ST79: Essentials for ASCs
January 21, 2015
© 3M 2015 All Rights ReservedToday’s meeting time: 1 - 2 p.m. CST
2015 All Rights Reserved.© 3M
House Keeping
FROM THE GO TO WEBINAR PAGE:
Click on the orange box with a white arrow to expand your control panel (upper right-hand corner of your screen).
Type a question in the question box and click send.
2015 All Rights Reserved.© 3M
How do I get a CE Certificate?
Next week, all of today’s meeting participants will be sent an email containing instructions for obtaining a CE certificate for today’s meeting.
The email will be sent to the email address you provided when you logged-in to today’s meeting. If there are others listening with you today who did not log-on, you may forward the CE certificate email to them.
2
DISCLOSURE STATEMENT
Dorothy Larson
Technical Service RepresentativeInfection Prevention Division3M Health Care
Sterilization Tech Line 1-800-441-1922 Option 2
© 3M 2015 All Rights Reserved
Learning Objectives
© 3M 2015 All Rights Reserved
Discuss AAMI ST79 and the need for the recommended practice.
Discuss the different types of steam sterilization processes and cycle parameters.
Describe the different types of sterilization monitoring devices, including PCDs.
Describe routine load release for implants and non-implants.
ANSI/AAMI ST79Comprehensive guide to steam sterilization and sterility assurance in health care facilities
•A voluntary, nonprofit organization
•Writes recommended practices and standards
•Purpose is to assist health care professions and industry with the use, acceptance, and advancement of medical technology
© 3M 2015 All Rights Reserved
3
ANSI/AAMI ST79Comprehensive guide to steam sterilization and sterility assurance in health care facilities
•Promote Sterility Assurance
•Quality of Care
•Policy and Procedure Development
© 3M 2015 All Rights Reserved
How do AAMI Standards relate to ANSI and ISO Standards?
ANSI/AAMI•ANSI - American National Standards Institute•Coordinates the development and promotion of U.S. voluntary
standards, representing the U.S. in setting international standards
ANSI/AAMI/ISO•International Standards Organization
•Writes standards that apply across the world
© 3M 2015 All Rights Reserved
EFFICACY OF STEAM STERILIZATION(AAMI ST79 INTRODUCTION – OVERVIEW)
1 • Lowering and limiting bioburden
2• Appropriately preparing devices
3 • Selecting the correct sterilization parameters
4• Maintaining the sterility of sterilized items until used
Why is the AAMI ST79 Recommended Practice needed?
© 3M 2015 All Rights Reserved
4
• Types of steam sterilization processes• Cycle parameters• Terminal and Immediate Use Steam Sterilization
© 3M 2015 All Rights Reserved
Types of Steam Sterilization Processes• GRAVITY DISPLACEMENT
– Incoming steam displaces residual air through a port or drain in or near the bottom of the sterilizer chamber
• DYNAMIC-AIR-REMOVAL
Prevaccuum
» Series of pressure and vacuum excursions
Steam-flush pressure-pulse (SFPP)
» Series of steam flushes and pressure pulses above atmospheric pressure
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 2.25, 2.26 © 3M 2015 All Rights Reserved
xxxxxxxxxxxx
xxxxxxxxxxxx
Cycle Selection: Gravity or Dynamic-air-removal
“The reusable medical device manufacturer is responsible for ensuring that the device can be effectively cleaned and sterilized.”
“The written IFU of the device manufacturer should always be followed.”
