SOP_021 Version 3.0 Page 1

Shared Resources: Flow Cytometry and Cell Sorting

Standard Operating Procedure For:

Independent End User Training and Maintaining Optimal Instrument Performance:

DHVI BD LSRII (L01/L02) and BD LSRFortessa (F01)

Page 2: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 2 of 27

Standard Operating Procedure

SOP Number: SOP Flow_021 Effective Date: 1/1/2017

Version Issue Number: 3.0

Owner: Derek W. Cain

Author(s): Patti McDermott Derek W. Cain

SOP Flow_021 Title: Independent End User Training and

Maintaining Optimal Instrument Performance

DHVI BD LSRII (L01/L02) and BD LSRFortessa (F01)

By signing the “Approved By” section below, the person attest that he/she has personally conducted a review of the document for completeness and accuracy and approves the contents of the SOP document.

Approved By: Derek W. Cain, PhD

Title: Managing Director of DHVI Flow Facility Signed:

Date: 1/17/2017

Page 3: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 3 of 27

SOP Revision Document History:

Title & Version

Replaces

Effective Date

Description of Change

1 7/1/2013 New SOP

2 1 7/1/2016 Revised to make SOP consistent with other Facility SOPs

3 2 1/1/2017 Revised SOP with Core research info, updated contact list and additional training information, removed old

attachments and renumbered

Page 4: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 4 of 27

1.0 Purpose / Scope

The purpose of this training packet is to provide materials to support and enhance the hands-on training experience for Independent End Users.

2.0 Expectations

Trainees must have been recommended by their manager/mentor to undergo training, have at least 10 instrument contact hours with an experienced flow cytometer user, have a need to use DHVI Flow Facility instruments at least 5 hours/month, and have a strong desire to learn how to safely and correctly operate a flow cytometer to acquire comprehensive polychromatic flow data. Trainees must complete two BD Biosciences online courses, “Introduction to Flow Cytometry” and “BD FACSDiva Software v8 Overview,” PRIOR to initiating hands-on training.

Trainees will be expected to complete all three components of the DHVI Flow Facility Training program: Classroom group orientation (~2hr), Hands-on session (~2 hr), and a Practical Exam (~2 hr). If the trainee misses one of the sessions or fails the practical exam, they will not be certified by the assigned training Operator and will have to repeat the training program at a later date.

The Pin Code to access Bay 5 will be given to the Trainee at the classroom session. If the Independent End User is the last person to leave the room, they need to make sure that the door(s) is closed and locked.

All approved Independent End Users of the DHVI Flow Facility instruments are expected to comply fully with the Policies and Procedures for DHVI Flow Facility Users detailed in SOP Flow_010 (on website) that is covered in the classroom session of the training.

3.0 Cytometer Overview

Your trainer will orient you to the critical components of the cytometer: Fluidics, Optics, Computers, and the BDFACS Diva software (see Attachments #2 and 3).

4.0 Operation and Maintenance

4.1 Instrument Startup

4.1.1 Turn on Fluidics Cart (L02 only).

4.1.2 Turn on Cytometer (needs 30 minute laser warm-up).

4.1.3 Be sure Sheath tank is full:

4.1.3.1 Filling L02 Sheath Tank: Unscrew large cap and carefully pour sheath into the tank up to the top line on the tank.

4.1.3.2 Filling L01/F01 Sheath Tank: Depressurize the sheath tank by slowly pulling up on the pressure relief valve.

4.1.3.3 Remove the lid of the sheath tank, and fill up to the top weld line inside the tank.

Page 5: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 5 of 27

4.1.3.4 DO NOT OVERFILL.

4.1.3.5 Replace the lid making sure that you lock it down from side to side, not top to bottom.

4.1.4 Making up Sheath Fluid (PBS):

4.1.4.1 Sheath containers are located under the sink between Bay 5 and Bay 6. Fill the sheath container with dH2O to the “Fill Line” on the back of the container.

4.1.4.2 Put a large “X” on the cap and the sides of the sheath container to indicate that it has been diluted. DO NOT “X” THE LABEL.

4.1.4.3 If you spill the sheath while pouring, please clean it up.

4.1.4.4 Additional sheath is located under the desk behind the L02.

4.1.5 Waste Tank: Empty Waste tank if full. There are two (2) 10 L waste tanks used with the L01/L02/F01 because the waste needs to sit in the bleach solution for at least 24 hours before disposal in the sink.

