tct-541 coronary computered tomography for systematic screening of coronary stent fractures in...

1
TCT-539 The longitudinal elongation of the contemporary drug eluting stent in a tapered bench-top model Masaki Takenaka 1 , Haruo Hirayama 2 , Kenji Hurusawa 2 , Mamoru Nanasato 2 , Takamitu Nin 2 , Yukihiko Yoshida 2 , Ruka Yoshida 2 , Hiroki Kamiya 2 1 Nagoya daini red cross hospital, Nagoya, Aichi prifecture, 2 Nagoya daini red cross hospital, Nagoya, Aichi Background: The characteristics of new generation drug eluting stents (DES) are the combination of thinner struts and fewer connectors. These may affect longitudinal stent deformation. In tapered vessels, a diameter mismatch between vessel and stent at the proximal edge could lead to longitudinal stent elongation after fully stent expansion. The aim of this study was to analyze the longitudinal integrity among different DESs. Methods: Six different DESs (Cypher select, Taxus Liberte, Xience V, Nobori, Taxus Element, and Endeavor) were examined using a tapered bench-top silicon model. The diameter of each stents is 3.5mm and the length is 23 or 24mm. The proximal stent edge was adjusted to 5mm away from the marker point of 4.0mm in diameter. First, every stents were deployed at nominal pressure. Second, they were dilated with the 4.0mm diameter non-compliant balloon at 10atm and 20atm at the same condition. Intravascular ultrasound was used to measure the luminal diameter and microscope was used to measure the longitudinal stent length of their overall, the length from the marker point to the proximal stent edge, and the uniformity of stent struts was evaluated. We used each three stents and their data were averaged. Results: Results show the table. The uniformity of stent struts were not kept in Taxus Element stent. stent length (mm) DES pre deploy stent deploy post balloon dilatation at 10atm post balloon dilatation at 20atm Cypher select 23.550.05 24.030.06 (0.48) 24.050.27 (0.50) 24.390.11 (0.84) Taxus Liberte 25.270.18 24.800.21 (0.47) 25.150.11 (0.12) 25.340.35 (0.07) Xience V 24.020.04 24.530.48 (0.51) 24.790.33 (0.77) 25.830.59 (1.81) Nobori 24.790.02 24.890.17 (0.10) 25.070.12 (0.28) 25.460.16 (0.67) Taxus Element 23.080.19 23.160.03 (0.08) 23.910.39 (0.83) 24.530.38 (1.45) Endeavor 25.090.02 25.570.04 (0.48) 25.550.01 (0.46) 26.320.35 (1.23) Stent prox length from the marker point (mm) stent deploy post balloon dilatation at 10atm post balloon dilatation at 20atm 5.690.21 (0.69) 5.550.37 (0.55) 6.160.21 (1.16) 5.180.29 (0.18) 5.500.15 (0.50) 5.730.31 (0.73) 5.670.37 (0.67) 6.080.26 (1.08) 7.000.52 (2.00) 5.160.21 (0.16) 5.390.52 (0.39) 6.140.25 (1.14) 5.120.18 (0.12) 5.810.34 (0.81) 6.420.08 (1.42) 5.170.09 (0.17) 5.450.13 (0.45) 6.160.58 (1.16) Conclusions: Old generation DESs have more longitudinal strength than new generation DESs. On the other hand, new generation DESs showed longitudinal stent elongation. We should be careful when we use a new generation DESs, such as Xience V and Taxus Element, for the treatment of the ostial lesion. TCT-540 Longitudinal Drug Eluting Stent Elongation Phenomenon due to high pressure additional dilatation; Insights from bench testing Takenori Domei 1 , Kisaki Amemiya 2 , Tomoaki Ito 3 , Masashi Iwabuchi 4 1 Kokura memorial hospital, Kitakyusyu, Japan, 2 Kokura Memorial Hospital, Kitakyushu, Fukuoka, 3 Kokura memorial hospital, Kitakyusyu, Fukuoka, 4 Kokura Memorial Hospital, Kitakyusyu, Japan Background: Recently, stent longitudinal deformation is becoming to be one of the clinical issue. The objective of this study was to evaluate how post stent dilatation with large balloon has an impact on stent elongation. Methods: A Silicon tube modelwas used. Two types of drug eluting stents (Taxus Element 3.0mm, Boston scientific, Massachusetts and Nobori 3.0mm, Terumo, Tokyo) were deployed into the model as follows; 10mm of one side of the stent was attached to the model wall and the 10mm on the opposite side of the stent was free from the vessel wall (shown in Fig.1). Then, in both types of stents, post stent dilatation was performed with three different types of 3.5mm balloon at rated balloon