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 7.2.2 © 3M 2015 All Rights Reserved
5
Healthcare Facility ResponsibilityOBTAIN AND ADHERE TO MANUFACTURER’S WRITTEN
INSTRUCTIONS
• Cleaning
• Cycle type
• Exposure time
• Temperature
• Drying time (if recommended)
Reconcile cycle parameter differences between the Sterilizer Manufacturer, Device Manufacturer, and Packaging Manufacturer
AORN RP for Sterilization in Perioperative Practice Setting, RP VI.b. and VII.a.2. © 3M 2015 All Rights Reserved
Section 8.6 Sterilization Parameters
© 3M 2015 All Rights Reserved
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 8.6
Terminal Sterilization
“Terminal sterilization: process whereby
product is sterilized within its sterile barrier
system.” (ST79, section 2.133)
“The process of sterilizing a packaged item.*”
© 3M 2015 All Rights Reserved
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 (Consolidated Text), Section 2.133*CSSD Dictionary and Reference Guide
6
Immediate-Use Steam Sterilization (IUSS)
• April/2010 Summit included representatives from 12 key organizations, accreditation bodies and agencies
• Produced a multi-society Position Statement
• Agreed on terminology transition from “Flash Sterilization” to the term Immediate Use Steam Sterilization (IUSS)
© 3M 2015 All Rights Reserved
http://www.aami.org/publications/standards/ST79_Immediate_Use_Statement.pdfANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013
Multi-society Position Statement on Immediate-Use Steam Sterilization (IUSS)EVOLUTION OF FLASH STERILIZATION• “Immediate use” i.e. item is removed from sterilizer and aseptically transferred to the sterile field• Agreed on terminology transition from “Flash Sterilization” to the term Immediate Use Steam Sterilization (IUSS)• Personnel should be knowledgeable and competent • Standardized practices should be implemented• Critical reprocessing steps (cleaning, decontamination, and aseptic transport) must be followed• Device manufacturer's written reprocessing instructions should be followed
© 3M 2015 All Rights Reserved
http://www.aami.org/publications/standards/ST79_Immediate_Use_Statement.pdfANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013
Immediate-Use Steam Sterilization (IUSS)
© 3M 2015 All Rights Reserved
“Process designed for the cleaning, steam sterilization, and delivery of patient care items for immediate use. Previously known as flash sterilization.”
Little or No dry time
No storage Item(s) transferred immediately to the point of use
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 2.61, 8.6.2.1, and IntroductionAORN RP for Sterilization in Perioperative Practice Setting, RP VII
Sterilization of unwrapped trays of instruments for immediate use is not recommended.“Implantables should not be sterilized for immediate use (CDC, 2008).”
7
Immediate-Use Steam Sterilization (IUSS)
“Unless a device manufacturer specifically recommends the use of gravity cycles, dynamic air removal cycles should be the cycle of choice.” (8.6.2.1)
© 3M 2015 All Rights Reserved
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013
Immediate-Use Steam Sterilization (IUSS)AORN RP for Sterilization“Immediate use steam sterilization (IUSS) should be kept to a minimum and should be used only in selected clinical situations and in a controlled manner.”(Sterilization RP VII)“Biological and chemical indicators should be used to monitor sterilizer efficacy and to assess whether conditions of sterilization have been achieved. (Sterilization RP VII.c.2.)“A Class 5 chemical integrating indicator or a class 6 indicator should be used within each sterilizer container or tray. Class 6 indicators are cycle-specific and should be used only in the specific cycles for which they are labeled." (Sterilization RP VII.c.3.)
© 3M 2015 All Rights Reserved
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 2.61, IntroductionAORN RP for Sterilization in Perioperative Practice Setting, RP VII
CMS and IUSS2014 MEMO TO STATE SURVEY AGENCY DIRECTORS ON
IMMEDIATE USE STEAM STERILIZATION (IUSS) IN SURGICAL SETTINGS- IUSS replaces the term ‘flash’ sterilization
SURVEY PROCEDURES • Consistent with professionally acceptable standards of practice for IUSS. For example: IUSS reserved for immediate use needs Process to ensure IUSS is not used for implants; single-use devices; devices not
validated for the specific cycle Instruments are cleaned and disinfected according to MIFU prior to IUSS Personnel are trained and have had their competency verified
© 3M 2015 All Rights Reserved
http://www.cms.gov/Medicare/Provider-Enrollment-and-Certification/SurveyCertificationGenInfo/Downloads/Survey-and-Cert-Letter-14-44.pdf
8
• Sterilization Process Monitoring Devices• Process Challenge Devices• Documentation
© 3M 2015 All Rights Reserved
Quality control recommended practices• Association for the Advancement of Medical
Instrumentation (AAMI) • Comprehensive guide to steam sterilization and sterility assurance in
health care facilities, ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013 (Consolidated Text)
• Association of periOperative Registered Nurses (AORN)
• AORN Recommended Practices for Sterilization in PerioperativePractice Setting (2014)
• AORN Recommended Practices for Selection and Use of Packaging Systems for Sterilization (2014)
© 3M 2015 All Rights Reserved
Sterilization Process Monitoring
© 3M 2015 All Rights Reserved
Goal of Sterilization Process:
To kill microorganisms
You can’t see sterility!
