pulse wave velocity in normal pulmonary arteries...
TRANSCRIPT
Pulse Wave Velocity in Normal Pulmonary Arteries by CMR;
a Comparison between Respiratory Gated and Non-Respiratory Gated Acquisition
W. Bradlow, P Gatehouse, R. Hughes, D. Firmin, R. Mohiaddin
Cardiovascular Magnetic Resonance Unit, The Royal Brompton Hospital, London, UK.
Introduction
Pulse wave velocity (PWV), calculated by dividing the distance separating two
points in a vessel by the time taken for the pulse wave to move between them, is a
useful marker of arterial wall stiffness. PWV measurement in normal pulmonary
arteries is currently being investigated by cardiac magnetic resonance (CMR), a
technique proven in similar analysis of the aorta 1. We have previously reported a
flow sequence where data was acquired at end-expiration and whose high
temporal resolution and use of interleaved slices allowed pulmonary PWV to be
calculated (Bradlow et al, Euro CMR Zurich, 2005).This may subsequently
provide a means with which to study abnormalities in arteries subject to
pulmonary hypertension.
Due to changes in intrathoracic pressure, caval flow and lung inflation, pressure-
flow waves, and their corresponding wave velocities, are subject to respiratory
variation. Methods to address this issue have been suggested for aortic PWV by
other non invasive means 2, but have not been fully explored in the assessment of
pulmonary PWV by CMR. Aim
To compare PWV data acquired at end-expiration with that acquired throughout
the respiratory cycle in the pulmonary arteries of adult volunteers.
Methods
Non-segmented through-plane phase velocity mapping (at 85cm/s VENC) was
applied to three slices transecting the main, left and right pulmonary arteries
(MPA,LPA,RPA) during free breathing (Figure 1). The use of an external
electrocardiogram system (with short filtering delays) to trigger the sequence
minimised the period between R wave and first RF pulse. Reference data
acquisition for the three slices was interleaved for 80 cine frames per slice within
each cardiac cycle, and repeated with velocity encoding on the subsequent cycle.
Velocity images were reconstructed by subtracting both phase images and
background velocity errors (measured by imaging the same planes in a stationary
phantom). Using maximum gradient and RF performance, and inverting the
slice-selection gradient for velocity encoding, TR 1.9ms was achieved, giving
TR 5.7ms for each of the three vessels in turn. Voxel size was FE 1.8 mm by PE
2.4 mm by SLT 8mm, echo time 1.1ms and flip angle 15o.
14 volunteers were studied and the sequence repeated three times with
diaphragm navigator (respiratory gated to end-expiration) and without (non-
navigated). A central region of interest was applied to each artery in each phase
and the mean velocity within it calculated (in-plane motion of MPA was
followed manually). Mean velocities were plotted against time (Figure 2). Pulse
arrival time was derived automatically without operator intervention. Baseline
and peak velocities were found and a velocity halfway between the two
calculated (V ½). V ½ was identified on the leading edge of the flow wave and a
straight line fitted to data points within +/-4ms of it. Arrival time was defined by
the intersection of this line and V ½. Path length was measured from multislice
transverse images.
Wilcoxon signed rank test was used to compare median values for navigated and
non-navigated data.
References
1. Lehmann E. Clinical value of aortic pulse-wave velocity measurement. Lancet
1999; 354:528-529.
2. Asmar R. Factors affecting Pulse Wave Velocity. Arterial Stiffness and Pulse
Wave Velocity Clincal Applications. Elsevier; 1999.p 82.
Figure 1
Figure shows typical sites of diaphragm navigator (small arrow) and vessel
slices (arrowhead)
Arrival of Flow Waves
0
0.1
0.2
0.3
0.4
0.5
0.6
0.7
0.8
0 20 40 60 80 100 120 140 160 180 200 220 240
Time (ms) after R-wave
Sp
eed
(m
/s)
MPA
LPA
RPA
Figure 2
Note measurable delay between onset of flow in MPA and that in RPA/LPA
Results
No statistically significant difference was found between navigated and non-
navigated data (Figure 3) (LPA p = 0.826, LPA p = 0.258).For clarity, three
volunteers who gave erratic (>10m/s or negative) LPA PWV values (found in
both navigated and non-navigated acquisition) and one outlier (non navigated
RPA; 4.85 m/s) are not included in the graph. They were all included in the
analysis.
Figure 3
Discussion
These results might be expected when one considers that end expiration
constitutes the majority of the free breathing cycle. In addition, respiratory
variations are likely to be attenuated as the derived PWV represents an average
value for the whole examination. Interleaving acquisition of the three imaging
planes is important in canceling out differences noted when planes were acquired
separately. A major source of error remains in the pulse arrival time detection;
further work in progress aims to improve this.
Conclusion
No significant difference in pulmonary PWV derived by navigator end-
expiratory gating acquisition compared with non-navigated acquisition was
identified. This supports the future use of the simpler non-navigated
approach.