reducing reagent waste through process improvement and

91
Reducing Reagent Waste Through Process Improvement and Preventive Maintenance By Amy Rose Gobel A.B., Princeton University 2012 Submitted to the MIT Sloan School of Management and the Institute for Data, Systems, and Society in partial fulfillment of these requirements for the degrees of Master of Business Administration and Master of Science in Engineering Systems in conjunction with the Leaders for Global Operations Program at the MASSACHUSETTS INSTITUTE OF TECHNOLOGY June 2017 © Amy Rose Gobel, MMXVII. All rights reserved. The author hereby grants to MIT permission to reproduce and to distribute publicly paper and electronic copies of this thesis document in whole or in part in any medium now known or hereafter created. Author ................................................................................................................................. MIT Sloan School of Management and the Institute for Data, Systems, and Society May 12, 2017 Certified by.......................................................................................................................... Nikos Trichakis, Thesis Supervisor Assistant Professor, MIT Sloan School of Management Certified by.......................................................................................................................... Dan Whitney, Thesis Supervisor Senior Research Scientist Emeritus, MIT Leaders for Global Operations Approved by ........................................................................................................................ Maura Herson Director, MBA Program, MIT Sloan School of Management Approved by ........................................................................................................................ John N. Tsitsiklis Clarence J. Lebel Professor of Electrical Engineering, IDSS Graduate Officer

Upload: others

Post on 24-Apr-2022

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Reducing Reagent Waste Through Process Improvement and

Reducing Reagent Waste Through Process

Improvement and Preventive Maintenance

By

Amy Rose Gobel

A.B., Princeton University 2012

Submitted to the MIT Sloan School of Management and the Institute for Data, Systems, and

Society in partial fulfillment of these requirements for the degrees of

Master of Business Administration

and

Master of Science in Engineering Systems

in conjunction with the Leaders for Global Operations Program at the

MASSACHUSETTS INSTITUTE OF TECHNOLOGY

June 2017

© Amy Rose Gobel, MMXVII. All rights reserved.

The author hereby grants to MIT permission to reproduce and to distribute publicly paper and electronic copies of this thesis document in whole or in part in any medium now known

or hereafter created.

Author .................................................................................................................................

MIT Sloan School of Management and the Institute for Data, Systems, and Society May 12, 2017

Certified by .......................................................................................................................... Nikos Trichakis, Thesis Supervisor

Assistant Professor, MIT Sloan School of Management

Certified by .......................................................................................................................... Dan Whitney, Thesis Supervisor

Senior Research Scientist Emeritus, MIT Leaders for Global Operations

Approved by ........................................................................................................................ Maura Herson

Director, MBA Program, MIT Sloan School of Management

Approved by ........................................................................................................................

John N. Tsitsiklis Clarence J. Lebel Professor of Electrical Engineering, IDSS Graduate Officer

Page 2: Reducing Reagent Waste Through Process Improvement and

THIS PAGE IS INTENTIONALLY LEFT BLANK.

Page 3: Reducing Reagent Waste Through Process Improvement and

3

Reducing Reagent Waste Through Process Improvement and Preventive Maintenance

by

Amy Gobel Submitted to the MIT Sloan School of Management and the Institute for Data, Systems, and Society and the on May 12, 2017, in partial fulfillment of the requirements for the degrees of

Master of Science in Engineering Systems and

Master of Business Administration Abstract Quest Diagnostics has committed to reducing operating expenses by $1.3B between 2012 and 2017. A portion of the cost-saving initiative focuses on reagents – expensive liquids that are combined with patient samples to detect compounds of interest. This project aims to reduce reagent waste for high-volume diagnostic tests run on an instrument platform that generates a relatively high amount of reagent waste. Waste, in this context, means any reagent that does not generate unique patient results. Therefore critical components of the quality system, such as quality control and calibration tests, are designated waste even though they are a necessary expenditure. Quality control (QC) samples and mechanical errors accounted for 5.2% and 4.4%, respectively, of all reagent usage prior to the start of the project. Mechanical errors occur when the diagnostic testing platform encounters something unexpected, such as debris or a reading that indicates insufficient sample volume, which interrupts sample processing. The instrument jettisons this test and attempts to repeat the assay. Initial discussions with laboratory representatives revealed differing interpretations of quality control requirements. All sites using the platform of interest were then surveyed to gauge the extent of variation. All sites met quality control requirements but several exceeded them. The most pertinent variations are listed below.

1. Frequency: Several sites ran control samples more often than established in Standard Operating Procedure (SOP) requirements, increasing total QC usage by over 70%.

2. Container size: The choice of container determines the amount of “dead volume”, material that the instrument cannot access and must be discarded. Some sites used containers with 12.8 times the dead volume required in the smallest option.

3. Reuse policy: Some labs reuse containers of quality control materials across multiple batches. Reusing QC material further reduces the amount of dead volume discarded, but using new QC materials eliminates the possibility of evaporation between batches.

An interdisciplinary team of experts tasked with maintaining the SOPs has reviewed these results and will clarify the appropriate SOP interpretation to unify practices across laboratories.

Page 4: Reducing Reagent Waste Through Process Improvement and

4

In order to understand mechanical errors, I observed routine maintenance at four sites and found that business units did not consistently share best practices. Collaborating with vendor representatives and operators, I launched an Autonomous Maintenance (AM) pilot program in order to develop training materials capturing institutional knowledge and to test additional maintenance procedures. The AM activities generated 29 training documents, which were added to a national database of competency training materials. All operators certified to operate the testing platform will be required to review and pass comprehension quizzes on the training materials. As the Marlborough site continues to develop improvements to the maintenance procedures, these changes will be shared with the vendor and incorporated into training documents. Thesis Supervisor: Nikos Trichakis Title: Assistant Professor, MIT Sloan School of Management Thesis Supervisor: Dan Whitney Title: Senior Research Scientist Emeritus, MIT Leaders for Global Operations

Page 5: Reducing Reagent Waste Through Process Improvement and

5

Acknowledgements

It takes a village to raise a thesis. First and foremost, I would like to

thank Toni Kick for mentoring me through all dimensions of my time with Quest

Diagnostics. Thanks go as well to Carlene Wong and Todd Raymond for making

sure I was at home in the Marlborough office, to Craig Vorwald for his guidance

on the nuances of quality control procedures, and to Brian Dunn for the constant

support in learning TPM.

This thesis was only possible because of the support, access, and guidance

I received from the entire team at Marlborough, especially Denis Gallagher, Mike

Hellyar, Margherita Walkowski, Paula Arnone, Cynthia Lam, Phil Chalvire, Dan

Carty, Sean Spooner, and ace operator MaryEllen Marshalsea. What I learned at

Marlborough was magnified by the opportunity to visit the labs in Miramar,

West Hills, and Wood Dale, and I am in debt to those teams for their patience as

I learned from their innovations. I also owe countless thanks to the vendor team

for their support in the AM initiative and for answering my countless questions

about the platform.

On the home front, I thank the LGO team for letting me be one of the

lucky 45 to experience this challenging, enlightening program, and for supporting

me throughout these 24 months. I would also like to thank Jason Jay and the

MIT Sloan Sustainability Initiative for helping me apply a sustainability lens to

the process improvement activities.

Then of course I owe so much to my advisors: to Dan Whitney for asking

the right guiding questions and sharing his experience in the diagnostics industry,

and to Nikos Trichakis for taking a chance on advising an LGO thesis.

I send warm thanks to the LGO and Sloan friends I have gotten to know

and love over the last glorious two years. This community is unparalleled in

richness, and I am honored and humbled every day to be a part of it. Finally

with deep appreciation I thank my parents for their unquestioning, unwavering

support through all my adventures.

Page 6: Reducing Reagent Waste Through Process Improvement and

6

THIS PAGE IS INTENTIONALLY LEFT BLANK.

Page 7: Reducing Reagent Waste Through Process Improvement and

7

Contents 1 Introduction 11

1.1 Problem statement .............................................................................. 11

1.2 Overview of approach, major findings, and recommendations ............ 12

2 Company context 16

2.1 Quest Diagnostics business model ....................................................... 16

2.1.1 Value proposition ................................................................... 16

2.1.2 Regulatory context ................................................................ 18

2.2 Competitive landscape ......................................................................... 19

2.3 Recent performance and history of Invigorate savings ........................ 20

2.4 Summary of relevant organization structure ....................................... 20

2.5 Chapter summary ................................................................................ 21

3 Total Productive Maintenance overview 22

3.1 Brief history of TPM ........................................................................... 22

3.2 Summary of goals, pillars, wastes, and primary metrics ..................... 23

3.3 Drivers of success and failure .............................................................. 27

3.4 Chapter summary ................................................................................ 28

4 Process description 29

4.1 Description of instrument .................................................................... 29

4.1.1 Analytical principle ................................................................ 29

4.1.2 Main mechanical systems ....................................................... 30

4.1.3 Mechanical functions and common errors .............................. 32

4.1.4 Reagent waste ........................................................................ 35

4.2 Maintenance overview ......................................................................... 36

4.2.1 Operator maintenance ........................................................... 36

4.2.2 Vendor maintenance .............................................................. 37

4.3 Stakeholder analysis ............................................................................ 37

4.3.1 Primary groups and their interests ........................................ 37

4.3.2 Potential conflicts .................................................................. 39

4.4 Chapter summary ................................................................................ 39

5 Current state analysis 41

5.1 Process observation ............................................................................. 42

5.1.1 Activities completed .............................................................. 42

5.1.1.1 Observed components of routine operation .............. 42

5.1.1.2 Prepared quality control survey ............................... 44

5.1.2 Preliminary observations ....................................................... 45

5.2 Data collection and evaluation ............................................................ 48

Page 8: Reducing Reagent Waste Through Process Improvement and

8

5.2.1 Sources of data ....................................................................... 48

5.2.2 Preliminary observations ....................................................... 50

5.3 Chapter summary ................................................................................ 53

6 Countermeasures to eliminate quality control waste 54

6.1 CLIA regulations regarding quality control procedures ...................... 55

6.2 Quest Diagnostics quality control requirements .................................. 55

6.3 Results from quality control survey .................................................... 57

6.4 Tension between quality requirements and reagent cost ..................... 61

6.5 Implementing quality control procedure changes ................................ 67

6.5.1 Changes at an example laboratory ........................................ 67

6.5.2 Savings due to changes .......................................................... 67

6.5.3 Continuing to align quality control practices ........................ 68

6.6 Chapter summary ................................................................................ 69

7 Countermeasures to eliminate mechanical error waste 71

7.1 Goals for AM pilot program ................................................................ 71

7.2 CLIA regulations related to operation, maintenance, and equipment

modifications ............................................................................................. 72

7.3 AM team structure and activities ....................................................... 73

7.3.1 Team structure ...................................................................... 73

7.3.2 Autonomous Maintenance activities ...................................... 74

7.4 Outcomes ............................................................................................. 80

7.4.1 Impact on instrument performance at Marlborough .............. 80

7.4.2 Description of training materials and how training materials

were disseminated ........................................................................... 81

7.5 Discussion of implementation approach .............................................. 81

7.5.1 Challenge of improving processes in a lab with high

productivity goals ........................................................................... 81

7.5.2 Strategy for sustaining improvements ................................... 83

7.6 Chapter summary ................................................................................ 83

8 Conclusions 85

8.1 Summary of main findings ................................................................... 85

8.2 Recommendations for Quest Diagnostics ............................................ 86

8.2.1 Specific process/decision-based recommendations ................. 87

8.2.2 Organizational recommendations ........................................... 87

8.2.3 Data collection recommendations .......................................... 89

8.3 Areas of future investigation ............................................................... 90

Page 9: Reducing Reagent Waste Through Process Improvement and

9

List of Figures

2.1 Process map of Quest Diagnostics service ...................................................... 17

4.1 Map of main mechanical systems in instrument ............................................ 30

4.2 Path of sample processing within instrument ................................................ 32

6.1 Rate of quality control use in example laboratory ......................................... 68

7.1 Example One-Point Lesson ............................................................................ 75

7.2 Rate of mechanical errors in Marlborough ..................................................... 80

List of Tables

5.1 Reagent usage March-May ............................................................................. 51

5.2 Error code data for an example laboratory March-May ................................ 52

6.1 Frequency of quality control use .................................................................... 57

6.2 Quality control material acceptance criteria .................................................. 58

6.3 Reuse policy and container size for quality control material ......................... 58

6.4 Reused quality control material storage policies ............................................ 60

6.5 Approach to mechanical errors on quality control samples ............................ 61

Page 10: Reducing Reagent Waste Through Process Improvement and

10

THIS PAGE IS INTENTIONALLY LEFT BLANK.

Page 11: Reducing Reagent Waste Through Process Improvement and

11

Chapter 1 Introduction

This thesis aims to identify and eliminate the sources of unnecessary reagent

consumption on a high-volume diagnostic testing platform at Quest Diagnostics. The

leading sources of avoidable reagent use are excessive quality control and mechanical

errors. Quality control tests are a necessary part of testing procedures, and

regulations require a minimum frequency and type of quality control. Mechanical

errors arise when an abnormality such as debris or an unexpected reading interrupts

normal sample processing. Observations and analysis of instrument performance data

lead to two main findings:

1. Unnecessary quality control use arises when individual labs implement quality

control practices that go above and beyond company or regulatory

requirements; and

2. Mechanical errors arise because the manufacturer’s recommended

maintenance and training procedures do not sufficiently address debris

accumulation in the instrument.

Given these findings, the proposed countermeasures included establishing clearer

standards for quality control practices and increasing the intensity, frequency, and

consistency of maintenance activities beyond manufacturer requirements. In addition

to exploring the impact of these countermeasures, this work highlights the challenges

of maintaining consistency across a large company and promoting a culture of

continuous improvement in an environment with high productivity goals.

1.1 Problem statement Quest Diagnostics generates diagnostic information that allows individuals and

physicians to make informed decisions about healthcare options. The company

performs diagnostic testing in 29 domestic and four international laboratories, each

drawing patient samples from a large region – for example, an area including most of

New England – through a reverse logistics network. Annually the company serves

approximately one third of the US adult population, generating $7.5B in revenue.

Page 12: Reducing Reagent Waste Through Process Improvement and

12

Quest Diagnostics offers a wide range of tests, from the standard array of blood tests

performed for routine physical examinations to esoteric, highly specialized genetic

analyses. Because of its breadth of offerings and relatively low cost, the company

holds substantial market share within the field of diagnostics. However, it operates in

a shifting, highly competitive landscape. Competition from private physician

practices, in-hospital laboratories, and large commercial laboratories threaten to take

over business, and decreasing reimbursement rates from governmental healthcare

payers put downward pressure on revenues.

Within this challenging context, Quest Diagnostics launched an initiative called

Invigorate in 2012 to generate $1.3B in run-rate operating cost savings. One

component of the initiative focuses on reagents, expensive and highly specialized

liquids that are combined with patient samples in order to detect compounds of

interest. This thesis aims to reduce unnecessary reagent consumption on a high-

volume diagnostic testing platform.

1.2 Overview of approach, major findings, and

recommendations This section provides a brief overview of the following chapters, highlighting the

methodology, findings, and recommendations of the project.

Quest Diagnostics operates in a complex competitive landscape within a highly

regulated industry. Chapter 2 provides the relevant financial, industrial, and

regulatory context. Financially, the company has struggled to consistently increase

revenues at the desired rate over the last five years, leading to ever-increasing

scrutiny on operating costs. The Invigorate initiative generates savings from many

sources, including reducing reagent waste – the focus of this thesis – and increasing

employee productivity.

The company’s competition includes similar large-scale commercial laboratories,

laboratories located within hospitals, and laboratories run in private physician

practices. The two latter types of competitors generally offer faster turnaround times

because of the relatively short transportation between sample collection and sample

analysis, as both can be performed within the same building. However, they

generally operate at higher cost and offer a narrower range of test options. As a

result, Quest Diagnostics competes based on the price and the breadth of its

offerings while maintaining a competitive turnaround time for sample results.

Page 13: Reducing Reagent Waste Through Process Improvement and

13

The changes that the company can make in the interest of increased efficiency are

bounded by the highly regulated nature of the diagnostics industry. The Centers for

Medicare & Medicaid Services (CMS) oversees the implementation of the Clinical

Laboratory Improvement Act (CLIA), which sets standards for minimum

maintenance and quality control procedures. CMS has granted enforcement

authority to several non-governmental organizations, such as the College of

American Pathologists (CAP), that inspect and accredit laboratories such as those

operated by Quest Diagnostics.

Each Quest Diagnostics laboratory operates many testing platforms. Each

instrument represents a highly complex piece of mechanical and biomedical

engineering. As such, adequate maintenance plays a key role in ensuring quality test

results and meeting turnaround time expectations. Chapter 3 discusses the history

and main principles of Total Productive Maintenance (TPM), a philosophy of

maintenance developed in manufacturing settings in Japan starting in the early

1970s. TPM emphasizes proactive rather than reactive maintenance and empowers

front-line workers to develop improved maintenance procedures.

Chapter 4 then describes the processes related to the testing platform of focus.

Instrument operators are responsible for running patient and quality control samples;

calibrating the instrument to ensure accuracy over time; performing routine

maintenance; and troubleshooting mechanical errors that arise during any of the

other steps. To understand the instrument function, it is important to first

understand the ten major mechanical subsystems within the instrument, their failure

modes, and the maintenance for each.

The chapter also describes the main stakeholders related to the instrument and by

association to the project. Front-line operators play a key role because they perform

the main functions related to the instrument. Laboratory management makes key

decisions related to how operators spend their time, given performance goals

established by regional management. The operations occur within the context of

manufacturer and company-specific standards. Specifically, the instrument

manufacturer establishes a set of minimum operating requirements. Then a national

Best Practices Team (BPT) within Quest Diagnostics uses this baseline to establish

Standard Operating Procedures (SOPs) for the company.

While the main groups related to this project are united around the goal of providing

accurate patient results, their unique perspectives and constraints lead the main

organizational challenges of this project, specifically the tensions between

Page 14: Reducing Reagent Waste Through Process Improvement and

14

productivity goals and laboratory improvement initiatives, benefits of more

conservative quality control practices and costs of excessive reagent use on quality

control testing and manufacturer intentions and their capabilities.

Having established the background of the company, methodology, and process,

Chapter 5 discusses the current state of the diagnostic testing platform of interest.

