osics multifunction platform - yonsei...

252
www.yenista.com OSICS Multifunction Platform 8-Channel Modular Platform User Manual OSICS_UM_3v2.2

Upload: others

Post on 24-Jan-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

www.yenista.com

OSICS Multifunction Platform8-Channel Modular Platform

User Manual

OSICS_UM_3v2.2

Page 2: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions
Page 3: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 3

Legal Notice

Copyright

Copyright © 2013–2017 by Yenista Optics. Published by Yenista Optics. All rights reserved.This documentation is provided as a user manual to Yenista Optics’ customers and potential customers only. The contents of this document may not be reproduced in any part or as a whole, transcribed, stored in a retrieval system, translated into any language, or transmitted in any form or by any means (electronic, mechanical, magnetic, optical, chemical, photocopying, manual, or otherwise) without the prior written permission of Yenista Optics.

Trademarks

"Yenista Optics", "Yenista" and "TUNICS" are registered trademarks of Yenista Optics. Other trademarks mentioned in this publication are used for identification purposes only.

Product Warranty and Limitation of Warranty

This Yenista Optics instrument product is warranted against defects in material and workmanship for a period of one year from date of shipment. During the warranty period, Yenista will repair products that prove to be defective. For warranty service or repair, such product must be returned to Yenista. Please contact Yenista support services for the appropriate procedure. Yenista does not accept any equipment shipped back without following the procedure.Yenista Optics warrants that its software and firmware designated by Yenista for use with an instrument will execute its programming instructions when properly installed on that instrument. Yenista does not warrant that the operation of the instrument, software, or firmware will be uninterrupted or error-free.The foregoing warranty shall not apply to defects resulting from improper or inadequate maintenance by Buyer, Buyer-supplied software or interfacing, unauthorized modification or misuse, operation outside of the environmental specifications for the product, or improper site preparation or maintenance.

Disclaimers

No Warranties: This documentation is provided "as is" without any ex-touch or implied warranty of any kind.Limitation of Liability: Yenista Optics does not assume any liability arising out of the application or use of any products, or software described herein. In no event shall Yenista Optics or its affiliate companies be liable for any damages whatsoever (including, without limitation, consequential or incidental damages, damages for loss of profits, or otherwise) arising out of the use of the information provided in this documentation.The contents of this publication have been carefully checked for accuracy. However, Yenista Optics makes no warranty as to the completeness, correctness, or accuracy of the information contained within this documentation. In the interest of continued product improvement, Yenista Optics further reserves the right to make changes to any products described herein without notice. This publication is subject to change without notice.

Page 4: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

4 OSICS User Manual

Contact Information

Headquarters North American Office China Office

Yenista Optics4 rue Louis de Broglie22300 LannionFRANCE

Phone: +33 2 96 48 37 16Fax: +33 2 96 48 73 04Website: www.yenista.com

Yenista Optics Inc. 2393 Teller Road, Suite 102Newbury Park, CA 91320-6089USA

Phone: +1 805 367 4075Fax: +1 609 423 0891

Yenista OpticsNo.555 WuDing Road, Room 834Jing’an DistrictShanghai 200040CHINA

Phone: +86 21 3251 7155

Sales

Sales AmericasEmail: [email protected]: +1 805 367 4075

Sales ChinaEmail: [email protected]: +86 21 3251 7155

Sales Europe, Middle East and Africa Email: [email protected]: +33 296 483 716

Sales Asia and PacificEmail: [email protected]

Customer Support and Repair Services

Americas Email: [email protected]

ChinaEmail: [email protected]

Europe, Middle East and AfricaEmail: [email protected]

Asia and PacificEmail: [email protected]

Page 5: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 5

About This Manual

Subject This manual explains how to install, set-up and use the OSICS multifunction modular platform composed of the OSICS Mainframe and all the available OSICS pluggable modules.

Application Information in this document applies to the OSICS Mainframe version 3.04 and the following versions of OSICS modules:• OSICS ATN v. 1.06• OSICS BKR v. 1.06• OSICS DFB v. 2.37• OSICS ECL v. 3.04• OSICS SLD v. 1.02• OSICS SWT v. 1.06• OSICS SWT-APC v. 1.11• OSICS T100 v. 3.04• OSICS TLS-50 v. 3.11• OSICS TLS-AG v. 3.11

Intended Readers Users of this manual must be familiar with fiber optic technology.

Date 5 April 2017

Manual Reference OSICS_UM_3v2.2

Typographical Conventions

Symbols

bold Identifies interface objects such as menu names, labels, buttons and icons.

italic Identifies references to other sections or other guides.

monospace Identifies portions of program codes, command lines, or messages displayed in command windows.

IMPORTANT Identifies important information to which you must pay particular attention.

Identifies conditions or practices that could result in injury or loss of life.

Identifies conditions or practices that could result in damage to the product or other property.

WARNING

CAUTION

Page 6: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

About This Manual

6 OSICS User Manual

Page 7: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 7

Table of Contents

Legal Notice ........................................................................................................................................................ 3

Contact Information .......................................................................................................................................... 4

About This Manual............................................................................................................................................. 5

Table of Contents............................................................................................................................................... 7

Safety Information........................................................................................................................................... 13

1. Product Presentation................................................................................................................................ 151.1 Product Features.............................................................................................................................................................. 151.2 OSICS Mainframe Technical Specifications ........................................................................................................... 161.3 Product Overview ............................................................................................................................................................ 17

1.3.1 OSICS Front Panel .......................................................................................................................................... 171.3.2 OSICS Rear Panel ............................................................................................................................................ 181.3.3 OSICS Information and Warning Markings .......................................................................................... 19

2. Installing and Connecting the OSICS Multifunction Platform............................................................. 212.1 Unpacking and Installing the OSICS Mainframe ................................................................................................. 212.2 Connecting the OSICS Mainframe to a Power Source...................................................................................... 222.3 Handling OSICS Modules Into the OSICS Mainframe....................................................................................... 23

2.3.1 Installing an OSICS Module Into the OSICS Mainframe.................................................................. 232.3.2 Removing an OSICS Module From the OSICS Mainframe ............................................................. 25

2.4 Connecting Optical Devices to OSICS Modules .................................................................................................. 262.5 Installing the Remote Interlock .................................................................................................................................. 27

3. Turning On/Off and Initializing the OSICS Multifunction Platform ................................................... 293.1 Turning on the OSICS Multifunction Platform and Accessing the System Status Screen................... 293.2 Accessing the OSICS Setup Menus........................................................................................................................... 333.3 Checking the OSICS Version ....................................................................................................................................... 343.4 Modifying the OSICS Mainframe Password .......................................................................................................... 353.5 Turning Off the OSICS Multifunction Platform .................................................................................................... 36

4. Controlling the Optical Output of all OSICS Modules Simultaneously .............................................. 374.1 Enabling/Disabling the Light Source of All Modules......................................................................................... 374.2 Configuring the Optical Output Power of All Modules .................................................................................... 38

4.2.1 Selecting the Power Unit for All Installed Modules .......................................................................... 384.2.2 Setting the Optical Output Power of All Installed Modules .......................................................... 38

4.3 Selecting the Spectral Unit for all Installed Modules ........................................................................................ 394.4 Configuring the OSICS Mainframe Modulation .................................................................................................. 40

4.4.1 Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source ........................... 414.4.2 Setting Up Digital (TTL) Modulation Using an External Source.................................................... 434.4.3 Using the Mainframe as a Modulation Source for External Devices (Synchro Output) ...... 44

5. Setting a Working Configuration ........................................................................................................... 45

Page 8: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Table of Contents

8 OSICS User Manual

5.1 Saving a Working Configuration ............................................................................................................................... 475.2 Loading a Working Configuration ............................................................................................................................ 48

6. Using the OSICS ATN-HP Module........................................................................................................... 496.1 OSICS ATN-HP Presentation ....................................................................................................................................... 49

6.1.1 OSICS ATN-HP Technical Specifications................................................................................................ 496.1.2 OSICS ATN-HP Module Overview............................................................................................................ 50

6.2 Setting the OSICS ATN-HP Optical Attenuation ................................................................................................. 516.2.1 Setting the Attenuation to a Definite Value......................................................................................... 516.2.2 Performing Real-time Adjustments to the ATN Attenuation ........................................................ 526.2.3 Step-Adjusting the ATN Attenuation...................................................................................................... 53

6.3 Setting the ATN Wavelength ...................................................................................................................................... 546.4 Setting the ATN Attenuation Offset ......................................................................................................................... 55

7. Using the OSICS BKR Module.................................................................................................................. 577.1 OSICS BKR Presentation................................................................................................................................................ 57

7.1.1 OSICS BKR Technical Specifications ........................................................................................................ 577.1.2 OSICS BKR Module Overview .................................................................................................................... 58

7.2 Setting the OSICS BKR Optical Reflectance........................................................................................................... 597.2.1 Setting the Reflectance to a Definite Value ......................................................................................... 597.2.2 Performing Real-time Adjustments to the BKR Reflectance.......................................................... 607.2.3 Step-Adjusting the BKR Attenuation ...................................................................................................... 61

7.3 Setting the BKR Wavelength/Frequency ................................................................................................................ 627.4 Setting the BKR Attenuation Offset.......................................................................................................................... 63

8. Using the OSICS DFB Module.................................................................................................................. 658.1 OSICS DFB Presentation................................................................................................................................................ 65

8.1.1 OSICS DFB Technical Specifications ........................................................................................................ 658.1.2 OSICS DFB Module Overview .................................................................................................................... 67

8.2 Enabling/Disabling the Light Source of the Module ......................................................................................... 698.3 Setting the Optical Output Power of the DFB Module..................................................................................... 70

8.3.1 Specifying the Optical Output Power of the Module....................................................................... 708.3.2 Performing Real-time Adjustments to the Output Power.............................................................. 71

8.4 Setting the Optical Emission Wavelength/Frequency....................................................................................... 728.4.1 Specifying the Optical Emission Wavelength/Frequency ............................................................... 728.4.2 Step-adjusting the Wavelength/Frequency Value of the Module .............................................. 73

8.5 Configuring the OSICS DFB Signal Modulation................................................................................................... 748.5.1 Setting Up Digital (TTL) Modulation Using the DFB Module Internal Source ........................ 758.5.2 Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source ........................... 778.5.3 Setting Up Digital (TTL) Modulation Using an External Source.................................................... 808.5.4 Setting Up Analog Modulation Using an External Source ............................................................. 838.5.5 Setting Up Analog Modulation With the Internal Sinus Modulation......................................... 868.5.6 Using the DFB Module as a Modulation Source for External Devices (OUT 2 Output) ...... 90

8.6 Monitoring the OSICS DFB Module Parameters Using the Analog OUT 1 Output ............................... 918.7 Calibrating the DFB Module........................................................................................................................................ 92

8.7.1 Performing a Wavelength Calibration.................................................................................................... 928.7.2 Performing a Power Calibration................................................................................................................ 938.7.3 Asking for a Factory Recalibration ........................................................................................................... 94

Page 9: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Table of Contents

OSICS User Manual 9

8.8 Testing the Performance of the DFB Module....................................................................................................... 958.8.1 Preparing Performance Tests..................................................................................................................... 958.8.2 Measuring the Absolute Wavelength Accuracy.................................................................................. 968.8.3 Verifying the Wavelength Stability .......................................................................................................... 968.8.4 Verifying the Power Stability...................................................................................................................... 97

9. Using the OSICS ECL Module................................................................................................................... 999.1 OSICS ECL Presentation ................................................................................................................................................ 99

9.1.1 OSICS ECL Technical Specifications......................................................................................................... 999.1.2 OSICS ECL Module Overview ................................................................................................................... 101

9.2 Enabling/Disabling the Light Source of the Module ....................................................................................... 1039.3 Setting the Optical Output Power of the ECL Module.................................................................................... 104

9.3.1 Specifying the Optical Output Power of the Module..................................................................... 1049.3.2 Performing Real-time Adjustments to the Output Power............................................................ 105

9.4 Setting the Optical Emission Wavelength/Frequency..................................................................................... 1069.4.1 Specifying the Optical Emission Wavelength/Frequency ............................................................. 1069.4.2 Step-adjusting the Wavelength/Frequency Value of the Module ............................................ 106

9.5 Activating Coherence Control .................................................................................................................................. 1089.6 Activating the Auto-peak Find Function .............................................................................................................. 1099.7 Configuring the OSICS ECL Signal Modulation ................................................................................................. 110

9.7.1 Setting Up Digital (TTL) Modulation Using the ECL Module Internal Source ....................... 1119.7.2 Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source ......................... 1139.7.3 Setting Up Digital (TTL) Modulation Using an External Source.................................................. 1169.7.4 Setting Up Analog Modulation Using an External Source ........................................................... 1199.7.5 Using the ECL Module as a Modulation Source for External Devices (OUT 2 Output) ..... 122

9.8 Monitoring the OSICS ECL Module Parameters Using the Analog OUT 1 Output.............................. 1239.9 Calibrating the ECL Module....................................................................................................................................... 124

9.9.1 Performing a Wavelength Reference Calibration ............................................................................ 1249.9.2 Performing a Wavelength/Frequency (λ/f) Calibration.................................................................. 1249.9.3 Performing a Power Calibration.............................................................................................................. 1269.9.4 Asking for a Factory Recalibration ......................................................................................................... 127

9.10 Testing the Performance of the ECL Module...................................................................................................... 1289.10.1 Preparing Performance Tests................................................................................................................... 1289.10.2 Measuring the Absolute Wavelength Accuracy................................................................................ 1299.10.3 Checking the Wavelength Range and Output Power .................................................................... 1299.10.4 Validating the Side Mode Suppression................................................................................................ 130

10. Using the OSICS SLD Module ................................................................................................................ 13110.1 OSICS SLD Presentation.............................................................................................................................................. 131

10.1.1 OSICS SLD Technical Specifications ..................................................................................................... 13110.1.2 OSICS SLD Module Overview................................................................................................................... 132

10.2 Enabling/Disabling the Light Source of the Module ....................................................................................... 13410.3 Setting the Optical Output Power of the SLD Module ................................................................................... 135

11. Using the OSICS SWT Module ............................................................................................................... 13711.1 OSICS SWT Presentation............................................................................................................................................. 137

11.1.1 OSICS SWT Technical Specifications.................................................................................................... 13911.1.2 OSICS SWT Module Overview ................................................................................................................. 140

Page 10: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Table of Contents

10 OSICS User Manual

11.2 Selecting the OSICS SWT Input/Output ............................................................................................................... 14211.3 Configuring Sequential Switching........................................................................................................................... 143

11.3.1 Selecting the Number of Scans............................................................................................................... 14311.3.2 Setting the Timeframe................................................................................................................................ 14411.3.3 Launching/Stopping a Sequence ........................................................................................................... 145

11.4 Testing the Performance of the SWT Module.................................................................................................... 14611.4.1 Preparing Performance Tests................................................................................................................... 14611.4.2 Verifying the Repeatability........................................................................................................................ 14711.4.3 Verifying the Crosstalk................................................................................................................................ 148

12. Using the OSICS SWT APC Module....................................................................................................... 14912.1 OSICS SWT APC Presentation................................................................................................................................... 149

12.1.1 OSICS SWT APC Technical Specifications .......................................................................................... 15212.1.2 OSICS SWT APC Module Overview........................................................................................................ 153

12.2 Selecting the Operating Mode................................................................................................................................. 15412.3 Selecting the OSICS SWT APC Input/Output Channel (Switch Mode) ..................................................... 15512.4 Configuring the Full-Band Mode ............................................................................................................................ 156

12.4.1 Automatically Detecting the T100 and ECL Modules Connected to the SWT APC ............ 15612.4.2 Manually Configuring the T100 and ECL Modules Connected to the SWT APC ................. 157

12.5 Enabling/Disabling the Light Source of all Connected Modules (Full-Band Mode)............................ 15912.6 Setting the Optical Output Power of the SWT APC Module (Full-Band Mode).................................... 16012.7 Setting the Optical Emission Wavelength/Frequency (Full-Band Mode) ................................................ 161

12.7.1 Specifying the Optical Emission Wavelength/Frequency ............................................................. 16112.7.2 Step-adjusting the Wavelength/Frequency Value of the Module ............................................ 16212.7.3 Scanning a Predefined Wavelength/Frequency Range ................................................................. 163

12.8 Activating Coherence Control on all Connected T100 and ECL Modules (Full-Band Mode)........... 16512.9 Activating the Auto-peak Find Function (Full-Band Mode) .......................................................................... 16612.10 Configuring the Modulation for All Connected T100 and ECL Modules (Full-Band Mode)............. 167

12.10.1 Defining the Modulation Source for All Connected T100 and ECL Modules........................ 16712.10.2 Controlling the Modulation of All Connected T100 and ECL Modules ................................... 168

12.11 Monitoring the Optical Output Power of the SWT APC Module................................................................ 16912.12 Testing the Performance of the SWT APC Module .......................................................................................... 170

12.12.1 Preparing Performance Tests................................................................................................................... 17012.12.2 Verifying the Repeatability........................................................................................................................ 17112.12.3 Verifying the Crosstalk................................................................................................................................ 172

13. Using the OSICS T100 Module .............................................................................................................. 17313.1 OSICS T100 Presentation............................................................................................................................................ 173

13.1.1 OSICS T100 Technical Specifications .................................................................................................... 17313.1.2 OSICS T100 Module Overview................................................................................................................. 175

13.2 Enabling/Disabling the Light Source of the Module ....................................................................................... 17713.3 Setting the Optical Output Power of the T100 Module ................................................................................. 178

13.3.1 Specifying the Optical Output Power of the Module..................................................................... 17813.3.2 Performing Real-time Adjustments to the Output Power............................................................ 179

13.4 Setting the Optical Emission Wavelength/Frequency..................................................................................... 18013.4.1 Specifying the Optical Emission Wavelength/Frequency ............................................................. 180

Page 11: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Table of Contents

OSICS User Manual 11

13.4.2 Step-adjusting the Wavelength/Frequency Value of the Module ............................................ 18113.5 Activating Coherence Control .................................................................................................................................. 18213.6 Activating the Auto-peak Find Function .............................................................................................................. 18313.7 Configuring the OSICS T100 Signal Modulation............................................................................................... 184

13.7.1 Setting Up Digital (TTL) Modulation Using the T100 Module Internal Source..................... 18513.7.2 Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source ......................... 18713.7.3 Setting Up Digital (TTL) Modulation Using an External Source.................................................. 19013.7.4 Setting Up Analog Modulation Using an External Source ........................................................... 19313.7.5 Using the T100 Module as a Modulation Source for External Devices (OUT 2 Output)... 196

13.8 Monitoring the OSICS T100 Module Parameters Using the Analog OUT 1 Output ........................... 19713.9 Calibrating the T100 Module .................................................................................................................................... 198

13.9.1 Performing a Wavelength Reference Calibration ............................................................................ 19813.9.2 Performing a Wavelength/Frequency (λ/f) Calibration.................................................................. 19813.9.3 Performing a Power Calibration.............................................................................................................. 20013.9.4 Asking for a Factory Recalibration ......................................................................................................... 201

13.10 Testing the Performance of the T100 Module ................................................................................................... 20213.10.1 Preparing Performance Tests................................................................................................................... 20213.10.2 Measuring the Absolute Wavelength Accuracy................................................................................ 20313.10.3 Checking the Wavelength Range and Output Power .................................................................... 20313.10.4 Validating the Side Mode Suppression................................................................................................ 204

14. Using the OSICS TLS-50 Module ........................................................................................................... 20514.1 OSICS TLS-50 Presentation........................................................................................................................................ 205

14.1.1 OSICS TLS-50 Technical Specifications ................................................................................................ 20514.1.2 OSICS TLS-50 Module Overview............................................................................................................. 206

14.2 Enabling/Disabling the Light Source of the Module ....................................................................................... 20814.3 Setting the Optical Output Power of the TLS-50 Module ............................................................................. 20914.4 Setting the Optical Emission Wavelength/Frequency..................................................................................... 210

14.4.1 Specifying the Optical Emission Wavelength/Frequency ............................................................. 21014.4.2 Step-adjusting the Wavelength/Frequency Value of the Module ............................................ 21114.4.3 Scanning a Predefined Wavelength/Frequency Range ................................................................. 212

14.5 Setting the SBS Suppression Mode........................................................................................................................ 21414.5.1 Selecting the Modulation Mode............................................................................................................. 21514.5.2 Selecting the SBS Rate................................................................................................................................ 21614.5.3 Selecting the SBS Waveform.................................................................................................................... 21614.5.4 Selecting the Frequency Deviation ........................................................................................................ 21714.5.5 Selecting the Amplitude Deviation........................................................................................................ 21814.5.6 Enabling/Disabling the SBS Suppression ............................................................................................ 219

14.6 Testing the Performance of the TLS-50 Module ............................................................................................... 22014.6.1 Preparing Performance Tests................................................................................................................... 22014.6.2 Measuring the Absolute Wavelength Accuracy................................................................................ 22114.6.3 Checking the Wavelength Range and Output Power .................................................................... 22114.6.4 Asking for a Factory Recalibration ......................................................................................................... 222

15. Using the OSICS TLS-AG Module .......................................................................................................... 22315.1 OSICS TLS-AG Presentation....................................................................................................................................... 223

15.1.1 OSICS TLS-AG Technical Specifications ............................................................................................... 22315.1.2 OSICS TLS-AG Module Overview ........................................................................................................... 224

Page 12: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Table of Contents

12 OSICS User Manual

15.2 Enabling/Disabling the Light Source of the Module ....................................................................................... 22615.3 Setting the Operating Mode..................................................................................................................................... 227

15.3.1 Selecting the Operating Mode................................................................................................................ 22715.3.2 Setting the Grid ............................................................................................................................................. 228

15.4 Setting the Optical Output Power of the TLS-AG Module ............................................................................ 23015.5 Setting the Optical Emission Wavelength/Frequency..................................................................................... 231

15.5.1 Specifying the Optical Emission Wavelength/Frequency ............................................................. 23115.5.2 Step-adjusting the Wavelength/Frequency Value of the Module ............................................ 23215.5.3 Scanning a Predefined Wavelength/Frequency Range ................................................................. 23315.5.4 Fine Tuning the Wavelength/Frequency ............................................................................................. 235

15.6 Setting the SBS Suppression Mode........................................................................................................................ 23715.6.1 Selecting the Optical Frequency Deviation ........................................................................................ 23715.6.2 Enabling/Disabling the SBS Suppression ............................................................................................ 23815.6.3 Disabling the Frequency Control Dither Function ........................................................................... 238

15.7 Testing the Performance of the TLS-AG Module .............................................................................................. 24015.7.1 Preparing Performance Tests................................................................................................................... 24015.7.2 Measuring the Absolute Wavelength Accuracy................................................................................ 24115.7.3 Checking the Wavelength Range and Output Power .................................................................... 24115.7.4 Asking for a Factory Recalibration ......................................................................................................... 242

16. Performing Basic Maintenance Operations ......................................................................................... 24316.1 Cleaning the OSICS Multifunction Platform........................................................................................................ 24316.2 Cleaning Optical Connectors .................................................................................................................................... 24316.3 Replacing the Line Fuse .............................................................................................................................................. 24416.4 Packaging for Shipment.............................................................................................................................................. 244

17. Troubleshooting ..................................................................................................................................... 24517.1 OSICS Mainframe Error and Warning Messages............................................................................................... 24517.2 OSICS Modules Error and Warning Messages ................................................................................................... 246

Certification and Compliance ....................................................................................................................... 247

List of Figures ................................................................................................................................................. 249

Page 13: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 13

Safety Information

Safety Labels and Symbols on the Product

Laser Safety The OSICS multifunction platform is identified as a Class 1 M product and is equipped with a remote Interlock connector on which a remote on/off switch can be connected (see section Installing the Remote Interlock, p. 27).• The OSICS T100, ECL, DFB, SLD, TLS-AG and TLS-50 modules are classified as

Class 1M products (power values are given in sub-section Technical Specifications in the appropriate module section of the OSICS User Manual).