4.1.5.1 Emptying the Waste Tank for the L02:

4.1.5.1.1 Bring the spare waste tank to the sink. Remove the large cap. Turn on the cold water and pour the 10% bleach waste down the drain letting the water run for a couple minutes when finished to flush the pipes.

4.1.5.1.2 Put 1 L of bleach and 2 L of water into the waste tank, so that when the tank is full it will be a 10% final bleach solution.

4.1.5.1.3 The bleach is located under the sink between Bay 5 and Bay 6.

4.1.5.1.4 Additional bleach is located under the desk behind the L02.

4.1.5.1.5 Remove the RED filter waste cap from the full waste tank and put it on the empty waste tank going onto the fluidics cart. Unscrew the sensor and carefully pull it out of the tank and put it in the empty waste tank. Be careful not to break the filter on the tubing.

4.1.5.1.6 NOTE: The red filter waste cap ALWAYS goes on the waste tank on the fluidics cart!!

4.1.5.1.7 Put the large white cap and the small cap on the full waste tank and write the “empty after” date

Page 6: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 6 of 27

(the next day) on the tape. Put the waste tank in the bucket to the side of the fluidics cart.

4.1.5.2 Emptying the Waste Tank of the L01/F01:

4.1.5.2.1 Disconnect the orange waste line and the black waste sensor line from the white cap on the full waste tank.

4.1.5.2.2 Bring the spare waste tank to the sink. Remove the large cap. Turn on the cold water and pour the 10% bleach waste down the drain letting the water run for a couple minutes when finished to flush the pipes.

4.1.5.2.3 Remove the waste tank from the holder. Put the empty waste tank in the holder.

4.1.5.2.4 Put the white waste cap with the sensor connections on it on the empty waste tank.

4.1.5.2.5 Reconnect the waste and sensor lines.

4.1.5.2.6 Put the large white cap on the full waste tank and put the “empty after” date on the tape.

4.1.5.2.7 Put the full waste tank in the bucket to the side under the L01/F01 table.

4.1.6 Sheath Reservoir (L02 Only). Be sure the Sheath Reservoir has ~1/4” of sheath in it. See Attachment #5 for troubleshooting the sheath reservoir.

4.1.6.1 Filling Sheath Reservoir:

4.1.6.1.1 Put fluidics in Standby.

4.1.6.1.2 Unscrew the white cap and take sensor out of the tank.

4.1.6.1.3 Using the funnel (located at the sink), put ~1/4” of sheath fluid in the tank and replace the sensor.

4.1.6.2 Low Sheath/Empty Sheath Alarm

4.1.6.2.1 When the low/empty sheath alarm goes off, the cytometer STOPS pulling sheath from the fluidics cart and starts pulling it from the sheath reservoir.

4.1.6.2.2 Put the cytometer in standby and refill the sheath tank.

4.1.6.2.3 After you refill the tank, you need to restart the fluidics cart.

4.1.6.2.4 NOTE: If you do not restart the cart, the sensor doesn’t sense the sheath and will drain the

Page 7: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 7 of 27

reservoir causing what can be a lengthy troubleshooting procedure that you will be billed at Operator rates.

4.1.7 Check Sheath Filter for air bubbles (L01/L02/F01)

4.1.7.1 Hold the sheath filter so that the tubing connected to the stopcock/vent cap is facing you.

4.1.7.2 Tap the filter a couple of times to dislodge any bubbles that might be in there.

4.1.7.3 Remove air bubbles from sheath filter:

4.1.7.4 L02: Using the stopcock device.

4.1.7.4.1 Using a container to catch the sheath that comes out, roll back the stopcock to un-pinch the tubing.

4.1.7.4.2 Discard sheath into container until all bubbles are gone, and then roll back the stopcock to pinch the tubing to stop the flow of sheath.

4.1.7.5 L01/F01: Using the vent port on the top of the filter.

4.1.7.5.1 Using a container to catch the sheath that comes out, pinch off the tubing right below the vent cap, and remove the cap.

4.1.7.5.2 Discard sheath into container until all bubbles are gone, and then replace the vent cap so that it pinches off the tubing to stop the flow of sheath.