The current state analysis involved direct observation of four laboratories using the

platform and data related to instrument usage and performance. The main findings

were as follows:

(1) Excessive quality control and mechanical errors – that is, interruptions to

sampling processing – were the largest sources of avoidable reagent usage;

(2) Individual laboratories implemented quality control practices that exceeded

Quest Diagnostics operating procedures and regulatory requirements;

(3) Mechanical errors on the instruments were a major source of irritation for

operators;

(4) The most common mechanical errors may be addressed through increasing

maintenance beyond manufacturer requirements; and

(5) Laboratories do not consistently share best practices related to maintenance or

process improvements.

Chapter 6 reviews this project’s approach to the first two observations above. A

survey of all Quest Diagnostics laboratories using the target platform established

that quality control practices differ in frequency, acceptance criteria, reuse policy,

storage approach, container size, and approach to mechanical errors. Individual

laboratories established their own balance between the benefits and the costs of

quality control measures beyond those required by company SOPs and regulatory

requirements, leading to substantial differences in how much quality control material

and reagents laboratories used in their quality control procedures. The chapter

discusses the pros and cons of the various laboratory policies.

Ultimately, the BPT should establish quality control policies to a level of specificity

that encompasses the varying parameters observed across laboratories. Determining

the reliability of quality control policies, such as reusing material between batches,

may require further investigation. The potential reagent cost savings from adopting

a less conservative policy set a reasonable bound on the budget for such testing.

Just as Chapter 6 addresses the first leading source of waste, quality control,

Chapter 7 examines the second, mechanical errors. Preliminary data suggests that

Page 15: Reducing Reagent Waste Through Process Improvement and

15

maintenance beyond the manufacturer’s recommendations could reduce the rate of

mechanical errors. As discussed in Chapter 3, the maintenance philosophy of TPM

provides an established structure within which a company may improve maintenance

procedures. Because of the proven success of TPM methods and the accessibility of

an internal expert on the approach, this thesis implements a component of TPM

called Autonomous Maintenance (AM) on the platform within a Quest Diagnostics

laboratory.

An AM team consisting of the author, an instrument operator, and a rotating group

of representatives from the manufacturer, creates two types of outputs to address the

mechanical errors observed. First, we develop a series of training documents called

One-Point Lessons (OPLs) to capture and easily disseminate best practices related to

instrument operation, troubleshooting, and maintenance. Second, we establish a

process for testing supplementary maintenance procedures. In order to fully realize

the benefits from this work, Quest Diagnostics should include the OPLs as part of

operator competency training and continue to create space for ongoing testing of

additional maintenance procedures.

Chapter 8 consolidates the observations and leading recommendations. In addition

to the process-specific recommendations discussed in Chapters 6 and 7, Quest

Diagnostics should consider data collection and organizational changes that will

encourage greater consistency across laboratories and drive a culture of continuous

improvement. To facilitate the ongoing process improvement, Quest Diagnostics

should collaborate with the instrument manufacturer to develop consistent,

streamlined methods of tracking instrument performance and correlating individual

mechanical errors with reagent waste.

At an organizational scale, the company should foster innovation by providing more

incentives for process improvement that increase safety, promote efficiency, or reduce

costs, which operators can submit through the BPT. As laboratories continue to

innovate, the company should continue to perform this type of comparative analysis

to identify other variations that inevitably arise between labs. Finally, the BPT

should update SOPs and training materials through their standard review procedure

to incorporate the process improvements in a standardized way to ensure compliance

across the company. This type of structural change can apply more broadly to any

national organization with processes repeated in multiple locations within the

company.

Page 16: Reducing Reagent Waste Through Process Improvement and

16

Chapter 2 Company Context This chapter discusses the relevant company context to ground the subsequent

chapters in this thesis. Quest Diagnostics provides clinical diagnostic test results for

approximately one third of the adult population of the United States every year

through an extensive network of regional laboratories, dedicated transportation, and

patient sample collection centers. Despite their large market share, they face

significant competition and an industry context of decreasing reimbursement rates

for testing.

This challenging financial environment led Quest Diagnostics to focus on operational

excellence as a core part of their business strategy starting in 2012. The strategy

includes ambitious cost-saving targets achieved in part by reducing laboratory

headcount and eliminating waste. One component of the cost-savings initiative

targeted reagent waste, thus motivating this thesis. In addition, the company

context of increasing productivity requirements for laboratory personnel created the

defining organizational constraints within which this thesis operated.

2.1 Quest Diagnostics business model

2.1.1 Value proposition

Quest Diagnostics provides diagnostic information services that allow individuals and

physicians to make informed decisions about healthcare. Examples of this type of

information include analyses of cholesterol levels, indications of infectious diseases,

evidence of illegal drug usage, genetic data about predisposition to diseases, and

many others.

The process through which a patient interacts with Quest Diagnostics generally

proceeds is depicted in Figure 2.1 and described here.

1. A doctor needs information about certain aspects of a patient’s health and

orders one or several tests.

Page 17: Reducing Reagent Waste Through Process Improvement and

17

2. The patient then provides a type of sample as specified by the test

requirements, such as blood, urine, saliva, or exhaled air. The sample is

collected in a dedicated testing area in a doctor’s office or another commercial

location, including spaces that Quest Diagnostics owns and operates.

3. Quest Diagnostics operates a reverse logistics network that collects patient

samples from these locations daily and transports them to the nearest regional

laboratories that can perform the required testing.

4. Once in the lab, the sample is logged and transported to the relevant testing

area.

5. Instrument operators then process the sample as necessary and load the

sample into the appropriate instrument, as specified by the prescribed test.

6. The instrument generates diagnostic information. The operator reviews the

results to confirm that the results meet Quest Diagnostics quality

requirements. If there are no unusual patterns, trends or distributions in

patient results, the operator then releases the results to the prescribing

doctor. Otherwise, the operator repeats testing until acceptable results are

achieved.

7. Meanwhile, the laboratory archives the patient sample for a time period

prescribed by Quest Diagnostics policy in the event that additional testing

becomes necessary.

8. Finally, the doctor uses this information to determine the appropriate course

of action for the patient.

Figure 2.1: Process map of Quest Diagnostics service

At doctor’s office and/or designated sample collection

location

Within Quest Diagnostics regional laboratory

1. Doctor orders test(s) for a patient

2. Patient delivers sample at doctor’s office or other designated location

3. Quest Diagnostics collects sample and transports to regional laboratory

4. Laboratory employee logs sample and transports it to relevant testing area

5. Instrument operator runs the sample on the designated testing platform

6. Instrument operator reviews test results for quality then releases results

7. Laboratory employee archives patient sample

8. Doctor uses laboratory results to determine course of treatment for patient

Page 18: Reducing Reagent Waste Through Process Improvement and

18

Quest Diagnostics operates primarily in the United States, with 29 domestic

laboratories, including Puerto Rico. In addition, the company operates four

international laboratories in Mexico, India, and Brazil. This thesis will consider only

the domestic laboratories because of the shared regulatory environment and greater

similarities in operations.

This analysis focuses on the stages during which the operator interacts with the

instrument and patient sample to generate diagnostic results – steps 5 and 6 in

Figure 2.1. While other components of the Quest Diagnostic process involve

opportunities to reduce waste, they lie beyond the scope of this analysis.

Each laboratory operates a variety of testing platforms based on the capabilities

required within that region. While the mix of platforms may vary between labs, any

given type of test will always be run on the same testing platform. A single platform

may have the ability to run several different tests.

The instruments themselves represent significant investments in terms of the

immediate cost of the instrument, ongoing service/maintenance costs, and the

process of introducing a new piece of equipment into the laboratory. In 2015,

laboratory equipment accounted for 44% of the company’s property, plant, and

equipment assets [1].

Depending on the platform, Quest may choose to buy the equipment or lease it from

the vendor. In either case, the company may also decide to take on a service contract

with the equipment manufacturer. As a result, the instrument manufacturer usually

maintains close ties with a laboratory that acquires the manufacturer’s instrument

by providing routine maintenance, validated instrument modifications, and/or

ongoing operator training. The relationship with the instrument manufacturer

proved critical to this analysis because of the technical expertise they could provide

and the regulatory requirements that the manufacturer approve any process changes.

2.1.2 Regulatory context

The Clinical Laboratory Improvement Act (CLIA) of 1967 and the subsequent

Clinical Laboratory Improvement Amendments of 1988 established the primary rules

within which Quest Diagnostics operates in the United States. This law established

laboratory-specific guidelines (42 CFR 493), designated the Centers for Medicare &

Medicaid Services (CMS) as the regulatory agency legally and financially responsible

Page 19: Reducing Reagent Waste Through Process Improvement and

19

for enforcement, and established Medicare and Medicaid compensation guidelines for

tests [2]. CMS has granted deemed status to several non-governmental agencies,

which allows the organizations to perform inspections and accredit laboratories.

Quest Diagnostics laboratories receive accreditation from the College of American

Pathologists (CAP), an association of peer pathologists and laboratory employees.

The FDA regulates manufacturers of In Vitro Diagnostic products (IVDs). The

FDA has asserted that it has jurisdiction over laboratory developed tests (LDTs),

but presently has not regulated LDTs since they are regulated by CMS under

CLIA 1988. There are various legislative proposals that could allow the FDA to

regulate certain aspects of LDT development under three of the FDA Quality

System Regulations (QSRs), whereas IVD and other medical device

manufacturers are subject to approximately 25 QSRs. However, such legislation

and its regulatory implementation are at least six or more years in the

future. Quest Diagnostics does not do animal research testing and does not have

any laboratories that are subject to 21 CFR Part 58 regulations pertaining to

animal and basic pharmacology and toxicology research testing.

Chapters 6 and 7 will discuss the specifics of the regulations as they apply to this

project.

2.2 Competitive landscape Quest Diagnostics operates in a highly competitive landscape in an era of significant

downward pressure on prices. The company faces competition from other commercial

laboratories, such as the company’s largest competitor, Laboratory Corporation of

America; laboratories that operate within hospitals; and private physician

laboratories. The physician-owned and hospital-based laboratories can deliver results

rapidly because of the minimal amount of sample transportation required, so Quest

faces substantial pressure to test patient samples rapidly to compensate for the

additional time spent in transit. Quest provides the additional value of offering a

larger array of testing options than most hospitals or independent labs can

accommodate and providing results at a lower cost because of the scale of their

operations.

Quest also faces decreasing prices. Part of the Affordable Care Act of 2010 involved

decreasing prices that CMS paid for diagnostic services by 1.75%, decreasing the

profit margins that Quest could generate from Medicare and Medicaid patients [1].

Page 20: Reducing Reagent Waste Through Process Improvement and

20

Private insurance companies have followed suit by decreasing reimbursement rates,

further affecting testing revenues.

2.3 Recent performance and history of

Invigorate savings In this highly competitive context, Quest Diagnostics has struggled to maintain

consistent growth of revenue, income, and margin since 2010. To address the

financial turbulence in May 2012, the company brought on a new CEO, Steve

Rusckowski, with the goal of establishing the company on firmer footing.

Mr. Rusckowski addressed this challenge by establishing a five-point strategy of

restoring growth, reducing operating costs, removing unnecessary layers of

management, divesting of operations outside the core diagnostics information

business, and increasing shareholder dividends. The effort to reduce operating costs,

dubbed the Invigorate initiative, originally aimed to decrease annual operating costs

by $500 million per year by 2014 from a 2011 baseline. After reaching a savings rate

of $700 million in 2014, the goal further increased to a total of $1.3 billion in annual

savings by the end of 2017.

One component of Invigorate focused on reducing waste of reagents – the expensive

liquids combined with patient samples in testing procedures. The reagent savings

target motivated this thesis.

Another important source of savings comes from increasing productivity. Generally

this means increasing laboratory throughput while maintaining or decreasing

employee headcount. While productivity considerations play a role in this thesis,

they are nominally out of scope.

2.4 Summary of relevant organizational

structure To understand the rest of this analysis, we must first review decision-making

processes at a site level and a national level.

Quest laboratories operate in accordance with national Standard Operating

Procedures (SOPs). The groups that establish SOPs, the Best Practices Teams

(BPTs), consist of medical, regulatory, and laboratory experts from across the

company. Each BPT owns the SOPs for a set of clinically related testing procedures.

Page 21: Reducing Reagent Waste Through Process Improvement and

21

Individual laboratories then have responsibility for implementing the SOPs.

Laboratory Medical Directors may provide guidance about SOP interpretation at a

regional level. As discussed in Chapter 6, by delegating some of the burden of SOP

interpretation, the BPTs allow room for individual laboratories to develop local

practices that exceed Quest Diagnostics standards and regulatory requirements,

leading to unnecessary reagent usage.

Within individual labs, instrument operators run the diagnostic tests, review the

instrument output to confirm that results meet company quality standards, and

release results to the relevant physicians. Within a clinically related area of the lab,

all operators report to a shift supervisor. The supervisors for one or several areas

then report to a laboratory manager. Finally, the laboratory managers report to a

laboratory director. The laboratory director controls decisions related to general

policies or lab targets such as headcount or overtime allocation, but many day-to-

day decisions about the specifics of lab operations often happen between operators

and supervisors. As discussed in Chapter 7, this localized decision-making led

individual labs to create innovative process improvements or develop localized

expertise in some part of instrument operations, but that knowledge was not

consistently shared beyond the lab. Subsequent chapters discuss the efforts of this

thesis to encourage labs to share that knowledge.

2.5 Chapter summary Quest Diagnostics operates in a highly competitive industry in an era of decreasing

prices, and as a result, the company has increasingly focused on decreasing operating

costs as a means of increasing net income. The Invigorate cost-saving initiative

motivates this thesis by setting targets for reducing unnecessary reagent use in

diagnostic testing. Any changes to current operations must be performed within the

bounds of relevant regulations, primarily the Clinical Laboratory Improvement Act,

and the changes must be approved by the company’s internal Best Practices Team

to ensure consistent approaches.

Quest Diagnostics uses highly complex biomedical equipment to generate patient

results, and maintaining these instruments is a critical component of generating

consumer value and abiding by regulatory requirements. The next chapter provides a

brief overview of maintenance philosophies including Total Productive Maintenance,

an approach applied in this thesis to improve instrument performance.

Page 22: Reducing Reagent Waste Through Process Improvement and

22

Chapter 3 Total Productive Maintenance overview This project aims to reduce waste on a diagnostic testing platform used in many

Quest Diagnostics laboratories. Routine maintenance plays a critical role in

instrument performance, especially in preventing mechanical errors that lead to

reagent waste. This project specifically employed Autonomous Maintenance, a

defining element of the maintenance philosophy Total Productive Maintenance

(TPM). To illuminate why TPM was selected as a countermeasure, this chapter

discusses the history of TPM, its main principles, and the drivers of its success and

failure.

3.1 Brief history of TPM The role of maintenance in manufacturing has evolved significantly over the 20th

century. This section briefly discusses the main predecessors to TPM.

• Breakdown Maintenance: Prior to the 1950s, most manufacturing organizations

performed maintenance only when operating conditions had noticeably

deteriorated [3]. This reactive approach led to expensive, unplanned interruptions

to operations.

• Preventive Maintenance: Introduced in 1951, Preventive Maintenance involves

monitoring the health of the equipment and performing maintenance functions

like lubrication and minor repairs at set times throughout the day, instead of

waiting to perform maintenance once a breakdown has occurred [3].

• Predictive Maintenance: Like Preventive Maintenance, Predictive Maintenance

attempts to address incipient errors before they lead to instrument breakdowns.

Unlike the time-based schedule of Preventive Maintenance, it instead uses

Page 23: Reducing Reagent Waste Through Process Improvement and

23

physical indicators such as instrument temperature, vibration, or component

wear to determine when to perform maintenance.

• Maintenance Prevention: This approach, introduced in the 1960s, involves

designing or modifying instruments to reduce maintenance steps or the time

required to perform those steps [3].

• Productive Maintenance: This philosophy combines Predictive Maintenance and

Maintenance Prevention with a particular emphasis on maximizing the

productivity of equipment and the lifetime cost of maintenance.

• Total Productive Maintenance: In 1971, the Toyota Motor Company supplier

Nippondenso Company developed Total Productive Maintenance [4]. This

philosophy expands on Productive Maintenance by including the entire workforce

in support of improved instrument productivity. Since that point, TPM has been

treated as a necessary precondition for maintenance excellence. For example, the

Japan Institute for Plant Maintenance began recognizing plants with excellent

maintenance practices in 1964, but since 1971 only plants that had fully

implemented TPM have received the prize [4].

The components of TPM evolved since its foundation. In the late 1970s, the Chuo

Spring Company began delegating the responsibility of process improvement to small

groups within. Building off this success, the Tokai Rubber Company developed the

seven-step Autonomous Maintenance process that now constitutes a foundational

component of TPM [5].

3.2 Summary of goals, pillars, wastes, and

primary metrics The key elements that define a TPM program are as follows:

(1) Focus on maximizing Overall Equipment Effectiveness;

(2) Develop a company-wide preventive maintenance process;

(3) Engage designers, operators, and maintenance personnel in improving

equipment;

(4) Involve all levels of company up to top managements; and

(5) Implement preventive maintenance through autonomous small-group activities

[6].

These components combine to accomplish the following goals:

Page 24: Reducing Reagent Waste Through Process Improvement and

24

(1) Achieve zero defects;

(2) Run instruments at 100% of available capacity;

(3) Prevent the “six big losses” of instrument inefficiency;

(4) Give operators ownership of equipment maintenance;

(5) Improve operators’ abilities;

(6) Design instruments that minimize maintenance, breakdowns, and the time

between installing and achieving full functionality.

Zero defects: Realistically, no plant can achieve zero defects. However, by striving for

100% quality, TPM encourages operators to identify minor problems before they

escalate into larger, more impactful problems.

100% available capacity: The primary metric for instrument performance is Overall

Equipment Effectiveness (OEE), which is defined as follows:

OEE = Availability x Performance x Quality.

In turn,

Availability = (time instrument was running) / (time available for instrument

to run),

where the difference between the numerator and the denominator arises from

unplanned breakdowns that reduce the total amount of time the instrument can

function during a day. This number could also be reduced when critical parts are not

delivered on time, forcing the instrument to remain idle.

Then, given the total amount of time available to run,

Performance = (actual # jobs) / (ideal # jobs).

The ideal number of jobs is the theoretical instrument throughput in the allocated

time, given the theoretical minimum cycle time. Finally, given the number of jobs

performed,

Quality = (# successful jobs) / (# jobs).