pressure as shown in Fig 2. A total of 6 stents were tested, and stent length before and after post stent dilatation, difference of degree of stent elongation between stent types and balloon types were measured. Results: Stent elongation was observed in all stents, and the length of the stent free from the model wall before and after post stent dilatation was 10.090.26mm and 11.510.66mm, respectively. There were no significant difference between the lengths of the two types of stents (1.610.58 and 1.270.12, p0.36), but significant difference between the three types of post stent balloons were found (1.110.02 and 1.120.00 and 1.190.04, p0.0063). Conclusions: Stent elongation after post stent dilatation was observed in this ex vivo model. Stent elongation after post stent dilatation may be influenced not only by stent type but also the type of balloon used in post stent dilatation. TCT-541 Coronary Computered Tomography for Systematic Screening of Coronary Stent Fractures in Patients at High Risk Lisa Krenn 1 , Vanessa Berger-Kulemann 1 , Philipp Krisai 1 , Florian Wolf 1 , Georg Delle-Karth 1 , Gerhard Kreiner 1 , Clemens Gangl 1 , Markus Vertesich 1 , Sabine Scherzer 1 , Christian Roth 1 , Irene Lang 1 , Heinz Sochor 1 , Thomas Neunteufl 1 , Christian Loewe 1 , Berger Rudolf 1 1 Medical University of Vienna, Vienna, Austria Background: Stent fractures (SFs) contribute to in-stent-restenosis and life-threatening complications. Coronary computered tomography (CCT) may be an appropriate method for the detection of a SF. Aim: This study prospectively evaluated the incidence of SFs in high-risk patients using CCT and assessed the clinical relevance of this finding using catheter coronary angiography (CCA). Methods: Patients with two or more risk factors for a SF defined as (1) stent length 28 mm, (2) overlapping stents, (3) stent localization in the right coronary artery, or saphenous vein graft and (4) vessel angulation 75° before implantation or stent angulation 45° after implantation were invited to undergo a CCT 6 months after the procedure. To differentiate between SF and overlap failure all stents were identified on the CCT image by measuring the distance between edges and comparing these measurements with the known stent lengths. A CCA was recommended in patients with a stent gap. Patients without stent gap but with pathological findings in the CCT who underwent CCA served as controls. Results: In 14 out of 41 patients (34%) coronary CCT revealed a stent gap including 8 patients with a SF (20%) and 6 patients with an overlap failure (15%). In the following CCA all stent gaps were confirmed by optical coherence tomography. A clinically relevant stent-related pathology could be detected in 4 out of 8 patients (50%) with SF (in-stent-restenosis in two patients, chronic total occlusion and coronary aneurysm in one patient, respectively), but in none out of 7 controls (chi-square – p0.001). Conclusions: Stent gaps are frequent in high-risk patients. The majority of these gaps result from a SF, which is often associated with a clinically relevant pathology. Therefore, screening for SFs using CCT in high-risk patients might be beneficial. TCT-542 Axial Integrity of Coronary Stents: Evaluation Using Intravascular Ultrasound Danny Dvir 1 , Hironori Kitabata 2 , Israel Barbash 3 , Joshua Loh 2 , Rebecca Torguson 1 , Salem Badr 3 , Lowell Satler 4 , Augusto Pichard 5 , Ron Waksman 6 1 Washington Hospital center, washington, DC, 2 Medstar Washington Hospital Center, Washington, DC, 3 Washington Hospital Center, Washington, DC, 4 Washington hospital center, washington, DC, 5 washsington hospital center, Washington, USA, 6 Georgtown University, Washington, DC Background: Longitudinal stent deformation (i.e. the axial shortening or lengthening of a stent after implantation) is increasingly reported. However, the axial integrity of stents has not been systematically studied using intravascular ultrasound (IVUS). TUESDAY, OCTOBER 23, 8:00 AM–10:00 AM www.jacc.tctabstracts2012.com B156 JACC Vol 60/17/Suppl B | October 22–26, 2012 | TCT Abstracts/POSTER/Stent Fracture and Longitudinal Deformation POSTERS

