  • PRESSURECARETechnical Guide

  • 2 3

    INTRODUCTIONTrulife’s Oasis, Elite and Azure products have been developed in conjunction with highly experienced operating room personnel resulting in products which can help protect the surgical patient from nerve damage and the development of pressure sores.

    Trulife’s Pressurecare range are Class 1 medical devices registered in Ireland and in conformity with the Medical Devices Directives S1252 (93/42/EEC) as amended by Directive 2007/47EC.

    The Oasis, Elite and Azure ranges are manufactured to ISO 13485:2003 standards, each carrying the CE quality mark and are guaranteed against manufacturing defects for 2 years.

    The Trulife Pressurecare range first launched in 1995 has become a market leader in many countries around the world. In the course of this development process, the Trulife Pressurecare products have been subjected to a series of rigorous tests to ensure that the materials used in their manufacture provide the optimal level of benefits to the anaesthetised patient to ensure compatibility with everyday use in the operating room environment. The results of these tests are outlined in the following pages.

    Contents

    INTRODUCTION 3

    SECTION 1 PATIENT PROTECTION TEST 4

    Test 1.1 Interface Pressure 4

    Test 1.2 Pressure Relieving Properties of Silicone Gel 6

    SECTION 2 PATIENT COMPATIBILITY TESTS 7

    Test 2.1 Skin Compatibility 7

    Test 2.2 Bacteriological Data 7

    Test 2.3 Latex Content 7

    SECTION 3 GENERAL USE OF GEL PADS IN AN OPERATING ROOM ENVIRONMENT 8

    Test 3.1 Disinfectant Report 8

    Test 3.2 Flammabilty Test 8

    Test 3.3 Temperature Test 9

    Test 3.4 Freezing Test 10

    Test 3.5 Conductivity Test 10

    Test 3.6 X-Ray Attenuation 10

    Test 3.7 MRI 10

    SECTION 4 DURABILITY TESTS 11

    Test 4.1 Compression Test 11

    Test 4.2 Dynamic Roll Test 11

    Test 4.3 Stress Test 12

    CONCLUSION 12

    SUMMARY 13

    HOSPITAL COMMENTS ON THE TRULIFE PRESSURECARE RANGE 14

  • 4 5

    SECTION 1 PATIENT PROTECTION TEST

    TEST 1.1 INTERFACE PRESSURE

    Pressure, the amount of force exerted on a given area, is often measured in millimetres of mercury (mmHg). When a force or pressure greater than normal capillary pressure is exerted on a body over time, this can restrict blood flow to the area and cause serious tissue damage. It is important therefore, to reduce this pressure particularly at the more susceptible parts of the body.

    Our gel pads help relieve pressure by increasing the area of contact between the body part and the supporting surface (i.e. Trulife Pressurecare Pads).

    We use Force Sensitive Applications (FSA) to evaluate our products for their pressure relieving capabilities. FSA is essentially a clinical tool that allows us to evaluate and map the interface pressure between a person and the support surface (i.e. product) they are lying on. The Pressure Mapping System is a versatile tool that provides accurate information in an easy to interpret graphical format.

    HEELS

    INTERFACE PRESSURE MAPS

    26

    HEAD

    WITHOUTPROTECTION

    Average Pressure 8.64mmHg

    Average Pressure 65.2mmHg Average Pressure 13.9mmHg

    Tested in accordance with protocol for interface pressure management Ref: FSAT.0017

    ARMS

    Average Pressure 64.1mmHg

    Average Pressure 35.9mmHg Average Pressure 7.04mmHg

    INTERFACE PRESSURE MAPS

    10068_TRULIFE PRESSURECARE:Layout 1 21/08/2009 09:28 Page 26

    HEELS

    INTERFACE PRESSURE MAPS

    26

    HEAD

    WITHOUTPROTECTION

    Average Pressure 8.64mmHg

    Average Pressure 65.2mmHg Average Pressure 13.9mmHg

    Tested in accordance with protocol for interface pressure management Ref: FSAT.0017

    ARMS

    Average Pressure 64.1mmHg

    Average Pressure 35.9mmHg Average Pressure 7.04mmHg

    INTERFACE PRESSURE MAPS

    10068_TRULIFE PRESSURECARE:Layout 1 21/08/2009 09:28 Page 26

    When interpreting the Interface Pressure Maps, the lower the average interface pressure, the higher the degree of pressure relief of the product. For more detailed testing information, please contact your distributor.

