real-time measurement of dentinal fluid flow during amalgam and composite restoration
TRANSCRIPT
Real-time measurement of dentinal fluid flow duringamalgam and composite restoration
Sun-Young Kim a, Jack Ferracane b, Hae-Young Kim c, In-Bog Lee a,*aDepartment of Conservative Dentistry and Dental Research Institute, School of Dentistry, Seoul National University,
28-2 Yeongeon-Dong, Jongro-Gu, Seoul 110-749, South KoreabDepartment of Restorative Dentistry, Division of Biomaterials and Biomechanics, School of Dentistry,
Oregon Health & Science University, Portland, OR, USAcDepartment of Dental Hygiene, College of Health Science, Eulji University, Seoul, South Korea
j o u r n a l o f d e n t i s t r y 3 8 ( 2 0 1 0 ) 3 4 3 – 3 5 1
a r t i c l e i n f o
Article history:
Received 8 October 2009
Received in revised form
23 December 2009
Accepted 25 December 2009
Keywords:
Dentinal fluid flow
Permeability
Amalgam
Composite
Etch-and-rinse
Self-etch
s u m m a r y
Objectives: This study examined changes in the dentinal fluid flow (DFF) during restorative
procedures and compared permeability after restoration among restorative materials and
adhesives.
Methods: A class 1 cavity was prepared and restored with either amalgam (Bestaloy), or
composite (Z-250) with one of two etch-and-rinse adhesives (Scotchbond MultiPurpose: MP
and Single Bond 2: SB) or one of two self-etch adhesives (Clearfil SE Bond: CE and Easy Bond:
EB) on an extracted human third molar which was connected to a sub-nanoliter fluid flow
measuring device (NFMD) under 20 cm water pressure. DFF was measured from the intact
tooth state through the restoration procedures to 30 min after restoration, and re-measured
at 3 and 7 days post-restoration.
Results: Inward flow during cavity preparation was followed by outward flow after prepa-
ration. In amalgam restoration, the outward flow changed into an inward flow during
amalgam filling, which was followed by a slight outward flow after finishing. In composite
restoration, MP and SB showed an inward flow and outward flow for the rinsing and drying
steps, respectively. Application of a hydrophobic bonding resin in the MP and CE systems
caused a decrease in the flow rate. Air-drying of solvent for the CE and EB systems caused a
sudden outward flow, whereas light-curing of the adhesive and composite caused an abrupt
inward flow.
Conclusions: Each restorative step clearly changed the direction and the rate of the DFF
during restoration, which could be well identified with NFMD.
# 2010 Elsevier Ltd. All rights reserved.
avai lab le at www.sc iencedi rect .com
journal homepage: www.intl.elsevierhealth.com/journals/jden
1. Introduction
The hydrodynamic theory is the most widely accepted
explanation of dentin hypersensitivity.1 This theory proposes
that stimuli applied to the affected tooth cause the dentinal
tubular fluid to move in either an outward or inward direction.
This movement stimulates the mechano-receptors of the
* Corresponding author. Tel.: +82 2 2072 3953; fax: +82 2 2072 3859.E-mail address: [email protected] (I.-B. Lee).
0300-5712/$ – see front matter # 2010 Elsevier Ltd. All rights reserveddoi:10.1016/j.jdent.2009.12.008
sensory nerves in the dentin or pulp. There have been many
reports about different aspects of the dentinal fluid flow (DFF)
due to various stimuli applied to exposed dentin. For example,
cold and air-blast stimuli caused an outward flow of the
dentinal tubular fluid while heat caused an inward flow.
Osmotic pressure through the application of a hypertonic
solution caused an outward flow while tactile stimuli caused a
.
j o u r n a l o f d e n t i s t r y 3 8 ( 2 0 1 0 ) 3 4 3 – 3 5 1344
slightly inward flow. Mechanical loading corresponding to
occlusal force caused an inward flow, followed by an outward
flow when the mechanical loading was removed.2–5
While an amalgam restoration involves a relatively simple
procedure for cavity filling, a composite restoration involves
more complicated steps, including acid-etching, rinsing and
drying, adhesive application and light-curing, composite
filling and additional light-curing, and finishing and polishing.
These sequential restoration procedures can act on dentin as
hydrodynamic stimuli, causing the dentinal fluid to move in
either an outward or inward direction. Thus far, investigations
that show consecutive DFF during composite restoration are
rare.6,7 Furthermore, a comparison of DFF during an amalgam
restoration to that during a composite restoration has not
been done.