4.1.8 Priming Cytometer to remove Air Bubbles

4.1.8.1 NOTE: This has to be done every time the cytometer is turned on because of the flow cell emptying and refilling with sheath

Page 8: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 8 of 27

4.1.8.2 Remove the tube of dH2O from SIP and press prime 3 or more times, if necessary, to remove any air inside the flow cell.

4.1.8.3 Note: Never prime the cytometer with a tube on the SIP – otherwise, a potential clog may enter into the flow cell.

5.0 Cytometer Setup and Tracking (CS&T)

CS&T needs to be run EVERY time the cytometer is turned on to ensure consistent laser performance from day to day. Lasers are light bulbs, with use they can blow out at any time.

5.1 Log into the Cytometer Controller computer (see Quick Fact sheet on wall next to Workstation computer) => launch the FACSDiva software by double clicking on the desktop icon => click OK in the Administrator window:

5.2 Note: If you start the software before the cytometer, you will need to manually connect them by selecting Instrument menu => Connect.

5.3 The instrument Status will probably say Fluidics Not Ready, just click “clear”.

5.4 Making CS&T Beads:

5.4.1 Beads are located in the refrigerator behind the L01. Vortex the beads and place 3-4 drops of beads/500 mL of PBS.

5.4.2 Label tube with: CS&T, lot # (found on CS&T vial), date, and your three initials.

5.5 Running CS&T Beads:

5.5.1 Select Cytometer in Workspace menu => CS&T.

5.5.2 The cytometer is now disconnecting from the Diva software and connecting to the CS&T software.

5.5.3 Check the Load Tube Manually box (if not already checked).

5.5.4 Make sure the CS&T Lot ID matches the CS&T Beads being used.

5.5.5 Put the Cytometer in Low and Run, then vortex and load the CS&T beads.

5.5.6 Click Run in the CS&T software, then click OK. CS&T will take about 8 minutes to complete.

5.5.7 When completed, remove the CS&T beads and put back into the refrigerator.

5.5.8 If CS&T:

5.5.8.1 Passed: Exit CS&T Program => after cytometer connects to the Diva software, proceed as usual.

Page 9: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 9 of 27

5.5.8.2 Passed with Warnings (PWW): Same as step 5.5.8.1, but notify the Flow Facility about the warning.

5.5.8.3 Failed: Try making fresh beads, double check Lot ID, shutdown, and restart the computer and cytometer and repeat process. If still failed, please contact the Flow Facility before continuing.

5.5.9 If no Flow Facility personnel are available: Exit the CS&T software and run the CS&T beads in the manual CS&T experiment located in the Diva Browser.

5.5.9.1 Browser => Manual CS&T folder => Double click CS&T experiment to open it => Expand the Yearly specimen => Copy and paste the last tube into the specimen and rename it with the correct date.

5.5.9.2 Run the CS&T beads to make sure that the dim and bright bead populations show up in the FSC/SSC plot (indicates that the blue laser and FSC/SSC PMTs are working) and that all three fluorescent peaks show up in their relative histograms (indicates all lasers/PMTs are working). If all look good, record 1,000 events and continue on as usual with your experiment.

5.5.10 NOTE: It is important that you report any warnings/failures to the Flow Facility so that we can monitor the lasers to prevent a blowout that could shut down the cytometer for use for a lengthy period of time.

5.5.11 NOTE: To see if CS&T has been performed for that day: Click on the CS&T Reports folder on the desktop => Config folder => Year Folder => Month Folder => Day Folder => .pdf of the CS&T Report.

5.5.12 NOTE: The Reset Target Values and Define Baseline procedures will always be done by the Flow Facility. If you get a notice saying that the Baseline has expired, please contact the Flow Facility.

6.0 Data Collection

6.1 Creating a New Experiment in Diva Browser: Note the time – this is the Actual Start time of your session.

6.1.1 Make a Yearly, Monthly and/or Daily folder, if necessary, and name accordingly in the Browser Frame: right click on Administrator => add a New Folder for the Year and name it accordingly => right click on the Yearly folder => add a folder for the Month and name it accordingly => right click on the Monthly folder => add a folder the Day and name it accordingly.