Your safety may be compromised if you do not use the OSICS multifunction platform in accordance with the instructions given in this manual and on the labels located on the product.

WARNING

Safety Label DescriptionClass 1M Laser ProductThe product may have a built-in light source (depending on the OSICS modules installed), which makes it a class 1M laser product, as defined by the standard on safety of laser products (see section Certification and Compliance, p. 247). This means that it is safe under all conditions of normal use except when viewed with optical instruments such as microscopes and telescopes. Class 1M lasers produce large-diameter beams, or beams that are divergent.

Laser Radiation LabelThis label indicates the location of the laser output.Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.Caution while Handling Label(OSICS ECL and T100 modules only). This label is affixed on the module side panel, as pressing the sides could result in potential damage to the product.

Electrical Safety LabelThis label indicates an injury hazard. It appears on a location that requires special instructions for proper use: you must consult the documentation before using the element of the product where this symbol is marked (see section Connecting the OSICS Mainframe to a Power Source, p. 22).

Class 1M Laser ProductInvisible Laser RadiationDo not view directly with

optical instrumentsIEC 60825-1:2007

Complies with FDAperformance standards forlaser products except for

deviations pursuant to LaserNotice N°50 (June 24, 2007)

CAUTION

DO NOT PRESSFRAGILE OPTICALPARTS INSIDE

FUSE 4.0A, 250V, SLOW(T)

TO AVOID ELECTRIC SHOCK, THEPROTECTIVE GROUNDING CONDUCTORMUST BE CONNECTED TO GROUND

100-240V ±10% CATII ~50-60Hz POWER MAX 220W

!

Page 14: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Safety Information

14 OSICS User Manual

• The OSICS ATN, BKR and SWT modules may be compatible with powerful lasers (output power values are given in sub-section Technical Specifications in the appropriate module section of the OSICS User Manual).

Ventilation

Electrical Safety

• Disable the laser output before connecting or disconnecting a fiber optic cable on the module (see section Enabling/Disabling the Light Source of All Modules, p. 37).

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

• To ensure proper environment conditions:• Make sure the location where the OSICS Mainframe will be

installed meets the environmental characteristics listed in section OSICS Mainframe Technical Specifications, p. 16.

• Do not install the OSICS Mainframe near any source of heat or cold.

• To ensure proper ventilation and cooling, make sure there is sufficient clearance below and at the back of the OSICS Mainframe in the place where it will be installed.

• If you plan to rack-mount the product (depends on version), make sure to allow at least a 15 mm (0.6 inches) gap between two products.

CAUTION

Power Supply Safety Requirements

• Make sure the product power source does not apply more than 265 Volts RMS between the supply conductors or between either of the supply conductors and the ground.

• To avoid the possibility of injury, make sure the socket outlet in which the power supply cable will be plugged is equipped with a protective ground contact, and that the electrical installation fulfills the local safety requirements.

• The OSICS Mainframe has a chassis connected to ground via the power supply cable. A protective ground connection by way of the grounding conductor in the power cable is essential for safe operation.

• The OSICS Mainframe is not designed for outdoor use. To avoid the possibility of injury, do not expose the product to rain or excessive moisture. Do not operate the product in the presence of flammable gases or fumes.

• To avoid personal injury, never remove the protective cover of the chassis to perform servicing or maintenance operations. You must refer to your Yenista Optics service representative.

• To avoid fire hazard, only use the correct fuse type, voltage and current ratings (see section OSICS Mainframe Technical Specifications, p. 16).

WARNING

WARNING

Page 15: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 15

1. Product Presentation

1.1 Product Features

The OSICS multifunction platform is an 8-channel modular instrument designed for fiber-optic component and system testing, particularly for DWDM system evaluations.It is composed of the following elements:• The OSICS Mainframe

The OSICS Mainframe contains eight multipurpose slots that can host up to eight pluggable modules such as tunable lasers, transmission lasers and passive optical functions and sources.

The OSICS Mainframe provides built-in front panel controls and display to drive both system common controls and module parameters. All system operations are software-controlled from the OSICS Mainframe front panel.

Any OSICS module can be inserted in any multipurpose module-slot and in any combination. The OSICS Mainframe automatically detects OSICS modules after plug-in.

The OSICS Mainframe is designed either for benchtop use (with or without the front legs deployed) or rack-mount using two rack handles (optional).

You can remotely operate the OSICS system via the available communication interfaces.

The OSICS Mainframe can automatically save the working configuration of all eight modules simultaneously. This allows restoring the operating parameters of the OSICS Mainframe and modules instantly as you left them at the last system turn-off.

• The OSICS pluggable modulesAll OSICS modules are of the same physical dimension and size. OSICS modules are plug-and-play units that require no additional hardware or software.

Page 16: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Product Presentation

16 OSICS User Manual

1.2 OSICS Mainframe Technical Specifications

The following table gives an overview of technical specifications of the OSICS Mainframe. Detailed specifications of all OSICS modules are detailed in OSICS User Manual.

General CharacteristicsNumber of slots 8Supported pluggable modules All OSICS modules in any combinationElectrical SpecificationsFuse type 4.0 A 250 V SLOW/(T)Power Supply 100-240 V AC ±10 %Power Supply Frequency 50-60 Hz ±5 %Maximum Power Consumption 120 WInterfacesControl Panel Screen, rotary knob, on/off and control buttonsRemote Control • GPIB connector for IEEE 488 communication

• USB-B connector for RS-232C communication

• SubD-9 for remote interlockAnalog Monitoring Multi-purpose BNC output for current, power or

temperature monitoring of each pluggable module

Modulation • Input:• External TTL modulation (Mod. In BNC

connector)• Internal TTL modulation (on-board

oscillator)• Output: Synchro BNC connector

Environmental SpecificationsEquipment Type Test and measurementEquipment Location Indoor use onlyOperating Temperature Range +15 °C to 35 °C (+59 °F to +95 °F) ambient

temperatureWarm up Time (at room temperature) 1 hour (typ.)Humidity 80 % to 50 % relative humidity

(with maximum operating temperature ranging from 31 °C to 40 °C)

Maximum Operating Altitude 2000 metersPhysical SpecificationsDimensions (W x H x D) 448 x 145 x 387 mmWeight 7.2 kg

Page 17: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Product Presentation

OSICS User Manual 17

1.3 Product Overview• The OSICS Mainframe is delivered with the following accessories:

• A power supply cable

• A USB key containing the OSICS Mainframe USB driver and user documentation

• OSICS modules are delivered with a certificate of calibration.

1.3.1 OSICS Front Panel

The front panel of the OSICS Mainframe is composed of:• The control panel, with display and control buttons• The modules area, with eight slots for pluggable OSICS modules.

Figure 1: Mainframe – Front Panel

Control Panel• Screen

The screen displays the operating parameters and readings (updated in real-time) of all plugged modules.

• On/Off ButtonThe On/Off button turns on/off the OSICS Mainframe.The button lights green if the OSICS Mainframe is turned on.

• Rotary knobThe rotary knob enables you to navigate through the menus and to modify the value of selected parameters: • Turning the knob enables you to navigate within menus or to modify the selected

parameter values.• Pressing the knob enables you to select a menu or to validate a selection or a

value.

• Left/Right Control ButtonsThe control buttons allow you to command the functions displayed on screen right above them.

Modules areaAny OSICS module can be placed in any of the eight multipurpose slots. For more details, see section Handling OSICS Modules Into the OSICS Mainframe, p. 23.

Multipurpose Slots (x8)ControlButtons

Rotary knob

Screen

On/OffButton Slot 1 Slot 2 Slot 3 Slot 4 Slot 5 Slot 6 Slot 7 Slot 8

Page 18: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Product Presentation

18 OSICS User Manual

1.3.2 OSICS Rear Panel

Figure 2: Mainframe – Rear Panel

MAINS

The MAINS part includes the following elements:• Power Switch

• Power Cable ConnectorThe OSICS Mainframe unit is equipped with a self-regulating power supply (for details, see section OSICS Mainframe Technical Specifications, p. 16).

• Fuse DrawerThe fuse holder contains a fuse to protect the OSICS Mainframe from overcurrent (for fuse type, see section OSICS Mainframe Technical Specifications, p. 16). For details on how to replace the fuse, see section Replacing the Line Fuse, p. 244.

Remote Interfaces

• USB-B (RS-232C)This port enables you to perform RS-232C remote control operations. For more information, see OSICS Programming Guide.

• IEEE 488 (GPIB)This port enables you to perform IEEE 488.2 remote control operations. For more information, see OSICS Programming Guide.

Inputs

• Mod. InYou can use this BNC connector as input for external TTL-modulated signals.The green LED located to the right side of the connector is lit when the Mod. In connector receives an external TTL-modulated signal.

The OSICS Mainframe can provide an internal TTL-modulated signal to the OSICS T100, OSICS ECL and OSICS DFB modules:

Power SwitchFuse Drawer

MAINS

LabelsCooling Fan

Remote interlock Input(SubD-9 connector)

GPIB port

USB-B portPower Cable

Connector

Synchro (output)Mod. in (input) OUT 2

connectors

OUT 1connectors

Page 19: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Product Presentation

OSICS User Manual 19

• OSICS T100: see section Configuring the OSICS T100 Signal Modulation, p. 184.• OSICS ECL: see section Configuring the OSICS ECL Signal Modulation, p. 110.• OSICS DFB: see section Configuring the OSICS DFB Signal Modulation, p. 74.

• Remote Interlock (SubD-9 connector)On/off switch for laser safety (see section Installing the Remote Interlock, p. 27).

Outputs

• SynchroThis BNC connector provides a modulated-signal enabling the OSICS Mainframe to synchronize measurements on external devices or instruments (see section Using the Mainframe as a Modulation Source for External Devices (Synchro Output), p. 44).

• OUT 1• For T100 and ECL pluggable modules, this BNC analog-output connector enables

channel monitoring of power and current:• OSICS T100: see section Monitoring the OSICS T100 Module Parameters Using

the Analog OUT 1 Output, p. 197.• OSICS ECL: see section Monitoring the OSICS ECL Module Parameters Using

the Analog OUT 1 Output, p. 123.

• For DFB modules, this BNC analog-output connector enables channel monitoring of power, current and temperature: see section Monitoring the OSICS DFB Module Parameters Using the Analog OUT 1 Output, p. 91.

• For SWT-APC modules, this BNC analog-output connector enables channel monitoring of the power: see section Monitoring the Optical Output Power of the SWT APC Module, p. 169.

• OUT 2This BNC analog-output connector enables you to use the T100, ECL or DFB modules as an external modulation source for external devices:• OSICS T100: see section Using the T100 Module as a Modulation Source for

External Devices (OUT 2 Output), p. 196.• OSICS ECL: see section Using the ECL Module as a Modulation Source for External

Devices (OUT 2 Output), p. 122.• OSICS DFB: see section Using the DFB Module as a Modulation Source for External

Devices (OUT 2 Output), p. 90.

Cooling Fan

The cooling fan extracts warm air from inside. A cover grid protects it.

1.3.3 OSICS Information and Warning Markings

Label Description

Identification of the Product

Indicates serial number, model, options (if any), CE marking and date of manufacture.

MANUFACTURED

OPTIONS

MODEL

SERIAL NUMBER

Made in Francexxxxxxxx xxxx

OSICS

xxxxxxxxxxxx

xxxxxxxxxxxx

Page 20: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Product Presentation

20 OSICS User Manual

Manufacturer Identification

Contact information of the manufacturer.

Class 1M Laser Product

The product may have a built-in light source (depending on the OSICS modules installed), which makes it a class 1M laser product, as defined by the standard on safety of laser products (see section Certification and Compliance, p. 247). This means that it is safe under all conditions of normal use except when viewed with optical instruments such as microscopes and telescopes. Class 1M lasers produce large-diameter beams, or beams that are divergent.Laser Radiation LabelAffixed on OSICS modules, this label indicates the location of the laser source. Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.Caution while Handling Label(OSICS ECL and T100 modules only). This label is affixed on the module side panel, as pressing the sides could result in potential damage to the product.

Electrical Safety Label

This label indicates an injury hazard. It appears on a location that requires special instructions for proper use: you must consult the documentation before using the element of the product where this symbol is marked (see section Connecting the OSICS Mainframe to a Power Source, p. 22).WEEE symbol for recycling

See section Certification and Compliance, p. 247.

EFUP Label (RoHS China)

The product has an environment friendly use period of 40 years. See section Certification and Compliance, p. 247.Warranty Seal

The product cover must not be open, otherwise the warranty is not valid anymore.

Label Description

Manufactured by:

4, rue Louis de Broglie www.yenista.com22300 LANNION Tel : +33 2 96 48 37 15France Fax : +33 2 96 48 73 04

Class 1M Laser ProductInvisible Laser RadiationDo not view directly with

optical instrumentsIEC 60825-1:2007

Complies with FDAperformance standards forlaser products except for

deviations pursuant to LaserNotice N°50 (June 24, 2007)

CAUTION

DO NOT PRESSFRAGILE OPTICALPARTS INSIDE

FUSE 4.0A, 250V, SLOW(T)

TO AVOID ELECTRIC SHOCK, THEPROTECTIVE GROUNDING CONDUCTORMUST BE CONNECTED TO GROUND

100-240V ±10% CATII ~50-60Hz POWER MAX 220W

!

Warranty voidif seal broken.

Page 21: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 21

2. Installing and Connecting the OSICSMultifunction Platform

2.1 Unpacking and Installing the OSICS Mainframe

Subject When used as a stand-alone bench-top instrument, the OSICS Mainframe is designed so that the feet leave enough room under the product to enable proper ventilation. The apertures (under the product) and the fan (behind the product) are used for the inside ventilation.

Before Starting

Procedure 1. Open the package with care and remove the protective foam.

2. Pull out the OSICS Mainframe vertically from its packaging: hold it by its two retractable handles located on both sides of the product and keep it horizontal.

3. Set the OSICS Mainframe on a flat stable surface free of excessive vibration.

4. Allow the flow of air to circulate freely around the OSICS Mainframe and remove any equipment or object that could block the air flow. Ventilation holes are located on the bottom and back sides of the OSICS Mainframe.

5. On the rear panel (see Figure 2, p. 18), make sure the power switch is set to O.

6. To tilt the OSICS Mainframe upward, deploy the two retractable legs located below it.

Figure 3: Retractable Legs

• To ensure proper environment conditions:• Make sure the location where the OSICS Mainframe will be

installed meets the environmental characteristics listed in section OSICS Mainframe Technical Specifications, p. 16.

• Do not install the OSICS Mainframe near any source of heat or cold.

• To ensure proper ventilation and cooling, make sure there is sufficient clearance below and at the back of the OSICS Mainframe in the place where it will be installed.

• If you plan to rack-mount the product (depends on version), make sure to allow at least a 15 mm (0.6 inches) gap between two products.

CAUTION

IMPORTANT When unpacking, handle the device with care and do not damage the original shipping container in case the OSICS Mainframe needs to be returned to Yenista Optics.

Page 22: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Installing and Connecting the OSICS Multifunction Platform

22 OSICS User Manual

2.2 Connecting the OSICS Mainframe to a Power Source

Subject The OSICS Mainframe is equipped with a self-regulating power supply. For voltage specifications, see section OSICS Mainframe Technical Specifications, p. 16.

Before Starting

Procedure 1. Make sure the power switch is set to O.

2. Connect one end of the provided power supply cable to the mains socket located on the rear panel and plug the other end to the proper voltage wall socket outlet (to know the voltage requirement, see section OSICS Mainframe Technical Specifications, p. 16).

• Make sure the product power source does not apply more than 265 Volts RMS between the supply conductors or between either of the supply conductors and the ground.

• To avoid the possibility of injury, make sure the socket outlet in which the power supply cable will be plugged is equipped with a protective ground contact, and that the electrical installation fulfills the local safety requirements.

• The OSICS Mainframe has a chassis connected to ground via the power supply cable. A protective ground connection by way of the grounding conductor in the power cable is essential for safe operation.

• The OSICS Mainframe is not designed for outdoor use. To avoid the possibility of injury, do not expose the product to rain or excessive moisture. Do not operate the product in the presence of flammable gases or fumes.

WARNING

Page 23: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Installing and Connecting the OSICS Multifunction Platform

OSICS User Manual 23

2.3 Handling OSICS Modules Into the OSICS Mainframe

OSICS modules can be installed in the OSICS Mainframe in any of the available 8 channel-slots and in any combination alongside other types of modules supported by the OSICS platform.

2.3.1 Installing an OSICS Module Into the OSICS Mainframe

Subject This section explains how to install an OSICS module into a empty slot of the OSICS Mainframe. All OSICS modules can be installed in any of the available 8-channel slots of the OSICS Mainframe and in any combination alongside other types of OSICS modules.

Before Starting • Open the package with care and remove the OSICS module from its packaging.

• Make sure you have a medium-small size flat blade screwdriver.

Procedure 1. Turn the OSICS Mainframe off (see section Turning Off the OSICS Multifunction Platform, p. 36) and unplug the power supply cable.

2. Unpack the OSICS module from its wrapping.

3. On the OSICS Mainframe front panel, remove the cover plate of the slot where you want to install the module: loosen the captive screw using a medium-small flat blade screwdriver and remove the plate.

Figure 4: Module Installation – Plate Removal

IMPORTANT When unpacking, handle the device with care and do not damage the original shipping container in case the OSICS module needs to be returned to Yenista Optics.

Do not perform any hardware installation or maintenance operation while the system is in operation.

CAUTION

Page 24: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Installing and Connecting the OSICS Multifunction Platform

24 OSICS User Manual

4. Hold the module with both hands by its top and bottom edges in front of the open slot to align the top and bottom edges of the module with the slot guides on the OSICS chassis and insert the module into the open slot.

5. Put your thumb on the captive screw notch of the module to push it in so that it slides in freely.

6. Apply pressure to seat the module in the OSICS Mainframe backplane connector.

7. Tighten the captive screw of the module to secure it in place.

Figure 5: Module Installation – Module Insertion

To avoid damaging the internal optical components, do not squeeze the module sideplates.

CAUTION

Page 25: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Installing and Connecting the OSICS Multifunction Platform

OSICS User Manual 25

2.3.2 Removing an OSICS Module From the OSICS Mainframe

Subject This section explains how to remove an OSICS module from the OSICS Mainframe.

Before Starting Make sure you have a medium-small size flat blade screwdriver.

Procedure 1. Turn the OSICS Mainframe off (see section Turning Off the OSICS Multifunction Platform, p. 36) and unplug the power supply cable.

2. On the OSICS Mainframe front panel, loosen the captive screw of the module you want to remove using a medium-small flat blade screwdriver.

3. Put your index and middle finger inside the groove located behind the captive screw notch of the module, and pull the module out as illustrated in the following figure:

Figure 6: Module Installation – Module Removal

4. Hold the module with both hands by its top and bottom edges and put it in its original container.

5. On the OSICS Mainframe front panel, put a cover plate on the empty slot and tighten the captive screw to secure it in place.

Do not perform any hardware installation or maintenance operation while the system is in operation.

Do not extract the module by pulling on any of the front-panel protruding elements, such as the laser protection cap or connectors because this may loosen or damage those elements permanently.

Do not operate the OSICS Mainframe if a channel-slot is left open. Always put back the protection cover plate on an empty slot.

CAUTION

CAUTION

CAUTION

Page 26: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Installing and Connecting the OSICS Multifunction Platform

26 OSICS User Manual

2.4 Connecting Optical Devices to OSICS Modules

Subject You can connect external optical components and instruments to the OSICS modules.

Before Starting Make sure you have the appropriate fiber type corresponding to the modules you want to connect. For details on the appropriate optical fiber type, see the Technical Specifications section corresponding to the modules you want to connect in the appropriate module section of this manual)

Procedure 1. Disable the laser output before connecting or disconnecting a fiber optic cable on the module (see section Enabling/Disabling the Light Source of All Modules, p. 37).

2. Make sure fiber-optic connectors are clean.

3. On the front panel, lift the protective cap and connect the optical component(s) to the OSICS module via the FC/APC fiber-optic connector.

4. In case of frequent connect/disconnect, use an intermediate jumper to prevent premature failure of the OSICS module internal connector.

• Keeping fiber-optic connectors clean at all times is essential to achieve optimum system performance (see section Cleaning Optical Connectors, p. 243).

• The standard output connector is a FC/APC connector. Never connect another type of connector to the optical output.

CAUTION

The use of optical instruments with this product increases eye hazard. Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

On OSICS SWT, do not connect several laser sources at the same time on the Com and Chn channels, or you may severely damage your optical equipment.

CAUTION

Page 27: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Installing and Connecting the OSICS Multifunction Platform

OSICS User Manual 27

2.5 Installing the Remote Interlock

Subject The OSICS Mainframe is equipped with a Remote Interlock connector at the rear panel.You can connect an external remote interlock switch to the Remote Interlock SubD-9 connector located on the rear panel.Remote interlock sets the system current and power to 0 mA and 0 mW, respectively.

Before Starting Make sure that the external switch is electrically isolated from other circuits, including earth ground.

Procedure • To connect the remote interlock switch, use a connector that meets the following pin assignment:

• Pins 1-2: shorted• Pins 6-7: connect to external switch

Figure 7: Remote Interlock pin assignment

• If the remote interlock switch is open, the laser output of all modules installed on the OSICS Mainframe is switched off.

In the System Status screen (see System Status Screen Description section, p 30), the message INTERLOCK is displayed for all installed modules to indicate that the OSICS laser output is switched off.

• If the remote interlock switch is closed, the OSICS system laser output is switched back on.

External switch

Page 28: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Installing and Connecting the OSICS Multifunction Platform

28 OSICS User Manual

Page 29: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 29

3. Turning On/Off and Initializing the OSICSMultifunction Platform

3.1 Turning on the OSICS Multifunction Platform and Accessing the System Status Screen

Subject Turning on the OSICS multifunction platform also turns on all installed modules.After the OSICS multifunction platform is turned on, the System Status screen is automatically displayed.

Before Starting Make sure the OSICS Mainframe is properly installed and connected (see section Installing and Connecting the OSICS Multifunction Platform, p. 21).

Procedure 1. On the rear panel, set the MAINS power switch to I (see Figure 2, p. 18).

2. On the front panel, press the on/off button (see Figure 1, p. 17).

• The OSICS Mainframe performs the initialization routine, the message Initializing appears.

• The OSICS Mainframe prompts you to enter a password.

Figure 8: Password Screen

3. Enter your password as follows (default password is 0000):

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

To modify the password, see section Modifying the OSICS Mainframe Password, p. 35.

If you forget your password, contact the Yenista Optics customer support service (see section Contact Information, p. 4).

Do not turn the OSICS Mainframe off before the system is fully initialized.

CAUTION

ENTER

Password :

0000

Press the knob to change the digit.

Page 30: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Turning On/Off and Initializing the OSICS Multifunction Platform

30 OSICS User Manual

4. Select ENTER by pressing the right control button.

After password identification, each module is initialized, turned on (the green LED is lit) and set to the power-up wavelength.At the end of the initialization routine, the STARTUP configuration is loaded on the OSICS Mainframe and on each installed module (for more details on working configurations, see section Setting a Working Configuration, p. 45).

The System Status screen appears.

System Status Screen Description

The System Status screen provides a general overview of the state, configuration and operating parameters of the system.

