  • OWNER'S MANUAL

    New Versa Chair

    English

  • MANUAL OVERVIEW

    EQUIPMENT MANUAL (INSTRUCTIONS FOR USE)

    GMDN: 47144 Dental Chair, electric

    Technical Name: Dental Chair

    Brand: Dabi Atlante

    Commercial Name: New Versa Chair

    Supplier/ Manufacturer: Alliage S/A Indústrias Médico Odontológica C.N.P.J. 55.979.736/0001-45 - Insc. Est. 582.002.897.114 Rod. Abrão Assed, Km 53+450m - Cx. Postal 782 CEP 14097-500 Ribeirão Preto - S.P. - Brazil Telephone +55 (16) 3512-1212

    Technical manager: Daniel R. de Camargo CREA-SP: 5062199650

    ANVISA Registration No.: 10101130085

    WARNING For greater security: Read and understand all instructions contained in these Instructions for Use before installing or operating this Equipment. Note: These Instructions for Use must be read by all operators of this Equipment.

    2

    Owner’s Manual – New Versa Chair

  • 3

    TABLE OF CONTENTS

    EQUIPMENT IDENTIFICATION............................................................................... 5 Principles and fundamentals applied to the product operation ........................................ 5 Equipment Purpose .................................................................................................. 5 Equipment Description .............................................................................................. 5 Equipment Indication ................................................................................................ 5

    MODULES, ACCESSORIES, OPTIONS AND CONSUME MATERIALS .......................... 6

    TECHNICAL SPECIFICATIONS ............................................................................... 7 General features ...................................................................................................... 7 Electromagnetic emissions ........................................................................................ 9 Dimensions (mm) .................................................................................................. 13 Packing symbols .................................................................................................... 14 Product Symbols .................................................................................................... 14

    EQUIPMENT INSTALLATION ................................................................................ 15

    EQUIPMENT OPERATION ..................................................................................... 18 Multifunctional Pedal ............................................................................................... 20

    PRECUTIONS, RESTRICTIONS AND WARNINGS ................................................... 21 Transport, storage and operation conditions .............................................................. 21 Sensitivity to predictable environmental conditions in normal use situations .................. 21 Precautions and warnings “during equipment installation” ........................................... 21 Recommendations for equipment maintenance .......................................................... 22 Precautions and warnings “during equipment use” ..................................................... 22 Precautions and warnings “after” equipment use ........................................................ 22 Precautions and warnings during equipment “cleaning and disinfection” ........................ 23 Precautions in case of change in the equipment operation ........................................... 23 Precautions for reducing environmental impact .......................................................... 23 Precautions in case of equipment disuse ................................................................... 24 Precautions to be taken against foreseeable or unusual risks, related to equipment

    deactivation and abandonment .............................................................................. 24

    CONSERVATION AND CORRECTIVE AND PREVENTIVE MAINTENANCE ................. 25 Additional procedures for reuse ................................................................................ 25 Cleaning ............................................................................................................... 25 Disinfection ........................................................................................................... 25 Preventive Maintenance .......................................................................................... 26 Corrective Maintenance ........................................................................................... 26

    UNFORESEEN EVENTS - TROUBLESHOOTING ....................................................... 26

    EQUIPMENT WARRANTY ..................................................................................... 27

    FINAL CONSIDERATIONS .................................................................................... 27

    3

  • EQUIPMENT IDENTIFICATION

    Dear Client Congratulations on your excellent choice. For buying equipment with “Dabi Atlante”

    quality, you can be absolutely sure that you are purchasing products with technology compatible with the best in class in the world.

    This manual gives you an overview of your equipment. It describes important details that can guide you in its correct use, as well as in solving small problems that may eventually arise.

    We advise you to read it thoroughly and keep it for future reference.

    Product Identification Technical Name: Dental Chair Commercial Name: New Versa Chair Brand: Dabi Atlante

    4

    Owner’s Manual – New Versa Chair

  • 5

    EQUIPMENT IDENTIFICATION

    Principles and fundamentals applied to the product operation Chair for accommodation of the patient during dental treatment, activated by the

    control pedal, where commands are sent from the electronic board to two motors responsible for the movement of the seat mechanism and chair back.