    “Studies comparing the pressure-reducing abilities of standard foam procedure bed mattresses to gel mattresses (i.e. visco-elastic polymer) have found gel mattresses to be more effective”. 1.

    1. 2011 Perioperative Standards and Recommended Practices AORN “Recommended Practices for Positioning the Patient in the Perioperative Practice Setting”

    27

    Average Pressure 11.4mmHg Average Pressure 19.7mmHg

    Average Pressure 17.8mmHg Average Pressure 24.9mmHg

    Any positioningdevice used toreduce pressure onpatient's body partsand prevent tissuedamage should havethe documentedability to reducecapillary interfacepressures to32mmHg or less (5)

    5 Hendrick, J. & thomson et al. "Pressure reduction products: making appropriate choices" Journal of Enterostomal Nursing 20 (Nov/Dec 1993)

    Average Pressure 12.5mmHg Average Pressure 23.1mmHg

    INTE

    RFA

    CE

    PR

    ESSU

    RE

    MA

    PS

    10068_TRULIFE PRESSURECARE:Layout 1 21/08/2009 09:28 Page 27

    “ “

    Any positioning device used to reduce pressure on patient’s body parts and prevent tissue damage should have the documented ability to reduce capillary interface pressures to 32mmHg or less2

    2. 2011 Perioperative Standards and Recommended Practices AORN J. & thomson et al. Pressure reduction products: making appropriate choices

    Support Surface with no protection

    Support Surface with azure

    Support Surface with oaSiS elite

    Support Surface with oaSiS

  • 6 7

    SECTION 2 PATIENT COMPATIBILITY TESTSTEST 1. 2 PRESSURE RELIEVING

    PROPERTIES OF SILICONE GEL

    The material used in the Trulife range is a medical grade silicone; it is softer than both skin and underlying tissue. This means that pressure is relieved by the slight movement of the silicone gel and resulting dissipation of pressure/force across the product.

    Silicone Gel Versus PU/Polymer Gel

    Tests were carried out on Trulife’s products by the School of Physics, Dublin Institute of Technology, Kevin Street, Dublin, Ireland. They studied the ‘cushioning’ / pressure relieving properties of six product samples made using different types of films and silicone fillings. The test samples were flat, donut and heel pad shapes filled with either silicone or polyurethane gel and using films of various thicknesses.

    The results of these tests show that the flat, silicone filled samples had superior pressure relieving properties when compared with specimens made from polyurethane gel.

    Other studies support the benefits of gel. The recommended practices from AORN state that “foam and gel mattresses are effective for preventing both skin changes and pressure sore formation” 3..

    3. 2008 Perioperative Standards and Recommended Practices AORN “Recommended Practices for Positioning the Patient in the Perioperative Practice Setting”

    TEST 2 .1 SKIN COMPATIBILITY

    Silicone is widely acceptable to most biomedical applications and poses no risk to the user.

    Irritancy potential tests were performed by CYTOX, Gottleib-Keim-Strabe 60, 95448 Bayreuth, Germany. The tests were carried out in accordance with ISO 10993-10 (2009) and ISO 10993-1 (2009) which require clinical trials to be conducted in accordance with the principles of good clinical practice. Suppliers of the materials used to manufacture Trulife’s Pressurecare products have confirmed that these materials do not support microbiological growth.

    TEST 2.2 BACTERIOLOGICAL DATA

    Silicone provides a unique balance of chemical and mechanical resistance and due to its pure state displays exceptional biocompatibility.

