spatiotemporal image correlation ... - hmr biophotonics lab · cence speckle microscopy has been...
TRANSCRIPT
Spatiotemporal Image Correlation Spectroscopy (STICS) Theory,Verification, and Application to Protein Velocity Mapping inLiving CHO Cells
Benedict Hebert,* Santiago Costantino,* and Paul W. Wiseman*y
*Department of Physics and yDepartment of Chemistry, McGill University, Montreal, Quebec, Canada
ABSTRACT We introduce a new extension of image correlation spectroscopy (ICS) and image cross-correlation spectroscopy(ICCS) that relies on complete analysis of both the temporal and spatial correlation lags for intensity fluctuations from a laser-scanning microscopy image series. This new approach allows measurement of both diffusion coefficients and velocity vectors(magnitude and direction) for fluorescently labeled membrane proteins in living cells through monitoring of the time evolution ofthe full space-time correlation function. By using filtering in Fourier space to remove frequencies associated with immobilecomponents, we are able tomeasure the protein transport even in the presence of a large fraction (.90%) of immobile species.Wepresent thebackground theory, computer simulations, andanalysis ofmeasurements on fluorescentmicrospheres to demonstrateproof of principle, capabilities, and limitationsof themethod.Wedemonstratemappingof flowvectors formixedsamplescontainingfluorescent microspheres with different emission wavelengths using space time image cross-correlation. We also present resultsfrom two-photon laser-scanningmicroscopy studies ofa-actinin/enhanced green fluorescent protein fusion constructs at the basalmembrane of living CHO cells. Using space-time image correlation spectroscopy (STICS), we are able to measure protein fluxeswith magnitudes of mm/min from retracting lamellar regions and protrusions for adherent cells. We also demonstrate themeasurement of correlated directed flows (magnitudes of mm/min) and diffusion of interacting a5 integrin/enhanced cyanfluorescent protein and a-actinin/enhanced yellow fluorescent protein within living CHO cells. The STICS method permits us togenerate complete transportmaps of proteinswithin subregions of the basalmembrane even if the protein concentration is too highto perform single particle tracking measurements.
INTRODUCTION
Fluorescence fluctuation techniques have been among the
most successful methods for quantitative measurements
inside living cells. They provide key insights into the dy-
namics and interactions of intracellular and transmembrane
proteins. The original fluorescence correlation spectroscopy
(FCS) method is based on temporal autocorrelation analysis
of fluorescence intensity fluctuations collected in time from
a tiny focal volume defined by the microscope focus of the
excitation laser beam within a sample (Elson and Magde,
1974; Magde et al., 1972). The magnitude and time decay of
the fluorescence intensity fluctuations contain information on
the concentration and dynamics of the fluorescent molecules
in the observation volume. Since the introduction of FCS,
there have been many improvements in both the technology
and computer processing power that have made possible the
analysis of increasingly complex systems. For example, the
introduction of confocal optics allowed for single molecule
detection (Rigler et al., 1993), two-photon fluorescence cross-
correlation allows measurement of the dynamics of interact-
ing molecules (Heinze et al., 2000), fluorescence correlations
between two adjacent focal volumes have been used to
determine velocity magnitude and direction in microstruc-
tured channels (Dittrich and Schwille, 2002), and scanning
FCS has been utilized to investigate protein-membrane
interactions (Ruan et al., 2004).
An imaging analog of FCS, image correlation spectros-
copy (ICS), was introduced to examine the distribution and
aggregation of cell membrane components (Petersen et al.,
1993). ICS involves spatial correlation analysis of fluores-
cence fluctuations within an image sampled using a
laser-scanning microscope (LSM). The image pixels are
effectively spatially parallel intensity measurements from
many confocal excitation volumes across the surface imaged.
The image cross-correlation spectroscopy (ICCS) technique
has also been introduced to measure transport properties
(Wiseman et al., 2004, 2000) and co-localization of two
different labeled molecules (Brown and Petersen, 1998). One
of the advantages of the image correlation techniques is that
the specific imaging timescales allow for measurements of
slow transport properties, even in quasistatic systems, as in
the case of transmembrane proteins in cells. A recent ex-
tension of the ICS family is intensity subtraction analysis,
which uses sequential uniform intensity subtraction from
confocal images to extract information about the brightest
population in a system containing a distribution of aggregate
sizes (Rocheleau et al., 2003).
One important problem in biophysics is characterizing the
motion and interactions of membrane proteins, extracellular
matrix components, and intracellular messengers involved in
the regulation of cell migration at the molecular level. As
recent studies have shown, the molecular partners involved
Submitted October 20, 2004, and accepted for publication February 7, 2005.
Address reprint requests to Paul W. Wiseman, Tel.: 514-398-5354; E-mail:
� 2005 by the Biophysical Society
0006-3495/05/05/3601/14 $2.00 doi: 10.1529/biophysj.104.054874
Biophysical Journal Volume 88 May 2005 3601–3614 3601
in cell migration are numerous and their interactions
complex (Lauffenburger and Horwitz, 1996). Cell migration
is a dynamic, integrated process that is coordinated both
spatially and temporally. Although numerous components
are known to interact before, during, and after the formation
of focal adhesions, less is known about the exact timing,
the number of components, and the transport mechanisms
involved in these interactions. New biophysical techniques
have begun to reveal important quantitative aspects of the
molecular mechanisms concerned. For example, fluores-
cence speckle microscopy has been used to investigate actin
polymerization at the front edge of migrating newt lung
epithelial cells (Ponti et al., 2004), and single particle
tracking (SPT) has also revealed movements of adhesion
proteins in apical cell membranes (Sheetz et al., 1989).
ICS analysis can quantify diffusion coefficients and flow
speeds of fluorescently labeled adhesion macromolecules
within the plasma membrane of living cells (Wiseman et al.,
2000). It relies on correlating the amplitude of fluorescence
intensity fluctuations arising from spontaneous variations in
molecular number within the illumination volume defined by
the focal spot of the LSM. These fluctuations arise as particle
clusters move in and out of the volume by diffusion, by flow,
or by a combination of both. However, the ICS technique is
not currently sensitive to the direction in which flowing
fluorescent entities exit the correlation volume; it only fol-
lows the temporal correlation of fluctuations irrespective of
the spatial direction of entry or exit from the focus. As a con-
sequence, ICS can measure only the magnitude of the velocity,
but not its direction.
We introduce a novel technique called space-time image
correlation spectroscopy (STICS), which is an extension of
temporal ICS. In contrast with ICS, STICS does not separate
the spatial fluctuation analysis from the temporal; instead, it
relies on a complete calculation of both the temporal and
all-spatial correlation lags for intensity fluctuations from
an LSM-sampled image series. Monitoring of the two-
dimensional average spatial correlations for every time lag
yields information on both directed flow and diffusion.
Further analysis is needed if a large immobile fraction is
present. Fourier-filtering the zero-frequency components in
time efficiently removes the immobile population contribu-
tion to the spatial correlations. Since fluorescence measure-
ments in crowded biological membranes often involve
a labeled macromolecule population undergoing diffusion
and/or directed motion in the presence of immobile fluo-
rescent species, STICS is ideal for isolating the fluctuations
due the mobile component population.
We have recently reported the first application of STICS
for cellular measurements (Wiseman et al., 2004). In this
study we present a full systematic treatment of the method to
illustrate its capabilities and limitations. We will first present
application of STICS for analyzing computer-simulated
images and fluorescent microsphere samples to demonstrate
the general approach and to determine the detection limits.