Continuous fluid movement through a dentin disc during a
bonding procedure,6 and DFF and cuspal displacement in
response to a resin composite restorative procedure has been
reported.7 The Flodec system (DeMarco Engineering, Geneva,
Switzerland) has been predominantly used for the study of
DFF; however, the results could not show clearly the details of
DFF for each restorative procedure due to the limitation of the
resolution (2 nL) of the instrument.6,7
In our study, a sub-nanoliter scaled fluid flow measuring
device (NFMD) capable of discriminating a volume change of
0.2 nL was fabricated to trace the DFF in real-time during
restorative procedures on an extracted tooth. The tooth was
restored with amalgam or resin composite under physiologic
pulpal pressure8 and DFF was measured during and after the
restorative procedure using the NFMD. The aim of this study
was to observe the aspects of DFF during the restoration
procedure and to compare the permeability after restoration
Fig. 1 – (a) Schematic diagram of the sub-nanoliter scaled denti
preparation.
with the different restorative materials and adhesives used.
The tested null hypotheses were that the DFF pattern did not
show any change according to each restorative procedure of
amalgam or composite restoration and that the permeability
after restoration did not show any difference among the
restoration materials and the measurement time.
2. Materials and methods
2.1. Structure and working mechanism of the NFMD
The NFMD fabricated for this study consisted of three parts: a
glass capillary and photo-sensor to detect the fluid movement;
a servomotor, lead screw, and ball nut to track the fluid
movement; and a rotary encoder and computer software to
record the data (Fig. 1a).
A water filled glass capillary with an internal diameter of
0.5 mm was connected between a water reservoir and the
tooth. A photo-sensor consisting of an infrared-light emitting
diode (EL-1CL3H, Kodensh Korea Corp., Iksan, Korea) and a
photo transistor (ST-1CL3H, Kodensh Korea Corp., Iksan,
Korea) detected the movement of an air bubble trapped within
the capillary. The voltage from the photo-sensor was input to a
comparator composed of an OP-amp (LM17741, National
Semiconductor, Santa Clara, CA, USA), which compared the
voltage from the photo-sensor with a preset reference voltage
corresponding to that of the water–air interface.
The output voltage from the comparator was input to the
servo-amplifier, which supplied the servo motor with power.
The motor rotated the lead screw, which had 1 mm of pitch,
and moved a connected ball nut to which the photo-sensor
nal fluid flow measurement system (NFMD). (b) Specimen
j o u r n a l o f d e n t i s t r y 3 8 ( 2 0 1 0 ) 3 4 3 – 3 5 1 345
was attached so that the photo-sensor was always placed at
the interface between the water and the air bubble. This
continuous negative feedback mechanism makes the move-
ment of the air bubble traceable in real-time.
The rotation of the screw was measured using a rotary
encoder (E40S-1000-3-2, Autonics, Seoul, Korea) which outputs
1000 pulses per rotation. The number of pulses was stored on a
computer at a rate of two data points/s using a data acquisition
device (PCI 6016, National Instrument, Austin, TX, USA).
As the movement distance per rotation of the lead screw is
1 mm and the resolution of the encoder is 1000 pulses per
rotation, the position resolution of the NFMD is 1 mm.
Therefore, the minimum measurable volume of water
movement is 0.196 nL [(0.25 mm � 0.25 mm � p) � 1 mm, the
diameter of the capillary = 0.5 mm].
2.2. Specimen preparation
Fifty caries-free human third molars were used in this study.
The project was approved by the Institutional Review Board of
Seoul National University Dental Hospital (CRI09005). The
third molars were extracted for surgical reasons independent
of this study. Extracted teeth were stored in 1% chloramine T
solution at 4 8C and were used within three months following
their extraction.
The root was removed 3 mm below the CEJ using a low-
speed diamond saw (Isomet, Buehler, IL, USA). The pulp tissue
in the pulp chamber was carefully removed without altering
the pre-dentin surface by using thin tissue forceps and
endodontic files. A sandblasted plexiglass square (10 mm
per side and 2 mm thick) with a hole drilled at its center was
used to mount each tooth. A metal tube 0.9 mm in diameter
was inserted into the hole, and the plexiglass was attached to
the tooth using an adhesive (Adper Scotchbond MultiPurpose,
3M ESPE, St. Paul, MN, USA) and a flowable composite
(Denflow, Vericom, Anyang, Korea) to ensure that one end
of the metal tube was located in the pulp chamber. The outer
surface of the bonded interface between the plexiglass and the
tooth on the top surface, and between the plexiglass and the
metal tube on the bottom surface was covered by nail varnish.
The specimen was placed in a rubber mold except for one third
Table 1 – The restoration materials and procedures of each gr
Restoration type Group code M
Amalgam AM Copalite varnish (C
TX, USA), Bestaloy
(Dongmyung, Seou
Composite – etch-and-rinse adhesive MP Adper Scotchbond
St. Paul, MN, USA)
(3M ESPE, St. Paul,
SB Adper Single Bond
St. Paul, MN, USA)
Composite – self-etch adhesive CE Clearfil SE Bond (K
Okayama, Japan), F
EB Adper Easy Bond (
Seefeld, Germany)
E, acid-etching for 15 s; P, primer application; A, adhesive application; PA
primer application; SEA, application of self-etch adhesive; LC, light-curin
of the exposed metal tube and was embedded with epoxy resin
to 1 mm above the CEJ (Fig. 1b).
