phototherapy device e in nigeria: current irradiance

1
PHOTOTHERAPY DEVICE EFFECTIVENESS IN NIGERIA: CURRENT IRRADIANCE LEVELS AND SIMPLE STRATEGIES FOR IMPROVEMENT B. K. CLINE 1, 2 , H. J. VREMAN 3 , K. L. FABER 4 , H. H. LOU 1 , K. M. DONALDSON 1, 2 , E. AMUABUNOSI 5 , G. OFOVWE 5 , V. K. BHUTANI 3 , T. M. SLUSHER 6 1 D-REV: DESIGN FOR THE OTHER 90%, PALO ALTO, CA; 2 HASSO PLATTNER INSTITUTE OF DESIGN, STANFORD UNIVERSITY, STANFORD, CA 3 DIVISION OF NEONATOLOGY AND DEVELOPMENTAL MEDICINE, DEPARTMENT OF PEDIATRICS, STANFORD UNIVERSITY, STANFORD, CA; 4 PEDIATRICS, UNIVERSITY OF LOUISVILLE, LOUISVILLE, KY 5 PEDIATRICS, UNIVERSITY OF BENIN, BENIN CITY, EDO, NIGERIA; 6 PEDIATRICS, UNIVERSITY OF MINNESOTA, MINNEAPOLIS, MN CONCLUSIONS Poor irradiance levels are frequently delivered by PT devices at the 16 hospitals surveyed in Nigeria. With simple, low cost adjustments, the efficacy of PT devices can often be improved substantially. More efficacious PT should decrease both the incidence of kernicterus and the need for exchange blood transfusions. There is a need for educational outreach to ensure that effective treatment is provided from existing PT devices. ACKNOWLEDGEMENTS The authors gratefully acknowledge the doctors and nurses in Nigeria who spoke with us and provided access to their facilities. Supported in part by NCIIA Grant 6885-09. RESULTS SUMMARY OF RESULTS • Mean irradiance of all PT devices (n=76) was 7.6 ± 5.9 μW/cm2/nm. Only one device (1%) had an irradiance exceeding the American Academy of Pediatrics intensive phototherapy standard of 30 μW/cm2/nm (Pediatrics 2004). 29 devices (38%) had irradiances below 5 μW/cm2/nm. • Devices were found to be at an excessive distance from the patient. The mean distance was 39.6 cm ± 12.6 cm SD (range 20 to 79 cm), compared to the minimum recommended distance of 10 to 15 cm (Pediatrics 2004). • Spectral irradiance of n=25 devices was significantly improved (p<0.001) from 9 ± 6.6 SD before adjustments to 27.3 ± 15.2 SD after adjustments. • Commercially manufactured devices (n=68) had statistically significantly higher (p<0.01) spectral irradiance (8.1 ± 5.9 SD) than homemade devices (2.3 ± 2.8 SD). • Suboptimal illumination spectrua were qualitatively observed. Several devices contained white or green lamps. Fig. 1. n = 76 phototherapy devices were evaluated in six southern and central Nigerian (indicated by red dots). 1 Map of Nigeria. www.un.org/Depts/Cartographic/map/profile/nigeria.pdf Table 1. Adjustments found to improve irradiance levels. Fig. 2. Distribution of irradiance of all measured devices (n=76) ranged from 0.3 to 30 μW/cm2/nm. Fig. 3. Distribution of distance to patient of all overhead devices (n=73) ranged from 20 to 79 cm. Fig. 4. Observed barriers to device efficacy included excessive distance between patient and device (left) and use of suboptimal spectral emissions (right). METHODS INTRODUCTION OBJECTIVE To survey the efficacy of PT devices in resource-diverse Nigerian hospitals and identify strategies to improve performance. Neonatal hyperbilirubinemia is a major cause of neonatal morbidity and mortality in resource limited settings. Recent surveys, including a 2011 survey of phototherapy devices in Nigerian hospitals (Owa et al., Indian Journal of Pediatrics 2011), have shown that phototherapy (PT) devices for neonatal hyperbilirubinemia in these settings frequently deliver suboptimal treatment. There is a need for identification of means for improving effectiveness of treatment delivered by existing devices, in order to reduce the incidence of kernicterus and the need for blood exchange transfusions. • PT devices were evaluated in hospitals