6.1.2 Right click the Daily Folder => New Experiment => 10 Color Template (L01/L02) or 8 Color Template (F01) => OK.

Page 10: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 10 of 27

6.1.3 The Experiment should default open, but if it does not, double click on the Experiment Book to open it.

6.1.4 Right click on the Experiment and rename it with the date (yymmdd), run number, resource (L01/L02/F01), your 3 initials, and your experiment number (e.g. 170101001 L02 PMM005).

6.1.5 Right click on the Specimen and rename it with the date, your initials, resource, and experiment number (e.g. 170101 L02 PMM005).

6.1.6 Expand the Specimen and right click on Tube_001 and rename it with your initials_001 (e.g. PMM_001).

6.1.7 Click the Acquisition Tube Pointer Box to the left of the tube in the Browser Frame => click the Parameters Tab in the Cytometer Frame => delete any unwanted fluorescent parameters by holding down the control key => select the dot next to the parameter you want to delete => click on delete (this saves on memory).

6.1.8 NOTE: We do suggest collecting one empty parameter for gating out autofluorescence if needed.

6.2 Creating Global Worksheets:

A new experiment (created from the icon and not the menu) will default with an empty Normal Worksheet. Normal Worksheets are tube specific, so we recommend using a Global Worksheet which is experiment specific.

6.2.1 To create a Global Worksheet, right click the Experiment book => New Global Worksheet.

6.2.2 Click on the worksheet to add plots.

6.2.3 Make the number of plots needed to satisfy your experimental design/panel, and change the labels accordingly.

6.2.4 You will need at least one dot plot of FSC/SSC, one dot plot with two parameters for compensation and a histogram for each fluorescent parameter you are using (including the dump parameter if you are using one) to set your voltages.

6.2.4.1 To add a dot plot to the worksheet: Click on the Dot plot icon in the Worksheet toolbar, and then click on the worksheet.

6.2.4.2 Click on the dot plot axes to change the parameters (i.e. FSC/SSC).

6.2.5 NOTE: If you have FSC-H (Height) checked, the first dot plot will always default to FSC-A vs FSC-H.

Page 11: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 11 of 27

6.2.6 Note: To make all your plots the same size, select all the plots making sure that the plot of the size you want is selected last and has the yellow sizing boxes on it. Go to the icon in the Worksheet toolbar that says “make plots the same size” (3rd icon from the right).

6.3 Optimizing Instrument Settings: Voltage/Threshold/Compensation

6.3.1 In the Browser Frame, highlight the Unstained tube (should be tube 1 on your protocol sheet) and position the Acquisition Tube Pointer in the box at the left of the tube.

6.3.2 Select Low and Run.

6.3.3 Put the Unstained tube on the SIP and click Acquire in the Acquisition Dashboard Frame.

6.3.4 In the Cytometer Frame, click on the Parameter Tab and adjust the FSC/SSC voltage to place your population of interest in view on the FSC/SSC dot plot.

6.3.5 If you are unsure what your population of interest is, you can back gate using one of your positive controls.

6.3.5.1 Put your positive control on the sip => draw a gate on the positive population => right click on the FSC/SSC plot => Show Population => select the gate around your positive population => look for the color of your gate in the FSC/SSC plot.

6.3.5.2 Adjust the FSC/SSC voltage to place your population of interest on scale.

6.3.5.3 So that you don’t waste your positive control sample, when your population of interest is in view you can delete the gate and put your unstained tube back on to finish the adjustments.

6.3.6 Adjust the Threshold in the Cytometer Frame to exclude most of the debris (each tick mark is equal to 10,000 channels).

6.3.6.1 If you are running Kappa beads, the threshold has to be at 5000 or you will not see the beads.

6.3.7 Draw a Gate around your population of interest (cells or beads).

6.3.8 Select all the other plots by holding down the shift key => right click on one plot => Show Population => select the gate in the FSC/SSC plot.

6.3.9 Using your single color control tubes, adjust the Voltages of the fluorescence channels in the Cytometer Frame => Parameter Tab so that you have an optimal signal-to-noise voltage while making sure that your positive population is on scale.

Page 12: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 12 of 27

6.3.9.1 Tip: For a weak signal (dim antibody labeling), adjust the voltage to get the best signal to noise ratio.

6.3.9.2 Tip: For a strong signal (intense antibody labeling), adjust the voltage so that the positive population is on scale.