The prepared specimen was stored in distilled water and
was connected to a water reservoir which contained distilled
water. A hydrostatic pressure of 20 cm H2O was applied to the
specimen from 24 h before the experiment was conducted.8
2.3. Measurement of the DFF during and after therestorative procedures
The prepared specimen was connected to the glass capillary
by silicon tubing filled with distilled water (Fig. 1). A
hydrostatic pressure of 20 cm H2O was applied throughout
all procedures with a water reservoir to simulate physiological
pulp pressure. Each specimen had a stabilizing time of 10 min
after being connected to the NFMD. After confirming that the
fluctuation level of the DFF was within �5 nL for a further
10 min, the restoration procedure was started.
A class 1 cavity with a mesio-distal width of 5 mm, a bucco-
lingual width of 3 mm, and an occlusal depth of 2 mm from a
central pit was prepared using an air-driven high speed
handpiece (MACH-QD, NSK, Tokyo, Japan) with 200,000–
300,000 rpm under water-coolant. The prepared cavity was
blot-dried with a wet cotton pellet and was then allowed to
remain exposed to air for 7–10 min. The average rate of the DFF
occurring for 5 min just before starting the restorative
procedure, which consisted of the varnish application for
the amalgam and the adhesive application for the composite,
was set as a baseline flow rate, i.e. internal reference, for
comparison with subsequent flow rate measurements after
restoration.
The prepared cavity was restored with either an amalgam
preceded by varnish (Copalite, Cooley & Cooley, Houston, TX,
USA), or a resin composite preceded by one of two etch-and-
rinse adhesives or one of two self-etch adhesives. The
materials and procedures used in this study are shown in
Table 1. An LED curing unit (Elipar Freelight 2, 3M ESPE, St.
Paul, MN, USA) was used to polymerize the composite and
adhesive (irradiance measured at 700 mW/cm2). The DFF
measurement was performed continuously from the intact
tooth state, throughout the cavity preparation and restorative
oup.
aterials Procedure
ooley & Cooley, Houston,
amalgam alloy
l, Korea)
Copalite application, amalgam filling
MultiPurpose (3M ESPE,
, Filtek Z-250
MN, USA)
E, P, A, LC (10 s), CF and LC (40 s):
incrementally by two horizontal layers
2 (3M ESPE,
, Filtek Z-250
E, PA, LC (10 s), CF and LC (40 s):
incrementally by two horizontal layers
uraray Medical Inc.
iltek Z-250
SEP, A, LC (10 s), CF and LC (40 s):
incrementally by two horizontal layers
3M ESPE AG,
, Filtek Z-250
SEA, LC (10 s), CF and LC (40 s):
incrementally by two horizontal layers
, application of mixed agent with primer and adhesive; SEP, self-etch
g; CF, composite filling.
Fig. 2 – A representative curve of consecutive DFF during
amalgam restoration. Upward (positive slope) movement
vs time on graph indicates outward DFF, whereas
downward (negative slope) movement indicates inward
DFF. Gentle air stream and amalgam filling caused
significant changes in the flow rate compared to the
baseline flow rate (P < 0.05). CP, cavity preparation; ECP,
end of cavity preparation; V, copalite varnish application;
ga, gentle air stream; AF, amalgam filling; and EAC, end of
amalgam condensing.
j o u r n a l o f d e n t i s t r y 3 8 ( 2 0 1 0 ) 3 4 3 – 3 5 1346
procedures, and to 30 min after the restoration in real-time. A
timetable was made for all steps of the cavity preparation,
adhesive procedure, and filling procedure. The timetable was
used to indentify the events in DFF corresponding to each step
during the restorative procedures. To determine the rate of
DFF after restoration, the average flow rate was calculated for
the last 5 min of the 30 min recorded after restoration.