    Equipment Purpose The purpose of this equipment is to accommodate the patient in a comfortable way for dental practices.

    Equipment Description The product was conceived in order to maximize the professional dental surgeon's and

    the patient's experience, by bringing features that bring benefits on a daily basis, such as:

    - Lateral lifting system: greater accessibility and ease of practice.

    - Seated patient entry: easier for patients with mobility limitations.

    - Trendelenburg position: automatic position with the chair movement.

    - Large tray: space to accommodate all instruments. The equipment was constructed using a range of specific materials for each function,

    namely: steel, cast iron and aluminum in its structure; polyurethane, PVC and leather in its upholstery, steel, plastic and integral-skin in the finishing; plastic or glass in the reflector.

    In addition, copper, acrylic, silicone, silicon, tin, among others is used.

    Equipment Indication This equipment is for dental use only, and must be used and handled by a qualified

    person (duly regulated professional, according to the country local legislation) observing the instructions contained in this manual.

    It is the user's obligation to only use the equipment in perfect condition and to protect him/herself, patients and third parties against possible dangers.

    5

  • MODULES, ACCESSORIES, OPTIONS AND CONSUME MATERIALS

    The drawing illustrates the equipment with all optional items, your equipment will consist only of the items chosen during your purchase option.

    01 02 03

    05

    04 08

    06 07

    01 - Headrest

    02 - Chest backrest

    03 - Fixed arms

    04 - Seat

    05 - Footrest

    06 - Joystick

    07 - Integral Skin Base *08 - Multifunctional pedal

    * Optional item

    6

    Owner’s Manual – New Versa Chair

  • 7

    TECHNICAL SPECIFICATIONS

    General features

    Product Classification:

    As per NBR IEC 60601-1

    Feeding

    127/220 V~ Selectable

    Frequency

    50/60 Hz

    Type of protection against electric shock

    Class I Equipment

    Degree of protection against electric shock

    Type B Applied Part

    Mode of operation

    Continuous, with intermittent charge - 1min. work and 4 min. rest

    Protection against harmful water penetration

    IP00 - The entire chair for exams, except the command pedal

    IP01 - Command pedal

    Input power

    530VA

    Protection Fuses

    F1 and F2 127V - T10A/250V

    F1 and F2 220V - T5A/250V

    Lifting capacity

    200 kg distributed load (patient mass + accessories and equipment)

    7

  • TECHNICAL SPECIFICATIONS

    The parts of this equipment are not subject to degradation due to wear, corrosion, fatigue or aging of materials, during its shelf-life.

    Emergency stop The chair has an automatic movement stop button - “Emergency Stop”. This key, when pressed, automatically blocks all movements and functions of the

    chair, and you just have to activate it again for the release and normal operation of the equipment, maintaining the previously recorded programs and positions.

    Overcurrent protection The equipment is equipped with a safety device for overcurrent, by means of 2 fuses

    and internal overcurrent protection for equipment attached to the chair.

    Electronic System The chair has an integrated electronic system with low voltage, which allows total

    interface with reflectors, equipment and electronic water units Dabi Atlante, providing greater safety to the patient and the professional, in addition to greater durability of the components.

    Parts lists and circuit diagrams Dabi Atlante declares that the provision of Circuit Schemes, Parts List or any other

    information that provides technical assistance by the user, may be requested, as long as previously agreed, between the user and the Company Dabi Atlante.

    The materials used in the construction of the equipment are Biocompatible.

    8

    Owner’s Manual – New Versa Chair

  • 9

    TECHNICAL SPECIFICATIONS

    Electromagnetic emissions

    Electromagnetic emissions

    New Versa Chair is intended for use in the electromagnetic environments specified below. The customer or the user of the New Versa Chair should assure that it is used in such an environment.

