spiration valve system - olympus€¦ · valve with anchors for 0% migration and expectoration....

8
0000 SPIRATION ® VALVE SYSTEM For the Treatment of Emphysema or Air Leaks.

Upload: others

Post on 23-Oct-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

  • 0000

    SPIRATION® VALVE SYSTEMFor the Treatment of Emphysema or Air Leaks.

  • ENGINEERED FOR AIRWAY MANAGEMENT

    Inspired by aerodynamics, the Spiration Valve redirects air away from diseased or damaged lung to healthier

    tissue, all while allowing trapped air and secretions to escape, so that patients may breathe easier.

    Unlike a stent, the unique design of the Spiration Valve minimizes contact with the bronchial wall, maintains

    position to redirect air even in complex patients, and facilitates removal when needed.

    The Spiration Valve System is intended to treat severely diseased lung in patients with heterogeneous

    emphysema and evidence or markers of low collateral ventilation such as complete fi ssures, or damaged lung

    resulting in air leaks by limiting airfl ow to selected areas.

    The procedure is considered minimally invasive and can be performed through a fl exible bronchoscope.

    On inhalation, the Spiration Valve conforms to the airway and redirects air

    to healthier tissues of the lung.

    On exhalation, the Spiration Valve fl exibly constricts with the airways,

    allowing trapped air and secretions to escape naturally along the bronchial

    wall between the umbrella struts.

  • Removable

    The fi rst and only

    valve with anchors

    for 0% migration and

    expectoration. 2,14,15

    The fi rst and only

    valve with a center rod

    to facilitate removal.

    The fi rst and only valve with

    umbrella struts to minimize tissue

    contact and allow secretions

    to escape naturally along the

    bronchial wall.

    Catheters for 2.0mm and 2.6mm

    working channels.

    5mm

    10mm

    8mm

    6mm

    11mm

    9mm

    7mm

    12mm

    10mm

    9mm

    12mm

    12mm

    Stays in Place

    Greater Airway Access

    Largest Range of Valve Sizes

    Now available in 9mm size for large airways.

    Redirects Air

  • THE CONFIDENCE YOU NEED FOR THE OUTCOMES YOU WANT.· Target lobe will have severe, highly heterogeneous emphysema with complete fissures. 1,2,3

    · Multiple valves are placed to occlude all the airways leading to the targeted lobe and enable

    atelectasis or significant lobar reduction.

    · Significant lobar reduction can relieve hyperinflation, enable healthier tissue to expand,

    and make breathing easier. 3

    FOR THE TREATMENT OF EMPHYSEMA

  • From reliable, non invasive evidence of low collateral

    ventilation...

    SeleCT enables clinicians to submit HRCTs and receive quantitative

    measures of fi ssure integrity.

    · More accurate and reproducible than visual CT analysis. 16

    · Comparable accuracy to Chartis. 16

    · Avoids an invasive procedure just to confirm collateral ventilation. 16

    · Not dependent on anatomy, coughing, or mucus where direct

    bronchoscopic measure may be unreliable or not possible. 16

    · Provides useful measures beyond complete fissures such as

    emphysema severity, heterogeneity and lobar volume that may

    improve the prediction of lung volume reduction. 16

    ...to 0% migration or expectoration 2,14,15 for effective lobar

    occlusion

    Complete occlusion is a prerequisite for signifi cant volume reduction,

    which correlates with improvement in clinical outcomes.17

    · FEV1 was improved by 21.4% at 90 days, with statistically significant

    differences also seen across IVC, RV, RV/TLC, 6MWD, mMRC, SGRQ,

    BODE and ADO. 17

    · 99% technical success. 2,14

    · Now available in 9mm size for large airways.

    by

    THE RIGHT PATIENT. THE RIGHT PRODUCT. THE RIGHT OUTCOMES.

  • THE PRECISION YOU NEED FOR THE OUTCOMES YOU WANT.· Persistent air leaks impact inpatient and outpatient resources utilization, cost, and morbidity. 4,5,6,7

    · An air leak present by day 5 should be considered for thoracic surgical consultation or treatment. 5,8,9,10

    · A substantial reduction in an air leak using Spiration Valves may accelerate the resolution of an air leak.

    Complete cessation of an air leak may not be achievable, or necessary, for successful treatment. 9,11,12,13

    THE PRECISION YOU NEED FOR THE OUTCOMES YOU WANT.THE PRECISION YOU NEED FOR THE OUTCOMES YOU WANT.

    FOR THE TREATMENT OF AIR LEAKS

  • A proven isolation technique9,18 to identify the source of air

    leak(s)...

    From the leader in valve therapy for air leaks with over two thousand

    procedures supported worldwide.

    1. Assess

    Block main bronchus to determine if the leak can be stopped or

    reduced and the length of time it takes to see a change in the water

    seal monitor.

    2. Isolate

    Systematically work from proximal to distal.