The restored specimen was continuously connected to the
water reservoir with silicon tubing while maintaining a
hydrostatic pressure of 20 cm H2O and stored in a 100%
humidity plastic box, and the flow rate was re-measured at 3
days and 7 days after restoration. The average flow rate at 3
days and 7 days was determined for 10 min after stabilization
was achieved. Reductions in the flow rate were indicated as a
percent of the decreased flow rate at 30 min, 3 days, and 7 days
post-restoration with respect to the baseline flow rate [%
reduction in flow rate = 100 � (flow ratebase line � flow ratepost-
restoration)/flow ratebase line].9
The temperature and humidity of the environment were
24 � 0.5 8C and 30 � 5%, respectively. Under consideration of
1% of maximum flow rate reduction difference among five
materials or among three repeated measurements, an
appropriate sample size of eight per group was obtained with
power of 0.82 and type one error level 0.05. Finally a sample
size of 10 teeth per group was used to avoid possible reduction
of power because of failure cases during experiment proce-
dure. A repeated measure ANOVA was conducted to analyze
whether there were differences in the flow rate for each
restorative step during restoration and whether there were
differences in the reduction of the flow rate among the
restoration materials and the post-restoration measurement
time (30 min, 3 days, and 7 days). A paired T-test was
conducted to compare the flow rate for specific restorative
step with the baseline flow rate. The level of significance was
a = 0.05. The statistical package SPSS version 14.0 (SPSS Inc.,
Chicago, IL, USA) was used for all analytical procedures.
3. Results
The specific behavior of the DFF during the restorative
procedures according to the restoration methods are shown
in Figs. 2–6. Cavity preparation with an air-driven high speed
handpiece under water cooling showed an inward flow in
most cases. The direction of the DFF then reversed to an
outward movement as soon as cavity preparation was
complete. As for the amalgam restoration, the outward flow
after cavity preparation changed to an inward flow when
amalgam condensation was performed, which changed again
to a slight outward flow once amalgam condensing was
complete (Fig. 2).
The changes in DFF during resin composite bonding
procedure of each adhesive system are shown in Figs. 3b–6b
in detail. For the etch-and-rinse adhesive systems, MP and SB
(Figs. 3 and 4), the etching gel application itself caused a
decrease in the rate of the outward flow. Water rinsing caused
an inward flow, while blot-drying caused an outward flow. The
use of air-blast to evaporate solvent from the adhesives during
the bonding procedure caused an abrupt increase in outward
flow rate, even when it was performed gently. In the case of
MP, the adhesive resin application decreased the rate of
outward flow to nearly zero or even caused a slight inward
flow.
For the self-etch adhesive systems, CE and EB (Figs. 5 and 6),
the application of a self-etching primer or a self-etching
primer/adhesive did not show a specific change in the flow
direction or the flow rate. An air-dry procedure abruptly
increased the rate of outward flow. In the case of CE,
application of the solvent free bonding resin decreased the
rate of outward flow to nearly zero or even caused a slight
inward flow.
In all resin composite restorations, all three light-curing
procedures (one for adhesive curing; two for composite curing)
caused an abrupt inward flow, and a rapid compensatory
outward flow occurred after the final light-curing. The rate of
outward flow then gradually decreased to a low rate (Figs. 3–6).
Fig. 7 shows the mean reduction in the flow rate in percent
compared to the baseline at 30 min, 3 days, and 7 days after
restoration. There was no significant difference according to
the three repeated measurement times (P = 0.469) nor among
the different types of the materials at any time period
(P = 0.226).
4. Discussion
This study is the first to measure a continuous dentinal fluid
flow during an entire restorative procedure, including cavity
preparation. While most previous studies showed individual
dentinal fluid movements responding to a single stimulus,
such as air-blasts or hot and cold conditions, this study
followed the DFF as it occurs in events such as rinsing, drying,
and adhesive applications as well as restorative procedures.
Fig. 3 – (a) A representative curve of consecutive DFF during
composite restoration with MP. (b) Magnified view of
consecutive DFF during bonding procedure of MP. Upward
(positive slope) movement vs time on graph indicates
outward DFF, whereas downward (negative slope)
movement indicates inward DFF. The steps of acid-
etching, rinse, gentle air stream, adhesive resin
application, and light-curing caused significant changes in
the flow rate compared to the baseline flow rate (P < 0.05).
CP, cavity preparation; ECP, end of cavity preparation; E,
acid-etching; r, rinse; d, blot-dry; P, primer application; ga,
gentle air; A, adhesive resin application; LC, light-curing;
C1, composite filling of first layer; and C2, composite filling
of second layer.
j o u r n a l o f d e n t i s t r y 3 8 ( 2 0 1 0 ) 3 4 3 – 3 5 1 347
Moreover, this study is the first to compare the DFF behavior
during amalgam and composite restorative procedures, as
well as comparing different types of resin adhesives.
Although an outward flow was driven by a simulated pulpal
pressure of 20 cm water, the DFF direction during the cavity
preparation process was mostly inward. The inward flow
during the cavity preparation might be caused from the
movement of dentinal tubular fluid due to heat produced by
the friction of the bur against the tooth structure,10 or caused
by water entering the dentinal tubule due to the pressure of
water-coolant application from a high speed handpiece. The
inward flow changed to an outward flow shortly after the
cavity preparation was complete, possibly due to hydrostatic
pulpal pressure and water evaporation.