selected by convenience sampling across 4 Nigerian states and the Federal Capital Territory. The sampled 16 urban and semi-urban hospitals included both public and private facilities. • Spectral irradiance was measured for each device as found, using a GE Healthcare BiliBlanket Light Meter I or II, or Minolta Fluoro-Lite 451 meter. • For a subset of devices where consent was obtained, adjustments were made and the optimized irradiance levels were subsequently recorded. Irradiance data are reported as mean level ± SD μW/cm2/nm. • Statistical comparisons were made using Student’s t-test. !" !" !" !" $" %!" &'" &%" %!" %$" (" )" %" !" )" )" !" *" %!" %*" &!" &*" )!" !+* ,* , %! , %* , &! , &* , )! , )* , $! , $* , *! , ** , '! , '* , (! , (* ! #$ %&'()&* %(*+,-)& .)/0 !"# !%# &'# "# (# &# &# (# &(# )(# !(# %(# *(# (+* ,* , &( , &* , )( , )* , !( ! #$ %&'()&* +,,-.(-/)& 0123)4 5 3/46 Adjustments found to improve irradiance levels: Decreasing distance between device and patient Turning all switches on for devices with multiple switches Replacing broken, missing, and non-blue lamps with new blue lamps Improving alignment of device relative to the patient, ensuring that illumination footprint is centered about the patient Removing dust from the plastic screen between the lamps and patient (for devices with such a screen) !" $!" %!" &!" '!" !() *) * $! * $) * %! * %) * &! * &) * '! * ') * )! * )) ! #$ %&'()&* +,,-.(-/)& 0123)4 5 3/46 +,,-./-012 -32, -.456782076 +,,-./-012 92:;,2 -.456782076 Fig. 5. Distribution of irradiance of devices that were adjusted (n=25). Design Revolution Kajama Dosso Kandi Parakou Cotonou Zinder Daura Goure Diffa Bosso Bol Kaura Namoda Nguru Magaria Gumel Damasak Potiskum Biu Bajoga Biliri Kari Kontagora Bida Badeggi Kanfanchan Baro Owo Auchi Ondo Ore Oyo Iseyin Iwo Wukari Bali Ganye Tungo Ibi Kisi Kamba Dan Gulbi Sabon Benin Gwari Funtua Buea Warri Sapele Banyo Ndali Oron Wawa Illela Wurno Koko Gummi Maradi Mokolo Mora Guider Bamenda Ngaoundéré Nkambe Foumban Léré Pala Bama Dikwa Damboa Baga Gombi Mubi Numan Lau Pankshin Amper Kwolla Oturkpo Ankpa Takum Abong Ogoja Gembu Gboko Ikom Yoko Kontcha Ilaro Bohicon Hadejia Wudil Azare B ir n i N k o n n i Gaya Shaki Sokoto Katsina Dutse Damaturu Birnin Kebbi Bauchi Jalingo Minna Ilorin Akure Ado-Ekiti Lokaja Enugu Awka Makurdi Abeokuta Port Harcourt Benin City Asaba Ikeja Jos Owerri Yenagoa Calabar Yola Umuahia Uyo Gombe Gusau Lafia Abakaliki Douala Bafoussam Kumba Nikongsamba Garoua Maroua Aba L a g o s Ogbomosho Ife O n i t s h a Kano Maiduguri Oshogbo Ibadan Zaria Kaduna Abuja Porto- Novo N'Djamena Kainj Reservoir N i g e r N i g e r B e n u e B é n o u é K o m a d u g u K o m a d u g u G a n a Y e d s e r a m S a n a g a M b a m K a d u n a G o n g o l a Hadejia S o k o t o Za m fara G ulbin K a S o k o t o Gulf of Guinea Bight of Benin Bight of Biafra Lake Chad N i g e r N i g e r D e l t a Q u é m é Lake Bamendjing SOKOTO ZAMFARA KATSINA KANO JIGAWA YOBE BORNO KEBBI KADUNA BAUCHI ADAMAWA KWARA KOGI OYO PLATEAU NASSARAWA TARABA BENUE NIGER ABUJA CAPITAL TERRITORY OSUN OGUN ONDO EDO GOMBE DELTA ENUGU EBONY IMO RIVERS BAYELSA ABIA CROSS RIVERS AKWA IBOM LAGOS EKITI A N A M B R A N I G E R CHAD CHAD CAMEROON BENIN C.A.R. Map No. 4228 UNITED NATIONS October 2004 Department of Peacekeeping Operations Cartographic Section 0 50 100 150 200 km 0 50 100 mi National capital State capital Town, village Major airport International boundary State boundary Highway Main road Secondary road Railroad NIGERIA The boundaries and names shown and the designations used on this map do not imply official endorsement or acceptance by the United Nations. NIGERIA

