Page 1: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Shielding Design for PET Clinics

Robert L. Metzger, Ph.D.

Page 2: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Introduction

Positron Emission Tomography is enjoying explosive growth due to its ability to stage and track cancer lesions.

Patients receive ~0.55 GBq (15 mCi) of 18F labeled 2-Fluoro-2-Deoxy-D-Glucose (FDG) and rest in a “Quiet Room” for 45 minutes to allow the FDG to localize in the lesions.

Page 3: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

A Typical PET Scan

After voiding, the patients are then scanned for approximately 45 minutes in the PET scanner, or more commonly now, a combined PET/CT unit.

The PET/CT unit overlays the PET image on the CT image to create a precise registration of the tumor location.

Page 4: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

PET/CT Study

                                                                       

Page 5: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Malignant Melanoma

                                                                       

Page 6: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Malignant Melanoma

Page 7: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

PET Clinic Layout

Quiet Rooms

PET/CT

Hot Lab

PET/CT Control

Patient Bathroom

Page 8: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

PET Clinic Layout

This clinic layout utilizes distance rather than shielding to protect the technologist. The quiet areas are >20 feet distant from the tech at the PET/CT control.

Page 9: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Example Clinic Layout

Page 10: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Shielding Problems in PET

There was no published shielding guide for PET clinics. The AAPM Guide (handout) just published.

PET started as a research tool with low patient volumes and tightly controlled research environments. Little shielding was necessary.

PET has now exploded into clinical practice where patient volumes are high, and facilities frequently crammed into existing spaces.

Page 11: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Shielding Problems in PET

The 511 keV photons can easily penetrate shielding used for conventional Nuclear Medicine (140 keV) and diagnostic X-Ray (<120 kVp).

PET patients can fog film stored in dry laser printers and dark rooms. Loaded cassettes are quite vulnerable.

Page 12: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Shielding Problems in PET

NCRP 49 has attenuation coefficients for 511 keV photons, but they are narrow beam coefficients and PET is a broad beam environment. There are no buildup factors provided in NCRP 49. This has led to undershielding of some facilities (see “Shielding Factors on pg. 6 of the handout).

Page 13: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Shielding Problems in PET

The hot lab “L” block, syringe shields, and shadow shields for 511 keV photons are much more costly than those available for 140 keV Tc-99m. Some facilities have substituted Tc syringe shields and “L” blocks to save money.

These problems have resulted in doses >5 rem in one month for a technologist and > 2 rem per year for adjacent tenants in some facilities.

These doses have also resulted in a regulatory crack-down in some states.

Page 14: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

AAPM Shielding Guide

Provides conservative shielding estimates for clinics using Monte Carlo derived broad beam shielding curves for lead, concrete, and iron.

Considers patient self attenuation, decay of 18F throughout the scanning process.

Uses conservative usage for the quiet rooms and scanner.

Uses NCRP 147 occupancy factors for surrounding areas.

Page 15: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Design Dose Limits

Effective Dose Equivalent in Unrestricted areas must not exceed 1 mSv/yr (100 mrem/year).

The occupational limit is 50 mSv/yr, but AAPM uses a design guide of 5 mSv/yr (500 mrem/yr) based on ALARA and pregnant worker limits.

NCRP 147 uses the same criteria.

Page 16: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Source Term

AAPM Draft Guide uses a patient dose rate of 0.092 Sv m2/MBq hr or 5.1 mrem per hour at 1 meter for a freshly dosed patient with 15 mCi of FDG (pg. 8).

Represents about 65% of the point source gamma constant and is consistent with anterior patient measurements.

Page 17: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Scanner Source Term

After the patient rests for approximately 45 minutes, they void, and are then scanned. The radioactive decay for one hour resting is 0.74 and the patient voids ~15% of the dose, and the reduction during the scan is 0.91, leaving about ½ of the delivered dose.

AAPM uses mean doses to come to the same number (pg 10).

30% of the dose is in the patient’s head with the remainder distributed in the body.

Page 18: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Scanner Attenuation

AAPM uses a 15% reduction in dose due to scanner attenuation, but does not use this factor in the example problems.

AAPM Guide was written before PET/CT became prevalent.

Attenuation provided by the PET/CT double gantry is significant.

Use manufacturer’s data for scanner attenuation.

Page 19: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Usage

AAPM Guide assumes 100% occupancy of the quiet room with a dosed patient.

It does not consider multiple quiet rooms that are now common.