Figure 9: System Status Screen

Menu bar

The menu bar is displayed at the upper part of all screens. It includes the following tabs:• The Main menu tab enables you configure the OSICS Mainframe (see section

Accessing the OSICS Mainframe Main Setup Menu, p. 33)

• One tab for each OSICS module, from left to right (1 to 8), to configure each module individually (see section Accessing an OSICS Module Setup Menu, p. 33).

To select a tab, turn the rotary knob to highlight the wanted tab and press the knob to enter the tab menu.

Mode area

The mode area displays the following read-only information:• Current system operating-mode (for more details, see OSICS Programming Guide).

• Working configuration currently active (for more details, see section Setting a Working Configuration, p. 45).

Parameters area

The Parameters area is a read-only table that displays the key operating-parameters of the installed modules, as described in the following table.

Main 1 2 3 4 5 6 7 8System Status

Mode : LOCAL Config. : STARTUP

CH Type F (GHz) P (dBm) Info

1 ATN 193415 55.00 dB2 BKR 230610 02.04 dB

3 TLS 193700.0 +13.01 0454 ECL 202562.5 -00.03 A

5 SLD 199861.7 +07.006 ECL 187370.3 -00.01 A7 SWT Shut8 SWT Ch1

Menu bar

Mode area

Parametersarea

Page 31: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Turning On/Off and Initializing the OSICS Multifunction Platform

OSICS User Manual 31

Column Name

Description of Value Displayed

CH Slot number, corresponding to the modules installed in the OSICS Mainframe.

Type Type of module installed in the corresponding slot.If the slot is empty, the field is left blank.

λ (nm)/f (GHz)

Light emission wavelength or frequency of the module, depending on the selected wavelength unit (see section Selecting the Spectral Unit for all Installed Modules, p. 39).

P (dBm)/P (mW)

Output power information:• <value>: output power value of the module in dBm or mW

depending on the selected power unit (see section Selecting the Power Unit for All Installed Modules, p. 38).

• Disabled: the laser output of the corresponding module is disabled.If no specific configuration has been defined (see section Setting a Working Configuration, p. 45), each OSICS module is in the disable (idle) state after system initialization.

• INTERLOCK: the laser safety Interlock switch is open. In this case, all modules are disabled (for more details, see section Installing the Remote Interlock, p. 27).

Info Additional information on module operating status, as explained in the following lines.

A T100 and ECL modules only. The Auto peak find function is enabled (see sub-section Activating the Auto-peak Find Function in the appropriate module section of this manual).

B TLS modules only. The Fine Tuning and SBS functions are enabled.BO2SBS2OBS2SBO2O

SWT 2x(1x1) modules only. Selected output mode (see section Selecting the OSICS SWT Input/Output, p. 142):• BS2S: both shut.• BO2O: both open.• BO2S: Ch A-B open, Ch 1-2 shut.• BS2O: Ch A-B shut, Ch 1-2 open.

Bar/Cross

SWT 2x2 module only. Selected output mode (see section Selecting the OSICS SWT Input/Output, p. 142).

C T100 and ECL modules only. The Coherence control function is enabled (see section Activating Coherence Control in the appropriate module section of this manual).

CHn SWT module only. Selected channel number (see section Selecting the OSICS SWT Input/Output, p. 142).

D TLS modules only. The Frequency Control Dither function is disabled (see section Disabling the Frequency Control Dither Function, p. 238).

F TLS-AG module only. The Fine Tuning function is enabled (see sub-section Fine Tuning the Wavelength/Frequency, p. 235).

LIM The selected power value cannot be set due to a maximum current-limit reached within the laser diode.

Page 32: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Turning On/Off and Initializing the OSICS Multifunction Platform

32 OSICS User Manual

M T100, ECL and DFB modules only. The modulation option is set to ON or ON INVERTED (see sub-section Configuring the Signal Modulation in the appropriate module section of this manual).

o ATN modules only. An offset is set (see section Setting the ATN Attenuation Offset, p. 55).

S TLS modules only. The SBS Suppression function is enabled (see sub-section Setting the SBS Suppression Mode in the appropriate module section of this manual).

s SWT module only. A sequence is running (see section Configuring Sequential Switching, p. 143).

Shut SWT 1x1 modules only. Selected output mode (see section Selecting the OSICS SWT Input/Output, p. 142).

<value> TLS modules only. • On TLS-50 and TLS-AG used in Grid mode: channel value.• On TLS-AG used in Manual or High Resolution mode: current position

value relative to the lowest frequency.

Column Name

Description of Value Displayed

Page 33: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Turning On/Off and Initializing the OSICS Multifunction Platform

OSICS User Manual 33

3.2 Accessing the OSICS Setup MenusSubject Each element of the OSICS platform can be configured individually:

• The OSICS Mainframe setup menu (Main Setup menu) enables you to configure all the operating parameters of the OSICS Mainframe and some system parameters that affect all installed modules.

• The OSICS module setup menu (<Module Name> Setup menu) enables you to set the operating parameters of a selected OSICS module.

Procedures Accessing the OSICS Mainframe Main Setup Menu

1. From the System Status screen, turn the rotary knob to highlight the Main tab.

2. Press the knob to enter the highlighted menu.

The Main Setup menu appears.

Accessing an OSICS Module Setup Menu

1. From the System Status screen, turn the rotary knob to highlight the tab corresponding to the slot number in which the wanted module is installed (see Figure 1, p. 17).

2. Press the knob to enter the highlighted menu.

The <Module Name> Setup menu appears.

Setup Menu Description

Figure 10: Setup Menu – Examples

Menus/Sub-menus and Symbols• : indicates the location of the cursor operated using the rotary knob.

To move the cursor, turn the rotary knob.• : indicates a menu without sub-menus.

To enter this type of menu: turn the rotary knob to put the cursor before the menu and press the rotary knob.

• : indicates a menu with one or more sub-menus.

: indicates that the menu is expanded.

• To expand a menu: turn the rotary knob to put the cursor before the menu and press the rotary knob.

• To enter a sub-menu: turn the rotary knob to put the cursor before the sub-menu and press the rotary knob to enter into it.

Main

EXIT

1 2 3 4 5 6 7 8

ECL Unit Setup

ESCAPE

Set Power

Step

Calibration

Coherence

Analog OutInformation

Adjust PowerSet λ/f

Modulation

Auto peak find

Main

EXIT

1 2 3 4 5 6 7 8Main Setup

ESCAPE

Optical Output : DISABLEDPower

Set unit dBm/mWSet power reference

Modulation

SourceSet frequency

Set unit GHz/nm

GPIB address : 10Configuration : STARTUP

LoadSave

PasswordInformation

OSICS Mainframe Setup Menu OSICS ECL Unit Setup Menu

Page 34: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Turning On/Off and Initializing the OSICS Multifunction Platform

34 OSICS User Manual

ESCAPEReturns to the previous displayed menu. Press the left control button (see Figure 1, p. 17) to operate this command.

EXITExits the current menu and displays the System Status screen. Press the right control button (see Figure 1, p. 17) to operate this command.You can only browse tabs from the System Status screen.

3.3 Checking the OSICS Version

Subject The Information menu displays information about the version of your OSICS Mainframe or OSICS modules.

Procedure Checking the Firmware Version of the OSICS Mainframe

1. Access the Main Setup menu (see section Accessing the OSICS Setup Menus, p. 33).

2. Turn the rotary knob to put the cursor before the Information menu and press the knob to enter it.

The information screen appears with the OSICS Mainframe FPGA and firmware revision number.

Checking the Version Information of an OSICS Module

1. Access the <Module Name> Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Information menu and press the knob to enter it.

The information screen appears with the OSICS module information.

Page 35: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Turning On/Off and Initializing the OSICS Multifunction Platform

OSICS User Manual 35

3.4 Modifying the OSICS Mainframe Password

Subject For laser safety reasons, a four digit password controls the access to the OSICS Mainframe.The default password is 0000. This section explains how to modify it.

Procedure 1. Access the Main Setup menu (see section Accessing the OSICS Setup Menus, p. 33).

2. Turn the rotary knob to put the cursor before the Password menu and press the knob to enter it.

The Password sub-menu appears and prompts you to enter the actual password.

3. Enter the current password as follows:

a. Turn the rotary knob to put the cursor under the digit to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate.

The screen prompts you to enter a new password.

Figure 11: Main Setup – New Password

4. Enter the new password as follows:

a. Turn the rotary knob to put the cursor under the digit to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Repeat steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate.

A confirmation message appears.

5. Press the knob to confirm the password change.

Main 1 2 3 4 5 6 7 8Main Setup - Password

EXITESCAPE

Password :

1234

Enter a new password.

ENTER

Page 36: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Turning On/Off and Initializing the OSICS Multifunction Platform

36 OSICS User Manual

3.5 Turning Off the OSICS Multifunction Platform

Before Starting Disable the light source of all modules (see section Enabling/Disabling the Light Source of All Modules, p. 37)

Procedure 1. On the front panel, press the on/off button (see Figure 1, p. 17).

The message Saving Configuration appears on screen. It indicates that the current system configuration is being saved in the STARTUP memory (the STARTUP configuration will be loaded at the next system start)

2. Wait until the green LED of the on/off button lights off.

3. On the rear panel, set the power switch to O.

4. If you want to turn on the OSICS Mainframe again, wait at least 30 seconds.

Page 37: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 37

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN4. Controlling the Optical Output of all OSICSModules Simultaneously

From the OSICS Mainframe Main Setup menu, you can control optical output parameters of all installed modules simultaneously.

4.1 Enabling/Disabling the Light Source of All ModulesSubject This procedure only applies to modules with an optical source.

You can enable/disable the optical output in two different ways:• From the wanted module front panel, you can individually enable or disable the light

source of the module by using the Enable button. For more details on this procedure, see sub-section Enabling/Disabling the Light Source of the Module in the appropriate module section of this manual.

• From the OSICS Mainframe Main Setup menu, you can simultaneously enable or disable the light source in all the modules installed in the OSICS Mainframe, as detailed in the following procedure. This function overrides the enable/disable state individually set in all modules.

Before Starting Connect an optical device to the OSICS modules (see section Connecting Optical Devices to OSICS Modules, p. 26).

Procedure Enabling the Light Source of All Installed Modules

1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Optical Output : DISABLED menu.

3. Press the rotary knob to enable the optical output of all installed modules.

The OSICS Mainframe beeps and the light source is enabled on all modules simultaneously (the green Enable LED of all modules lights on).

The menu displays Optical Output : ENABLED.

Disabling the Light Source of All Installed Modules

1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Optical Output : ENABLED menu.

3. Press the rotary knob to disable the light source of all installed modules.The optical output is disabled on all modules simultaneously (the green Enable LED of all modules lights off).

• Do not enable the laser while there is no fiber connected to the optical output connector.

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 38: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Controlling the Optical Output of all OSICS Modules Simultaneously

38 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

4.2 Configuring the Optical Output Power of All Modules

4.2.1 Selecting the Power Unit for All Installed Modules

Subject The following procedure explains how to modify the power unit (dBm or mW) for all modules installed in the OSICS Mainframe. You cannot set the power unit of an individual module.

Procedure 1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Expand the Power menu and enter the Set unit dBm/mW sub-menu.

The Set unit dBm/mW sub-menu appears and displays the current unit setting.

Figure 12: Main Setup – Set unit dBm/mW sub-menu

3. Turn the rotary knob to put the cursor before the wanted unit and press it to validate your selection.

The Main Setup menu is automatically displayed and the optical power value of each module is now set to the selected unit.

4.2.2 Setting the Optical Output Power of All Installed Modules

Subject You can set the optical output power in two different ways:• From the wanted module menu, you can set the module output power individually.

For more details on this procedure, see section Setting the Optical Output Power of the Module in the appropriate module section of this manual.

• From the OSICS Mainframe Main Setup menu, you can set all modules output power to the same value simultaneously, as detailed in the following procedure.

As modules of different types and different power specifications may be installed in the system, the power setting may exceed one of the modules maximum power-rating. To prevent any possible power-overshoots, the power of each module is software-limited within the acceptable range for that particular module.

Procedure 1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Expand the Power menu and enter the Set power reference sub-menu.

The Power reference sub-menu appears and displays the current power setting.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Set unit dBm/mW

ESCAPE

Power Unit :

dBm

mW

Select a new unit.

Page 39: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Controlling the Optical Output of all OSICS Modules Simultaneously

OSICS User Manual 39

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 13: Main Setup – Power reference sub-menu

3. Modify the power value as follows:

a. Turn the rotary knob to put the cursor under the digit or +/- sign you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new power value.

The Main Setup menu is automatically displayed and the optical power value of each module is now updated (according to the individual modules power limits).

4.3 Selecting the Spectral Unit for all Installed ModulesSubject The following procedure explains how to modify the spectral unit (nm or GHz) for all

modules installed in the OSICS Mainframe. You cannot set the spectral unit of an individual module.

Procedure 1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set unit GHz/nm menu and press the knob.

The Set unit GHz/nm sub-menu appears and displays the current unit setting.

Figure 14: Main Setup – Set unit GHz/nm sub-menu

3. Turn the rotary knob to put the cursor before the wanted unit and press it to validate your selection.The Main Setup menu is automatically displayed and the wavelength/frequency value of each module is set to the selected unit.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Power reference

ESCAPE

Power reference :

P = +03.00 dBm ENTER

Modify the value.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Set unit GHz/nm

ESCAPE

Wavelength of Frequency Unit :

GHz

nm

Select a new unit.

Page 40: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Controlling the Optical Output of all OSICS Modules Simultaneously

40 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

4.4 Configuring the OSICS Mainframe Modulation

Presentation The OSICS Mainframe digital modulation feature includes the following capabilities:• External TTL-modulation by applying an external signal to the Mod. in BNC

connector located at the rear panel of the OSICS Mainframe (see Figure 2, p. 18).

• Internal TTL-modulation using the built-in oscillator generating signals from 123 Hz to 1 MHz.

For details on signal levels, see section OSICS Mainframe Technical Specifications, p. 16.The OSICS Mainframe works as a TTL-modulated signal generator for DFB, T100 and ECL modules and also for external devices, if connected to the Synchro output BNC connector.

Modulation Scheme

The modulated-signal generator of the OSICS Mainframe provides the same modulation for all the modules in a synchronous way, as illustrated in the following figure.

Figure 15: Mainframe Modulation Scheme

For more details on modulation of OSICS modules, see the following sections:• DFB module: section Configuring the OSICS DFB Signal Modulation, p. 74

• ECL module: section Configuring the OSICS ECL Signal Modulation, p. 110

• T100 module: section Configuring the OSICS T100 Signal Modulation, p. 184

Page 41: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Controlling the Optical Output of all OSICS Modules Simultaneously

OSICS User Manual 41

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

4.4.1 Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source

Subject This modulation setting uses the OSICS Mainframe’s built-in internal oscillator as a source of TTL-level modulated signals (see Figure 15, p. 40). The main purpose of this setting is to provide simultaneous and synchronous modulation to several modules installed in the OSICS Mainframe, or to an external device connected to the OSICS Mainframe Synchro output connector.The range of the frequency of the modulation provided by the OSICS Mainframe is defined to encompass the frequency requirements of all types of OSICS modules. The frequency settings for the lower and upper bounds may therefore lie outside of the modulation capabilities of a particular module, in which case the module's own frequency range sets the modulation specifications.The OSICS Mainframe signal generator can provide a TTL signal with a repetition frequency in the range 123 Hz–1 MHz, but for optimum modulation performance, we recommend applying modulation frequencies within the specified range of the installed modules.

Procedure Step 1 – Set the OSICS Mainframe modulation source to INTERNAL:

1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Enter the Source sub-menu as follows:

a. Turn the rotary knob to put the cursor before the Modulation menu and press the knob to expand it.

b. Turn the rotary knob to put the cursor before the Source sub-menu and press the knob to enter it.

The Modulation Source sub-menu appears and displays the current selected modulation source.

Figure 16: Main Setup – Modulation Source (Internal)

3. Turn the rotary knob to put the cursor before INTERNAL and press it to validate your selection.

The Main Setup menu is automatically displayed and the modulation is set to the selected source.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

EXTERNAL

Select a new value.

Page 42: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Controlling the Optical Output of all OSICS Modules Simultaneously

42 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Step 2 – Set the frequency of the OSICS Mainframe internal modulated-signal:

1. In the Main Setup menu, expand the Modulation menu and select the Set frequency sub-menu.

The Modulation frequency sub-menu appears and displays the current selected modulation frequency-value.

Figure 17: Main Setup – Modulation Frequency

2. Modify the frequency value as follows:

The internal TTL-signal generator can provide a TTL-modulated signal in the range 123 Hz to 1 MHz.

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new frequency value.

The frequency is updated to the "real frequency" as shown in Figure 18, p. 42. This screen displays the current modulation frequency produced by the OSICS Mainframe oscillator, which may slightly differ from the values you have set.

Figure 18: Main Setup – Modulation Real Frequency

3. Press the knob to go back to the Main Setup menu.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation frequency

ESCAPE

Frequency value :

f = 0014700 Hz

Modify the value.

ENTER

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation frequency

ESCAPE

The real frequency is :

14706 Hz

Please press the knob.

Page 43: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Controlling the Optical Output of all OSICS Modules Simultaneously

OSICS User Manual 43

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN4.4.2 Setting Up Digital (TTL) Modulation Using an External Source

Subject External digital-modulation is achieved by applying a digital signal to the BNC Mod. In connector located on the OSICS Mainframe rear panel (see Figure 15, p. 40). This allows all the modules (or an external device connected to the OSICS Mainframe Synchro output connector) to be modulated in a synchronous way by means of an external source. The OSICS Mainframe distributes the external modulated-signal to all T100, ECL and DFB installed modules in order to synchronize all modules with a single modulated signal, amplitude, and frequency.For optimum stability of the modulated signals, we recommend applying a digital modulation in the specified range of the modules.

Before Starting Make sure you have a digital-signal generator.

Procedure Step 1 – Set the OSICS Mainframe modulation source to EXTERNAL:

1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Enter the Source sub-menu as follows:

a. Turn the rotary knob to put the cursor before the Modulation menu and press the knob to expand it.

b. Turn the rotary knob to put the cursor before the Source sub-menu and press the knob to enter it.

The Modulation source sub-menu appears and displays the current selected modulation source.

Figure 19: Main Setup – Modulation Source (External)

3. Turn the rotary knob to put the cursor before EXTERNAL and press it to validate your selection.

The Main Setup menu is automatically displayed and the modulation is set to the selected source.

Step 2 – Set the TTL-level signal generator:

1. On the OSICS Mainframe rear panel, connect your external digital signal-generator to the Mod. in BNC connector (see Figure 2, p. 18).

2. Apply the TTL signal not exceeding the maximum voltage of +5 V and repetition frequency between and .

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

EXTERNAL

Select a new value.

Page 44: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Controlling the Optical Output of all OSICS Modules Simultaneously

44 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

The green LED located next to the Mod. In connector is lit to indicate that an external signal-generator is connected.

4.4.3 Using the Mainframe as a Modulation Source for External Devices (Synchro Output)

Subject On the OSICS Mainframe rear panel (see Figure 2, p. 18) the Synchro BNC output connector enables you to use the Mainframe as a modulation source for external devices.

Procedure 1. Set up the wanted digital modulation to the <Module Name> module using one of the following methods:

• OSICS Mainframe internal-modulation source, as explained in section Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source, p. 41.

• OSICS Mainframe external-modulation source, as explained in section Setting Up Digital (TTL) Modulation Using an External Source, p. 43.

2. Connect the appropriate device to the Synchro BNC connector located on the rear panel of the OSICS Mainframe.

The selected modulation frequency is applied to the connected device.

Page 45: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 45

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN5. Setting a Working Configuration

Definitions The working configuration is composed of the following parameters:• The OSICS Mainframe settings • The configuration settings of each installed OSICS module

You can save your working configuration to easily restore it at the next system startup.The OSICS Mainframe features three types of configuration:• DEFAULT configuration: the factory-set operating parameters on the OSICS

Mainframe and all installed OSICS modules. You cannot modify this configuration. Default settings are listed in the following Stored Configuration Parameters section, p 45.

• STARTUP configuration: the operating parameters (on the OSICS Mainframe and all installed modules) as you left them at the system turn-off. This configuration is automatically saved when the OSICS Mainframe is turned off and automatically loaded at system startup.The first time you turn on the OSICS Mainframe, the STARTUP configuration is identical to the DEFAULT configuration.Saved parameters are listed in the following Stored Configuration Parameters section, p 45.

• A, B, C and D configurations: the operating parameters on the OSICS Mainframe and all installed modules as you decided to save them. You can save four different sets of operating parameters. Parameters you can save are listed in the following Stored Configuration Parameters section, p 45.

Stored Configuration Parameters

STARTUP/A/B/C/DStored Settings

DEFAULT Configuration Settings

OSICS Mainframe Parameters

Optical outputThe STARTUP configuration never save the ENABLED state. For laser safety reasons, all modules are always in DISABLED state after system turn-on and initialization.

DISABLED

Power unit mWPower setting 0Spectral unit nm (wavelength)Modulation source InternalFrequency of the modulation source 200 HzGPIB address 10

Page 46: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Setting a Working Configuration

46 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

OSICS Module Parameters

Optical output DisabledOperating Mode(SWT and SWT-APC only)

• SWT APC: Switch• SWT 2x(1x1): BS2S• SWT 1x2, 1x4: Ch1• SWT 2x2: Bar• SWT 1x1: Shut

Power unit mWPower setting • DFB, ECL, T100: 0 mW

• TLS-AG, TLS-50: 10 mW• SWT, SWT APC: none• ATN: 55 dB (attenuation)• BKR: 55 dB (reflectance)• SLD: 5 mW

Spectral unit nm (Wavelength)Wavelength/Frequency setting • ECL, T100: center wavelength

• DFB: minimum wavelength• TLS-AG, TLS-50: CH1• SWT, SWT APC: none• ATN, BKR: 1550 (SMF)• SLD: maximum wavelength

Modulation source Internal (T100, ECL, DFB only)Modulation control(T100, ECL, DFB only)

OFF

Frequency of the modulation source (T100, ECL, DFB only)

200 Hz

Coherence Control function (T100, ECL only)

OFF

Auto peak find function(T100, ECL only)

ON

Step in menu 0.1 nmStep in frequency 50 GHzAnalog out (T100, ECL, DFB only) Power

STARTUP/A/B/C/DStored Settings

DEFAULT Configuration Settings

Page 47: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Setting a Working Configuration

OSICS User Manual 47

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN5.1 Saving a Working Configuration

Subject Four user-defined configurations are available to save your working configuration (A, B, C and D).If you save your configuration on an existing one, it is replaced by the new one.

Before Starting Set the parameters of the OSICS Mainframe and all installed modules parameters as you want them to be saved.

Procedure 1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Expand the Configuration menu and enter the Save sub-menu.

The Save configuration sub-menu appears and displays the configurations available for saving.

Figure 20: Main Setup – Save configuration

3. Turn the rotary knob to put the cursor before the wanted configuration name, and press it to validate your selection.

The Saving configuration... message appears.

When the configuration is saved, the System Status screen is displayed and the saved configuration is loaded.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Save configuration

ESCAPE

Save configuration to :

ABCD

Select a target memory.

Page 48: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Setting a Working Configuration

48 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

5.2 Loading a Working Configuration

Subject You can load one of the following configuration: DEFAULT, STARTUP, A, B, C or D.

Before Starting Make sure you have already saved a user-defined configuration (see section Saving a Working Configuration, p. 47).

Procedure 1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Expand the Configuration menu and enter the Load sub-menu.

