documentri

89
Radiographic Films Radiographic Films

Upload: ayyappan-subramanian

Post on 21-May-2017

212 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: DocumentRi

Radiographic FilmsRadiographic Films

Page 2: DocumentRi

Radiographic FilmRadiographic Film

Base

Base must be :-

1. Transparent - To allow white light to go through

2. Chemically inert

3. Must not be susceptible to expansion and contraction

4. High tensile strength

5. Flexibility

Page 3: DocumentRi

Radiographic FilmRadiographic Film

Base

Subbing

Subbing

Subbing layer is the adhesive between the emulsion and base

Page 4: DocumentRi

Radiographic FilmRadiographic Film

Base

Subbing

SubbingEmulsion AgBr

Emulsion AgBr

Supercoat

Supercoat

Page 5: DocumentRi

What are the advantages of Double Coated Film?

•Improve contrast

• Reduce the exposure time

Page 6: DocumentRi

Image formationImage formationWhen radiation passes through an object it is differentially absorbed depending upon the materials thickness and any differing densitiesThe portions of radiographic film that receive sufficient amounts of radiation undergo minute changes to produce the latent image (hidden image)

1. The silver halide crystals are partially converted into metallic silver to produce the latent

image2. The affected crystals are the amplified by the

developer, the developer completely converts the affected crystals into black metallic silver

3. The radiograph attains its final appearance by fixation

Page 7: DocumentRi

Film Types

Grain Size Speed Quality Film factor Coarse Fast Poor 10Medium Medium Medium 35Fine Slow Good 90Ultra Fine V.Slow V.Good 200

Film emulsion produced by mixing solutions of nitrate and salt such as potassium bromide.

• The rate and temperature determine the grain structures

1. Rapid mixing at low temperature - Finest grain structure

2. Slow mixing at high temperature - Large grain structure

Page 8: DocumentRi

Processing FilmProcessing Film

Page 9: DocumentRi

Processing Systems

Dev

elop

er

Stop

bath

Fixe

r Running water

Manual System

Page 10: DocumentRi

DevelopmentDevelopment Metallic Silver converted into Black metallic silver

3-5 min at 20OC 68 F

Main ConstituentsMain ConstituentsDeveloping agent metol-hydroquinoneAccelerator keeps solution alkalineRestrainer ensures only exposed silver halides convertedPreservative prevents oxidation by air

Processing Systems

Replenishment Replenishment

Purpose – to ensure that the activity of the developer and the

developing time required remains constant

Guideline – 1. After 1m2 of film has been developed,

about 400 ml of replenisher needs to be added

Page 11: DocumentRi

Stop BathStop Bath3% Acetic acid - neutralises the developer

Processing Systems

Page 12: DocumentRi

FixerFixer• Sodium thiosulphate or ammonium thiosulphate

• Functions:- 1. Removes all unexposed silver grains 2. Hardens the emulsion gelatin

• Clearing time - The time taken for the radiography to loose its milky appearance.

• Fixing time - Twice the clearing time

Processing Systems

Page 13: DocumentRi

Processing Systems

Running waterRunning water• Films should be washed in a tank with constant running water

for at least 20 minutes.

• Insufficient washing the film can caused the yellow fog appears.

Page 14: DocumentRi

SENSITOMETRYSENSITOMETRY

Page 15: DocumentRi

Characteristic CurvesCharacteristic Curves• Increasing exposures applied to

successive areas of a film• After development the densities are

measured• The density is then plotted against the

log of the exposure

Characteristic curve

Sensitometric curve

Hunter & Driffield curve

Page 16: DocumentRi
Page 17: DocumentRi

Characteristic CurvesCharacteristic Curves

Log Relative Exposure

Density (Log)

Toe

Shoulder

Straight line section

Page 18: DocumentRi

Characteristic Curves Information which can be obtained from a

films characteristic curve• The position of the curve on the exposure axis

gives information about the films speed• The gradient of the curve gives information on the

films contrast

Page 19: DocumentRi

Characteristic CurvesCharacteristic Curves

Log Relative Exposure

Density (Log)

