stability of the electron energy recycler department meeting april 5 th, 2006 a. shemyakin, m....
TRANSCRIPT
Stability of the electron energy Stability of the electron energy
Recycler department meetingApril 5th, 2006
A. Shemyakin, M. Sutherland
2
MotivationMotivation
In calibrations of the electron energy by cooling of pbars with RF off, we saw the energy shifting by more than 1 keV at the same settings
Sometimes adjusting the Pelletron voltage seemed to improve the cooling rate
Part of the problem is combination of a drift of the charging current and a finite amplification of the HV regulation loop
3
Regulation circuitryRegulation circuitry
-30
-20
-10
0
10
20
30
40
50
90 110 130 150Chain current, microA
Nee
dle
cu
rren
t, m
icro
A
Gri
d v
olt
age,
V
4.316
4.318
4.32
4.322
4.324
4.326
4.328
4.33
4.332
Ter
min
al v
olt
age,
MV
R:GRIDV
R:TPSPRB
R:GVMVLT
Terminal
Chain current Needle
current
Resistor chainPelletron HV regulation circuitry
0
5
10
15
20
25
30
35
40
45
50
-25 -20 -15 -10 -5 0
Grid voltage, V
Nee
dle
cu
rren
t (R
:TP
SP
RB
),m
icro
A The terminal voltage (GVMVLT), grid voltage (GRIDV, and the needle current (TPSPRB) as functions of the chain current. 4-Apr-2006. The voltage changes as 0.15 kV/A.
Needle current as a function of the grid voltage
Generation voltmeter (GVM)
4
Dispersion – based energy deviation Dispersion – based energy deviation measurementmeasurement
U-bend
BYD08
BYA07
BYA05
CollectorGun
1050 mm
The idea is to use the high dispersion in U-bend to trace the energy changes, while Autotunes keep the beam positions constant in BPMS A05 and A07
5
BPM driftsBPM drifts
RMS of the BPM noise corresponds to ~20 eV
BXD08S
BYD08S
36 hours
0.02 mm
BPM readings of a calibration signal. 32.05 kHz modulation, 1 Hz recording. Standard deviations in 2 hours are 1.8 m for X and 3.4 m for Y.
6
CalibrationCalibration
-2.0
-1.5
-1.0
-0.5
0.0
0.5
1.0
1.5
2.0
2.5
3.0
4.310 4.315 4.320 4.325 4.330 4.335
GVM, MV
BY
D08
S,
mm
-1.980
-1.970
-1.960
-1.950
-1.940
-1.930
BX
D08
S,
mm
R:BYD08S
R:BXD08S
BPM D08 reading as a function of the Pelletron voltage. 14-Mar-06, Mary
The Y derivative is 0.219 mm/kV and is consistent with the bending radius.
The X derivative is -0.001 mm/kV.
7
Data from 3-hour runData from 3-hour run
GVM readings and BYD08S, BXD08S (multiplied by the calibration coefficient). 1 Hz data. 29-Mar-2006, Mary.
Deviation of energy with time. All points
9 9.5 10 10.5 11 11.5 124
3
2
1
0
Time, hour
dE, k
eV
GVM
dEx
dEy
8
Data with 10 sec averagingData with 10 sec averaging
9 9.5 10 10.5 11 11.5 124
3
2
1
0
Time, hour
dE, k
eV
The drift of the GVM value is 0.9 kV and is consistent with the drop of the chain current by 6 A.
The drift measured by BYD08S is 2.7 keV.
9
Comparison of energy deviations estimated in 3 Comparison of energy deviations estimated in 3 BPMsBPMs
Deviation of energy estimated by 3 BPMs
9 9.5 10 10.5 11 11.5 124
3
2
1
0
1
Time, hour
dE, k
eVi3 0 1 50 An energy change without
any response in GVM
10.2 10.22 10.24 10.26 10.283
2.5
2
1.5
Time, hour
dE, k
eV
6 min
1.5kV
BYD08
BYD07
BXD07
Changes of the beam positions in three BPMs multiplied by corresponding dispersion coefficients and adjusted to begin from zero.
10
Possible reasons for the discrepancy Possible reasons for the discrepancy
Drift of the GVM electronics Kermit is preparing a test of its stability
Drift of the GVM rotation frequency Kermit may try to measure the frequency
Changes in the distance between GVM and the terminal GVM measures the electric field, not the potential If the distance change, the measured voltage
deviates as 5.2 kV/mm Axially symmetrical thermal effect is 710-6/K The distance stays the same within 0.05 mm from
opening to opening of the Pelletron tank Non-axially symmetric motion is still not excluded
11
ConclusionConclusion
The long-term stability of the Pelletron voltage is poor, at least after turning the Pelletron on. Kermit plans to determine stability of electronics.
The procedure of measuring the energy deviation by the beam position in a high- dispersion region works to some extend. A beam motion upstream the bend seems to be an issue.
One can consider implementing similar procedure in the full line First, make an absolute energy calibration with pbars Save e-beam settings as “energy calibration” file In normal operation, compare readings in BYR01S with
the calibration
12
180 deg bend as spectrometer180 deg bend as spectrometer
Fix the beam positions in BPMs C90, Q01 with an Autotune
Measure shifts in BYR01S