The Load configuration sub-menu appears and displays the current configuration.

Figure 21: Main Setup – Save configuration

3. Turn the rotary knob to put the cursor before the wanted configuration name, and press it to validate your selection.

The Moving... message appears.

When the configuration is loaded, the System Status screen is displayed and the selected configuration is loaded.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Load configuration

ESCAPE

Select a configuration to load :

DEFAULTSTARTUPABCD

Page 49: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 49

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN6. Using the OSICS ATN-HP Module

This section explains how to operate the OSICS ATN-HP via the ATN-HP Unit Setup menu. For more details on how to install and initialize the module into the OSICS Mainframe, see section Installing and Connecting the OSICS Multifunction Platform, p. 21.

6.1 OSICS ATN-HP Presentation

The OSICS ATN-HP is an optical attenuator module that comes in a 1x1 configuration (one input / one output).

Figure 22: ATN-HP Configuration

6.1.1 OSICS ATN-HP Technical Specifications

Using the module in a reverse configuration could create higher return loss value and attenuation might not be corrected calibrated. We recommend to use the attenuator in a standard input/output configuration.

Input OutputOSICS ATN-HP

In Out

CAUTION

SMF PMF

Wavelength Range 1250–1650 nm 1440–1650 nm*1

1250–1510 nm*2

*1: On PM15 fiber.

Attenuation Range IL to 60 dB typ.

Calibrated Range Up to 40 dB at 1310 nm and 1550 nm

Up to 40 dB at 1550 nm*1,1625 nm*1 and 1310 nm*2

Attenuation Accuracy (typ.)*3 ±0.3 dB

Insertion Loss < 2 dB (1 dB typ.)Attenuation Setting Resolution 0.1 dB

Polarization Dependent Loss*4 < 0.1 dB

PER n/a ≥ 18 dB

Return Loss*5 > 50 dB

Maximum Input Power 2 W (+33 dBm)

Optical Connectors*6 FC/APC narrow key

Specifications apply for wavelengths not equal to any water absorption line.

Page 50: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ATN-HP Module

50 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

6.1.2 OSICS ATN-HP Module Overview

Figure 23: ATN-HP Module – Front and Rear Panels

Front Panel Optical Connectors

Fiber optic connectors are standard input/output FC/APC type connectors (narrow key). For more details on the appropriate fiber type and laser power compatibility, see section OSICS ATN-HP Technical Specifications, p. 49.

The label located on the optical output connector cap indicates the location of the laser output.

*2: On PM13 fiber.*3: Inside calibrated range.*4: Total PDL including both FC-APC connectors.*5: RL at 1550 nm for SMF & PM15, RL at 1310 nm for PM13.*6: PMF: slow axis is aligned to the connector key.

Openingfor air output

Top Notch

Bottom Notch

96-pin VMEConnector

Input Optical ConnectorFC/APC type

Retaining Screw

Output Optical ConnectorFC/APC type

If you use high power lasers (> 500 mW), please be aware of risks and procedures needed to handle such high power. Make sure optical connections are perfectly clean to avoid any dust that will result in permanent damage of the optical connector (see section Cleaning Optical Connectors, p. 243). Warranty will not apply in case of burnt connectors.

The use of optical instruments with this product increases eye hazard. Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

CAUTION

WARNING

Page 51: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ATN-HP Module

OSICS User Manual 51

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INRetaining Screw

The captive screw enables you to lock/unlock the module inserted into the slot of the OSICS Mainframe.

Rear Panel Notches

The top and bottom notches enable you to slide the module into the OSICS Mainframe slot (see section Installing an OSICS Module Into the OSICS Mainframe, p. 23).

96-pin VME Connector

The rear panel of the OSICS modules interfaces with the OSICS Mainframe backplane by means of the 96-pin VME connector.

Opening

The opening ensures proper flow of cooled air drawn from the OSICS Mainframe chassis.

Side Panel The labels affixed on the side panel of the OSICS ATN-HP are described in section OSICS Information and Warning Markings, p. 19.

6.2 Setting the OSICS ATN-HP Optical Attenuation

You can change the optical attenuation for each module individually using the ATN Unit Setup menu.You can configure the total attenuation using one of the following method:• Setting the attenuation to a definite value: see section Setting the Attenuation to a

Definite Value, p. 51

• Adjusting the power level in real-time: see section Performing Real-time Adjustments to the ATN Attenuation, p. 52

6.2.1 Setting the Attenuation to a Definite Value

Procedure 1. Access the ATN Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set Attenuation menu and press the knob.

The Set Attenuation menu appears and displays the current attenuation setting.

The OSICS Mainframe does not support "hot-swap" of modules; for more details, see section Handling OSICS Modules Into the OSICS Mainframe, p. 23).

CAUTION

Page 52: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ATN-HP Module

52 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Figure 24: ATN Unit Setup – Set Attenuation menu

3. Modify the value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new value.

The Unit Setup menu is automatically displayed and the optical attenuation is now set.

6.2.2 Performing Real-time Adjustments to the ATN Attenuation

Subject The Adjust Attenuation feature lets you adjust the attenuation of the ATN module in real-time (within the accuracy determined by the smallest decimal-digit of the power value).

Procedure 1. Access the ATN Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Adjust Attenuation menu and press the knob.

The Adjust Attenuation menu appears and displays the current attenuation setting.

Figure 25: ATN Unit Setup – Adjust Attenuation menu

3. To change the attenuation of the ATN-HP module in real-time:

Main

EXIT

1 2 3 4 5 6 7 8ATN Unit Setup - Set Attenuation

ESCAPE

Atn = +10.00 dB ENTER

Attenuation :

Modify the value.

Main

EXIT

1 2 3 4 5 6 7 8ATN Unit Setup - Adjust Attenuation

ESCAPE

ATN = +10.00 dB

Attenuation :

Modify the value.

Press the knob to change the digit.

Page 53: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ATN-HP Module

OSICS User Manual 53

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INa. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit. Rotating the knob increments or decrements the digit above the cursor. Going above "9" increases the next significant digit; going below "0" decreases the next significant digit.

c. Perform steps a. and b. again with the other digits to modify.

The ATN module attenuation is modified in real-time.

4. Press one of the following control buttons:

• ESCAPE: to validate the displayed value and go back to the ATN-HP Unit Setup menu.

• EXIT: to display the System Status screen.

6.2.3 Step-Adjusting the ATN Attenuation

Subject The following procedure explains how to modify the attenuation step by step using the attenuation increment defined in the Set Step menu.The initial attenuation value is originally defined in the Set Attenuation menu (see section Setting the Attenuation to a Definite Value, p. 51).

Before Starting Verify the selected attenuation value (see section Setting the Attenuation to a Definite Value, p. 51)

Procedure 1. Access the ATN Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Step menu and enter the Set Step sub-menu.

The Set Step sub-menu appears and displays the current value of the attenuation step in dB.

Figure 26: ATN Unit Setup – Set Step sub-menu

3. Modify the step value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

Main

EXIT

1 2 3 4 5 6 7 8ATN Unit Setup - Set Step

ESCAPE

Step : +01.00 dB

Modify the value.

ENTER

Page 54: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ATN-HP Module

54 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new step value.

The ATN Unit Setup menu is automatically displayed.

4. To modify the wavelength/frequency defined in the Set Attenuation menu (see section Setting the Attenuation to a Definite Value, p. 51), expand the Step menu, move the cursor before the Step + or Step - sub-menus and press the knob to step-adjust the value.

5. Repeat step 4. until the wanted attenuation value is reached.

6.3 Setting the ATN Wavelength

Subject The Set λ/f menu enables you to set the reference wavelength. Each module is factory-calibrated at different wavelengths depending on the module version (SMF or PMF): see section OSICS ATN-HP Technical Specifications, p. 49.

Procedure 1. Access the ATN Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set λ/f menu and press the knob.

The Set Wavelength/Set Frequency menu appears and displays the current value of wavelength or frequency according to the selected unit.

Figure 27: ATN Unit Setup – Set Wavelength menu

3. Turn the rotary knob to put the cursor before the wanted value and press it to validate your selection.

The Unit Setup menu is automatically displayed and the wavelength/frequency value is now set.

Main

EXIT

1 2 3 4 5 6 7 8ATN Unit Setup - Set Wavelength

ESCAPE

Wavelength :

1550 nm1300 nm

Page 55: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ATN-HP Module

OSICS User Manual 55

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN6.4 Setting the ATN Attenuation Offset

Subject Two independent offset can be set for the two calibrated wavelengths.The final attenuation of the module is given by the following formula:Total attenuation = <Attenuation setting> - <offset>The offset can be set between -10 dB and +10 dB with a 0.01 dB step.If the offset is different from 0, then the o character appears in the Info column of the System Status screen (see Figure 9, p. 30).If an offset is set, the attenuation setting is adjustable from<insertion loss> + <offset value> to 60 dB + <offset value>

Procedure 1. Access the ATN Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Offset menu and press the knob.

The Set Offset menu appears and displays the current selected wavelength and offset.

Figure 28: ATN Unit Setup – Set Offset menu

3. Modify the offset value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new step value.

The ATN Unit Setup menu is automatically displayed and the offset set to the selected value.

Main

EXIT

1 2 3 4 5 6 7 8ATN Unit Setup - Set Offset

ESCAPE

Offset (λ=1300 nm) :

+00.00 dB

Modify the value.

ENTER

Page 56: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ATN-HP Module

56 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Page 57: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 57

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN7. Using the OSICS BKR Module

This section explains how to operate the OSICS BKR via the BKR Unit Setup menu. For more details on how to install and initialize the module into the OSICS Mainframe, see section Installing and Connecting the OSICS Multifunction Platform, p. 21.

7.1 OSICS BKR Presentation

The OSICS BKR module is an optical variable reflector that generates known values of reflectance and operates throughout a large wavelength range.

Figure 29: OSICS BKR Principle

7.1.1 OSICS BKR Technical Specifications

OSICS BKR

Wavelength Range1250–1650 nm on SMF1240–1520 nm on PMF

Reflectance Range Up to 55 dBCalibrated Range Up to 40 dB at 1310 and 1550 nm

Reflectance Accuracy (typ.)*1

*1: Inside calibrated range and up to 35 dB.

±0.3 dB

Insertion Loss ≤ 4 dB (3 dB typ.)

Reflection Setting Resolution*2

*2: From 1 to 10 dB; 0.1 dB for 10 to 40 dB.

0.02 dB

Polarization Dependent Loss 0.2 dBSpeed 0.1 second / 3 dB (typ.)Maximum Input Power 0.2 W (+23 dBm)

Optical ConnectorsFC-APC on SMF-28

FC-APC on PMF PM13All specifications are tested at 23°C +/- 2°C; optical connector included.

Page 58: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS BKR Module

58 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

7.1.2 OSICS BKR Module Overview

Figure 30: OSICS BKR Module – Front and Rear Panels

Front Panel Optical Connectors

The fiber optic connector is standard input/output FC/APC type connectors (narrow key). For more details on the appropriate fiber type and laser power compatibility, see section OSICS BKR Technical Specifications, p. 57.

The label located on the optical output connector cap indicates the location of the laser source.

Retaining Screw

The captive screw enables you to lock/unlock the module inserted into the slot of the OSICS Mainframe.

Rear Panel Notches

The top and bottom notches enable you to slide the module into the OSICS Mainframe slot (see section Installing an OSICS Module Into the OSICS Mainframe, p. 23).

Openingfor air output

Top Notch

Bottom Notch

96-pin VMEConnector

Retaining Screw

Input/Output Optical ConnectorFC/APC type

The use of optical instruments with this product increases eye hazard. Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

The OSICS Mainframe does not support "hot-swap" of modules; for more details, see section Handling OSICS Modules Into the OSICS Mainframe, p. 23).

WARNING

CAUTION

Page 59: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS BKR Module

OSICS User Manual 59

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN96-pin VME Connector

The rear panel of the OSICS modules interfaces with the OSICS Mainframe backplane by means of the 96-pin VME connector.

Opening

The opening ensures proper flow of cooled air drawn from the OSICS Mainframe chassis.

Side Panel The labels affixed on the side panel of the OSICS BKR are described in section OSICS Information and Warning Markings, p. 19.

7.2 Setting the OSICS BKR Optical Reflectance

You can change the optical reflectance for each module individually using the OSICS BKR Unit Setup menu.You can configure the total attenuation using one of the following method:• Setting the reflectance to a definite value: see section Setting the Reflectance to a

Definite Value, p. 59

• Adjusting the reflectance level in real-time: see section Performing Real-time Adjustments to the BKR Reflectance, p. 60

7.2.1 Setting the Reflectance to a Definite Value

Procedure 1. Access the BKR Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set Attenuation menu and press the knob.

The Set Attenuation menu appears and displays the current reflectance setting.

Figure 31: BKR Unit Setup – Set Attenuation menu

3. Modify the value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

Main

EXIT

1 2 3 4 5 6 7 8BKR Unit Setup - Set Attenuation

ESCAPE

Atn = +42.33 dB ENTER

Attenuation :

Modify the value.

Page 60: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS BKR Module

60 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new value.

The Unit Setup menu is automatically displayed and the optical reflectance is now set.

7.2.2 Performing Real-time Adjustments to the BKR Reflectance

Subject The Adjust Attenuation feature lets you adjust the attenuation of the BKR module in real-time (within the accuracy determined by the smallest decimal-digit of the power value).

Procedure 1. Access the BKR Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Adjust Attenuation menu and press the knob.

The Adjust Attenuation menu appears and displays the current reflectance setting.

Figure 32: BKR Unit Setup – Adjust Attenuation menu

3. Modify the value as follows, to update the BKR module reflectance in real-time:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit. Rotating the knob increments or decrements the digit above the cursor. Going above "9" increases the next significant digit; going below "0" decreases the next significant digit.

c. Perform steps a. and b. again with the other digits to modify.

The BKR module attenuation is modified in real-time.

4. Press one of the following control buttons:

• ESCAPE: to validate the displayed value and go back to the BKR Unit Setup menu.

• EXIT: to display the System Status screen.

Main

EXIT

1 2 3 4 5 6 7 8BKR Unit Setup - Adjust Attenuation

ESCAPE

ATN = +42.33 dB

Attenuation :

Modify the value.

Press the knob to change the digit.

Page 61: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS BKR Module

OSICS User Manual 61

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN7.2.3 Step-Adjusting the BKR Attenuation

Subject The following procedure explains how to modify the reflectance step by step using the reflectance increment defined in the Set Step menu.The initial reflectance value is originally defined in the Set Attenuation menu (see section Setting the Reflectance to a Definite Value, p. 59).

Before Starting Verify the selected reflectance value (see section Setting the Reflectance to a Definite Value, p. 59)

Procedure 1. Access the BKR Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Step menu and enter the Set Step sub-menu.

The Set Step sub-menu appears and displays the current value of the reflectance step in dB.

Figure 33: BKR Unit Setup – Set Step sub-menu

3. Modify the step value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new step value.

The BKR Unit Setup menu is automatically displayed.

4. To modify the wavelength/frequency defined in the Set Attenuation menu (see section Setting the Reflectance to a Definite Value, p. 59), expand the Step menu, set the cursor before the Step + or Step - sub-menus and press the knob to step-adjust the value.

5. Repeat step 4. until the wanted reflectance value is reached.

Main

EXIT

1 2 3 4 5 6 7 8BKR Unit Setup - Set Step

ESCAPE

Step : +01.00 dB

Modify the value.

ENTER

Page 62: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS BKR Module

62 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

7.3 Setting the BKR Wavelength/Frequency

Subject The Set λ/f menu enables you to set the reference wavelength. Each module is factory-calibrated at different wavelengths depending on the module version (SMF or PMF): see section OSICS BKR Technical Specifications, p. 57.

Procedure 1. Access the BKR Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set λ/f menu and press the knob.

The Set Wavelength/Set Frequency menu appears and displays the current value of wavelength or frequency according to the selected unit.

Figure 34: BKR Unit Setup – Set Wavelength menu

3. Turn the rotary knob to put the cursor before the wanted value and press it to validate your selection.

The Unit Setup menu is automatically displayed and the wavelength/frequency value is now set.

Main

EXIT

1 2 3 4 5 6 7 8BKR Unit Setup - Set Wavelength

ESCAPE

Wavelength :

1550 nm1300 nm

Page 63: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS BKR Module

OSICS User Manual 63

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN7.4 Setting the BKR Attenuation Offset

Subject Two independent offsets can be set for the two calibrated wavelengths./frequencies.The final reflectance of the module is given by the following formula:Total reflectance = <Reflectance setting> - <offset>The offset can be set between -10 dB and +10 dB with a 0.01 dB resolution.If the offset is different from 0, then the o character appears in the Info column of the System Status screen (see Figure 9, p. 30)If an offset is set, the reflectance setting is adjustable from<insertion loss> + <offset value> to 55 dB + <offset value>

Procedure 1. Access the BKR Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Offset menu and enter one of the calibrated wavelengths/frequencies.

The Set Offset menu appears and displays the current selected wavelength/frequency and offset.

Figure 35: BKR Unit Setup – Set Offset menu

3. Modify the offset value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new step value.

The BKR Unit Setup menu is automatically displayed and the offset set to the selected value.

Main

EXIT

1 2 3 4 5 6 7 8BKR Unit Setup - Set Offset

ESCAPE

Offset (λ=1300 nm) :

+00.00 dB

Modify the value.

ENTER

Page 64: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS BKR Module

64 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Page 65: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 65

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN8. Using the OSICS DFB Module

This section explains how to operate the OSICS DFB via the DFB Unit Setup menu. For more details on how to install and initialize the module into the OSICS Mainframe, see section Installing and Connecting the OSICS Multifunction Platform, p. 21.

8.1 OSICS DFB PresentationThe OSICS DFB is a distributed feedback laser diode module designed for fiber-optic component and system testing. OSICS-DFB incorporates proven Yenista Optics distributed feedback laser diode technology in a compact-size modular format.Emission wavelengths are factory-set on user-selected wavelength. The center wavelength is chosen at the time of order to match one of the ITU-T grid channels. For additional flexibility, the center-wavelength can be finely tuned (see section OSICS DFB Technical Specifications, p. 65).

8.1.1 OSICS DFB Technical Specifications

CWDMSMF

CWDM PM13

CWDM PM15

DWDM PM15

LANWDMSMF

LANWDMPM13

Models Channels See OSICS DFB Datasheets

Wavelength of first channel 1270 nm1310 nm

1450 nm 1529.553 nm 1295.56 nm / 231.4 THz

Wavelength of last channel 1650 nm 1650 nm 1611.787 nm 1309.14 nm / 229.0 THz

Wavelength Channel center wavelength grid ±3 nm Grid matchedTuning range*1 1.6 nm (1.8 nm typ.)

Accuracy*2 ±0.03 nm

Stability over 1 hour*2,*3,*4 ±0.005 nm

Stability over 24 hours*2,*3,*4 ±0.005 nm typ.Power

Maximum10 mW (for channels from 1270 to 1570 nm) 8 mW (for channels from

1590 to 1650 nm)+13 dBm 10 mW

Stability over 1 hour*2,*3,*4 ±0.01 dB

Stability over 24 hours*2,*3,*4 ±0.01 dB typ.Optical Isolation > 30 dBRIN*5

(Relative Intensity Noise)< -140 dB/Hz < -130 dB/Hz

Spectrum Laser line width < 10 MHzSMSR*2 (Side Mode Suppression Ratio)

> 30 dB (40 dB typ.) > 35 dB (45 dB typ.) > 30 dB (40 dB typ.)

Page 66: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

66 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

ModulationTTL

Internal1 Hz to 890 kHz

1 Hz to 890 kHzExternal 16 Hz to 890 kHz

Analog (external/front panel) 150 Hz to 150 MHz

SBS Suppression (internal)*6

Waveform sineFrequency Range 10 kHz to 100 kHz

Modulation Depth 0 to 15%

Interfaces on Module Front Panel*7

Enable key with status LED Power up laserOptical connector FC/APC narrow keyOptical fiber SMF PM13 PM15 SMF PM13Fiber alignment to connector key n/a Slow axis n/a Slow axis

PER (Polarization Extinction Ratio) n/a > 17 dB SMF >17 dB

Electrical connector(analog modulation) Coaxial SMB – 50 Ω

Others Laser safety Class 1 MDimensions (WxHxD) 35 x 128 x 230 mmWeight 1.1 kg

*1: For DFB DWDM and LANWDM: Location of channel center: lower boundary of the range + 0.4 nm < channel center < upper boundary of the range - 0.4 nm.

*2: After warm-up and at maximum power.*3: At a constant temperature.*4: Measured with an APC terminated jumper on a powermeter.*5: RIN within the range 100 MHz–20 GHz measured at 10 dBm output power with RBW = 30 kHz.*6: SBS = Stimulated Brillouin Scattering.*7: See OSICS Mainframe Data Sheet for details on OSICS common specifications and interfaces on the rear panel.

CWDMSMF

CWDM PM13

CWDM PM15

DWDM PM15

LANWDMSMF

LANWDMPM13

Page 67: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 67

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN8.1.2 OSICS DFB Module Overview

Figure 36: DFB Module – Front and Rear Panels

Front Panel On LED

This LED indicates that the module is turned on (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

Enable Button

This button enables you to turn on/off the laser inside the module. The green LED of the button is lit if the optical output is enabled, and the power value is displayed in the System Status screen (see section System Status Screen Description, p. 30).

Modulation Input Connector (Coaxial SMB type)

This connector enables you to directly modulate the laser diode chip of the module (see section Setting Up Analog Modulation Using an External Source, p. 83).

Optical Output Connector

The standard output connector is a FC/APC type connector.For more details on the appropriate fiber type, see section OSICS DFB Technical Specifications, p. 65.

The label located on the optical output connector cap indicates the location of the laser source.

Openingfor air output

Top Notch

Bottom Notch

96-pin VMEConnector

On LED

Enable Button

Modulation Connector

Optical OutputConnector

Retaining Screw

Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.WARNING

Page 68: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

68 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Retaining Screw

The captive screw enables you to lock/unlock the module inserted into the slot of the OSICS Mainframe.

Rear Panel Notches

The top and bottom notches enable you to slide the module into the OSICS Mainframe slot (see section Installing an OSICS Module Into the OSICS Mainframe, p. 23).

96-pin VME Connector

The rear panel of the OSICS modules interfaces with the OSICS Mainframe backplane by means of the 96-pin VME connector.

Opening

The opening ensures proper flow of cooled air drawn from the OSICS Mainframe chassis.

Side Panel The labels affixed on the side panel of the OSICS DFB are described in section OSICS Information and Warning Markings, p. 19.

The OSICS Mainframe does not support "hot-swap" of modules; for more details, see section Handling OSICS Modules Into the OSICS Mainframe, p. 23).

CAUTION

Page 69: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 69

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN8.2 Enabling/Disabling the Light Source of the Module

Subject You can enable/disable the light source in two different ways:• From the OSICS Mainframe Main Setup menu (see section Enabling/Disabling the

Light Source of All Modules, p. 37), you can simultaneously enable or disable the light source in all the modules installed in the OSICS Mainframe. This function overrides the enable/disable state individually set in all modules.

• From the DFB front panel, you can individually enable or disable the light source of the module by using the Enable button as explained in the following procedure. This action does not affect the enable/disable state of other modules installed in the OSICS Mainframe.

For laser safety reasons, the light source in modules are disabled after OSICS initialization.

Before Starting Connect an optical device to the OSICS module (see section Connecting Optical Devices to OSICS Modules, p. 26).

Procedures Enabling the Light Source of the Module

• On the DFB module front panel press the Enable button.