9
2015 All Rights Reserved.© 3M
Monitoring Tools• Physical monitors
• Bowie-Dick tests
• Chemical indicators
- External, Internal
• Biological indicators
• Process challenge devices (PCDs)test packs or challenge packs
Sterilization Process Monitoring
2015 All Rights Reserved.© 3M
Critical VariablesCritical Variables
• Parameters identified as being essential to the sterilization process (and required monitoring)
• For example, the critical variables for steam sterilization to be effective are:
Time Temperature Steam
ANSI/AAMI/ISO 11140-1:2005/(R)2010 (Mfrs. Standard)
2015 All Rights Reserved.© 3M
Physical Monitors
Verify that parameters of sterilization cycle are met• Recording charts, Gauges
• Printouts
• Digital displays
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 10.5.1
Limitations• Generally only monitor one
location in the sterilizer• Conditions not revealed in the
temperature recording, such as:
• Improper packaging
• Improper loading procedures
"If the interpretation of the physical monitors suggests inadequate steam processing, the contents of the load should not be dispensed or used.”
10
2015 All Rights Reserved.© 3M
Physical Monitors•Marked with correct date and sterilizer identification at beginning of cycle
• Read & verified by initialing at end of cycle
• If not correct, do not release load
Sterilizers that do not have recording devices should not be used.
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013 , Section 10.5.1
2015 All Rights Reserved.© 3M
Chemical Indicator Classifications
CLASS 1 PROCESS INDICATORS – EXTERNAL CI
CLASS 2 INDICATORS FOR USE IN SPECIFIC TESTS
INTERNAL CIS: CLASS 3, 4, 5, 6
CLASS 3 SINGLE VARIABLE INDICATORS
CLASS 4 MULTI-VARIABLE INDICATORS
CLASS 5 INTEGRATING INDICATORS
CLASS 6 EMULATING INDICATORS
Ref: ANSI/AAMI/ISO 11140-1, 2005 Sterilization of health care products-Chemical indicators -Part 1: General requirements
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 10.5.2.1
2015 All Rights Reserved.© 3M
Chemical IndicatorsCLASS 1: PROCESS INDICATORSEXPOSURE CONTROL (E.G., INDICATOR TAPES)
Use with individual units to indicate that the unit has been directly exposed to the sterilization process
– Distinguish between processed and unprocessed units
– Designed to react to one or more of the critical process variables
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 10.5.2.1
11
2015 All Rights Reserved.© 3M
Chemical Indicators
CLASS 2: INDICATORS FOR SPECIFIC TESTS
TESTING STERILIZER PERFORMANCE
Bowie-Dick Test monitors efficacy of air
removal and steam penetration in 270-275°F
vacuum assisted sterilizers
BOWIE-DICK TESTS CAN DETECTAir leaks, Inadequate air removal, Inadequate steam penetration, Presence of non-condensable gases (air or gases from boiler additives)
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 10.5.2.1
2015 All Rights Reserved.© 3M
Bowie-Dick Test
2015 All Rights Reserved.© 3M
Bowie-Dick Test• Uniform color change – Use• If not uniform, shut down and
report to supervisor• Remain in use• Retest• Call for repair
• Re-qualify if major repair
12
2015 All Rights Reserved.© 3M
Chemical Indicators – Internal CIsCLASS 3: SINGLE VARIABLE INDICATORS (INTERNAL CIS)
Designed to react to one of the critical variables
Indicates exposure to a sterilization process at a stated value (SV) of the chosen variable
Do not provide as much information as Class 4, 5, or 6
CLASS 4: MULTI-VARIABLE INDICATORS (INTERNAL CIS)
• Designed to react to two or more of the critical variables
• Indicates exposure to a sterilization cycle at stated values of the chosen variables
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 10.5.2.1
2015 All Rights Reserved.© 3M
Chemical Indicators – Internal CIsCLASS 5: INTEGRATING INDICATORS (INTERNAL CIS)
• Designed to react to all critical variables • Stated values equivalent to the performance requirements for BIs (ISO 11138 series: 2006)
• Testing conducted in a resistometer, under ideal saturated steam sterilization conditions
• Response must correlate to a BI at three time/temperature relationships • 250°F/121°C, 275°F/135°C, and at one or more temperatures in
between• Stated value at 250°F/121°C must be >16.5 min
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 &A3:2012 & A4:2013, Section 10.5.2.1
2015 All Rights Reserved.© 3M
Chemical Indicators CLASS 6: EMULATING INDICATORS (INTERNAL CIS)
–Designed to react to all critical variables
–Stated value generated from the critical variables of the specific sterilization process
–Cycle verification indicators
“Class 6 emulating indicators are cycle-specific; that is, they should be used only in the specific cycles for which they are labeled.”
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 10.5.2.1 and 10.5.2.2
13
Biological Indicators
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 2.7
Definition• Test systems containing viable microorganisms providing a
defined resistance to a specified sterilization process
Biological IndicatorsWith enzyme-based early-readout (AAMI)
• Fluorescence in 30 minutes to 3 hours• Visual color change in 24 to 48 hours
Conventional• Visual color change in 10 - 48 hours
Manufacturer’s incubation instructions should be followedANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 10.5.3.1“Periodic verification of the early-readout with spore growth should be performed in accordance with the manufacturer’s instructions and facility policy and procedures.”