A world-class manufacturing facility will achieve Availability ¿ 90%, Performance ¿

95%, and Quality ¿ 99%, for an OEE ¿ 85%. Plants that receive recognition from

JIPM generally have an OEE exceeding 90%, demonstrating the difficulty of

achieving high values for this metric [4].

Page 25: Reducing Reagent Waste Through Process Improvement and

25

Chapter 5 discusses the difficulty of establishing a consistent method of determining

OEE for the target testing platform. Ultimately, the project relied on weekly quality

data and information about the frequency of mechanical errors to track instrument

performance over time.

Six big losses: TPM identifies six categories of problem that affect OEE. These “six

big losses” are as follows:

1. Breakdown: When a mechanical failure has forced an instrument to stop

processing, TPM identifies the lost time as a breakdown loss.

2. Setup and adjustment: This type of loss includes the time spent changing

instrument parameters, swapping out dies, or otherwise preparing the

instrument for function when it is not in a breakdown state.

3. Idling and minor stoppage: Small errors, such as a material jam or blocked

sensor may interrupt a process. If the operator can resolve the issue quickly,

the loss is defined as a minor stop instead of a breakdown, which completely

halts production.

4. Reduced speed: An operator may choose to run an instrument at a

processing speed lower than the instrument’s potential because of quality or

safety concerns associated with the higher speed. The reduced speed also

reduces the number of jobs the instrument can complete.

5. Quality defects and rework: The process may generate jobs that do not

meet the required specifications. In some processes, the operator can salvage

the job, but that requires extra time and effort. Anything that does not come

out correct on the first try counts as a defect.

6. Startup and yield: Processes may undergo a phase during startup when

jobs do not meet specifications. For example, if the instrument generates

faulty jobs as it warms up, those jobs count as startup or yield loss.

The first two losses – (1) breakdown and (2) setup and adjustment – affect

Availability. Similarly, (3) idling and minor stoppage and (4) reduced speed detract

from Performance. Finally, (5) quality defects and rework and (6) startup and yield

determine Quality. As discussed above, other reasons for delay, such as a shortage of

parts, can be classified as one of these types of losses.

Page 26: Reducing Reagent Waste Through Process Improvement and

26

Operator ownership of maintenance: TPM promotes the idea that when operators

gain responsibility and control of maintenance procedures, they perform maintenance

more effectively. If every mechanical error feels like an embarrassment, the operators

will go to greater lengths to prevent that error from occurring again. This greater

ownership occurs in the context of an Autonomous Maintenance (AM) program,

which leads operators to modify the work environment, the instrument, and the

maintenance procedures to drive towards zero defects. In the context of this project,

instrument modifications were not permitted under CLIA regulations. Chapter 7

discusses the specific activities of the AM team in greater detail.

Improved operator skillset: Tracing a defect to its true root cause requires a deep

understanding of the instrument. Therefore TPM involves extensive operator

training, most of it hands-on and self-driven. For example, operators are instructed

to touch every single component of the instrument and record any questions they

have about the instrument. The team records the questions and incorporates the

answers into training documents in the form of One-Point Lessons, which are

discussed in greater detail in Chapter 7.

Maintenance prevention and early equipment maintenance: The maintenance

performed during the functional lifecycle of an instrument captures only a part of the

total maintenance involved. In the manufacturing context in which TPM was

developed, instruments usually experience a high rate of defects immediately after

installation compared to the rate during normal functioning. This phenomenon arises

because the operators must determine all of the necessary operating parameters,

identify any imperfectly installed components, and integrate the instrument with the

rest of the operating environment. TPM includes early equipment maintenance,

which focuses on stabilizing the instrument as quickly as possible after installation,

thereby reducing the number of defects at the beginning of an instrument’s life.

In a medical context, start-up costs are usually associated with the extended

validation process required to confirm that the instrument operates according to

specifications and provides accurate results. Early equipment maintenance in this

context would focus on guaranteeing that the instrument requires the minimum

number of validation tests before starting patient testing.

Looking farther upstream in the instrument lifecycle, TPM also includes

maintenance prevention, discussed in Section 3.1, which aims to modify the

instrument design to eliminate maintenance steps completely. This component of

Page 27: Reducing Reagent Waste Through Process Improvement and

27

TPM requires collaboration with the instrument manufacturer and therefore often

falls outside the scope of what a company can directly accomplish in a TPM

program. This is especially true of a regulated industry like healthcare, where design

changes can be prohibitively expensive because of the cost of gaining regulatory

approval.

3.3 Drivers of success and failure Companies that effectively implement TPM show dramatic improvements in the

common measurements of plant performance – safety, quality, delivery, inventory,

and productivity – as well as employee morale. However, many organizations that

begin a TPM program do not succeed. The path from initial TPM activity to

significant results usually takes three to five years, and organizations often fail to

sustain TPM activities throughout that time [4].

The primary barriers to successful implementation are as follows [7]:

• Behavioral: Employees and/or management do not maintain the commitment,

involvement, and support to sustain TPM.

• Human and cultural: The company culture does not promote the individual

empowerment required in TPM or limits employees’ abilities to adapt to a new

approach to maintenance.

• Strategic: A lack of clear organizational goals or a poor framework for generating

decisions leads to widespread frustration and disillusionment with TPM.

• Operational: Poor organization around TPM tools, such as a lack of standard

operating procedures or a chaotic work environment, can interfere with the main

work of TPM, draining energy from the program.

• Technical: Employees do not gain a complete understanding of the TPM

principles and therefore are unable to effectively use TPM tools.

Of these, the behavioral barriers are the most commonly cited cause for TPM failure

[8]. Specifically, when top management does not maintain consistent support of TPM

throughout the implementation phase, employees will quickly return to the status

quo. The next largest barriers are organizational: when TPM is applied

inconsistently throughout an organization, the uneven application creates friction

Page 28: Reducing Reagent Waste Through Process Improvement and

28

that eventually causes the initiative to grind to a halt. The least important is

technical [7]. With a basic understanding of the principles of TPM and a clear,

coherent vision, an organization can succeed in implementing TPM. This ranking

agrees with the experiences of the author in trying to organize an autonomous

maintenance pilot project, as discussed in Chapter 7.

3.4 Chapter summary Approaches to maintenance range from the highly reactive Breakdown Maintenance,

which occurs in response to a system failure, to the proactive and holistic Total

Productive Maintenance, which aims to prevent breakdowns and streamline

maintenance procedures. An effective TPM program has the combined benefits of

increasing operator engagement, decreasing the frequency of instrument breakdowns,

and improving the rate of correctly performed jobs. However, TPM is notoriously

difficult to implement because it requires a significant cultural shift, which

companies struggle to maintain without the appropriate employee or management

support. Chapter 7 discusses the process of implementing Autonomous

Maintenance, a component of TPM, and the challenges associated with that pilot

project in this historical context.

In order to understand how Autonomous Maintenance applies here, we must first

understand the mechanics of the diagnostic testing platform that serves as the focus

of this thesis. The following chapter reviews the major components of the instrument

in detail and highlights the primary causes for mechanical errors, which an effective

maintenance program must address.

Page 29: Reducing Reagent Waste Through Process Improvement and

29

Chapter 4 Process description Having established the relevant company context and history of maintenance, we

now turn to the diagnostic testing platform that serves as the focus for this thesis. In

this chapter, we briefly discuss the analytical principle of the test, which involves

combining the patient sample with the appropriate reagents, creating a luminescence

reaction that allows the instrument to measure the concentration or identify the

presence of the compound of interest. In order to perform this sequence of automated

functions, the instrument contains several mechanical subsystems, of which 10 prove

critical to understanding the testing process. Each subsystem can lead to mechanical

errors that interrupt normal sample processing, wasting the reagent consumed during

the test and taking away from the time available for sample processing. We discuss

the leading causes of mechanical errors within each subsystem, their underlying

cause, and the routine maintenance procedures intended to reduce their occurrence.

It is also important to understand the organizational structures that interact with

the instrument. This chapter concludes by discussing the main stakeholders and how

their interests inform the resources available for this project.

4.1 Description of instrument

4.1.1 Analytical principle

The function of the instrument is to determine the presence of a target compound in

the sample. At a chemical level, the process generally proceeds as follows. The

sample is combined with a test-specific reagent in a test-specific reaction well and

incubated to allow the reagent to react with any of the target compound present in

the sample. Depending on the test, the target compound may bind to compounds

embedded in the well walls so that rinsing the well removes any unbound chemicals.

A second generic reagent provides a catalyst so that the bound compound

luminesces. By measuring the amount of light emitted, the instrument can measure

the amount of target compound present. The sample is considered reactive (the

compound is present) above a threshold high level of luminescence; non-reactive (the

compound is not present) below a threshold low level; or borderline (additional

Page 30: Reducing Reagent Waste Through Process Improvement and

30

testing may be required to determine the presence of the compound) between the

high and low levels.

4.1.2 Main mechanical systems

The instrument performs the sequence of tasks described in the previous section

using a sophisticated automated system that consists of the subsystems depicted in

Figure 4.1.

Figure 4.1: Map of main mechanical systems in instrument

(a) View from top of the instrument

(b) View from front of instrument

The mechanical subsystems interact with the patient sample as follows, with the

sequence of events involved in sample processing also shown in Figure 4.2:

i. Sample loading area: The operator loads samples through this portal.

ii. Reagent supply area: The operator loads reagent kits consisting of the liquid

reagents and the associated reaction wells into this refrigerated chamber on

the instrument. As soon as the instrument determines the type of test

Well Shuttle

Incubator

Wash Reagent

Proboscis

Signal Reagent

Proboscis

Luminometer

Reagent Supply Area

Sample Supply Area

Reagent Proboscis

Sample Proboscis

êFront

Wash Reagent

Proboscis

Signal Reagent

Proboscis Sample

Supply Area

Solid Waste Container

Incubator Reagent Supply

Area

Liquid Waste

Container

éTop

Page 31: Reducing Reagent Waste Through Process Improvement and

31

scheduled for the sample in question, it dispenses a reaction well into the

incubator through the well shuttle.

iii. Sample proboscis: The sample proboscis first picks up a disposable plastic

cone with the large end creating a seal around the body of the proboscis and

with a hole in the small end to allow sample to enter. By using the disposable

tip, the instrument minimizes the risk of cross-contamination between

samples. With the tip, the proboscis aspirates a specific volume of the sample,

as determined by the test to be performed, and dispenses the sample into the

reaction well dispensed in the previous step. The proboscis then discards the

tip in the solid waste container (see solid waste container description below).

iv. Reagent proboscis: The reagent supply area presents the appropriate type of

reagent for the test in question. Then, following the same process as the

sample proboscis, the reagent proboscis picks up a new plastic tip, aspirates a

specific volume of reagent, dispenses it into the reaction well with the sample,

and ejects the tip into the solid waste container.

v. Incubator: While the reagents react with the samples, the reaction wells sit

for a fixed amount of time as dictated by the specific test in the incubator,

which maintains the wells at body temperature. The incubator also transports

the well to the other subsystems involved in sample processing.

vi. Wash reagent proboscis: Following incubation, this proboscis dispenses the

wash reagent that removes any material not bound to the sides of the reaction

wells and removes the residual liquid.

vii. Liquid waste container: The residual liquid from the wash step is deposited in

a liquid waste container, which the operator empties regularly.

viii. Signal reagent proboscis: After the wash step, this proboscis dispenses the

reagent that catalyzes the luminescence reaction.

ix. Luminometer: The luminometer detects the amount of luminescence in the

sample relative to an internal reference.

x. Solid waste container: Finally, the instrument deposits the used reaction well

in this waste container, which the operator empties regularly. This container

is also used for plastic tips used on the reagent and sample proboscises.

Page 32: Reducing Reagent Waste Through Process Improvement and

32

Figure 4.2: Path of sample processing within instrument

4.1.3 Mechanical functions and common errors

This instrument was selected as the target for this project because of the relatively

high fraction of reagent waste that the testing platform generated over the previous

year. This section will discuss the major mechanical function of the main subsystems

and the dominant types of mechanical errors observed in each subsystem, each of

which lead to reagent waste. Whenever a mechanical error occurs, the instrument

attempts to re-run the test on that sample. If the instrument cannot automatically

schedule a repeat test, the operator can manually re-run the test at a later point.

The instrument reports an error code when sample testing is interrupted so that the

operator can determine the cause of the error.

i. Sample loading area: This component involves (1) a rotating set of positions

where samples can be loaded and (2) sensors that read sample barcodes in

order to determine which test or tests have been ordered for each sample. The

mechanics of this subsystem are relatively robust and are only affected if gross

contamination clogs the movement of the rotating components or blocks the

sensors.

ii. Reagent supply area: Reagent kits consist of containers of the liquid reagent

and sleeves containing nested stacks of the reaction wells. The reagents do not

require any advanced preparation, and the lifespan of a single reagent pack on

the instrument is long compared to the rate of reagent consumption, so

reagent quality concerns or expiration rarely generate waste at Quest

Diagnostics labs. However, if the reagent kits are agitated before being loaded

onto the instrument, bubbles at the top of the reagent container can cause

issues for the reagent proboscis.

1. Operator loads sample into Sample Supply Area. 2. Reagent Supply dispenses reaction well to incubator through

Well Shuttle. 3. Sample Proboscis aspirates patient sample. 4. Sample Proboscis dispenses patient sample into reaction well. 5. Reagent Proboscis aspirates reagent in Reagent Supply Area. 6. Reagent Proboscis dispenses reagent into reaction well. 7. Reaction well stays in Incubator for designated length of time. 8. Wash Reagent Proboscis rinses residual liquid from reaction

well. 9. Signal Reagent Proboscis dispenses signal reagent into

reaction well. 10.  Luminometer measures luminescence in reaction well. 11.  Incubator ejects reaction well into Solid Waste Container.

1

2 3 4

5 6

89

10

11

Page 33: Reducing Reagent Waste Through Process Improvement and

33

The more common mechanical error for the reagent supply area relates to the

reaction wells. The reagent kits hold the wells in a sleeve open at the bottom

and top. A screw with a tip fitted to the diameter of the sleeve moves down a

fixed number of mechanical “steps” while pushing on the top well to cause a

single well to fall out the bottom of the sleeve into the incubator. A sensor

within the incubator confirms that a single well has been dispensed. Moisture

in the instrument can lead the wells to stick together rather than nesting

loosely, preventing a well from dropping or causing two wells to drop

simultaneously. In each instance, the instrument will abort the test and

discard any wells in the well shuttle into the solid waste container.

iii. Sample proboscis: The sample proboscis consists of an external sleeve and an

internal hollow piston covered with a plastic cap. The proboscis is connected

to a pump that draws air through the hollow piston in order to aspirate and

dispense sample liquid. The piston cap protects the narrow piston from

exposure to liquid, and it creates small openings between the sleeve and the

piston through which air must pass before it can be drawn up through the

piston. During use, a disposable plastic tip fits onto the external sleeve. In

this way, sample only comes into contact with the disposable tip rather than

the piston, piston cap, or external sleeve.

The main mechanical issues from the sample proboscis derive from debris

accumulating on the proboscis that compromises airflow. As the proboscis

moves within the instrument, sample can splash up onto the piston cap and

dry, leaving residue. The instrument software determines aspirated and

dispensed liquid volume by measuring the pressure changes detected at the

pump. Therefore if the debris disrupts airflow, it can lead to deviations from

the expected pressure profile. Then, because the instrument cannot guarantee

the accuracy of the volume measurement, it abandons the test.

Sample integrity issues can create the same effect. For example, insufficient

sample volume, heterogeneity in sample consistency, or bubbles in or on the

sample can also lead to unusual pressure profiles, causing the instrument to

stop processing the sample.

Mechanical errors also occur from interactions with the plastic tips: the

proboscis may fail to pick up the tip on the first try, it may fail to eject the

tip after processing, or the tip may have smaller than usual opening, impeding

airflow.

Page 34: Reducing Reagent Waste Through Process Improvement and

34

iv. Reagent proboscis: The function and primary error types of the reagent

proboscis are identical to those of the sample proboscis. The reagent proboscis

generally aspirates and dispenses a greater amount of liquid than does the

sample proboscis, so debris may accumulate more rapidly than in the sample

proboscis.

v. Incubator: The incubator is a heated compartment containing several

concentric rings with holes where the reaction wells sit. The rings rotate the

wells among positions dedicated for different steps in the process. The

compartment has holes through which sample and the different types of

reagent are dispensed so that the well remains inside the incubator through

these steps.

The wells also are moved between rings for different stages of the incubation.

To move the well between rings, a small metal rod pushes up through the

holes in the ring to lift the well out of its position and into a shuttle. Inside

the shuttle, metal tabs grip the well until the shuttle moves the well into

position over the appropriate ring, at which point the tabs release. This

process also repeats to allow the instrument to measure the luminescence: the

rod lifts the well into the shuttle, which holds the well in position for

measurement and then transports the well to the disposal chute leading to the

solid waste container.

Thus the incubator involves many moving parts that must coordinate

precisely in order to process samples. If any of these movements breaks down,

the incubator can jam, which interrupts all of the tests in the incubator at the

time – as many as 100. Mechanical jams can occur when sample or reagent

splashes out of the wells onto mechanical components; when a proboscis is out

of alignment or is leaking, causing liquids to fall onto mechanical components;

when debris prevents the many position sensors from verifying the location of

moving components; when fibers are left behind after maintenance activities;

when a previous jam in the well transport system causes a well to get caught

in moving components and shatter, leaving behind plastic chips; and when the

solid waste container overflows and backs up into the incubator.

vi. Wash reagent proboscis: The wash reagent proboscis consists of two hollow

metal straws, one of which dispenses wash reagent and one of which aspirates

the mixture for transport to the liquid waste container. The design is simpler,

Page 35: Reducing Reagent Waste Through Process Improvement and

35

and the wash reagent is both more dilute and more homogeneous than

samples or the reactive reagent. As a result, wash reagent proboscis does not

usually experience the same problems with blockage seen in the reagent and

sample proboscises. Instead, the primary mechanical issue facing the wash

reagent proboscis is the integrity of the fluidics system. If a connection leaks

or a crack appears in a line, then air bubbles can infiltrate the lines, leading

to leakage. If the instrument detects that the incorrect amount was dispensed,

then it will abort the test in progress. In addition, leakage that falls into the

incubator can contribute to mechanical jams.

vii. Liquid waste container: The liquid waste container collects the liquid waste

from sample processing and does not involve a high level of mechanical

complexity. The primary problem arising from this subsystem occurs when

the waste container fills up during sample processing, which will cause the

instrument to stop processing additional samples.

viii. Signal reagent proboscis: The signal reagent proboscis consists of two metal

straws that dispense the two possible types of signal reagent. Like the wash

reagent proboscis, the primary source of mechanical errors with this

subsystem lies in air intrusions into the fluidics system.

ix. Luminometer: To generate accurate results, the luminometer requires

complete exclusion of outside light. The instrument isolates the luminometer

by performing the measurement in a closed compartment within the

incubator. However, if the compartment cover is not seated properly, external

light sources may infiltrate the chamber.

x. Solid waste container: Similar to the liquid waste container, the solid waste

container involves little mechanical complexity. However, if not emptied

regularly, the wells can accumulate, back up into the disposal chute, and

eventually jam the moving components of the incubator.