The green LED on the button is lit.In the System Status screen, the output power is displayed in the P field (in dBm or mW depending on the selected power unit) for the corresponding module.

Disabling the Light Source of the Module

• On the DFB module front panel press the Enable button.

The green LED on the button is turned off.In the System Status screen, the message Disabled is displayed for the corresponding module.

• Do not enable the laser while there is no fiber connected to the optical output connector.

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 70: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

70 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

8.3 Setting the Optical Output Power of the DFB Module

8.3.1 Specifying the Optical Output Power of the Module

Subject You can set the optical output power in two different ways:• From the OSICS Mainframe Main Setup menu, you can simultaneously set the

optical power of all modules to the same value: see section Setting the Optical Output Power of All Installed Modules, p. 38.

• From the DFB Unit Setup menu, you can set the output power of the module individually as explained in the following procedure.

The power value is software-limited to stay within the specified range.

Before Starting • Set the optical power unit from the OSICS Mainframe Main Setup menu (see section Selecting the Power Unit for All Installed Modules, p. 38). You cannot set the power unit of an individual module.

Procedure 1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set Power menu and press the knob.

The Set Power menu appears and displays the current power setting.

Figure 37: DFB Unit Setup – Set Power menu

3. Modify the power value as follows:

a. Turn the rotary knob to put the cursor under the digit of +/- sign you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new power value.

The Unit Setup menu is automatically displayed and the optical power value of the module is now modified (according to the module power limits).

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Power

ESCAPE

Power :

P = +03.00 dBm ENTER

Modify the value.

Page 71: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 71

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN8.3.2 Performing Real-time Adjustments to the Output Power

Subject The following procedure explains how to adjust the output power of the module in real-time (within the accuracy determined by the smallest decimal-digit of the power value).The starting power is the value defined in the Set Power menu.

Before Starting Enable the optical output of the module (see section Enabling/Disabling the Light Source of the Module, p. 69).

Procedure 1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Adjust Power menu and press the knob.

The Adjust Power menu appears and displays the current power setting, with the first digit highlighted.

Figure 38: DFB Unit Setup – Adjust Power menu

3. Adjust the power value as follows:

a. Turn the knob clockwise to increase the digit value or anticlockwise to decrease it and press the knob to validate the selected digit.Rotating the knob increments or decrements the digit above the cursor. Going above "9" increases the next significant digit; going below "0" decreases the next significant digit.

b. Perform steps a. and b. again with the other digits to modify.

The module output power is updated in real time when you change the value in this menu.

4. Press one of the following control buttons:

• ESCAPE: to validate the displayed value and go back to the DFB Unit Setup menu.

• EXIT: to display the System Status screen.

The optical power value of the module is now modified (according to the module power limits).

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Adjust Power

ESCAPE

Measured Power :

P = +02.10 dBm

Modify the value.

Press the knob to change the digit

Page 72: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

72 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

8.4 Setting the Optical Emission Wavelength/Frequency

8.4.1 Specifying the Optical Emission Wavelength/Frequency

Subject The following procedure explains how to set the emission wavelength or optical frequency of the OSICS DFB module according to its specifications. Even though the module lets you set wavelength values outside this range, performance of your module is guaranteed only within the wavelength range established by the module specifications

Before Starting • Set the spectral unit from the OSICS Mainframe Main Setup menu (see section Selecting the Spectral Unit for all Installed Modules, p. 39). You cannot set the spectral unit of an individual module.

Procedure 1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set λ/f menu and press the knob.

The Set Wavelength menu appears and displays the emission wavelength or frequency.

Figure 39: DFB Unit Setup – Set Wavelength/Frequency menu

3. Modify the wavelength/frequency value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new wavelength value.

The Unit Setup menu is automatically displayed and the wavelength/frequency value of the module is now modified (according to the module limits).

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Wavelength

ESCAPE

Wavelength :

λ = 1556.02 nm ENTER

Modify the value.

Page 73: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 73

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

8.4.2 Step-adjusting the Wavelength/Frequency Value of the Module

Subject The following procedure explains how to modify the wavelength/frequency of the module step by step using the increment defined in the Set Step sub-menu.You can define a valid step value within the range 0.001–2.000 nm. The starting wavelength is the value originally defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 72).

Procedure 1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Step menu and enter the Set Step sub-menu.

The Set Step sub-menu appears and displays the current value of the wavelength/frequency step in nm or GHz .

Figure 40: DFB Unit Setup – Set Step sub-menu

3. Modify the step value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new wavelength value.

The Unit Setup menu is automatically displayed.

4. To increase or decrease the wavelength/frequency value defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 72) step by step, expand the Step menu and put the cursor before the Step + or Step - sub-menus and press the knob to step-adjust the value.While the module is updating the wavelength/frequency, the message Moving… appears along with the new wavelength/frequency value after step-change.

5. Repeat step 4. until the wanted wavelength/frequency is reached.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Step (nm)

ESCAPE

Step:

01.00 nm ENTER

Modify the value.

Page 74: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

74 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

8.5 Configuring the OSICS DFB Signal ModulationPresentation In addition to the OSICS Mainframe built-in modulation source (see section Configuring

the OSICS Mainframe Modulation, p. 40), the OSICS DFB module holds a built-in digital modulation source.The OSICS DFB module supports the following modulation schemes:• Internal modulation:

• Digital (TTL) modulation using the OSICS DFB internal source• Digital (TTL) modulation using the OSICS Mainframe internal source

• External modulation:• Digital (TTL) modulation using an external source connected to the OSICS

Mainframe• Analog modulation using an external source connected to the OSICS DFB

The OUT 2 BNC connector can be used as an external modulation source for other devices.

Modulation Scheme

Figure 41: Mainframe and DFB Module Modulation Scheme

Page 75: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 75

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN8.5.1 Setting Up Digital (TTL) Modulation Using the DFB Module Internal Source

Subject Digital modulation capabilities of the DFB module is within the 1 Hz–890 kHz frequency range (see the DFB specifications).Internal digital modulation uses the DFB module's built-in TTL-level signal generator (see Figure 41, p. 74). The DFB module’s built-in TTL signal-generator can provide a TTL modulation signal in the range of 1 Hz–555 kHz.The resulting modulated optical output signal follows a pattern as shown in the following figure.

Figure 42: TTL-Modulated Output Power Signal (DFB Module)

Procedure Step 1 – Set the module modulation source to INTERNAL:

1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the TTL Modulation menu and enter the Source sub-menu.

The Modulation source sub-menu appears and the cursor automatically points to the current modulation source setting.

Figure 43: DFB Unit Setup – Modulation Source (Internal)

3. Turn the rotary knob to put the cursor before INTERNAL and press it to validate your selection.

The DFB Unit Setup menu is automatically displayed and the modulation is set to the selected source.

Time Reference

Power Setting

Optical OutputPower

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

MAINFRAME

Select a new value.

Page 76: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

76 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Step 2 – Set the frequency of the module INTERNAL modulation source:

1. In the DFB Unit Setup menu, expand the TTL Modulation menu and enter the Set Frequency sub-menu.

The Set Frequency sub-menu appears and displays the current selected modulation frequency.

Figure 44: DFB Unit Setup – Set Frequency

2. Modify the frequency value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps 1. and 2. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new frequency value.

The frequency is updated to the "real frequency". This screen displays the actual modulation frequency produced by the OSICS DFB oscillator, which may slightly differ from the values you have set.

3. Press the knob to validate the displayed real frequency.

Step 3 – Set the modulation control:

1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the TTL Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Main

EXIT

2 3 4 5 6 7 8Unit Setup - Set Frequency

ESCAPE

Frequency :

f = 0016700 Hz

Modify the value.

ENTER

1

Page 77: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 77

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 45: DFB Unit Setup – Modulation Control (On)

3. Turn the rotary knob to put the cursor before the wanted modulation:

• ON for standard modulation.

• ON INVERTED for reverse modulation

4. Press the rotary knob to validate your selection.

Modulation is enabled. In the System Status screen, the letter M is displayed in the channel Info field of the corresponding module.

5. To stop the modulation: in the Control sub-menu, select OFF.

8.5.2 Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source

Subject This modulation setting uses the OSICS Mainframe’s built-in internal oscillator as a source of TTL-level modulated signals (see Figure 41, p. 74). The main purpose of this setting is to provide simultaneous and synchronous modulation to several modules installed in the OSICS Mainframe, or to an external device connected to the OSICS Mainframe Synchro output connector.The range of the frequency of the modulation provided by the OSICS Mainframe is defined to encompass the frequency requirements of all types of OSICS DFB modules. The frequency settings for the lower and upper bounds may therefore lie outside of the modulation capabilities of a particular module, in which case the module's own frequency range sets the modulation specifications.The OSICS Mainframe signal generator can provide a TTL signal with a repetition frequency in the range 123 Hz–1 MHz.The resulting signal-shape of the modulated output power is similar to the internally modulated pattern of the OSICS DFB, as shown in Figure 42, p. 75.

Procedure Step 1 – Set the OSICS Mainframe modulation source to INTERNAL:

1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Enter the Source sub-menu as follows:

a. Turn the rotary knob to put the cursor before the Modulation menu and press the knob to expand it.

Main

EXIT

1 2 3 4 5 6 7 8DFB Unit Setup - Modulation control

ESCAPE

TTL Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Page 78: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

78 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

b. Turn the rotary knob to put the cursor before the Source sub-menu and press the knob to enter it.

The Modulation Source sub-menu appears and displays the current selected modulation source.

Figure 46: Main Setup – Modulation Source (Internal)

3. Turn the rotary knob to put the cursor before INTERNAL and press it to validate your selection.

The Main Setup menu is automatically displayed and the modulation is set to the selected source.

Step 2 – Set the frequency of the OSICS Mainframe internal modulated-signal:

1. In the Main Setup menu, expand the Modulation menu and select the Set frequency sub-menu.

The Modulation frequency sub-menu appears and displays the current selected modulation frequency-value.

Figure 47: Main Setup – Modulation Frequency

2. Modify the frequency value as follows:

The internal TTL-signal generator can provide a TTL-modulated signal in the range 123 Hz to 1 MHz.

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new frequency value.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

EXTERNAL

Select a new value.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation frequency

ESCAPE

Frequency value :

f = 0014700 Hz

Modify the value.

ENTER

Page 79: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 79

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INThe frequency is updated to the "real frequency" as shown in Figure 48, p. 79. This screen displays the current modulation frequency produced by the OSICS Mainframe oscillator, which may slightly differ from the values you have set.

Figure 48: Main Setup – Modulation Real Frequency

3. Press the knob to go back to the Main Setup menu.

Step 3 – Set the module modulation source to MAINFRAME:

1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the TTL Modulation menu and enter the Source sub-menu.

The Modulation source sub-menu appears and the cursor automatically points to the current modulation source setting.

Figure 49: DFB Unit Setup – Modulation Source (Mainframe)

3. Turn the rotary knob to put the cursor before MAINFRAME and press it to validate your selection.

The DFB Unit Setup menu is automatically displayed and the modulation is set to the selected source.

Step 4 – Set the modulation control:

1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the TTL Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation frequency

ESCAPE

The real frequency is :

14706 Hz

Please press the knob.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

MAINFRAME

Select a new value.

Page 80: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

80 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Figure 50: DFB Unit Setup – Modulation Control (On)

3. Turn the rotary knob to put the cursor before the wanted modulation:

• ON for standard modulation.

• ON INVERTED for reverse modulation

4. Press the rotary knob to validate your selection.

Modulation is enabled. In the System Status screen, the letter M is displayed in the channel Info field of the corresponding module.

5. To stop the modulation: in the Control sub-menu, select OFF.

8.5.3 Setting Up Digital (TTL) Modulation Using an External Source

Subject External digital-modulation is achieved by applying a digital signal to the BNC Mod. In connector located on the OSICS Mainframe rear panel (see Figure 41, p. 74). This allows all the modules (or an external device connected to the OSICS Mainframe Synchro output connector) to be modulated in a synchronous way by means of an external source. The OSICS Mainframe distributes the external modulated-signal to all installed modules in order to synchronize all modules with a single modulated signal, amplitude, and frequency.For optimum stability of the modulated signals, we recommend applying a digital modulation in the specified range of the modules.

Before Starting Make sure you have a digital-signal generator.

Procedure Step 1 – Set the OSICS Mainframe modulation source to EXTERNAL:

1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Enter the Source sub-menu as follows:

a. Turn the rotary knob to put the cursor before the Modulation menu and press the knob to expand it.

b. Turn the rotary knob to put the cursor before the Source sub-menu and press the knob to enter it.

The Modulation source sub-menu appears and displays the current selected modulation source.

Main

EXIT

1 2 3 4 5 6 7 8DFB Unit Setup - Modulation control

ESCAPE

TTL Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Page 81: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 81

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 51: Main Setup – Modulation Source (External)

3. Turn the rotary knob to put the cursor before EXTERNAL and press it to validate your selection.

The Main Setup menu is automatically displayed and the modulation is set to the selected source.

Step 2 – Set the TTL-level signal generator:

1. On the OSICS Mainframe rear panel, connect your external digital signal-generator to the Mod. in BNC connector (see Figure 2, p. 18).

2. Apply the TTL signal not exceeding the maximum voltage of +5 V and repetition frequency between 1 Hz and 890 kHz.

The green LED located next to the Mod. In connector is lit to indicate that an external signal-generator is connected.

Step 3 – Set the module modulation source to MAINFRAME:

1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the TTL Modulation menu and enter the Source sub-menu.

The Modulation source sub-menu appears and the cursor automatically points to the current modulation source setting.

Figure 52: DFB Unit Setup – Modulation Source (Mainframe)

3. Turn the rotary knob to put the cursor before MAINFRAME and press it to validate your selection.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

EXTERNAL

Select a new value.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

MAINFRAME

Select a new value.

Page 82: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

82 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

The DFB Unit Setup menu is automatically displayed and the modulation is set to the selected source.

Step 4 – Set the modulation control:

1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the TTL Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Figure 53: DFB Unit Setup – Modulation Control (On)

3. Turn the rotary knob to put the cursor before the wanted modulation:

• ON for standard modulation.

• ON INVERTED for reverse modulation

4. Press the rotary knob to validate your selection.

Modulation is enabled. In the System Status screen, the letter M is displayed in the channel Info field of the corresponding module.

5. To stop the modulation: in the Control sub-menu, select OFF.

Main

EXIT

1 2 3 4 5 6 7 8DFB Unit Setup - Modulation control

ESCAPE

TTL Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Page 83: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 83

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

8.5.4 Setting Up Analog Modulation Using an External Source

Subject The SMB connector located on the front panel of the DFB module is directly connected to the laser chip and allows you to couple an external modulating-current to the system bias current (see Figure 41, p. 74).By modulating this external current, you can modulate the optical power output in a frequency range from 150 Hz to 150 MHz.

Figure 54: External Analog Modulation Setup

The relationship between the voltage applied at the SMB input and the current added to the diode is given by: Iin = Vin/Rinternal

where:• Iin is the added current• Vin is the voltage applied to the SMB input• Rinternal is the 50 Ω internal impedance of your DFB module

TTL modulation can also be applied directly on the module's front panel using the input SMB connector generally used for the analog modulation. The modulation is applied directly on the laser chip. In this case, simultaneous modulation for all the modules is not possible.Laser diode current level is available in section OSICS DFB Technical Specifications, p. 65.

Before Starting Make sure you have a 50 Ω coaxial cable to connect your HF-generator to the 50 Ω input SMB connector.

Procedure 1. Set the module modulation control to off as follows:

OSICS Module Laser Diode Source

Rinternal

50 Ω

Waveform Generator

Ibias

Polarization Current

Iin

Idiode

Vin

Coaxial cable 50Ω

Mod.SMB Connector

50Ω

Before applying an external analog-modulation, make sure the modulation control is set to off.

CAUTION

Page 84: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

84 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

a. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

b. Expand the TTL Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Figure 55: DFB Unit Setup – Modulation Control (Off)

c. Turn the rotary knob to put the cursor before OFF and press it to validate your selection.

Modulation control is disabled and the DFB Unit Setup menu is displayed.

2. Connect your HF-generator to the 50 Ω input SMB connector using a 50 Ω coaxial cable.

3. Since the SMB input is directly connected to the laser diode chip, make sure that the input voltage is such that the following conditions are always met:

• The diode laser is forward-biased.

• The current flowing through the diode is lower than the maximum allowed current (Imax) indicated in the system acceptance test report.

The best way to fulfill the above conditions is to monitor the DFB module bias-current (Ibias) via the rated output voltage measured at the OUT 1 port located on the rear panel of the OSICS Mainframe (see Figure 2, p. 18), and to compute the total current using the following formula:

Idiode = Ibias + Vin/Rinternal

Where:

• Idiode is the total current through the diode• Ibias is the bias current• Vin is the voltage applied to the SMB input• Rinternal is the 50 Ω internal impedance of the DFB moduleYou must monitor Vin to ensure that the total current remains positive and lower than the maximum allowed current (Imax), as shown in the following figure:

0 < Idiode < Imax

The SMB input is directly connected to the laser diode chip via a 50 Ω resistor. It is essential to make sure that at all times the diode is forward-biased to avoid damage or possible destruction of the diode.

Main

EXIT

1 2 3 4 5 6 7 8DFB Unit Setup - Modulation control

ESCAPE

TTL Modulation control :

OFF

ON

Select a new value.

ON INVERTED

CAUTION

Page 85: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 85

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 56: Resulting Diode Current with External Analog Modulation Source

If the voltage applied to the SMB input is an AC signal with no DC component and with peak-to-peak amplitude is Vpp, the above condition becomes:Vpp < 2 x Rinternal (Imax - Ibias)Vpp < 2 x Rinternal (Ibias)

Ibias

tIbias

tVin

t

Idiode

Idiode Ibias

Diode modulated current

Current set on OSICSwithout modulation

Input modulated current

Imax

IA

IB

IA < Imax

IB > 0

0

Page 86: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

86 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

8.5.5 Setting Up Analog Modulation With the Internal Sinus Modulation

Subject Sinus Modulation is an internal modulation feature implemented on the DFB module and designed to limit the effect of Brillouin scattering. The DFB module can generate a sinusoidal signal with a user-defined frequency in the range of 10 kHz–100 kHz and a user-defined modulation-rate in the range of 0%–15% (of the diode current value) to modulate the current of the diode.The following figure shows the resulting sinusoidal modulation of the laser diode bias-current for a given modulation rate (< 15%) and frequency. The new diode current fluctuates symmetrically around the original value of the bias-current (Ibias).

Ibias is the mean average of the new modulated current.

Figure 57: Diode current resulting from sinusoidal modulation

8.5.5.1 Specifying the Sinus Frequency

Subject To use the Sinus frequency, you must first set the internal frequency (in the range 10 kHz–100 kHz). The frequency unit cannot be modified; it is defined in kHz.

Procedure 1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Sinus Modulation menu and enter the Set Frequency sub-menu.

The Set Sinus Frequency sub-menu appears and displays the current frequency of the sinus modulation signal.

Ibias

Max. Rate = 15 %

Timereference

Diode current

Page 87: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 87

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 58: DFB Unit Setup – Sinus Modulation (Set Frequency)

3. Modify the frequency value as follows:

A valid sinusoidal frequency value may range from 10 kHz to 100 kHz.

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps 1. and 2. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new frequency value.

The sinus frequency is updated.

Main

EXIT

2 3 4 5 6 7 8DFB Unit Setup - Set Sinus Frequency

ESCAPE

Frequency :

f = 030.0 kHz

Modify the value.

ENTER

1

Page 88: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

88 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

8.5.5.2 Specifying the Sinus Rate

Subject The sinus rate is defined as the amplitude of the sinusoidal modulating-signal. This rate value allows to calculate the amplitude of the modulation as a direct percentage of the diode bias-current (Ibias). The rate can range from 0% to 15% of the diode bias-current. Setting this rate value to 15% means that the resulting diode current is modulated ±15% around the mean average (Ibias).

Procedure 1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Sinus Modulation menu and enter the Set Rate sub-menu.

The Set Sinus Rate sub-menu appears and displays the current rate of the sinusoidal modulation signal.

Figure 59: DFB Unit Setup – Sinus Modulation (Set Rate)

3. Modify the rate value as follows:

A valid sinus rate value may range from 5 % to 15 %.

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease the value and press the knob to validate the selected digit.

c. Perform steps 1. and 2. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new rate value.

The sinus rate is updated.

Main

EXIT

2 3 4 5 6 7 8DFB Unit Setup - Set Sinus Rate

ESCAPE

Rate :

Rate = 10.0 %

Modify the value.

ENTER

1

Page 89: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 89

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN8.5.5.3 Turning on/off the Sinus Modulation

Subject This section explains how to switch on and off the sinus modulation.

Before Starting Make sure the internal frequency and modulation rate are set (see section Specifying the Sinus Frequency, p. 86 and section Specifying the Sinus Rate, p. 88).

Procedure 1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Sinus Modulation menu and enter the Control sub-menu.

The Modulation Control sub-menu appears and displays the current state of the modulation function.

Figure 60: DFB Unit Setup – Sinus Modulation Control

3. Turn the rotary knob to put the cursor before the ON setting:

4. Press the rotary knob to validate your selection.

The sinus modulation is implemented immediately.

In the System Status screen, the message S appears in front of the corresponding module channel.

5. To stop the modulation: in the Control sub-menu, select OFF.

Main

EXIT

1 2 3 4 5 6 7 8DFB Unit Setup - Modulation control

ESCAPE

Sinus Modulation control :

OFF

ON

Select a new value.

Page 90: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

90 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

8.5.6 Using the DFB Module as a Modulation Source for External Devices (OUT 2 Output)

Subject The OSICS Mainframe rear panel (see Figure 2, p. 18) features eight channel ports, each dedicated to a specific channel-slot and identified by its channel number (CH1 to CH8).As they are numbered from left to right, channel ports do not face across their corresponding slots.Each channel can produce two types of output signals on BNC connectors:• OUT 1: analog signal for constant monitoring of the laser diode current, output

power or laser chip temperature. For more details, see section Monitoring the OSICS DFB Module Parameters Using the Analog OUT 1 Output, p. 91.

• OUT 2: analog signal from the DFB module (see Figure 41, p. 74) that you can use as an external modulation source for other devices. The following procedure explains how to set the OUT 2 analog signal.

You can also use the OSICS Mainframe as a modulation source for external devices with the Synchro BNC output connector, as explained in section Using the Mainframe as a Modulation Source for External Devices (Synchro Output), p. 44.

Procedure 1. Set up the wanted digital modulation to the DFB module using one of the following methods:

• DFB module internal modulation source, as explained in section Setting Up Digital (TTL) Modulation Using the DFB Module Internal Source, p. 75.

• OSICS Mainframe internal modulation source, as explained in section Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source, p. 77.

• OSICS Mainframe external modulation source, as explained in section Setting Up Digital (TTL) Modulation Using an External Source, p. 80.

2. Connect the wanted appropriate device to the OUT 2 BNC connector corresponding to the channel slot number of the module.

The selected modulation frequency is applied to the connected device.

Page 91: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 91

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN8.6 Monitoring the OSICS DFB Module Parameters Using the Analog OUT 1 Output

Subject The OSICS Mainframe rear panel (see Figure 2, p. 18) features eight channel ports, each of them is dedicated to a specific channel-slot and identified by its channel number (CH1 to CH8).As they are numbered from left to right, channel ports do not face across their corresponding slots.Each channel can produce two types of output signals on BNC connectors:• OUT 2: analog signal from the DFB module that you can use as an external

modulation source for other devices. For more details, see section Using the DFB Module as a Modulation Source for External Devices (OUT 2 Output), p. 90.

