quick reference guide: sedline

7
SedLine ® Sedation Monitor Quick Reference Guide See the SedLine Sedation Monitor Operator’s Manual for complete instructions, including warnings, indications for use, intended use, cautions, and notes.

Upload: phungdung

Post on 05-Jan-2017

383 views

Category:

Documents


12 download

TRANSCRIPT

SedLine® Sedation MonitorQuick Reference Guide

See the SedLine Sedation Monitor Operator’s Manual for complete instructions, including warnings, indications for use, intended use, cautions, and notes.

3 Masimowww.masimo.com2 Masimowww.masimo.com

Display Window

Display Window

Electrode Status Display (See Page 7) - Provides electrode conductance status of the Sensor to indicate whether electrode adjustment is warranted.

Artifact (ARTF)- Measures the level of physiological (not brain-related) and environmental noise that the system detects. This value ranges from 0 to 100%. The numeric value of ARTF is a moving average.

Electromyograph (EMG) - Measures the level of detected muscle activity, such as grimacing or jaw clenching. This value ranges from 0 to 100%. The numeric value of EMG is a moving average.

Suppression Ratio (SR) - Measures the level of EEG signal suppression. This value ranges from 0 to 100%. The numeric value of SR is a moving average.

Patient State Index (PSI) (See Page 9) - Measures anesthetic depth. This value ranges from 0 to 100. The numeric value of PSI is a moving average. The trend display for PSI is color-coded, with the green band representing the range of values that is within user-configurable alarm limits.

EEG Display (See Page 8) - Reflect electrical activity of the frontal and pre-frontal cortices of the brain.

Density Spectral Array (DSA) (See Page 10) - Utilizes color to represent the power levels of EEG between 0 and 30 Hz (Upper limit can be adjusted to 40 Hz).

1

2

3

4

5

6

7

SedLine is a patient-connected, 4-channel processed electroencephalograph (EEG) monitor designed specifically for intraoperative or intensive care use. It displays electrode status, EEG waveforms and Density Spectral Array (DSA).

General knowledge of electroencephalograph (EEG) monitoring and an understanding of the features and functions of the SedLine Sedation Monitor are prerequisites for proper use.

Do not operate the SedLine Sedation Monitor without completely reading and understanding these instructions.

Notice Purchase or possession of this device does not carry any express or implied license to use with replacement parts which would, alone or in combination with this device, fall within the scope of one of the relating patents.

Caution: Federal law restricts this device to sale by or on the order of a physician.

Caution: For professional use. See instructions for use for full prescribing information, including indications, contraindications, warnings, precautions and adverse events.

For further information contact: Masimo Corporation 40 Parker Irvine, CA 92618, USA Tel.: 949-297-7000 Fax.: 949-297-7001 www.masimo.com

Patents: www.masimo.com/patents.htm

SedLine, Masimo, and are registered trademarks of Masimo Corporation. All other trademarks and registered trademarks are property of their respective owners. Printed in the USA.© 2015 Masimo Corporation.

EU authorized representative for Masimo Corporation:

MDSS GmbH Schiffgraben 41 D-30175 Hannover, Germany

EC REP

5 Masimowww.masimo.com4 Masimowww.masimo.com

Applying the Sensor

To connect the Sensor and place it on a patient, follow the steps below:

1. Prep skin with alcohol and dry before applying Sensor.

2. Peel the release liner from the center foam and discard.

3. Center CB/CT/L1/R1 electrodes on forehead above nose. (A)

4. Press the white lining around the electrodes firmly onto the forehead to secure the Sensor. Do not press directly on the electrode,

otherwise the gel may leak out. (B)

5. Peel release liner from R2 foam and discard.

6. Apply R2 in hairless area just above temple. (C)

7. Repeat steps 5 and 6 for L2.

8. Align the white end of the Patient Cable with the arrow on the Sensor connector.

9. Push together until the Sensor clicks into place. (D)

10. Ensure the monitor is connected and powered up.

11. If any of the electrodes are displayed as red, yellow, or blue on the Electrode Status Display on the monitor, push gently and wiggle

the electrodes until all electrodes are displayed as yellow or green.