4.1.4 Reagent waste

The instrument performs the process described above for three types of samples:

patient samples, quality control samples, and calibration samples. Quality control

materials are used to confirm that the instrument is continuing to generate

consistent results over time, as described in greater detail in Chapter 6. Calibration

samples are used to fix the instrument’s luminescence readings to known values,

Page 36: Reducing Reagent Waste Through Process Improvement and

36

thereby correcting for any drift that happens over time or to accommodate for any

differences in reagents that would lead to subtly different results.

The outcome of all of these tests will either be a reactive, nonreactive, borderline, or

incomplete result. The reactive and nonreactive tests indicate the presence or

absence of the target compound, as discussed in Section 4.1.1. A borderline result

indicates a luminescence level that does not strongly indicate either the presence or

the absence of the target compound and triggers a repeat test to confirm its

reactivity. An incomplete result arises when a mechanical error interrupts normal

sample processing. In this instance, a well dispensed by the instrument does not

generate a luminescence reading.

According to the definitions in this analysis, reagent is considered waste when it does

not lead to reactive or nonreactive results for patient samples. Thus quality control

and calibration samples are counted as waste for the purposes of this analysis, even

though they are critical components of a quality program and are required by CLIA

and Quest SOP’s. Likewise, borderline and incomplete results are both considered

waste. However, because borderline results arise from the chemical state of the

patient sample, they were not a focus of this analysis.

The platform of focus for this study was selected because preliminary data indicated

that it generated a relatively high fraction of waste and, anecdotally, it experienced a

relatively high rate of mechanical errors.

4.2 Maintenance overview In addition to running patient, quality control, and calibration samples as part of

normal operations, instrument operators also perform routine preventive

maintenance and troubleshooting. When preventive or reactive maintenance goes

beyond the training of operators, representatives from the instrument manufacturer

also perform service. This section discusses the types of maintenance performed by

each group.

4.2.1. Operator maintenance

The operators clean the major instrument subsystems at a frequency recommended

by the manufacturer and determined by the approximate rate of debris accumulation

on that subsystem. Cleaning includes all proboscis tips; all moving components of

the incubator, including the metal rods that lift and lower reaction wells; and

surfaces where samples might spill, such as the sample supply area. After manually

Page 37: Reducing Reagent Waste Through Process Improvement and

37

cleaning the proboscises, the operator also runs an instrument program that confirms

effective function of all proboscises.

In addition to cleaning, operators must also maintain adequate supplies of

consumables like the plastic tips for the reagent and sample proboscises and the

reagent kits, and they must regularly empty the solid and liquid waste containers.

The testing platform includes software that guides the operator through all the steps

involved in maintenance with detailed instructions of the cleaning supplies and

procedures and diagrams of the targets of cleaning.

Troubleshooting usually involves replicating a set of routine maintenance activities

to address the potential source of debris. The operators also have the authority to

replace certain parts, such as the reagent and sample proboscises. The user interface

of the testing platform allows an operator to look up the recommended

troubleshooting process for any condition code that the system generates. Beyond a

limited set of activities, a vendor representative must intervene.

4.2.2. Vendor maintenance

Engineering representatives from the manufacturer perform more complex

troubleshooting operations and more comprehensive preventive maintenance

activities. They may replace components such as connectors and fluid lines for the

signal reagent system.

4.3 Stakeholder analysis

4.3.1. Primary groups and their interests

Despite this project’s narrow focus – reagent waste on a single testing platform – it

nonetheless affected many groups within Quest Diagnostics at both a local and a

national scale. Aligning the interests of these overlapping groups proved a leading

challenge throughout the project. This section describes the main groups, their

interests, and the leading sources of friction.

i. Operators: The instrument operators prioritized completing their work

effectively and preserving quality of patient testing results. Beyond

successfully completing their work, they also expressed a desire to minimize

the frustrations involved in their work, such as the hassle of repeating dozens

of patient samples because of a mechanical error.

Page 38: Reducing Reagent Waste Through Process Improvement and

38

ii. Laboratory management: The supervisors and managers at a local level

prioritize testing quality as well as general laboratory metrics. Result

turnaround time and employee productivity (e.g., tests per full-time

employee) received the most attention during the thesis.

iii. Regional management: Higher tiers of laboratory management set the

performance goals for the laboratory within the larger context of the

Invigorate initiative. They therefore placed more value on long-term

investments like training and improvement initiatives like Total Productive

Maintenance than local managers and supervisors did.

iv. Manufacturer: The manufacturer of the testing platform had three main

interests. First, they wanted to strengthen the relationship with Quest

Diagnostics to drive more business in the future. Second, they wanted to

preserve their internal norms related to work-life balance. They resisted

requests that significantly increased the workload of any one employee and

argued for longer timelines for delivering on requests for access to data or

information about instrument performance. Third, reagent serves as a

significant source of revenue for the manufacturer, so they will suffer in the

short term from any process improvements that decrease the amount of

reagents that Quest Diagnostics purchases. This short-term impact is

balanced against the vendor’s drive to become a strategic business partner for

Quest Diagnostics, thus improving the likelihood of future revenue

opportunities. While this issue did not arise explicitly during the thesis, it

served as an important backdrop for understanding the manufacturer’s motivations.

v. Best Practices Team (BPT): The national BPT that oversaw the standard

operating procedures (SOPs) for the target testing platform wanted primarily

to establish guidelines that would guarantee the quality of test results. They

acknowledged the value from reducing reagent waste but resisted any actions

that could potentially reduce result quality.

vi. Reagent waste team: The national working group tasked with generating

savings within the Invigorate program focused on savings opportunities that

would not compromise result quality.

Page 39: Reducing Reagent Waste Through Process Improvement and

39

4.3.2. Potential conflicts

The main conflicts that arose among these groups are the tensions between (1)

productivity goals and laboratory improvement initiatives, (2) reagent costs from

excessive quality control testing and benefits from more conservative quality

measures, and (3) manufacturer intentions and capabilities. Each conflict is discussed

briefly below.

i. Productivity goals vs. laboratory improvement initiatives: The regional

management championed improvement activities such as Total Productive

Maintenance. However, implementing such activities requires a significant

investment of equipment and operator time. While laboratory management

supported the premise of these improvement initiatives, they were constrained

by conflicting requirements related to employee productivity and turnaround

time, which limited their ability to dedicate staff to the thesis work.

ii. Reagent costs from excessive quality control testing vs. benefits from more

conservative quality measures: Quality control is a critical and unavoidable

part of testing. Applying quality control practices that increase the frequency

or quantity of quality control material consumed may add certainty to the

results, but it always increases reagent costs. The reagent waste team

identified a few opportunities to reduce reagent costs that involved adopting a

less conservative quality control policy. Because the BPT prioritizes quality so

highly, these suggestions sparked significant resistance, regardless of whether

data indicated that using the more conservative approach actually resulted in

higher quality.

iii. Manufacturer intentions vs. capabilities

Finally, the manufacturer intended to use this thesis as an opportunity to

build a stronger relationship with Quest Diagnostics. However, the analysis

revealed gaps in their capabilities, such as a lack of data correlating an

interrupted test to a specific condition code, that could not be resolved on the

timescale of the thesis.

4.4 Chapter summary The target instrument for this thesis involves 10 major mechanical subsystems

that must function in concert to generate patient results. A mechanical error at

any one of the subsystems has the potential to interrupt sample processing,

forcing the instrument to discard the sample, which then must be retested.

Page 40: Reducing Reagent Waste Through Process Improvement and

40

Certain types of errors, including most errors in the incubator, can interrupt

sample processing for all tests in progress at the time – as many as 100. Routine

operator maintenance and vendor maintenance activities are designed with the

intention of minimizing the occurrence of these errors.

Any changes to current procedures must balance the interests of the primary

groups involved with the instrument. The balance between productivity goals and

laboratory improvement activities constrains the activities of the Autonomous

Maintenance team, as discussed in Chapter 7. The balance between reagent

costs from excessive quality control and the benefits from more conservative

quality control measures informs the recommendations about quality control

procedures discussed in Chapter 6. Finally, the tension between the

manufacturer’s intentions and their capabilities limited the possible depth of

quantitative analysis in this thesis. The next chapter discusses the current state

of the process, given the available data.

Page 41: Reducing Reagent Waste Through Process Improvement and

41

Chapter 5 Current state analysis Building on the process details discussed in the prior chapter, this chapter discusses

the current state of the process as of the beginning of the project. Establishing the

current state involves process observation and data analysis. Observed processes

include standard operation, quality control, routine maintenance, and

troubleshooting. Observation occurred at four laboratories in distinct geographical

areas of the United States to understand the basis for variations in preliminary data

about reagent waste. For the purposes of the following discussion, the term “waste” refers to any reagent that does not contribute to a unique patient result. Therefore

reagent used in quality control and calibration – necessary and legally mandated

parts of a quality system – is termed waste, despite their role in the process. When

observations of quality control procedures revealed variations in implementation, all

sites using the testing platform were surveyed to gather comprehensive information

about quality control practices.

In parallel with direct observations, the process was also analyzed using three forms

of instrument data: (1) usage counters, which track how many wells generate

patient, quality control, calibration, or incomplete results; (2) condition code reports,

which record the frequency of mechanical errors on the instrument; and (3)

incubator maintenance charts, which show how long the incubator was opened for

maintenance, thus providing a crude metric for maintenance adequacy. The

preliminary observations from the data analysis include:

1. Quality control is a bigger source of waste than mechanical errors

2. Three instrument subsystems drive the majority of mechanical errors

3. The most common condition codes may be addressed through

improved maintenance.

Thus quality control and mechanical errors arise as the major sources of waste to be

tackled in this thesis. Chapter 6 goes on to discuss how to address the variations in

Page 42: Reducing Reagent Waste Through Process Improvement and

42

SOP interpretations, and Chapter 7 discusses how to unify maintenance and

operation around national best practices.

5.1 Process observation Observations of the process at four laboratories across Quest Diagnostics led to a

preliminary assessment of some of the drivers of reagent waste, specifically,

individual laboratories exceeding quality control requirements and a lack of

consistent dissemination of best practices in maintenance and operation.

5.1.1 Activities completed

This section discusses in detail the types of observations performed and the content

of the survey of quality control practices.

5.1.1.1 Observed components of routine operation

As described in Chapter 4, the interactions between the operators and the testing

platform consisted of (i) routine operation, (ii) quality control, (iii) routine

maintenance, (iv) troubleshooting, and (v) calibration. As discussed below in

Section 5.2.2, calibration comprised a small fraction of overall reagent usage, so

calibration was not a focus during the observations.

Steps (i)-(iv) were observed as described below in detail. The sections are separated

into processes that could be observed directly and those that required conversations

with the operators, lab manager, or other employees to evaluate.

A critical component of the analysis involves comparisons across different sites.

Anecdotally different sites experienced dramatically varying rates of reagent waste.

To understand the basis for these differences, observation was repeated at four

laboratories in distinct geographical regions: Marlborough, MA; Wood Dale, IL;

West Hills, CA; and Miramar, FL.

i. Routine operation

- Observed:

• Routine operations starting from delivery of patient samples to the

testing area, continuing through sample processing and result

interpretation, and concluding with returning samples to relevant

storage; and

• The inventory policy by noting the location and quantities of

consumable materials.

Page 43: Reducing Reagent Waste Through Process Improvement and

43

- Discussed:

• The general sample mix and the turnaround time requirements for each

lab;

• The layout of the lab and any recent or anticipated changes to the

number, generation, and organization of the instruments;

• The operator’s experience of running the instruments and qualitative

impressions of the main sources of waste; and

• The occurrence of reagent expiration, either while in storage or while

loaded on the instrument.

• Operators’ experience of reagent expiration.

ii. Quality control

- Observed:

• The containers used for quality control; and

• The location and storage approach for quality control materials.

- Discussed:

• The frequency of quality control samples;

• The acceptance criteria for quality control results;

• Strategy for addressing quality control failures; and

• Relevant lab history leading to quality control policies.

iii. Routine maintenance

- Observed all weekly maintenance happening within a one-week timescale,

including removing waste, replacing consumables, and cleaning instrument

components.

- Discussed:

• Training history of operator performing maintenance; and

• Operator’s understanding of basis for maintenance procedures.

iv. Troubleshooting:

- Observed operators troubleshooting mechanical errors on the instrument when

the opportunity arose during routine operation.

- Discussed what the operators perceived to be the most common reasons for

interruptions to sample processing and their strategies for addressing the

issues.

Page 44: Reducing Reagent Waste Through Process Improvement and

44

In addition to these structured observations, informal conversations were held with

key laboratory personnel to understand the main challenges facing that laboratory,

relevant process improvement initiatives, and unique qualities of the client base that

influence operations.

5.1.1.2 Prepared quality control survey

The laboratory observation highlighted variations in how laboratories interpreted the

SOPs for the testing platform. In order to understand the full range of quality

control practices for this instrument, a survey was designed with input from the

BPT and circulated to all laboratories using the instrument.

The questions within the survey included the following:

1. How often does your lab run positive quality control materials?

2. How often does your lab run negative quality control materials?

3. What size container does your lab use for quality control materials?

4. Which statement applies most closely to your lab’s practices:

a. “We prepare new containers of quality control material for each

use.”

b. “We prepare a container of quality control material and reuse it for

multiple rounds of quality control testing, but we do NOT replenish

the container with new quality control material.”

c. “We prepare a container of quality control material and reuse it for

multiple rounds of quality control testing, and we replenish the

container from the same lot of parent quality control material as

needed.” 5. Does your lab refrigerate quality control material between uses?

6. Does your lab cover the container of quality control material between

uses? If so, what kind do you use?

7. For qualitative assays, does your lab track quality control results

qualitatively or quantitatively?

8. How does your lab respond when a mechanical error occurs on a

quality control sample, leading to an incomplete result?

The responses to the survey are discussed in Chapter 6.

Page 45: Reducing Reagent Waste Through Process Improvement and

45

5.1.2 Preliminary observations

Preliminary observations from visiting the four laboratories can be summarized as

follows:

1. Mechanical errors are a major source of irritation for operators;

2. Operator knowledge levels about instrument function, operation, and

maintenance varied within and across labs;

3. Individual operators/labs develop local process improvements, but do

not consistently share with other labs; and

4. Interpretation of quality control requirements in SOPs varied across

labs.

i. Mechanical errors are major source of irritation for operators

Operators at all sites perceived the instrument to be sensitive to mechanical

errors, which fall into two general categories: transient errors and shutdown

errors. For the purposes of this thesis, transient errors are defined as mechanical

errors that affect only the sample currently being processed, and shutdown errors

are those that affect all samples being processed by the machine at that time

(i.e., including those in the incubator). In all cases of a mechanical error, the test

is re-run until a result can be delivered to the requesting provider.

A common example of a shutdown error is an incubator crash. The incubator

contains many moving mechanical parts, and if a jam occurs at any point in the

incubator, the instrument rarely can recover without operator intervention. At

that point, all sample processing halts, and the operator must remove the jam

before the machine can continue processing. The instrument may have up to 100

samples in the incubator at one time, and all these are forfeited because of the

potential for contamination when the operator opens the instrument to address

the jam.

An example of a transient error occurs when the opening for airflow in the

reagent proboscis becomes slightly occluded. The blockage may occasionally

interfere with airflow through the proboscis and therefore with the instrument’s ability to calculate the quantity of reagent aspirated or dispensed. When the

instrument cannot calculate this quantity to a high degree of certainty, it will

jettison the reaction currently being processed and try again with a new aliquot

of sample. This error may occur periodically until the proboscis is cleaned during

routine maintenance, but sample processing will otherwise continue without

interruption.

Page 46: Reducing Reagent Waste Through Process Improvement and

46

There are two addenda to the definitions as outlined. First, a transient error can

also escalate into a shutdown error. In the example of a reagent proboscis above,

a more serious blockage can cause sequential errors of the same type. When such

an error repeats for a threshold number of times, the instrument stops all sample

processing until the issue is resolved, usually through operator maintenance.

Second, a shutdown error does not always lead to the loss of all samples in

progress. If the shutdown error derives from an issue with the reagent proboscis,

the incubator does not need to be opened, so there is no potential for

contamination. In this instance, the only tests that must be discarded are those

that incubate longer than the time established by the standard operating

procedure and therefore may have invalid results. If the operator resolves the

problem quickly, few if any tests may be discarded.

As a rule, the instrument discards tests whenever there is a mechanical concern

for contamination or incorrect results, per standard industry practice.

Operators perceive the instruments to be sensitive to mechanical errors because

of the impact of shutdown errors. While transient errors generate more reagent

waste than shutdown errors, the shutdown errors have an outsized effect on the

operators because shutdown errors must be addressed immediately, creating

disruptions to the operator’s regular schedule; and increase the effective workload

for a given shift, increasing the difficulty of meeting turnaround time

requirements.

However, operators generally acknowledged that the current instrument

experiences fewer mechanical issues than the previous generation of the

instrument, which Quest Diagnostics is phasing out of operation.

ii. Operator knowledge levels about instrument function, operation, and

maintenance varied within and across labs

The level of operator skill varied within labs. Generally skill and depth of

knowledge increased with experience, and operators who had attended a weeklong

training session hosted by the instrument vendor had considerable expertise. The

experts were not necessarily responsible for routine maintenance and operation: in

order to provide development opportunities and to encourage workforce

flexibility, Quest Diagnostics promotes cross-training, so the experts were usually

responsible for training their colleagues on the instrument so that more people

were available to operate or maintain the instrument.