• OUT 1: analog signal for constant monitoring of the DFB laser diode current (ILD), output power (POUT), or laser chip temperature (θ).

The scale-factors for the OUT 1 signal are the following (FPGA v.6.00):

• Laser diode current (ILD): 10 V / 300 mA

• Optical output power (POUT): 10 V / 30 mW

• Laser chip temperature (θ): 0 V = 10°C; 10 V = 45°C

The following procedure explains how to set the OUT 1 analog signal.

Before Starting Make sure you have an oscilloscope.

Procedure 1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Analog out menu and press the knob.

The Analog Out menu appears and displays the monitoring signal currently active on the OUT1 connector.

Figure 61: DFB Unit Setup – Analog Out setting

3. Turn the rotary knob to put the cursor before the parameter you want to monitor and press it to validate your selection.

The selected parameter is validated and the DFB Unit Setup menu is displayed.

4. To monitor the selected signal, connect your oscilloscope to the OUT 1 BNC connector corresponding to the channel slot number of the module.

Main

EXIT

2 3 4 5 6 7 8DFB Unit Setup - Analog Out

ESCAPE

Analog Out :

Select a new signal.

1

CURRENT

POWER

TEMPERATURE

Page 92: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

92 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

8.7 Calibrating the DFB Module

8.7.1 Performing a Wavelength Calibration

Subject The Wavelength Calibration function enables you to apply a wavelength offset to the DFB emission wavelength. The value entered as wavelength offset is a "d " offset between the calibrated wavelength and the wavelength announced by the DFB display.You can use the wavelength offset in either of two ways:• To apply an arbitrary offset to the displayed wavelength, as required by your

particular application.

• To apply a calibration offset computed as the difference between the wavelength reading on the wavemeter and the value displayed by the DFB module.

The wavelength offset applies on the DFB tuning mechanism and changes the value of the actual emission wavelength. The wavelength value displayed by the DFB is unchanged. For example, if the wavelength setting of the DFB module is 1560.000 nm, and you set an offset of +0.100 nm, the actual emission wavelength at the optical output is offset +0.100 nm but the DFB display is unchanged and still shows 1560.000 nm.You can set a wavelength offset from -0.200 nm to +0.200 nm.

Before Starting Enable the optical output of the module (see section Enabling/Disabling the Light Source of the Module, p. 69).

Procedure 1. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Calibration menu and enter the Wavelength calibration sub-menu.

The Wavelength calibration sub-menu appears prompting you to enter the wavelength offset in nm.

Figure 62: DFB Unit Setup – Wavelength calibration

3. Modify the wavelength value as follows:

A valid wavelength offset value can range from -0.200 nm to +0.200 nm.

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease the value and press the knob to validate the selected digit.

Main

EXIT

2 3 4 5 6 7 8DFB Unit Setup - Wavelength calibration

ESCAPE

Wavelength offset :

Modify the value.

ENTER

1

dλ = +0.100 nm

Page 93: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 93

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INc. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the value.

A confirmation message is displayed

4. Press the knob to confirm.

The message Calibrating… is displayed during the calibration process. It may take up to two minutes.Once the calibration is performed, the Unit Setup menu is displayed.

8.7.2 Performing a Power Calibration

Subject This section explains how to perform a one-point power calibration sequence using the Power calibration function. The purpose of this calibration is to correct the discrepancies between the power displayed by the OSICS DFB module and the power measured at your reference power meter.The new calibration data sets the new system calibration default.

Before Starting Make sure you have a power meter.

Procedure 1. Set the DFB output power to 1 mW.

2. Connect the optical output of the OSICS DFB module to a power meter.

3. Enable the module optical output (see section Enabling/Disabling the Light Source of the Module, p. 69).

4. Access the DFB Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

5. Expand the Calibration menu and enter the Power calibration sub-menu.

The message Moving... appears, and the Power calibration sub-menu appears, prompting you to correct the power value.

Figure 63: DFB Unit Setup – Power calibration

6. Measure the power level on the power meter.

A valid power value can range from 0.6 mW to 1.2 mW.

7. Enter the power value measured by the power meter and select ENTER.

The message Moving... appears while the power value is set. A confirmation message appears.

Main

EXIT

2 3 4 5 6 7 8DFB Unit Setup - Power Calibration

ESCAPE

Power :

P = +06.771 dBm

Correct the value.

ENTER

1

Page 94: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

94 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

8. Press the knob to confirm.

The message Moving... appears while the power value is set. The power calibration is now complete and the values are stored in the module's non-volatile memory.

The DFB Unit Setup menu is displayed.

The new calibration data sets the new system calibration default.

8.7.3 Asking for a Factory Recalibration

Subject To externally verify the performance of the OSICS, refer to the procedure described in section Performing a Wavelength Calibration, p. 92 and section Performing a Power Calibration, p. 93. If the performance does not meet warranted specifications, a complete calibration of the instrument may be required.We recommend performing factory calibration of OSICS DFB modules over one-year intervals.

Procedure Contact the Yenista Optics customer support service (see section Contact Information, p. 4) for a complete calibration, as some calibration steps require the adjustment of internal components including the laser head.

Page 95: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 95

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN8.8 Testing the Performance of the DFB Module

Use the procedures described in this section to validate warranted specifications of the OSICS DFB module, as needed.During the performance verification, the settings for wavelength and power may not agree with the measured values. If this occurs, you may calibrate wavelength and power settings as explained in section Performing a Wavelength Calibration, p. 92 and section Performing a Power Calibration, p. 93.Notwithstanding performance tests described in this section, we recommend performing calibration of OSICS DFB modules over one-year intervals (see section Asking for a Factory Recalibration, p. 94).

8.8.1 Preparing Performance Tests

Before Starting Make sure you have the following required equipment for conducting performance tests on OSICS DFB modules:

The various types as well as the number of fiber-optic connectors needed may vary depending on the type of equipment used and the distance between the test equipment and the OSICS DFB modular optical source under test.

Procedure 1. In the OSICS Mainframe, install the DFB modules you want to test.

2. Turn-on the OSICS multifunction platform (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

3. Let the OSICS multifunction platform and other test equipments run continuously for 2 hours at room temperature (within the specified operating temperature range) to achieve proper equipment warm-up.

Equipment Minimum RequirementOptical power meter Wavelength range: 1260 nm to 1640 nmOptical spectrum analyzer (OSA) Wavelength range: 1260 nm to 1640 nmWavemeter Wavelength range: 1260 nm to 1640 nmJumper (x2) FC/APC to FC/PC or to FC/APCIn-line optical adapter FC to FC

• Disable the laser output before connecting or disconnecting a fiber optic cable on the module (see section Enabling/Disabling the Light Source of All Modules, p. 37).

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 96: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

96 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

8.8.2 Measuring the Absolute Wavelength Accuracy

Subject This test procedure enables you to verify that the actual measured wavelength stays within acceptable deviation range from the wavelength setting.

Before Starting • Make sure you have a wavemeter with ±1 pm resolution and a NIST-traceable calibration (see section Preparing Performance Tests, p. 95) and jumper.

Procedure 1. Connect the OSICS DFB module to the wavemeter.

2. Set the OSICS DFB output power to maximum (the value of the maximum output depends on your DFB model).

3. Set the wavelength of the OSICS DFB tunable laser to the minimum value of the wavelength range. The minimum wavelength depends on the DFB model you are using (see section Specifying the Optical Emission Wavelength/Frequency, p. 72).

4. Set the wavelength step to +0.1 nm.

5. Enable the laser output by pressing the Enable button on the module front panel.6. Compute the wavelength deviation of the DFB module and check that it is not greater

than ±0.03 nm compared with the wavelength reading of the wavemeter.

The absolute wavelength accuracy is the deviation between the wavelength setting and the wavelength reading provided by the wavemeter at each measurement step.

7. Increment the DFB wavelength setting one wavelength step (+0.1 nm) and repeat measurement step 6. to check that the wavelength deviation is not greater than ±0.03 nm.

8. Stop when you reach the DFB maximum wavelength limit.

If the minimum and maximum wavelength limits do not meet the specifications, you can calibrate these values as explained in section Performing a Wavelength Calibration, p. 92.

8.8.3 Verifying the Wavelength Stability

Subject This procedure enables you to verify the DFB module wavelength stability in the long-term run, typically over 1-hour continuous use.

Before Starting Make sure you have an appropriate wavemeter (see section Preparing Performance Tests, p. 95) and jumper.

Procedure 1. Disable the DFB laser output.

2. Connect the OSICS DFB module to the wavemeter.

3. Set the OSICS DFB output power to maximum (the value of the maximum output depends on your DFB model, see section OSICS DFB Technical Specifications, p. 65).

4. Set the wavelength of the OSICS DFB laser to the center wavelength (the value of the center wavelength depends on your DFB model, see section OSICS DFB Technical Specifications, p. 65).

5. Enable the laser output by pressing the Enable button on the DFB front panel.

Page 97: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

OSICS User Manual 97

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN6. Record the wavelength reading on the wavemeter every 30 seconds during 1 hour.

7. After one-hour sampling:

a. Pick the maximum and the minimum wavelength values from the wavelength records.

b. Subtract the minimum wavelength value from the maximum recorded wavelength value to get the wavelength stability.

c. Verify that the calculated wavelength stability value is within the specification range of ±0.005 nm/h.

8.8.4 Verifying the Power Stability

Subject This procedure enables you to verify if the DFB module power stability is within specifications in the long-term run, typically over 1-hour continuous use.

Before Starting Make sure you have an appropriate power-meter (see section Preparing Performance Tests, p. 95) and jumper.

Procedure 1. Disable the DFB laser output.

2. Connect the OSICS DFB module to the power-meter.

3. Set the OSICS DFB output power to maximum (the value of the maximum output depends on your DFB model, see section OSICS DFB Technical Specifications, p. 65).

4. Set the wavelength of the OSICS DFB laser to the center wavelength (the value of the center wavelength depends on your DFB model, see section OSICS DFB Technical Specifications, p. 65).

5. Enable the laser output by pressing the Enable button on the DFB front panel.

6. Record the wavelength reading on the power-meter every 30 seconds during 1 hour.

7. After one-hour sampling:

a. Pick the maximum and the minimum power values from the wavelength records.

b. Subtract the minimum power value from the maximum recorded power value to get the power stability.

c. Verify that the calculated wavelength stability value is within the specification range of ±0.01 db/h.

Page 98: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS DFB Module

98 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Page 99: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 99

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN9. Using the OSICS ECL Module

This section explains how to operate the OSICS ECL via the ECL Unit Setup menu. For more details on how to install and initialize the module into the OSICS Mainframe, see section Installing and Connecting the OSICS Multifunction Platform, p. 21.

9.1 OSICS ECL Presentation

The OSICS ECL incorporates proven Yenista Optics TUNICS tunable external cavity laser technology in a compact-size modular format. The optical layout is a modified Littman-Metcalf configuration. Double-pass reflection on the grating provides maximum dispersion, while the very short cavity, which maximizes mode spacing, yields a spectrally pure and truly single-mode operation.The high-performance optical isolator and angle-polished output fiber connector also contribute to spectral purity by protecting the laser cavity from any interference that may be generated by your equipment.In the ECL configuration, however, the dihedral rear reflector in the self-aligned cavity acts as a 2D corner cube, so that the resonator remains perfectly in tune, overriding any small misalignments and thus guaranteeing long-term stability.

9.1.1 OSICS ECL Technical Specifications

ECL 1560

Tuning Characteristics

Wavelength Range (nm) P = 0 dBm 1520–1600

P = 6 dBm 1530–1580

Wavelength Setting Accuracy*5 ±0.2 nm

Wavelength Setting Resolution 0.01 nm (0.001 nm option R)

Wavelength Stability*1 ±0.01 nm/h (±0.01 nm / 24 h typical)

Wavelength Setting Repeatability ±0.01 nm typical

Tuning Speed (step-bystep)*2*3 10 nm/s typical

Laser Output Characteristics

SSSER ≥ 45 dB / 0.1 nm typical

SMSR*4 ≥ 45 dB

Output Power Stability*5 ±0.01 dB/h (±0.01 dB / 24 h typical)

Relative Intensity Noise*6 <-140 dB/Hz

Spectral Width (FWHM) 150 kHz typical (coherence control off)

> 100 MHz (coherence control on)

Laser Safety Classification Class 1M

Page 100: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

100 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Permissible Values

Absolute Maximum Ratings

Interfaces

Analog Modulation 50 Hz to 50 MHz (external)

Digital Modulation 50 Hz to 1 MHz (internal & external)

Output Fiber Type SMF or PMF (option M)

Output Connector FC/APC narrow key

Physical Specifications

Dimensions (W x H x D) 35 mm (1.4 in) x 128 mm (5.0 in) x 230 mm (9.0 in)

Weight 1.07 kg (2.36 lb)

All specifications are given after 60 minutes warm-up and apply for wavelengths not equal to any water absorption line.

*1: At constant temperature.*2: With the high resolution option (R) the tuning speed is 2.5 nm/s typ.*3: The kinematic chain of the laser does not allow for swept operation.*4: Measured over a 0.1 nm bandwidth ±1nm from the signal.*5: Measured at 0 dBm output power.*6: RIN within the range 100 MHz–3 GHz measured at +3 dBm output power with RBW = 30 kHz

over 1530–1580 nm.

ECL 1560

ECL Model Maximum Power Laser Diode Current LevelP= 0 dBm 1 mW 200 mAP= 6 dBm 4 mW 200 mA

Minimum output power > 0.2 mWMaximum output power < 15 mWUpper and lower limits of operating wavelength range

1500-1620 nm

Maximum current 200 mA

Page 101: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 101

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN9.1.2 OSICS ECL Module Overview

Figure 64: ECL Module – Front and Rear Panels

Front Panel On LED

This LED indicates that the module is turned on (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

Enable Button

This button enables you to turn on/off the laser inside the module. The green LED of the button is lit if the optical output is enabled, and the power value is displayed in the System Status screen (see section System Status Screen Description, p. 30).

Modulation Input Connector (Coaxial SMB type)

This connector enables you to directly modulate the laser diode chip of the module (see section Setting Up Analog Modulation Using an External Source, p. 119).

Optical Output Connector

The standard output connector is a FC/APC type connector.For more details on the appropriate fiber type, see section OSICS ECL Technical Specifications, p. 99.

The label located on the optical output connector cap indicates the location of the laser source.

Openingfor air output

Top Notch

Bottom Notch

96-pin VMEConnector

On LED

Enable Button

Modulation Connector

Optical OutputConnector

Retaining Screw

Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.WARNING

Page 102: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

102 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Retaining Screw

The captive screw enables you to lock/unlock the module inserted into the slot of the OSICS Mainframe.

Rear Panel Notches

The top and bottom notches enable you to slide the module into the OSICS Mainframe slot (see section Installing an OSICS Module Into the OSICS Mainframe, p. 23).

96-pin VME Connector

The rear panel of the OSICS modules interfaces with the OSICS Mainframe backplane by means of the 96-pin VME connector.

Opening

The opening ensures proper flow of cooled air drawn from the OSICS Mainframe chassis.

Side Panel The labels affixed on the side panel of the OSICS ECL are described in section OSICS Information and Warning Markings, p. 19.

The OSICS Mainframe does not support "hot-swap" of modules; for more details, see section Handling OSICS Modules Into the OSICS Mainframe, p. 23).

CAUTION

Page 103: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 103

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN9.2 Enabling/Disabling the Light Source of the Module

Subject You can enable/disable the light source in two different ways:• From the OSICS Mainframe Main Setup menu (see section Enabling/Disabling the

Light Source of All Modules, p. 37), you can simultaneously enable or disable the light source in all the modules installed in the OSICS Mainframe. This function overrides the enable/disable state individually set in all modules.

• From the ECL front panel, you can individually enable or disable the light source of the module by using the Enable button as explained in the following procedure. This action does not affect the enable/disable state of other modules installed in the OSICS Mainframe.

For laser safety reasons, the light source in modules are disabled after OSICS initialization.

Before Starting Connect an optical device to the OSICS module (see section Connecting Optical Devices to OSICS Modules, p. 26).

Procedures Enabling the Light Source of the Module

• On the ECL module front panel press the Enable button.

The green LED on the button is lit.In the System Status screen, the output power is displayed in the P field (in dBm or mW depending on the selected power unit) for the corresponding module.

Disabling the Light Source of the Module

• On the ECL module front panel press the Enable button.

The green LED on the button is turned off.In the System Status screen, the message Disabled is displayed for the corresponding module.

• Do not enable the laser while there is no fiber connected to the optical output connector.

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 104: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

104 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

9.3 Setting the Optical Output Power of the ECL Module

9.3.1 Specifying the Optical Output Power of the Module

Subject You can set the optical output power in two different ways:• From the OSICS Mainframe Main Setup menu, you can simultaneously set the

optical power of all modules to the same value: see section Setting the Optical Output Power of All Installed Modules, p. 38.

• From the ECL Unit Setup menu, you can set the output power of the module individually as explained in the following procedure.

The power value is software-limited to stay within the specified range.

Before Starting • Set the optical power unit from the OSICS Mainframe Main Setup menu (see section Selecting the Power Unit for All Installed Modules, p. 38). You cannot set the power unit of an individual module.

Procedure 1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set Power menu and press the knob.

The Set Power menu appears and displays the current power setting.

Figure 65: ECL Unit Setup – Set Power menu

3. Modify the power value as follows:

a. Turn the rotary knob to put the cursor under the digit of +/- sign you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new power value.

The Unit Setup menu is automatically displayed and the optical power value of the module is now modified (according to the module power limits).

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Power

ESCAPE

Power :

P = +03.00 dBm ENTER

Modify the value.

Page 105: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 105

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN9.3.2 Performing Real-time Adjustments to the Output Power

Subject The following procedure explains how to adjust the output power of the module in real-time (within the accuracy determined by the smallest decimal-digit of the power value).The starting power is the value defined in the Set Power menu.

Before Starting Enable the optical output of the module (see section Enabling/Disabling the Light Source of the Module, p. 103).

Procedure 1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Adjust Power menu and press the knob.

The Adjust Power menu appears and displays the current power setting, with the first digit highlighted.

Figure 66: ECL Unit Setup – Adjust Power menu

3. Adjust the power value as follows:

a. Turn the knob clockwise to increase the digit value or anticlockwise to decrease it and press the knob to validate the selected digit.Rotating the knob increments or decrements the digit above the cursor. Going above "9" increases the next significant digit; going below "0" decreases the next significant digit.

b. Perform steps a. and b. again with the other digits to modify.

The module output power is updated in real time when you change the value in this menu.

4. Press one of the following control buttons:

• ESCAPE: to validate the displayed value and go back to the ECL Unit Setup menu.

• EXIT: to display the System Status screen.

The optical power value of the module is now modified (according to the module power limits).

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Adjust Power

ESCAPE

Measured Power :

P = +02.10 dBm

Modify the value.

Press the knob to change the digit

Page 106: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

106 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

9.4 Setting the Optical Emission Wavelength/Frequency

9.4.1 Specifying the Optical Emission Wavelength/Frequency

Subject The following procedure explains how to set the emission wavelength or optical frequency of the OSICS ECL module according to its specifications. Even though the module lets you set wavelength values outside this range, performance of your module is guaranteed only within the wavelength range established by the module specifications.

Before Starting • Set the spectral unit from the OSICS Mainframe Main Setup menu (see section Selecting the Spectral Unit for all Installed Modules, p. 39). You cannot set the spectral unit of an individual module.

Procedure 1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set λ/f menu and press the knob.

The Set Wavelength menu appears and displays the emission wavelength or frequency.

Figure 67: ECL Unit Setup – Set Wavelength/Frequency menu

3. Modify the wavelength/frequency value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new wavelength value.

The Unit Setup menu is automatically displayed and the wavelength/frequency value of the module is now modified (according to the module limits).

9.4.2 Step-adjusting the Wavelength/Frequency Value of the Module

Subject The following procedure explains how to modify the wavelength/frequency of the module step by step using the increment defined in the Set Step sub-menu.You can define a valid step value within the range 00.00–50.00 nm.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Wavelength

ESCAPE

Wavelength :

λ = 1556.02 nm ENTER

Modify the value.

Page 107: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 107

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INThe starting wavelength is the value originally defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 106).

Procedure 1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Step menu and enter the Set Step sub-menu.

The Set Step sub-menu appears and displays the current value of the wavelength/frequency step in nm or GHz .

Figure 68: ECL Unit Setup – Set Step sub-menu

3. Modify the step value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new wavelength value.

The Unit Setup menu is automatically displayed.

4. To increase or decrease the wavelength/frequency value defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 106) step by step, expand the Step menu and put the cursor before the Step + or Step - sub-menus and press the knob to step-adjust the value.While the module is updating the wavelength/frequency, the message Moving… appears along with the new wavelength/frequency value after step-change.

5. Repeat step 4. until the wanted wavelength/frequency is reached.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Step (nm)

ESCAPE

Step:

01.00 nm ENTER

Modify the value.

Page 108: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

108 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

9.5 Activating Coherence Control

Subject The Coherence Control function induces broadening of the OSICS ECL tunable laser linewidth. This feature is especially useful in helping reduce interferometric noise.Coherence control consists in modulating the frequency with a pseudo-noise signal that widens the OSICS ECL optical signal linewidth. The specified linewidth for the OSICS ECL module is 150 kHz (typical value) as indicated in the module specifications without coherence control. This linewidth is brought to 100 MHz when the coherence control function is on.The OSICS ECL module features a built-in modulation source used to implement coherence control function on-board. You do not need to connect any external modulation device to the OSICS system.The following procedure explains how to activate the coherence control function to produce the linewidth broadening required.

Procedure 1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Coherence menu and press the knob.

The Coherence menu appears and the cursor points to the current setting.

Figure 69: ECL Unit Setup – Coherence Control menu

3. Turn the rotary knob to put the cursor before the wanted setting, and press it to validate your selection.

The new state of the coherence control function is implemented immediately. The Unit Setup menu is automatically displayed.

In the OSICS System Status screen, the message C is displayed in the OSICS module channel Info field if the coherence control function is currently enabled on this module.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Coherence

ESCAPE

Coherence :

OFF

ON

Select a new value.

Page 109: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 109

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN9.6 Activating the Auto-peak Find Function

Subject The Auto peak find function generates current pulses in order to stabilize the cavity operating conditions and stabilize the linewidth.When implemented, the Auto peak find function generates a small power fluctuation during a brief instant that may disrupt certain applications.If enabled, the Auto-peak Find is performed 1 second immediately after the laser output is enabled and then every 10 minutes afterward.

Procedure 1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Auto peak find menu and press the knob.

The Auto peak find menu appears and the cursor points to the current setting.

Figure 70: ECL Unit Setup – Auto peak find menu

3. Turn the rotary knob to put the cursor before the wanted setting, and press it to validate your selection.

The new state is implemented immediately. The Unit Setup menu is automatically displayed.

In the OSICS System Status screen, the message A is displayed in the OSICS module channel Info field if the Auto peak find function is currently enabled on this module.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Auto peak find

ESCAPE

Auto peak find :

OFF

ON

Select a new value.

Page 110: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

110 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

9.7 Configuring the OSICS ECL Signal ModulationPresentation In addition to the OSICS Mainframe built-in modulation source (see section Configuring

the OSICS Mainframe Modulation, p. 40), the OSICS ECL module holds a built-in digital modulation source.The OSICS ECL module supports the following modulation schemes:• Internal modulation:

• Digital (TTL) modulation using the OSICS ECL internal source• Digital (TTL) modulation using the OSICS Mainframe internal source

• External modulation:• Digital (TTL) modulation using an external source connected to the OSICS

Mainframe• Analog modulation using an external source connected to the OSICS ECL

The OUT 2 BNC connector can be used as an external modulation source for other devices.

