technical seminar on outsourcing sweeteners testing validation... · 2017-09-12 · initial method...

54

Upload: others

Post on 06-Jul-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Confirm to Fit for your Purpose??

What is method validation??

Confirmation, through the provision of objective evidence, that the requirements for a specific intended use or application have been fulfilled

ISO 9000

Confirmation by examination and provision of objective evidence that the particular requirementsfor a specific intended use are fulfilled

ISO/IEC 17025

Verification, where the specified requirements are adequate for an intended use

VIM

What do we get from Method Validation??

To have better understanding on the method performance

such as limitation, scope, working range……

How well of your method?

Demonstrate whether your method

is fit for a particular purpose

Confidence on your results

When I need to have Method Validation??

ISO/IEC 17025

Non-standard methods;

Laboratory-designed /developed methods;

Standard methods used outside their intended scope;

Amplifications and modifications of standard methods.

NATA Technical Note 17

For facilities involved in medical testing, elements of methods

endorsed “research use only” or “not for diagnostic”.

Eurachem

Validation is also required when it is necessary to demonstrate

the equivalence of results obtained by two methods, e.g. a

newly developed method and an existing standard / regulatory

method.

Method Verification

NATA Technical Note 17 - October 2013

To demonstrate by meeting system suitability specifications established for the

method, as well as demonstration of accuracy and precision or other method

parameters for the type of method.

Standard Methods published by international

organization which already validate the

method by collaborative studies.

Verification of methods by the facility must include statistical correlation with

existing validated methods prior to use.

A laboratory using standard methods has to confirm that it has ability to carry

out those methods.

Verification is usually carried out by comparing the performance data obtained

by the laboratory when performing a standard method with those claimed by the

same method.HOKLAS SC No. 20 - 8 July 2014

What should I consider before Method Validation??

a) Purpose of measurement (what is to be identified and

why??)

b) What are the likely sample matrices??

c) Are there any interferences expected, and, if so,

should they be determined??

d) What is the scope (what are the expected

concentration levels or ranges)??

e) Are there any specific legislative or regulatory

requirements??

f) Are there any specific equipment accommodation and

environmental conditions that need to be

considered??

g) What type of equipment is to be used?? Is the method

for one specific instrument, or should it be used by all

instruments of the same type??

h) Method used for the preparation, sub-sampling,

procedure and including equipment to be used??

NATA Technical Note 17 - October 2013

Detection Limit / Quantitation Limit

Linearity and Range

Sensitivity

Selectivity

Accuracy

Precision

Ruggedness

Measurement Uncertainty

Method Performance Characteristics

NATA Technical Note 17 - October 2013

Method Validation vs Method Verification

Limit of Detection (LOD)

Method Detection Limit (MDL)

Instrumental Detection Limit (IDL)

• LOD based on visual evaluation

7 independent replicates, randomised

• LOD based on the standard deviation of the blank

Blanks n ≥ 20 or independent sample blank with n ≥ 10 each

LOD = Blank value + 3s

• LOD based on the range in which the calibration equation applies

yLOD = a+3sy/x= a +bxLOD xLOD = 3 sy/x/b

• LOD based on signal-to-noise

There is no need to estimate the LOD or LOQ for methods that will always be

applied to measure analyte concentrations much greater than the LOQ.

NATA Technical Note 17 - October 2013

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Limit of Detection (LOD)

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

If the test sample measured duplicate

(n=2), sample blank sample measured

once (nb=1)

𝑠𝑜′ = 𝑠𝑜

1

2+1

1

False Positive Rate and False Negative Rate

Limit of Detection (LOD) for qualitative test

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Limit of Quantitation (LOQ)

Common practices are

LOQ = SD x 10; or

= LOD x 3; or

= Blank value + 10 x SD; or

= 50% above the lowest

fortification level used to

validate the method; or

= LOD x 10

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

NATA Technical Note 17 - October 2013

The lowest level of analyte that can be determined

with acceptable performance.

Linearity and Range

i) Confirm the relationship;

ii) Demonstrate the instrument working range is

compatible with the interval stated in the method

scope;

iii) Verify that the proposed instrument calibration

procedure (single point, bracketing, or multiple

points) is adequate.