By using filtering in Fourier space to remove frequencies
associated with immobile components, we can detect
directional protein transport even in the presence of a large
fraction (.90%) of immobile species. We also present
results from two-photon laser-scanning microscopy STICS
studies of actinin/enhanced green fluorescent protein (EGFP)
fusion constructs at the basal membrane of living CHO cells.
We illustrate cases of measuring protein diffusion, protein
flow in random directions, and net flow within living cells
using STICS. We are able to quantify protein fluxes with
magnitudes of mm/s from retracting lamellar regions and
protrusions for adherent cells plated on fibronectin. We also
demonstrate the measurement of correlated directed flows
(magnitudes of mm/min) of interacting a5 integrin/enhanced
cyan fluorescent protein (ECFP) and a-actinin/enhanced
yellow fluorescent protein (EYFP) within living CHO cells
using two-color STICS.
MATERIALS AND METHODS
Computer simulations
A simulation program of laser-scanning microscopy of point emitters in
a two-dimensional system was written in IDL (RSI, Denver, CO) to test ICS
and its STICS derivative. This program allows the user to set a wide variety
of experimental system and instrumental collection parameters, including
the flow and diffusion characteristic times of several simulated fluorescent
particle populations, their densities, the laser beam characteristics, white-
noise levels, interacting fractions, image size, pixel size, time step, and
number of images. The simulations were run on standard desktop PCs.
Fluorescent microsphere preparations
Fluorescent microspheres with a radius of 0.1 mm were obtained from
Molecular Probes (Eugene, OR). The microspheres contained two different
fluorophores with the following spectral properties: 505/515 nm and
580/605 nm (absorption/emission). For sample preparation, both types of
Fluospheres were diluted by a factor of 5 in doubly distilled water, and drops
of the solution were deposited on coverslips. Particle flow was generated by
convective currents near the edge of the drop. These samples were imaged
using two-photon fluorescence microscopy (details below) to provide
images of two independent particle populations with no spectral overlap.
Using the image analysis freeware program ImageJ (National Institutes
of Health, Bethesda, MD), a separate single-channel image series of fluores-
cent spheres was then added to both channels of the image series of the
independent fluorescent particles in the mixed sample. This effectively
created an artificial interacting fraction between the two independent
fluorescent sphere populations, as the added signal from this third
superimposed image series appeared in both channels. In this way, we
were creating pixels with identical intensity in both channels, thus artificially
creating a cross-channel signal that mimicked two interacting red and green
microspheres. The fraction and the flow direction of this artificial (added)
interacting population were systematically varied before applying STICS
analysis.
Cell culture
CHO-K1 cells (a mutant cell line that is deficient in expression of the a5
integrin; Schreiner et al., 1991) along with CHO-K1 cells transfected with
visible fluorescent protein (VFP) fusions with a-actinin and/or a5 integrin
3602 Hebert et al.
Biophysical Journal 88(5) 3601–3614
were cultured in minimum essential medium supplemented with 10% fetal
bovine serum, and glutamine. For VFP-expressing cells, 0.5 mg/mL
neomycin (G418) was also added to the growth media. Cells were
maintained in a humidified, 8.5% CO2 atmosphere at 37�C. Cells were
lifted with trypsin and plated on 40-mm diameter No. 1.5 coverslips. The
coverslips were precoated with an integrin-activating extracellular matrix
protein (2, 5, or 10 mg/mL fibronectin) or with a non-integrin activating
matrix (200 mg/mL poly-d-lysine). For imaging, the samples were
maintained in CCM1 medium at 37 6 0.2�C in a Bioptechs FCS2 closed
incubation chamber (Bioptechs, Butler, PA), in combination with
a Bioptechs objective heater. Nontransfected CHO cells were used as
control samples to determine autofluorescence background levels. Cell
samples that had been fixed with 4% paraformaldehyde in PBS for 20 min
at room temperature were also prepared for each type of cell line studied.
The fixed cells were imaged in each experiment to provide a control for
any contributions from mechanical vibrations, stage translations, and laser
fluctuations.
Two-photon laser-scanning microscopy
Two-photon microscopy of the fluorescent microspheres was conducted
using an Olympus Fluoview 300CLSM/IX70 inverted microscope (Olym-
pus, Melville, NY), coupled with a Tsunami (model 3960) pulsed femto-
second Ti:sapphire laser (Spectra Physics, Mountain View, CA) pumped
by a Millennia XsJS laser. The microspheres were excited at 800 nm and
point detection was achieved with two external photomultiplier tubes
(Hamamatsu, Bridgewater, NJ). For imaging our microspheres, a 720
DCSPXR excitation dichroic mirror, a 555-dclp emission beam splitter, and
HQ525/50 HQ610/75 emission filters (all from Chroma Technology,
Brattleboro, VT) were employed for light detection. All images were
collected using a PlanApo Olympus 603 (NA 1.40) oil immersion objective
lens. Images were collected with a typical optical zoom setting of 23 cor-
responding to x and y pixel dimensions of 0.23 mm/pixel. Image time-series of
100 frames with a time delay of 0.45s between frames were collected.
Two-photon imaging of cells was conducted using a Biorad RTS2000MP
video-rate-capable two-photon/confocal microscope (Biorad, Hertfordshire,
UK), coupled with a MaiTai pulsed femtosecond Ti:sapphire laser (Spectra
Physics, Mountain View, CA) tunable over from 780 to 920 nm. The
microscope uses a resonant galvanometer mirror to scan horizontally at the
NTSC line-scan rate. Point detection is employed using one or two
photomultiplier tube(s) with fully open confocal pinholes when imaging. For
imaging EGFP in cells, the laser was tuned to a wavelength of 890 nm, and
a 560 DCLPXR dichroic mirror and an HQ528/50 emission filter were
employed for light detection. For imaging cells expressing both ECFP and
EYFP fusion proteins, the laser was tuned to 880 nm, and a D500LP dichroic
mirror and HQ485/22 and HQ560/40 emission filters were used for detection
and separation of the emitted fluorescence. All filters were from Chroma
Technology.
All image time-series were collected using a PlanApo Nikon (Nikon,
Tokyo, Japan) 603 oil immersion objective lens (NA 1.40), which was
mounted in an inverted configuration. Images having dimensions of 480
(height) 3 512 (width) pixels were collected with a typical optical zoom
setting of 23 corresponding to x and y pixel dimensions of 0.118 mm/pixel.
Image series with time delays of 1, 5, or 10 s between sequential frames and
60, 120, or 150 frames in total were collected from single cells. Individual
image frames sampled from the cells were accumulated as averages of 32
video rate scans (i.e., ;1 s per frame).
Image autocorrelation andcross-correlation analysis
Microscope image time-series volumes were viewed, and image subsections
of 162, 322, 642, 1282, or 2562 pixels in size were selected from regions of
the cell and exported for image correlation analysis using a custom
Interactive Data Language (IDL 6.0, RSI) program written for the PC.
Correlation calculations for each image time-series and nonlinear least-
squares fitting of the spatial correlation functions were performed in a
Windows environment on a 1.3-GHz processor PC using programs written
in IDL. Discrete intensity fluctuation autocorrelation functions were
calculated from the image sections as has been previously described
(Wiseman et al., 2000). The equations used for the calculation and fitting of
the normalized intensity fluctuation autocorrelation and cross-correlation
functions (both spatial and temporal) are described below.
THEORY
ICS and ICCS have been introduced in previous contribu-
tions (Petersen et al., 1993; Wiseman and Petersen, 1999).
We will provide a summary of the basic concepts behind
these techniques to introduce the theory necessary for STICS
analysis.