    Emission Assay Compliance Electromagnetic environment - Guide

    RF Emissions ABNT NBR IEC CISPR 11

    Group 1 New Versa Chair uses RF energy only for its internal function. However, its RF emissions are very low and are not likely to cause any Interference with nearby electronic equipment. RF Emissions

    ABNT NBR IEC CISPR 11 Class B

    IEC Hormone Emissions

    61000-3-2

    Class A New Versa Chair is convenient for use in all establishments, including domestic establishments and those directly connected to a public low voltage electricity supply network that powers buildings used for domestic purposes. Voltage fluctuations/ Flicker

    emissions IEC 61000-3-3

    Complies

    9

  • TECHNICAL SPECIFICATIONS Manufacturer's guidelines and declaration - Electromagnetic immunity

    New Versa Chair is intended for use in the electromagnetic environments specified below. The customer or the user of the New Versa Chair should ensure that it is used in such an environment.

    Immunity Assay

    Test Level ABNT NBR IEC 60601

    Compliance Level Electromagnetic Environment Guidelines

    Electrostatic Discharge (ESD) IEC 6100-4-2

    ± 6 kV Contact ± 8 kV Air

    ± 6 kV Contact ± 8 kV Air

    Floors should be wooden, concrete or ceramic. If the floor is covered with synthetic material, the relative humidity should be at least 30%.

    Fast electrical transients/ pulse train ("brust") IEC 61000-4-4

    ± 2 kV on power lines ± 1 kV on the input/ output lines

    ± 2 kV on power lines ± 1 kV on the input/ output lines

    It is recommended that the quality of the energy supply be that of a typical hospital or commercial environment.

    Outbreaks IEC 61000-4-5

    ± 1 kV line(s) to line(s) ± 2 kV line(s) to land

    ± 1 kV line(s) to line(s) ± 2 kV line(s) to land

    It is recommended that the quality of the energy supply be that of a typical hospital or commercial environment.

    Voltage drops, short interruptions and voltage variations in the power supply input lines

    IEC 61000-4-11

    < 5% Ut (>95% Ut drop) for 0.5 cycle 40% Ut (60% Ut drop) for 5 cycles 70% Ut (30% Ut drop) for 25 cycles < 5% Ut (>95% Ut drop) for 5s

    < 5% Ut

    (>95% Ut drop) for 0.5 cycle

    40% Ut

    (60% Ut drop) for 5 cycles

    70% Ut

    (30% Ut drop) for 25 cycles

    < 5% Ut

    (>95% Ut drop) for 5s

    It is recommended that the quality of the energy supply be that of a typical hospital or commercial environment. If the New Versa Chair user requires continued operation during power interruptions, it is recommended that the New Versa Chair be powered by an uninterruptible power supply or a battery.

    Magnetic field generated by the electric frequency (50/60Hz)

    IEC 61000-4-8

    3 A/m 3 A/m If image distortion occurs, it may be necessary to position the equipment away from the supply frequency or install magnetic shielding. The frequency magnetic field should be measured at the Installation site to ensure that it is low enough.

    NOTE Ut is the supply voltage c.a. before applying the test level

    10

    Owner’s Manual – New Versa Chair

  • 11

    TECHNICAL SPECIFICATIONS Manufacturer's guidelines and declaration - Electromagnetic immunity

    New Versa Chair is intended for use in the electromagnetic environments specified below. The customer or the user of the New Versa Chair should ensure that it is used in such an environment.

    Immunity Assay

    Test Level ABNT NBR IEC 60601

    Compliance Level

    Electromagnetic Environment Guidelines

    Rf conducted IEC 61000-4-6

    Rf radiated IEC 61000-4-3

    3 vrms 150 kHz up to 80 MHz

    3 V/m 80MHz up to 2.5 GHz

    3 Vrms

    3 V/m

    It is recommended that portable and mobile RF communication equipment should not be used near any part of the New Versa Chair including cables, with less than the recommended separation distance, calculated from the equation applicable to the transmitter frequency.

    Recommended separation distance: d = 1.2√P

    d = 1.2√P 80 MHz to 800MHz

    d = 2.3√P 800 MHz to 2.5MHz

    Where P is the maximum nominal output power of the transmitter, and d is the recommended separation distance in meters (m). It is recommended that the field strength from the RF transmitter, as determined through on-site electromagnetic inspection, be less than the level of compliance in each frequency rangeb.