Instrument Working Range

Method Working Range

NATA Technical Note 17 - October 2013

The following parameters should be considered:

Correlation coefficient (r)

Residual plot

Calibration Curve

- Six calibration solution including blank or close to blank

- Evenly spaced over the range of interest

- The range should encompress 0-150% or 50-150% of the

concentration likely to be encountered (or concentration span

that exceeds the expected concentration range by ± 10% or

even ± 20%)

- The calibration standards should be run at least in duplicate

and preferably triplicate or more, in random order.

y = 1.112x + 0.0435R² = 0.9995

0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

0.00 0.20 0.40 0.60 0.80

-0.015

-0.01

-0.005

0

0.005

0.01

0.00 0.20 0.40 0.60 0.80Residuals

X Variable 1

X Variable 1 Residual Plot

Linearity and Range

Instrument Working Range

Method Working Rangei) Samples with known concentrations and sample blank should be available

ii) The samples used should be taken the entire measurement procedure

iii) The concentrations of the different samples should preferably cover the whole

range of interest

iv) The instrument should have been calibrated according to the suggested calibration

procedure.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Linearity and Range

Instrument Working Range

Method Working Rangei) Samples with known concentrations and sample blank should be available

ii) The samples used should be taken the entire measurement procedure

iii) The concentrations of the different samples should preferably cover the whole

range of interest

iv) The instrument should have been calibrated according to the suggested calibration

procedure.

The method working range needs to be established for each matrix

covered in the method scope.

This is because interferences can cause non-linear responses, and the

ability of the method to extract/recover the analyte may vary with the

sample matrix.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Sensitivity

• Change in instrument response which corresponds to a change

in the measured quantity

• Slope of calibration curve

• Not a particular important performance

• But useful to check of instrument performance and standards as

a quality assurance and quality control procedures.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

NATA Technical Note 17 - October 2013

Interferences

Wrong Identification Positive / Negative Bias

Matrix effect Background

Noise

Selectivity

The selectivity of a procedure must be established for in-house

developed methods, methods adopted from the scientific literature and

methods published by standardisation bodies used outside the scope

specified in the standard method.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Poor Extraction

Selectivity

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Checking Matrix Effect

• Pure standard vs standards taken through the analysis is

indicative of any losses of analyte which are related to the

method, while enhanced results may indicate reagent

contamination.

• Pure standard taken through the analysis compared with pure

standards added to extracted or digested extracts provides an

indication of matrix enhancement or suppression effects on the

detection system.

• Pure standards added to blank matrix after extraction or

digestion, compared to pure standards fortified in matrix prior to

extraction or digestion, provides an indication of losses of analyte

during processing.

NATA Technical Note 17 - October 2013

Peak A

rea R

ati

o

(Ma

trix

)

Peak Area Ratio (Solution)

Interferences

Wrong Identification

Use highly specific procedures /

techniques

Confirmatory

Techniques / Method

Matrix Effect / Background Noise

Selectivity

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

-Isotopic Internal Standard

Calibration

-Matrix Calibration

-Standard Addition

.

.

.

Overspiking with analyte suspectedSANCO/12571/2013 - 01/01/ 2014

Selectivity

“Confirmatory method / technique can be applied to

both the identity and concentration.”

“Inevitably there is a trade-off between costs and time taken for

analyte identification, and the confidence with which one can decide

if the identification has been made correctly.”

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

NATA Technical Note 17 - October 2013

“Confirmation of analyte identity and concentration for positive

samples may be achieved using a different detection system or

column or using a specific detection system or using an alternative

analytical technique. In such cases validation of the confirmatory

technique must also be performed.”

• EU (2002/657/EC)

• Requirement

• ID point (Identification Point) and Ion Ratio.

Selectivity

The variability of ion ratios should preferably be determined from calibration standards during the

initial method validation and subsequently during routine analysis. In certain cases, these data may

be used to set performance-based criterial, for individual analyte, rater than applying the fixed,

generic criteria given in Table 5.SANCO/12571/2013 - 01/01/ 2014

Selectivity

Clenbuterol in Pork LC-MS/MS analysis

Ion ratio determined by

277203 (q) and 277132

Standards Ion ration 0.328

with %RSD 2.95%

Method A by enzymatic digestion

with column A

Ion ratio 0.380

(within 30% of tolerance)

Results by quantifying 277203 7.94 ppb

Method B by perchlorate digestion

with column A

Ion ratio 0.331

Results by quantifying 227203 5.49 ppb

Accuracy

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Bias

Three general approaches:

i) Analysis of certified reference materials /

reference materials

ii) Recovery experiments using spiked samples

iii) Comparison with results obtained with another

method

To take account any variation between runs, bias must be determined

over several days and preferably throughout the measuring range

and, if applicable, through the use of a suitable combination of

different specimens.