6.3.10 After all the voltages have been set, select the Stopping Gate for your target population and the number of Events to Record in the Acquisition Dashboard Frame.

6.3.11 LEAVE THE STORAGE GATE “AT ALL EVENTS”!!!

6.3.12 You can now setup your experimental panel and name the parameters accordingly using the Experiment Layout tool in the Workspace menu.

6.3.12.1 Experiment Menu => Experiment Layout => Label => label your parameters with the correct antibody name and fluorochrome you are using (e.g. CD3-FITC) by either typing the info in the correct box or by selecting an existing label from the drop down menu => ok.

6.3.13 Put your unstained tube on the SIP and click Acquire and then Record.

6.3.14 You can now put the fluidics in High if you choose.

6.3.15 After the sample is finished recording, remove it from the SIP.

6.3.16 Expand the unstained tube in the Browser Frame => right click on its Cytometer Settings and Copy.

6.3.17 Paste the Cytometer Settings into the Experiment Cytometer Settings under the experiment book by right clicking on the Cytometer Settings => Paste.

6.3.18 Select any remaining tubes in your specimen => right click and paste the cytometer settings into them as well. The voltages for your experiment are now set.

6.3.19 Begin manual compensation: Click Next Tube in the Acquisition Dashboard frame to create a new tube or go to the next tube.

6.3.20 Put your first Compensation tube on the SIP, click Acquire, and then Record.

6.3.21 When the sample is finished recording, remove it from the SIP.

6.3.22 Put the fluidics in Standby to stop the flow of sheath.

6.3.23 In your Fluorescent Parameter dot plot, change the parameter on the “X” axis to the positive control you are working with and the parameter on the “Y” axis to your next positive control (every time you change a parameter in the plot, you will need to redraw the gates).

Page 13: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 13 of 27

6.3.24 Note: If using the same gates for two different parameters, always make sure that you move the gate used on the positive population to the other positive population. If not, you will invert the numbers to be matched up.

6.3.25 Click on the Rectangle Gate in the Worksheet Toolbar, and draw it around the Positive population. Make sure it encloses the whole population including those events below the axis (use the biexponential view if needed).

6.3.26 Draw another Rectangle Gate around the Negative population, making sure that you get those events below the x and y axes.

6.3.27 If you have not already created a Population Hierarchy View and a Statistics View box, do so now.

6.3.27.1 Right click on the plot => Show Population Hierarchy.

6.3.27.2 Right click on the plot => Create Statistics View.

6.3.28 The only information you need in the statistics view at this time are the Populations you are matching up and the Parameters you are looking at with their Means (MFI).

6.3.29 To Edit the information you do not need, right click on the Statistics View Box => Edit Statistics => Header Tab => deselect all the header information by unchecking the All box at the top of the column => Populations Tab deselect all the populations you are not looking at as well as the % and # columns by deselecting the box at the top of the column => Statistics Tab select all the fluorescent parameters you are using and their means => OK.

6.3.30 The goal of compensation is to ensure that signals from neighboring channels are appropriately subtracted. For example, if channels X and Y are correctly compensated, then in channel X the MFI of a positive population in channel Y should be the same as the MFI of a negative population in channel Y. To compensate your control tubes, you need to match up the Means of the Positive population to that of the Negative population for all other parameters by adjusting the percent compensation values to add or subtract the Spectral Overlap.

6.3.30.1 NOTE: The positive population value should not be smaller than the negative value or be a negative number when compensation is applied because you will be over compensating that population.

6.3.31 In the Browser, highlight the tube you want to compensate (it should turn Blue) => click the Cytometer Settings Tab in the Inspector Frame => click the Compensation Tab and make sure that the Enable Compensation box is checked.

Page 14: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 14 of 27

6.3.31.1 Make sure that the tube you are trying to compensate is highlighted. If it is not, you will be adjusting compensation for the wrong tube.

6.3.32 Adjusting the compensation can be done a number of ways. You can use the Arrows to make Small Increments of .10, you can hold down the Control Key and use the Arrows for Medium Increments of 1 or you can Type in the Values for Large Increments.

6.3.33 Note: I DO NOT suggest using the Slide Ruler - you will be just chasing your tail.

6.3.34 To help you remember what fluorochrome you need to subtract/add compensation from, look at the top of the compensation table. It has Fluorochrome - % Fluorochrome.

