Page 1: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Practice In Service: Sterilization Guide

Page 2: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Implant Direct: Implants & Sterilization

● Sterile Packaging: Implants and included materials are sold sterile by gamma irradiation.

● Do not re-sterilize.

● Shelf-Life YYYY-MM-DD: Dental implants are considered sterile for five years from the date of initial sterilization. The product expiration date is indicated by the hourglass symbol on the product label, followed by the year and month of expiration.

1

Page 3: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Implants and Labeling

● Look for the “use by date” symbol on the implant label.

2

Page 4: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Implants & Handling

• Do not use if packaging is damaged.

• Implants must be stored in a dry place, at room temperature, in their original packaging.

• Dental implants are provided in sterile vials.

• The fixture-mount or carrier are intended to be used to transport the implant to the prepared surgical site.

• Appropriate sterilized insertion tools, with retention, are used to transport and place the mount-free implants.

• Do not handle implant surfaces directly.

• Users are advised to visually inspect vials to insure seals and contents are intact and in their original packaging prior to use.

3

Page 5: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Abutments & Prosthetics Sterilization

5

Page 6: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Implant Direct: Abutments & Sterilization

● All abutments and accessories are supplied non-sterile unless they are explicitly marked as sterile.

● Sterilization is required for all prosthetic components prior to being used intraorally.

6

Page 7: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Abutments: Packaging for Sterilization

Enclose the product in FDA cleared steam sterilizable pouches and sterilize in a FDA cleared sterilizer.

7

Page 8: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Sterilization of Non-Sterile Prosthetic Components

TITANIUM, GOLD, ANDPLASTICS ABUTMENTS

ZIRCONIA ABUTMENTS ZEST PROSTHETICS

Steam sterilize using a gravity displacement cycle for 30 minutes at

121°C (250°F), 15-20 psig.

Steam sterilize Zirconia abutments using a pre-vacuum cycle for 4

minutes at a temperature of 132°C (270°F).

Please follow sterilization instructions provided with the product.

Dry time: 15 minutes. Dry time: 20 minutes.

8

Wrapped in a double pouch. Multiple piece products should be disassembled to individual components prior to sterilization.

When sterilizing multiple devices in one steam sterilization cycle, ensure that the sterilizer manufacturer’s maximum load requirements are followed.

Page 9: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Surgical Instruments: Cleaning & Sterilization

9

Page 10: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Surgical Instruments, Drills & Trays

• Surgical instruments, drills and trays for implant surgery must be sterilized prior to use.

• Sterilization must be done according to the instructions for use.

• These surgical devices are reusable and provided non-sterile. Must be cleaned and sterilized according to the instructions for use.

10

Page 11: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Surgical Drills

Surgical drills (Stainless Steel) are used in a sequence to create an osteotomy.

11

Page 12: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Surgical Trays

● The surgical trays are designed to hold various surgical devices in order to organize, steam sterilize, and transport them between uses.

● The trays are to be enclosed in an FDA cleared steam sterilizable wrap and sterilized in an FDA cleared sterilizer using the sterilization method specified below.

● The trays are intended for sterilization of non-porous loads, and the trays are recommended not to be stacked during sterilization.

● Implant Direct Sybron Manufacturing LLC does not make any lumen claims.

12

Page 13: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

FDA Guidance Document for Reprocessing Medical Devices in Health Care Settings

CAUTION: Universal precautions for the handling of contaminated or biohazardous materials should be observed. The surgical devices are not designed for cleaning and disinfection in the fully assembled state. The surgical devices must be removed from the tray and the tray disassembled for adequate cleaning results. After cleaning and disinfection, per instructions to follow, the tray is reassembled and the instruments and drills can be loaded into the tray, per the packaging for sterilization instructions, and sterilized.

NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12, AAMI TIR 30, and FDA draft guidance document for processing/reprocessing medical devices in health care settings.

Surgical devices are packaged non-sterile and should be cleaned and sterilized (Note 3) prior to each use. Disassemble surgical devices prior to cleaning. Wipe devices with soft cloth prior to sterilization.

13

Page 14: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Processing / Reprocessing Instructions (during surgery)

● Point of Use: ○ Tray of distilled water: placed used instruments, drills,

etc. in a basin of distilled water immediately after use to prevent body fluids, hard and soft tissues from drying on the surgical device and minimize the cleaning and reprocessing time.