Density obtained in a photographic emulsion does not vary linearly with applied exposure

The steeper the slope the greater the contrast

Page 20: DocumentRi

Characteristic Curves Information which can be obtained from a

films characteristic curve• The position of the curve on the exposure axis

gives information about the films speed

Page 21: DocumentRi

Characteristic Curves

Log Relative Exposure

Density

A B C D E

Film A is faster than Film B

Film B faster then C

Page 22: DocumentRi

Characteristic Curves Information which can be obtained from a

films characteristic curve• The position of the curve on the exposure axis

gives information about the films speed• The gradient of the curve gives information on the

films contrast• The position of the straight line portion of the curve

against the density axis will show the density range range within which the film is at its optimal

Page 23: DocumentRi

Changing DensityChanging Density

Log Relative Exposure

DensityDensity achieved 1.5

Density required 2.5

Determine interval between logs

1.8 - 1.3 = 0.5

2.5

1.5

1.3 1.8

Antilog of 0.5 = 3.18

Therefore multiply exposure by 3.18(measured density is lower than the required density)(measured density is lower than the required density)

Original exposure 10 mA mins

New exposure 31.8mA mins

Page 24: DocumentRi

Changing FilmChanging Film

Log Relative Exposure

DensityObtain Logs for Films A and B at required density

Interval between logs = 0.15

1.7 1.85

Antilog of 0.15 = 1.42

Multiply exposure by 1.42

Original exposure 10 mA mins

New exposure 14.2 mA mins

2.5

A B

Page 25: DocumentRi

RADIOGRAPHIC DEFINITIONDEFINITION is the sharpness of DEFINITION is the sharpness of the dividing line between areas of the dividing line between areas of different densitydifferent density

Page 26: DocumentRi

Radiographic Definition

Geometric unsharpness Inherent unsharpness• FFD/SFD too short• OFD too large• Source size too large• Vibration/movement• Poor screen contact

• Coarse grain film• Salt screens• Wavelength too short

Page 27: DocumentRi

Geometry Unsharpness ( Ug)Geometry Unsharpness ( Ug)• Controlled by focal spot, focal to film distance ( FFD), object to film distance (OFD)

Inherent unsharpness (Ui) Inherent unsharpness (Ui) • Controlled by the type of films being used (slow or fast), type of screens and amount of backscatter

Radiographic DefinitionRadiographic Definition

Page 28: DocumentRi

Geometry of Image FormationGeometry of Image Formation

Page 29: DocumentRi

Penumbra Ug)

Ug= F x ofd fod

(Ug = 0.25mm)

Focal spot size, F

ofd

fodffd

Page 30: DocumentRi

Source size as small as possible

Source to object distance as large as

possible

Object to film distance as small as

possible

Penumbra (Ug)

Page 31: DocumentRi

Penumbra = S x OFD FFD - OFD

S = 4mmOFD = 25mmFFD = 275

= 4 x 25 275 - 25

Penumbra = 0.4mm

Penumbra CalculationsPenumbra Calculations

Page 32: DocumentRi

Penumbra CalculationsPenumbra Calculations

= 4 x 25 0.25

+ 25

Min FFD = S x OFD Penumbra (0.25)

S = 4mmOFD = 25mmFFD = 275

+ OFD

Min FFD = 425mm

Page 33: DocumentRi

Inherent Unsharpness

Exposed radiographwith crack like indication

Stray electrons fromexposed crystals

Adjacent crystalsaffected by stray electrons

- -

-

--

--

- -

-

Page 34: DocumentRi

Inherent Unsharpness

Large film grain size increased inherent Unsharpness

Short wavelength increased inherent Unsharpness

Loose film crystal distribution increased inherent Unsharpness

Page 35: DocumentRi

Intensifying ScreensIntensifying Screens

Radiographic film is usually sandwiched between two intensifying screensThere are three main types of intensifying screens

•Lead screens

•Fluorescent screens

•Fluorometallic screens

Page 36: DocumentRi

Film placed between 2 intensifying screensIntensification action achieved by emitting particulate radiation (electrons)

