1 Copenhagen, Denmark 24 27 September 2018 Stability studies for in vitro diagnostics Mark Lanigan Dossier Assessments Prequalification of Diagnostics Team WHO

Upload: others

Post on 13-Jun-2020

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

1 Copenhagen, Denmark 24 – 27 September 2018

Stability studies for in vitro diagnostics

Mark Lanigan

Dossier Assessments

Prequalification of Diagnostics Team

WHO

Page 2: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• What is stability?

– Why it is important; regulatory framework

• How is stability determined?

– Types of stability studies

– Experience of WHO Prequalification Dossier

Assessment team

2

Presentation overview

Page 3: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• In brief, stability is…

…the ability of an IVD (or its reagents) to

maintain its performance characteristics over

a defined time interval

3

What is stability?

Page 4: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• ISO 13485:2003. Medical devices – Quality management

systems – Requirements for regulatory purposes

– ISO 23640:2011. In vitro diagnostic medical IVDs -

Evaluation of stability of in vitro diagnostic

reagents

– EP25-A Evaluation of Stability of In Vitro Diagnostic

Reagents; Approved Guideline

– ASTM International D4169-14

4

Stability – regulatory framework

Page 5: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• WHO also recommends:

• TGS-2 Establishing stability of in vitro diagnostic

medical devices

• Technical Specifications Series, e.g. TSS-3

Malaria rapid diagnostic tests

• But why are these necessary?

5

Stability – WHO guidance

Page 6: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

6

The WHO Prequalification Programme:

• WHO PQ identifies IVDs suitable in lower- and middle-income countries

• Extremes of operating conditions

• Diversity of user skills

Conditions not likely to be experienced where IVDs have been validated and

regulated

Page 7: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

7

The WHO Prequalification Programme:

• Very few stability studies

take into account use of an

IVD in LIMC

• Satisfactory to NRA

• Have not taken into account

extremes of temperature

and humidity and prolonged

transportation times

Page 8: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• What studies should a manufacturer perform?

• WHO experience with stability studies

8

How should stability be determined?

Page 9: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

9

Stability: three components

Shelf life • How long the IVD can be stored

• Conditions of storage

In-use stability • Stability of components reagents

after opening, reconstitution

Transport stability • Conditions experienced between

manufacture and end-user

Page 10: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

10

Stability: three components

Shelf life How long the IVD can be

stored

Conditions of storage

In-use stability Stability of components

reagents after opening,

reconstitution

Transport stability Conditions experienced

between manufacture

and end-user

Page 11: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

11

A test kit and its components

• Antibodies/antigens,

• Colour-forming reagents,

• Stabilizers,

• Other reagents

Page 12: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

12

Stability must be determined using the final product

Annex to TGS-2 Establishing

component stability for an IVD

Case study: single-use buffer

vials for rapid diagnostic tests

www.who.int

Page 13: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• Stability studies should be based on risk assessment and

take into account, as a minimum:

– Variability of components

– The users’ environment

– Extremes of conditions during transport:

• temperature, humidity, ambient pressure, vibration

13

Stability: risk-based

Page 14: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

25/09/2018 14

Transport stability • Conditions experienced

between manufacture and

end-user

Page 15: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

Challenge at extreme conditions that simulate shipping

15

Transport simulation

High extreme

Temperature, humidity

Low extreme

Temperature, humidity

Normal Storage

Conditions

Normal

Return to

Normal Storage

Conditions

Normal

Determine

performance

Determine

performance

Page 16: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• Actual transport:

– Provides an indication of likely conditions

– Conditions difficult to maintain and/or control

– Conditions are often not recorded

• For example:

– Kits loaded in insulated chest – freighted overnight in

an air-conditioned train carriage

– Does not reflect extreme conditions

16

Why not use a real shipping event?

Page 17: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• Actual transport is useful as a first step; however,

• Shipping stability should be based on simulated

transport challenge using controlled conditions

• Kits should also be subjected to drop-shock testing

[ASTM International D4169-14]

17

Why not use a real shipping event?

Page 18: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• Transport conditions that do not reflect usage in resource-

limited lower- and middle-income countries

• No end-of-life testing

– Kits tested before and after transport challenge

– But effect on shelf-life not determined

• Product performance determined using inappropriate

testing panels

– Specimens that would be unlikely to show that

performance had degraded

18

Other WHO PQ experiences…

Page 19: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

25/09/2018

| Title of the presentation

19

Shelf life • How long the IVD can be stored

• Conditions of storage

Page 20: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

e.g. “…the test can be stored at 5 - 35°C for 12 months…”

20

Determining Shelf life

35 ± 2°C

Normal Storage

Conditions

Determine

performance

T=0

3, 6, 9, 12, 15 months

2-8°C

Determine performance periodically, e.g.

3, 6, 9, 12, 15 months

Does the IVD

remain stable?

Shelf life

= 12 months

Page 21: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

21

Does the IVD remain stable?

• Stability is…

…the ability of an IVD (or its reagents)

to maintain its performance

characteristics over a defined time

interval

Page 22: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

22

Does the IVD remain stable?

Time = at manufacture

Time = near the end of shelf life

Will the IVD will fulfil its

intended use?

‘Positive’ specimen

Page 23: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

23

What testing panel to use?