Biological Indicators - Positive BI ControlIncubate a positive BI control each day a test vial is incubated in each incubator or auto-reader
• From same lot number as test BI
Purpose is to verify the test system is working and ensure• Correct incubation conditions
• Viability of spores
• Capability of medium to promote growth
• Proper functioning of auto-reader and incubator
Good Science
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Sections 10.7.2.3, 10.7.3.3, 10.7.4.3
14
Process Challenge Device (PCD) Definition“Item designed to constitute a defined resistance to a sterilization process and used to assess performance of the process.”
• ‘Process challenge device’ is used in place of ‘test’ or ‘challenge pack’ in AAMI ST79
Representative of load and creates the greatest challenge
• Equal to or greater than the most difficult item to sterilize
• Placed in most difficult area in sterilizer for the sterilant to penetrate
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 2.95© 3M 2015 All Rights Reserved
Process Challenge Device (PCD)
Depending on the application (type of testing being done) the PCD may contain…
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 10.5.4
A BI A BI & a Class 5 Integrating CI
A Class 5 Integrating CI
A Class 6 Emulating CI
© 3M 2015 All Rights Reserved
Process Challenge Device (PCD)• User assembled
• Challenge test pack or tray (e.g., AAMI 16-towel pack, IUSS containment device)
• Commercially available, FDA Cleared• Not available for IUSS
• Not available for Table-top sterilizers
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Section 10.5.4© 3M 2015 All Rights Reserved
15
Documentation“Documentation demonstrates compliance with regulatory
and accrediting agency requirements and identifies trends and quality improvement opportunities.”
AORN Recommended Practices for Sterilization in Perioperative Practice Setting RP XVIII
Documentation
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Sections 10.3.2
• Sterilizer and load records• Load contents• Date and time of cycle• Exposure time and temperature• Initials of operator• Biological results• Chemical indicator results
• Records of sterilizer repair, maintenance and calibration
• Maintain records for time determined by risk manager
© 3M 2015 All Rights Reserved
IUSS Sterilization AORN Documentation• Log or database to trace sterilized items used on patients should include:• Information on each load
• Item(s) processed• Patient on whom the items were used• Type of cycle (e.g., gravity-displacement, dynamic-air-removal)• Cycle parameters used (e.g., temperature, duration of cycle)• Monitoring results• Day and time cycle was run• Operator information (i.e., person who initiated the cycle, person who
retrieved the item from the sterilizer)• Reason for IUSS
AORN Recommended Practices for Sterilization in Perioperative Practice Setting RP VII.g.1.© 3M 2015 All Rights Reserved
16
•Load Release – Nonimplants and Implants•Testing a Combination Sterilizer•Routine Efficacy Monitoring
© 3M 2015 All Rights Reserved
2015 All Rights Reserved.© 3M
• Physical monitors• External process indicator (Class 1) on every package• Internal CI (Class 3, 4, 5 or 6) inside every package• Optional monitoring of load with a PCD containing a
• BI• BI and a Class 5 CI • Class 5 integrating indicator• Class 6 emulating indicator
• Evaluation of all data by an experienced, knowledgeable person• Do not distribute load if any data suggests a sterilization process failure
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Sections 10.6.1,10.6.2
Routine Load Release Nonimplants
2015 All Rights Reserved.© 3M
• Physical monitors• External process indicator (Class 1) on every package• Internal CI (Class 3, 4, 5 or 6) inside every package• A PCD containing a BI and a Class 5 integrating indicator
- Class 6 CI in a PCD may be used in addition to the BI PCD
• Evaluation of all data by an experienced, knowledgeable person• Do not distribute load if any data suggests a sterilization process failure
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Sections 10.6.1,10.6.2
Routine Load Release Implants
17
2015 All Rights Reserved.© 3M
Routine Sterilizer Efficacy Monitoring“Combination” sterilizer - Example
Cycle Temp. Exposure/ Dry Time
Monitor with BI PCD?
Prevacuum“Terminal” Mixed load
270°F 4 min. exp., 30 min. dry
Gravity-Displacement “Terminal”
250°F30 min. exp., 45 min. dry
Prevacuum“Immediate-Use Sterilization”
270°F 4 min. exp., 1 min. dry
Which cycles should be routinely monitored with a BI PCD?
