quantitative quality assurance in hull & east yorkshire · • photon counting full field...
TRANSCRIPT
Quantitative Quality Assurance in
Hull & East Yorkshire
Dr. Craig Moore & Dr. Tim Wood
Radiation Physics Department
What imaging modalities do we
have in Hull & East Yorks? • Lots of Computed Radiography (CR)
• Some Direct Radiography (DR)
• CT
• MRI
• Full Field Digital Mammo (FFDM)
• Cone Beam CT – Dental
– Radiotherapy imaging
• Digital Fluoro labs
• In this talk I’ll try to give you a flavour of the quantitative QA measurements we do on some of the above (with some results!!)
What do we measure for CR and
DR?
• We (try to!!) conform
to the requirements of
IPEM Report 32 part
7
– MTF
– NNPS
– SNR
– Variance as a function
of dose
Computed Radiography
• Have been measuring
MTF and NNPS of
our CR systems for
approx 5 years
• Only have Agfa CR in
Hull
• Originally wrote own
software in Matlab
• But have since moved
to IQWorks
Measurement of CR MTF • First need to derive the
system transfer
properties (STP) of the
system, i.e. flat field
images at doses of:
– 1 uGy
– 4 uGy
– 12 uGy
– 50 uGy
• Always with 1 mm Cu
filtration and 70 kVp
y = 4265.1Ln(x) + 13221
R2 = 0.9981
0.00E+00
5.00E+03
1.00E+04
1.50E+04
2.00E+04
2.50E+04
3.00E+04
3.50E+04
0.00 10.00 20.00 30.00 40.00 50.00 60.00
Dose (uGy)
me
an
PV
Measurement of CR MTF
• Acquire ‘MTF image’ using a tungsten metal
edge tool, 50 x 50 mm and 1 mm thick
50mm
50mm
Measurement of CR MTF
tube
CR cassette
1.5m
MTF edge tool
Typical results…
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5 5
Frequency (cycles mm -1)
MT
F
Current
Baseline
Measurement of CR NNPS
• Use the 4 uGy STP
image
• Use IQWorks (or
Matlab) to do the
number crunching
Typical Results…
Measurement of SNR and Variance
• Use STP images
– Linearise
– Chop out central 5 cm
square portion of
image
– Calc mean, standard
deviation and variance
– SNR = mean/SD
– Plot STP corrected
variance against dose
STP Corrected Variance
y = 0.0019x1.3266
R2 = 0.9924
0.00E+00
1.00E-02
2.00E-02
3.00E-02
4.00E-02
5.00E-02
6.00E-02
-1.00 1.00 3.00 5.00 7.00 9.00 11.00 13.00 15.00
Dose (uGy)
ST
P C
orr
ecte
d V
ari
an
ce
Find relationship and quote
the exponent
If only quantum noise
the exponent should
be unity
CR results over time…
• Over the last 5 years, we have seen no significant changes in – MTF
– NNPS
– SNR
– Variance
– STP
– DDI calibration (Agfa’s ‘SAL log’ number)
• Do these quantitative QA tests take any extra time? – No!
– only one extra exposure required (MTF edge) as STP images would be taken anyway for detector response
Getting the images off the
system…… • For Agfa CR we set up a
DICOM node on each system
and send the image to our
laptops
– Must have DICOM server
running on laptop
– We use K-PACS
– Images must always be
acquired as:
• System diagnosis
• Flat field
• Speed class 200
General DR measurements
• We only have one
general Philips DR
system
• Measure MTF, NNPS
etc in the same
manner as for CR
• Grid removed
MTF edge on
detector housing
Typical DR results…
• STP is logarithmic (usually linear for a DR
system)
0
0.2
0.4
0.6
0.8
1
1.2
0 1 2 3 4 5 6
Spatial Frequency lp/mm
MT
F
Current
Baseline
Typical CR
0.00E+00
1.00E-05
2.00E-05
3.00E-05
4.00E-05
5.00E-05
6.00E-05
7.00E-05
0 0.5 1 1.5 2 2.5 3 3.5 4
Spatial Frequency lp/mm
NN
PS
mm
2
NNPS current
NNPS Baseline
Typical CR NNPS
y = -320.39Ln(x) + 2286.7
R2 = 0.9999
0
500
1000
1500
2000
2500
0.00 10.00 20.00 30.00 40.00 50.00 60.00
Dose (uGy)
Mean
Pix
el
Valu
e
DR Variance
• Variance as function of dose
y = 0.0012x1.2578
R2 = 0.9967
0
0.02
0.04
0.06
0.08
0.1
0.12
0.14
0.16
0.18
0.2
0.00 10.00 20.00 30.00 40.00 50.00 60.00
Dose (uGy)
ST
P C
orr
ec
ted
Va
ria
nc
e
Hot off the press…
• Last week we found significant difference
in MTF and NNPS from baseline of our DR
system
• Service engineer attending this Friday
– Will be interesting to see if we find any
genuine issues!!!