12. For additional details, see the SedLine Sensor Instructions for Use.

• It is recommended that the Sensor be applied prior to induction to assess patient’s PSI baseline and to monitor the brain’s

response to induction agents.

• The Sensor is designed to fit patients with head circumference of 20 to 25 inches (50 to 64 cm).

1

2

3

4

5

6

7

8

9

10

11

12

Applying the Sensor

Recommendations:

7 Masimowww.masimo.com6 Masimowww.masimo.com

Electrode Status Display

Electrode Status Display

The electrode status display provides electrode conductance status of the Sensor.

Display Description Action

Green - Electrode impedance is within acceptable range.

No electrode adjustment necessary.

Yellow - Electrode impedance is marginal but acceptable.

Minor electrode adjustment may be required.

Red - Electrode impedance is out of acceptable range.

1. Gently push/wiggle electrodes until all are yellow and/or green.

2. Re-prepping of electrode may be necessary.

Blue - Unreliable connection or disconnection of Sensor electrodes.

Confirm that all electrodes, particularly electrodes CT and CB, of the Sensor are properly connected.

Light Gray - Impedance values are unavailable due to Sensor, Patient Cable, or Module not adequately detected.

1. Replace the Patient Cable.

2. The Module may need to be replaced.

Dark Gray with Cyan X - Gel-bridging detected on the electrode.

1. Clean any gel that has leaked outside of the electrodes on the patient's forehead.2. Confirm that all electrodes of the Sensor are properly connected.3. The Sensor may need to be replaced.

Dark Gray - Electrode monitoring disabled. All electrode icons will be dark gray.

1. Confirm that Sensor is plugged into Patient Cable. 2. Confirm that Patient Cable is connected to Module.

Electrode Status Examples

Gel-bridging between R1 and CB

Impedance Value unavailable on CT

Unreliable connection on L1

Unacceptable impedance on CB

High but acceptable impedance on L2

9 Masimowww.masimo.com8 Masimowww.masimo.com

EEG and PSI

Patient State Index (PSI)

Numeric Value When there is insufficient EEG data to calculate a PSI, the PSI numeric value will be replaced by dashes (--).

Trend PSI trend conveys the PSI numeric values over a period of time. The vertical axis range is 0 to 100 and it is configurable by the user. The horizontal axis represents time, and the timeframe displayed is configurable by the user.

Blue conveys that the PSI reading is

below the lower limits.

Green conveys that the PSI reading is within limits.

Yellow conveys that the PSI reading is

above the upper limits.

Expanding and Contracting Trend Pinch any trend display to expand or contract the timeframe displayed for all trending parameters.

EEG Display

• The EEG waveforms reflect electrical activity of the frontal and pre-frontal cortices of the brain.

• The display is configured to contain 4 data input sources.

• These input sources are acquired from electrodes on the Sensor: L1, R1, L2, and R2.

Default speed of waveforms is 30 mm/sec.Tap to configure speed.

Vertical axis displays the amplitude for different electrodes.

The PSI is the result of... not limited to (1) changes in power in various EEG frequency bands; (2) changes in symmetry and synchronization between critical brain regions; and (3) the inhibition of regions of the frontal cortex.

Default amplitude of waveforms is 5 uV/mm.Tap to configure amplitude.

Horizontal axis represents time.

11 Masimowww.masimo.com10 Masimowww.masimo.com

DSA

Frequency Bands

Understanding the DSA

Periods of artifact (ARTF) are conveyed by vertical white lines.

Density Spectral Array (DSA)

Frequency Scale The DSA represents power between 0-30 Hz as well as 0-40 Hz.

The “L” and “R” horizontal color graphs represent the activities of the EEG from the left and right frontal scalp regions, respectively.

95% spectral edge frequency (SEF) is displayed on each of the left and right color graphs as a white trend line.

Periods of suppression are conveyed by vertical black lines with a blue bar at the O Hz timeline for both the left and the right sides. (Periods of no data are conveyed by vertical black lines only.)

Beta

(13 - 35 Hz)

Alpha

(8 - 12 Hz)

Theta

(4 - 8 Hz)

Delta

(0.5 - 4 Hz)

Representative EEG Waveforms

36000/7600B-1215

www.masimo.com