If multiple rooms are used, the occupancy cannot be 100%.

Scanners can scan a patient every 30 to 45 minutes, so, at maximum, the scanner cannot do more than 16 patients per 8 hour shift.

Practically, 10 is the maximum given patient setup times.

Page 20: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Usage

Quiet rooms can produce no more than one patient every 45 minutes or 10.7 (call it 10) per shift.

When multiple rooms are in use usage factors of ~0.65 are common.

Scanners are considered to be continuously occupied.

Page 21: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Occupancy Factors

Occupancy factors for surrounding areas are drawn from NCRP 147 (not NCRP 49).

NCRP 147 values are more realistic. Caution must be used when choosing

the 1/40th occupancy factor.

Page 22: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Draft Guide Limitations

The guide does not manage layered shields that typically comprise floor and ceiling shielding. Treating each layer individually and summing the attenuation causes overshielding as 511 keV photons are assumed to be incident on each layer.

Does not discuss the hot lab much.

Page 23: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Shielding

The 511 keV photons from 18F and the mobile nature of the source (patient) create some unique shielding design problems for a PET clinic.

New clinics are commonly sandwiched into existing imaging centers that are densely populated. Areas above and below the clinics are routinely occupied by other offices.

Page 24: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

PET Clinic Shielding

Inadequate structural shielding in some facilities has led to high doses to non-occupationally exposed personnel both within the facility and adjacent to it.

Improper hot lab shielding (“L” block and syringe shields for 99mTc) has led to high doses to the Nuclear Medicine technologists.

Page 25: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Hot Lab Shielding

Page 26: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

NCRP 49

Don’t use it! NCRP 49 HVLs are narrow beam

attenuation values, while the PET patient represents a broad beam condition.

Buildup in concrete is high at 511 keV.

Page 27: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Attenuation Curves(from AAPM Draft Guide on PET Shielding)

Lead

0.0001

0.0010

0.0100

0.1000

1.0000

0 5 10 15 20 25 30 35 40 45 50

Thickness(mm)

Transmission

Monte Carlo Simulation(Broad Parallel Beam)

Constant TVL 16.6 mm

Lead

Figure 1.

Thickness (mm)

Page 28: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Attenuation Curves

Concrete

0.0001

0.0010

0.0100

0.1000

1.0000

0 5 10 15 20 25 30 35 40 45 50

Thickness(cm)

Transmission

Monte Carlo Simulation(Broad Parallel Beam) Constant TVL 17.6 cm

Page 29: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Wall Shielding

Wall shielding is commonly required for the hot lab, quiet rooms, and scanner room.

Many designs use distance rather than shielding for the interior spaces as technologists dislike closing off their patients in the quiet rooms. Doors, when provided, are rarely closed.

Page 30: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Wall Shielding

Wall shielding can be easily calculated using point kernel techniques with buildup factors or from the AAPM Draft Guide on Pet Shielding Design.

Source terms and occupancy factors may be taken from the draft guide or from actual experience.

Page 31: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Wall Shielding

The height of the wall shielding is controversial. Some references say the shielding should extend to the floor above rather than the typical 7 foot height.

Not practical. AC, electrical, call button, CCTV hardware, intercom, etc. run in the interstitial space above the false ceiling.

Streaming is not significant at 450 KeV.

Page 32: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Example 1 Wall Shielding

See Page 9-10, Example 1 – Quiet Room.

15 mCi FDG, 40 pts/wk (one shift), uptake time one hour, 4 m to fully occupied uncontrolled area (T=1)

Weekly dose is 105.9 Sv (Eq 3). Limit is 20 Sv, therefore required

attenuation is 0.189.

Page 33: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Determining Pb Thickness

Draft Guide: Table 4, 1.2 cm lead. Point Kernel: I/I0=Be-x

= 1.79 cm-1 for Pb B = 1.35 at 1.2 cm I/I0= 0.16 – Good Agreement NCRP49: No Buildup – Off by about

30%.

Page 34: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Ceiling and Floor Shielding

When occupied areas exist above and/or below the quiet rooms and the scanning room, it is sometimes necessary to add sheet lead to the concrete deck. The floor thickness alone may not be sufficient to meet the non-occupational limit of 1 mSv (100 mrem) per year (25 mrem in some European countries).

Page 35: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Detector Locations

The Draft Guide recommends that the dose limits be applied at 0.5 meter above the floor above (height of a low chair), and 1.7 meters (5.6 ft) above the floor below.