● Directly after use, wipe excess soil from the surgical devices.

● Automated cleaning may not be effective. Thorough manual cleaning process is recommended.

14

Page 15: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Preparation for Manual Cleaning (post-surgery)

● If the tray contains surgical devices, remove them from the tray; clean and disinfect separately before sterilizing. Reassemble them in the cleaned tray.

● Disassemble the tray into its individual components: base, lid, and insert (if applicable).

● See next slides for disassembly process.

15

Page 16: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Surgical Tray: Disassembly

DISASSEMBLE:

1. Open the lid of the tray and fold back.

2. Remove surgical devices from the insert and base, if any.

3. Rotate the insert up and unsnap the insert from the base.

4. Remove the insert from the assembly.

5. Unsnap the lid from the base to disassemble.

16

Page 17: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Cleaning & Disinfection

Agents: Acceptable cleaning agents are specified in the cleaning instructions ahead.

Do not use cleaning agents containing hydrogen peroxide, chlorine or chloride. The active agents are corrosive to the surgical devices. Using a detergent outside of the recommended pH may adversely affect the finish of the surgical devices. Ensure surgical drills and instruments, made with different materials, are not placed together in liquid during cleaning.

17

Page 18: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Preliminary Manual Cleaning

● For heavily soiled surgical devices, or those that have accumulated organic debris, a preliminary manual cleaning is advised.

● Use a soft brush (no metal bristles) to completely remove residue.

18

Page 19: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Cleaning Process for Surgical Instruments & Drills (See Note 4)

Step

1 Wipe Surgical instruments and Drills using a soft, disposable non-shedding wipe.

2 Soak in lukewarm enzymatic solution, per manufacturer's instructions, for 20 minutes

3 Wash using broad spectrum cleaning solution

4 Rinse with water and dry

5 Prepare a neutral pH enzyme detergent solution per manufacturer's instructions.Sonicate for 10 Minutes at a frequency of 45 to 50 kHz

6 Rinse for 3 minutes with Reverse Osmosis (RO) or Deionized Water (DI)

19

Page 20: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Cleaning Process for Surgical Instruments & Drills (See Note 4)

Step

7 Repeat step 5 & 6 with fresh RO/DI water.

8 Dry using a clean, disposable, absorbent, non-shedding wipe

9Inspection: If soil, tissue, or fluid visible on device, repeat steps 1-8.

If no visible soil, tissue or fluid, proceed to Steam sterilization (next slide)

20

Page 21: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Steam Sterilization for Surgical Devices

Packaging for Sterilization: Enclose the surgical device in FDA cleared steam sterilizable packaging and sterilize in an FDA cleared sterilizer.

To avoid deformation, the surgical trays should not touch the walls of the steam sterilizer. When sterilizing multiple surgical devices in one steam sterilization cycle, ensure that the sterilizer manufacturer’s maximum load requirements are followed.

21

Page 22: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Surgical Device Sterilization Parameters

Sterilization PackagingEnclose instruments and drills in FDA cleared steam sterilizable wraps or pouches

Steam Sterilization Parameters

GravityTime: 15 minutesTemperature: 132°C (270°F)Dry Cycle: 30 minutesPre-vacuumTime: 4 minutesTemperature: 132°C (270°F)Dry Cycle: 20 Minutes

22

Page 23: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Surgical Tray: Cleaning & Sterilization

23

Page 24: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Surgical Tray Sterilization

● After cleaning, reassemble the surgical tray for sterilization.

● See the following slide for assembly instructions

24

Page 25: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Surgical Tray Assembly

ASSEMBLE:

1. Snap on the lid to the base and keep both the parts perpendicular to each other.

2. Place the insert in between the base and the lid and snap in.

3. Assemble surgical devices in the insert and base, if needed.

4. Close the lid by lowering the lid. It automatically locks onto the base when lowered.

25

Page 26: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

After individually cleaning and disinfecting all tools and the tray, the instruments can be loaded into the assembled tray, packaged, and sterilized.

26

Surgical Tray Sterilization Parameters

Page 27: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Surgical Tray Sterilization Parameters

Sterilization Packaging Double Wrap assembled trays in FDA cleared steam sterilizable wraps

Steam Sterilization Parameters

GravityTime: 15 minutesTemperature: 132°C (270°F)Dry Cycle: 30 minutesPre-vacuumTime: 4 minutesTemperature: 132°C (270°F)Dry Cycle: 20 Minutes

27

Page 28: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Torque Wrenches and Other Products

The information provided here is for readability and convenience purposes only and does not supercede the Instructions for use included with the product. For any questions, please refer to the instructions for use.