2015 All Rights Reserved.© 3M
Combination Sterilizer – Sterilizers > 2 cu. ft.EXAMPLE, Cont.
Bowie-Dick Test Pack - (prevacuum)Run empty-chamber warm-up cycle first, to pre-heat the chamber.
Empty chamber
IUSS cycle: BI PCD - Place a BI inside container, along with a CI
No instrumentsEmpty chamber
Terminal cycles: BI PCD First load(Each different cycle type)
Implant Loads: Include PCD that contains a BI + Class 5 CI
Implantloads
Frequency of Monitoring with a BI PCD: Weekly, preferably daily, AAMI ST79 Section 10 - Table 6
2015 All Rights Reserved.© 3M
Table-Top Steam Sterilizers• ROUTINE EFFICACY MONITORING WITH A BI PCD
• Representative of the package or tray routinely processed• Most difficult to sterilize
• Contains items normally present during routine sterilization• Full load in cold point (check with sterilizer Mfr.)• Bowie-Dick Test – Check with Sterilizer Manufacturer
Frequency of Monitoring with a BI PCD: Weekly, preferably daily, AAMI ST79 Section 10 - Table 6
Routinely Process: User-assembled BI PCD:
Peel pouches BI + CI inside peel pouch
Wrapped instruments BI + C inside wrapped tray
Peel pouchesWrapped instruments
BI + CI inside wrapped tray
18
2015 All Rights Reserved.© 3M
Sterilizers larger than 2 cubic feet • AAMI 16 towel pack or commercially available disposable, FDA cleared BI PCD• Full load on bottom shelf over drain
Table top sterilizers• BI PCD should be representative of the package or tray routinely processed, and
most difficult to sterilize• Contains items normally present during routine sterilization• Full load in cold point (check with sterilizer mfr.)
Immediate use steam sterilization cycles• BI PCD placed on bottom shelf over the drain• Otherwise empty chamber• Each type of tray configuration used should be tested
Routine Efficacy Monitoring with a BI PCDFrequency of Monitoring with a BI PCD: Weekly, preferably daily
ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013, Sections 10.5.3.2, 10.6.1,10.6.2, 10.7.3.1, 10.7.3.2, 10.7.4.1, 10.7.4.2
2015 All Rights Reserved.© 3M
Key Learnings:
According to ANSI/AAMI ST79, for sterilizers larger than 2 cubic feet, routine sterilizer efficacy monitoring with a BI PCD is done in a/an ___________ chamber.
According to ANSI/AAMI ST79, for immediate-use steam sterilization, routine sterilizer efficacy monitoring with a BI PCD is done in a/an _______ chamber.
Facility policies and procedures on sterilization process monitoring should be based on the recommended practices and guidelines from AAMI, AORN, and the CDC.
Take the time to do proper documentation.
fully loaded
empty
2015 All Rights Reserved.© 3M
19
2015 All Rights Reserved.© 3M
References• Association for the Advancement of Medical Instrumentation (AAMI)
• Comprehensive guide to steam sterilization and sterility assurance in health care facilities, ANSI/AAMI ST79:2010 & A1:2010 & A2:2011 & A3:2012 & A4:2013 (Consolidated Text)
• Sterilization of health care products-chemical indicators-Part 1: general requirements. ANSI/AAMI/ISO 11140-1, 2005/(R) 2010
• Multi-society Position Statement on Immediate-Use Steam Sterilization, accessed 11.13.14 at http://www.aami.org/publications/standards/ST79_Immediate_Use_Statement.pdf
2015 All Rights Reserved.© 3M
How to Purchase AAMI Standards for Your Reference Library
• AAMI documents can be purchased through AAMI by credit card using the following four options:
• Internet: http://marketplace.aami.org • Call: 1-877-249-8226• Fax: 301-206-9789• Mail: AAMI Publications, P.O. Box 0211, Annapolis Junction, MD 20701-0211
2015 All Rights Reserved.© 3M
References• Centers for Medicare & Medicaid Services
• Center for Clinical Standards and Quality/Survey & Certification Group• http://www.cms.gov/Medicare/Provider-Enrollment-and-
Certification/SurveyCertificationGenInfo/Downloads/Survey-and-Cert-Letter-14-44.pdf
• Association of periOperative Registered Nurses (AORN) • Perioperative Standards and Recommended Practices (2014)
• Recommended Practices for Sterilization in Perioperative Practice Setting• Recommended Practices for Selection and Use of Packaging Systems for
Sterilization
20
2015 All Rights Reserved.© 3M
References• The International Association of Healthcare Central Service Materiel Management
• CSSD Dictionary and Reference Guide (2010)