Upload: lisa-krenn

Post on 28-Nov-2016

213 views

Category:

Documents


1 download

TRANSCRIPT

TCT-539

The longitudinal elongation of the contemporary drug eluting stent in atapered bench-top model

Masaki Takenaka1, Haruo Hirayama2, Kenji Hurusawa2, Mamoru Nanasato2,Takamitu Nin2, Yukihiko Yoshida2, Ruka Yoshida2, Hiroki Kamiya2

1Nagoya daini red cross hospital, Nagoya, Aichi prifecture, 2Nagoya daini redcross hospital, Nagoya, Aichi

Background: The characteristics of new generation drug eluting stents (DES) arethe combination of thinner struts and fewer connectors. These may affectlongitudinal stent deformation. In tapered vessels, a diameter mismatch betweenvessel and stent at the proximal edge could lead to longitudinal stent elongationafter fully stent expansion. The aim of this study was to analyze the longitudinalintegrity among different DESs.Methods: Six different DESs (Cypher select, Taxus Liberte, Xience V, Nobori,Taxus Element, and Endeavor) were examined using a tapered bench-top siliconmodel. The diameter of each stents is 3.5mm and the length is 23 or 24mm. Theproximal stent edge was adjusted to 5mm away from the marker point of 4.0mmin diameter. First, every stents were deployed at nominal pressure. Second, theywere dilated with the 4.0mm diameter non-compliant balloon at 10atm and 20atmat the same condition. Intravascular ultrasound was used to measure the luminaldiameter and microscope was used to measure the longitudinal stent length oftheir overall, the length from the marker point to the proximal stent edge, and theuniformity of stent struts was evaluated. We used each three stents and their datawere averaged.Results: Results show the table. The uniformity of stent struts were not kept in TaxusElement stent.

stent length (mm)

DES pre deploy stent deploypost balloon dilatation

at 10atmpost balloon

dilatation at 20atm

Cypher

select

23.55�0.05 24.03�0.06 (�0.48) 24.05�0.27 (�0.50) 24.39�0.11 (�0.84)

Taxus

Liberte

25.27�0.18 24.80�0.21 (��0.47) 25.15�0.11 (��0.12) 25.34�0.35 (�0.07)

Xience V 24.02�0.04 24.53�0.48 (�0.51) 24.79�0.33 (�0.77) 25.83�0.59 (�1.81)Nobori 24.79�0.02 24.89�0.17 (�0.10) 25.07�0.12 (�0.28) 25.46�0.16 (�0.67)Taxus

Element

23.08�0.19 23.16�0.03 (�0.08) 23.91�0.39 (�0.83) 24.53�0.38 (�1.45)

Endeavor 25.09�0.02 25.57�0.04 (�0.48) 25.55�0.01 (�0.46) 26.32�0.35 (�1.23)

Stent prox length from the marker point (mm)

stent deploypost balloon

dilatation at 10atmpost balloon

dilatation at 20atm

5.69�0.21 (�0.69) 5.55�0.37 (�0.55) 6.16�0.21 (�1.16)5.18�0.29 (�0.18) 5.50�0.15 (�0.50) 5.73�0.31 (�0.73)5.67�0.37 (�0.67) 6.08�0.26 (�1.08) 7.00�0.52 (�2.00)5.16�0.21 (�0.16) 5.39�0.52 (�0.39) 6.14�0.25 (�1.14)5.12�0.18 (�0.12) 5.81�0.34 (�0.81) 6.42�0.08 (�1.42)5.17�0.09 (�0.17) 5.45�0.13 (�0.45) 6.16�0.58 (�1.16)

Conclusions: Old generation DESs have more longitudinal strength than new generationDESs. On the other hand, new generation DESs showed longitudinal stent elongation. Weshould be careful when we use a new generation DESs, such as Xience V and TaxusElement, for the treatment of the ostial lesion.

TCT-540

Longitudinal Drug Eluting Stent Elongation Phenomenon due to highpressure additional dilatation; Insights from bench testing

Takenori Domei1, Kisaki Amemiya2, Tomoaki Ito3, Masashi Iwabuchi41Kokura memorial hospital, Kitakyusyu, Japan, 2Kokura Memorial Hospital,Kitakyushu, Fukuoka, 3Kokura memorial hospital, Kitakyusyu, Fukuoka, 4KokuraMemorial Hospital, Kitakyusyu, Japan