    Interference may occur in the vicinity of equipment marked with the following symbol:

    NOTE 1 At 80MHz and 800MHz, the highest frequency range is applied.

    NOTE 2 These guidelines may not be applicable in all situations. Electromagnetic propagation is affected by

    absorption and reflection from structures, objects and people.

    a - Field strength from fixed transmitters, such as radio base stations for telephones (mobile or cordless phone) and mobile ground radios, amateur radio, AM and FM radio broadcasts and TV broadcasts cannot be theoretically predicted precisely. In order to assess the electromagnetic environment generated by fixed RF transmitters, an electromagnetic field survey should be considered. If the field strength measured at the site where the New Versa Chair will be used exceeds the applicable RF compliance level defined above, the New Versa Chair should be observed to verify that it is functioning normally. If abnormal performance is detected, additional measures may be necessary, such as reorienting or relocating the New Versa Chair.

    b - Above the frequency range of 150kHz to 80 MHz, it is recommended that the field strength is less than 3 V/m.

    11

  • TECHNICAL SPECIFICATIONS

    Recommended separation distances between mobile, portable RF communication equipment and the New Versa Chair

    The New Versa Chair is intended for use in an electromagnetic environment in which RF disturbances are controlled. The customer or user of the New Versa Chair can help prevent electromagnetic interference by maintaining a minimum distance between mobile and portable RF communication equipment (transmitters) and the New Versa Chair as recommended below, according to the maximum output power of the communication equipment.

    Maximum output power w

    Separation distance according to the transmitter frequency

    150 kHz up to 80 MHz d= 1.2√p

    80 kHz up to 800 MHz d= 1.2√p

    800 kHz up to 2.5 GHz d= 2.3√p

    0.01 0.12 0.12 0.23

    0.1 0.38 0.38 0.73

    1 1.2 1.2 2.3

    10 3.8 3.8 7.3

    100 12 12 23

    For transmitters with a maximum nominal output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation applicable to the transmitter frequency, where P is the maximum nominal output power of the transmitter in watts (W), according to the transmitter manufacturer.

    NOTE 1 At 80MHz and 800MHz, the separation distance for the highest frequency range is applied.

    NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by the absorption and reflection of structures, objects and people.

    12

    Owner’s Manual – New Versa Chair

  • 13

    TECHNICAL SPECIFICATIONS

    Dimensions (mm)

    13

  • TECHNICAL SPECIFICATIONS

    Packing symbols

    Maximum stacking, it determines the maximum amount of cartons that can be stacked during transportation and storage “according to packaging”.

    It determines that the packaging must be stored or transported with the arrow side up.

    It determines that the packaging must be stored and transported with caution (it must not fall or be impacted).

    Product Symbols

    Caution: It indicates important

    instruction for product operation. Failure to follow it can cause dangerous malfunction

    Note: It indicates useful

    information for product operation

    Important: It indicates safety instructions for product operation. Failure to follow it can result in serious danger to the patient

    Grounding (at various points on the equipment) indicates the condition of being grounded

    Spitting Position Activation / Return to last position

    It determines that the packaging must be stored or transported protected from humidity (do not expose it to rain, water splashes or floors.

    It determines that the

    packaging must be stored or transported protected from light.

    It determines the temperature limits within which the packaging must be stored or transported.

    On position

    Off position

    Type B

    Seat up

    Seat down

    Backrest up

    Backrest down

    It determines the working

    positions ”1 and 2”

    It determines the working

    positions ”3 and 4”

    14

    Owner’s Manual – New Versa Chair

  • 15

    TECHNICAL SPECIFICATIONS

    Product Symbols

    Determines the position back to zero Reflector activation:

    Authorized representative in the European community

    Reversing the rotation direction of the electric micro motor

    Warning – refer to manual ! Emergency stop

    EQUIPMENT INSTALLATION

    The installation of this equipment requires the need for specialized

    technical assistance (Dabi Atlante).

    This information is also part of the equipment Installation and Maintenance Manual that is in the possession of the authorized technical representative Dabi Atlante.

