Preliminary Performance Measurements for Streak Camera with
Large-Format Direct-Coupled CCD Readout*
15th Topical Conference on High-Temperature Plasma DiagnosticsSan Diego, CA April 19 - 22, 2004
R. A. Lerche, J. W. McDonald , R. L. Griffith, D. S. Andrews,G. Vergel de Dios, A. W. Huey, P. M. Bell, O. L. LandenLawrence Livermore National Laboratory
P. A. Jaanimagi, R. BoniLaboratory for Laser Energetics, University of Rochester
* This work was performed under the auspices of the U.S. Department of Energy by University of California Lawrence Livermore National Laboratory under contract No. W-7405-Eng-48.
HTPD-2004-SC - 1ral 040418
HTPD-2004-SC - 2ral 040418
Abstract
Livermore’s ICF Program has a large inventory of optical streak cameras built in the 1970s and 1980s. The cameras are still very functional, but difficult to maintain because many of their parts are obsolete. The University of Rochester’s Laboratory for Laser Energetics is leading an effort to develop a fully automated, large-format streak camera that incorporates modern technology. Preliminary characterization of a prototype camera shows spatial resolution better than 20 lp/mm, temporal resolution of 12 ps, line-spread function of 40 m (fwhm), contrast transfer ratio (CTR) of 60% at 10 lp/mm, and system gain of 101 CCD electrons per photoelectron. A dynamic range of 175 for a 2 ns window is determined from system noise, linearity and sensitivity measurements.
HTPD-2004-SC - 3ral 040418
Summary: We have characterized a new prototype streak camera (ROSS)
1. The streak camera ROSS (Rochester Optical Streak System) Design effort led by Laboratory for Laser Energetics LLNL collaborated on design Streak tube: Photonis P-510 Direct coupled CCD (2K x 2K E2V 42-40 backside) LLNL input optics used for these tests
2. Camera performance Spatial magnification: 1.3 Spatial resolution: > 20 lp/mm (limiting visual) Temporal resolution: 12 ps @ 2 ns window Sensitivity: 101 CCD e- / photoelectron System noise: 5 e- / pixel Dynamic range: ~225 (at best resolution)
3. Camera appears to meet NIF optical streak camera requirements
HTPD-2004-SC - 4ral 040418
The compact streak camera has a direct-coupled large-format CCD readout
Prototype camera configuration
S-20 Streak tubePhotonis P-510
CCD
Input optics (for our tests*):LLNL input hardwareTek C-27 lens1-mm slitMag: 1.167
Streak tube: Photonis P-510S-20 photocathode
CCD:Spectral InstrumentsSI-800E2V 42-40 backside2K x 2K (13.5 m pixels)
Prototype streak camera (ROSS)(12” x 7” x 28” without input optics)
* The prototype input optics module for the ROSS is not yet available.
HTPD-2004-SC - 5ral 040418
Prototype camera magnification is 1.35, field-of-view is 20.5 mm
Sample image for magnification and FOV measurements
Test conditions:Light: Collimated (532 nm)Mask: 10-m slit every 1.5 mmCCD: 2K x 2K (13.5-m pixels)
Time (CCD pixels)
Sp
ace
(CC
D p
ixel
s)
0 20480
2048
Notes:CCD (27 mm square) records central (high-res) region of 60-mm dia streak tube image.
Magnification and FOV are referred to photocathode of the streak tube.