Generalized spatiotemporal correlation function
ICS is based on the correlation of fluorescence intensity
fluctuations measured from an observation area defined by
the diffraction-limited focal spot of the exciting laser beam in
a laser-scanning microscope. The intensity fluctuations in
fluorescence are recorded in an image series as the laser
beam is repeatedly rastered across the sample. Spatial and
temporal correlation is then applied to the image time-series.
We define a generalized spatiotemporal intensity fluctu-
ation correlation function which is a function of spatial lag
variables j and h and of a temporal lag variable t for
detection channels a and b:
rabðj;h; tÞ ¼Ædiaðx; y; tÞdibðx1 j; y1h; t1 tÞæ
ÆiaætÆibæt1t
; (1)
where dia(b)(x,y,t) is the intensity fluctuation in channel a(b)at pixel position (x,y) and time t with diaðbÞðx; y; tÞ ¼iaðbÞðx; y; tÞ � ÆiaðbÞæt, and Æ. . .æ in the denominator represent
spatial ensemble averaging over images at time t and t1t in
the time-series, and the numerator is also an ensemble aver-
age over all pixel fluctuations in pairs of images separated
by a lag-time of t. White-noise sources contribute to the nu-
merator only at zero lag (temporal and spatial), whereas they
will contribute to the average intensities in the denominator.
Correction methods dealing with white-noise and background
correlation have been reported (Wiseman and Petersen,
1999). The zero-lag amplitude value has not been weighed
for all fits in the current work.
Spatial correlation and cross-correlation
ICS has traditionally treated the cases for spatial and
temporal correlations separately. The spatial correlation
function rab(j, h, 0) is defined by evaluating Eq. 1 with zero
time-lag:
Velocity Mapping of Proteins in Cells 3603
Biophysical Journal 88(5) 3601–3614
rabðj;h; 0Þt ¼Ædiaðx; y; tÞdibðx1 j; y1h; tÞæ
ÆiaætÆibæt: (2)
These functions are typically calculated by Fourier methods
and fit to standard Gaussian functions by nonlinear least-
squares methods (Petersen et al., 1993; Wiseman and
Petersen, 1999). The Gaussian fit function for the spatial
correlation of the nth image is given as a two-dimensional
spatial correlation function,
rabðj;h; 0Þn ¼ gabð0; 0; 0Þn exp �j21h
2
v22p ab
( )1 gNabn (3)
(note that in this fitting equation and those that follow, the fit
parameters are highlighted in bold type). Fit parameters are
the zero-lag amplitude gab(0,0,0)n, the two-photon correla-
tion radius v2p ab, which is proportional to the laser-beam
horizontal radius, and the offset gNabn. For an ideal system
of non-interacting particles, the zero-lag amplitude
gab(0,0,0)n is inversely proportional to the mean number of
independent fluorescent particles in the correlation area
defined by the focus of the laser (Petersen et al., 1993). When
a ¼ b ¼ 1 or 2, Eq. 2 defines a spatial autocorrelation
function for one detection channel, and when a ¼ 1 and
b ¼ 2, Eq. 2 defines a spatial cross-correlation function be-
tween two detection channels.
Temporal correlation and cross-correlation
The temporal correlation function is given by evaluating the
generalized correlation function at zero spatial lags:
rabð0; 0; tÞ ¼Ædiaðx; y; tÞdibðx; y; t1 tÞæ
ÆiaætÆibæt1t
: (4)
Its decay will essentially depend on the temporal persistence
of the average spatial correlation of intensity fluctuations
between images in the time-series separated by a lag-time of
t as measured from an ensemble of focal spots (correlation
areas) within a sampled image area. The same equality and
inequality relationships hold for the a and b subscripts in
defining temporal auto- and cross-correlation functions as
was outlined above for the spatial case.
Decay models for correlation functions
The rate and shape of the decay of the correlation functions
will reflect any dynamic process that contributes fluctuations
on the timescale of the measurement. The actual decay
models for fluorescence correlation will depend on both the
underlying dynamics of the fluctuating process and the
geometry of the focal spot (the point-spread function;
Thompson, 1991). We consider four separate functional
forms that are analytical solutions for the generalized
intensity fluctuation correlation function appropriate for
specific cases of two-dimensional transport phenomena as
measured within a membrane system illuminated by a TEM00
laser beam with Gaussian transverse intensity profile. See
below.
Two-dimensional diffusion
rabð0; 0; tÞ ¼ gabð0; 0; 0Þ 11t
td
� ��1
1 gN ab: (5)
Two-dimensional flow
rabð0; 0; tÞ ¼ gabð0; 0; 0Þexp � jvf jtÆv2p abæ
� �2( )
1 gNab: (6)
Two-dimensional diffusion and flow for a single population
rabð0; 0; tÞ ¼ gabð0; 0; 0Þ 11t
td
� ��1
3exp � jvf jtÆv2p abæ
� �2
11t
td
� ��1( )
1 gN ab: (7)
Two-dimensional diffusion and flow for twopopulations (i 5 1, 2)
rabð0; 0; tÞ ¼ gabð0; 0; 0Þ1 11t
td1
� ��1
1 gabð0; 0; 0Þ2 exp � jvf2jtÆv2p abæ
� �2( )
1 gN ab:(8)
The highlighted fit-parameters are the zero-lag amplitude
gab(0,0,0), the offset gNab, the characteristic diffusion decay
time td, and the mean speed of the particles jvfj,
jvf j ¼v2p ab
tf; (9)
where tf is the characteristic flow time. The effective e�2
correlation radius Æv2p abæ is calculated by averaging the
individual v2p ab obtained from fitting Eq. 3 for every image
in the time-series. The best fit characteristic diffusion time
combined with the average correlation radius allows calcu-
lation of the diffusion coefficient:
Dexp ¼Æv2p abæ
2
4td: (10)
It is important to distinguish v2p, the effective two-photon
e�2 correlation radius, and v0, the e�2 beam radius, because
v2p is equal to v0 divided by the square-root of 2. Note that
in Eq. 9, the mean speed jvfj is directionally blind (a velocitymagnitude). ICS is not sensitive to the direction in which
the particles escape the correlation area, because the basic
3604 Hebert et al.
Biophysical Journal 88(5) 3601–3614
analysis does not include non-zero spatial lags with the
temporal lags (see Eq. 4).
Space-time image correlation andcross-correlation spectroscopy (STICS)
The object of the current work is to extend ICS and ICCS to
obtain flow vectors, or essentially to determine the direction
in which the particles are exiting the correlation areas if
directed flux is present. To achieve this, one must combine
the spatial information imbedded in the two-dimensional
spatial correlations with the time-dependent transport
measured by the temporal correlation. For this we define
a discrete approximation to the full space-time correlation
function as
r0abðj;h;DtÞ ¼
1
N�Dt+
N�Dt
t¼1
Ædiaðx;y; tÞdibðx1j;y1h; t1DtÞæÆiaætÆibæt1Dt
;
(11)
where N is the total number of images in the time-series. The
value r9ab represents the average cross-correlation function
for channels a and b, for all pairs of images separated by
a lag-time of Dt. This generalized space-time correlation
function can be considered as a time-series, where the images
are averaged two-dimensional spatial (cross-)correlation
functions, and the time variable is actually the lag-time
(Dt) between all images pairs for which the correlation was
computed.
For an image time-series collected using an LSM, r9aa(j, h,0) is the average spatial autocorrelation function from each
image (Eq. 1 averaged for each image n in the series; see Fig.
1). It will appear as a two-dimensional Gaussian with peak
value at (j¼ 0,h¼ 0). Assuming that the temporal resolution
is sufficiently high for intensity fluctuations to be correlated
between successive images, r9aa(j, h, 1), r9aa(j, h, 2), . . . arealso going to appear as Gaussian spatially distributed
correlations. However, if some particles havemoved between
frames, the correlation function is going to change depending
on the kind ofmicroscopicmotion undergone by the particles.