Upload: others

Post on 10-Jan-2022

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: PHOTOTHERAPY DEVICE E IN NIGERIA: CURRENT IRRADIANCE

PHOTOTHERAPY DEVICE EFFECTIVENESS IN NIGERIA: CURRENT IRRADIANCE LEVELS AND SIMPLE STRATEGIES FOR IMPROVEMENT

B. K. CLINE1, 2, H. J. VREMAN3, K. L. FABER4, H. H. LOU1, K. M. DONALDSON1, 2, E. AMUABUNOSI5, G. OFOVWE5, V. K. BHUTANI3, T. M. SLUSHER6

1D-REV: DESIGN FOR THE OTHER 90%, PALO ALTO, CA; 2HASSO PLATTNER INSTITUTE OF DESIGN, STANFORD UNIVERSITY, STANFORD, CA3DIVISION OF NEONATOLOGY AND DEVELOPMENTAL MEDICINE, DEPARTMENT OF PEDIATRICS, STANFORD UNIVERSITY, STANFORD, CA; 4PEDIATRICS, UNIVERSITY OF LOUISVILLE, LOUISVILLE, KY

5PEDIATRICS, UNIVERSITY OF BENIN, BENIN CITY, EDO, NIGERIA; 6PEDIATRICS, UNIVERSITY OF MINNESOTA, MINNEAPOLIS, MN

CONCLUSIONSPoor irradiance levels are frequently delivered by PT devices at the 16 hospitals surveyed in Nigeria. With simple, low cost adjustments, the e�cacy of PT devices can often be improved substantially. More e�cacious PT should decrease both the incidence of kernicterus and the need for exchange blood transfusions. There is a need for educational outreach to ensure that e�ective treatment is provided from existing PT devices.

ACKNOWLEDGEMENTSThe authors gratefully acknowledge the doctors and nurses in Nigeria who spoke with us and provided access to their facilities. Supported in part by NCIIA Grant 6885-09.

RESULTS SUMMARY OF RESULTS

• Mean irradiance of all PT devices (n=76) was 7.6 ± 5.9 μW/cm2/nm. Only one device (1%) had an irradiance exceeding the American Academy of Pediatrics intensive phototherapy standard of 30 μW/cm2/nm (Pediatrics 2004). 29 devices (38%) had irradiances below 5 μW/cm2/nm. • Devices were found to be at an excessive distance from the patient. The mean distance was 39.6 cm ± 12.6 cm SD (range 20 to 79 cm), compared to the minimum recommended distance of 10 to 15 cm (Pediatrics 2004).• Spectral irradiance of n=25 devices was significantly improved (p<0.001) from 9 ± 6.6 SD before adjustments to 27.3 ± 15.2 SD after adjustments. • Commercially manufactured devices (n=68) had statistically significantly higher (p<0.01) spectral irradiance (8.1 ± 5.9 SD) than homemade devices (2.3 ± 2.8 SD).• Suboptimal illumination spectrua were qualitatively observed. Several devices contained white or green lamps.

Fig. 1. n = 76 phototherapy devices were evaluated in six southern and central Nigerian (indicated by red dots).1 Map of Nigeria. www.un.org/Depts/Cartographic/map/profile/nigeria.pdf

Table 1. Adjustments found to improve irradiance levels.

Fig. 2. Distribution of irradiance of all measured devices (n=76) ranged from 0.3 to 30 μW/cm2/nm.

Fig. 3. Distribution of distance to patient of all overhead devices (n=73) ranged from 20 to 79 cm.

Fig. 4. Observed barriers to device e�cacy included excessive distance between patient and device (left) and use of suboptimal spectral emissions (right).

METHODS

INTRODUCTION

OBJECTIVETo survey the e�cacy of PT devices in resource-diverse Nigerian hospitals and identify strategies to improve performance.

Neonatal hyperbilirubinemia is a major cause of neonatal morbidity and mortality in resource limited settings. Recent surveys, including a 2011 survey of phototherapy devices in Nigerian hospitals (Owa et al., Indian Journal of Pediatrics 2011), have shown that phototherapy (PT) devices for neonatal hyperbilirubinemia in these settings frequently deliver suboptimal treatment. There is a need for identification of means for improving e�ectiveness of treatment delivered by existing devices, in order to reduce the incidence of kernicterus and the need for blood exchange transfusions.

• PT devices were evaluated in hospitals selected by convenience sampling across 4 Nigerian states and the Federal Capital Territory. The sampled 16 urban and semi-urban hospitals included both public and private facilities. • Spectral irradiance was measured for each device as found, using a GE Healthcare BiliBlanket Light Meter I or II, or Minolta Fluoro-Lite 451 meter. • For a subset of devices where consent was obtained, adjustments were made and the optimized irradiance levels were subsequently recorded. Irradiance data are reported as mean level ± SD μW/cm2/nm. • Statistical comparisons were made using Student’s t-test.