Page 47: Reducing Reagent Waste Through Process Improvement and

47

The skill levels of some of the less experienced operators suggested that there was

an opportunity to strengthen and unify training procedures, especially as they

related to maintenance. For example, one operator spent time cleaning a

component that, because of the types of testing performed in Quest Diagnostics

laboratories, is never used and therefore has no impact on instrument

performance.

iii. Individual operators/labs develop local process improvements, but do not

consistently share with other labs

Individual labs developed innovative ways to improve operation and maintenance

of the instrument. Encountering ergonomic stress when removing debris from a

deep shelf on the instrument, one laboratory created a tool that helped sweep off

debris and eliminate the strain. The same lab had collaborated with the vendor

to apply labels to key components within the instrument with helpful reminders

about proper operation and maintenance (e.g., “Wait until light turns green

before opening compartment”). A different lab designated a cart to hold

maintenance equipment for the instrument to reduce the time spent collecting the

necessary tools every day. Each of these ideas was developed locally and was not

shared outside that laboratory even though other labs would benefit from those

insights.

iv. Interpretation of quality control requirements in SOPs varied across labs

All of the labs met the minimum quality control requirements as specified by

SOPs; however, the precise interpretations varied across labs, with some labs

choosing to exceed requirements significantly. The variations relevant for this

study in waste reduction were as follows:

1. Frequency: some labs ran quality control samples more often than was

required by the SOP.

2. Container size: the size of the container determines the amount of

“dead volume”, i.e., liquid the instrument cannot access, that must be

discarded after testing.

3. Acceptance criteria: some tests on the target platform yielded

qualitative results and others yielded quantitative results. Laboratories

differed in the acceptance criteria for quality control samples, with

Page 48: Reducing Reagent Waste Through Process Improvement and

48

some tracking all tests through quantitative methods, and others using

a qualitative or quantitative methods depending on the type of test.

4. Reuse policy: one lab prepared new containers of quality control

material for every use, while another used a single container of quality

control material for multiple batches. Yet another maintained a

container of quality control material that was replenished periodically

with quality control material from the original parent batch.

5. Storage approach: reusing quality control materials creates the

possibility of evaporation of quality control material between batches.

Some labs opted to cap and refrigerate the quality control containers in

between batches while others did not cap or refrigerate.

6. Approach to mechanical errors on quality control samples: occasionally

a transient error would occur on a quality control sample. One lab

espoused the policy of rerunning all samples from the affected assay

when this occurred. A different lab repeated several patient samples

and treated the results as sufficient quality control if they agreed with

the results as reported previously.

5.2 Data collection and evaluation

5.2.1 Sources of data

The qualitative observations described in Section 5.1 highlighted the need to

quantify the sources of waste and measure the impact of mechanical errors on

reagent usage. The data sources used to generate the results discussed in Chapters

6 and 7 consisted of usage counters, condition code reports, and incubator

maintenance charts.

i. Usage counters

The instrument vendor tracked the outcome of every well dispensed by each copy of

the testing platform at Quest Diagnostics. Outcomes fell into four categories:

1. Patient – completed tests of patient samples;

2. Quality control – completed tests of quality control samples;

3. Calibration – completed tests performed during instrument calibration;

and

Page 49: Reducing Reagent Waste Through Process Improvement and

49

4. Incomplete – any of the three previous types of samples that were

interrupted during processing because of a mechanical error.

Because of the relatively infrequent communication between the instruments and the

vendor servers, usage counters could only be recorded on a weekly basis. They

therefore served to provide a general sense of large trends over time or across

business units but were not valuable in evaluating countermeasures tested on shorter

timescales.

The usage counters do not take into consideration the amount of quality control

material that remained in the sample container after testing and was discarded. We

were unable to measure this quantity directly, instead opting to calculate the

minimum required dead volume depending on sample container and reuse policy.

ii. Condition code reports

The usage data provides insight into the frequency of mechanical errors but not their

cause. To understand the drivers of mechanical errors, the vendor provided reports

of the condition codes generated by the instruments during sample processing.

Condition codes ranged in content and severity, and most condition codes did not

indicate that a test had been lost. When the instrument software linked a condition

code to a test in progress, it is assumed that the test was interrupted, leading to an

incomplete result. This correlation did not hold 100% of the time, so vendor software

specialists periodically reviewed the condition codes to confirm data accuracy.

We could not, however, overcome the other weakness of the condition code data:

those data alone did not indicate whether the condition code led to a transient error

or a shutdown error. Therefore the number of incomplete tests captured by the usage

data is expected to exceed the number of condition codes.

The vendor further summarized the condition code reports by sorting the codes into

subsystems. Most condition codes were linked to a physical subsystem within the

instrument such as the incubator, reagent proboscis, or sample proboscis. The

condition codes were sorted by relevant subsystem to highlight instrument

components that required more or different maintenance. As noted above, the

frequency of condition codes did not correlate exactly with the number of tests lost

because a shutdown error would result in one instance of a condition code but many

– up to 100 – lost tests.

Page 50: Reducing Reagent Waste Through Process Improvement and

50

Instrument documentation provided some insight into potential root causes for the

mechanical errors and troubleshooting steps. Instrument engineers from the vendor

also drew from their experience to suggest operator actions that could prevent the

most common codes.

iii. Incubator maintenance charts

The vendor was able to record automatically the time that the instrument incubator

was open. The vendor’s maintenance procedures for the incubator required at least

15 minutes of operator time, as verified by the vendor and confirmed independently

by a team within Quest Diagnostics, so these data provided a crude indication of the

quality of incubator maintenance. The data did not confirm that maintenance was

being performed correctly, but if the incubator were open for fewer than 15 minutes,

they would strongly indicate that maintenance was performed incorrectly.

Because of the relatively small amount of information provided by these charts, they

were not included in subsequent analyses. Instead, they were used as an illustrative

tool of the importance of complete maintenance during site observations.

5.2.2 Preliminary observations

Preliminary results from these data sources established the current conditions for the

target platform and guided the choice of countermeasures discussed in Chapters 6

and 7. The primary insights from these preliminary data were as follows:

1. Quality control is a bigger source of waste than mechanical errors

2. Three instrument subsystems drive the majority of mechanical errors

3. The most common condition codes may be addressed through

improved maintenance

i. Quality control is a bigger source of waste than mechanical errors

Recall that while quality control is a critical part of the patient testing process, it

qualifies as waste for this analysis because it is not a patient test result. Usage

counter data from March-May 2016 showed that within an example laboratory,

quality control tests accounted for 7.1% of all reagent usage at the site, almost twice

as much as incomplete tests from mechanical errors, which accounted for 3.6% of

reagent use, as shown in Table 5.1. Calibration tests accounted for a negligible

0.16% of reagent used, and there was no indication of excessive calibration or

frequent calibration failures that would prompt further investigation. Therefore

calibration is treated as an insignificant source of waste for the remainder of the

thesis.

Page 51: Reducing Reagent Waste Through Process Improvement and

51

Table 5.1: Reagent usage March-May

This table shows the fraction of reagent that generated four types of results: patient,

quality control, calibration, and incomplete. The incomplete results arise when a

mechanical error interrupts normal sample processing. The rates remain stable from

month to month, and the usage patterns at an example laboratory are consistent

with the overall picture within Quest Diagnostics. However, during this time period,

the example lab consistently used quality control at a slightly higher rate than

average, and mechanical errors occurred at a slightly lower rate.

Month / Laboratory Patient Quality Control Calibration Incomplete

Marc

h

Company Average 90.1% 5.3% 0.2% 4.5%

Example Lab 87.2% 7.7% 0.2% 4.9%

Apri

l Company Average 90.5% 5.2% 0.2% 4.1%

Example Lab 89.3% 6.9% 0.1% 3.7%

May Company Average 90.3% 5.1% 0.2% 4.4%

Example Lab 91.4% 6.3% 0.2% 2.1%

All

Company Average 90.3% 5.2% 0.2% 4.3%

Example Lab 89.1% 7.1% 0.2% 3.6%

Nationally over the same time period, quality control and incomplete tests accounted

for 5.2% and 4.3% of tests, respectively. The results for the example lab were largely

consistent with the company-wide averages over the same time period, indicating

that instrument behavior in the example laboratory is representative of normal

instrument operation, with two exceptions: the example laboratory consistently

performed slightly more quality control tests and experienced slightly fewer

mechanical errors than the national average. These subtle differences aside, the usage

data suggests that insights from activities in example might be applicable at a

national scale.

ii. Three instrument subsystems drive the majority of mechanical errors

During March-May 2016, prior to the start of the internship, three instrument

subsystems accounted for 89.7% of mechanical errors, based on the condition code

reports. As shown in Table 5.2, the sample proboscis, reagent supply area, and the

reagent proboscis accounted for 49.5%, 20.3%, and 19.9% of mechanical errors,

respectively. As discussed in Section 5.1.1 above, operators perceive incubator

crashes to generate the most waste, probably because of the outsized impact that

Page 52: Reducing Reagent Waste Through Process Improvement and

52

any one incubator crash can have, in terms of both lost tests and pressure on

operators. However, incubator crashes occur infrequently compared with transient

errors on the reagent proboscis, accounting for only 1.7% of the mechanical errors.

Table 5.2: Error code data for an example laboratory March-May

This table shows the relatively frequency of mechanical errors occurring on the three

instances of the testing platform in the example lab. The mechanical errors are

sorted by the subsystem where the error occurred. The “Other” subsystem includes

software complications or other errors that are not associated with a major

instrument subsystem.

Instrument

1 Instrument

2 Instrument

3 Total

Sample Proboscis 39.3% 67.4% 35.2% 49.5%

Reagent Supply 27.5% 13.9% 22.2% 20.3%

Reagent Proboscis 17.4% 12.9% 31.5% 19.9%

Luminometer 11.1% 0.6% 8.9% 6.1%

Other 2.6% 2.1% 0.9% 1.9%

Incubator 1.5% 2.4% 1.0% 1.7%

Sample Supply 0.7% 0.8% 0.3% 0.6%

iii. The most common condition codes may be addressed through improved

maintenance

Review of the condition codes for the reagent and sample proboscis with guidance

from vendor engineers revealed that the most common codes for these subsystems

can be addressed through improved maintenance. 90% and 54% of the codes for the

reagent proboscis and sample proboscis, respectively, arise because debris inhibits

smooth airflow through the proboscis, triggering transient errors and potential

shutdown errors. By increasing the intensity or frequency of maintenance on these

subsystems, the laboratory may reduce the frequency of these errors, thereby

reducing reagent waste and improving turnaround time.

Of the reagent supply errors, 74% came from a single condition code: two reaction

wells were dispensed into the incubator at once. This occurs primarily because

humidity or some other binding agent causes wells to stick together. Other than

ensuring that the reagent kits are stored in low-humidity conditions throughout the

chain of custody, there are no operator-based actions that can reduce the frequency

of this error. Therefore the reagent supply condition codes are not considered good

targets for maintenance-based countermeasures and are not discussed further in the

thesis. However, based on this result, the vendor began discussions with the reagent

Page 53: Reducing Reagent Waste Through Process Improvement and

53

kit manufacturer to test upstream process changes that could reduce the frequency

of these errors.

5.3 Chapter summary Routine operations including quality control, maintenance, and troubleshooting were

observed at four geographically separated laboratories, leading to the following

observations:

1. Mechanical errors are a major source of irritation for operators;

2. Operator knowledge levels varied within and across labs;

3. Individual operators/labs develop local process improvements, but do

not have a clear channel for sharing with other labs; and

4. Interpretation beyond the minimum quality control requirements in

standard operating procedure (SOPs) varied across labs.

Despite the first observation above, analysis of the usage data indicated that quality

control samples actually constitute a larger fraction of reagent use than mechanical

errors, i.e., they lead to more waste. This observation motivates the efforts discussed

in Chapter 6 to identify unnecessary quality control practices that lead to reagent

costs without commensurate benefits to the quality of results.

Nevertheless, mechanical errors occur in approximately 4.3% of tests. These errors

occur predominantly in three of the instrument subsystems described in Chapter 4:

the sample proboscis, the reagent proboscis, and the reagent supply. The most

common mechanical error associated with the reagent supply system usually arises

from manufacturing conditions, thus falling outside the scope of this thesis. However,

the leading errors for the other two systems may be preventable through improved

maintenance procedures that go above and beyond manufacturer requirements. This

observation motivates the Autonomous Maintenance team activities discussed in

Chapter 7 that intend to reduce instrument errors through enhanced maintenance.

Page 54: Reducing Reagent Waste Through Process Improvement and

54

Chapter 6 Countermeasures to eliminate quality control waste The previous chapter established that quality control was one of the two major

sources of reagent waste on the testing platform. In addition, laboratories varied

significantly in their interpretation of SOPs related to quality control procedures.

This chapter discusses the approaches used to investigate the source of these

variations and develop recommendations for aligning practices with the least

wasteful approach that meets quality control requirements.

The chapter begins with a summary of the relevant requirements for quality control

– both regulator-enforced and internally established – which set the foundation for

what these procedures must achieve. Then we review the results of the quality

control survey, which showed variations in quality control frequency, container size,

acceptance criteria, reuse policy, storage method, and approach to mechanical errors.

These variations led to large differences in how much quality control material was

consumed and how much reagent was used in quality control testing, both of which

qualify as waste.

The different approaches are then discussed in terms of their potential impact on

waste and result quality. Based on this analysis, one example laboratory opted to

reduce the frequency of negative quality control sample testing to be consistent with

the SOP, resulting in a reduction in reagent use in quality control testing by nearly

half.

This analysis reveals the importance of creating unified procedures, and the BPT

must establish those guidelines, ideally to be incorporated into the SOPs. The

chapter closes with recommendations for how Quest Diagnostics can sustain

consistency and continue to elevate performance across the company.

Page 55: Reducing Reagent Waste Through Process Improvement and

55

6.1 CLIA regulations regarding quality control

procedures As discussed in Chapter 2, Quest Diagnostics is a laboratory regulated under the

Clinical Laboratories Improvement Act, or CLIA. Laboratory practices must meet or

exceed the requirements set forth in CLIA, and understanding these regulations is a

critical foundation for understanding quality control practices at Quest Diagnostics.

The components of CLIA relevant to quality control are summarized here.

i. Quality control procedures must monitor accuracy and precision of

tests that may be subject to instrument drift, environmental changes,

or variance in instrument or operator performance over time.

ii. Procedures must follow or exceed the number and frequency specified

by the equipment manufacturer subject to the constraints listed below.

1. For quantitative assays (i.e., those that report a concentration

of target compound in the patient sample), quality control

samples must be run at two different concentrations every day

patient testing is performed.

2. For qualitative assays (i.e., those that report either a positive or

negative diagnosis), positive and negative quality control

samples must be run every day patient testing is performed.

iii. The laboratory must establish criteria for accepting quality control

results, and the criteria for quantitative tests must be based on

statistical parameters (e.g., mean and standard deviation) for each

batch.

iv. The laboratory must follow the manufacturer’s specifications for using

reagents and be responsible for the results.

v. The laboratory must document all quality control tests performed.

6.2 Quest Diagnostics quality control

requirements Quest Diagnostics fulfills CLIA requirements by establishing standard operating

procedures (SOPs) for quality control for each type of test performed on the target

platform, as summarized in this section. In addition to meeting CLIA requirements

Page 56: Reducing Reagent Waste Through Process Improvement and

56

and manufacturer specifications, Quest Diagnostics quality control requirements

intend to lead to accurate patient results in a high-volume, rapid-turnaround-time

environment.

The Quest Diagnostics SOPs for the target platform require:

i. For qualitative tests, negative and positive control samples must be

run at the beginning of every day of testing, and positive control

samples must be run intermittently throughout testing and after the

final patient test of the day, thus bracketing each batch of patient

samples between control samples.

ii. Similarly for quantitative tests, the low level (target compound is not

present) and the high level (target compound is present at a clinically

significant concentration) control samples must be run at the beginning

of every day of testing, and high level control samples must be run

intermittently throughout testing and after the final patient test of the

day, again bracketing patient samples between control samples.

iii. Passing negative quality controls must have a negative result, positive

quality controls must have a positive result, and quantitative results

must fall within plus or minus of the identified standard deviations of

the mean for the quality control material batch as tracked by Quest

Diagnostics.

iv. Patient results that are not bracketed by successful quality control

tests must be rerun, and operators must document corrective actions

taken; however, the laboratory director or designated authority may

override the rejection of runs and document the justification.

Quest Diagnostics does not establish policies regarding a preferred type of container

to be used for quality control material, a policy on whether quality control materials

may be reused across multiple batches, or a set of criteria for overriding the rejection

of runs; and CLIA quality control regulations do not require a laboratory to establish

rules in these areas.

Note that because of the mechanics of the testing platform, the instrument always

yields a numeric result representing the factor by which the measured luminesce

exceeds the detection level. Therefore results on this platform of “reactive” (positive)

and “nonreactive” (negative) are supported by a numeric measure. This creates the

Page 57: Reducing Reagent Waste Through Process Improvement and

57

possibility – if approved by the appropriate internal medical experts – of using

quantitative acceptance criteria for qualitative assays, despite the qualitative nature

of the assay.

Maintaining SOPs for each testing area (e.g., hematology, immunology, etc.) falls to

a interdisciplinary teams consisting of experts in medicine, regulations, and

laboratory operations called the Best Practices Team, or BPT.

6.3 Results from quality control survey As discussed in Section 5.1, the four laboratories observed in this study performed

testing consistent with the SOPs; however, because Quest Diagnostics has not

established a standard for all elements of operations, such as a preferred test tube for

quality control testing, laboratory practices differed in subtle but significant ways. A

survey was sent to all laboratories using the target platform to gauge the extent of

variation in quality control frequency, container size, acceptance criteria, reuse

policy, storage method, and approach to mechanical errors. The survey questions are

shown in Section 5.1.1.1.

The results of this survey are as follows:

Table 6.1: Frequency of quality control use

This table shows the fraction of Quest Diagnostics laboratories running the indicated

type of quality control material at the specified frequency. All Quest Diagnostics

laboratories either meet or exceed quality control requirements. The laboratories

exceeding quality control requirements are shown in bold italics.

Testing

frequency

Type of QC

material

Intermittently

throughout

patient testing

Every shift Once per day

Positive/high values 100%

Negative/low values 25% 25% 50%

Recall that the SOPs require positive / high values of quality control intermittently

throughout patient testing and negative / low values of quality control once per day.

The boldfaced values in Table 6.1 indicate the labs with practices consistent with

the SOP. 50% of labs test negative / low values of quality control several times more

often than is required.

Page 58: Reducing Reagent Waste Through Process Improvement and

58

Table 6.2: Quality control material acceptance criteria

This table shows the fraction of laboratories employing the indicated criteria for

determining whether to accept the quality control results.