Generally lead of 0.02mm to 0.15mmFront screen shortens exposure time and improves quality by filtering out scatterBack screen acts as a filter only

Lead Intensifying ScreensLead Intensifying Screens

Page 37: DocumentRi

Film placed between 2 intensifying screens

Intensification action achieved by emitting Light radiation (Visible or UV-A)Intensification action twice that of lead

screensNo filtration action achievedSalt used calcium tungstate

Salt Intensifying ScreensSalt Intensifying Screens

Page 38: DocumentRi

Film placed between 2 intensifying screens

Intensification action achieved by emitting light

radiation (Visible or UV-A) and particulate radiation

electrons)

High cost

Front screen acts as a filter and intensifier

Salt used calcium tungstate

Fluorometallic Intensifying ScreensFluorometallic Intensifying Screens

Page 39: DocumentRi

ScatterScatter

• Radiation emitted from any other source than that giving the primary desired rectilinear propagation

• Scatter will lead to poorer contrast and definition and create spurious indications

• It may also cause radiological protection problems

Page 40: DocumentRi

ScatterScatter• Internal scatter originating within the

specimen• Side scatter walls and nearby

objects in the path of the primary beam

• Back scatter materials located behind the film

Page 41: DocumentRi

ScatterScatter• Internal scatter originating within the

specimen

Page 42: DocumentRi

ScatterScatter• Side scatter walls and nearby objects in the path

of the primary beam

Page 43: DocumentRi

ScatterScatter• Back scatter materials located

behind the film

Page 44: DocumentRi

Control of ScatterControl of Scatter

• Collimation• Protection from back scatter• Beam filtration• Blocking• Grids• Increased beam energy

Page 45: DocumentRi

Sensitivity

IQI sensitivity Defect sensitivity

Page 46: DocumentRi

IQI sensitivityIQI sensitivity

The image on a radiograph which is used to determine the quality level

Defect sensitivity Defect sensitivity

Ability to assist the sensitivity and locate a defect on a radiograph(Depend on the defect orientation)(Depend on the defect orientation)

Page 47: DocumentRi

Image Quality IndicatorImage Quality Indicator

Page 48: DocumentRi

Image Quality Indicators

IQI’s / Penetrameters are used to measure radiographic sensitivity and the quality of the radiographic technique used. They are not used to measure the size of defects detected

Standards for IQI’s include:BS 3971BS EN 462DIN 62

Page 49: DocumentRi

7FE12

Step / Hole type IQI Wire type IQI

Image Quality Indicators

Page 50: DocumentRi

Image Quality Indicators Image Quality Indicators

Page 51: DocumentRi

1 Hole visible = 4T

2 Holes visible = T

3 Holes visible = 2T

IQI Sensitivity

Minimum Penetrmeter Thickness 0.5mm(2% of the weld thickness)Minimum Diameter for 1T Hole 0.5mmMinimum Diameter for 2T Hole 1.0mmMinimum Diameter for 4T Hole 2.00mm

Penetrmeter Design

4T diaT dia

2T dia

17 12mm

38mm T

ASME Image Quality Indicators ASME Image Quality Indicators

Page 52: DocumentRi

Wire Type IQI

Step/Hole Type IQI

Image Quality Indicators

Page 53: DocumentRi

Placement of IQIPlacement of IQI

• IQI must be placed on the maximum thickness of

weld

• Thinnest required step or wire must be placed at the

extreme edge of section under test

• IQI must be placed at the source side

• In case of access problem , IQI has to placed on the film side of the object, letter ‘FS’ should be placed beside the IQI.