    Suppliers of the materials used to manufacture Oasis, Elite and Azure Gel Pads have confirmed that these materials are latex free and do not support microbiological growth.

    TEST 2 .3 LATEX CONTENT

    Suppliers of the materials used to manufacture Oasis, Elite and Azure gel pads have confirmed that these materials are latex-free, plasticizer and fibreglass free.

  • 8 9

    SECTION 3 GENERAL USE OF GEL PADS IN AN OPERATING ROOM ENVIRONMENT

    TEST 3. 3 TEMPERATURE TEST

    Thermal stability is very important when the material is in contact with the body, as the temperature of the body and its environment is continually changing.

    The test replicates the treatment that Trulife’s Pressurecare range would receive in a hospital setting when placed in a preheated oven for a period of time to heat it up prior to it being used under a patient. The temperature of the product is monitored while heating in the oven and while cooling down to room temperature. The test results indicate that it takes a minimum of 30 minutes to heat the product from a temperature of 20˚C / 68°F to 50˚C / 122°F. In isolation, the product will not return to room temperature for at least 3 hours. This time will vary according to room and patient temperature.

    Tests were performed by the Research & Development Laboratory of Trulife Ltd, Dublin 24, Ireland.

    The Trulife Pads may be preheated to 40 degrees, or heated to body temperature, in an oven or used in conjunction with a heating mattress. Please follow instructions of the oven or mattress manufacturer as appropriate. (Refer to good clinical practices). Please follow oven or mattress manufacturer instructions.

    0 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25

    OASIS TEMPERATURE VS. TIME TEST IN PREHEATED OVEN

    TIME (HOURS)

    SERIES 1: OA012 OASIS HEADRING

    SERIES 1: OA030 OASIS TABLE PAD

    0˚C / 32˚F

    10˚C / 50˚F

    20˚C / 68˚F

    30˚C / 86˚F

    40˚C / 104˚F

    50˚C / 122˚F

    60˚C / 140˚F

    70˚C / 158˚F

    80˚C / 176˚F

    0 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25 2.5

    OASIS TEMPERATURE VS. TIME TEST FOR COOL DOWN STAGE

    TIME (HOURS)

    SERIES 1: OA012 OASIS HEADRING

    SERIES 1: OA030 OASIS TABLE PAD

    0˚C / 32˚F

    10˚C / 50˚F

    20˚C / 68˚F

    30˚C / 86˚F

    40˚C / 104˚F

    50˚C / 122˚F

    60˚C / 140˚F

    70˚C / 158˚F

    80˚C / 176˚F

    90˚C / 194˚F

    TEM

    PERA

    TURE

    TEM

    PERA

    TURE

    0 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25

    OASIS TEMPERATURE VS. TIME TEST IN PREHEATED OVEN

    TIME (HOURS)

    SERIES 1: OA012 OASIS HEADRING

    SERIES 1: OA030 OASIS TABLE PAD

    0˚C / 32˚F

    10˚C / 50˚F

    20˚C / 68˚F

    30˚C / 86˚F

    40˚C / 104˚F

    50˚C / 122˚F

    60˚C / 140˚F

    70˚C / 158˚F

    80˚C / 176˚F

    0 0.25 0.5 0.75 1 1.25 1.5 1.75 2 2.25 2.5

    OASIS TEMPERATURE VS. TIME TEST FOR COOL DOWN STAGE

    TIME (HOURS)

    SERIES 1: OA012 OASIS HEADRING

    SERIES 1: OA030 OASIS TABLE PAD

    0˚C / 32˚F

    10˚C / 50˚F

    20˚C / 68˚F

    30˚C / 86˚F

    40˚C / 104˚F

    50˚C / 122˚F

    60˚C / 140˚F

    70˚C / 158˚F

    80˚C / 176˚F

    90˚C / 194˚F

    TEM

    PERA

    TURE

    TEM

    PERA

    TURE

    TEST 3.1 DISINFECTANT REPORT

    The Trulife range is easy to clean / disinfect between surgical cases and is more hygienic than many disposable foam options which can facilitate bacterial growth by absorbing fluids. The disinfectant market is a very heterogeneous one comprised of a number of suppliers. Thus, testing types of chemistries, rather than brands, is the most logical approach. The three primary chemistries are:

