audiometry is at least as accurate as manual audiometry
![Page 15: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/15.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Conclusions Is in-situ audiometry valid and reliable?
corrections for coupling required (REDD)ambient noise is controlledopen tips more reliable than closed tips
Is automated audiometry valid and reliable?
Is automated in-situ audiometry valid and reliable?
Most likely
![Page 16: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/16.jpg)
National Acoustic Laboratories, Sydney, Australia Keidser et al.
Acknowledgement
The in-situ audiometry study was partly sponsored by Siemens Instruments
The automated audiometry study was partly sponsored by Australian Hearing
The self-fitting hearing aid concept is now being further explored through the Hearing CRC
![Page 17: The audiometric threshold: measured in-situ, automated, and by the hearing aid Gitte Keidser, Harvey Dillon, Anna O’Brien, Dan Zhou, Lyndal Carter, Ingrid](https://reader036.vdocuments.mx/reader036/viewer/2022070409/56649e8f5503460f94b93bc4/html5/thumbnails/17.jpg)
Thank you for listening!
For further information:www.nal.gov.au