Background: Recently, stent longitudinal deformation is becoming to be one of theclinical issue. The objective of this study was to evaluate how post stent dilatation withlarge balloon has an impact on stent elongation.Methods: A Silicon tube modelwas used. Two types of drug eluting stents (TaxusElement 3.0mm, Boston scientific, Massachusetts and Nobori 3.0mm, Terumo, Tokyo)were deployed into the model as follows; 10mm of one side of the stent was attached tothe model wall and the 10mm on the opposite side of the stent was free from the vessel wall(shown in Fig.1). Then, in both types of stents, post stent dilatation was performed with threedifferent types of 3.5mm balloon at rated balloon pressure as shown in Fig 2. A total of 6 stentswere tested, and stent length before and after post stent dilatation, difference of degree of stentelongation between stent types and balloon types were measured.Results: Stent elongation was observed in all stents, and the length of the stent free fromthe model wall before and after post stent dilatation was 10.09�0.26mm and11.51�0.66mm, respectively. There were no significant difference between the � lengthsof the two types of stents (1.61�0.58 and 1.27�0.12, p�0.36), but significant difference

between the three types of post stent balloons were found (1.11�0.02 and 1.12�0.00 and1.19�0.04, p�0.0063).Conclusions: Stent elongation after post stent dilatation was observed in this ex vivomodel. Stent elongation after post stent dilatation may be influenced not only by stent typebut also the type of balloon used in post stent dilatation.

TCT-541

Coronary Computered Tomography for Systematic Screening of CoronaryStent Fractures in Patients at High Risk

Lisa Krenn1, Vanessa Berger-Kulemann1, Philipp Krisai1, Florian Wolf1,Georg Delle-Karth1, Gerhard Kreiner1, Clemens Gangl1, Markus Vertesich1,Sabine Scherzer1, Christian Roth1, Irene Lang1, Heinz Sochor1,Thomas Neunteufl1, Christian Loewe1, Berger Rudolf11Medical University of Vienna, Vienna, Austria

Background: Stent fractures (SFs) contribute to in-stent-restenosis and life-threateningcomplications. Coronary computered tomography (CCT) may be an appropriate methodfor the detection of a SF. Aim: This study prospectively evaluated the incidence of SFs inhigh-risk patients using CCT and assessed the clinical relevance of this finding usingcatheter coronary angiography (CCA).Methods: Patients with two or more risk factors for a SF defined as (1) stent length �28mm, (2) overlapping stents, (3) stent localization in the right coronary artery, or saphenousvein graft and (4) vessel angulation �75° before implantation or stent angulation �45° afterimplantation were invited to undergo a CCT 6 months after the procedure. To differentiatebetween SF and overlap failure all stents were identified on the CCT image by measuring thedistance between edges and comparing these measurements with the known stent lengths. ACCA was recommended in patients with a stent gap. Patients without stent gap but withpathological findings in the CCT who underwent CCA served as controls.Results: In 14 out of 41 patients (34%) coronary CCT revealed a stent gap including 8patients with a SF (20%) and 6 patients with an overlap failure (15%). In the followingCCA all stent gaps were confirmed by optical coherence tomography. A clinicallyrelevant stent-related pathology could be detected in 4 out of 8 patients (50%) with SF(in-stent-restenosis in two patients, chronic total occlusion and coronary aneurysm in onepatient, respectively), but in none out of 7 controls (chi-square – p�0.001).Conclusions: Stent gaps are frequent in high-risk patients. The majority of these gapsresult from a SF, which is often associated with a clinically relevant pathology. Therefore,screening for SFs using CCT in high-risk patients might be beneficial.

TCT-542

Axial Integrity of Coronary Stents: Evaluation Using IntravascularUltrasound

Danny Dvir1, Hironori Kitabata2, Israel Barbash3, Joshua Loh2,Rebecca Torguson1, Salem Badr3, Lowell Satler4, Augusto Pichard5,Ron Waksman6

1Washington Hospital center, washington, DC, 2Medstar Washington HospitalCenter, Washington, DC, 3Washington Hospital Center, Washington, DC,4Washington hospital center, washington, DC, 5washsington hospital center,Washington, USA, 6Georgtown University, Washington, DC

Background: Longitudinal stent deformation (i.e. the axial shortening or lengthening ofa stent after implantation) is increasingly reported. However, the axial integrity of stentshas not been systematically studied using intravascular ultrasound (IVUS).

TUESDAY, OCTOBER 23, 8:00 AM–10:00 AM www.jacc.tctabstracts2012.com

B156 JACC Vol 60/17/Suppl B | October 22–26, 2012 | TCT Abstracts/POSTER/Stent Fracture and Longitudinal Deformation

PO

ST

ER

S