The simplest case is to imagine the particles as stationary, then
the correlation stays unchanged forDt¼ 0 toN and centered at
(j ¼ 0, h ¼ 0). If we now consider the particles as diffusing,
they will tend to exit the correlation area in a symmetric
fashion, thus broadening the correlation Gaussian in every
direction, analogous to a tracer diffusion experiment. The
peak will stay centered at (j ¼ 0, h ¼ 0) but its value will
decrease hyperbolically (see Eq. 5). Finally if the particles
are flowing uniformly, the spatial correlation Gaussian peak
is going to maintain its original shape as a function of time,
but its peak valuewill be shifted to lag positions (j¼�v3Dt,h ¼ �vy 3 Dt) where vx and vy are the x and y velocities ofthe particles. This is consistent with the observation that for
a flowing population, the temporal autocorrelation function
raa(0, 0, t) decays as a Gaussian. The negative signs in the
expression for j and h arise from the fact that the Gaussian
correlation peak moves in a direction opposite to the flow.
FIGURE 1 Schematic illustration of the
algorithm used to compute the discrete approx-
imation to a generalized spatiotemporal corre-
lation function for a simulated system with flow
and diffusion. The Gaussian autocorrelation
peaks for each image are shown in the left
column (Dt ¼ 0), the cross-correlation for a
lag-time of 1 time-unit in the middle column
(Dt¼ 1), and for the second longest lag-time of
N–1 time-units in the right column (Dt ¼ N–1,
where N is the number of frames in the image
series). The open arrows on the simulation
images represent the direction of the flow. The
averaged Gaussian correlation functions r9aa(j,
h, Dt) are shown at the bottom for Dt¼ 0,1 and
N–1. The separation of the Gaussian correlation
peak due to flow (FG) from the Gaussian
correlation peak arising from the diffusing
population (DG) is clearly seen.
Velocity Mapping of Proteins in Cells 3605
Biophysical Journal 88(5) 3601–3614
This analysis is only valid as long as the particles undergoing
concerted motion stay within the bounds of the analyzed
region. The two-population combined case of a flowing and
diffusing population is illustrated in Fig. 1, where the
diffusion spatial Gaussian correlation peak (DG) broadensand stays centered at (j¼ 0, h¼ 0) and the flowing Gaussian
correlation peak (FG) shifts in a direction opposite to the flowof the particles (as indicated by the open arrows on the
simulated images). In this case the flowing and diffusing
populations were equally represented (in terms of density and
intensity); however, in the cell system in this study the
actively transported subpopulation is usually a small fraction
of the total dynamic species population. This effectively
makes tracking the flowing Gaussian difficult, as it is hard to
resolve near the zero-lags origin due to the diffusing and
immobile populations. A solution to this problem is presented
in the next section.
Immobile population removal
The most general case is a combination of diffusion, flow,
and immobile populations. The challenge is to extract the
velocity direction by following the flow Gaussian correlation
peak, without influence from the correlations of the im-
mobile or slowly diffusing populations (which effectively
remain centered at (0, 0) spatial lags). The immobile
population contribution to r9aa(j, h, Dt) can be removed by
Fourier-filtering in frequency space the DC component for
every pixel trace in time before running the space-time
correlation analysis. The intensity values in channel a for
a given pixel value, say ia(0, 0, t), contains contributions tothe signal of interest from dynamic and immobile component
signals and spurious white-noise sources. The signal from
dynamic components (flow and diffusion) contribute in-
tensity fluctuations that change as a function of time for
a given pixel trace. However, an immobile component only
adds a constant intensity offset to the single pixel intensity
trace through time, so removing the DC frequency
component eliminates this contribution from the correlation
analysis. For a given pixel location (x,y), the corrected
intensities are given by
i0aðx;y; tÞ ¼F
�1
f fFtfiaðx;y; tÞg3H1=Tðf Þg; (12)
where T is the total acquisition time of the image series,
H1/T(f) is the Heavyside function which is 0 for f, 1/T and 1
for f . 1/T, Fð�1Þi denotes the (inverse) Fourier-transform
with respect to variable i, and f is the pixel temporal-
frequency variable.
RESULTS AND DISCUSSION
Simulations
Using our laser-scanning microscopy simulation program
(see Materials and Methods) we ran three different sets of
simulations for point particles, exhibiting i), pure flow with
vx¼�0.12 mm/s and vy¼ 0.08 mm/s; ii), pure diffusion withD¼ 0.01 mm2/s; and iii), a two-population (one flowing, onediffusing) combination with vx ¼ �0.12 mm/s, vy ¼ 0.08
mm/s, and D ¼ 0.01 mm2/s. Fig. 2 A shows the results of
STICS analyses (without any Fourier-filtering) performed on
these simulations. All the populations are set to have the
same total number of particles (density of 50 particles/mm2),
with equal quantum yields. All simulations are 128 3 128
pixels at a resolution of 0.06 mm/pixels, and 500-frames-
long, with a time step of 0.02 s/frame. The beam radius is set
at 0.4 mm. The consequences of flow or diffusion on the
evolution of the two-dimensional contour plot of the cor-
relation function are evident. As expected in simulation i, theGaussian flow correlation peak shifts with a direction oppo-
site to the flow, and in simulation ii, the central Gaussian
correlation peak broadens due to diffusion. In the combined
case of simulation iii, the flow and diffusion the Gaussian
peaks are initially superimposed and they separate after the
flowing population has moved by more than a correlation
radius.
The temporal-decay of the zero spatial lags center of these
contours is r9aa(0, 0, t) (where channel a denotes a single
collection channel for these simulations); i.e., the temporal
autocorrelation function (Eq. 4) This is shown for simu-
lations i–iii (Fig. 2 B). Once again, we can clearly distinguishbetween the various types of transport. The Gaussian profile
of Eq. 6 (squares) is fitted to give a velocity magnitude of
0.152 6 0.004 mm/s and the fit to the hyperbolic profile of
Eq. 5 (triangles) yields a diffusion coefficient of 0.008 6
0.002 mm2/s via Eq. 10. Notice that the velocity magnitude is
close to the input value, but there is no directional infor-
mation to be gathered from the ICS analysis. The analysis
of the combined case (open circles) provides information
on the motion of both populations, with a fit (Eq. 9) velo-
city of 0.144 6 0.004 mm/s, and a diffusion coefficient of
0.009 6 0.002 mm2/s.