    ♦ Phenolics

    ♦ Quaternary ammonia

    ♦ Quaternary ammonia and alcohol blends (Alcohol, chlorine)

    This list covers the major families of detergents.

    If there is something specific that you wish to confirm please contact your local distributor

    Cleaner/Disinfectant (Manufacturer) Main Chemistry Tested Mix Ratio

    Santex A Water, Sulphates no -

    Safeseat Ethanol no -

    Incidur Spray (Henkel Hygiene GmbH) Aldehydes, alcohols, quatemany compounds yes neat

    Incidin Plus (Acidic) Glucoprotamin, quatermany ammonium yes 200 to 1

    Pursept (Corrosive) Ethanol, Glyoxal, Quatermany Compounds yes 100 to 1

    Cutasept F (Bode) Alcohol, porpan no -

    Cutasept G (Bode) Alcohol, porpan no -

    Milton (Sodium hypochlorite, chlorine) Sodium hypochlorite, chlorine yes 100 to 1

    Terminator New Centuary Sales no -

    Mikrozid (S&M) Ethanol, propanol yes 5 to 1

    Bacillol AF (Bode) Ethanol, propanol no -

    Bigusept Fluid (Bode) Alcohol no -

    Vikron Potassium Peroxomonosulphate, Sulphamic Acid, Sodium alkyl benzene sulphonate yes 100 to 1

    Please note that it is not advisable to soak the Oasis, Elite and Azure pads in a cleaning solution overnight as this may affect the long-term durability of the products.

    TEST 3.2 FLAMMABILTY TEST

    Flammability tests were carried out by BTTG Fire Technology Services, Atlantic St, Broadheath, Attrinchau WA14 5DW, UK, in accordance with section 2 of BS 7175. Results indicate that when tested against this standard for ignitability in its “as received” condition, Trulife’s Pressurecare range are fire rated as self-extinguishing.

  • 10 11

    TEST 3.4 FREEZING TEST

    Tests were performed by the Research and Development Laboratory of Trulife Ltd, Dublin 24, Ireland investigating the effect of freezing temperatures on Trulife’s Operating Room Products when stored at temperatures of -18ºC / 0º F for 7 days without any deterioration in their condition. The results of the test indicate that damage is unlikely to occur as a result of freezing temperatures. We recommend that the products can be cooled to -12ºC without any adverse effects.

    TEST 3.5 CONDUCTIVITY TEST

    Modern surgical procedures, including High Frequency Techniques, require the interface pad to be non-conductive. According to BS 2050 (Specification for Electrical Resistance of conducting and anti-static products made from flexible polymeric material), accessory pads can be classed as non-conductive if their surface resistance is greater than 106 ohms. Trulife Pressurecare products were tested at the National Electronics test centre at Eolas in Glasnevin using a HP 4329A High Resistance Meter and a HP16008A Resistivity Cell according to the procedure laid down in BS2050.

    Results showed the Trulife Pressurecare pads to be within the non-conductive range with values from 3.0 x 109 to 2.3 x 1013 ohms.

