Principle of USG imaging, construction of transducers and USG controls
DR. DEV LAKHERA
![Page 2: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/2.jpg)
Topics
• Properties of sound wave
• Propagation of sound wave
• Transducer components
• Workings of a transducer
• Interaction between sound and matter
• Ultrasonic image display
• USG controls
![Page 3: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/3.jpg)
Sound as a wave
• Mechanical
• Require a medium for transport
• Normal auditory frequency – 20Hz-20 KHz
• Ultrasonic - > 20 KHz
• Diagnostic imaging – 1 MHz – 20 MHz
![Page 4: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/4.jpg)
• Longitudinal waves travel with alternate compression and rarefaction.
![Page 5: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/5.jpg)
• Wavelength (λ) – bet two compression bands
• Time (T) to complete a single cycle is called the period.
• frequency (f ) -number of complete cycles in a unit of time.
![Page 6: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/6.jpg)
Propagation of sound (velocity)
Depends on:
Density
Resistance to compression
1540 m/sec
![Page 7: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/7.jpg)
• Sound travels slowest in gases, intermediate in liquids and fastest through solids
• Body tissues behave as liquids (avg 1540 m/sec)
• V = Freq x Wavelength (velocity constant in a medium)
• Freq – inc wavelength - dec
![Page 8: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/8.jpg)
Propagation of sound (intensity)
Depends on:
Amplitude of oscillation
Db
Defines the Brightness of the image
Irrespective of the Freq the Amp remains constant
The Higher the Amp the brighter the image and the lower the more darker the images
Returning Waves
![Page 9: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/9.jpg)
Frequency
Higher the freq Lower the penetration and Higher the resolution
Low the freq higher the penetration and lower the resolution
![Page 10: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/10.jpg)
Formation of USG image
1. Electrical Energy converted to Sound waves
2. The Sound waves are reflected by tissues
3. Reflected Sound waves are converted to electrical signals and later to Image
![Page 11: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/11.jpg)
Transducer
• Device that converts one form of energy into another
![Page 12: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/12.jpg)
Components
• Piezoelectric crystal
• Electrodes with conducting material
• Backing block
• Coaxial cable
• Casing
![Page 13: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/13.jpg)
• Backing block
![Page 14: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/14.jpg)
Piezoelectric crystals
• Piezoelectric effect- certain materials on application of electric energy change their physical dimensions
• Naturally occurring: Quartz
• PZT- Lead zirconate titanate
![Page 15: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/15.jpg)
• Dipoles –geometric pattern
• Plating electrodes
• Voltage applied in a pulse causes this crystal to vibrate
![Page 16: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/16.jpg)
Receive
• Echoes reflect back and from each tissue interface and cause physical compression of crystal element
• Dipole change their orientation
• Causes generation of voltage received and displayed
![Page 17: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/17.jpg)
Characteristics of a USG Beam
• Fresnel zone- Determined by radius of transducer
• Fraunhofer zone (divergent part)
• Fresnel zone- Increases with frequency and diameter
![Page 18: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/18.jpg)
Advantage of high frequency beams
• Superior superficial resolution , longer frensel zone
• Tissue absorption increases with increasing frequency so low frequency beam required to penetrate thick parts
• Larger transducers however reduce side to side resolution.(now reduced due to focused transducers)
![Page 19: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/19.jpg)
Phase array transducer
Linear transducer
Convex transducer- C60
![Page 20: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/20.jpg)
Interactions between sound and matter
Reflection
Refraction
Attenuation
Scattering
![Page 21: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/21.jpg)
REFLECTION
Images are produced by the reflected portion of beam
Percentage of reflected beam depends upon
1.Tissue’s acoustic impedance
2.Beam’s angle of incidence
![Page 22: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/22.jpg)
Acoustic Impedance
• How much resistance an ultrasound beam encounters as it passes through a tissue.
• Acoustic impedance depends on:
the density of the tissue (d, in kg/m3)
the speed of the sound wave (c, in m/s)
• Amplitude of returning echo is proportional to the difference in acoustic impedance between the two tissues
![Page 23: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/23.jpg)
Acoustic Impedance
• Two regions of very different acoustic impedances, the beam is reflected or absorbed
• Acoustic impedance of tissue is constant (speed of transmission is constant)
Examples of impedance
for bodily tissues (in
kg/(m2s)):
•air 0.0004 × 106
•lung 0.18 × 106
•fat 1.34 × 106
•water 1.48 × 106
•kidney 1.63 × 106
•blood 1.65 × 106
•liver 1.65 × 106
•muscle 1.71 × 106
•bone 7.8 × 106
![Page 24: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/24.jpg)
• Tissue - air interface – 99.9 % beam is reflected
• Coupling agent is needed
• Ultrasound gel
![Page 25: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/25.jpg)
Angle of incidence
• Higher the angle of incidence lesser is the reflection
![Page 26: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/26.jpg)
Specular reflectorDiaphragm
Wall of urine-filled bladder
Endometrial stripe
![Page 27: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/27.jpg)
Echogenicity (caused by Reflection)
Anechoic Hypo-Echoic Hyper-Echoic
![Page 28: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/28.jpg)
Scattering
• Redirection of sound in several directions
• Caused by interaction with small reflector or rough surface
• Only portion of sound wave returns to transducer
![Page 29: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/29.jpg)
Refraction
• Sound passes from one medium to other at an angle change in velocity but frequency is constant so there is a change in wavelength.