Modulation Scheme

Figure 71: Mainframe and ECL Module Modulation Scheme

Page 111: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 111

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN9.7.1 Setting Up Digital (TTL) Modulation Using the ECL Module Internal Source

Subject Digital modulation capabilities of the ECL module is within the 50 Hz–1 MHz frequency range (see the ECL specifications).Internal digital modulation uses the ECL module's built-in TTL-level signal generator (see Figure 71, p. 110). The ECL module’s built-in TTL signal-generator can provide a TTL modulation signal in the range of 153 Hz–1 MHz, but for optimum modulation performance we recommend applying modulation frequencies within the specified range of the ECL module.The resulting modulated optical output signal follows a pattern as shown in the following figure.

Figure 72: TTL-Modulated Output Power Signal (ECL Module)

Procedure Step 1 – Set the module modulation source to INTERNAL:

1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Source sub-menu.

The Modulation source sub-menu appears and the cursor automatically points to the current modulation source setting.

Figure 73: ECL Unit Setup – Modulation Source (Internal)

3. Turn the rotary knob to put the cursor before INTERNAL and press it to validate your selection.

The ECL Unit Setup menu is automatically displayed and the modulation is set to the selected source.

Time Reference

Power Setting

2 x Power Setting

Optical OutputPower

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

MAINFRAME

Select a new value.

Page 112: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

112 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Step 2 – Set the frequency of the module INTERNAL modulation source:

1. In the ECL Unit Setup menu, expand the Modulation menu and enter the Set Frequency sub-menu.

The Set Frequency sub-menu appears and displays the current selected modulation frequency.

Figure 74: ECL Unit Setup – Set Frequency

2. Modify the frequency value as follows:

The ECL module’s built-in TTL signal-generator can provide a TTL modulation signal in the range of 153 Hz–1 MHz, but for optimum modulation performance we recommend applying modulation frequencies within the specified range of the ECL module.

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps 1. and 2. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new frequency value.

The frequency is updated to the "real frequency". This screen displays the actual modulation frequency produced by the OSICS ECL oscillator, which may slightly differ from the values you have set.

3. Press the knob to validate the displayed real frequency.

Step 3 – Set the modulation control:

1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Main

EXIT

2 3 4 5 6 7 8Unit Setup - Set Frequency

ESCAPE

Frequency :

f = 0016700 Hz

Modify the value.

ENTER

1

Page 113: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 113

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 75: ECL Unit Setup – Modulation Control (On)

3. Turn the rotary knob to put the cursor before the wanted modulation:

• ON for standard modulation.

• ON INVERTED for reverse modulation

4. Press the rotary knob to validate your selection.

Modulation is enabled. In the System Status screen, the letter M is displayed in the channel Info field of the corresponding module.

5. To stop the modulation: in the Control sub-menu, select OFF.

9.7.2 Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source

Subject This modulation setting uses the OSICS Mainframe’s built-in internal oscillator as a source of TTL-level modulated signals (see Figure 71, p. 110). The main purpose of this setting is to provide simultaneous and synchronous modulation to several modules installed in the OSICS Mainframe, or to an external device connected to the OSICS Mainframe Synchro output connector.The range of the frequency of the modulation provided by the OSICS Mainframe is defined to encompass the frequency requirements of all types of OSICS ECL modules. The frequency settings for the lower and upper bounds may therefore lie outside of the modulation capabilities of a particular module, in which case the module's own frequency range sets the modulation specifications.The OSICS Mainframe signal generator can provide a TTL signal with a repetition frequency in the range 123 Hz–1 MHz, but for optimum modulation performance, we recommend applying modulation frequencies within the specified range of the ECL module.The resulting signal-shape of the modulated output power is similar to the internally modulated pattern of the OSICS ECL, as shown in Figure 72, p. 111.

Procedure Step 1 – Set the OSICS Mainframe modulation source to INTERNAL:

1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Enter the Source sub-menu as follows:

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation control

ESCAPE

Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Page 114: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

114 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

a. Turn the rotary knob to put the cursor before the Modulation menu and press the knob to expand it.

b. Turn the rotary knob to put the cursor before the Source sub-menu and press the knob to enter it.

The Modulation Source sub-menu appears and displays the current selected modulation source.

Figure 76: Main Setup – Modulation Source (Internal)

3. Turn the rotary knob to put the cursor before INTERNAL and press it to validate your selection.

The Main Setup menu is automatically displayed and the modulation is set to the selected source.

Step 2 – Set the frequency of the OSICS Mainframe internal modulated-signal:

1. In the Main Setup menu, expand the Modulation menu and select the Set frequency sub-menu.

The Modulation frequency sub-menu appears and displays the current selected modulation frequency-value.

Figure 77: Main Setup – Modulation Frequency

2. Modify the frequency value as follows:

The internal TTL-signal generator can provide a TTL-modulated signal in the range 123 Hz to 1 MHz, but for optimum modulation performance we recommend applying modulation frequencies within the specified range of the ECL module.

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

EXTERNAL

Select a new value.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation frequency

ESCAPE

Frequency value :

f = 0014700 Hz

Modify the value.

ENTER

Page 115: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 115

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INc. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new frequency value.

The frequency is updated to the "real frequency" as shown in Figure 78, p. 115. This screen displays the current modulation frequency produced by the OSICS Mainframe oscillator, which may slightly differ from the values you have set.

Figure 78: Main Setup – Modulation Real Frequency

3. Press the knob to go back to the Main Setup menu.

Step 3 – Set the module modulation source to MAINFRAME:

1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Source sub-menu.

The Modulation source sub-menu appears and the cursor automatically points to the current modulation source setting.

Figure 79: ECL Unit Setup – Modulation Source (Mainframe)

3. Turn the rotary knob to put the cursor before MAINFRAME and press it to validate your selection.

The ECL Unit Setup menu is automatically displayed and the modulation is set to the selected source.

Step 4 – Set the modulation control:

1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation frequency

ESCAPE

The real frequency is :

14706 Hz

Please press the knob.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

MAINFRAME

Select a new value.

Page 116: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

116 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

2. Expand the Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Figure 80: ECL Unit Setup – Modulation Control (On)

3. Turn the rotary knob to put the cursor before the wanted modulation:

• ON for standard modulation.

• ON INVERTED for reverse modulation

4. Press the rotary knob to validate your selection.

Modulation is enabled. In the System Status screen, the letter M is displayed in the channel Info field of the corresponding module.

5. To stop the modulation: in the Control sub-menu, select OFF.

9.7.3 Setting Up Digital (TTL) Modulation Using an External Source

Subject External digital-modulation is achieved by applying a digital signal to the BNC Mod. In connector located on the OSICS Mainframe rear panel (see Figure 71, p. 110). This allows all the modules (or an external device connected to the OSICS Mainframe Synchro output connector) to be modulated in a synchronous way by means of an external source. The OSICS Mainframe distributes the external modulated-signal to all installed modules in order to synchronize all modules with a single modulated signal, amplitude, and frequency.For optimum stability of the modulated signals, we recommend applying a digital modulation in the specified range of the modules.

Before Starting Make sure you have a digital-signal generator.

Procedure Step 1 – Set the OSICS Mainframe modulation source to EXTERNAL:

1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Enter the Source sub-menu as follows:

a. Turn the rotary knob to put the cursor before the Modulation menu and press the knob to expand it.

b. Turn the rotary knob to put the cursor before the Source sub-menu and press the knob to enter it.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation control

ESCAPE

Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Page 117: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 117

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INThe Modulation source sub-menu appears and displays the current selected modulation source.

Figure 81: Main Setup – Modulation Source (External)

3. Turn the rotary knob to put the cursor before EXTERNAL and press it to validate your selection.

The Main Setup menu is automatically displayed and the modulation is set to the selected source.

Step 2 – Set the TTL-level signal generator:

1. On the OSICS Mainframe rear panel, connect your external digital signal-generator to the Mod. in BNC connector (see Figure 2, p. 18).

2. Apply the TTL signal not exceeding the maximum voltage of +5 V and repetition frequency between 50 Hz and 1 MHz.

The green LED located next to the Mod. In connector is lit to indicate that an external signal-generator is connected.

Step 3 – Set the module modulation source to MAINFRAME:

1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Source sub-menu.

The Modulation source sub-menu appears and the cursor automatically points to the current modulation source setting.

Figure 82: ECL Unit Setup – Modulation Source (Mainframe)

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

EXTERNAL

Select a new value.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

MAINFRAME

Select a new value.

Page 118: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

118 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

3. Turn the rotary knob to put the cursor before MAINFRAME and press it to validate your selection.

The ECL Unit Setup menu is automatically displayed and the modulation is set to the selected source.

Step 4 – Set the modulation control:

1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Figure 83: ECL Unit Setup – Modulation Control (On)

3. Turn the rotary knob to put the cursor before the wanted modulation:

• ON for standard modulation.

• ON INVERTED for reverse modulation

4. Press the rotary knob to validate your selection.

Modulation is enabled. In the System Status screen, the letter M is displayed in the channel Info field of the corresponding module.

5. To stop the modulation: in the Control sub-menu, select OFF.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation control

ESCAPE

Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Page 119: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 119

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

9.7.4 Setting Up Analog Modulation Using an External Source

Subject The SMB connector located on the front panel of the ECL module is directly connected to the laser chip and allows you to couple an external modulating-current to the system bias current (see Figure 71, p. 110).By modulating this external current, you can modulate the optical power output in a frequency range from 50 Hz to 50 MHz.

Figure 84: External Analog Modulation Setup

The relationship between the voltage applied at the SMB input and the current added to the diode is given by: Iin = Vin/Rinternal

where:• Iin is the added current• Vin is the voltage applied to the SMB input• Rinternal is the 50 Ω internal impedance of your ECL module

TTL modulation can also be applied directly on the module's front panel using the input SMB connector generally used for the analog modulation. The modulation is applied directly on the laser chip. In this case, simultaneous modulation for all the modules is not possible.Laser diode current level is available in section OSICS ECL Technical Specifications, p. 99.

Before Starting Make sure you have a 50 Ω coaxial cable to connect your HF-generator to the 50 Ω input SMB connector.

Procedure 1. Set the module modulation control to off as follows:

OSICS Module Laser Diode Source

Rinternal

50 Ω

Waveform Generator

Ibias

Polarization Current

Iin

Idiode

Vin

Coaxial cable 50Ω

Mod.SMB Connector

50Ω

Before applying an external analog-modulation, make sure the modulation control is set to off.

CAUTION

Page 120: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

120 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

a. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

b. Expand the Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Figure 85: ECL Unit Setup – Modulation Control (Off)

c. Turn the rotary knob to put the cursor before OFF and press it to validate your selection.

Modulation control is disabled and the ECL Unit Setup menu is displayed.

2. Connect your HF-generator to the 50 Ω input SMB connector using a 50 Ω coaxial cable.

3. Since the SMB input is directly connected to the laser diode chip, make sure that the input voltage is such that the following conditions are always met:

• The diode laser is forward-biased.

• The current flowing through the diode is lower than the maximum allowed current (Imax) indicated in the system acceptance test report.

The best way to fulfill the above conditions is to monitor the ECL module bias-current (Ibias) via the rated output voltage measured at the OUT 1 port located on the rear panel of the OSICS Mainframe (see Figure 2, p. 18), and to compute the total current using the following formula:

Idiode = Ibias + Vin/Rinternal

Where:

• Idiode is the total current through the diode• Ibias is the bias current• Vin is the voltage applied to the SMB input• Rinternal is the 50 Ω internal impedance of the ECL moduleYou must monitor Vin to ensure that the total current remains positive and lower than the maximum allowed current (Imax), as shown in the following figure:

0 < Idiode < Imax

The SMB input is directly connected to the laser diode chip via a 50 Ω resistor. It is essential to make sure that at all times the diode is forward-biased to avoid damage or possible destruction of the diode.

Main

EXIT

1 2 3 4 5 6 7 8

ESCAPE

Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Unit Setup - Modulation control

CAUTION

Page 121: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 121

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 86: Resulting Diode Current with External Analog Modulation Source

If the voltage applied to the SMB input is an AC signal with no DC component and with peak-to-peak amplitude is Vpp, the above condition becomes:Vpp < 2 x Rinternal (Imax - Ibias)Vpp < 2 x Rinternal (Ibias)

Ibias

tIbias

tVin

t

Idiode

Idiode Ibias

Diode modulated current

Current set on OSICSwithout modulation

Input modulated current

Imax

IA

IB

IA < Imax

IB > 0

0

Page 122: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

122 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

9.7.5 Using the ECL Module as a Modulation Source for External Devices (OUT 2 Output)

Subject The OSICS Mainframe rear panel (see Figure 2, p. 18) features eight channel ports, each dedicated to a specific channel-slot and identified by its channel number (CH1 to CH8).As they are numbered from left to right, channel ports do not face across their corresponding slots.Each channel can produce two types of output signals on BNC connectors:• OUT 1: analog signal for constant monitoring of the laser diode current. For more

details, see section Monitoring the OSICS ECL Module Parameters Using the Analog OUT 1 Output, p. 123.

• OUT 2: analog signal from the ECL module (see Figure 71, p. 110) that you can use as an external modulation source for other devices. The following procedure explains how to set the OUT 2 analog signal.

You can also use the OSICS Mainframe as a modulation source for external devices with the Synchro BNC output connector, as explained in section Using the Mainframe as a Modulation Source for External Devices (Synchro Output), p. 44.

Procedure 1. Set up the wanted digital modulation to the ECL module using one of the following methods:

• ECL module internal modulation source, as explained in section Setting Up Digital (TTL) Modulation Using the ECL Module Internal Source, p. 111.

• OSICS Mainframe internal modulation source, as explained in section Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source, p. 113.

• OSICS Mainframe external modulation source, as explained in section Setting Up Digital (TTL) Modulation Using an External Source, p. 116.

2. Connect the wanted appropriate device to the OUT 2 BNC connector corresponding to the channel slot number of the module.

The selected modulation frequency is applied to the connected device.

Page 123: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 123

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN9.8 Monitoring the OSICS ECL Module Parameters Using the Analog OUT 1 Output

Subject The OSICS Mainframe rear panel (see Figure 2, p. 18) features eight channel ports, each of them is dedicated to a specific channel-slot and identified by its channel number (CH1 to CH8).As they are numbered from left to right, channel ports do not face across their corresponding slots.Each channel can produce two types of output signals on BNC connectors:• OUT 2: analog signal from the ECL module that you can use as an external

modulation source for other devices. For more details, see section Using the ECL Module as a Modulation Source for External Devices (OUT 2 Output), p. 122.

• OUT 1: analog signal for constant monitoring of the ECL laser diode current (ILD) or output power (POUT).

The scale-factors for the OUT 1 signal are the following (FPGA v.3.00):

• Laser diode current (ILD): 10 V / 260.4 mA

• Optical output power (POUT): 10 V / 15 mW

The following procedure explains how to set the OUT 1 analog signal.

Before Starting Make sure you have an oscilloscope.

Procedure 1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Analog out menu and press the knob.

The Analog Out menu appears and displays the monitoring signal currently active on the OUT1 connector.

Figure 87: ECL Unit Setup – Analog Out setting

3. Turn the rotary knob to put the cursor before the parameter you want to monitor and press it to validate your selection.

The selected parameter is validated and the ECL Unit Setup menu is displayed.

4. To monitor the selected signal, connect your oscilloscope to the OUT 1 BNC connector corresponding to the channel slot number of the module.

Main

EXIT

2 3 4 5 6 7 8ECL Unit Setup - Analog Out

ESCAPE

Analog Out :

Select a new signal.

1

CURRENT

POWER

Page 124: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

124 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

9.9 Calibrating the ECL Module

9.9.1 Performing a Wavelength Reference Calibration

Subject The Wavelength Reference function allows the OSICS ECL to perform its own internal calibration based on a built-in mechanical reference.

Procedure 1. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Calibration menu and enter the Wavelength Reference sub-menu.

The Wavelength reference sub-menu appears prompting you to confirm whether you want to run the wavelength reference calibration.

3. Press the rotary knob to run the wavelength reference calibration.

The message Calibrating… appears. The referencing process may take up to two minutes. Once the calibration is complete, the ECL Unit Setup menu is displayed.

9.9.2 Performing a Wavelength/Frequency (λ/f) Calibration

Subject In the λ/f calibration sub-menu, first the system internally calibrates the ECL module wavelength or frequency (depending on the unit setting) with a mechanical reference (as explained in section Performing a Wavelength Reference Calibration, p. 124), then the system enters the two-point wavelength/frequency calibration sequence.The purpose of this function is to correct the discrepancies between the wavelength/frequency displayed by OSICS and the wavelength/frequency reading of your reference wavemeter. This calibration is made at two different wavelengths/frequencies using the ECL Unit Setup menu.The new calibration wavelengths/frequencies sets the new system calibration default.

Before Starting Make sure you have a wavemeter.

Procedure 1. Connect the optical output of the OSICS ECL module to the wavemeter.

2. Enable the module optical output (see section Enabling/Disabling the Light Source of All Modules, p. 37).

3. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

4. Expand the Calibration menu and enter the λ/f calibration sub-menu.

The system internally calibrates the ECL module wavelength or frequency with a mechanical reference (as explained in section Performing a Wavelength Reference Calibration, p. 124).The Wavelength/Frequency calibration sub-menu appears, showing the first wavelength/frequency value.

Page 125: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 125

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 88: ECL Unit Setup – λ/f calibration, first wavelength

This is the first step of the two-point wavelength/frequency calibration method.The first wavelength/frequency is the first factory calibration wavelength/frequency. The value you enter in this field is the wavemeter reading.

5. Enter the value measured at the wavemeter as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the value.

The message Moving... appears while the first wavelength/frequency is being set.

The Wavelength/Frequency calibration window appears, showing the second wavelength value.

Figure 89: ECL Unit Setup – λ/f calibration, second wavelength

This is the second step of the two-point wavelength/frequency calibration method.

6. Measure the second wavelength/frequency using the wavemeter.

7. Enter the wavelength/frequency reading in this field and select ENTER.

A confirmation message appears.

8. Press the knob to confirm.

While updating the emission wavelength/frequency, the system displays the Moving… message until the wavelength/frequency is set.

Main

EXIT

2 3 4 5 6 7 8ECL Unit Setup - Wavelength calibration

ESCAPE

First wavelength value :

Correct the value.

ENTER

1

λ = 1540.461 nm

Main

EXIT

2 3 4 5 6 7 8ECL Unit Setup - Wavelength calibration

ESCAPE

Second wavelength value :

Correct the value.

ENTER

1

λ = 1589.694 nm

Page 126: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

126 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Once both calibration points are entered, the new calibration values are implemented immediately and stored in the system's flash-memory, and the current wavelength/frequency setting of the ECL module is updated to match the new calibration data. The new calibration wavelengths/frequencies sets the new system calibration default.

The ECL Unit Setup menu is displayed.

9.9.3 Performing a Power Calibration

Subject This section explains how to perform a two-point power calibration sequence using the Power calibration function. The purpose of this calibration is to correct the discrepancies between the power displayed by the OSICS ECL module and the power measured at your reference power meter.Also, calibration is made at two different wavelengths to account for small wavelength-dependence of the power meter.The new calibration data sets the new system calibration default.

Before Starting Make sure you have a power meter.

Procedure 1. Set the ECL output power to 1 mW.

2. Connect the optical output of the OSICS ECL module to a power meter.

3. Enable the module optical output (see section Enabling/Disabling the Light Source of the Module, p. 103).

4. Access the ECL Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

5. Expand the Calibration menu and enter the Power calibration sub-menu.

The message Moving... appears, and the Power calibration sub-menu appears, prompting you to set the power at the first wavelength value.

Figure 90: ECL Unit Setup – Power calibration, power value at first wavelength

6. Measure the power level at the first wavelength on the power meter.

A valid power value can range from 0.6 mW to 1.2 mW.

7. Enter the power value measured by the power meter and select ENTER.

The message Moving... appears while the power value is set. The Power calibration window appears, showing the second window. The ECL tunable laser

Main

EXIT

2 3 4 5 6 7 8ECL Unit Setup - Power calibration

ESCAPE

First power value :

Correct the value.

ENTER

1

P = +09.160 dBm

First wavelength : 1540.000 nm.

Page 127: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 127

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INmoves to a second wavelength in the tuning range and displays the power level corresponding to the calibration data.

Figure 91: ECL Unit Setup – Power calibration, power value at second wavelength

8. Measure the power level at the second wavelength on the power meter.

A valid power value can range from 0.6 mW to 1.2 mW.

9. Enter the power value measured by the power meter in this field and select ENTER.

A confirmation message appears.

10. Press the knob to confirm.

The message Moving... appears while the power value is set. The power calibration is now complete and the values are stored in the module's non-volatile memory.

The ECL Unit Setup menu is displayed.

The new calibration data sets the new system calibration default.

9.9.4 Asking for a Factory Recalibration

Subject To externally verify the performance of the OSICS, refer to the procedure described in section Performing a Wavelength Reference Calibration, p. 124 and section Performing a Power Calibration, p. 126. If the performance does not meet warranted specifications, a complete calibration of the instrument may be required.We recommend performing factory calibration of OSICS ECL modules over one-year intervals.

Procedure Contact the Yenista Optics customer support service (see section Contact Information, p. 4) for a complete calibration, as some calibration steps require the adjustment of internal components including the laser head.

Main

EXIT

2 3 4 5 6 7 8ECL Unit Setup - Power calibration

ESCAPE

Second power value :

Correct the value.

ENTER

1

P = +09.160 dBm

Second wavelength : 1590.000 nm.

Page 128: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

128 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

9.10 Testing the Performance of the ECL Module

Use the procedures described in this section to validate warranted specifications of the OSICS ECL module, as needed.During the performance verification, the settings for wavelength and power may not agree with the measured values. If this occurs, you may calibrate wavelength and power settings as explained in section Performing a Wavelength/Frequency (λ/f) Calibration, p. 124 and section Performing a Power Calibration, p. 126.Notwithstanding performance tests described in this section, we recommend performing calibration of OSICS ECL modules over one-year intervals (see section Asking for a Factory Recalibration, p. 127).

9.10.1 Preparing Performance Tests

Before Starting Make sure you have the following required equipment for conducting performance tests on OSICS ECL modules:

The various types as well as the number of fiber-optic connectors needed may vary depending on the type of equipment used and the distance between the test equipment and the OSICS ECL modular optical source under test.

Procedure 1. In the OSICS Mainframe, install the ECL modules you want to test.

2. Turn-on the OSICS multifunction platform (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

3. Let the OSICS multifunction platform and other test equipments run continuously for 2 hours at room temperature (within the specified operating temperature range) to achieve proper equipment warm-up.

Equipment Minimum RequirementOptical power meter Wavelength range: 1520 nm to 1600 nmOptical spectrum analyzer (OSA) Wavelength range: 1520 nm to 1600 nmWavemeter Wavelength range: 1520 nm to 1600 nmJumper (x2) FC/APC to FC/PC or to FC/APCIn-line optical adapter FC to FC

• Disable the laser output before connecting or disconnecting a fiber optic cable on the module (see section Enabling/Disabling the Light Source of All Modules, p. 37).

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 129: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

OSICS User Manual 129

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

9.10.2 Measuring the Absolute Wavelength Accuracy

Subject This test procedure enables you to verify that the actual measured wavelength stays within acceptable deviation range from the wavelength setting.

Before Starting • Make sure you have a wavemeter with ±1 pm resolution and a NIST-traceable calibration (see section Preparing Performance Tests, p. 128) and jumper.

