TIMEPIX AND TOT MODETaken from Llopart’s 2007 thesis: Paper X and

J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183

Ryan Badman, July 15th

Page 2: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

2

OUTLINE

Introduction to Timepix Calibration with charge (test pulse) Calibration with energy

Page 3: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

3

IMAGE OF TIMEPIX

Fig. 1

Page 4: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

4

TIMEPIX LAYOUT Analog part has one band-gap

circuit to internally generate a stable reference voltage to be used by the 13 on-chip global DACs Reference voltage has sensitivity of -

0.22 mV/ deg C and a power supply of less than 1mV/V

Has eight 8-bit current DACs and four 8-bit voltage DACs, and a single 14-bit voltage DAC

Digital part has all input/output control logic, the IO wire-bonding pads, and a 24-bit fused blown registry for unique chip identification

36 million transistors total in the entire chip, and a read-out time of less than 300us

Page 5: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

5

CHARGE COLLECTION IN A SINGLE CHANNEL Electronics for each individual pixel Any charge collected by the pixel anode is integrated and compared to the

global threshold. If the preamplifier voltage crosses the threshold then the output of the

discriminator generates a pulse whose width corresponds to the length of the time the preamplifier output voltage remains over threshold

Minimum detectable charge is expected to be 650 e- and the noise is 100 e-rms

Page 6: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

6

TIMEPIX ABILITIES Timepix uses an external clock

with frequency up to 100MHz Each pixel has its own

preamplifier, discminator with hysteresis, and a 14-bit counter

Each pixel has 4 modes: Masked mode: pixel is off Medipix mode: incoming

particles are counted TOT mode: Counter is

incremented continuously as long as signal is over TH, used to measure particle energy. (TOT = time over threshold)

Timepix mode: counter is continuously incremented from the time the fist hit arrives until the end of the shutter

Fig. 11 shows a full matrix response to 10 test pulses of about 2.3 ke-

Page 7: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

7

TOT CALIBRATION

TOT vs. Qin (test pulse) In TOT mode the energy resolution is better than

5% if the input charge is greater than 1 keV above threshold

Fig. 6, TOT measurements on the right, and the measured energy resolution with 1% and 5% markers on the right. Fig 1 (above, another Timepix schematic.

Page 8: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

8

TIME-WALK

Defined as the difference between the time measured from an input charge that is 1ke- over threshold and an infinite input charge.

The faster the preamp peaking time, the better the time-walk value

One method of compensating the time walk is to arrange the pixel matrix in a chess pattern where pixels alternate between TOT mode and arrival mode, and then use knowledge of the input charge in the neighboring pixels.

Measured time-walk per pixel is less than 50ns.

Time-walk of 1 pixel at the matrix center for five different thresholds. The thicker lines represent Preamp 255 (1.8uA) and the thin with Preamp 127 (900nA). Y axis is time of arrival.

Page 9: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

9

TIMEPIX EXAMPLE RESULTS

Fig. 12 shows TOT measurements of a single cosmic background particle on the left after interacting in the gas of a GEM detector, and on the right the arrival time measurements of the cosmic particle obtained with a Micromegas gas gain grid coupled to the Timepix chip.

Page 10: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

10

TOT MODE ENERGY CALIBRATION Each pixel needs

to be calibrated for proper energy measurement.

This calibration is nonlinear in lower energy ranges, and linear in higher.

The calibration can be described by a surrogate function containing four parameters.

Fig. 2 (a is gain, b the y-intercept, and t relates to threshold)

Page 11: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

11

HOW THE FIT PARAMETERS VARY PER PIXEL (MY RESULTS, TH400)

Results for (col 0-40) x 256 pixel Note: These our for test pulse, not for Jakubek’s energy calibration

Page 12: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

12

COMPARE TO RICHARD’S (SAME BINNING)

TH300, Col (0-40) x 256

Page 13: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

13

PROBLEM WITH ENERGY CALIBRATION Calibration requires measurement of at least 4 spectral

lines and at minimum 5 least-squares fits per pixel. However, fitting spectral peaks with Gaussians in the non-linear portion close to threshold gives incorrectly shifted results.

Page 14: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

14

IMPROVEMENTS TO ENERGY CALIBRATION

The solution to the bad peak shape is to combine a Gaussian with the inverse of the surrogate function for the nonlinear region (called model M); and just a Gaussian in the linear region.

Fig. 3. The TOT spectrum of 55Fe (the same spectrum as shown in Fig. 2 on the previous slide) fitted with the model M.

Page 15: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

15

IRRADIATION WITH MULTI-ENERGY CALIBRATION SOURCE. Jakubek irradiated the Timepix chip to obtain TOT spectra for each

pixel in single pixel clusters. The threshold was just above the noise edge, sensor bias voltage

was 100V, the clock frequency was 10 MHz, and Ikrum was 1.

Page 16: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

16

CALIBRATED GLOBAL SPECTRUM

The calibration quality was tested without the Fe layer and the following global spectrum was observed.

Fig. 6. Calibrated spectra of dual energy source (241Am+In) for different cluster sizes (the cluster size is denoted by label). Peaks for different cluster sizes are well aligned. Peaks are fitted with Gaussians. The energy resolution (sigma of Gaussian) is 2.3keV for both peaks.

Page 17: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

17

DISTORTION CORRECTION Since the cluster

volume spectra is distorted (its supposed to be independent of bias voltage), the calibration function is not accurate for high ionization charges.

The difference between estimated and measured cluster heights is a linear function so the calibration function can be corrected above 0.9MeV (up to 1.2MeV).

Fig. 12. Cluster volume spectra of 5.5MeV alpha particles. The distorted response of pixels was corrected above 0.9MeV and the spectrum shape restored (compare with Fig. 10).

Page 18: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

18

CLUSTER HEIGHT VS BIAS VOLTAGE

Fig. 8. Maximum signal seen by a pixel in the cluster (cluster height) as a function of the bias voltage for 5.5MeV alphas. Fig. 9. Dependence of the

cluster height on the bias voltage for 5.5MeV alpha particles. When the pixel response reaches about 0.9MeV the cluster height starts to grow unexpectedly.

Page 19: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

19

EXAMPLE OF DISTORTION

Fig. 10. Cluster volume spectra for sensor bias voltage of 39V (top) and 72V (bottom) showing distortion caused by the deviation of pixel response from the calibration function f

Page 20: T IMEPIX AND TOT MODE Taken from Lloparts 2007 thesis: Paper X and J. Jakubek, Nucl. Instr. and Meth. A (2010), doi:10.1016/j.nima.2010.06.183 Ryan Badman,

20

SUMMARY

Timepix offers more control over individual pixels than its predecessor Medipix2.

The small periphery will cause no loss in coverage if the chip is chosen for the VELO upgrade.

The power consumption, radiation tolerance, pixelated format, small pitch, and convenient programmability are also favorable for the VELO upgrade.

Jakubek’s correction successfully corrects the cluster height measurements for higher pixel responses up to 1.2 MeV.

We will come up with a better energy calibration?