That is, low chair above, and tall standing person below.

Very conservative, even unrealistic.

Page 36: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Example 4

See Page 11 - 12. Quiet room, 15 mCi FDG, 40 pts per

week, uptake time 1 hr, 4.3 m floor to floor clear height, 10 cm concrete deck, uncontrolled area above, T=1.

D = (4.3 –1) + 0.5 = 3.8 meters. Eq. 3: 117 Sv (one week)

Page 37: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Determining Concrete x

Draft Guide Table 4: 17 cm concrete. Point Kernel: I/I0=Be-x

= 0.204 cm-1 for Concrete B = 7.5 at 17 cm I/I0= 0.23 – Reasonable Agreement NCRP49: No Buildup – Gross

overestimate of shielding provided by concrete deck.

Page 38: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Example 4

Guide calls for 0.65 cm (slightly more than ¼ inch) of lead to be applied to the ceiling above.

Sums attenuation provided by layered lead and concrete.

Approach commonly leads to ½ inch lead requirements.

Inaccurate. Better approach described later.

Page 39: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Ceiling and Floor Shielding

The ceiling and floor shield consists of lead suspended under the support trusses of the concrete deck, forming a layered shield.

Much of the shielding cost of a PET clinic is driven by these layered shields as they frequently require structural reinforcement to support the weight.

Page 40: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Imaging Room

Correctly assumes about ½ of the delivered dose remains in patient.

Assumes a 15% reduction in dose by scanner gantry, but does not use this in example calculations.

The guide was developed before the advent of PET/CT units with multi-slice CT scanners.

Actual dual gantries of the modern PET/CT units provide substantial attenuation.

Page 41: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Example 2

See page 11. 15 mCi FDG original dose, 1 hr uptake,

imaging time 30 min, 40 pts per week, 3 m to uncontrolled area with occupancy of 1.0.

Calculates 59.7 Sv and calls for 0.8 cm Pb (1/2 inch Pb practically).

Very conservative.

Page 42: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

AAPM Guide Assumptions

Occupancy of the quiet room(s) with a dosed patient is 1.0. This is physically impossible. The patient must have a blood sugar test, be made comfortable, have an IV started, have the procedure described, and only then given the dose.

Page 43: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

AAPM Guide Assumptions

Single Quiet Room. Virtually all facilities now have 2 to 3 quiet rooms. The objective is to keep the scanner scanning. With the resting period of 45 minutes and a scanning time of 30 minutes for new units, multiple quiet rooms are necessary for efficient utilization of the scanner.

Page 44: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

AAPM Guide Assumptions

511 KeV. The guide correctly considers patient self-attenuation. But, since the self-attenuation is due to Compton Scattering, the energy is also reduced. The shielding curves are for 511 KeV photons incident on a shield wall under broad beam conditions. MCNP models indicate the actual energy is 350 – 450 keV incident on the shielding.

Page 45: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

AAPM Guide Assumptions

Limited to No Scanner Attenuation. The double gantries of the modern PET/CT units provide significant attenuation to the sides, top, and bottom. The guide describes an attenuation factor of 15%, but then does not use it for the example calculation of the wall shielding.

Page 46: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

CT

Where distances are large (e.g. control booth), the shielding needed for the CT may dominate. Common for new PET/CT rooms as the space requirements are large for the dual gantry scanners.

Use NCRP 147 to calculate CT shielding requirements.

Look at both PET and CT shielding requirements and pick.

Page 47: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Testing

NCRP 147 and many regulatory agencies ask for tests to ensure that the erected shielding is adequate.

May be performed with pressurized ion chambers (e.g. Victoreen 451P) or large volume ion chambers (e.g. 1800 cc Radcal chamber)

Alternately, monitor badges may be used (very cheap).

Page 48: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Testing

LLD is an issue. 100 mrad/year is 0.05 mrad/hour for full occupancy. The survey instrument must be able to accurately measure at 0.05 mR/hr or integrate to achieve this LLD.

Page 49: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Monitor Badges

Place away from areas frequented by patients.

Factor for area occupancy and shift change (if any).

May stop after compliance is determined. Some regulatory agencies have been

erecting their own monitor badges where shielding is suspect.

Page 50: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Ceiling and Floor Shielding

Point Kernel methods that calculate the attenuation provided by each layer and then sum them to obtain the total attenuation, tend to overestimate the shielding requirements when the layers are thick (in mfp). They assume 511 keV photons are incident on each layer.