    For your safety, the Chair leaves the factory positioned at 220V~, so make sure that your network is compatible before connecting it, if it is 127V~ request the presence of an authorized Dabi Atlante technician.

    - This equipment can only be unpacked and installed by an authorized Dabi Atlante technician, under penalty of loss of warranty, since only he/ she has the information, the

    appropriate tools and the training necessary to perform this task. - Dabi Atlante is not responsible for damage or accidents caused by poor installation

    carried out by an unauthorized Dabi Atlante technician. - Only after the equipment has been installed and properly tested by the authorized

    representative Dabi Atlante technician, will it be ready to start work operations.

    15

  • EQUIPMENT INSTALLATION

    IMPORTANT: Dabi Atlante dental office is dimensioned so that the stability of the set consisting of

    dental chair, equipment, reflector and water unit remains stable under normal use conditions. If, at the client's discretion, there is a need to install other accessories to the office, Dabi Atlante informs that it is the client's responsibility to assess the stability, and suggests that in this case, the base of the chair is fixed to the floor through the holes already arranged at its base. This makes the added weight not to compromise the set stability.

    How to fix the chair base to the floor The base of the chair must be fixed to the floor using S-10 Plastic Bushing, Hexagonal

    Wood Steel Screw 5/16 ”x 2” ¼, New Versa Office Fixation Plate, indicated for this purpose, provided by Alliage.

    Failure to attach the base of the chair to the floor can cause instability and in extreme cases the chair fall. There is no safe level of use of the chair unless its base is attached to the floor.

    In the event of a fall or instability of the chair not affixed to the floor that damages it or causes damage to operators or patients, the responsibility will be the sole responsibility of the owner.

    16

    Owner’s Manual – New Versa Chair

  • 17

    EQUIPMENT OPERATION Main switch

    Turn on the main switch located on the back of the product.

    Turn off the office after use or at the end of the day.

    The main switch has an internal LED that stays on/ off when the office is on/ off.

    If the Chair is in any movement (seat and/or back, working position) and at this moment there is a power outage or accidental disconnection of the main switch, when the power supply returns or main switch is turned on, the key back-to-zero must be activated and wait for the Chair to complete this movement, for its proper operation.

    17

    Main switch

  • EQUIPMENT OPERATION

    Seat Movement Activate the joystick vertically up/ down to lower/ raise the seat. To stop the movement, just stop activating the joystick.

    Backrest Movement Activate the joystick laterally to the left/ right to raise/ lower the backrest. To stop the movement, just stop activating the joystick.

    Longitudinal movement of the headrest - Optional

    Rotate the handle located on the back of the Chair back counterclockwise to unlock the

    backrest, place it in the desired location and rotate the handle clockwise to fix it.

    18

    Owner’s Manual – New Versa Chair

  • 19

    EQUIPMENT OPERATION

    Positioning the bi-articulated headrest

    Rotate the handle counterclockwise to release the mechanism. Position the backrest in the desired location and turn the handle clockwise to lock the mechanism. Make sure that the assembly has been sufficiently tightened to prevent movement during use.

    Multiarticulated Headrest

    Press and hold the pneumatic lock to release the set brake. Position the backrest in the desired position, when reaching the desired position, release the lock to lock the set.

    19

    Handle

    Pneumatic lock

  • EQUIPMENT OPERATION

    Multifunctional Pedal The Multifunctional Pedal can be used in two situations, when the hand piece is out of the holder, the pedal assumes the same functions as the electric Micro Motor pedal, and when the hand piece is in the tip holder, the pedal resumes its function as the Chair pedal.

    1 - Chip Blower

    2 – Back to Zero

    3 - Micro Motor Memory

    4 - Reflector

    5 - Dry tip actuation (pushing the pedal sliding lever to the left)

    6 - Counterclockwise (implant removal) by pushing the pedal sliding lever to the left

    7 – Working position

    8 - Peristaltic Pump

    9 - Spitting Position

    10 - Spray tip actuation (by pushing the pedal sliding lever to the right)

    11 - Clockwise (implant insertion) by pushing the pedal sliding lever to the right

    20

    Owner’s Manual – New Versa Chair

  • 21

    PRECAUTIONS, RESTRICTIONS AND WARNINGS

    Transport, storage and operation conditions The equipment must be transported and stored under the following conditions:

    - Carefully, so as not to fall or receive impacts. - With moisture protection, do not expose to rain, water splash or wet floor. - Keep in a place protected from rain and direct sunlight and in its original packaging. - When transporting, do not move it on uneven surfaces and protect the packaging from direct rain and respect the maximum stacking informed on the outside of the packaging. - Room temperature range for transport or storage -12ºC to + 50ºC. - Room temperature range recommended by Dabi Atlante + 10ºC to + 35ºC.