Low-reactivity

specimen e.g. low concentration of

target antigen

Medium-reactivity

specimen

Highly-reactivity specimen

Very high concentrationof analyate

Page 24: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• Depending on the IVD the testing panel should include:

– A non-reactive [negative] specimen

– Low-reactivity [positive] specimens

– A medium-reactivity [positive] specimen

Control materials on their own may not be sufficient

24

What testing panel to use?

Page 25: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• Depending on the IVD this should include:

– All claimed analytes

• e.g. Plasmodium falciparum and P.vivax

– Whole blood specimens

• Preferably capillary whole blood

25

What testing panel to use?

Page 26: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

26

Ongoing concerns with shelf life studies…

– Claimed shelf life studies:

– Not provided for all kit components

– Did not testing using whole blood specimens

– Tested only a single claimed analyte

• e.g. Plasmodium falciparum but not P.vivax

– Used highly-reactive specimens

– Inappropriate claims:

• “…stored at room temperature…”; “10-30°C”

• “…stable for at least 2 years…”

Page 27: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

25/09/2018 27

In-use stability • Stability of components reagents

after opening, reconstitution

Page 28: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

28

The opened test kit

Page 29: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

29

Test cassette

…, etc.

Stored for e.g. 1 hour at

claimed operating conditions Determine

performance

Stored for e.g. 2 hours Determine

performance

Determine

performance

Stored for e.g. 24 hours?

Page 30: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

30

Buffer in bottle

Open at Time = 0

Determine performance, recap and return to storage

Reopen at e.g. Time = 3 months

Determine performance, recap

and return to storage

Time = 12 months open, recap, test

Time = 15 months open, recap, test

Page 31: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• How does in-use stability of kit components affect over-all performance of the IVD?

• e.g. buffer tested to only 2 months

– Is there sufficient buffer to cover the claimed shelf life of the product (e.g. 12 months) ?

31

In-use stability claim

Page 32: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

• Is in-use stability reflected in the Instructions for Use?

– e.g. “…the cassette should be used immediately after opening…”

– e.g. “…Assay buffer is stable after opening until the end of shelf life…”

32

In-use stability claim

Page 33: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

– In use stability studies do not reflect actual routine use of device

• Buffers only opened once

– Studies not provided for some components (e.g. test cassette)

• “Unnecessary” for single use devices

• Not provided for reagent bottles/buffers

– Inappropriate instructions:

• “…use buffer immediately after opening…”

33

Ongoing concerns with is-use stability studies…

Page 34: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

Reporting requirements

• Each study should include:

• Detailed testing protocol

• Both summary and detailed results

• Clear acceptance criteria and conclusion(s)

• Semi-quantitative scoring of results

e.g. -,w,+,++,+++

Reporting “positive” or “negative” or “10/10

positive” etc. not sufficient

• Preferably photographic records of results

Page 35: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

PQ guidance, sample dossiers and

technical specifications: Technical

guidance series TGS 1 - Standards applicable to the WHO Prequalification of in vitro diagnostic medical device

TGS 2 - Establishing stability of in vitro diagnostic medical devices

Annex to TGS 2 - Establishing component stability for in vitro diagnostic medical devices. Case study: single-use-buffer vials for rapid diagnostic tests - Draft

TGS 3 - Principles of performance studies

TGS 4 - Test method validation for in vitro diagnostic medical devices

TGS 5 - Designing instructions for use for in vitro diagnostic medical devices

TGS 6 - Panels for quality assurance and quality control of in vitro

diagnostic medical devices

TGS 7 - Risk management for manufacturers of in vitro diagnostic

medical devices - Draft open for public comment

Page 36: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

Sample Product Dossier for a CD4 IVD

Sample Product Dossier for an IVD intended for HIV self-testing

Sample Product Dossier for a qualitative nucleic acid-based testing technology

for HIV-1 and HIV-2 to detect HIV-1 and HIV-2

Sample Product Dossier for a quantitative nucleic acid-based

testing technology to measure HIV-1 RNA

PQ guidance, sample dossiers and

technical specifications: Sample

dossiers

Page 37: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

PQ guidance, sample dossiers and

technical specifications: Technical

specifications series

TSS 1 - Technical Specification Series to WHO Prequalification - Diagnostics

Assessment: HIV rapid diagnostic tests for professional use and/or self-testing

TSS 2 - Technical Specification Series for submission to WHO Prequalification -

Diagnostic Assessment: IVD to identify Glucose-6-phosphate dehydrogenase

(G6PD) activity

TSS 3 - Technical Specification Series for submission to WHO Prequalification –

Diagnostic Assessment: Malaria rapid diagnostic tests

TSS 4 - Technical Specification Series for submission to WHO Prequalification –

Diagnostic Assessment: IVDs used for the detection of high-risk Human

Papillomavirus (HPV) types in cervical cancer screening

TSS 5 - Technical Specification Series for submission to WHO Prequalification –

Diagnostic Assessment: Rapid diagnostic tests used for surveillance and

detection of an outbreak of cholera

Page 38: Stability studies for in vitro diagnostics · – In use stability studies do not reflect actual routine use of device • Buffers only opened once – Studies not provided for some

38

Thank you

Any questions?