    TEST 3.6 X-RAY ATTENUATION

    The Trulife Oasis, Elite and Azure pads are x-ray translucent, however, higher radiation levels may be needed (See table below)

    Attenuation Assessment

    Product Product Tested: Materials & Thickness Exposure Factor’s during Test’s Product Attenuation in mm of Aluminium

    OA030 10mm thick, silicone

    100kVp, 40mAsec’s

    2-4mm

    OA212 75mm thick,silicone >10mm Aluminium

    EL012

    At centre, 10mm silicone & foam 2-4mm

    At edge, 50mm silicone & foam 4-6mm

    EL141

    At centre, 40mm silicone & foam 4-6mm

    At edge, 140mm silicone & foam >10mm

    EL216 75mm thick, silicone & foam 6-8 mm

    AZ500 60mm, 2 types of silicone >10mm

    AZ600 10mm, 2 types of silicone 2-4mm

    FP012 75mm silicone, foam & fabric 2-4mm

    FP050 90mm silicone, foam & fabric 2-4mm

    TEST 3.7 MRI

    For medical implants and devices, the objectives of Magnetic Resonance Imaging (MRI) testing determines the presence of magnetic field interactions, heating, and artifacts in association with the use of an MRI system. Accordingly, assessments of magnetic field interactions (deflection angle and torque), MRI-related heating, and artifacts were conducted at 3-Tesla on the Trulife range. Trulife Pressurecare range does not cause significant signal loss or distortion during MRI imaging and is considered to be an acceptable device that may be used for patient positioning or patient comfort migration.

    Project Conducted by: Frank G. Shellock, Ph.D. President and CEO Shellock R & D Services, Inc.

    For more detailed testing information, please contact your local distributor.

    SECTION 4 DURABILITY TESTS

    TEST 4 .1 COMPRESSION TEST

    Silicone shows resistance to weathering which is important for the long term life of the products.

    When used in a normal environment a Trulife. Pressurecare product is repeatedly compressed and relaxed when a patient is placed on and removed from the pad.

    The compression test assesses the impact of a static bar being applied to and removed from the products continuously at a predetermined rate dependent on the size and specific application of the product. It is carried out at room temperature in an enclosed area.

    We require a product to undergo a minimum of 20,000 cycles at a rate of 15 cycles per minute without showing any signs of damage to successfully pass this test. Over a 2 year life span, this would simulate 27 compressive actions on the product for each day of the year.

    TEST 4 . 2 DYNAMIC ROLL TEST

    During normal use, the gel pads are likely to be pulled, pushed and rolled into position on an operating table. The Dynamic Roll Test assesses the likelihood of the silicone separating from the film during normal positioning and storage over the life of the product. During this test, the product is attached to a jig in a flat position and is subjected to a rolling action which effectively folds it in half and then rolls it back to its original flat position.

    The machine runs at 10 cycles per minute and the test is continued for a minimum of 5000 cycles. It is carried out at room temperature in a controlled environment. This is an accelerated test which simulates the product manipulation and positioning for each day of usage. A product is expected to withstand this treatment without showing any signs of damage before being added to the Trulife Pressurecare range.

  • 12 13

    TEST 4 . 3 STRESS TEST

    In the marketplace, Trulife’s Gel Range is likely to be pulled and manipulated repeatedly during everyday use. This will create high local stresses in a product and will have a tendency to encourage the separation of film and gel. The Stress Test measures the degree of adhesion between the film and gel of the Trulife Operating Room Range and therefore enables us to confidently predict the ability of the product to perform its function over the required lifetime.

    A sample piece is subjected to a 50 Newton Load cell on a Lloyd PVM3 Tensile Tester as illustrated in the diagram. This machine runs at a speed of 200 mm per minute and the test is conducted at room temperature. The force, extension and appearance of the sample are recorded over a 5 minute test period. A minimum value of 30 Newtons is required for a product to pass this test.

    CONCLUSION

    Silicone has been a world leader in the medical industry and will continue to be a prominent material in the future.

    Trulife uses it in its manufacturing, as the properties and adaptability far exceeds any other material in its field including polyurethanes. This is mainly due to its pressure relieving quality and its compatibility with skin.

    The Trulife Pressurecare range is a simple and cost effective solution with a comprehensive offering to cover all surgical procedures.