Acceptance

criteria

% of

labs

Same as test 60%

Quantitative 40%

CLIA regulations set the sample acceptance criteria for quality control results: for

qualitative tests, the positive sample yields a positive result and the negative sample

yields a negative result; and for quantitative tests, the numeric results must fall

within an acceptable range of the established mean value. Standard industry practice

sets the tolerance window according to the Westgard rules, specifically, plus or

minus two standard deviations, as established by historical results, of the accepted

mean. As shown in Table 6.2, 60% of Quest Diagnostics laboratories set their

acceptance criteria (qualitative vs. quantitative) based on whether the test

performed generates qualitative or quantitative results. The remaining 40% use

quantitative criteria for all tests, regardless of the type of result reported, which

exceeds the minimum standards for tracking.

Table 6.3: Reuse policy and container size for quality control material

This table indicates the fractions of laboratories that prepare fresh containers of

quality control materials for every use (“No”), those that reuse a container of

quality control material and dispose of it when only the dead volume remains (“Yes

no repl”), and those that reuse a container of quality control material and replenish

it from a parent container of quality control material when the level gets low (“Yes

w repl”). The results are further broken down by the size of the container used by

those labs, with the dead volume associated with each container shown on the right.

Reuse Policy Dead volume

(uL) Container Size No

Yes no repl

Yes w repl

0.5mL 15% 100

12x75mm 45% 15% 300

13x75mm 5% 5% 300

16x100mm 5% 4500

16x85mm 5% 5% 450

Page 59: Reducing Reagent Waste Through Process Improvement and

59

Quest Diagnostics does not specify a preferred container for quality control material

during testing and laboratories have established a variety of procedures. The choice

of container greatly impacts the amount of “dead volume”, i.e., the quantity of

material that the instrument cannot access and must be discarded after testing.

Quality control materials are also tracked as part of the reagent waste initiative, and

reducing the amount of discarded quality control material would generate savings.

In addition, Quest Diagnostics does not establish a company policy on quality

control material reuse across batches, so each laboratory establishes its own policies.

The choice of container size combines with the laboratory reuse policy to drive large

variations in the amount of quality control material consumed by a lab. For

example, as shown in Table 6.3, the 15% of labs that reuse quality control

materials in a 12x75mm test tube and replenish the container from a master

container will discard about 300uL of quality control material once in several days,

whereas the 5% of labs that prepare a 16x85mm test tube of quality control material

for every batch discard at least 450uL several times per day. The volume of quality

control material varies depending on the test but is always less than 300uL. Thus

the laboratory that reuses and replenishes a quality control container will discard as

little as 1% of the quality control batch, compared with the laboratory that does not

reuse quality control material, which discards over 60% of the quality control

material batch.

The primary concern associated with reusing quality control material is the

possibility of contamination or evaporation leading to incorrect results.

Contamination could affect negative quality control samples, yielding incorrect

results. Evaporation could lead to elevated concentrations of the target compound in

the positive control sample, which could lead to results exceeding control limits for

tests tracked quantitatively. The best ways to prevent both contamination and

evaporation are refrigerating the quality control material and capping the container

between batches, but laboratories vary in their treatment of quality control

materials. Table 6.4 shows the fraction of labs that refrigerate and/or cap the

quality control containers between uses. The values do not sum to 100% because the

25% of labs that do not reuse quality control materials are excluded from this table.

Page 60: Reducing Reagent Waste Through Process Improvement and

60

Table 6.4: Reused quality control material storage policies

This table shows results for only the laboratories that indicated that they reused

quality control materials across multiple uses. The results indicate the fraction of

laboratories refrigerating and/or capping the reused quality control containers

between uses.

Cap

Refrigerate No Yes

No 5% 10%

Yes 5% 55%

Because mechanical errors occur on average once in every 23 samples (4.3% of the

time), occasionally a transient error occurs on a quality control sample. The SOPs

require that labs must successfully test positive quality control samples

intermittently throughout patient testing or else all patient samples from the

affected assay must be repeated. However, the laboratory director may override the

patient retest requirement based on professional judgment. Labs tended to interpret

this rule in one of four ways. When transient errors occurred on quality control

samples, labs would do one of these four things:

1. Address the mechanical error and rerun quality control samples as soon

as possible;

2. Rerun all patient samples;

3. Rerun 5-10 positive patient samples on a separate machine and treat

agreement between the first and second round of results as quality

control; or

4. Choose between options 2 and 3 depending on the type of mechanical

error.

Table 6.5 shows the frequency of these policies within Quest Diagnostics

laboratories. The labs that opted to rerun all patient samples consumed 70-100 times

the reagent in addressing the error as the labs that opted to rerun only the relevant

quality control sample.

Page 61: Reducing Reagent Waste Through Process Improvement and

61

Table 6.5: Approach to mechanical errors on quality control samples

This table shows the fraction of laboratories employing one of four common

approaches to addressing a situation in which a mechanical error occurs during a

quality control sample.

Approach

% of

labs

Repeat QC 40%

Rerun 5-10 samples 15%

Rerun all 25%

Either 5-10 or all 20%

6.4 Tension between quality requirements and

reagent cost Setting quality control policies involves evaluating a range of costs and risks,

including reagent costs and the likelihood of incorrect results under the policy.

Generally a more conservative approach adds greater certainty with respect to the

quality of the patient results, but when this added benefit is marginal, changing to a

less conservative approach is a responsible way to reduce reagent costs. Similarly,

any action that can increase result quality without adding significantly to reagent

costs appears prudent. However, the trade-offs with evenly matched risks and

benefits require a level of industry expertise beyond the scope of this thesis. Any

recommendations presented in this thesis require the review of medical and quality

experts prior to implementation. This section includes a discussion of the risks and

benefits of variations in laboratory quality control practices and either a

recommendation or a deferral to the Best Practice Team.

Frequency:

The appropriate frequency of the negative quality control sample is influenced by (1)

the balance among the cost of quality control testing and the consequences of a

quality control test not meeting acceptance criteria; (2) the ease of identifying

potential instrument issues independent of quality control testing, and (3) the

complexity of testing positive and negative quality control tests with different

frequencies. Overall, these considerations suggest significant reagent cost savings and

few additional operational costs associated with a reduced frequency of negative

quality control testing.

First, we evaluate the trade-offs associated with testing costs. In an extreme

simplification of the diagnostic testing platform, we can model the instrument as a

Page 62: Reducing Reagent Waste Through Process Improvement and

62

system that can generate outputs (quality control results) that may drift out of the

control limits (plus or minus two standard deviations of the mean). For the purposes

of this cost assessment, we assume that if the instrument goes out of control, it

requires operator intervention to return to control. If the instrument does go out of

control, all samples in the previous batch must be reevaluated. In this case, the

expected cost of quality control testing after every batch is as follows:

Ctotal = b * Cneg + b * pneg * Crerun

Ctotal – total cost of quality control in a day

b – number of batches in a day

Cneg – cost of running a single negative quality control test

pneg – probability negative quality control sample exceeds control limits

Crerun – cost of rerunning a batch.

If the laboratory chooses to run the negative quality control every other batch, the

equation changes in three ways. First, the number of quality control tests decreases

by a factor of two. Second, the probability that any quality control test will exceed

the control limits increases by a factor of two because if the instrument goes out of

control during either batch, the quality control result will exceed control limits.

Third, the cost of rerunning the batches will double because every time a quality

control result exceeds the limits, two batches must be reevaluated. Thus the

expected cost of quality control testing becomes the following:

Ctotal = (b / 2) * Cneg + (b / 2) * (pneg * 2) * (Crerun,* 2)

= (b / 2) * Cneg + b * 2 * pneg * Crerun

A generalized form of this for quality control testing performed after any number g

of batches in a grouping is thus as follows:

Ctotal = (b / g) * Cneg + b * g * pneg * Crerun

To minimize the total cost with respect to the number of batches in a grouping, we

can then simply take the derivative of the cost equation and solve for the g where

the derivative equals 0:

dCtotal / dg = - b * Cneg / g2 + b * pneg * Crerun = 0

b * pneg * Crerun = b * Cneg / g2

g2 = (b * Cneg) / (b * pneg * Crerun)

g = sqrt(Cneg / (pneg * Crerun))

Thus the number of batches to be run before quality control testing is large when

the probability of the instrument drifting out of the control limits is small relative to

Page 63: Reducing Reagent Waste Through Process Improvement and

63

the ratio of Cneg/Crerun. Between April and December in one example laboratory, the

negative quality control sample exceeded the control limits for 0.145% of all tests

(0.00145), compared to a Cneg/Crerun ratio of approximately 0.07, an order of

magnitude larger. In this case, the ideal grouping is constrained by the frequency of

testing required by CLIA.

Moreover, the laboratories use an additional technique to identify potential

deviations that is more sensitive than the negative quality control sample. Positive

results occur much less often than negative results, and Quest Diagnostics tracks the

relative occurrence rates for the tests performed on this platform. If the positive

result rate significantly exceeds historical averages, a technologist will review the

entire batch and may rerun it, regardless of successful quality control tests.

Statistically, these average rates are much more likely to identify this malfunction

than a single control sample.

Finally, some laboratory employees expressed concern about a quality control process

in which positive and negative quality control samples were evaluated with different

frequencies because of the added complexity for the operators. This is not of great

concern, given the skill level and general procedural competency of the operators:

they consist primarily of college-educated technologists who are required to interpret

test results with a high level of sophistication. Adding a checklist or creating

separate sample carriers for the first quality control samples of the day – both simple

changes – could reinforce the different processes.

A simple reagent cost optimization model indicates that Quest Diagnostics should

run as few negative quality control samples as possible, given the historical frequency

of negative quality control samples exceeding the control limits. In this model, the

risks associated with this change would likely be small, given the secondary method

of evaluating quality and the high operator skill level. As a result, Quest Diagnostics

laboratories should consider running negative control samples at the minimum

frequency of once per day in all laboratories. This process change must be reviewed

by medical, laboratory, and quality experts within Quest Diagnostics before

implementation.

Acceptance criteria:

The survey results indicate a fairly even split in approaches to quality control

acceptance criteria. Labs that prefer uniformly quantitative methods argue that an

unusually high or low result in a qualitative test may nevertheless indicate a problem

Page 64: Reducing Reagent Waste Through Process Improvement and

64

with the instrument or with the quality control material and should therefore be

investigated.

An alternative approach could harness the benefits of quantitative tracking without

the costs of over-conservative acceptance criteria: labs could record the quantitative

results but use the qualitative results as the acceptance criteria unless there are

additional reasons to mistrust the results. For example, if a high result for a positive

control occurred concurrently with an unusually large number of positive patient

tests, the affected assays could be rerun. Without any indication of impact to patient

results, the offending quality control sample could be rerun alongside a second

quality control sample from a new batch to confirm alignment. A large difference in

result would indicate a problem with initial quality control material rather than the

instrument. Agreement between the samples yielding a normal value may indicate

transient errors affecting the results, so retesting the patient samples from the

affected test would be appropriate. Agreement between the samples yielding a high

value may indicate a need for recalibration, and retesting the patient samples would

again be appropriate. This approach would maintain quality standards by requiring

laboratories to investigate abnormal quality control results but would give them the

flexibility to accept patient results on the basis of qualitative criteria when

investigation does not lead to any reason to suspect the quality of the results. This

flexibility would eliminate the cost and time pressures associated with unnecessarily

running several hours’ worth of patient tests.

Reuse policy:

Reusing quality control materials increases the risk of contamination or evaporation.

However, if these risks are minimized through laboratory practices such as capping

and refrigerating the test tube in between batches, a consistent policy of reusing

quality control materials would reduce reagent costs by reducing dead volume

discarded between tests.

In order to determine the feasibility of reuse, I recommend Quest Diagnostics

perform a side-by-side comparison of laboratories using the three different

approaches: no reuse, reuse without replenishment, and reuse with replenishment.

One example laboratory that followed a policy of no reuse at the time of this thesis

maintains records of the outcome of every quality control test, indicating, for

example, whether the quality control sample exceeded control limits, a mechanical

error interrupted quality control sample processing, or the instrument required

recalibration. Between April and December 2016, this laboratory positive quality

control samples exceeded control limits in 0.132% of tests. Comparable data should

Page 65: Reducing Reagent Waste Through Process Improvement and

65

be obtained from laboratories with the other two reuse policies. If the laboratory

with a no-reuse policy experiences a significantly lower rate of abnormal quality

control results than the other laboratories, then we could conclude that evaporation

or contamination from reuse affects the results of quality control samples. On the

other hand, if there are no significant differences in abnormal quality control results,

we could conclude that reusing quality control materials has negligible impact on

result quality and therefore would offer a low-risk approach to reducing reagent

waste. Identifying the superior approach would allow Quest Diagnostics to improve

laboratory operations, either through improved quality or through reagent cost

savings.

Storage approach:

As discussed in the previous section, the storage approach goes hand in hand with

the reuse policy. Refrigerating and capping are irrelevant if fresh quality control

material is prepared for every use; but with a reuse policy, they are easy ways to

minimize the possibility of contamination and evaporation. The added complexity of

maintaining caps and remembering to refrigerate is relatively small and could be

minimized by using standard test tubes for the quality control material so that test

tube caps are readily available or color-coding the caps to differentiate them.

Container size:

As with storage approach, the best container size depends on the reuse policy.

Without reuse, the best container is the 0.5mL cup used by 10% of labs because it

yields the smallest amount of dead volume – 100uL per sample. With reuse, the best

container is the smallest container that also has a cap, i.e., the 12x75mm test tube.

Making the quality control container uniform across Quest Diagnostics involves

essentially no cost other than the time associated with making the recommendation,

especially because 60% of labs already use the 12x75mm container, as shown in

Table 6.3. With this change, dead volume would consume a factor of 15 less than

the largest container currently in use requires.

Approach to mechanical errors on quality control samples:

Determining the best response to a transient mechanical error on a quality control

sample requires expertise beyond the scope of this thesis. This section will explore

the different approaches observed in Quest Diagnostics laboratories and highlight the

importance of unifying guidance from the Best Practice Team.

As shown in Table 6.5, the labs respond in four different ways, with responses split

fairly evenly:

Page 66: Reducing Reagent Waste Through Process Improvement and

66

1. Rerun quality control samples;

2. Rerun all patient samples;

3. Rerun 5-10 positive patient samples; or

4. Choose between options 2 and 3.

The first option represents the best choice because it requires only one additional

test. However, it is not always possible. For example, if the mechanical error

happens close to the end of the acceptable interval within which a quality control

test must be performed (see Table 6.1 which describes the recommended quality

control test interval), the fastest repeat test might not yield a result until an hour

after the previous quality control result or later because of the extended incubation

time required on the target platform. At this point, the options available include

numbers 2-4 above.

Option 2, rerunning all patient samples, represents the most conservative approach.

This strategy incurs large costs in terms of reagent use, operator time, and increased

turnaround time.

Option 3, rerunning 5-10 confirmatory patient samples, represents a more cost- and

time-sensitive strategy, representing an order of magnitude decrease in the reagent

required compared to Option 2. The 5-10 patient samples, usually with positive

results (i.e., target compound is present) to replicate the positive control, are

retested on a separate instrument. If the results match the original diagnoses, the

agreement could be interpreted as sufficient evidence that the precision of the

instrument is sufficient; and if the patient samples are bracketed by successful

quality control samples, then accuracy of the results can be confirmed as well. To

bolster the argument in favor of this approach, the operator could evaluate the type

of mechanical error based on the information provided in the condition code and

determine whether the error might have affected the outcome of other results. Recall

that in general, transient errors will cause the instrument to jettison a test and

report no result because of the uncertainty in the quality of the result. Therefore

most transient errors would be unlikely to generate inaccurate patient results and

would instead provide no result.

The question of whether repeating 5-10 patient samples is acceptable quality control

process ultimately falls to medical, laboratory, and quality experts within Quest

Diagnostics, who will evaluate the risks and costs associated with such a policy. In

general, creating greater consistency among these quality control practices can lead

to reduced reagent waste or improved reliability of patient testing.

Page 67: Reducing Reagent Waste Through Process Improvement and

67

6.5 Implementing quality control procedure

changes

6.5.1 Changes at an example laboratory

As discussed in the previous section, many of the variations observed in quality

control procedures involve the expertise of the BPT and therefore could not be

implemented during the timeframe of the internship. However, one facility plans to

modify the frequency of negative quality control samples, per the recommendations

above.

One facility evaluated the following considerations before implementing the change.

First, the reagent costs of running a negative quality control sample after every

batch were compared to the reagent costs of running the negative quality control

sample at the minimum required frequency. The potential reagent cost savings were

considered large enough to justify further exploration. Second, the frequency of

negative quality control results exceeding the specified control limits was evaluated.

The low frequency suggested that the additional negative quality control samples did

not provide meaningful information about result quality above and beyond the other

quality control checks performed for instrument results. Third, the impact on the

process of reporting results was evaluated and determined to be minimal. Finally,

the policy change was confirmed by the appropriate medical and laboratory experts

and implemented.

6.5.2 Savings due to changes

The example facility immediately saw a significant decrease in the fraction of reagent

consumed in quality control. The change went into effect mid-October, and Figure

6.1 shows the steep drop in the fraction of reagent used on quality control from 6-

7% in March-June down to 3-4% by November.

Page 68: Reducing Reagent Waste Through Process Improvement and

68

Figure 6.1: Rate of quality control use in example laboratory

The usage counter data showed that the rate of quality control testing decreased

from 6-7% during the four months preceding the project down to 3-4% by November.

Prior to the project the laboratory had been running negative quality control

material every 4 hours of patient testing. Based on the observations of the project,

the lab modified its policies in mid-October to run negative controls daily, consistent

with the standard operating procedures for the platform.

6.5.3 Continuing to align quality control practices

Variations in quality control practices arose on the target platform, and they are

likely to reoccur without measures to maintain alignment. Furthermore, variations

are likely to exist across the many other platforms used in Quest Diagnostics

laboratories, so any procedures to maintain consistency would benefit multiple areas

within the lab. In order to sustain consistent practices and continue to improve

quality control processes, I recommend the following:

1. Continue identifying non-uniformities in laboratory operating

procedures;

2. Provide more incentives for ongoing improvements; and

3. Update SOPs regularly with new best practices.

This thesis demonstrates the value of comparative analysis across Quest Diagnostics

laboratories. Quest Diagnostics should continue to perform this type of comparative

0%

1%

2%

3%

4%

5%

6%

7%

8%

9%

Dec Feb Apr Jun Aug Oct Dec

Instrument 1 Instrument 2 Instrument 3

Page 69: Reducing Reagent Waste Through Process Improvement and

69

evaluation to identify variations that arise in quality control practices on the target

platform and other platforms.

