sample scanner laser position sensitive photodiode tip cantilever afm working in ifj atomic force...

30
The AFM detection of ligand-receptor interaction on a surface of living cells Małgorzata Lekka

Upload: brooke-obrien

Post on 21-Jan-2016

224 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

The AFM detection of ligand-receptor interaction on a surface of living cells

Małgorzata Lekka

Page 2: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

Sample

Scanner

Laser

Position Sensitive Photodiode

Tip

Cantilever

AFM working in IFJ

atomic force microscopy

FN

FL

Page 3: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

scanner linearization

spring constant

A–B signal

geometry: = 10º

tip – shape

Hoh et al.

force spectroscopy

Page 4: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

scanner linearization

scanner

photodiodelaser

sin2max

n

s

Page 5: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

without protein (conA)

nominal value

= 7 kHz, k = 0.01 N/m

measured value

= 5.8 kHz, k = 0.007 N/m

resonant frequency of a thermally excited cantilever

with protein (conA)

0.3 mg/ml

= 5.8 kHz, k = 0.007 N/m

1 mg/ml

= 5.4 kHz, k = 0.006 N/m

Sader et al.

!

Page 6: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

TkkF B min

k = 0.03 N/mRT

Fmin 11 pN

Page 7: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

probe

Standard TGT01

Si3N4 + APTES (4%)

+ glutaraldehyde (GL; 2.5%)

Si3N4

+ APTES (4%) + GL (2.5%) + conA (0.3 g/ml)

R = 54 +/- 7 nm R = 275 +/- 10 nm

Page 8: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

probe size protein concentration immobilization procedure

Moy et al.

Grandbois et al.

number of molecules on probe

Single molecular pair

Page 9: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

bond type energy [kJ/mol]

bond length [nm]

force [nN]

covalent 250 (for S–S)

0.20 4.42

noncovalent ionic 20 0.25 0.27

Van der Waals 2 0.35 0.02

hydrogen 740

0.300.30

0.080.45

rough estimation of the bond strength

2

2xkE

+ hydrophobic forces

Page 10: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL
Page 11: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL
Page 12: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

Kim et al.

Distribution of vitronectin receptors on a living MC3T3-E1 cell (murine osteoblastic cell)

Page 13: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

Grandbois et al.

AFM tip functionalized with Helix pomatia

N-acetylgalactosamine in membrane of group A of RBC

mixed red blood cells

Page 14: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

No

of e

vent

s

Force [nN]

Page 15: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

Tees et al.

Page 16: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

Lee et al.

Eb 10 -20 [J] koff[s-1] xb [Å]

AGD – αIIbβ3

RGD – αIIbβ3

- 2.67

- 2.64

47.58

1.53

1.09

1.03

Page 17: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

No

of e

vent

s

Force [nN]

5 s

0.3 s

retraction velocity 3.5 μm/s

ConA-CaY

Page 18: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

How to check what it is measured ?

not functionalized AFM probe

measurements of known ligand-receptor pair

HCV 29 cellssilicon nitride tip

ConA–PC-3 ConA–ASA ConA–CaY

Force [pN] 116 17 790 32 940 39

Page 19: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

CaY – Con A+ 1 mg/ml Con A

non-specific interaction

interaction between ligand – receptor pair

blocking of the binding sites

CaY– Con A free amount of ligand in solution

all or certain number of binding sites can be blocked

Page 20: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

different types of interaction characteristic for cancerous cells

HCV 29 non–malignant transitional epithelial cells of ureter

T24 transitional cell cancer of urine bladder

AFM, contact mode

Page 21: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

SNAPHA-L

ConA

lectinslectins

sialic acidsialic acid

NN-acetylglucosamine-acetylglucosamine

mannose, glucosemannose, glucosecarbohydratescarbohydrates

Page 22: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

binding force Cell lineCell line LectinLectin Binding force [pN]Binding force [pN]

HCV29HCV29 ConAConA 43.343.3 3.43.4

PHA-LPHA-L 59.959.9 77..11

SNASNA 167.2 167.2 5.5 5.5

T24T24 ConAConA 123123.6 .6 1818..11

PHA-LPHA-L 152152..66 88..22

SNASNA 76.276.2 1010.9.9

Page 23: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

verification

50 µg/ml ConA

Page 24: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL
Page 25: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL
Page 26: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

conclusions AFM allows detecting molecular interaction on a surface of living cell

The spatial arrangement of functional carbohydrate groups on cell surface was attributed to the density of all types of the carbohydrate structures (mannose, N-acetylglucosamine, sialic acids).

The maximum range of force distribution (presented in histograms up to 1.2 nN), the size of the adhesion spot (i.e. one single point on the distribution map ~ 0.95 μm2), the number of bonds (2–3 for cancerous cells) suggested that ligands present on a surface of T24 cells formed groups composed of several single carbohydrate chains involved in adhesion process in the lectin recognition.

Page 27: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

Institute of Medical Biochemistry Medical CollegeJagiellonian University

Piotr LaidlerJoanna DulińskaMaryla Łabędź

The Henryk NiewodniczańskiInstitute of Nuclear PhysicsPolish Academy of Sciences

Zbigniew StachuraMałgorzata LekkaJanusz LekkiJan Styczeń

PhD studentsJoanna GrybośKateryna LebedGrażyna Pyka

Atomic Force Microscopy

Page 28: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

Histogram

max

min

)()()(x

xdxrxfxfrC

Autocorrelation function

bin size

large number of data

force between single pair: CaY-ConA

F = 960 +/- 110 pN

Page 29: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL

more complex interactionsF const

single interactionF const

FF

FFF2

2

2

22 2

m = m2

F 2

= m · F + F0

F0 – non-specific force

F · F0

f1(F, F2, F0 )·

f2(F, F2, F0 )

Page 30: Sample Scanner Laser Position Sensitive Photodiode Tip Cantilever AFM working in IFJ atomic force microscopy FNFN FLFL