    Summary

    ♦ Re-distributes weight effectively ♦ The products can be cooled to -12°C

    ♦ High load-bearing capacity ♦ High tensile strength and elasticity

    ♦ Supports weight without bottoming out ♦ Hypoallergenic

    ♦ Silicone performs most like human skin ♦ Reduce heat and friction

    ♦ Medical Grade Silicone is highly durable and chemically stable compared to many other gels

    ♦ Can be moulded into any shape or form

    ♦ Over time its mechanical properties (softness etc) are likely to last much longer.

    ♦ Non-conductive

    ♦ Easily cleaned ♦ Cost effective

    ♦ X-ray translucent ♦ Won’t leak, flow or bottom out

    ♦ Reusable ♦ Easily repaired if cut

    ♦ Acceptable for use with MRI ♦ Fire rated as self-extinguishing

    ♦ Latex, fibre glass and plasticizer free ♦ Does not harden over time

    ♦ Does not support bacterial growth ♦ Retains it’s original shape after deformation

    ♦ Shock-absorbing ♦ Resistant to UV light

    ♦ Pads may be heated to body temperature 40°C ♦ Market Leader in many Countries

  • 14 15

    Pozycjonery żelowe są niezastąpionym produktem w pozycjonowaniu pacjenta na stole operacyjnym w każdej dogodnej pozycji. Dzięki swoim właściwościom przeciwdziałają powstawaniu odleżyn, oraz chronią przed ryzykiem podrażnienia nerwów.

    Produkty te ńie zawierają lateksu, który jest jednym z czynników alergennych. Pozycjonery są przezierne dla promieni RTG, dzięki czemu nie ma konieczności usuwania ich spod pacjenta podczas operacji. Powłoka, z której są wykonane pozycjonery jest łatwa do czyszczenia i dezynfekcji, co w dużym stopniu ułatwia pracę personelu.

    Jesteśmy bardzo zadowoleni z pozycjonerów żelowych TRULIFE, znacznie ułatwiają one pracę naszego zespołu.

    Agnieszka Bartczak, Ward Nurse Anesthesiology and Intensive Care Department

    Warszawa, Polska

    ‘Trulife Gel Positioners are indispensable products in the positioning of patients in the Operating Room. The positioners protect the patient from pressure sores and nerve injuries. They are kind to the skin because they are latex free.

    The positioners are x-ray translucent, therefore, there is no need to remove them from the patient during surgery. The products are easy to clean and very hygienic, which helps the staff.

    We are very happy with the Trulife Gel Positioners because they make our team’s job easier’.

    Agnieszka Bartczak, Ward Nurse Anesthesiology and Intensive Care Department

    Warsaw, Poland

    Die Anwendung der Lagerungshilfsmitteln der Firma Trulife ist ein wichtiger Bestandteil meiner täglichen Arbeit im OP zur optimalen Patientenpositionierung. Die Produkte entsprechen allen Anforderungen, die ein Lagerungshilfsmittel erfüllen muss.

    Sie sind leicht anzuwenden und gewähren einen sicheren Dekubitusschutz. Desweiteren sind alle Lagerungshilfsmittel latexfrei, röntgenstrahlendurchlässig, gut zu desinfizieren, widerstandsfähig und reparaturfreundlich. Trulife bietet mit seinem umfangreichen Sortiment eine Lösung für nahezu jede Lagerung.

    Auch Extremlagerungen, wie zum Beispiel die OP-Lagerung von Adipositaspatienten in Bauchlage oder Seitenlage sind sicher mit den Trulifeprodukten umzusetzen.

    – Katrin Schmidt, Fachkrankenschwester für Anästhesie und Intensivmedizin

    Deutschland

    ‘In order to correctly position patients in the Operating Room I use Trulife Gelpads on a daily basis. These products meet all the requirements a positioning pad has to fulfil.

    They are easy to use and give the patient great protection. Furthermore, all pads are latex free, radiograph radiolucent, easy to disinfect, robust and easy to repair.