    WARNING: The equipment must be used only with the power input cable supplied by Dabi Atlante, otherwise it may cause an increase in emissions or a decrease in the equipment immunity.

    The Equipment maintains its safety and effectiveness condition, as long as it is

    maintained (stored) as mentioned in this instruction for use. In this way, the equipment will not lose or change its physical and dimensional characteristics.

    Sensitivity to predictable environmental conditions in normal use situations

    - The equipment was designed not to be sensitive to interference such as magnetic fields, external electrical influences, electrostatic discharges, pressure or pressure variation, as long as the equipment is installed, maintained, cleaned, stored, transported and operated in accordance with this instruction for use. .

    Precautions and warnings “during equipment installation” - The equipment must be installed only by authorized Dabi Atlante technicians or

    technical assistance. - Position the unit in a place where it will not be wet.

    - Install the equipment in a place where it will not be damaged by pressure, temperature, humidity, direct sunlight, dust or salts.

    - The equipment must not be subjected to inclination, excessive vibrations, or shocks (including during transport and handling).

    - This equipment is not designed for use in an environment where vapors, anesthetic mixtures which are flammable with air, or oxygen and nitrous oxide can be detected.

    - Check the equipment voltage when making the electrical installation.

    - Before first use and/or after long work interruptions such as holidays, clean and disinfect the equipment.

    This information is also part of the equipment Installation and Maintenance Manual that is in the possession of the authorized technical representative Dabi Atlante.

    21

  • PRECAUTIONS, RESTRICTIONS AND WARNINGS

    Recommendations for equipment maintenance Your Dabi Atlante equipment was designed and enhanced within the standards of

    modern technology. Like all devices, it needs special care, which is often overlooked for different reasons and circumstances.

    So here are some important reminders for your daily life. Try to observe these small rules, which, incorporated into the work routine, will save time and avoid unnecessary expenses.

    Precautions and warnings “during equipment usage” - The equipment must be operated only by duly qualified and trained technicians

    (Dental Surgeons, Trained Professionals) - If any maintenance is needed, use only Dabi Atlante Authorized Technical Assistance

    services. - The equipment was manufactured to withstand continuous and intermittent

    operation; therefore, follow the cycles described in these Instructions for Use. - Although this equipment has been designed in accordance with the standards of

    electromagnetic compatibility, it can, in very extreme conditions, cause interference with other equipment. Do not use this equipment in conjunction with other devices that are very sensitive to interference or with devices that create high electromagnetic

    disturbances. - Do not subject the plastic parts to contact with chemical substances, used in dental

    treatment routines. Such as: acids, mercury, acrylic liquids, amalgams, etc. - When moving the chair, never lift it by the leg support part, lift it by the structure in

    the seat region.

    Dabi Atlante shall not be responsible for: - Use of equipment other than that for which it is intended. - Damage caused to the equipment, the professional and/or the patient due to

    incorrect installation and erroneous maintenance procedures, different from those described in these Instructions for Use that accompany the equipment or by its incorrect operation.

    Precautions and warnings “after” equipment use - Turn off the office's main switch when not in use for a long time. - Keep the equipment clean for the next operation. - Do not modify any part of the equipment. Do not unplug the cable or other

    connections unnecessarily. - After using the equipment, clean and disinfect all parts that may be in contact with

    the patient.

    22

    Owner’s Manual – New Versa Chair

  • 23

    PRECAUTIONS, RESTRICTIONS AND WARNINGS

    Precautions and warnings during equipment “cleaning and disinfection”

    - Before cleaning the equipment, turn off the main switch. - Avoid spilling water or other liquids inside the equipment, which could cause short circuits.