Performing STICS on simulation i yields peak positions
for the two-dimensional correlation Gaussian (Fig. 2 C,squares). Fitting for the x- and y-peak displacements
provides an estimate of vx ¼ �0.119 6 0.002 mm/s and
vy ¼ 0.0792 6 0.0005 mm/s. The analysis of the central
diffusion peak in simulation ii provides an estimate of the
accuracy of our results for these simulations. The x and yvelocities should be zero and they are found to be vx ¼0.006 6 0.01 mm/s and vy ¼ �0.007 6 0.006 mm/s (Fig. 2
C, triangles). Moreover, the STICS analysis applied to
simulation iii without modification is not accurate, because
of the perturbation to the velocity analysis due to the presence
of the second Gaussian correlation peak for the diffusing
population, which is centered at (j ¼ 0, h ¼ 0). In the
situation where both flowing and diffusing populations are
present, there are two cases where STICS analysis can still be
performed accurately (see Fig. 3). First, if the diffusing
population is fast compared to the directional flow, then its
3606 Hebert et al.
Biophysical Journal 88(5) 3601–3614
effects on the flow correlation peaks will be short-lived as the
central Gaussian will decay quickly. Second, if the diffusing
population is slow, it can be considered as quasi-immobile
and its intensity contribution is also going to be mostly
eliminated by the Fourier-filtering described previously. To
determine the timescales where these problems would arise,
we performed simulations where we systematically varied
the diffusion coefficient while keeping the velocity of the
particles fixed. These simulations showed that the STICS
analysis is still valid when the characteristic diffusion time is
approximately five times faster or slower than the flow
characteristic time (see Fig. 3). In simulation iii (Fig. 2 A),the diffusion is neither fast nor slow compared to the flow,
and the time-evolving Gaussian correlation peak can be fit to
FIGURE 2 STICS and ICS analysis results for computer-generated
simulations. (A) A contour plot of the STICS correlation functions for
the case of i), flow (vx ¼ �0.12 mm/s and vy ¼ 0.08 mm/s); ii), diffusion(D¼ 0.01 mm2/s); and iii), two-populations’ flow and diffusion (vx ¼�0.12
mm/s, vy ¼ 0.08 mm/s, and D ¼ 0.01 mm2/s). (B) The ICS temporal
autocorrelation functions (Eq. 4) and best fits for the same simulations. (C)
Plots of the peak position of the STICS Gaussian correlation peaks from A as
a function of time. The peak position versus time gives an estimate of the
concerted velocity of the particles in case i, of vx ¼ �0.119 6 0.002 mm/s
and vy ¼ 0.0792 6 0.0005 mm/s and in case iii, of vx ¼ �0.096 6 0.007
mm/s and vy ¼ 0.068 6 0.006 mm/s, respectively. All simulations were
128 3 128 pixels with 500 frames at a density of 50 particles/mm2, using
0.02 s/frame, 0.06 mm/pixel, and an e�2 radius of 0.4 mm.
FIGURE 3 STICS analysis results for computer simulations of two
populations, one flowing (vx ¼ �0.12 mm/s and vy ¼ 0.08 mm/s) and one
diffusing (varying diffusion coefficients). (A) The x velocity as resolved by
STICSwith and without the immobile population filtering. (B) The y velocity
as resolved by STICS with and without the immobile population filtering. On
both graphs, each point and error bar represents the average result of 100
simulations with standard deviation. The shaded regions show the set
velocities in the simulation with an acceptable error of610%. The computer
simulations were 128 3 128 pixels with 100 frames at a density of
100 particles/mm2, using 0.1 s/frame, 0.06 mm/pixel, and an e�2 radius of
0.4 mm.
Velocity Mapping of Proteins in Cells 3607
Biophysical Journal 88(5) 3601–3614
give vx ¼ �0.096 6 0.007 mm/s and vy ¼ 0.068 6 0.006
mm/s (Fig. 2 C, circles) after Fourier-filtering. Notice that thex and y velocities are slightly smaller in this combined case
than for the flow-only case. This is because the remnants of
the diffusing population contribution effectively weight the
flowing Gaussian back toward the zero-lags center when we
try to fit the position of the Gaussian peak. In such a scenario,
if one can assume that the total flow is dominated by the
directional flux (as opposed to separate flows in random
directions), then one can scale the x and y velocities from
STICS analysis according to the total velocity obtained by
ICS analysis to get vx ¼ �0.118 6 0.008 mm/s and vy ¼0.083 6 0.007 mm/s. Note that the temporal ICS analysis
will be sensitive to all flow processes present, which will all
contribute to the decay of the correlation function. For the
case of the adhesion protein transport at the membrane in
cells that we report in this study, the second scenario of faster
diffusion (tD , tf) was usually observed.
If the flowcharacteristic time is slow compared to the image
acquisition rate, then some analysis artifacts are introduced by
the Fourier-filtering, which complicates the case of a slowly
flowing population. If the particles do not move more than
a correlation radius over the time of acquisition of the entire
image-series, then removing the DC offset will spatially
anticorrelate the intensities over a short distance in the
direction of the flow. In otherwords, the centralGaussian peak
is reduced in width, and accompanied by two diametrically
opposed depressions aligned with the flow direction. Never-
theless, these artifacts are of no real consequence in the
determination of the flow direction because a Gaussian can
still effectively be fit to the correlation functions. Moreover,
we can neglect signal fluctuations due to bleaching in flow/
diffusion measurements as long as the characteristic times
associated with these processes are shorter than the bleaching
time (simulation results not shown). This was the case for the
cell measurements reported below.
Fluorescent microspheres
Fluorescent microsphere samples containing a mixture of
flowing spheres emitting at two different wavelengths
(referred to as red and green) were prepared and imaged
by two-photon laser-scanning microscopy (see Materials and
Methods) to generate an image series of two independent
particle populations (referred to as non-interacting). By
adding another image series of flowing microspheres to both
independent channels in the collected time-series, we could
effectively introduce an artificial interacting population. The
direction of flow of the interacting population added by
image processing was chosen by the user and is thus
independent of the direction of flow of the original non-
interacting particle populations. A typical image from this
image-processed time-series is shown in Fig. 4 A. It has anequal density of red and green microspheres, with ;40% of
each population interacting. Such an image time-series was
then analyzed with two-color STICS, yielding directional
flow information for the red and green populations, as well as
for the interacting fraction. We can recover the flow
directions of the non-interacting red and green microsphere
populations to within 8� in the presence of the interacting
population, as compared with the recovered flow directions
of the original image time-series (i.e., analysis performed
without the addition of the interacting population). More-
over, we can find the direction of flow of the interacting
population to within 5�. Fig. 4 B shows the one-to-one
FIGURE 4 (A) A composite image of fluorescent microspheres consisting
of two different fluorescent particle populations (red and green) and an
added interacting particle population (yellow overlap). The composite image
was made by adding an independent image of fluorescent spheres (our
artificial interacting population) to both detection channel images. For each
image time-series constructed, the velocities of the non-interacting popula-
tions remained constant whereas the interacting fraction’s direction of flow
was systematically changed. (B) A plot of the interacting population velocity
magnitudes in x and y measured by STICS in the composite image series
(i.e., with the added population and the red and green microspheres present)
versus the velocity magnitude of the introduced interacting population
measured by STICS in its original image series (i.e., without the red and
green particles present).
3608 Hebert et al.
Biophysical Journal 88(5) 3601–3614
relationship between the recovered velocity magnitudes (in xand y) for the added population as measured by STICS from
the image time-series before the addition and the velocity
magnitude of the added (interacting) population as measured
by STICS in the dual-channel constructed image time-series
in the presence of the non-interacting microsphere popula-
tions. The data are plotted for several experiments in which
the direction of flow of the added interacting population was
different in each case. The magnitudes of the velocities
measured by STICS analysis on the dual-channel image
time-series differ by ,10% from the original values (as
measured separately before the addition of the artificial
interacting population; see Table 1). Note that the density of
the added interacting particle population was five times lower
than the density of each independent fluorescent particle
population.
The STICS analysis for the single non-interacting
populations is influenced by both the fraction that is flowing
independently and the movements of the interacting fraction.
As long as the overall contribution to the image intensity
from the interacting population does not exceed that of the
non-interacting population, STICS can detect the differences
in flow direction between the populations. In the case where
this effect becomes dominant (i.e., equal contribution from
interacting and non-interacting species), one can fit two
Gaussians in r9aa(j, h, t) (a ¼ 1 or a ¼ 2) to extract the two
flow directions. Conversely, the STICS analysis of the
interacting population can also be influenced by the single
non-interacting populations if these flow in the same
direction and random spatial cross-correlations occur. These
effects account for the errors in magnitude and direction of
the measured velocities in the constructed image series as
compared with the velocities measured from the original
image series of the independent microspheres.
