scanning electron microscopy

24
Scanning electron microscopy Tracy Furutani November 13, 2008

Upload: rafal

Post on 06-Jan-2016

70 views

Category:

Documents


4 download

DESCRIPTION

Scanning electron microscopy. Tracy Furutani November 13, 2008. Uses photons of visible light Resolution 1000 nm Color images Lenses of glass Aberration in lenses difficult to correct Sample in air. Uses electrons Resolution 1 nm Color meaningless Lenses of solenoid coils - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Scanning electron microscopy

Scanning electron microscopy

Tracy Furutani

November 13, 2008

Page 2: Scanning electron microscopy

Light microscope vs SEM

• Uses photons of visible light

• Resolution 1000 nm• Color images• Lenses of glass• Aberration in lenses

difficult to correct• Sample in air

• Uses electrons• Resolution 1 nm• Color meaningless• Lenses of solenoid

coils• Stigmator allows

control of beam• Sample in vacuum

Page 3: Scanning electron microscopy

Types of electron microscopy• Transmission electron microscopy -- first invented

(1931), works like light microscope but with electrons, needs thin samples

• Scanning electron microscopy -- have beam of electrons “scan” across a surface to image it

• Reflection electron microscopy -- detects the pattern of scattered electrons

• Scanning transmission electron microscopy -- combining SEM sample interaction with TEM imaging

• Scanning tunneling microscopy -- electrons “tunnel” through vacuum to “feel” surface

Page 4: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.

Page 5: Scanning electron microscopy
Page 6: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.

Page 7: Scanning electron microscopy

Some definitions

• Stigmation -- correcting asymmetries in horizontal versus vertical focus. If the image has astigmatism there will be “streakiness” in the image.

• Collimation -- the creation of parallel path particles (in this case, electrons)

Page 8: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.

Page 9: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.

Page 10: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.

Page 11: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.

Page 12: Scanning electron microscopy
Page 13: Scanning electron microscopy
Page 14: Scanning electron microscopy

Sample preparation

• Coated (sputtered) versus noncoated

• Stubs versus vises

• Carbon tape versus carbon paint

Page 15: Scanning electron microscopy

Some techniques to improve performance

• Spot size -- literally the size of the path of the focused electrons as they strike the sample

• Minimize this to improve focus at high magnification by a) decreasing the working distance or b) increasing the current on the focusing lens

• Trade off: smaller area of coverage, lower beam current means worse contrast

Page 16: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.

Page 17: Scanning electron microscopy

More techniques to improve performance

• Depth of field -- how much relief can be on the sample with it still staying in focus

• Maximizing this allows you to see detail on rough surfaces by a) decreasing the aperture size, b) decreasing the magnification or c) increasing the working distance

• Trade off: lower magnification

Page 18: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.

Page 19: Scanning electron microscopy

Even more techniques to improve performance

• Signal-to-noise ratio -- the contrast between interacting and noninteracting surfaces

• Maximize this to gain more fine detail by a) using a high beam current or b) using a slower scan rate

• Trade off: Much larger spot size

Page 20: Scanning electron microscopy

Energy dispersive X-ray spectroscopy

• Known as EDS, EDX or EDXRF

• Used for elemental analysis

• Electrons from SEM beam excite electrons in atoms in sample

• Atoms in sample emit characteristic energy X-rays as electrons drop back to ground state

Page 21: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.

Page 22: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.

Page 23: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.

Page 24: Scanning electron microscopy

QuickTime™ and aTIFF (LZW) decompressor

are needed to see this picture.