Improvements beyond the status quo come from the front lines. . Potential

innovations from the people who perform the task every day spark a healthy

reexamination of current processes. Quest Diagnostics should foster ongoing process

improvement. Quest Diagnostics should foster ongoing process improvement. For

example, the company could provide financial incentives to cost savings measures

that adhere to company quality standards and that the BPT subsequently approves.

As individuals generate process improvements, the BPT will need to update the SOP

on a regular basis. The BPT can include these updates as part of their regular

review process.

Together, these recommendations will help Quest Diagnostics align laboratory

practices around common best practices that continue to drive savings and

improvements in quality.

6.6 Chapter summary

The previous chapter identified quality control as one of the two primary sources of

reagent waste within Quest Diagnostics. Laboratory observation and surveys reveal

that laboratories across Quest Diagnostics take subtly different approaches to quality

control practices, leading to significant variations in the amount of quality control

material consumed at each site. The primary sources of variation are quality control

frequency, container size, acceptance criteria, reuse policy, storage method, and

approach to mechanical errors.

The optimal approach may be identified using existing data. For example, we

determined that the example laboratory could decrease the frequency of running

negative quality control samples without significantly increasing the cost of repeat

testing and while remaining in compliance with manufacturer guidance and CLIA

regulations. Similar intra- or inter-laboratory evaluations could determine the

effectiveness of quality control reuse, for example.

Achieving greater uniformity in operating procedures can help Quest Diagnostics

reduce reagent costs or improve reliability of patient testing. The company may be

able to reduce variations on an ongoing basis through the following

recommendations:

Page 70: Reducing Reagent Waste Through Process Improvement and

70

1. Continue identifying non-uniformities in laboratory operating

procedures;

2. Provide more incentives for ongoing improvements; and

3. Update SOPs regularly with new best practices.

The next chapter discusses the countermeasures for the second major source of

waste: mechanical errors.

Page 71: Reducing Reagent Waste Through Process Improvement and

71

Chapter 7 Countermeasures to eliminate mechanical error waste As discussed in Chapter 5, the two primary sources of reagent waste were quality

control and mechanical errors. Chapter 6 discussed strategies to reduce the former,

and this chapter will discuss strategies for the latter.

The current state analysis in Chapter 5 identified that operator skill levels varied,

that laboratories did not necessarily share best practices, and that mechanical errors

may be reduced through a maintenance program that goes above and beyond

manufacturer requirements. Thus the goals of this part of the project were to

standardize and elevate operator skill levels and to create maintenance standards

that exceeded manufacturer requirements. This chapter discusses the CLIA

regulatory framework within which these modifications can be made.

The author launched an Autonomous Maintenance (AM) pilot project within the

Marlborough laboratory in order to achieve the goals of this section. The AM team

completed Steps 0-2 of an AM program, creating several training documents related

to best practices and generating a set of trial maintenance procedures. The training

materials will be circulated within the company to achieve more consistent operator

skill levels. The trial maintenance procedures did not show a reduction in mechanical

errors because of the relatively short period of implementation.

Like the previous chapter, this chapter ultimately highlights the need for ongoing

incorporation of best practices into SOPs and training materials, and for incentives

related to process improvement.

7.1 Goals for AM pilot program The goal of this section of the project was to reduce reagent waste from mechanical

errors. Chapter 5 discussed a few key observations related to instrument

mechanical performance. Specifically, operator performance varied within and across

labs, the vast majority of mechanical errors occurred on four instrument subsystems,

Page 72: Reducing Reagent Waste Through Process Improvement and

72

and the most common of these errors could be addressed through maintenance above

the manufacturer’s recommended maintenance program. Together, these

observations suggest two strategies for tackling mechanical errors.

1. Increase the quantity or quality of information about best maintenance

practices provided to operators to ensure uniform performance of

maintenance; and

2. Develop additional cleaning and inspection procedures above and

beyond standard maintenance procedures to reduce the most frequent

mechanical errors.

As with the quality control standards, the countermeasures for mechanical errors

were developed with the intention of disseminating the results at a national scale to

maximize the impact of the proposed changes. Because these goals aligned with the

goals of Autonomous Maintenance, as discussed in Chapter 4, this project used the

structure of Autonomous Maintenance in order to address mechanical errors.

7.2 CLIA regulations related to operation,

maintenance, and equipment modifications CLIA regulations were less salient in tackling mechanical errors than in resolving

variations in quality control processes because the challenge here related more to the

mechanics of improving instrument performance than to varying interpretations of

regulatory requirements. Nevertheless, CLIA requirements form an important

foundation for our approach.

The relevant elements of CLIA requirements for equipment maintenance are as

follows:

- For equipment unmodified from the manufacturer’s original design, the

laboratory must perform maintenance as specified by the manufacturer

and with at least the frequency specified by the manufacturer. (493 CFR

1254.a)

- Quality control testing must be performed after any major preventive

maintenance or after replacement of any critical part that may influence

test performance. (493 CFR 1256.6)

- Laboratory personnel must document all maintenance performed. (493

CFR 1425.3)

Page 73: Reducing Reagent Waste Through Process Improvement and

73

- The laboratory must document evidence that operators have the skills

required to perform preventive maintenance and troubleshooting. (493

CFR 1423.4.ii.D)

The above regulations influence Quest Diagnostics maintenance procedures in the

following ways. First, the target platform, as installed in Quest Diagnostics

laboratories, remains in its original and unmodified form. As a result, any additional

maintenance activities must be performed with the guidance of the manufacturer.

Second, significant preventive maintenance activities trigger a round of quality

control testing. Ideally, then, preventive maintenance should occur during normal

breaks between rounds of quality control to minimize the amount of quality control

waste generated from maintenance activities. Third, supplementary documentation

must accompany any additional maintenance performed. Fourth, new training

materials must come with a mechanism for ensuring operator comprehension. The

countermeasures discussed in this chapter work within the guidelines set forth above.

7.3 AM team structure and activities To tackle waste from mechanical errors, this project launched an Autonomous

Maintenance (AM) program within the Marlborough laboratory focused on the

target platform. AM was an attractive approach for two reasons. First, the company

had recently hired an expert in Total Productive Maintenance (TPM) with the

intention of incorporating TPM into the company’s maintenance processes and

continuous improvement efforts. Therefore the expertise required for a successful

launch was readily at hand. Second, the strategies of the AM approach aligned with

the goals of this project, specifically, to increase operator skill levels and to develop

additional maintenance procedures.

This section discusses the structure of the team and activities within the AM

program.

7.3.1 Team structure

As discussed in Chapter 3, the recommended organization involves overlapping

small groups throughout the company: groups of 5-7 individuals focus on a particular

platform or area of the work space, and the leader of any group also participates in a

group organized at the next higher level of the organization. In other words, the

recommended structure for the AM team would consist of 5-7 operators focused on

the target platform, and the operator designated as the AM team leader would also

be part of a supervisors’ and managers’ AM team focused on a different platform.

Page 74: Reducing Reagent Waste Through Process Improvement and

74

However, resource constraints resulted in an AM team with one operator, a rotating

group of vendor representatives, and the author as the team leader, and no TPM

organization representative beyond the working group. Section 7.5 discusses the

impact of these limitations. The AM team met weekly for nine weeks to perform

activities that required access to the instrument. Outside of the weekly meetings,

team members performed additional activities related to team goals, as discussed in

the following section.

7.3.2 Autonomous Maintenance activities

The components of an Autonomous Maintenance program are well documented and

well characterized. The Quest AM team followed several of the procedures for Steps

0, 1, and 2 of Autonomous Maintenance with the goal of reducing mechanical error

frequency, as discussed in this section.

Step 0: Education

During Step 0, the goal of the AM team is to lay the foundation for subsequent

activities. The activities consist of establishing the structure of the AM team,

learning about the tools of AM, evaluating the current state of the instrument, and

developing a deeper understanding of how the instrument works, each of which will

be discussed in this section.

Establishing AM team structure: The first AM team meeting established individual

roles and responsibilities, guidelines and restrictions, and norms. For each type of

activity within a given AM step, the team members each took on responsibility of

maintaining that particular aspect, for example, collecting the supplies needed for

the group activity or performing safety audits before beginning work. The guidelines

and restrictions established the scope of activities. To minimize disruption to

production schedules and to maintain continuity across team meetings, we only

worked on a single instrument within the Marlborough lab with the assumption that

any findings on that instrument would be generalizable to other instruments and

other labs. We also set the boundaries of the type of activities that could be

performed based on Quest quality standards, e.g., that our activities could not

compromise the quality of patient results or impact compliance with CLIA

regulations, e.g., that the manufacturer had to approve any changes to maintenance

procedures. Finally, the team norms established expectations for how the team

would operate, such as participating actively, meeting deadlines, and fulfilling the

established responsibilities.

Page 75: Reducing Reagent Waste Through Process Improvement and

75

Learning tools of AM: This stage involved familiarizing the team with three tools

critical to AM: the principles of 5S, Quick Kaizens, and One-Point Lessons (OPLs).

The 5S principles are intended to empower operators to eliminate wasted energy,

time, and materials in their work. After receiving training from an expert in the

principles of 5S, the AM team applied those principles to create a standardized cart

of maintenance supplies to streamline the maintenance process.

The AM team leader trained the rest of the team on Quick Kaizens – improvements

to the efficiency, safety, or quality of current operations – and OPLs – concise

training documents that share (1) basic knowledge, (2) troubleshooting tips, or (3)

improvement ideas. Because these documents are designed to capture and share best

practices, the AM team identified them as a primary mechanism for sharing the

knowledge gained through AM activities and for disseminating best practices learned

at other sites. An example of an OPL is shown in Figure 7.1.

Figure 7.1: Example One-Point Lesson

Operators wash incubator rings as a part of routine maintenance. This lesson

demonstrates proper technique for drying the rings in order to avoid leaving behind

fibers, which can cause incubator jams.

Page 76: Reducing Reagent Waste Through Process Improvement and

76

Measuring baseline equipment condition: Before we could measure our progress, we

had to know where we were starting. We began by coordinating with the

manufacturer to perform a condition audit of the instrument. The AM team

established an audit checklist for all functional components of the instrument, and

an engineer from the manufacturer evaluated the instrument based on those criteria.

The audit revealed a small number of components that were not performing at

optimal levels and required separate inspection or replacement. The audit also

identified areas of the instrument where current maintenance practices as specified in

manufacturer instructions were not sufficient to prevent debris from accumulating to

levels that could potentially impact instrument performance.

Quantifying baseline instrument performance: The primary metric AM uses to

measure performance is Operational Equipment Effectiveness (OEE), as discussed in

Chapter 3. The metric is a product of Availability, Performance, and Quality,

where the capitalization is intended to denote the AM definitions for each of these

terms. Availability represents the amount of time the instrument operates relative to

the amount of time allocated for the instrument to operate. Performance represents

the number of samples processed during that time compared to the instrument’s theoretical throughput. Quality represents the fraction of successful jobs performed

during that time.

We were unable to identify reliable data sources for a reliable calculation of OEE

and therefore relied on other sources of information in order to measure progress.

Specifically, no data sources existed that could easily calculate Availability; and

Performance and Quality metrics were not available at the daily timescale most

useful for OEE. The AM team tested tracking Availability by asking operators to

record the amount of downtime for a target instrument, but this did not prove

successful: because of their large workload, operators struggled to record consistent,

accurate data about the duration of instrument downtime and its causes. Daily data

about Performance and Quality was only available through a labor-intensive process

of collection and processing, which we determined to be unsustainable for the AM

team and for the operators.

Therefore instead of looking at the holistic measure of OEE, we used two of the data

sources described in Chapter 5 to measure instrument performance. First, the

usage counter data substituted for a Quality metric at a weekly scale, showing the

fraction of tests used for patient results. Second, the condition code data replaced a

Performance metric by showing the number of breakdowns over the course of a

week. While the condition code data did not provide information about the amount

Page 77: Reducing Reagent Waste Through Process Improvement and

77

of downtime due to mechanical errors, the more important metric for the purposes of

this thesis was the impact on reagent waste, which could be inferred from the data.

Learning instrument mechanics and safety: The final components of Step 0, learning

the instrument mechanics and establishing safety guidelines, laid the foundation for

subsequent team activities. With the guidance of vendor engineers, the AM team

first established a safety checklist for the instrument and the surrounding area to

guarantee that we could perform work on the instrument with no risks to our health

and safety and no risks of damaging the equipment. Parts of the instrument with

inherent risks were identified as a potential target for Quick Kaizens or, if the AM

team could not eliminate the risk, as a point of design feedback for the

manufacturer. The AM team reviewed the safety checklist before performing any

activities.

Then the AM team discussed the mechanics of every part of the instrument.

Following standard AM practice, the team forced themselves to evaluate every

component by creating detailed sketches of the instrument, which promoted

attention to detail and a methodical review of components.

Step 1: Clean to inspect

During Step 1, the goal of the AM team is to expose hidden defects within the

instrument in anticipation of updating cleaning and inspection standards in Step 2.

Activities in Step 1 include cleaning and inspecting the instrument over time;

identifying and, where possible, eliminating areas of the instrument that are difficult

for the operator to access; and tracking any Sources of Contamination (SOCs). Hard

to access areas can be eliminated by making the areas more accessible or identifying

procedure changes to prevent contamination from reaching the area, thereby

precluding the need for access. Note that in this context, the capitalization of SOCs

is intended to denote the AM-specific definition of contamination, which is simply

any substance found where it is not intended, and not to connote any danger for

contaminating samples.

Cleaning to inspect: Just as drawing helped the AM team more deeply investigate

different instrument components, manually cleaning every instrument component

encouraged the AM team to thoroughly review the current instrument condition.

The AM team performed this activity for three weeks. The first cleaning established

a spotless baseline condition so that any Contamination that accumulated over the

course of the week could be easily identified and traced to its source. The approach

follows the same logic as a car owner who wants to keep her driveway clean to detect

Page 78: Reducing Reagent Waste Through Process Improvement and

78

oil leaks: if the driveway is dirty and covered with oil splatters, she will struggle to

notice any new drops, but if the driveway starts out clean, she can easily detect any

new drops and address the oil leak immediately.

Identifying hard-to-access areas: In anticipation of a Step 2 goal of reducing the total

time required for maintenance, Step 1 involves identifying areas of the instrument

that are difficult, uncomfortable, or unpleasant to access. Examples include parts of

the instrument that are difficult to clean, such as a shelf where disposed consumables

accumulate that is located at knee height and is the length of an arm, so, without

modifications to the instrument or tools used, cleaning the shelf requires the operator

to lean over or kneel down and insert an arm into the space; or information stored in

the instrument that is not readily accessible through the user interface. This activity

involved tracking these hard-to-access areas. When doing so did not involve an

unacceptable or infeasible modification of the instrument, the AM team developed

countermeasures to improve accessibility and documented the actions in the form of

Quick Kaizens or OPLs. Otherwise, the observations were submitted to the

manufacturer as potential targets for design changes.

Tracking Sources of Contamination: The process of cleaning to inspect the

instrument in this stage then allowed the AM team to track Sources of

Contamination. During the initial cleaning, the AM team recorded all debris

observed before cleaning it. During subsequent cleaning, the AM team noted whether

the debris had recurred. If not, then the debris was considered a one-time occurrence

that did not require additional maintenance measures to address. An example of this

type of debris is dust on horizontal, non-functional surfaces in the instrument, which

had accumulated since installation but would not have any impact on instrument

performance. In this instance, because a small amount of dust had accumulated over

a relatively long time, no countermeasures were necessary. If the debris appeared on

subsequent AM team sessions, however, a root cause analysis was required.

The root cause analysis allowed the AM team to categorize the debris as either

normal or abnormal. Consider, for example, debris that accumulates under a moving

component on the instrument, and the engineering representative from the

manufacturer identifies that the debris comes from friction between moving parts. If

the parts are designed to create friction, then this debris would be considered a

normal SOC, and an adequate countermeasure would involve monitoring the debris

and cleaning it regularly. If, on the other hand, the parts were not designed to wear,

the debris would indicate an abnormal SOC. Abnormal SOCs provide evidence of a

hidden defect, which requires immediate resolution. In this instance, the hidden

Page 79: Reducing Reagent Waste Through Process Improvement and

79

defect would be a problem with component alignment that would trigger remedial

action from a service engineer provided by the manufacturer. Following service, the

AM team would recommend that maintenance procedures also include checking for

this debris and triggering service if debris is observed, thereby establishing a process

for rapidly identifying the defect if it returns.

For every SOC observed, this process identifies:

1. A description of the debris and its location within the instrument;

2. A description of the root cause of the problem; and

3. A solution for how to address it going forward.

Step 2: Tentative Cleaning and Inspection Standards

During this step, the AM team synthesized the understanding of the instrument and

the observations of SOCs to develop cleaning standards that go above and beyond

the standard manufacturer recommendations. Traditionally this stage also includes a

goal of decreasing the total operator time spent on maintenance to ¡2% of the total

labor time. While the AM team could reduce the total amount of debris

accumulation through eliminating hidden defects and addressing hard-to-access

areas, we could not eliminate cleaning steps that are part of the vendor’s validated

process for performing instrument maintenance.

Over the four weeks available for this step, the AM team developed supplemental

cleaning standards. We prioritized improving maintenance for instrument subsystems

with the highest number of mechanical errors, as determined using the condition

code data. To address the mechanical errors, we combined an assessment of the most

frequent condition codes for that subsystem with observations of debris accumulation

inside the instrument to identify the specific component or process that the cleaning

should target. After developing the cleaning standards and receiving manufacturer

approval to proceed with them, we trained all operators responsible for maintenance

on the new procedures and, in compliance with CLIA regulations, added

documentation of the new maintenance steps to the standard maintenance tracking

system.

The new maintenance standards applied only to the instrument that was the focus of

the AM team’s activities to allow direct comparison of performance between the

“test” instrument, where the new standards were applied, and the “control” instrument, where only the original maintenance procedures were used.

Page 80: Reducing Reagent Waste Through Process Improvement and

80

7.4 Outcomes

7.4.1 Impact on instrument performance at

Marlborough

The goal of Step 2 was to develop cleaning and inspection standards that improved

instrument performance. Unfortunately, both usage counter data and condition code

data indicate that the AM team did not accomplish that goal, as shown in Figure

7.2. In retrospect, this outcome is unsurprising given the short timeframe available

to design and implement potential changes.