!"# !"# !"# !"#

$"#

%!"#

&'"#

&%"#

%!"#

%$"#

("#

)"#

%"#

!"#

)"# )"#

!"#

*"#

%!"#

%*"#

&!"#

&*"#

)!"#

!+*# ,#*# ,#%!# ,#%*# ,#&!# ,#&*# ,#)!# ,#)*# ,#$!# ,#$*# ,#*!# ,#**# ,#'!# ,#'*# ,#(!# ,#(*#

!"#$"%&'()&*"

%(*+,-)&".)/0"

!"#$

!%#$

&'#$

"#$

(#$&#$ &#$

(#$

&(#$

)(#$

!(#$

%(#$

*(#$

(+*$ ,$*$ ,$&($ ,$&*$ ,$)($ ,$)*$ ,$!($

!"#$"%&'()&*"

+,,-.(-/)&"0123)453/46"

Adjustments found to improve irradiance levels: • Decreasing distance between device and

patient • Turning all switches on for devices with

multiple switches • Replacing broken, missing, and non-blue lamps

with new blue lamps • Improving alignment of device relative to the

patient, ensuring that illumination footprint is centered about the patient

• Removing dust from the plastic screen between the lamps and patient (for devices with such a screen)

!"#

$!"#

%!"#

&!"#

'!"#

!()# *#)# *#$!# *#$)# *#%!# *#%)# *#&!# *#&)# *#'!# *#')# *#)!# *#))#

!"#$"%&'()&*"

+,,-.(-/)&"0123)453/46"

+,,-./-012#

-32,#

-.456782076#

+,,-./-012#

92:;,2#

-.456782076#

Fig. 5. Distribution of irradiance of devices that were adjusted (n=25).

Design Revolution

Kajama

Dosso

Kandi

Parakou

Cotonou

Zinder

Daura

Goure

Diffa

Bosso

Bol

KauraNamoda

Nguru Magaria

Gumel

Damasak

Potiskum

Biu

Bajoga

Biliri

Kari

Kontagora

BidaBadeggi

Kanfanchan

Baro

Owo AuchiOndo

Ore

OyoIseyin

IwoWukari

Bali

Ganye

TungoIbi

Kisi

Kamba

Dan Gulbi

Sabon Benin Gwari

Funtua

Buea

Warri

Sapele

Banyo

Ndali

Oron

Wawa

Illela

Wurno

Koko

Gummi

Maradi

Mokolo

Mora

Guider

Bamenda

Ngaoundéré

Nkambe

Foumban

Léré Pala

Bama

Dikwa

Damboa

Baga

Gombi Mubi

NumanLau

Pankshin Amper

Kwolla

OturkpoAnkpa Takum Abong

Ogoja

Gembu

Gboko

Ikom

Yoko

Kontcha

Ilaro

Bohicon

Hadejia

WudilAzare

Birni Nkonni

Gaya

Shaki

Sokoto

Katsina

Dutse Damaturu

Birnin Kebbi

Bauchi

Jalingo

Minna

Ilorin

AkureAdo-Ekiti

Lokaja

Enugu

Awka

Makurdi

Abeokuta

PortHarcourt

Benin City

Asaba

Ikeja

Jos

Owerri

YenagoaCalabar

Yola

Umuahia

Uyo

Gombe

Gusau

Lafia

Abakaliki

Douala

Bafoussam

Kumba

Nikongsamba

Garoua

Maroua

Aba

Lagos

Ogbomosho

Ife

Onitsha

Kano Maiduguri

OshogboIbadan

Zaria

Kaduna

N'Djamena

Abuja

Porto-Novo

N'Djamena

KainjReservoir

Niger

Niger

Benue

Bénoué

Komadugu

Komadugu G

ana

Yedsera

m

Sanaga

Mbam

Kaduna

Gon

gola

Hadejia

Sokoto

Zamfara

Gulbin Ka

Sokoto

Gul f of Guinea

Bight of Benin

Bight of Biafra

Lake Chad

Nig

er

Ni ge r D

e l t a

Quém

é

Lake Bamendjing

SOKOTO

ZAMFARA

KATSINA

KANO

JIGAWA Y O B EB O R N O

K E B B I

KADUNA B A U C H I

ADAMAWA

KWARA

K O G I

O Y O

P L AT E AU

NASSARAWA

T A R A B A

B E N U E

N I G E R

ABUJA CAPITAL TERRITORY

OSUN

O G U NONDO

E D O

GOMBE

DELTA

ENUGU

EBONY

IMO

RIVERS

BAYELSA

ABIACROSS RIVERS

AKWA IBOM

LAGOS

EKITI

AN

AM

BR

A

N I G E RC H A D

CHAD

C A M E RO O N

B E N I N

C.A

.R.

Map No. 4228 UNITED NATIONSOctober 2004

Department of Peacekeeping OperationsCartographic Section

0 50 100 150 200 km

0 50 100 mi

National capitalState capitalTown, villageMajor airportInternational boundaryState boundaryHighwayMain roadSecondary roadRailroad

NIGERIA

The boundaries and names shown and the designations used on this map do not imply official endorsement or acceptance by the United Nations.

NIGERIA