Velocity mapping of protein transport inliving cells
We measured the phenomena of directed and nondirected
transport in living cells expressing adhesion protein EGFP
fusion constructs using STICS. We first measured a-actinin/
EGFP constructs expressed in CHO-K1 cells plated on
fibronectin. The protein a-actinin is a cytoplasmic molecule
that binds to the integrins at the membrane and also links the
actin cytoskeleton (Lauffenburger and Horwitz, 1996). We
have previously determined that a-actinin is more mobile in
the peripheral regions of the CHO cells where there is active
lamellar extension, retraction, and membrane ruffling when
the cells are activated on fibronectin (Wiseman et al., 2004).
We focused our measurements on such active peripheral
areas (see Fig. 5).
Fig. 6 shows the ICS and STICS analysis results for
a typical 643 64 pixel region from the cell periphery (Fig. 5
A). As is evident from Fig. 6 B, the temporal autocorrelation
function can be fit very well by Eq. 5, which yields a
diffusion coefficient of (9 6 1) 3 10�4 mm2/s. We show
contour plots of the Gaussian correlation peaks for different
time lags in Fig. 6 A for i), the unmodified image time-series
(without the immobile population removed); and ii), the
filtered image time-series (with the immobile population
removed). As expected, in both cases the correlation peaks
stay centered at zero spatial lags (indicated by the whitecrosshairs). Fitting for the displacement of the Gaussian
yields a very small velocity vSTICS ¼ (1.2 6 0.8) 3 10�3
mm/s (from vx ¼ (�0.9 6 0.8) 3 10�3 and vy ¼ (�0.8 6
0.7) 3 10�3 mm/s, see Fig. 6 C), which can be attributed to
either a real but very slow concerted flux of the proteins, or to
an experimental artifact such as a slow stage drift. These
values are on the order of the precision of our measurements,
which was assessed by applying the STICS analysis to cells
fixed in 4% paraformaldehyde. The corresponding values
vx ¼ (0.4 6 0.3) 3 10�3 and vy ¼ (0.2 6 0.3) 3 10�3 mm/s
establish our detection limits. These results illustrate a mem-
brane region consisting mainly of protein-diffusing and im-
mobile proteins, and show how the random walk is manifest
in both the ICS and STICS analyses.
The same analyses were applied to a different region from
the periphery of another cell (Fig. 5 B) and reveal different
protein transport. Fig. 7 shows our results for a 128 3 128
pixel region in which clusters of a-actinin are clearly
TABLE 1 STICS-measured parameters for a microsphere-image time-series (see Fig. 3)
First population
velocity (mm/min)
Second population
velocity (mm/min)
Added population
velocity (mm/min)
Interacting population
velocity (mm/min)
Set # vx vy vx vy vx vy vx vy
1 0.11 6 0.04 �1.53 6 0.05 0.14 6 0.05 �1.48 6 0.05 �0.01 6 0.05 �3.9 6 0.1 �0.01 6 0.09 �4.2 6 0.2
2 0.27 6 0.05 �1.24 6 0.04 0.30 6 0.03 �1.22 6 0.05 1.39 6 0.04 0.09 6 0.04 1.21 6 0.07 �0.25 6 0.08
3 0.03 6 0.01 �1.0 6 0.1 0.16 6 0.03 �1.23 6 0.05 �1.26 6 0.03 2.27 6 0.03 �1.08 6 0.04 2.13 6 0.04
4 0.03 6 0.06 �1.08 6 0.06 0.11 6 0.09 �0.98 6 0.04 �1.44 6 0.03 0.60 6 0.01 �1.33 6 0.03 0.52 6 0.03
5 0.02 6 0.05 �1.27 6 0.03 0.03 6 0.05 �1.20 6 0.05 �0.31 6 0.03 �1.05 6 0.02 �0.22 6 0.05 �1.02 6 0.04
6 0.12 6 0.02 �0.51 6 0.05 0.17 6 0.03 �0.63 6 0.05 0.21 6 0.02 0.98 6 0.02 0.19 6 0.02 1.00 6 0.02
The column Added population velocity refers to the STICS analysis results before image addition when applied to the single-channel image time-series that is
subsequently added to the dual-channel image time-series to create an interacting population. The column Interacting population velocity refers to the two-
color STICS analysis results after image addition for the co-localized (interacting) population in the composite image. These values should, in theory, be
equal to the values in the column Added population velocity.
Velocity Mapping of Proteins in Cells 3609
Biophysical Journal 88(5) 3601–3614
resolved, and these clusters can be observed to flow in
a directed fashion on what appear to be defined linear tracks.
However, the ICS (Fig. 7 B) and STICS (Fig. 7, A and C)analyses yield very different values for flow: vICS ¼ (136 1)
3 10�3 mm/s and vSTICS ¼ (1.1 6 0.7) 3 10�3 mm/s (from
vx ¼ (�0.676 0.02)3 10�3 and vy ¼ (�0.96 0.8)3 10�3
mm/s). The total velocity value for ICS is approximately 10
times higher than the velocity value measured by STICS.
This is due to the fact that STICS only measures the net
resultant directed component (here the majority, but not all of
the clusters, were observed to be traveling to the left and
down in the image series), whereas ICS measures an average
total flow speed (and a small diffusion coefficient in this
case). Hence the combination of ICS and STICS allows us to
distinguish between directional flow in one direction (see
also Fig. 8), or directional flow in many random directions as
was the case here. Visual tracking of the resolved clusters
shows that the directions are random, with more moving
toward the lower left of the image. In this case, single particle
tracking (SPT) analysis will, in principle, provide more in-
formation about the range of transport (Saxton and Jacobson,
1997). However, it proved difficult to track the clusters with
the fluorescence signal/noise and for the density of expres-
sion of EGFP proteins typical for these transfected cells (SPT
data not shown).
The true advantage of STICS emerges in situations where
no bright clusters are clearly resolved (hence SPT would be
impossible), but concerted flux of protein can be detected
by correlation analysis. Fig. 8 shows analyses results for a
128 3 128 pixel region of a basal membrane of a CHO cell
FIGURE 5 Two-photon LSM images
of the basal membrane of CHO cells
expressing EGFP-labeled a-actinin. The
regions analyzed with ICS and STICS are
shown as open squares and the STICS
analysis results are shown in Figs. 5–7.
(A) A 642 pixel region where the temporal
autocorrelation function is best fit to a
single-population diffusion model (Eq.
5). (B) A 1282 pixel region where the
temporal autocorrelation function is best
fit to a two-population flow/diffusion
model (Eq. 8). (C) A 1282 pixel region where the temporal autocorrelation function is best fit to a two-population flow/diffusion model (Eq. 8). All
images are 512 3 480 pixels at a resolution of 0.118 mm/pixel, and a total of 180, 360, and 120 frames at a resolution of 5, 5, and 15 s/frame for A–C,
respectively.
FIGURE 6 In vivo ICS and STICS
analysis of protein diffusion in a peripheral
basal membrane region of a CHO cell (Fig.
5 A) expressing EGFP-labeled a-actinin.
(A) Contour plots of space-time correlation
functions from STICS analysis (Eq. 11) as
a function of lag-time for i) with and ii)
without the immobile population contribu-
tion present. (B) A plot of the ICS temporal
autocorrelation function and best fit to a
single-population diffusion model (Eq. 5).
The recovered diffusion coefficient was
D¼ (96 1)3 10�4mm2/s. (C) Peak trackingplot of the STICS correlation peaks reveals
that they stay centered at zero spatial lags,
within the precision of our measurement.