    Trulife offers an extensive range of gelpads to suit every patient’s position. Even extreme positions like positioning an adipose patient, prone and lateral positions are safe and possible.”

    – Katrin Schmidt, Specialist Nurse, Anesthesia and Intensive Care

    Germany

    Zespól bloku operacyjnego Kliniki Neurochirurgii chciałby podzielić się opinią na temat pozycjonerów żelowych. Pozycjonery żelowe, które mamy przyjemność kupować od firmy Empireum zapewniają naszym pacjentom ogromną wygodę i bezpieczeństwo podczas długotrwałych i ciężkich zabiegów operacyjnych. Dzięki temu, że umożliwiają właściwe ułożenie ciała poprzez odpowiednie dopasowanie to zapobiegają powstawaniu odleżyn.

    Są łatwe do mycia i czyszczenia przy użyciu standartowych środków dezynfekcyjnych, lekkie i miłe w dotyku, wszystko to na pewnov jest ogromną zasługą materiału z którego są wykonane. Jesteśmy zadowoleni z zakupu wyżej wymienionych pozycjonerów i mamy nadzieję na dalszą owocną współpracę.

    – Z poważaniem, Nurse Neurosurgery Clinic

    Polska

    The Trulife Gel Positioners, which we are pleased to buy from the company Empireum, provide our patients with comfort and eliminate the dangers of pressure sores during long and difficult surgeries. They allow the correct positioning of the patient and are hygienic.

    Due to the material they are made from, they are easy to clean using standard detergents and disinfectants. They are light to lift and feel nice to touch. We are happy with Trulife Gel Positioners and we are looking forward to further fruitful cooperation with Trulife’.

    – Z Powazanlem, Nurse Neurosurgery Clinic

    Poland

    Hospital comments on the Trulife Pressurecare Range

    Eine angenehme Lagerung mit weichen, vorgewärmten Materialien nimmt dem Patienten im Vorfeld der Operation die Angst, lindert Schmerzen und lenkt vom Geschehen ab. Mit vielseitig einsetzbaren Systemen sowie gut ausgearbeiteten Standards können Lagerungen schnell und professionell durchgeführt werden, was zu einer erheblichen Zeitersparnis im OP führt und eventuelle Folgekosten durch Dekubiti reduziert.

    Ich habe mich für das umfangreiche System von Trulife entschieden, weil ich damit jede Positionierung individuell durchführen kann.

    Durch die Vielzahl der Polster habe ich für jede Lagerung eine Lösung und muss nicht verschiedene Systeme mischen.

    – Sabine Graf-Redecker, Fachkrankenschwester für Anästhesie und Intensivmedizin

    Deutschland

    ‘Trulife Gelpads offer patients a pleasant positioning experience. The soft and warm pads reduce the fear and the pain of the patient and distract him/her from the operation.

    With manifold systems and correct frameworks we can position the patient quickly and professionally so that we save time and follow-up costs caused by decubitus. I use the Trulife Gelpads because they suit every position and I don’t have to use different systems.

    – Sabine Graf-Redecker, Specialist Nurse, Anesthesia and Intensive Care

    Germany

  • ENQUIRIES ADDRESS TELEPHONE FAX

    UK: 41 Amos Road, Meadowhall , Sheffield, S9 1BX, UK +44 114 26 18100 +44 1142610074

    International: Trulife, 3013 Lake Drive, Citywest Business Campus, Dublin 24, Ireland +353 1 4511755 +353 1 4660015

    Ireland: Trulife, 3013 Lake Drive, Citywest Business Campus, Dublin 24, Ireland +353 1 4511755 +353 1 4660015

    USA: 2010 East High St., P.O. Box 89, Jackson, MI 49204, USA +1 800 788 2267 +1 517 789 3299

    Canada: 39 East Davis St, Trenton, Ontario, K8V 4K8, Canada +1 6132 392 6528 +1 613 392 4139

    www.trulife.com

    TL/P

    C02/

    1012