6.3.35 The positive control will ALWAYS be in the - % Fluorochrome column.

6.3.36 Rule of Thumb: if unsure what numbers you need to match up in your statistics box, place your thumb over the positive control and match up all the other fluorochrome columns.

6.3.37 When you finish with the compensation of your first control tube, right click its Cytometer Settings and Copy Spectral Overlap.

6.3.38 Go up to the Global Cytometer Settings (under your experiment book) and right click and Paste Spectral Overlap With Zeros.

6.3.38.1 It is IMPORTANT that you paste with zeros because if you subtract a value to a zero, the software will keep the value.

6.3.39 Go to your next tube and expand it and Paste Spectral Overlap With Zeros into its cytometer settings.

6.3.40 Repeat steps 6.4.34 thru 6.4.42 for the remaining tubes.

6.3.41 When you are finished with the last compensation tube, Paste the Spectral Overlap With Zeros into the Global Cytometer Settings and to all of the previous as well as prior tubes. Click Yes To All. This will guarantee that all tubes have the same cytometer settings. Your samples are now compensated.

6.4 Data Acquisition: Running your Experiment Samples

6.4.1 To keep the software from locking up during acquisition, please make up additional experiments for runs with more than 100 samples. If additional experiments are needed to complete the run, after optimizing the first experiment duplicate the experiment without data.

6.4.2 Put your sample tube on the SIP.

Page 15: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 15 of 27

6.4.3 Click Acquire, wait until the sample gets going and then click Record. Ensure that you are collecting enough events for necessary analysis.

6.4.4 When the sample is finished recording, remove it from the SIP and click Next Tube to go on to the next sample.

6.4.4.1 If you need to Stop Recording before the set number of events have been reached, just click Stop Recording and then Stop Acquiring.

6.4.4.2 There are a number of ways to tell when you sample has finished recording: a bell will ding (if not muted), the blue bar in the acquisition control frame will have reached the end, and the acquisition tube pointer box as well as the tube in the browser will turn green.

6.4.5 Repeat steps 6.5.2 thru 6.5.4 with the rest of your samples.

6.4.6 Standardizing your Experiments: To standardize your experiments, you can run CS&T beads after your first experiment to capture where the peaks of your fluorescence parameters fall. CS&T beads should then be run as the first tube in future experiments to set your voltages to the mean peak of the first experiment. (Ask Flow Facility staff for help with this approach if needed).

7.0 Instrument Shutdown/Data Management

7.1 Determine if you are, or are not, the last person scheduled to run for that day by looking at the calendar in CoreResearch. If there is a session behind you, please double click on that session to see if the User has put “CANCEL” in the Experiment Information box.

7.1.1 IF YOU ARE NOT the last person, then on high run 2 minutes 10% bleach and then 2 minutes dH2O => leave the tube of dH2O on the SIP => put the fluidics in Low and Standby => top off sheath tank => empty Waste (if necessary).

7.1.2 IF YOU ARE the last person, then on high run 5 minutes 10% bleach and then 5 minutes dH2O => leave the tube of NO MORE THAN 1 mL of dH2O on the SIP => put the fluidics in Low and Standby => turn off the cytometer => turn off the wet cart (L02 only) => close Diva software and log out of the cytometer controller computer (do not shutdown computer) => top off sheath tank => empty waste (if necessary).

7.1.2.1 The tube of water is to keep the sheath from crystallizing in the injection probe.

7.1.2.2 You CANNOT have more than 1 mL in the tube because when the cytometer is turned off, or left in standby mode, there is a small amount of fluid that back drips into the

Page 16: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 16 of 27

tube. Too much fluid in the tube could overflow into the air line causing a LENGTHY service call to be placed.

7.1.3 Please DO NOT shutdown the computers, we need access to the export drive.

7.2 Exporting Data

When acquisition is completed you need to Export both your Experiment and your FCS files. We also suggest exporting your Experimental Template if you plan on using it more than once.

7.2.1 If you get an error message when trying to export your data, make sure that the “F” drive is connected to the controller by clicking “My Computer”. If “F” drive says disconnected, click on the Map_Drive icon on the desktop and follow the prompts.