Amalgam restoration showed a relatively simple DFF
behavior without abrupt outward or inward flow (Fig. 2). For
example, the outward flow after cavity preparation changed to
a slight inward flow upon filling amalgam into the cavity; that
inward flow changed to a slight outward flow once amalgam
burnishing was complete. The pressure from amalgam
condensing and burnishing did not appear to cause an abrupt
change of the DFF, possibly because the force application is
localized and does not affect a majority of the tubules at any
given time. Although varnish is used to reduce the permeabil-
ity of dentin during amalgam restoration, its use is not likely to
improve the sealing of dentinal tubule significantly. Instead,
smear layer and/or smear plug was proved to provide most of
the sealing of cavity in previous studies.11–13 This study also
showed that varnish applied after cavity preparation could not
cause a significant change of the flow rate compared to the
baseline flow rate (Fig. 2).
The DFF changes measured during composite restoration
were far more complicated compared to placement of
amalgam (Figs. 3–6).
An air-blast, even when it was performed gently, increased
the rate of outward flow abruptly. The air-blast is often
performed to evaporate the organic solvent, to prevent
adhesive pooling, or to make a thick adhesive layer thin
during the bonding procedure. There are two possible
mechanisms to explain why an air-blast causes an abrupt
outward flow. First, an air-blast would increase the water
evaporation from the cavity surface, which would establish an
air–liquid interface at the dentinal tubule.14 This would cause
an elevated capillary phenomenon resulting in abrupt
outward movement of dentinal tubular fluid.15 The elevated
water height by the capillary action is given by h = 2s/grr [h:
elevated water height (m), s: water surface tension (0.07 N/m),
g: gravity (9.81 m/s2), r: water density (1000 kg/m3), r: inner
radius of capillary].16 A small capillary with a diameter of 1–
5 mm, similar to dentinal tubule, has as great an absorption
force as a 7 m water column when the mean diameter of the
dentinal tubule is assumed to be 2 mm. Second, an air-blast
would act as cold stimuli that would decrease the surface
temperature, which would shrink the volume of dentinal
tubular fluid and cause an outward movement of this fluid.
Based on this study, the DFF during the composite
restoration showed a complicated behavior with more
frequent changes in the direction of fluid movement when
compared to that occurring during the amalgam restoration.
Although it is likely that most postoperative sensitivity
associated with composite restoration is due mainly to
polymerization shrinkage,17 this frequent change in the DFF
may contribute to postoperative sensitivity afterwards by
stimulation of the sensory nerve ending during the restoration
process in advance.18
The differences between etch-and-rinse adhesives (MP and
SB) and self-etch adhesives (CE and EB) were sharply reflected
in the DFF curves during the bonding procedure. Rinsing and
drying after acid-etching moved the dentinal fluid inward and
outward, respectively, in the form of a valley-shaped flow
curve during the etch-and-rinse adhesive bonding procedure
(Figs. 3 and 4). On the other hand, the self-etch adhesive
maintained an outward flow until light-curing, although there
was a slight difference in the flow rate between each step of
the bonding procedure (Figs. 5 and 6).
A difference in the DFF behavior was also found between an
adhesive system which used a separate hydrophobic bonding
resin (i.e. MP and CE) and the simplified adhesive system in
Fig. 4 – (a) A representative curve of consecutive DFF during composite restoration with SB. (b) Magnified view of consecutive
DFF during bonding procedure of SB. Upward (positive slope) movement vs time on graph indicates outward DFF, whereas
downward (negative slope) movement indicates inward DFF. The steps of acid-etching, rinse, gentle air stream, and light-
curing caused significant changes in the flow rate compared to the baseline flow rate (P < 0.05). CP, cavity preparation; ECP,
end of cavity preparation; E, acid-etching; r, rinse; d, blot-dry; sb, SB application; ga, gentle air; LC, light-curing; C1,
composite filling of first layer; and C2, composite filling of second layer.