Figure 7.2: Rate of mechanical errors in Marlborough

The usage counter data did not indicate that the frequency of mechanical errors on

the test platform, Instrument 3, decreased significantly after the beginning of the

AM program: note the results for October and November are comparable to those of

the preceding months, except February, which experienced an abnormally high rate

of errors for reasons beyond the scope of this thesis.

Because the maintenance countermeasures were not having a significant positive

impact on instrument performance during the final week of the pilot program, a

longer-term approach was established in order to continue the process of improving

maintenance. While the AM team meetings could not continue absent the AM team

leader, a smaller working group was arranged: the operator representative and the

operator from the AM team allocate an hour each week to work on ongoing process

0% 1% 2% 3% 4% 5% 6% 7% 8% 9%

10%

Jan Apr Jun Aug Oct Dec

Instrument 1 Instrument 2 Instrument 3

Page 81: Reducing Reagent Waste Through Process Improvement and

81

improvement activities. They use this time to review the condition code data from

the prior week, discuss with vendor representatives how to address the most common

mechanical errors, and train other operators on the applicable modifications to the

maintenance procedures. The results of these efforts cannot be quantified at this

point because the improvement activities are ongoing.

7.4.2 Description of training materials and how

training materials were disseminated

While the tentative cleaning and inspection standards were not able to demonstrate

improvement in timeframe of the thesis, the AM activities had the positive effect of

increasing operator knowledge of the instrument.

Over the course of the AM sessions, the team collectively prepared 29 OPLs, which

were approved for general circulation by the vendor representatives. The OPLs will

be circulated through a national competency website. Individual laboratory managers

will be responsible for identifying personnel within their laboratory who are certified

to run the target platform and adding those operators to a central database. Then a

representative from the BPT can easily distribute OPLs to all operators trained on

the target platform over several weeks. The OPLs will be associated with

comprehension questions, which will verify that the operators absorbed the

information captured in the OPLs.

This process will allow widespread dissemination of the best practices captured

through the AM team meetings. Moreover, it creates a pathway through which sites

can share best practices related to instrument operation or maintenance. As

discussed in Chapter 5, one of the main preliminary observations from this project

was that labs did not regularly share best practices. By establishing the national

competency testing website as the means of communication and OPLs as the format,

this process will demonstrate to labs how these best practices can be shared and

encourage them to follow suit.

7.5 Discussion of implementation approach

7.5.1 Challenge of improving processes in a lab with

high productivity goals

A primary constraint for the AM team was the amount of time the operator on the

team could dedicate to the group activities. Because the laboratory faced increasing

pressure to maintain productivity and meet turnaround time requirements the

operator was usually responsible for running an instrument or set of instruments

Page 82: Reducing Reagent Waste Through Process Improvement and

82

concurrently with AM team meetings. Moreover, the operator did not have dedicated

time outside of the team meetings to explore improvement activities, such as

developing OPLs or implementing Quick Kaizens.

There are several potential strategies for addressing this type of constraint.

1. Ask the operator to “make time” during normal working hours;

2. Provide overtime compensation for process improvement activities;

or

3. Include process improvement as part of job expectations.

Consider the first option. Every job has stretches of time when the worker can

operate at a slower pace. One option for addressing the time limitations would

involve asking the operator to identify these slower moments as opportunities to

work on separate process improvement activities. However, given the high

expectations for work quality and turnaround time, adding to additional

responsibilities is likely to lead to operator burnout.

The second option decreases the burden on the operator during working hours by

providing overtime compensation for process improvements completed outside of

normal working hours. While this eliminates the stress associated with the multi-

tasking the first approach requires, it nevertheless adds strain on the employee. An

employee with significant demands outside of work, such as a child or an ailing

family member, would have significantly less latitude to contribute to process

improvements than ideal. The Marlborough lab followed this approach during the

AM pilot program, and the operator found time for the extracurricular work only on

occasion, despite the compensation incentive.

The final approach would involve including process improvement activities as part of

explicit job requirements for operators and allocating time during standard working

hours for these activities. This approach is likely to decrease the apparent

productivity in the short term as employees spend less of the total week performing

their primary front-line tasks. However, the time spent improving processes,

increasing safety, or strengthening training materials represents an investment that,

when implemented correctly, are likely to pay dividends in more efficient operations,

lower productive hours lost to workplace injuries, or lower rates of mechanical

failures.

Managerial support can be achieved by demonstrating the potential positive impact

of “quick wins”, i.e., improvement activities that provide a rapid return on the

Page 83: Reducing Reagent Waste Through Process Improvement and

83

investment of time, under either of the first two regimes as a way of justifying

additional investment. While this approach may require greater investment on the

part of company management, the returns are likely to outweigh what they would be

under the first two scenarios without the potential employee burnout likely under

those approaches.

7.5.2 Strategy for sustaining improvements

Just as with quality control processes, the gains from this investigation of mechanical

errors are best secured through ongoing comparative analysis across sites and

through incentives to contribute process improvement ideas, here in the form of

OPLs.

Comparative analysis: As observed in the sites visited, individual sites will continue

to develop innovative solutions to tackle process challenges. Continued comparative

analysis across sites will help to unite laboratory operations around a single best

practice.

Process improvement incentives: Ultimately, the most efficient process would involve

individual laboratories broadly communicating when they develop a process

improvement. For example, at the time of implementation when a laboratory group

communicates the process improvement internally, it could circulate the

improvement to the comparable groups in other laboratories within Quest

Diagnostics. In this paradigm, the labs self-identify when they have made

improvements and submit their improvements in the form of OPLs to some

approving body, such as the BPT, after receiving approval from the manufacturer.

Quest could begin the transition to this type of process immediately by creating

financial incentives for individuals or laboratories for developing OPLs that get

incorporated into the centralized database. This type of incentive would encourage

labs across the company to invest more in the long-term process improvements that

pay dividends across the company, rather than focusing exclusively on cutting

headcount and reducing waste.

7.6 Chapter summary As discussed in Chapter 5, mechanical errors account for approximately 4.3% of all

reagent use in Quest Diagnostics laboratories, all of which is considered waste for

purposes of this thesis. Many of those mechanical errors arise because of debris

accumulation at various subsystems within the instrument and therefore may be

addressable through more frequent or more intensive maintenance beyond what is

Page 84: Reducing Reagent Waste Through Process Improvement and

84

required by the manufacturer. The AM team attempted to develop such cleaning

standards that go above and beyond manufacturer requirements. While the AM

team was not able to demonstrate significant reductions in mechanical errors over

the course of the thesis, the approach established a good model for continuing efforts

to improve maintenance practices.

The secondary goal of the AM team activities was to standardize operations around

best practices. This motivation arises from the observation discussed in Chapter 5

that individual labs develop innovative solutions to operational challenges but do not

have a clear channel for disseminating these innovations to other laboratories.

Through the OPLs, the AM team documented best practices identified by individual

labs in conjunction with the manufacturer. By sharing the OPLs on a national

competency platform, Quest Diagnostics can more effectively disseminate best

practices of operators across the company and unify operations around these

innovative best practices.

The greatest challenge the AM team encountered was the constraints associated with

operator availability during the thesis. This challenge is the result of a different

component of the Invigorate initiative that emphasizes increasing employee

productivity instead of reducing reagent waste. In order to balance the potentially

conflicting goals of improving processes and increasing productivity, Quest

Diagnostics could include process improvement activities as part of employee job

expectations, thereby creating a structure for managerial support of these activities;

and they should offer incentives for improvement ideas to further inspire innovation.

Page 85: Reducing Reagent Waste Through Process Improvement and

85

Chapter 8 Conclusions This thesis aims to reduce waste on a high-volume diagnostic testing platform at

Quest Diagnostics. In this concluding chapter, we review the main findings of the

project, the countermeasures implemented, the recommended follow-up actions for

the company, and the most fruitful areas of future research.

8.1 Summary of main findings This section will briefly review the primary findings discussed in the previous

chapters.

1. Quality control and mechanical errors were largest sources of waste.

Data provided by the manufacturer indicated that quality control

samples and incomplete tests accounted for 5.2% and 4.3% of all tests,

respectively, from March through May 2016 across all Quest

Diagnostics laboratories. These represented the main sources of reagent

waste for the target platform.

2. Individual laboratories implemented quality control practices that

exceeded Quest Diagnostics operating procedures and regulatory

requirements

Survey results from all laboratories using the target platform indicated

that laboratories varied in their implementation of standard operating

procedures (SOPs). The variations included differences in quality

control frequency, container size, acceptance criteria, reuse policy, and

approach to mechanical errors.

The different approaches fell along a cost-conservatism spectrum that

universally met SOP requirements but occasionally exceeded those

requirements without substantial benefit to test quality. For example,

several laboratories ran negative/non-reactive quality control samples

more often than required, which, other laboratories argued, provided

Page 86: Reducing Reagent Waste Through Process Improvement and

86

little meaningful information about the accuracy of the instrument

results.

3. Mechanical errors were a major source of irritation for operators

The observed rate of mechanical errors may understate the impact of

these errors on the operator experience of the instrument. Anecdotally,

operators reported that the instrument experienced a relatively high

rate of errors, including occasional shutdown errors that generally

compromised all of the tests currently in progress.

4. The most common mechanical errors may be addressed through

improved maintenance.

Many of the most common mechanical errors arose because of

accumulation of debris that interrupted normal operation of the

instrument component. Thus increasing frequency or intensity of

maintenance beyond the manufacturer’s recommendation may reduce

the frequency of these mechanical errors. This observation suggested

that a program of Autonomous Maintenance could benefit Quest

Diagnostics by creating an independent team dedicated to improving

maintenance practices.

5. Laboratories do not share best practices related to maintenance or

process improvements.

Subtle differences between individual operators’ practices and between

overall lab practices related to maintenance and operation suggested a

lack of knowledge sharing both within and among laboratories. This

observation bolstered the argument for an Autonomous Maintenance

program, which involves creating training documents to capture and

disseminate best practices.

8.2 Recommendations for Quest Diagnostics This project developed countermeasures to address the opportunities that the above

observations revealed. This section discusses three sets of recommendations based on

the project analysis: (1) the actions Quest Diagnostics must take in order to lock in

the opportunities identified through this project, (2) the organizational changes

Quest Diagnostic should make to build from the learnings in this project, and (3) the

technical changes to data collection that will facilitate process improvement going

forward.

Page 87: Reducing Reagent Waste Through Process Improvement and

87

8.2.1 Specific process/decision-based

recommendations

The assessment of sources of reagent waste lead to two concrete recommendations to

address the immediate causes:

1. Add specificity to quality control procedures

2. Share best operation and maintenance practices across the company

The first recommendation will create greater consistency among laboratories in their

interpretation of quality control procedures. Depending on the best practice

ultimately determined by internal medical, laboratory, and quality experts,

standardizing these procedure beyond the level currently specified in SOPs may also

reduce reagent costs by avoiding over-conservative quality control practices that

provide little meaningful insight into result quality. The second recommendation will

reduce mechanical errors by elevating instrument operation to a consistently high

level.

Add specificity to quality control procedures

The most reliable way to unify laboratory quality control practices for the target

platform involves the Best Practice Team incorporating the recommended

approaches into SOPs for the platform.

Share best operation and maintenance practices across the company

This project began documenting and circulating best practices related to operation

and maintenance of the target instrument. In order to reap the full benefit of these

materials, Quest Diagnostics should circulate the training documents to everyone

involved in operating the instruments. The Best Practice Team is a natural

authority to oversee this training.

The One-Point Lessons capture many best practices, but operators will continue to

develop new ideas for improving processes on the target platform as well as other

platforms. The BPTs for different testing areas should encourage operators to

continue developing OPLs to be included in operator training requirements.

8.2.2 Organizational recommendations

This project highlights the opportunities related to aligning procedures and the

challenges of implementing change on one testing platform. To capture those

Page 88: Reducing Reagent Waste Through Process Improvement and

88

opportunities and overcome those challenges, Quest Diagnostics should make four

general organizational changes:

1. Continue performing comparative analysis

2. Establish process improvement objectives on par with clinical goals

3. Create incentives for problem-solving

4. Regularly update SOPs with best practices

Continue performing comparative analysis

Through four laboratory visits, this project revealed variations between laboratories

that could lead to substantial cost and reagent savings. Comparable variations

probably exist on other testing platforms, and Quest Diagnostics should continue the

type of comparative analysis performed in this thesis to identify and minimize these

variations.

Establish process improvement objectives on par with clinical goals

A primary challenge for the Autonomous Maintenance team activities arose because

of the higher priority placed on the clinical goals of turnaround times for patient

results and employee productivity despite top management’s support for process

improvement activities. A solution involves holding laboratory managers accountable

to process improvement goals in addition to their traditional goals. This would

encourage middle management to create more space for improvement activities. It

would also provide middle management with leverage against increasing productivity

requirements when those requirements compromise the process improvement goals.

Create incentives for problem solving

As observed in this project, operators have many ideas about process improvements

that can improve working conditions, increase productivity, or cut costs, all of which

create value for Quest Diagnostics. However, these local improvements do not

consistently spread to all operators within a laboratory or to all laboratories.

To encourage greater dissemination of these improvement ideas, Quest Diagnostics

should establish employee incentives for approved ideas. The process might proceed

as follows: the employee drafts a One-Point Lesson based on the proposed

improvement; the BPT reviews the OPL in the context of existing OPLs; if the BPT

approves the OPL, the lesson is incorporated to operator competency training, and

the creator of the OPL receives a cash bonus. Such a process would harness

employee creativity and benefit the company at a national scale at relatively low

cost.

Page 89: Reducing Reagent Waste Through Process Improvement and

89

Regularly update SOPs with best practices

The SOPs enshrine process requirements for all laboratories. In order to guarantee

that all laboratories follow the current best practices, the BPT for clinical testing

areas should incorporate newly identified best practices into SOPs as part of their

regular SOP review procedure.

8.2.3 Data collection recommendations

The improvement initiatives discussed in this analysis, especially those related to

reducing the rate of mechanical errors, rely on consistent, accurate data. Increasing

data precision and ease of access will help inform ongoing process improvement

activities and facilitate better decisions about how to make improvements. In fact,

the current lack of timely, relevant data will quickly become a roadblock as

operators try to improve maintenance processes.

To remove this barrier, we recommend that Quest Diagnostics prioritize generating

easy access to the following types of data in collaboration with the instrument

manufacturer:

1. Daily OEE data; and

2. Definitive mapping between condition codes and lost reagent.

Daily OEE data

An Autonomous Maintenance program cannot reach its full potential without daily

Operational Equipment Effectiveness data, which this project lacked. Each

component – Availability, Performance, and Quality – proved impossible or highly

labor-intensive to access on a daily basis. Given the time constraints already placed

on operators, any data collection step requiring more than a few minutes is

unsustainable.

Quest Diagnostics should therefore prioritize generating easy, daily OEE data. This

task will require collaboration with the instrument manufacturer, who has already

built a relationship with several laboratory personnel over the course of the AM pilot

project. While the manufacturer’s current data sources do not accommodate an OEE

calculation, there were several types of data that approached the correct information.

For example, the usage counter information was available on a weekly basis, but not

on a daily basis because of how frequently the instrument, which stored test-by-test

usage information, updated the manufacturer’s servers with new data. In this case, a

small software change could easily lead to daily Quality data.

Page 90: Reducing Reagent Waste Through Process Improvement and

90

Definitive mapping between condition codes and lost reagent

One key challenge of this project related to the difficulty of identifying which

condition codes that arose on the instrument led to incomplete tests. This project

relied in part on code-by-code input from vendor engineers and in part on

generalized observations about what made a condition code more likely to lead to an

incomplete result. Neither approach provided adequate consistency, ease, or

specificity for consistent process improvement of this type.

Unfortunately, the onus of this change lies again with the manufacturer’s software

team. The instrument software must be modified to correlate an incomplete result,

which is recorded in the usage counter data, with a mechanical cause.

8.3 Areas of further investigation In addition to the analyses of quality control procedures discussed in Section 8.2.1,

the above analysis suggests two follow-up areas of investigation:

1. Maintenance prevention through design improvements

2. Repeat comparative analysis on other instruments

Maintenance prevention through design improvements:

The observation that 89.7% of impactful mechanical errors arose from three

subsystems suggests that those subsystems are relatively prone to failure. While the

AM activities discussed in this project aimed to reduce the rate of mechanical errors

through additional maintenance procedures, a more efficient approach would involve

designing more robust components.

Comparative analysis for additional platforms:

The analysis of this instrument yielded many insights into potential improvements.

This type of analysis could be repeated across other platforms to generate

comparable reagent savings.

Together, these recommendations will allow Quest Diagnostics to address the

immediate causes of reagent waste, perform analyses to further improve

instrument performance and effectiveness of maintenance, and foster a culture of

continuous improvement.

Page 91: Reducing Reagent Waste Through Process Improvement and

91

Bibliography

[1] Quest Diagnostics, “2015 Annual Report on Form 10-K.” 2016.

[2] B. Wang, “Medical Equipment Maintenance: Management and Oversight,”

Synthesis Lectures on Biomedical Engineering, vol. 7, no. 2, pp. 1–85, Oct.

2012.

[3] Abhishek Jain, Rajbir Bhatti, and Harwinder Singh, “Total productive

maintenance (TPM) implementation practice: A literature review and

directions,” Lean Six Sigma Journal, vol. 5, no. 3, pp. 293–323, Jul. 2014.

[4] S. Nakajima, Introduction to TPM: Total Productive Maintenance.

Cambridge, MA: Productivity, 1988.

[5] Masaji Tajiri and Fumio Gotoh, Autonomous maintenance in seven steps:

implementing TPM on the shop floor. Portland, OR: Productivity, 1999.

[6] K. Shirose, TPM for Workshop Leaders. Cambridge, MA: Productivity, 1992.

[7] R. Attri, S. Grover, and N. Dev, “A graph theoretic approach to evaluate the

intensity of barriers in the implementation of total productive maintenance

(TPM),” International Journal of Production Research, vol. 52, no. 10, pp.

3032–3051, May 2014.

[8] H. M. Lazim, N. Ahmad, K. Hamid, and T. Ramayah, “Total employees

participation in maintenance activity: a case study of autonomous

maintenance approach,” Malaysia Labour Review, vol. 3, no. 2, pp. 47–62,

2009.