    Care for cleaning upholstery (PVC laminated "courvin" or leather) Never use any type of alcohol, micro abrasive material, detergents that contain solvents such as ether or stain-remover for cleaning upholstery, as this may cause stains.

    Procedure for cleaning PVC laminate (courvin) upholstery: - We recommend the use of the product BactSpray (manufactured by AEROPAC

    INDUSTRIAL LTDA and distributed by DCMA PRODUTOS MÉDICOS ODONTOLÓGICOS LTDA), or cloth moistened in water with mild soap.

    Procedure for cleaning leather upholstery: - For cleaning and hydrating leather upholstery, we recommend the use of products

    specifically intended for this purpose.

    - Never wet or rub the leather thoroughly. If any dirt that may stain the leather reaches it, wipe it off immediately with a cloth moistened with water with mild soap.

    Precautions in case of change in the equipment operation - If the equipment has any abnormality, check if the problem is related to any item listed

    in the topic unforeseen events (failures, causes and solutions). If it is not possible to solve the problem, turn off the equipment, unplug the power cable from the outlet and contact your representative (Dabi Atlante).

    Precautions for reducing environmental impact Alliage S/A aims to achieve an environmental policy to promote the supply of environmentally conscious medical and dental products that continuously minimize environmental impact and are friendlier to the environment and human health. In order to maintain a minimal environmental impact, observe the recommendations below:

    - After installation, submit the recyclable materials for recycling process. - During the life cycle of the equipment, turn it off when it is not in use. - In order to prevent environmental contamination, disposable consumables must follow the

    normal dental office procedure for biomedical waste. Biomedical waste includes non-acute materials that may cause disease or suspicion of

    harboring pathogenic organisms, which must be stored in a yellow bag properly labeled with a biohazard symbol, stored in a puncture-resistant, watertight container until collection and incineration.

    The dental office packaging is composed of wood, cardboard and plastic, which are 100% recyclable materials.

    23

  • 24

    PRECAUTIONS, RESTRICTIONS AND WARNINGS Precautions in case of equipment disuse

    In order to avoid environmental contamination or misuse of the dental office, when it is disused, they must be discarded (according to current legislation) in an appropriate place, as the materials existing inside may contaminate the environment. For the European Economic Area (EEA), this product is subject to Directive 2012/19/EU, as well as the corresponding national laws. This directive requires that the product must be taken to a special waste collection site at the end of its shelf-life. Applies to both the device and accessories. Contact your dealer if final product destination is required.

    This equipment must not be disposed of as household waste.

    Precautions to be taken against foreseeable or unusual risks, related to equipment deactivation and abandonment

    In order to avoid environmental contamination or misuse of the Equipment and its accessories/ parts after disuse, it must be disposed of in an appropriate site (according to local country legislation).

    Pay attention to the country local legislation for the conditions of installation and waste disposal.

    Precautions to be taken against foreseeable or unusual risks, related to equipment deactivation and abandonment

    - In order to avoid environmental contamination or misuse of the Equipment and its accessories/ parts after disuse, it must be disposed of in an appropriate site (according to local country legislation).

    - Pay attention to the country local legislation for the conditions of installation and waste disposal.

    24

  • CONSERVATION AND CORRECTIVE AND PREVENTIVE MAINTENANCE

    Additional procedures for reuse The equipment is reusable in indeterminate quantities, i.e., unlimited, requiring only

    cleaning and disinfection.

    Cleaning Important: To perform cleaning or any type of maintenance, make sure that the

    equipment is disconnected from the mains.

    The cleaning procedure below must be done at the beginning of the day and after each patient. Always turn off the main switch before performing daily maintenance procedures.

    To clean your equipment, we recommend the use of the product “BactSpray” (MS Reg. No.: 3.2079.0041.001-5) or one that has similar characteristics:

    Active Ingredient: Benzalkonium Chloride (Tri-quaternary Ammonium) 50% Sol. ................................................ 0.329%

    Product Composition: Butylglycol, Decyl Polyglucose, Sodium Benzoate, Sodium Nitrite, Essence, Deodorized Propane / Butane,

    Demineralized water. For more information on cleaning procedures, refer to the

    manufacturer's instructions on the product.

