simultaneous confirmation and quantification using …€¦ · figure 4. spectrum shows a...
TRANSCRIPT
S IM U LTA N EOUS CO N F I RMAT IO N A N d QUA N T I F IC AT IO N US INg X E V O T Q M S: P RO dU C T IO N CO N F I RMAT IO N ( P IC )
Marian Twohig, Gordon Fujimoto, Joanne Mather, and Robert S. Plumb Waters Corporation, Milford, MA, U.S.
INT RODUCT ION
Tandem quadrupole mass spectrometers are used extensively in the
pharmaceutical industry for analyte quantification. This is primarily
performed by multiple reaction monitoring (MRM) as the matrices
are complex and the specificity of MRM gives the best signal-to-
noise ratios.
As well as performing quantification, these instruments are often
used for initial qualitative information, with the instrument operated
in scan mode. This information is used to confirm the identity of the
peak of interest that is being quantified.
In complex matrices, situations can arise where closely-related
compounds, e.g., metabolites or matrix interferences, can give rise
to signals even in MRM mode. This can lead to ambiguity and may
require a second qualitative experiment. Product ion confirmation
provides a means of verifying that the signal from the MRM peak is
from the compound of interest.
With conventional instrumentation, these experiments require
separate full-scan analyses. Many conventional tandem quadrupole
MS instruments are unable to perform MRM and scan experiments
simultaneously, in the timeframe of an LC peak, while maintaining
data quality. The Waters® Xevo™ TQ Mass Spectrometer is equipped
with a novel collision cell design. The collision gas is always on,
allowing both quantification (MRM) and characterization to be
performed simultaneously on the peak as it elutes from the LC or
UPLC® column while maintaining good data quality.
The new ScanWave™ mode of operation allows ions within the
collision cell to be accumulated and then separated according to
their mass-to-charge (m/z) ratio. Synchronizing the release of these
ions with the scanning of the second quadrupole mass analyzer
greatly improves duty cycle, which significantly enhances the signal
intensity of full-scan spectra for both MS and product ions.
Figure 1. Xevo TQ Mass Spectrometer with the ACQUITY UPLC® System.
EX PERIMENTAL
Product ion confirmation on Xevo TQ MS
The Xevo TQ MS can simultaneously acquire a product ion con-
firmation (PIC) scan along with an MRM chromatogram to obtain
additional information about an eluting peak. A PIC scan is enabled
in the MRM method, where a scan is used to collect either:
n MS scan
n Enhanced MS scan using ScanWave mode
n Product ion scan
n Enhanced product ion scan using ScanWave DS mode
In PIC mode, the Xevo TQ MS will switch from MRM to scan after
the apex of an LC peak as long as a minimum intensity threshold
is achieved. The trigger to start will occur after four consecutive
downward scans have been detected. If the minimum intensity
criteria is met, an MS or MS/MS spectrum is acquired using the
final resolving quadrupole (MS2) to perform the scan before
switching back to MRM mode (Figure 2). The threshold ensures
that the PIC scan is of sufficient quality to be beneficial to the user.
The high data collection rate of the Xevo TQ MS is such that the
area of the MRM peak can still be accurately determined, since PIC
is triggered after the peak top is detected and the definition of the
peak itself is not affected. Consequently, quantitative and qualita-
tive data are acquired simultaneously.
Figure 2. Schematic showing Product Ion Confirmation (PIC) switching after the peak top.
Figure 3 shows an example of an MRM chromatogram (3A) obtained
from the quantification of the corticosteroid fluticasone, m/z 501.
Qualitative confirmation of the peak of interest is provided by the
resulting PIC spectrum operated in ScanWave DS mode (3B).
The scan range for the PIC is selected by the software, in this
case m/z 40 to 511.
Figure 3. Chromatogram from the analysis of fluticasone, with MRM 501 > 293, and an example of the ScanWave DS PIC spectrum.
A PIC spectrum using ScanWave DS is displayed in Figure 4A. Here
it is been compared with a PIC spectrum using conventional product
ion scan (DS), 4B, and a combined spectrum (20 scans) from a
ScanWave DS of fluticasone, 4C. The spectral quality is maintained
when a PIC spectrum in ScanWave DS mode (4A) is compared to a
combined ScanWave DS spectrum (4C).
The data show that a four-fold signal enhancement was observed
when ScanWave DS mode (4A) is used to collect the PIC spectrum
compared to a conventional product ion spectrum (4B). This is due
to the more efficient duty cycle that is achieved in ScanWave mode.
This extra sensitivity available with ScanWave mode allows for high
quality spectra to be obtained even at low levels.
Switches here and acquires PIC Scan
Switches back to MRM data acquisition
MRM Trace
B
m/z40 60 80 100 120 140 160 180 200 220 240 260 280 300 320 340 360 380 400 420 440 460 480 500
%
0
100 293
205
109185155
121147
275
265
251
217235
313
501361333 389 481
Time0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80
%
0
Flu_1_9_015d 1: MRM of1 Channel ES+ 501.3 > 293.2 (Fluticasone)
4.09e7
MRM forFluticasone
PIC spectrumfrom MRM peakat Rt = 1.80 min
O
O
OH
CH 3
H
C H 3
F
F
H
C H 3
OS
F
O
C H 3
A100
Figure 4. Spectrum shows a comparison of a PIC spectrum for ScanWave DS, a regular product ion PIC spectrum and a combined spectrum acquired by ScanWave DS for fluticasone m/z 501 (Vertical axis linked).
CONCLUSION
The Xevo TQ MS can be used to perform quantification of fluti-
casone with simultaneous characterization of the MRM peak as it
elutes from the chromatographic system. This eliminates the need
for separate injections when qualitative confirmation of MRM peaks
is required and reduces the total analysis time in these situations.
When used routinely, product ion confirmation increases user
confidence in qualitative results from complex matrixes, and thus
reduces the need for re-analysis.
m/z50 75 100 125 150 175 200 225 250 275 300 325 350 375 400 425 450 475 500
%
0
100
%
0
100
%
0
100Flu_1_9_015d 2 (1.813)
1: Product Ions of 501 ES+501.3 > 293.2 (Fluticasone)
5.26e7
293
205
109185155135
275
251217
313
501361333 389 481
Flu_1_9_014d 2 (1.814)
1: Product Ions of 501 ES+501.3 > 293.2 (Fluticasone)
5.26e7
293
205109 155
275251
313
359333 389
Flu_1_9_017d 961 (1.796)ScanWave DS of 501ES+
5.26e7293
205
109
95155121 185
275
251217
313
501361333 389 481
PIC spectrum ScanWave DS
PIC spectrum DS
Spectrum ScanWave DS
A
B
C
Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com
Waters, UPLC, and ACQUITY UPLC are registered trademarks of Waters Corporation. Xevo, ScanWave, and The Science of What’s Possible are trademarks of Waters Corporation. All other trademarks are the property of their respective owners.
©2008 Waters Corporation. Printed in the U.S.A.October 2008 720002829EN LB-CP