• Causes a change in direction
• spatial distortion
![Page 30: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/30.jpg)
Absorption
• Due to frictional forces opposing the movement of particles in a medium
• Utrasonic energy Thermal energy
• Depends on 1) frequency
2) viscosity of the medium
3) relaxation time of the medium
![Page 31: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/31.jpg)
• The deeper the wave travels in the body, the weaker it becomes
• The amplitude of the wave decreases with increasing depth
Attenuation
![Page 32: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/32.jpg)
Ultrasonic display
• Electronic representation of data
• A – mode
• M – mode
• 2D B mode
•
![Page 33: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/33.jpg)
Amplitude Modulation (A- mode)
• line through the body with the echoes plotted on screen as a function of depth
• Stronger echoes produce larger spinkes
![Page 34: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/34.jpg)
Motion mode (M- Mode)
• pulses are emitted in quick succession
• organ boundaries that produce reflections move relative to the probe
• commonly in cardiac and fetal cardiac imaging
![Page 35: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/35.jpg)
B-mode or 2D mode: (Brightness mode)
• Most used imaging mode
• Produces a picture of a slice of tissue
• Brightness depends upon the amplitude or intensity of the echo
![Page 36: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/36.jpg)
USG Imaging Controls
• TGC- Time gain compensator
• Near gain
• Far gain
• Intensity
• Coarse gain
• Reject
• Delay
• Enhancement
![Page 37: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/37.jpg)
![Page 38: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/38.jpg)
Time gain compensator
TGC adjusts the degree of amplification of
echoes
![Page 39: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/39.jpg)
Amplitude
• Intensity control- Increases the potential difference between transducer
• Coarse gain – Increases the height of all echoes proportionately
• Reject control- It helps remove echoes below
a minimum amplitude
![Page 40: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/40.jpg)
• Delay – Regulates the depth at which the TGC begins to augment the weaker signal
• Q-scan - automatically optimises key imaging parameters
![Page 41: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/41.jpg)
• General mode
• Resolution mode (high frequency setting)
• Penetration mode
• (low frequency setting)
![Page 42: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/42.jpg)
• 2D – sets to default B-mode
• Depth
• Zoom
![Page 43: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/43.jpg)
• Power doppler
Uses amplitude of Doppler signal to detect moving matter
• Pulse wave doppler
Emits USG in pulses
Lower velocity
• Continuous wave doppler
Transducer emits and receives continuously
High velocity
Color flow
Type of power doppler emits pulses
Directional color coding
![Page 44: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/44.jpg)
Tissue Harmonic Imaging-
Selectively removes phase
aberrations generated by
variation of velocity between
interfaces
![Page 45: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/45.jpg)
Speckle reduction filter
• ultrasound speckle image degradation and loss of contrast.
• Interaction of generated acoustic fields grainy appearance
![Page 46: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/46.jpg)
THANK YOU
![Page 47: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/47.jpg)
Types of Resolution
• Axial Resolution• specifies how close together two objects can be along the
axis of the beam, yet still be detected as two separate objects
• frequency (wavelength) affects axial resolution
![Page 48: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/48.jpg)
Types of Resolution
• Lateral Resolution• the ability to resolve two adjacent objects that are
perpendicular to the beam axis as separate objects
• beamwidth affects lateral resolution
![Page 49: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/49.jpg)
Types of Resolution
• Spatial Resolution• also called Detail Resolution
• the combination of AXIAL and LATERAL resolution
• some customers may use this term
![Page 50: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/50.jpg)
Types of Resolution• Contrast Resolution
• the ability to resolve two adjacent objects of similar intensity/reflective properties as separate objects
![Page 51: Principle of usg imaging, construction of transducers](https://reader033.vdocuments.mx/reader033/viewer/2022051710/5a65947b7f8b9ac2368b4633/html5/thumbnails/51.jpg)
Types of Resolution
• Temporal Resolution• the ability to accurately locate the position of moving
structures at particular instants in time
• also known as frame rate
• VERY IMPORTANT IN CARDIOLOGY