    3. Place Valve

    Once an airway is identified, size the airway and place a valve.

    4. Reassess

    Repeat process to isolate additional leaks as dynamics may have

    changed since valve placement.*

    ...to a minimally invasive solution with a center rod to facilitate

    removal upon resolution of the air leak.

    Treatment with the Spiration Valve System has demonstrated a favorable

    responder rate.12,13,18

    · 100% (n=9/9) of treated patients met the primary endpoint, as

    identified as air leak cessation allowing chest tube removal. 13

    · 77% (n=10/13) of treated patients were responders, as identified

    as successful chest tube removal without the need for further

    interventions. 12

    · 100% (n=7/7) of treated patients had improvement in the air leak,

    as defined as improvement allowing Heimlich valve use, improvement

    of 1 Cerfolio classification, or complete cessation of the air leak. 18

    Reassess

    Spiration Isolation Method

    *Treatment should be limited to no more than 3 segments by placing valves in segmental or sub-segmental bronchi in

    the target lung to avoid excessive isolation of tissues from ventilation.

    THE RIGHT PATIENT. THE RIGHT PRODUCT. THE RIGHT OUTCOMES.

  • Copyright © 2015 Spiration, Inc. d/b/a Olympus Respiratory AmericaARTG Identifiers: 188455, 182553, 181950

    WAND Reference Nos: 110725-WAND-6BJ53E, 110505-WAND-6B1TDA, 110509-WAND-6B2O34

    OLYMPUS EUROPA SE & CO. KGWendenstrasse 14-18, 20097 Hamburg, Germany

    KEYMED LTD.KeyMed House, Stock Road, Southend-on-Sea, Essex, SS2 5QH, UK

    OLYMPUS AUSTRALIA PTY. LTD.3 Acacia Place, Notting Hill, VIC 3168 Australia

    OLYMPUS RESPIRATORY AMERICA6675 185th Ave NE, Redmond, Washington 98052 USA

    OLI

    T-0

    356

    0-0

    3 R

    ev A

    A

    E04

    2859

    0

    SPIRATION VALVE SYSTEM TECHNOLOGY

    References

    1. Scuriba. NEJM 2010; 363: 1233-1244. 2. Springmeyer. Thorac Surg Clin 2009; 19(2):247-253.3. Eberhardt. Chest 2012; 142(4): 900-908.4. Brunelli. Ann Thorac Surg 2004; 77:1205-1210.5. Schoenenberger. Arch Surg 1991; Vol 126.6. Varela. Eur J Cardiothoracic Surg 2005; 27:329-33.7. Brunelli. Chest 2006; 130:1150-6. 8. Cerfolio. Ann Thorac Surg 1998; 66: 1726-1730.9. Mahajan. J Thorac Cardiovasc Surg 2013; 145:626-630. 10. Yarmos. Chest 2012; 141(4):1098-105.11. Instructions for use, Humanitarian Use Device (HDE), Spiration Valve System.12. Firlinger. Ann Thorac Surg 2013 Apr; 95(4):1243-9.13. Dooms. Eur Respir J Epub Nov 14, 2013; doi: 10.1183/09031936.00117613. 14. Elstad. Am J Respir Crit Care Med 2012; 185:A1112.15. Ninane. Eur Respir J 2012; 39:1319–1325.16. Schuhmann. Am J Respir Crit Care Med 2015; Jan 30 [Epub ahead of print]; doi: 10.1164/rccm.201407-1205OC.17. Eberhardt. Chest Epub March 29, 2012; doi:10.1378/chest.11-2886.18. Gillespie. Ann Thorac Surg 2011; 91:270-3.

    Deployment Catheter and Loader

    Model Name Article Number

    Catheter Working

    Length

    Bronchoscope

    Channel Inner

    Diameter Valve Size Compatibility

    IBV-C26N N5381300 1020mm 2.6mm or greater For 5, 6, 7 and 9mm Valves

    IBV-C20 N3521830 1140mm 2.0mm or greater For 5, 6 and 7mm Valves

    Spiration Valves

    Model Name Article Number Valve Size Cartridge Color Number Required Per Procedure

    IBV-V5 N3495330 5mm Blue

    Determined by number of target locationsIBV-V6 N3495430 6mm Yellow

    IBV-V7 N3495530 7mm Green

    IBV-V9 N5381200 9mm Grey

    Airway Sizing Kit

    Model Name Article Number Gauge Hole Glass Syringe Number Required Per Procedure

    IBV-VSK N5534500 Sized for appropriate

    valve selection

    500 microliters 1

    Required Ancillary Equipment Needed for Each Procedure

    · Flexible bronchoscope with a working channel inner diameter of 2.0mm or greater

    · Olympus balloon catheter B5-2C

    · Bronchoscopy forceps appropriate for valve removal

    · Standard 10cc sterile syringe with Luer-lock for use in preparing the balloon catheter

    · Sterile saline