3610 Hebert et al.
Biophysical Journal 88(5) 3601–3614
expressing EGFP-labeled a-actinin (Fig. 5 C). Here the ICSanalysis again detects flow and diffusion of two separate
populations (Fig. 8 B) with vICS ¼ (7.7 6 0.8) 3 10�3 mm/s
and a small diffusion coefficient D ¼ (66 1)3 10�5 mm2/s.
The STICS analysis also detects a directional flow (Fig. 8, Aand C) with vx ¼ (1.86 0.3)3 10�3 and vy ¼ (5.56 0.2)3
10�3 mm/s. This example illustrates the importance of remov-
ing the immobile population, since the Gaussian correlation
FIGURE 7 In vivo ICS and STICS
analysis of protein flux in random
directions in a peripheral basal mem-
brane region of a CHO cell (Fig. 5 B)
expressing EGFP-labeled a-actinin. (A)
Contour plots of space-time correlation
functions from STICS analysis as
a function of time for i) with and ii)
without the immobile population con-
tribution present. (B) A plot of the ICS
temporal autocorrelation function and
best fit to a two-population flow/dif-
fusion model (Eq. 8). The recovered
ICS velocity and diffusion were vICS ¼(136 1)3 10�3 mm/s andD¼ (86 1)
3 10�4 mm2/s. (C) Peak tracking plot
of the STICS correlation peaks reveals
that they stay centered at zero spatial
lags, within the precision of our mea-
surement, yielding a very small velocity
of vSTICS ¼ (1.1 6 0.7) 3 10�3 mm/s.
FIGURE 8 In vivo ICS and STICS
analysis of directed protein flow in
a peripheral basal membrane region of
a CHO cell (Fig. 5 C) expressing
EGFP-labeled a-actinin. (A) Contour
plots of space-time correlation func-
tions from STICS analysis as a function
of time for i) with and ii) without theimmobile population contribution pres-
ent. (B) A plot of the ICS temporal
autocorrelation function and best fit to
a two-population flow/diffusion model
(Eq. 8). The recovered velocity was
vICS ¼ (7.7 6 0.8) 3 10�3 mm/s and
a small diffusion coefficient was mea-
sured: D ¼ (6 6 1) 3 10�5 mm2/s. (C)Peak tracking plot of the STICS
correlation peaks (after the immobile
population removal) shows a net dis-
placement of the Gaussian center,
yielding velocities of vx ¼ (1.8 6 0.3)
3 10�3 and vy ¼ (5.5 6 0.2) 3 10�3
mm/s.
Velocity Mapping of Proteins in Cells 3611
Biophysical Journal 88(5) 3601–3614
peak in Fig. 8 A i) is dominated by immobile protein
population spatial correlations and thus stays centered at
zero spatial lags. However, after the immobile population
removal, one can see the Gaussian peak clearly moving away
from the zero lags center toward the bottom left corner in
a directed fashion and the residual central peak from the
diffusion population (Fig. 8 A ii).The formation and disassembly of adhesions as lamellar
protrusions are extended or retracted must result in some
form of transport of adhesion macromolecules into or out of
the transient lamellar extension. Fig. 9 shows the heteroge-
neous spatial distribution of a-actinin in another CHO cell,
close to the edge where lamellar protrusions are clearly
visible in the lower part of the image. During the course of
this 60-frame (300 s) image series, the protrusions retract and
adhesion structures disassemble. STICS analysis shows that
there is a net flux of proteins directed toward the interior of
the cell (Fig. 9, arrows), with the average flow rate of 0.226
0.04 mm/min (see Table 2), in agreement with rates of
filamentous actin retrograde flow (Vallotton et al., 2003).
ICS analyses on the same regions show greater variations in
the total velocity measured (50% spread versus 17% for
STICS). However, given the relatively small size of these
regions (162 or 322 pixels) it was hard to determine fit
parameters to the temporal autocorrelation function with
high precision as the signal/noise ratio depends on the
number of independent fluctuations sampled (Meyer and
Schindler, 1988). STICS is not as sensitive to spatial
sampling as ICS because we are simply tracking a Gaussian
peak, which is far easier than fitting a noisy temporal
autocorrelation curve with a three-parameter hyperbola or
Gaussian. Hence the discrepancy between the two analyses
is due to the better performance of STICS in this small spatial
sampling regime. In the case of ruffling membranes,
however, surface height variations are going to lead to
intensity variations due to the intensity distribution of the
point-spread function and we exclude such regions from
analysis. Such variations are detected as changes in the fit
radius of the spatial correlation functions (Wiseman et al.,
2004). Fluorescent speckle microscopy (FSM) has also been
used to investigate filamentous actin flow at the leading edge
of migrating cells (Vallotton et al., 2003). Tracking very
small amounts of a fluorescent derivative of the monomer
that forms the actin filaments allowed the authors to generate
retrograde actin flow maps with excellent spatial resolution
(1 mm2 regions, or;82 pixels in our case). The advantage of
FSM is that it does not average flow directions because it
follows the trajectories of single speckles, hence it can have
greater precision in regions where there is flow in several
directions. However, FSM requires specialized fluorescent
labeling techniques, whereas STICS can be used without
special labeling (i.e., standard VFP transfected cells) using
standard LSM imaging approaches with approximately the
same spatial resolution.
Another advantage of the STICS method is that it can be
carried out in a cross-correlation scheme, with dual color
labeling of different macromolecular species. If the zero
time-lag cross-correlation function is non-zero, it means that
the proteins are interacting in a common complex (Wiseman
et al., 2004). Furthermore, if we monitor the spatial cross-
correlation peaks in subsequent time lags, then we can
determine if the proteins are flowing or diffusing together.
We imaged CHO cells expressing a5 integrin/EYFP and
a-actinin/ECFP using two-photon microscopy and dual
channel detection, and analyzed the spatiotemporal inten-
sity fluctuations from both channels via two-color spatio-
temporal image cross-correlation spectroscopy (STICCS).
FIGURE 9 Two-photon STICS velocity map image of EGFP/a-actinin
flux in the vicinity of a retracting lamellar extension in a CHO cell plated on
fibronectin. The original image-series reveals that the extended lamellar
protrusions are retracting. Selecting regions of 16 3 16 or 32 3 32 pixels
and performing the STICS analysis reveals a net flow of the a-actinin toward
the cell interior (arrows). The spatial scale is shown as a bar and the velocity
scale is shown as an arrow. The original image-dimensions are 512 3 480
pixels at 0.118 mm/pixel and a total of 120 images at 5 s/frame.
TABLE 2 STICS-measured parameters for image regions of a CHO cell (see Fig. 7)
Region 1 2 3 4 5 6 7 8
vx (mm/min) �0.07 6 0.04 �0.06 6 0.01 0.06 6 0.06 0.01 6 0.05 0.052 6 0.006 0.17 6 0.02 0.24 6 0.05 0.09 6 0.04
vy (mm/min) 0.18 6 0.04 0.14 6 0.01 0.17 6 0.02 0.25 6 0.05 0.25 6 0.02 0.17 6 0.02 0.12 6 0.07 0.22 6 0.04
3612 Hebert et al.
Biophysical Journal 88(5) 3601–3614
The two-photon microscopy images have been corrected for
an 8% bleedthrough of the ECFP signal into the EYFP
detection channel, as had been determined from control mea-
surements on cells expressing just ECFP a-actinin or EYFP
a5 integrin alone. We found that a5 integrin and a-actinin
are actively transported as a complex in some regions of
the cell (see Fig. 10), with an average transport velocity of
0.16 6 0.03 mm/min. Moreover, non-zero temporal cross-
correlation functions were calculated for a5 integrin and
a-actinin outside of visible focal adhesions (for example, in
region 2 of Fig. 10), meaning that both components that are
known to interact in mature adhesions are also present as
co-transported microcomplexes throughout the cell as we
have reported recently (Wiseman et al., 2004). In this study,
however, we have added the directional measurement and
obtained vectors via STICCS analysis for the co-localized
flowing populations.