j o u r n a l o f d e n t i s t r y 3 8 ( 2 0 1 0 ) 3 4 3 – 3 5 1348
which primer and bonding resin were mixed in a single bottle
(i.e. SB and EB). In the case of the MP and CE, the application of
a hydrophobic bonding agent itself caused the flow rate to
decrease close to zero and finally reached a plateau or
sometimes a slight inward flow (Figs. 3b–5b). In SB and EB,
which contain a mixture of a primer and a bonding resin,
however, the application of the material itself did not affect
the outward flow rate (Figs. 4b and 6b). While these simplified
adhesives contain a higher concentration of hydrophilic resin
monomer which allows them to penetrate better into the
dentin, their ability to block the outward movement of
dentinal fluid may be weak. Previous studies reported that
simplified adhesives of the two-step etch-and-rinse type and
one-step self-etch type contain many water channels in the
adhesive layer, even after polymerization; hence, the adhesive
layer would behave as a semi-permeable membrane.19–21 This
study showed that the hydrophobic bonding resin application
itself has the effect of decreasing the rate of outward flow
immediately, which can contribute to stabilizing the initial
adhesion and sealing between the dentin and resin and cause
less pulpal reactions to bonding procedures.22,23
The inward flow during the light-curing is thought to result
from the thermal expansion of dentinal fluid caused mainly by
the heat from the light-curing unit and secondly by the
polymerization heat of the composite. In this study, light-
curing of each layer of composite was performed for 40 s, and
the temperature of the light-curing tip end increased by more
than 20 8C from room temperature, which caused an increase
of temperature by 4–5 8C within the pulp chamber in a
supplementary experiment results not shown. When the
cavity was sealed by adhesive and resin composites, fluid
expansion in the dentinal tubules and pulp chamber caused by
this temperature increase would produce an abrupt inward
flow. The return to the steep outward flow shortly after light-
curing seems to be due to a ‘‘rebounding effect’’ in which the
expanded fluid in the dentinal tubule and pulp chamber
contracts again as the temperature decreases.
Ideally, restoration of a cavity preparation should seal the
dentinal tubules completely to reduce concerns over bacterial
penetration and hypersensitivity. When the flow rate was
measured at 30 min, 3 days, and 7 days post-restoration, the
amalgam and composite restorations showed similar results,
reductions in permeability with the flow rate decreased by
88.7–97.8% compared to the baseline flow rate. Thus,
movement of fluids across the dentin surface was dramati-
cally reduced; the dentin surface was not completely sealed.
Within the limits of this study, no restorations may seal the
dentinal tubules completely under circumstance in which
physiologic pulp pressure continuously causes an outward
flow.
Fig. 5 – (a) A representative curve of consecutive DFF during composite restoration with CE. (b) Magnified view of consecutive
DFF during bonding procedure of CE. Upward (positive slope) movement vs time on graph indicates outward DFF, whereas
downward (negative slope) movement indicates inward DFF. The steps of air-dry, bond resin application, and light-curing
caused significant changes in the flow rate compared to the baseline flow rate (P < 0.05). CP, cavity preparation; ECP, end of
cavity preparation; SE, self-etch primer application; a, air-dry; A, bond resin application; LC, light-curing; C1, composite
filling of first layer; and C2, composite filling of second layer.
j o u r n a l o f d e n t i s t r y 3 8 ( 2 0 1 0 ) 3 4 3 – 3 5 1 349
A hydrostatic pressure of 15–20 cm H2O has been applied to
simulate the physiologic pulp pressure in many studies since
in vivo studies showed that normal physiologic pulp pressure
corresponded to a hydrostatic pressure of about 14 cm H2O
(range: 8–22 cm H2O).6,8,24–27 The studies applying a hydrostat-
ic pressure of 15–20 cm H2O including this study may over-
estimate DFF that might occur in vivo since under most in vivo
conditions, patients have received infiltration or block local
anesthetics containing epinephrine (1:100,000) which has
been shown to decrease pulpal blood flow up to 1–2 h.28–30
As pulpal blood flow falls, pulpal tissue pressure which is the
driving force for DFF would fall, so the DFF too would fall.27 It
would be needed to perform a further study considering the
effect of local anesthetics on permeability of restorations by
lowering or removing the hydrostatic pulp pressure during
restoration and afterward giving a physiologic pulp pressure.
The decrease in the flow rate according to the type of
restorative material or adhesive did not show any significant
differences in this study. This study determined the DFF rate
at relatively short time periods of 30 min, 3 days, and 7 days
after restoration. The permeability difference between the
amalgam and the composite restoration and among the
adhesives may not be shown in such a short-term observation.
A previous study which measured the decreased rate of
permeability after an amalgam and composite restoration
through a fluid transport system reported that the amalgam
restoration showed less permeability than the composite
restoration over a long period of 3 months while it did not
show a significant difference over a short period of 1 week.31
Recent studies in which microleakage was measured also
could not find a significant difference between amalgam and
composite restoration.32,33 Additional studies simulating the
intraoral condition by thermo-cycling and/or long-term
application of the hydrostatic pulpal pressure might be
necessary to investigate the difference in the permeability
between amalgam and composite restorations.