7.2.2 If “F” drive still doesn’t connect, try rebooting both the Controller computer and the Workstation computer.

7.2.3 If “F” drive still fails to connect, export to the local “D” drive and let the Flow Facility staff know so that they can trouble shoot and transfer your data to the correct drive.

7.3 Exporting Experiment:

7.3.1 Right click on the Experiment Book => Export => Export Experiment => OK.

7.3.2 NOTE: The experiment should export to the default location:

F:\BDExport\Experiments

If not, Browse =>”F” drive => BDExport => Experiments.

7.4 Exporting FCS Files:

7.4.1 Right click on the Experiment Book => Export => FCS => 3.0 format => OK => Save.

7.4.2 NOTE: The FCS should export to the default location:

F:\BDExport\FCS

If not, Browse =>”F” drive => BDExport => FCS.

7.5 Exporting Experiment Template:

7.5.1 Right click on the Experiment Book => Export => Experiment Template => Type => Select your Initials => Name the Template so that you will know what template it is in the future => Finish. (e.g. 7 color PBMC).

7.5.2 NOTE: If you check the Lock Template box, this ensures that the template cannot be overwritten by a template with the same name.

7.5.3 NOTE: We suggest that you keep a copy of every different template you make up, because to manage the Data Base, we

Page 17: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 17 of 27

need to delete experiments off of the Browser as well as off of the export drive. These templates stay on the “D” drive until the User leaves the lab or an upgrade to the software is made which will no longer read the older formatted templates.

7.6 Logging time in CoreResearch:

7.6.1 If your session began and ended within the window of your original CoreResearch reservation, you do not need to log your time in CoreResearch. The fundcode associated with your reservation will be automatically billed for the original reservation time.

7.6.2 However, if you began your session before the reservation start time and/or your session extended past your reservation end time, you will need to modify your actual usage. Here are the steps to do this: 7.6.2.1 On the cytometer workstation, open the designated

Internet Browser (not Firefox). CoreResearch should be the default tab. If not, use this link https://medschool.duke.edu/research/core-research-facilities. Log into CoreResearch (see section 11.14.1 in SOP_010 for more direction, if necessary).

7.6.2.2 Navigate to your reservation on the Reservation Calendar, and double-click on it.

7.6.2.3 On the menu at the top of the screen, click “Update Actual Usage.”

7.6.2.4 Change the “Actual Start Time” and/or “Actual End Time” to reflect your early start or late stop. Note: you will not be allowed to make the “Actual Start Time” later than the original reservation start time, nor the “Actual End Time” earlier than the original reservation end time.

7.7 Transferring Data to CDMS/Biotrue:

After you have exported your Experiment and FCS files, you need to Transfer your FCS folder of data to our Collaborative Data Management System (CDMS)/Biotrue from the Workstation computer.

7.7.1 Before you upload your FCS files to Biotrue, you need to put a copy of your protocol sheet in the FCS folder.

7.7.2 Go to the Workstation computer and log in if you haven’t already done so (log in information is on the Quick Fact sheet on the wall next to the Workstation computer).

7.7.3 Open the Biotrue Uploader located on the desktop and log in (log in information is on the Quick Fact sheet on the wall next to the Workstation computer).

7.7.4 Open the BDExport folder located on the desktop => FCS => Days folder => Put your protocol sheet into your FCS folder =>

Page 18: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 18 of 27

Drag your FCS folder into your Lab’s Biotrue dropbox => click Yes, please to upload.

7.7.5 NOTE: If you are unable to Export your data or transfer it to Biotrue, please contact the Flow Facility via email or phone to let us know so that we can assist you.

7.7.6 NOTE: It is the end Users’ responsibility to export and upload their FCS files to Biotrue. Facility staff will no longer be checking behind you to make sure you have uploaded your FCS Files.

7.7.7 NOTE: If you have recorded a LARGE amount of data, please do a property check on your FCS folder to see how big the file is. If it is greater than 500MB, you need to divide the files into two folders before uploading to Biotrue to prevent corruption of any of the files. Please name the second folder the same as the original and put the word “cont.” at the end.

7.8 If you are the last person scheduled for the day, log off of the Workstation computer. Please do not shut down the computer, just log off.