HTPD-2004-SC - 6ral 040418
System line-spread function (LSF)shows a 40 m resolution (fwhm)
1. Measure system magnification
2. Illuminate 10-m mask
3. Take spatial lineout
Mask: 11 um (at photocathode)
Binning: 2x2
FWHM: 2.0 super pixels (40 m)
40 mfwhm
Line-Spread Function
Gaussian
HTPD-2004-SC - 7ral 040418
We calculate the system contrast-transfer function (CTF) from the line-spread function
Calculation convolves LSF with square wave at various frequencies * =
Contrast Transfer Function
- Ronchi ruling measurements
HTPD-2004-SC - 8ral 040418
Ronchi ruling measurements at 8.6 lp/mm confirm 70% CTR estimated from LSF
Spatial Lineout and CTR
CTR @ 8.6 lp/mmMeasured: 68%LSF Prediction: 70%
8.6 lp/mm Ronchi Ruling
Space
Tim
e
HTPD-2004-SC - 9ral 040418
Contour plots show position dependence of spatial resolution in the streak camera image
1-mm slit 100-m slit
2x2 binningSuper-pix = 20 m
FW
HM
(C
CD
su
pe
r p
ixe
ls)
Po
sit
ion
(C
CD
su
pe
r p
ixe
l)
1000
020
0
10
Position (CCD super pixel)
8
0
4
Position (CCD super pixel)
HTPD-2004-SC - 10ral 040418
Contour plot shows position dependence of time resolution in the streak camera image
Contours showing position dependence of time resolution Test conditions:
Collimated light (530 nm)Slit: 100 mCCD: 2K x 2KSweep: Static (no sweep)Binning: 2x2
Time resolution:< 4 super pixels (fwhm)
over 90% of image
Position (CCD super pixel)
FW
HM
(C
CD
su
pe
r p
ixe
ls)
Po
sit
ion
(C
CD
su
pe
r p
ixe
l)
1000
0
0
5
10
Sweep (ns) t (ps)
2 126 23
12 4630 112
HTPD-2004-SC - 11ral 040418
Camera gain is high enough to detect individual photoelectrons
• Determine total ADUs in signal •Convert with CCD gain
• Determine number of pe- generated• Energy at photocthode times QE• Correct for streak camera time window
Gain = 101 CCD e- / pe-
2x2 = 6 CCD e-
* CCD gain = 1.09 e-/ADU
FWHM: 11.3 = 4.81 pix
Laser pulse
Time (ns)
Am
p
Noise for 2 sec exposureBinning Noise(e-)
1x1 5.132x2 5.973x3 6.694x4 7.92
Histogram of Background
Nu
mb
er o
f p
ixel
s
Counts (ADUs)
FWHM: 11.3 ADUsnoise: 4.81 ADUs
Image of swept slit(3 mm x 0.5 mm)
Tim
e
Space
HTPD-2004-SC - 12ral 040418
We use 20% temporal broadening to define the maximum usable current density
J0 = 2.2 amp/cm2
2. C-L current depends on geometry and voltage
1. 20% broadening occurs near 1% of Child-Langmuir space-charge limited current (J0) for laser pulse t = 45 ps
FWHM vs Current
1% C-L
Window
pc-15 kV
Extraction grid -12.5 kV
x
Expect reduced performance for J > 1% of C-L current (J0)
HTPD-2004-SC - 13ral 040418
We use SNR versus photoelectrons per resolution element as a figure-of-merit
SNR = s2N / [s2N (C/C)2 + (s/sb)2(b/C)2]1/2
s2 = # of detector pixels / image resolution elementN = # of photoelectrons per detector pixelsb
2 = number of detector pixels per binned pixelb = read plus dark current noise for one binned pixelC = # of CCD electrons / photoelectronC = standard deviation in C
For the ROSS streak camera we have:s2 = 32 pixels (4 space, 8 time)sb
2 = 1, 4, 9, 16b = 5.13, 5.97, 6.69, 7.92C = 101 CCD e- / pe-
C = Unknown
Time
Sp
ace
1 pixelsb2
(2x2)
Image PSF(32 pixels)
Maximize SNR by:Increasing s (more averaging)Increasing N (reduce sweep speed)
Increasing C (more efficient pe- detection)Decreasing read (improve CCD)
HTPD-2004-SC - 14ral 040418
A SNR plot establishes the dynamic range (DR) at ~60 for an image resolution element
SNR too low (avoid to ensure quality data)
SNR versus Photoelectrons per Detector Pixel
Sweep DR*2ns 1756ns 40512ns 855
* Binning: 2x2
Dynamic Range = f(s, sweep speed)
>1% C-L
2ns6ns
12ns
C = 101
HTPD-2004-SC - 15ral 040418
Summary: We have characterized a new prototype streak camera (ROSS)
1. The streak camera ROSS (Rochester Optical Streak System) Design effort led by Laboratory for Laser Energetics LLNL collaborated on design Streak tube: Photonis P-510 Direct coupled CCD (2K x 2K E2V 42-40 backside) LLNL input optics used for these tests
2. Camera performance Spatial magnification: 1.3 Spatial resolution: > 20 lp/mm (limiting visual) Temporal resolution: 12 ps @ 2 ns window Sensitivity: 101 CCD e- / photoelectron System noise: 5 e- / pixel Dynamic range: ~225 (at best resolution)
3. Camera appears to meet NIF optical streak camera requirements