A number of earlier studies reported significant differences
of permeability between different types of adhesive; this study
did not show any statistically significant difference among the
adhesives used. 6,25,34 The reason why this study had a
different result from the earlier studies is thought to be that
this study had several different experimental conditions. First,
the baseline used in the previous studies was an increased
permeable state of dentin produced by acid-etching, whereas
the baseline in this study was the flow rate of a prepared cavity
with a smear layer produced by cutting but before acid-
etching. Second, this study measured not only the permeabil-
ity of cured adhesive on dentin but also the permeability of a
class 1 cavity restored with composite. The cavosurface
margins of the class 1 cavity in this study consisted of
excellent enamel–resin adhesion of resin to acid-etched
enamel compared to dentin–resin adhesion; this type of
Fig. 6 – (a) A representative curve of consecutive DFF during composite restoration with EB. (b) Magnified view of consecutive
DFF during bonding procedure of EB. Upward (positive slope) movement vs time on graph indicates outward DFF, whereas
downward (negative slope) movement indicates inward DFF. Air-dry and light-curing caused significant changes in the
flow rate compared to the baseline flow rate (P < 0.05). CP, cavity preparation; ECP, end of cavity preparation; eb, EB
application; a, air-dry; LC, light-curing; C1, composite filling of first layer; and C2, composite filling of second layer.
j o u r n a l o f d e n t i s t r y 3 8 ( 2 0 1 0 ) 3 4 3 – 3 5 1350
adhesion could have a much higher sealing ability compared
to the types used in the previous studies, in which a dentin
disc or piece of exposed dentin produced by horizontal cutting
was used as a specimen. Third, polymerization shrinkage
stress or water expansion of a composite in a cavity can both
be considered as an influencing factor that affects the
permeability of the restoration. Therefore, it is difficult to
directly compare the permeability of the adhesives in this
study with previous studies. Within the limits of this study,
Fig. 7 – Percent reduction in the DFF rate of each group at
30 min, 3 days, and 7 days after restoration with respect to
the baseline rate of DFF (n = 10). There was no significant
difference according to the three repeated measurement
times (P = 0.469) nor among the different types of the
materials at any time period (P = 0.226).
the adhesives used here would not be expected to show
significant difference in dentin permeability, at least when
used in restorations with all margins in enamel.
The main influencing factors affecting water evaporation,
which caused an outward flow from the exposed dentin
surface, are temperature and humidity.14,15 The temperature
and humidity of a prepared cavity surface are not constant in
most clinical situations. A prepared cavity can be filled under
the conditions of room temperature and ambient humidity
when isolated by a rubber dam, but also under intraoral
circumstances of 37 8C and 100% humidity as the other
extreme. The behavior of DFF during a composite restoration
procedure can also be affected by the finishing and polishing
procedure using rotary instruments, as heat may be generated
during this process.
In a future study, it would be interesting to investigate
how changes in the temperature and humidity around the
cavity and clinical procedures such as finishing and polish-
ing can affect the DFF and permeability of composite
restorations.
Acknowledgement
This study was supported by a grant (no. 3-2008-0014) from the
Seoul National University Dental Hospital Research Fund.
j o u r n a l o f d e n t i s t r y 3 8 ( 2 0 1 0 ) 3 4 3 – 3 5 1 351
r e f e r e n c e s
1. Brannstrom M, Linden LA, Astrom A. The hydrodynamics ofthe dental tubule and of pulp fluid. A discussion of itssignificance in relation to dentinal sensitivity. CariesResearch 1967;1:310–7.
2. Brannstrom M, Linden LA, Johnson G. Movement of dentinaland pulpal fluid caused by clinical procedures. Journal ofDental Research 1968;47:679–82.
3. Brannstrom M. The hydrodynamic theory of dentinal pain:sensation in preparations, caries, and the dentinal cracksyndrome. Journal of Endodontics 1986;12:453–7.
4. Pashley DH, Matthews WG, Zhang Y, Johnson M. Fluid shiftsacross human dentin in vitro in response to hydrodynamicstimuli. Archives of Oral Biology 1996;41:1065–72.
5. Paphangkorakit J, Osborn JW. The effect of normal occlusalforces on fluid movement through human dentine in vitro.Archives of Oral Biology 2000;45:1033–41.
6. Hashimoto M, Ito S, Tay FR, Svizero NR, Sano H, Kaga M,et al. Fluid movement across the resin–dentin interfaceduring and after bonding. Journal of Dental Research2004;83:843–8.
7. Ratih DN, Palamara JE, Messer HH. Dentinal fluid flow andcuspal displacement in response to resin compositerestorative procedures. Dental Materials 2007;23:1405–11.
8. Ciucchi B, Bouillaguet S, Holz J, Pashley D. Dentinal fluiddynamics in human teeth, in vivo. Journal of Endodontics1995;21:191–4.
9. Kolker JL, Vargas MA, Armstrong SR, Dawson DV. Effect ofdesensitizing agents on dentin permeability and dentintubule occlusion. Journal of Adhesive Dentistry 2002;4:211–21.
10. Horiuchi H, Matthews B. In-vitro observations on fluid flowthrough human dentine caused by pain-producing stimuli.Archives of Oral Biology 1973;18:275–94.