7.9 Spray your cooler with 70% Ethanol before leaving the lab.

7.10 Accessing Biotrue in your Lab:

7.10.1 URL: https://cdms.duhs.duke.edu

7.10.1.1 Username: flowguest

7.10.1.2 Password: DHVIflow0311

7.10.1.3 Select DHVI Flow Cytometry Facility and your lab flow data folder.

8.0 Attachments

8.1 #1: DHVI Flow Facility Contact List

Page 19: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 19 of 27

8.2 #2: Overview of BD LSRII/Fortessa Instruments

8.3 #3: Overview of BD FACSDiva Software

8.4 #4: Cytometer-Specific Protocol Form/Template

8.5 #5: Troubleshooting Sheath Reservoir (L02 only)

Page 20: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 20 of 27

Attachment #1: DHVI Flow Facility Contact List

General Flow Facility Support 919-684-4130 CoreResearch Support Services 919-684-2243 option 4

MSRB2 Help

919-452-2709 GHRB Help

919-452-1751

Labs/Instruments A01 GHRB, 1071 919-684-8813

A02 MSRB2, 4036 919-613-5920 N01 MSRB2, 4038 919-681-0629 F01 MSRB2, 2045 Bay 6 919-684-4130 L01 MSRB2, 2045 Bay 5 919-684-4130 L02 MSRB2, 2045 Bay 6 919-684-4130

Role

Faculty/Staff

Office Phone

Facility Director

Derek Cain, PhD

919-681-8193

Senior Flow Operator

Patrice McDermott

919-684-4130

Flow Engineer Steven Slater 919-684-4130

Billing/Finance

Darlene McCain

919-681-1783

Scientific Advisor, CHAVI-ID Tony Moody, MD

919-668-2551

Scientific Director, DHVI Shared Resources Gregory Sempowski, PhD 919-684-4386

Page 21: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 21 of 27

Attachment #2: Overview of BD LSRII Instrument Figures/text captured from BD LSRII User’s Guide – See full PDF for details, on Facility website. Components

Page 22: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 22 of 27

Control Panel

Page 23: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 23 of 27

Fluid Control

Page 24: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 24 of 27

Sample Injection Port (SIP)

Page 25: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 25 of 27

Attachment #3: Overview of BD FACSDiva Software Go to Flow facility website Training Checklist (https://shared-resources.dhvi.duke.edu/training-materials) and click on the link to the Diva Training Manual. Workspace Components

Menu Bar

Workspace Toolbar Save Browser Frame - Search Field - Browser Toolbar -Folders -Experiments -Specimens -Tubes - Global Instrument Settings - Global Worksheet - Current Tube Pointer Plate Frame Cytometer Frame

-Status of the Instrument -Laser Delay/Area Scaling -Instrument Settings for Current Tube Inspector Frame -Experiment Information -Tube Information -Acquisition Information -Instrument Setting Information -Worksheet Information

-Plot Information -Plate Information -Keyword Information

Worksheet Frame -Worksheet Toolbar -Toggle between Global and Normal Worksheets -Select Arrow -Plots -Magnification Tools -Gating Tools -Editing Tools -Alignment Tools -Population Hierarchy -Statistics View

Acquisition Dashboard Frame

Page 26: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 26 of 27

Attachment #4: Cytometer-Specific Protocol Form/Template Download CURRENT version from Facility website: https://shared-resources.dhvi.duke.edu/dhvi-core-facilities/dhvi-flow-cytometry/downloads

Page 27: Standard Operating Procedure For: Independent End User Training … · 2017-10-02 · Independent End User Training (L01, L02, F01) 1/2017 Page 6 of 27 (the next day) on the tape

Independent End User Training (L01, L02, F01)

1/2017 Page 27 of 27

Attachment #5: Troubleshooting Sheath Reservoir (L02 Only)

1.0 Removing air from system when the sheath reservoir runs dry:

1.1 Put a tube of dH2O on SIP.

1.2 Press prime, and when you see bubble in the water tube press standby (this empties the flow cell).

1.3 While in standby, depressurize the sheath reservoir by loosening the white cap on the reservoir.

1.4 Toggle between prime and run waiting 2 seconds between toggles (repeat a dozen times).

1.5 Put it in standby, and tighten the white cap.

1.6 Wait 40 seconds and then put the machine in run.

1.7 Put the CS&T beads on to make sure that the FSC/SSC are the same as they were during the QC in the morning. If not, repeat the above steps.