• IQI material chosen should have similar radiation absorption/transmission properties to the test specimen

Page 54: DocumentRi

Ideally IQI should be placed on the source side

IQI sensitivity is calculated from the following formula

Sensitivity % = Thickness of thinnest step/wire visible x 100Object Thickness

IQI Sensitivity

Page 55: DocumentRi

Image Quality IndicatorsThickness BS 3971 DIN 54 109 BS EN 462-2 BS EN 462-1

(mm) STEP WIRE WIRE (DIN 62) STEP/HOLE WIRE1-6 7-12 13-18 4-10 9-15 15-21 1-7 6-12 10-16 H 1 H 5 H 9 H 13 W 1 W 6 W 10 W 13

0.050 70.063 7 60.08 6 50.10 5 7 7 40.125 6 4 6 6 6 30.150.16 5 3 5 5 5 20.20 4 2 7 4 4 4 10.25 3 1 6 7 3 3 7 30.300.32 2 5 6 2 2 6 6 20.350.40 1 4 5 1 1 5 5 10.50 6 3 4 4 40.600.63 5 2 3 3 30.750.80 4 1 7 7 2 2 6 7 20.901.00 3 6 6 1 1 5 6 11.201.25 2 5 5 4 51.50 1 41.60 4 3 41.80 32.00 6 2 3 2 6 32.50 5 1 2 1 5 23.003.20 4 1 4 14.00 3 35.00 2 26.30 1 1

Page 56: DocumentRi

IQI SensitivityA Radiograph of a 16mm thick butt weld is viewed under the correct conditions, 5 wires visible on the radiograph IQI pack 6-12 Din 62, what is the IQI sensitivity?

Sensitivity = Thickness of thinnest wire visible X 100 Total weld thickness

Sensitivity = 0.4 X 100 16

Sensitivity = 2.5 %

Page 57: DocumentRi

IQI SensitivityUsing the same IQI pack 6-12 Din 62, How many IQI wires must be visible to give an IQI sensitivity of 2 %

Thickness of thinnest wire visible = Sensitivity X Total weld thickness 100

= 2.0 X 16 100

= 0.32 6 wires visible

Page 58: DocumentRi

Radiographic Definition

Definition measured by the use of a type III I.Q.I.Alternative terms given

•Duplex type

•Cerl type B

•EN 462 part 5 EN 4

6 2-5

Page 59: DocumentRi

Exposure ControlExposure Control

Page 60: DocumentRi

Exposure control• For FFD/SFD change

T1 D1 2

T2 D2 2=

T1 = New exposure time

T2 = Original exposure time

D1 = New FFD

D2 = Original FFD

Page 61: DocumentRi

Exposure control• For FFD/SFD change

Example:

Calculate new exposure time for FFD = 600 mm

Original exposure at 500mm was 10 min

T1 =(600) 2

(500) 2 X 10 = 14.4 mins

Page 62: DocumentRi

Exposure calculation

E = M X Time (mA.min)

E = exposure (mA.min)M = Tube current (mA)T = Exposure time (min)

Page 63: DocumentRi

Exposure calculationIn one radiographic operation, an-x-ray machine is set at 5mA and the radiographic film is exposed for a period of 15 minutes. What is the total exposure received by the film?Solution:

Given,

Tube current (M) = 5mA

Exposure time (t) = 15 minutes

Exposure ( E) = M X T

= 5 X 15

= 75 mA.min

Page 64: DocumentRi

Radiographic TechniquesRadiographic Techniques

Page 65: DocumentRi

Radiographic Techniques

Single Wall Single Image (SWSI)- film inside, source outside

Single Wall Single Image (SWSI) panoramic- film outside, source inside (internal exposure)

Double Wall Single Image (DWSI)- film outside, source outside (external exposure)

Double Wall Double Image (DWDI)- film outside, source outside (elliptical exposure)

Page 66: DocumentRi

Single wall single image SWSI

IQI’s should be placed source side

Film

Film

Page 67: DocumentRi

Single wall single image SWSI panoramic

• IQI’s are placed on the film side

• Source inside film outside (single exposure)