0
0.2
0.4
0.6
0.8
1
1.2
0 1 2 3 4 5 6 7
Spatial frequency (lp mm-1)
MTF
(f)
Measured
Baseline
1.00E-06
1.00E-05
1.00E-04
0 0.5 1 1.5 2 2.5 3 3.5 4
Spatial frequency (lp mm-1)
NN
PS
Measured X
Baseline X
DR image transfer
• Not easy to transfer
images
– Same method as
Agfa CR but Philips
were very reluctant
to help us with this
FFDM Philips MicroDose
• Photon counting full field digital mammo system
• STP measured with 2 mm Al filter at the end of the collimator
• MTF measured in a similar way to CR/DR
• Edge tool placed on breast platform – No extra filtration
– 32 kV/12 mAs
– Approx 4cm from chest wall edge
• Easy to take images off the system directly to an external hard drive
Philips MicroDose MTF Modulation Transfer Function
0 1 2 3 4 5 6 7
Spatial Frequency (lp per mm)
0.0
0.2
0.4
0.6
0.8
1.0
1.2
MT
F(f
)
We did have an issue with one of
our MicroDose systems!!!
• Strange shape to the MTF
• Corresponding dip in LSF
• No issues with basic TOR MAX line pair test
• We went back to test the system but it passed
Line Spread Function
-6 -4 -2 0 2 4 6
Distance (mm)
-0.2
0.0
0.2
0.4
0.6
0.8
1.0
1.2
Va
lue
Modulation Transfer Function
0 1 2 3 4 5 6 7 8
Spatial Frequency (lp per mm)
0.0
0.2
0.4
0.6
0.8
1.0
1.2
MT
F(f
)
MicroDose Noise
• We also measure detector response with dose
• And quantum and structure variance
0
20
40
60
80
100
1 10 100 1000
Detector dose (uGy)
% o
f to
tal va
ria
nce
Quantum variance Stucture variance AKref Baseline quantum variance Baseline structure variance
y = 24.93x - 245.87
R2 = 1.00
0
2000
4000
6000
8000
10000
12000
14000
16000
18000
0 200 400 600 800
Detector Air Kerma (uGy)
Me
an
pix
el va
lue
Radiotherapy 2D on-board imaging
• We work very closely with our RT colleagues wrt QA and testing of kV imaging systems – 2D and 3D
• STP images – 1 mm Cu at tube port
– 70 kVp
– 1 to 20 uGy
• MTF edge tool placed on imager – Approx 70 uGy
• Same analysis and reporting as for diagnostic CR/DR
• Get images off the system with the help of RT physicists
kV tube
imager
RT 2D Imaging
• Huttner image for this detector scored 12 groups (1.8 lp/mm)
• This is much worse than this MTF suggests
• Possible image processing on the acquisition monitor not applied to image sent for MTF analysis
– Does this demonstrate the usefulness of doing quantitative QA?
Modulation Transfer Function
0 1 2 3 4 5 6 7
Spatial Frequency (lp per mm)
0.0
0.2
0.4
0.6
0.8
1.0
1.2
MT
F(f
)
0
0.0001
0.0002
0.0003
0.0004
0.0005
0.0006
0.0 0.5 1.0 1.5 2.0 2.5
Frequency lp/mm
NN
PS
RT 2D Imaging - Variance
y = 0.0176x2.0081
R2 = 1
0.00E+00
2.00E-02
4.00E-02
6.00E-02
8.00E-02
1.00E-01
1.20E-01
1.40E-01
0.00 1.00 2.00 3.00 4.00 5.00
Dose (uGy)
ST
P C
orr
ecte
d V
ari
an
ce
Genuine due to
structure noise?
RT 3D Imaging • Use Catphan phantom
• Measure the MTF with the bead tool
• Measure CT numbers etc
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0.9
1
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6
Spatial frequency (lp / mm)
MT
F
Measured Baseline
Cone Beam Dental Imaging
• Tests conform to
HPA-CRCE-010
– Guidance on safe use
of dental cone beam
CT equipment
• We use a specially
designed phantom
shaped like a jaw to
hold Sedentex inserts
CBCT Dental MTF
Modulation Transfer Function
0 1 2 3 4 5 6 7
Spatial Frequency (lp per mm)
0.0
0.2
0.4
0.6
0.8
1.0
1.2
MT
F(f
)
Dental CBCT
MRI Quantitative QA
• Use Leeds Test
Object MagIQ
phantom
• Measure:
– SNR
– Linearity
– Uniformity
– MTF 0
0.2
0.4
0.6
0.8
1
0 0.2 0.4 0.6 0.8 1 1.2 1.4 1.6 1.8 2
Spatial frequency (mm-1
)
MT
F
256 x 256 horizontal 256 x 256 vertical 384 x 384 horizontal
384 x 384 vertical 512 x 512 horizontal 512 x 512 vertical
Future Work
• Want to start quantitative QA on our digital fluoro labs – Just recently found out (we think!!) how to take
images off our Siemens and Philips systems
• Thorough CT analysis via automated Catphan measurements
• Non-CBCT digital dental
• Do we want to start measuring effective MTF, NNPS, DQE etc???? – Similar to clinical set up with an appropriate phantom
so not detector centric
– But would be much more time consuming
Conclusions
• We have implemented routine quantitative testing (e,g. MTF, NNPS, SNR, Variance) into our QA programme, specifically for the following modalities: – CR
– DR
– FFDM
– CBCT (RT and Dental)
– MRI
• No significant time penalty in doing these quantitative tests if software is already set up
• We haven’t seen any significant changes in results over the years – but is this to be expected??? – We rarely see gross changes in Leeds Test Objects either
• Vital we get cooperation from manufacturers to ensure we can easily acquire images in correct format and easily get them off the system