Several concentration levels must be incorporated in the

determination of bias. Otherwise the facility must be able to prove

that the method of analysis employed has the same trueness

throughout the measuring range.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

NATA Technical Note 17 - October 2013

Analysis of certified reference materials

NIST Special Publication 260-181, June 2014

A strong evidence to demonstrate the trueness and the realization of

traceability.

Analysis of certified reference materials

NIST Special Publication 260-181, June 2014

Analysis of certified reference materials

NIST Special Publication 260-181, June 2014

Analysis of certified reference materials

NIST Special Publication 260-181, June 2014

RM should only be used for one purpose

during a validation study.

For example, an RM used for calibration

shall not also be used to evaluate bias.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Recovery experiments using spiked samples

Matrix selection

At least five different common food matrixes (protein, carbohydrate,

oil, dietary fibre and water), and at least three food types

representative of each food matrix.HOKLAS SC-37- 8/07/ 2014

NIST Special Publication 260-181, June 2014

SANCO/12571/2013 - 01/01/ 2014

A practical default range of 60-140% may be used for individual

recoveries in routine multi-residue analysis.

SANCO/12571/2013 - 01/01/ 2014

HOKLAS SC-37- 8/07/ 2014

Recovery

Limitation by using spiked samples

The inherent problem with spike samples is that analyte introduced in such

a way will probably not be bound as strongly as that which is naturally

present in the test portion matrix and so the technique will give an

unrealistically high impression of the extraction efficiency.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Incurred tea leaves

DWM Sin, YL Wong. Analytical and Bioanalytical Chemistry To be published.

Wet overnight

Wet overnight

Extraction

efficiency of

incurred sample

w/o wetting

Comparison with results with another method

Reference method, such as standard method or

primary method, with smaller uncertainty than

candidate method

Method currently in routine use in the laboratory.

The significance of the bias may be estimated by

statistical analysis (t-test) of the results obtained.

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

NATA Technical Note 17 - October 2013

Precision

RepeatabilityVariability by same analyst, same equipment,

short timescale

ReproducibilityVariability by different laboratories,

Different analyst, different equipment, long time scale

Intermediate PrecisionSingle laboratory, different analysts,

different equipment, extended timescale

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Precision is generally dependent on analyte concentration,

and so should be determined at a number of concentrations

across the range of interest.

6 – 15 replicates for each material

Standard Deviation

sr , si

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Simultaneous Determination of Repeatability

and Intermediate Precision

6 – 15 groups of duplicate measurement

on different analyte/days/equipment for

each material

One-way ANOVA

Repeatability

=Within group precision

Intermediate Precision

= Square root of the sum of

squares of the within-group and

between group

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Repeatability standard deviation obtained

with at least 6 degrees of freedom

e.g. 7 times in a series with one test item,

4 times in a series with two test items,

3 times in a series with three test items.

NATA Technical Note 17 - October 2013

Precision Limits

enables the analyst to decide whether there is a significant

difference, at a specified level of confidence, between results

from duplicate analyses of a sample obtained under specified

conditions.

Repeatability Limit (r) = 2.8 x sr

Reproducibility Limit (R) = 2.8 x sR

Eurachem -The Fitness for Purpose of Analytical Methods, 2014

Y1-Y2 < 16.7 mg/L for lemonade@~430 mg/L

Statistics Tools for Verification

BS EN 12857:1999

Statistics Tools for Verification

F-test by comparing the precision of the two methods

Should use repeatability standard deviation

F = (Repeatability SD of lab)2 / (Repeatability SD of Std Method)2

If F<F(95%): statistically no significant difference

BS EN 12857:1999

Ruggedness

A measure of its capacity to remain unaffected by small,

but deliberate variations in method parameters.

Ruggedness provides an indication of the method’s

reliability during normal usage.

To know which procedures / factors affect the

analytical results significantly.

In some cases, information may be available during in-house

method development. Intermediate precision, by their

nature, take into account some aspects of a method’s

ruggedness.

NATA Technical Note 17 - October 2013

Validation is always a balance

between costs, risks and technical

possibilities.

ISO/IEC 17025