Film

Page 68: DocumentRi

Film

Double wall single image DWSI

• IQI’s are placed on the film side• Source outside film outside (multiple exposure)• This technique is intended for pipe diameters

over 100mm

Page 69: DocumentRi

Double wall single image DWSI

Radiograph

Identification

ID MR11

• Unique identificationEN W10

• IQI placingA B• Pitch marks

indicating readable film length

Page 70: DocumentRi

Film

Double wall double image DWDI elliptical exposure

• IQI’s are placed on the source side• Source outside film outside (multiple exposure)• A minimum of two exposures• This technique is intended for pipe diameters

less than 100mm

Page 71: DocumentRi

Double wall double image DWDI

Shot A Radiograph

Identification

ID MR12

• Unique identification EN W10

• IQI placing

1 2• Pitch marks indicating readable film length

4 3

Page 72: DocumentRi

Double wall double image (DWDI) perpendicular exposure

Film

• IQI’s are placed on the source side• Source outside film outside (multiple exposure)• A minimum of three exposures• Source side weld is superimposed on film side weld• This technique is intended for small pipe diameters

Page 73: DocumentRi

Density requirement 2.0 to 3.0Density unacceptable

Density1.2

Density1.2

Density3.0

Density3.0

Sandwich Technique

Page 74: DocumentRi

LEAD SCREENS

FILM AFILM B

FILM A: Fast film - Thicker sectionFILM B: Slow film - Thinner section

FILM AFILM B

Density2.0

Density2.0

Density3.0

Density3.0

Sandwich Technique

Density 2.0 to 3.0 acceptable

Page 75: DocumentRi

Interpretation conditionsInterpretation conditions

Page 76: DocumentRi

Viewing conditionsViewing conditions

• Darkened room

• Clean viewer

• Minimum adequate illumination from the viewer is 3000cd/m2

• Eyesight must be adjusted to the darkened conditions

• Comfortable viewing position and environment

• Avoid fatigue

Page 77: DocumentRi

Radiographic QualityRadiographic Quality Density - relates to the degree of darkness

Contrast - relates to the degree of difference in density between adjacent areas on a radiograph

Definition - relates to the degree of sharpness

Sensitivity - relates to the overall quality of the radiograph

Page 78: DocumentRi

Factors Influencing Sensitivity

Sensitivity

Contrast Definition

Page 79: DocumentRi

ContrastContrastSubject contrastSubject contrast :- Contrast arising from variation in

opacity within an irradiated area

Radiographic contrastRadiographic contrast :- The density difference on a radiography

between two areas- usually subject and

the background (overall)

Film contrastFilm contrast :- The slope of characteristic curve of the film at

specified density. ( Type of film being used, fine

grain or large grain)

Page 80: DocumentRi

Factors Influencing Sensitivity

Density

Sensitivity

Contrast Definition

Film Energy Subject contrast

Processing

Page 81: DocumentRi

Factors Influencing SensitivitySensitivity

Definition

Density Film Energy Object contrast

Processing

Time Temperature Type Strength Agitation

Contrast

Page 82: DocumentRi

Radiographic Contrast

Film Contrast Subject Contrast

Film type Density Processing Scatter Wavelength Screens

Page 83: DocumentRi

Radiographic Contrast

Poor contrast

Poor contrast

High contrast

Page 84: DocumentRi

Radiographic DensityRadiographic Density

Density = Log10

Incident lightTransmitted

light

* Greater contrast is achieved at higher density

Page 85: DocumentRi

Radiographic DensityRadiographic Density

Lack of Density

Under exposure

Developer temp too low

Exhausted developer

Developer too weak

Excessive Density

Over exposure

Excessive development

Developer temp too high

Too strong a solution

Page 86: DocumentRi

Measuring Radiographic DensityMeasuring Radiographic Density Density is measured by a densitometer A densitometer should be calibrated

using a density strip

4.0 3.5 3.0 2.5 2.0 1.5 1.0

Page 87: DocumentRi

Factors Influencing SensitivitySensitivity

Definition

Film speed

Screens Energy Vibration ProcessingGeometry

Contrast

Page 88: DocumentRi

Factors Influencing SensitivitySensitivity

Contrast Definition

Film speed

Screens Energy Vibration Processing

Time Temperature Type Strength Agitation

Geometry

Page 89: DocumentRi

What is a good radiograph? A good radiograph satisfies A good radiograph satisfies the inspection requirementthe inspection requirement