11. Pashley DH, O’Meara JA, Williams EC, Kepler EE. Dentinpermeability: effects of cavity varnishes and bases. Journal ofProsthetic Dentistry 1985;53:511–6.
12. Pashley DH, Depew DD. Effects of the smear layer. Copaliteand oxalate on microleakage. Operative Dentistry 1986;11:95–102.
13. Carrilho MR, Tay FR, Sword J, Donnelly AM, Agee KA,Nishitani Y, et al. Dentine sealing provided by smear layer/smear plugs vs. adhesive resins/resin tags. European Journalof Oral Sciences 2007;115:321–9.
14. Goodis HE, Tao L, Pashley DH. Evaporative water loss fromhuman dentine in vitro. Archives of Oral Biology 1990;35:523–7.
15. Matthews WG, Showman CD, Pashley DH. Air blast-inducedevaporative water loss from human dentine, in vitro.Archives of Oral Biology 1993;38:517–23.
16. Benenson W, Harris JW, Stocker H, Lutz H. Handbook ofphysics. 1st ed. New York: Springer; 2006. (p. 185, 795).
17. Bausch JR, de Lange K, Davidson CL, Peters A, de Gee AJ.Clinical significance of polymerization shrinkage ofcomposite resins. The Journal of Prosthetic Dentistry1982;48:59–67.
18. Pashley DH. Dynamics of the pulpo-dentin complex. CriticalReviews in Oral Biology and Medicine 1996;7:104–33.
19. Tay FR, Pashley DH, Suh BI, Carvalho RM, Itthagarun A.Single-step adhesives are permeable membranes. Journal ofDentistry 2002;30:371–82.
20. Tay FR, Pashley DH. Water treeing – a potential mechanismfor degradation of dentin adhesives. American Journal ofDentistry 2003;16:6–12.
21. Itthagarun A, Tay FR, Pashley DH, Wefel JS, Garcıa-Godoy F,Wei SH. Single-step, self-etch adhesives behave aspermeable membranes after polymerization. Part III.Evidence from fluid conductance and artificial cariesinhibition. American Journal of Dentistry 2004;17:394–400.
22. King NM, Tay FR, Pashley DH, Hashimoto M, Ito S, BrackettWW, et al. Conversion of one-step to two-step self-etchadhesives for improved efficacy and extended application.American Journal of Dentistry 2005;18:126–34.
23. Brackett WW, Ito S, Tay FR, Haisch LD, Pashley DH.Microtensile bond strength of self-etching resins: effect of ahydrophobic layer. Operative Dentistry 2005;30:733–8.
24. Sauro S, Pashley DH, Montanari M, Chersoni S, Carvalho RM,Toledano M, et al. Effect of simulated pulpal pressure ondentin permeability and adhesion of self-etch adhesives.Dental Materials 2007;23:705–13.
25. Chersoni S, Suppa P, Grandini S, Goracci C, Monticelli F, YiuC, et al. In vivo and in vitro permeability of one-step self-etch adhesives. Journal of Dental Research 2004;83:459–64.
26. Ozok AR, Wu MK, Ten Cate JM, Wesselink PR. Effect ofdentinal fluid composition on dentin demineralization invitro. Journal of Dental Research 2004;83:849–53.
27. Vongsavan N, Matthews B. Fluid flow through cat dentine invivo. Archives of Oral Biology 1992;37:175–85.
28. Olgart L, Gazelius B. Effect of adrenalin and felypressine(octapressin) on blood flow and sensory nerve activity in thetooth. Acta Odontologica Scandinavica 1977;35:69–75.
29. Kim S, Edwall L, Trowbridge H, Chien S. Effects of localanesthetics on pulpal blood flow in dogs. Journal of DentalResearch 1984;63:650–2.
30. Odor TM, Pitt Ford TR, McDonald F. Adrenaline in localanaesthesia: the effect of concentration on dental pulpalcirculation and anaesthesia. Endodontics and DentalTraumatology 1994;10:167–73.
31. Ozok AR, De Gee AJ, Wu MK, Wesselink PR. The influence ofresin composite and bonded amalgam restorations ondentine permeability in Class II cavities in vitro. DentalMaterials 2001;17:477–84.
32. Hersek N, Canay S, Akca K, Ciftci Y. Comparison ofmicroleakage properties of three different filling materials.An autoradiographic study. Journal of Oral Rehabilitation2002;29:1212–7.
33. Murray PE, Hafez AA, Smith AJ, Cox CF. Bacterialmicroleakage and pulp inflammation associated withvarious restorative materials. Dental Materials 2002;18:470–8.
34. Gregoire G, Guignes P, Millas A. Effect of self-etchingadhesives on dentin permeability in a fluid flow model.Journal of Prosthetic Dentistry 2005;93:56–63.