    WARNING: • This product can also be used for cleaning and disinfecting

    the water unit basin. • In order to eliminate safety risks or damage to the

    equipment, we recommend that, when cleaning, there is no liquid penetration inside the equipment.

    • The application of other cleaning chemicals based on solvents or sodium hypochlorite is not recommended, as they can

    damage the equipment.

    NOTE: “BactSpray” Registration with the Ministry of Health is carried out separately

    from the product described in this manual, as “BactSpray” is not manufactured by Dabi Atlante.

    Disinfection - To disinfect the equipment, use a clean, soft cloth moistened with 70% alcohol

    (except for upholstery). - Never use corrosive disinfectants or solvents.

    Note: Wear gloves and other protective systems when disinfecting.

    25

    Owner’s Manual – New Versa Chair

  • 26

    I

    CONSERVATION AND CORRECTIVE AND PREVENTIVE MAINTENANCE

    Preventive Maintenance The equipment must undergo routine measurements, according to the legislation in

    force in the country. However, never with a period exceeding 3 years. To protect your equipment, seek Dabi Atlante technical assistance for periodic

    preventive maintenance reviews.

    Corrective Maintenance If the equipment shows any abnormality, check if the problem is related to any of the

    items listed in item Unforeseen Events (situation, cause and solution). If it is not possible to solve the problem, turn off the equipment, and request Dabi Atlante technical assistance.

    UNFORESEEN EVENTS - TROUBLESHOOTING

    If you encounter any problem in the operation, follow the instructions below to check and fix the problem, and/or contact your representative.

    Unforeseen Events Probable Cause Solutions

    - Completely inoperative

    chair.

    - Anti-crush device activated.

    - Release anti-crush device

    - “Emergency stop” button activated.

    - Plug disconnected from

    the outlet.

    - Lack of electricity

    - Main switch off.

    - Fuse(s) blown.

    - Press the “Emergency stop” button again. - Connect the plug to the outlet.

    - Wait for the network to normalize.

    -Turn on main switch. -Turn off the chair power and request the presence of a Technician.

    26

  • EQUIPMENT WARRANTY This equipment is covered by the warranty periods from the date of installation, as

    specified below; provided that the defect has occurred under normal conditions of use and that the equipment has not been stored for more than 06 months from the invoice issuance date until the actual installation date.

    - WARRANTY DEADLINES: Check the warranty certificate; - LOSS OF WARRANTY: A) Attempted repair using an inappropriate tool or by unauthorized technicians; B) Installation of the equipment by an unauthorized technician; C) Damage resulting from improper storage or signs of violation; D) Misuse of the equipment; E) Use of cleaning product not recommended by the manufacturer; F) Falls or crashes that the equipment may suffer or lack of observation and

    compliance with the Owner's manual, which was delivered with the equipment. Repair or replacement of parts during the warranty period will not extend the warranty's validity period.

    - This warranty does not exempt the customer from paying the service fee for the visit and the travel expenses of the technician, except when the customer sends the equipment to carry out maintenance within the technical assistance establishment.

    “Consumer Protection Code - art. 50, sole paragraph”. - The Warranty Certificate accompanies the product and must be completed on the date

    of installation by the Dabi Atlante Authorized Technician. - Questions and information: Dabi Atlante Customer Service at (+55) 16 3512-1212. - Check the warranty term attached in this manual.

    FINAL CONSIDERATIONS Among the precautions you must take with your equipment, the most important is

    regarding spare parts. In order to guarantee the shelf-life of your device, only replace original Dabi

    Atlante parts. They are guaranteed the standards and technical specifications required by Dabi Atlante representative.

    We draw your attention to our network of authorized resellers. Only it will keep your equipment constantly new, as it has trained technical assistants and specific tools for the correct maintenance of your device.

    Whenever you need it, request the presence of a Dabi Atlante representative technician at the nearest retailer, or request it through Dabi Atlante Customer Service: + 55 (16) 3512-1212.

    27