28

Page 29: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Torque Wrench Sterilization (Product number: TW30, TW35)

● Before initial use and after each procedure, clean the DynaTorq Torque Wrench (TW30, TW35) with a wet towel or disinfectant wipe.

● Rinse under warm tap water and wipe dry.

● Before sterilization, place a drop of lubricating oil into the opening where the wrench head releases from the shaft.

● The DynaTorq Torque Wrench may be sterilized by steam autoclave. Sterilize for 20 minutes at 135° C (275° F), with the handle in the “broken” position.

● After sterilization, re-set the torque break in the handle, then manually “break” the torque again to ensure that the wrench is operating correctly.

29

Page 30: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Torque Wrench Sterilization (Product number: TWM)

● Before initial use and after each procedure, clean the torque wrench with a wet towel or disinfectant wipe.

● Rinse under warm tap water and wipe dry.

● The torque wrench may be autoclaved. Follow the autoclave manufacturer’s instructions. 20 minutes at 135° C (275° F) is recommended.

● Do not store in sealed autoclave bag.

● After autoclaving, rotate the setting dial to ensure that the wrench is operating correctly.

30

Page 31: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

● Clean, sterilize, and maintain wrenches, LT/RA Latch Driver, RB-06, and Zest system attachment tools according to specific instructions provided with the products.

● The cleaning and sterilization instructions supplied with these tools are different than the instructions in IFU-001.

31

Page 32: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Limitations of Reprocessing

● Proper processing/reprocessing has minimal effect on the surgical devices. End of life is determined by damage due to wear.

● Clean and sterilize the surgical devices using the allowable parameters. Not following the cleaning and sterilization instructions may accelerate normal product aging. Ensure surgical drills and instruments made with different materials are not touching during sterilization.

32

Page 33: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Drill Usage Management

● Implant Direct provides drill usage charts to assist practices with tracking the usage of drills.

● Download the InterActive drill usage chart here

● Download the Legacy drill usage chart here

● Please reach out to your Implant Direct representative for more information on the correct drill usage chart for your Implant Direct implant system.

33

Page 34: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Instructions For Use (IFU)

Users are advised to consult packaged instructions for use and the technical support sections of our website (www.implantdirect.com) for assistance.

For support or questions, call Implant Direct Customer Service & Support at 888-649-6425 or visit http://www.implantdirect.com/support

34

Page 35: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Symbol Key

Manufacturer’s symbol

Use by date

CE Mark

Sterilized using gamma radiation

Do not re-use

Caution: Federal law restricts this device to sale by or on the order of a dentist

Consult instructions for use

Do not use if packaging is damaged

35

Page 36: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Symbol Key, cont’d

Caution

Catalog Number

Non-sterile

Authorized Representative in Europe

36

Page 37: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Notes

NOTE 4: Wrenches (P/N: TW, TWM, TW30, TW35, BioTorq wrenches), LT/RA Latch Driver (P/N:58-100000), RB-06, and Zest attachment system tools are provided with different cleaning and sterilization instructions. Please use the instructions for use provided with the wrenches, LT/RA Latch Driver, and Zest tools as they are different than the instructions in this document. The cleaning and sterilization instructions provided with these tools do not apply to Implant Direct surgical drills, Instruments, or trays.

NOTE 5: Cleaning: Prolystica 2X concentrate Enzymatic Detergent and Prolystica 2X neutral pH detergent were used to validate the manual cleaning process as per the manufacturer’s dilution instructions. Cleaning validations were performed using the micro BCA protein assay with device extraction. Process validation consisted of contamination, cleaning, and sterilization. The haemoglobin level was well below the acceptance criteria level of <2.2 ug/cm2. The protein level was well below the acceptance criteria level of 6.4 ug/cm2.

37

Page 38: Sterilization Guide Practice In Service - Implant Direct · NOTE 3: Cleaning and sterilization instructions validated by Implant Direct Sybron Manufacturing LLC as per AAMI TIR 12,

Notes, Continued

NOTE 6: Please check that the sterilization packaging is compatible with the sterilization parameters for the device.

38