CONCLUSION
We have shown that STICS, along with ICS, are powerful
tools for the investigation of protein dynamics and inter-
actions, and that STICS provides a way of measuring full
directional velocity vectors in the case of concerted macro-
molecular flow. The applications are not limited by ex-
pression levels in the cell, since this technique does not rely
on optically resolving and tracking individual molecules, and
they do not require any special labeling approaches. Using
STICS with ICS, we can distinguish between diffusion,
protein flow in random directions, directed flux in a single
direction, or a combination of these transport modes. By
employing Fourier-filtering with the STICS analysis, we can
effectively perform these measurements even in cell mem-
brane environments where there are significant levels of
immobile proteins (.90% immobile fraction, simulation
results not shown). Gaining directional information helps in
understanding complex phenomena such as adhesion forma-
tion and disassembly, membrane protein transport, and tran-
sport in polarized cell systems. The application of STICCS
to double-label cross-correlation experiments also promises
new insights into detecting molecular interactions and co-
transport of macromolecules in cells.
We acknowledge Prof. A.R. Horwitz and Dr. C.M. Brown (University of
Virginia) for kindly providing the transfected cell lines used in these studies
and for numerous insightful discussions. We thank Efraim Feinstein and
Jonathan Rossner for the original work on the simulation program. We also
thank Prof. Mark Ellisman (University of California at San Diego) for
allowing P.W.W. to perform some of the two-photon imaging at the Na-
tional Center for Microscopy and Imaging Research.
B.H. acknowledges a post-graduate scholarship from the Natural Sciences
and Engineering Research Council of Canada. P.W.W. acknowledges
funding in support of this work from the Natural Sciences and Engineering
Research Council of Canada, the Canadian Foundation for Innovation, and
the Canadian Institutes of Health Research.
REFERENCES
Brown, C. M., and N. O. Petersen. 1998. An image correlation analysis ofthe distribution of clathrin-associated adaptor protein (AP-2) at theplasma membrane. J. Cell Sci. 111:271–281.
Dittrich, P. S., and P. Schwille. 2002. Spatial two-photon fluorescencecross-correlation spectroscopy for controlling molecular transport inmicrofluidic structures. Anal. Chem. 74:4472–4479.
Elson, E. L., and D. Magde. 1974. Fluorescence correlation spectroscopy. I.Conceptual basis and theory. Biopolymers. 13:1–27.
Heinze, K. G., A. Koltermann, and P. Schwille. 2000. Simultaneoustwo-photon excitation of distinct labels for dual-color fluorescencecross-correlation analysis. Proc. Natl. Acad. Sci. USA. 97:10377–10382.
Lauffenburger, D. A., and A. F. Horwitz. 1996. Cell migration: a physicallyintegrated molecular process. Cell. 84:359–369.
Magde, D., E. Elson, and W. W. Webb. 1972. Thermodynamic fluctuationsin a reacting system: measurement by fluorescence correlation spectros-copy. Phys. Rev. Lett. 29:705–708.
Meyer, T., and H. Schindler. 1988. Particle counting by fluorescencecorrelation spectroscopy. Simultaneous measurement of aggregation anddiffusion of molecules in solutions and in membranes. Biophys. J. 54:983–993.
Petersen, N. O., P. L. Hoddelius, P. W. Wiseman, O. Seger, and K. E.Magnusson. 1993. Quantitation of membrane receptor distributions byimage correlation spectroscopy: concept and application. Biophys. J. 65:1135–1146.
Ponti, A., M. Machacek, S. L. Gupton, C. M. Waterman-Storer, and G.Danuser. 2004. Two distinct actin networks drive the protrusion ofmigrating cells. Science. 305:1782–1786.
FIGURE 10 Two-photon auto- and
cross-correlation STICS velocity map
images of ECFP/a-actinin and EGFP/a5
integrin at the basal membrane of a CHO
cell on a fibronectin substrate. (A) STICS
velocity map for ECFP/a-actinin (channel
1); (B) STICS velocity map for EYFP/a5
integrin (channel 2); and (C) two-color
STICCS velocity map of channel 1 with
channel 2 showing flow vectors for co-
transported ECFP/a-actinin and EGFP/a5
integrin. The spatial scale is shown as a bar
and the velocity scale is shownas an arrow.
The original image dimensions are 5123
480 pixels at 0.118mm/pixel and a total of
120 images at 5 s/frame.
Velocity Mapping of Proteins in Cells 3613
Biophysical Journal 88(5) 3601–3614
Rigler, R., U. Mets, J. Widengren, and P. Kask. 1993. Fluorescencecorrelation spectroscopy with high count rate and low background:analysis of translational diffusion. Eur. Biophys. J. 22:169–175.
Rocheleau, J. V., P. W. Wiseman, and N. O. Petersen. 2003. Isolation ofbright aggregate fluctuations in a multipopulation image correlationspectroscopy system using intensity subtraction. Biophys. J. 84:4011–4022.
Ruan, Q., M. A. Cheng, M. Levi, E. Gratton, and W. W. Mantulin. 2004.Spatial-temporal studies of membrane dynamics: scanning fluorescencecorrelation spectroscopy (SFCS). Biophys. J. 87:1260–1268.
Saxton, M. J., and K. Jacobson. 1997. Single-particle tracking—applica-tions to membrane dynamics. Annu. Rev. Biophys. Biomol. Struct. 26:373–399.
Schreiner, C., M. Fisher, S. Hussein, and R. L. Juliano. 1991. Increasedtumorigenicity of fibronectin receptor deficient Chinese hamster ovarycell variants. Cancer Res. 51:1738–1740.
Sheetz, M. P., S. Turney, H. Qian, and E. L. Elson. 1989. Nanometre-levelanalysis demonstrates that lipid flow does not drive membrane glyco-protein movements. Nature. 340:284–288.
Thompson, N. L. 1991. Fluorescence correlation spectroscopy. In Topics inFluorescence Spectroscopy, Vol. 1: Techniques. J.R. Lakowicz, editor.Plenum Press, New York. 337–378.
Vallotton, P., A. Ponti, C. M. Waterman-Storer, E. D. Salmon, and G.Danuser. 2003. Recovery, visualization, and analysis of actin and tubulinpolymer flow in live cells: a fluorescent speckle microscopy study.Biophys. J. 85:1289–1306.
Wiseman, P. W., C. M. Brown, D. J. Webb, B. Hebert, N. L. Johnson, J. A.Squier, M. H. Ellisman, and A. F. Horwitz. 2004. Spatial mapping ofintegrin interactions and dynamics during cell migration by image cor-relation microscopy. J. Cell Sci. 117:5521–5534.
Wiseman, P. W., and N. O. Petersen. 1999. Image correlation spectroscopy.II. Optimization for ultrasensitive detection of preexisting platelet-derived growth factor-b receptor oligomers on intact cells. Biophys. J.76:963–977.
Wiseman, P. W., J. A. Squier, M. H. Ellisman, and K. R. Wilson. 2000.Two-photon image correlation spectroscopy and image cross-correlationspectroscopy. J. Microsc. Oxford. 200:14–25.
3614 Hebert et al.
Biophysical Journal 88(5) 3601–3614