This can dramatically increase the shielding cost, particularly when the second course of lead requires structural reinforcement of the ceiling.

Page 51: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Monte Carlo Model of the Quiet Room

We developed an MCNP model of the PET quiet room consisting of a MIRD phantom in a reclining position centered in a quiet room where the room and ceiling dimensions are taken from facility plans.

The 0.55 GBq (15 mCi) of 18F was equally split between the bladder and brain in the phantom.

Page 52: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

MCNP Model

Page 53: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Mercurad Model

Mercurad is a deterministic code developed by CEA specifically for layered shielding problems.

We developed a second model for this code using the same room materials and dimensions but a water sphere for the source term as human phantoms have not been ported to this code.

Page 54: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

MCNP Model of Scanner

A second model was developed for the scanner room using a MIRD phantom in a double gantry of a PET/CT scanner.

The phantom was loaded with 7.5 mCi of FDG. 30% of the dose was in the phantom’s head with the remainder in the body.

The scanner was developed from data provided by GE Healthcare for their current PET/CT unit.

Page 55: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

MCNP Scanner Model

Page 56: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Mercurad Scanner Model

A second model of the scanner room was developed for Mercurad using water filled spheres and cylinders to simulate the patient in the scanner.

The source term was 7.5 mCi, with 30% in the head and the rest distributed uniformly in the body.

Arrays of detectors were placed above and below the scanner room.

Page 57: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Mercurad Scanner Model

Page 58: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Example

A clinic design was chosen where the point kernel method indicated that two layers of ¼ inch lead would be needed to protect the office above the quiet rooms.

The actual room dimensions and concrete floor deck thickness was used and a volume detector was set 61 cm (chair height) above the second floor deck. The actual floor deck was corrugated, but only the thinner section was used in the calculation.

Page 59: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Scanner Example

A scanner room with a fully occupied OT clinic above and a cafeteria below was used to test the scanner model.

This room had a 4 inch ceiling and six inch concrete floor deck.

The areas above and below were uncontrolled and fully occupied.

Page 60: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

MCNP Runs

Moritz visualization was used to ensure that all of the source points were within the brain and bladder and that the volumes were adequately sampled.

After verification, 5 x 107 photons were run and the volume detector above the quiet room converged and passed all statistical tests.

Page 61: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Mercurad Run

A point detector set to read in exposure rate was set at the same location for the Mercurad model. An array of detectors were used for the scanner room as the maximum exposure is difficult to predict given the complex geometry.

The code converged in less than a minute. Results are expected to be higher than MCNP for

the quiet room as no patient self attenuation was considered in this model (source term was a ten cm sphere).

Page 62: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Moritz Source Verification

Page 63: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Scanner Room Results

Page 64: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Scanner Room Results

Page 65: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Mercurad Scanner Results

Distance from Patient's Head (cm)

0 50 100 150 200 250

Exp

osu

re R

ate

(m

R/h

r)

0.000

0.005

0.010

0.015

0.020

0.025

0.030

Floor Above Scanner RoomFloor Below Scanner Room

Page 66: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Testing

Once construction was complete, the exposure rates in the adjacent areas were measured before the facility went into full operation.

Exposure rates on the second floor were measured with a Radcal 1800 cc ion chamber.

Page 67: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Measurements

Actual exposure rates were measured with an 1800 cc ion chamber while the rooms were occupied with dosed patients.

Page 68: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Results (mR/wk)

Location MCNP Mercurad Observed

Office Above QR 1

1.96 2.28 0.29

Office Above QR 2

1.96 2.28 0.86

Page 69: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Discussion

The actual measurements were lower than predicted by either code.

The floor trusses, lead shielding support, electrical , plumbing, and lighting fixtures in the false ceiling all provided some attenuation and reduced the actual dose above the rooms.

Page 70: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Discussion

The Mercurad model (as expected) predicted slightly higher doses than the MCNP model as patient self attenuation was not considered.

Mercurad results should be factored for patient self attenuation.

Mercurad model development and execution was remarkably easy.

Page 71: Radiation Safety Engineering, Inc Shielding Design for PET Clinics Robert L. Metzger, Ph.D

Radiation Safety Engineering, Inc

Conclusion

The AAPM Draft Guide provides a conservative and simple shielding guide for PET clinics.

The guide overshields clinics with multiple quiet rooms, layered ceiling and floor shielding, and scanner rooms where modern PET/CT units are employed.