Procedure 1. Connect the OSICS ECL module to the wavemeter.

2. Set the OSICS ECL output power to 0 dBm.

3. Set the wavelength of the OSICS ECL tunable laser to the minimum value of the wavelength range. The minimum wavelength depends on the ECL model you are using (see section Specifying the Optical Emission Wavelength/Frequency, p. 106).

4. Set the wavelength step to +5 nm.

5. Enable the laser output by pressing the Enable button on the module front panel.6. Compute the wavelength deviation of the ECL module and check that it is not greater

than ±0.2 nm compared with the wavelength reading of the wavemeter.

The absolute wavelength accuracy is the deviation between the wavelength setting and the wavelength reading provided by the wavemeter at each measurement step.

7. Increment the ECL wavelength setting one wavelength step (+5 nm) and repeat measurement step 6. to check that the wavelength deviation is not greater than ±0.2 nm.

8. Stop when you reach the ECL maximum wavelength limit.

If the minimum and maximum wavelength limits do not meet the specifications, you can calibrate these values as explained in section Performing a Wavelength/Frequency (λ/f) Calibration, p. 124.

9.10.3 Checking the Wavelength Range and Output Power

Subject This procedure enables you to verify that the ECL module can achieve the rated output power at both ends of the ECL wavelength range.

Before Starting Make sure you have a powermeter with a NIST-traceable calibration (see section Preparing Performance Tests, p. 128) and jumper.

Procedure 1. Disable the ECL laser output.

2. Connect the ECL laser output to the optical power meter.

3. From the ECL Unit Setup menu, set the wavelength of the ECL tunable laser to the minimum wavelength that corresponds to the type and power option of the ECL module you are using (see OSICS ECL Technical Specifications).

4. Enable the ECL module laser output by pressing the Enable button.

Page 130: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS ECL Module

130 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

5. Set the output power value to the rated output power, as given in section OSICS ECL Technical Specifications and check that the ECL module can deliver the desired output power at the given wavelength.

6. Set the wavelength of the ECL tunable laser to the maximum wavelength that corresponds to the type and power option of the ECL module you are using (see section OSICS ECL Technical Specifications) and check that the ECL module can deliver the desired output power at the given wavelength.

7. Do one of the following, depending on the measured values:

• If the ECL power settings show small discrepancies compared to the measured powers, perform a two-point manual calibration of the ECL output power (see section Performing a Power Calibration, p. 126).

• If either of the rated output power at the wavelength range ends cannot be achieved, the ECL module requires servicing. Please contact your Yenista Optics customer support service (see section Contact Information, p. 4) immediately.

9.10.4 Validating the Side Mode Suppression

Subject This procedure enables you to verify the suppression of any wave emissions that are more than 1 nm away from the selected wavelength.

Before Starting Make sure you have an appropriate optical spectrum analyzer (see section Preparing Performance Tests, p. 128) and jumper.

Procedure 1. Disable the ECL laser output.

2. Connect the ECL laser output to the optical spectrum analyzer.

3. Set the wavelength of the ECL module to a value at mid-range of the wavelength range as set in the ECL module specifications (see OSICS ECL Technical Specifications).

4. Set the ECL laser output power to 0 dBm.

5. Configure the optical spectrum analyzer as follows:

• Center wavelength: same as ECL laser wavelength setting• Span: 20 nm• Resolution: 0.1 nm• Reference level: 0 dBm• Level scale: 10 dB/div• Scan: Repeat

6. Enable the ECL module laser output by pressing the Enable button.

7. Check that, at wavelength deviations larger than 1 nm away from the center wavelength of the laser peak, the power level of any residual emission is lower than-45 dB down from the peak level.

Page 131: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 131

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN10. Using the OSICS SLD Module

This section explains how to operate the OSICS SLD via the SLD Unit Setup menu. For more details on how to install and initialize the module into the OSICS Mainframe, see section Installing and Connecting the OSICS Multifunction Platform, p. 21.

10.1 OSICS SLD Presentation

The OSICS SLD is a broadband light source rack module designed for fiber-optic component and system testing, based on superluminescent light emitting diode.

10.1.1 OSICS SLD Technical Specifications

OSICS SLD 1300 OSICS SLD 1550

Central Wavelength*1*2

*1: Typical.*2: ±20 nm.

1320 nm 1540 nm

Spectral Width*3

*3: Guaranteed at output power = 5 mW.

-3 dB (FWHM) 50 nm-10 dB 80 nm-20 dB 100 nm-30 dB 125 nm

Total Output PowerHigh +10 mWLow +5 mW

Power Stability Over Temperature Range ±0.1 dBPower Stability at Constant Temperature ±0.01 dBOptical Fiber SMFOptical Interface FC/APC connector

Page 132: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SLD Module

132 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

10.1.2 OSICS SLD Module Overview

Figure 92: SLD Module – Front and Rear Panels

Front Panel On LED

This LED indicates that the module is turned on (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

Enable Button

This button enables you to turn on/off the optical output of the module. The green LED of the button is lit if the optical output is enabled, and the power value is displayed in the System Status screen (see section System Status Screen Description, p. 30).

Optical Output Connector

The standard output connector is a FC/APC type connector.For more details on the appropriate fiber type, see section OSICS SLD Technical Specifications, p. 131.

The label located on the optical output connector cap indicates the location of the laser source.

Openingfor air output

Top Notch

Bottom Notch

96-pin VMEConnector

On LED

Enable Button

Optical OutputConnector

Retaining Screw

Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.WARNING

Page 133: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SLD Module

OSICS User Manual 133

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INRetaining Screw

The captive screw enables you to lock/unlock the module inserted into the slot of the OSICS Mainframe.

Rear Panel Notches

The top and bottom notches enable you to slide the module into the OSICS Mainframe slot (see section Installing an OSICS Module Into the OSICS Mainframe, p. 23).

96-pin VME Connector

The rear panel of the OSICS modules interfaces with the OSICS Mainframe backplane by means of the 96-pin VME connector.

Opening

The opening ensures proper flow of cooled air drawn from the OSICS Mainframe chassis.

Side Panel The labels affixed on the side panel of the OSICS SLD are described in section OSICS Information and Warning Markings, p. 19.

The OSICS Mainframe does not support "hot-swap" of modules; for more details, see section Handling OSICS Modules Into the OSICS Mainframe, p. 23).

CAUTION

Page 134: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SLD Module

134 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

10.2 Enabling/Disabling the Light Source of the Module

Subject You can enable/disable the light source in two different ways:• From the OSICS Mainframe Main Setup menu (see section Enabling/Disabling the

Light Source of All Modules, p. 37), you can simultaneously enable or disable the light source in all the modules installed in the OSICS Mainframe. This function overrides the enable/disable state individually set in all modules.

• From the SLD front panel, you can individually enable or disable the light source of the module by using the Enable button as explained in the following procedure. This action does not affect the enable/disable state of other modules installed in the OSICS Mainframe.

For laser safety reasons, the light source in modules are disabled after OSICS initialization.

Before Starting Connect an optical device to the OSICS module (see section Connecting Optical Devices to OSICS Modules, p. 26).

Procedures Enabling the Light Source of the Module

• On the SLD module front panel press the Enable button.

The green LED on the button is lit.In the System Status screen, the output power is displayed in the P field (in dBm or mW depending on the selected power unit) for the corresponding module.

Disabling the Light Source of the Module

• On the SLD module front panel press the Enable button.

The green LED on the button is turned off.In the System Status screen, the message Disabled is displayed for the corresponding module.

• Do not enable the laser while there is no fiber connected to the optical output connector.

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 135: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SLD Module

OSICS User Manual 135

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN10.3 Setting the Optical Output Power of the SLD Module

Subject You can set the optical output power in two different ways:• From the OSICS Mainframe Main Setup menu, you can simultaneously set the

optical power of all modules to the same value: see section Setting the Optical Output Power of All Installed Modules, p. 38.

• From the SLD Unit Setup menu, you can set the output power of the module individually as explained in the following procedure.

The power value is software-limited to stay within the specified range.

Before Starting • Set the optical power unit from the OSICS Mainframe Main Setup menu (see section Selecting the Power Unit for All Installed Modules, p. 38). You cannot set the power unit of an individual module.

Procedure 1. Access the SLD Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set Power menu and press the knob.

The Set Power menu appears and displays the current power setting.

Figure 93: SLD Unit Setup – Set Power menu

3. Turn the rotary knob to put the cursor before the wanted setting, and press it to validate your selection.

The Unit Setup menu is automatically displayed and the optical power value of the module is now modified (according to the module power limits).

Main

EXIT

1 2 3 4 5 6 7 8SLD Unit Setup - Set Power

ESCAPE

Power :

+07.00 dBm (Low)+10.00 dBm (High)

Page 136: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SLD Module

136 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Page 137: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 137

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN11. Using the OSICS SWT Module

This section explains how to operate the OSICS SWT via the SWT Unit Setup menu. For more details on how to install and initialize the module into the OSICS Mainframe, see section Installing and Connecting the OSICS Multifunction Platform, p. 21.

11.1 OSICS SWT Presentation

Features The OSICS SWT is a full suite of fiber optic switch and shutter modules based on optical prism technology.• The 1x1 configuration is a shutter. It allows to block/pass an optical signal with high

isolation.

• The 2x(1x1) configuration is a 2x shutter. It features two 1x1 configurations.The first one is called Shutter A-B, the second one Shutter 1-2. They can be operated simultaneously as well as individually.

• The 2x2 switch configuration allows to cross (or not) two input and two output channels.

• The 1x2 and 1x4 switch configuration features one common input/output channel (Com), and 2 or 4 output/input channels (Ch1 and Ch2 or Ch1, Ch2, Ch3 and Ch4) to be used sequentially.

The OSICS SWT module is fully bi-directional: you can use the module as 1 input and 2 or 4 outputs (see Figure 94, p. 137), or 1 output and 2 or 4 inputs (see Figure 95, p. 137), depending on the needs of your measurement setup.• In a standard configuration, it allows you to direct a laser signal from the common

channel Com to any output channel Ch1 to Chn, as shown in the following figure:

Figure 94: SWT Standard Configuration

• In a reverse configuration, you can direct any signal one after another from Ch1 to Chn to the common channel Com as illustrated in the following figure:

Figure 95: SWT Reverse Configuration

The SWT optical-switch module can be combined with a variety of modular sources such as tunable external cavity laser (T100), distributed feedback laser diode (DFB) and tunable laser source (TLS), within the same OSICS multifunction platform.

Output

Output

OSICSSWT

InputCom

Ch1

Chn

Input

Input

Ch1

Chn

OSICSSWT

ComOutput

Page 138: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT Module

138 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Examples of SWT Implementations

Measurement of Two Laser Sources with a Wavemeter

If you want two laser sources to share the same measurement device (such as a wavemeter), you can connect two laser sources to Ch1 and Ch2, which can then be sequentially measured during a specific time frame (if you define one):

Figure 96: SWT Implementation Example – Two Laser Sources and a Wavemeter

Sequential Testing of Two or Four Devices

To sequentially test two or four devices, you can use two OSICS SWT modules. A laser source and the input of the devices-under-test are connected to the first module. The output of the devices-under-test are connected to the second module, which sequentially directs the signal to a unique measurement device:

Figure 97: SWT Implementation Example – Sequential Testing of Two or Four Devices

Testing of an Optical Amplifier

To test an optical amplifier, you need to connect both the non-amplified and amplified signals to a measurement device. In this setup two OSICS SWT modules are needed.

Figure 98: SWT Implementation Example – Testing of an Optical Amplifier

Laser Source

Laser Source

Ch1

Ch2

SWT 1x2or

SWT 1x4

ComWavemeter

Laser SourceCh1SWT

1x2

TestedDevice

TestedDevice

SWT1x2

Measurement Device

Ch2

Com ComCh1

Ch2

Laser SourceCh1SWT

1x2

EDFA SWT1x2

Yenista Optics OSA400 with

EDFA Characterization

FunctionCh2

Com ComCh1

Ch2

Amplified signal

Source signal

Page 139: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT Module

OSICS User Manual 139

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN11.1.1 OSICS SWT Technical Specifications

SMF*1

*1: Specifications apply for wavelengths not equal to any water absorption line

PMFAvailable Configurations 1x1; 2x(1x1); 1x2; 1x4; 2x2

Spectral Range (nm) 1260–1630 On PM15: 1480–1630On PM13: 1240–1510

Insertion Loss*2*3

*2: Typical values including connectors. Maximum insertion loss is 1.4 dB for SMF and 2.0 dB for PMF.*3: On the whole wavelength range.

≤ 1 dB ≤ 1.5 dB

Polarization Dependent Loss*2 PDL ≤ 0.1 dB PER ≥ 20 dB

Repeatability*3 ≤ 0.005 dB ≤ 0.01 dB

Return Loss*2 > 55 dB

On/Off ratio(1x1 Shutter Only) ≥ 80 dB

On/Off ratio(2x(1x1) Shutter Only) ≥ 65 dB

Crosstalk*3 ≥ 55 dB

Switching Time 30 ms typ.Guaranteed Lifetime(number of cycle) > 108

Input/Output Fiber Type SMF-28 fiber PM15 or PM13

Connector Type*4

*4: For PMF, slow axis is aligned to the connector key.

FC/APC narrow key

Dimensions W x H x D 35 x 128 x 230 mm (1.4 x 5.0 x 9.0 in)Weight 1 kg (2.21 lb)Temperature Range +15 to +35 °C (+59 to +95 °F)

Page 140: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT Module

140 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

11.1.2 OSICS SWT Module Overview

Figure 99: SWT Module – Front and Rear Panels

Front Panel Com Connector (on SWT 1x1, 1x2 and 1x4)

Fiber optic connector. It is a standard input/output FC/APC type connector. For more details on the appropriate fiber type, see section OSICS SWT Technical Specifications, p. 139.

Chn Connector and LEDs

Fiber optic connector. It is a standard input/output FC/APC type connector. For more details on the appropriate fiber type, see section OSICS SWT Technical Specifications, p. 139.

Openingfor air output

Top Notch

Bottom Notch

96-pin VMEConnector

Com Connector

Retaining Screw

Chn Connector

LED

Page 141: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT Module

OSICS User Manual 141

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INThe green LEDs indicates which channel (SWT 1x2, SWT1x4) or output mode is currently active. If the channel is enabled, the LED is lit and the corresponding information is displayed in the System Status screen (see section System Status Screen Description, p. 30).

The label located on the front panel indicates the locations of the laser output.

Retaining Screw

The captive screw enables you to lock/unlock the module inserted into the slot of the OSICS Mainframe.

Rear Panel Notches

The top and bottom notches enable you to slide the module into the OSICS Mainframe slot (see section Installing an OSICS Module Into the OSICS Mainframe, p. 23).

96-pin VME Connector

The rear panel of the OSICS modules interfaces with the OSICS Mainframe backplane by means of the 96-pin VME connector.

Opening

The opening ensures proper flow of cooled air drawn from the OSICS Mainframe chassis.

Side Panel The labels affixed on the side panel of the OSICS SWT are described in section OSICS Information and Warning Markings, p. 19.

On OSICS SWT, do not connect several laser sources at the same time on the Com and Chn channels, or you may severely damage your optical equipment.

Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

The OSICS Mainframe does not support "hot-swap" of modules; for more details, see section Handling OSICS Modules Into the OSICS Mainframe, p. 23).

CAUTION

WARNING

CAUTION

Page 142: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT Module

142 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

11.2 Selecting the OSICS SWT Input/Output

Subject This section explains how to select the channel through which the signal will be directed.You can only activate one output channel at a time.For SWT 1x1, this section explains how to open or shut the switch.

Procedure 1. Access the SWT Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Select Channel (for SWT 1x2 or 1x4) or Select Output Mode (for SWT 1x1, 2x2 or 2x(1x1)) menu and press the knob.

The Select Channel (for SWT 1x2 or 1x4) or Select Output Mode (for SWT 1x1, 2x2 or 2x(1x1)) menu appears and indicates the currently active SWT channel or output mode with a cross next to the selected channel.

Figure 100: SWT Unit Setup – Channel Selection menu (SWT 1x4 Example Screen)

3. Turn the rotary knob to put the cursor before the channel or output mode you want to activate and press it to validate your selection.

The corresponding channel is immediately activated and a cross appears next to the selected channel. On the SWT front panel, the LED corresponding to the selected channel or output mode is lit.On SWT 2x(1x1), the LED is lit if the switch is shut.

4. Press one of the following control buttons:

• ESCAPE: to go back to the SWT Unit Setup menu.

• EXIT: to display the System Status screen.

In the System Status screen, the selected channel number or output mode is displayed in the corresponding SWT channel Info field.

Main

EXIT

1 2 3 4 5 6 7 8SWT Unit Setup - Select Channel

ESCAPE

Select Channel:

Ch1Ch2Ch3 xCh4

Page 143: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT Module

OSICS User Manual 143

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN11.3 Configuring Sequential Switching

You can configure sequential switching for each SWT module individually from the Unit Setup menu.The sequence activates each channel one after the other from channel 1 to channel 2 (for SWT 1x2), from channel 1 to channel 4 (for SWT 1x4), from Bar to Cross (for SWT 2x2) or from Shut to Open (for SWT 2x(1x1)) according to the configured timeframe and number of scans. You cannot select a different order.The sequence can be executed once or continuously.

11.3.1 Selecting the Number of Scans

Subject You can select from 1 to 1000 scans in continuous.

Procedure 1. Access the SWT Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Sequence Switching menu and enter the Number of Scans sub-menu.

The sub-menu appears and displays the current configured number of scans.

Figure 101: SWT Unit Setup – Sequence Switching - Number of Scans

3. Modify the number of scan as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new step value.

The number of scans is set and the Unit Setup menu is automatically displayed.

Main

EXIT

1 2 3 4 5 6 7 8SWT Unit Setup - Number of Scans

ESCAPE

Number of scans: 0001 ENTER

Modify the value.

Page 144: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT Module

144 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

11.3.2 Setting the Timeframe

Subject You can set a different period of time for each channel or output mode individually, or the same period of time for all; from 0.1 s to 48 h.

Procedure 1. Access the SWT Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Sequence Switching menu and enter the Set Delays sub-menu.

The sub-menu appears and displays the available channels (for SWT 1x2 or 1x4) or output modes (for SWT 2x2 or 2x(1x1)).

3. Do one of the following:

• If you want to set the same period of time for all channels or output modes, turn the rotary knob to put the cursor before All and press the knob to validate your selection.

• If you want to set a different period of time for an individual channel or output mode, turn the rotary knob to put the cursor before the channel or mode you want to configure and press the knob to validate your selection.

The time setting sub-menu appears

Figure 102: SWT Unit Setup – Sequence Switching - Set delay (SWT 1x4 Example Screen)

4. Modify the time value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Press the ENTER control button to validate the time set.

The timeframe is set for the selected channel or mode. The channel or mode selection menu is displayed.

5. Perform steps 3. and 4. on each channel or mode for which you want to set a particular period of time.

6. Press the ESCAPE control button to validate your configuration and go back to the Unit Setup screen.

Main

ENTER

1 2 3 4 5 6 7 8SWT Unit Setup - Sequence Switching

ESCAPE

Ch 3 - set delay :

00h 00mn 00.1s

Press the knob to change the digit.

Page 145: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT Module

OSICS User Manual 145

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN11.3.3 Launching/Stopping a Sequence

Subject The Control sub-menu enables you to launch the sequence once or continuously.

Before Starting Make sure the sequence number of scans (see section Selecting the Number of Scans, p. 143) and time (see section Setting the Timeframe, p. 144) are set.

Procedure 1. Access the SWT Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Sequence Switching menu and enter the Control sub-menu.

Figure 103: SWT Unit Setup – Sequence Switching - Control

3. Turn the rotary knob to put the cursor before the wanted action:

• STOP: the SWT module executes the sequence once.

• CONTINUOUS: the SWT module executes the sequence continuously. The sequence will only stop if you select the STOP action.

4. Press the knob to start the sequence switching.

The sequence starts. In the System Status screen, an s and the currently activated channel or mode are displayed in the corresponding SWT module Info field.

5. To stop the sequence if you have selected CONTINUOUS: in the Control sub-menu, turn the rotary knob to put the cursor before STOP and press the knob to confirm your selection.

Main 1 2 3 4 5 6 7 8SWT Unit Setup - Control

Switching mode:

STOPSTARTCONTINUOUS

Select a new value.

EXITESCAPE

Page 146: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT Module

146 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

11.4 Testing the Performance of the SWT Module

Use the procedures described in this section to validate warranted specifications of the OSICS SWT module, as needed.

11.4.1 Preparing Performance Tests

Before Starting Make sure you have the following required equipment for conducting performance tests on OSICS SWT modules:

The various types as well as the number of fiber-optic connectors needed may vary depending on the type of equipment used and the distance between the test equipment and the OSICS SWT modular optical source under test.

Procedure 1. In the OSICS Mainframe, install the SWT modules you want to test.

2. Turn-on the OSICS multifunction platform (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

3. Let the OSICS multifunction platform and other test equipments run continuously for 2 hours at room temperature (within the specified operating temperature range) to achieve proper equipment warm-up.

Equipment Minimum RequirementOptical power meter (x2) Wavelength range: 1250 nm to 1650 nm

Resolution: < 0.001 dBDynamic range: > 70 dB

Laser source Wavelength range: 1250 to 1650 nmFiber optic coupler FC/APC connectors

• Disable the laser output before connecting or disconnecting a fiber optic cable on the module (see section Enabling/Disabling the Light Source of All Modules, p. 37).

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 147: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT Module

OSICS User Manual 147

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

11.4.2 Verifying the Repeatability

Subject This procedure verifies that the SWT module repeatability is within specification range over 100 measurements for each channel.The following figure illustrates the appropriate setup to perform the repeatability test:

Figure 104: SWT Performance Test – Repeatability

Before Starting Make sure you have the appropriate power-meter (see section Preparing Performance Tests, p. 146).

Procedure 1. Disable all laser source outputs and other modules in the OSICS Mainframe.

2. Connect the measurement equipments as indicated on Figure 104, p. 147.

3. If you are using an OSICS T100 or OSICS DFB as the external source, enable the laser output.

4. Enable the output channel under test (in our example Ch1) and record the power output value P1 and P2 provided respectively by OPM1 and OPM2 after the power reading has stabilized.

5. Select dBm as the power-meter measurement unit and hold the range on OPM2 for better accuracy.

6. Turn off and on Ch1 for 100 times, and record each time the power reading P1 and P2 (in dBm) on both power-meters OPM1 and OPM2.

7. Subtract OPM2 reading from OPM1 reading (P1-P2) for each of the 100 measurements and record the result.

8. Find the maximum (|P1-P2|max) and minimum (|P1-P2|min) values among the |P1-P2| records calculated in step 6. above.

9. Subtract the maximum and minimum values to get the module repeatability.

10. Verify that the calculated repeatability value is within the specification range of 0.02 dB.

11. Perform the same operations for the second OSICS SWT module channel.

Power-meterOPM1

CouplerLaser sourceIn

Out1

Out2OSICS

SWT 1xnPower-meter

OPM2

Ch1

Chn

Com

Page 148: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT Module

148 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

11.4.3 Verifying the Crosstalk

Subject The following figure illustrates the appropriate setup to perform the crosstalk test:

Figure 105: SWT Performance Test – Crosstalk

Before Starting Make sure you have the appropriate power-meter (see section Preparing Performance Tests, p. 146).

Procedure 1. Disable all laser source outputs and other modules in the OSICS Mainframe.

2. Connect the measurement equipments as indicated on Figure 105, p. 148.

3. Select dBm as the power-meter unit.

4. Select the channel (Ch1 in our example) that you want to measure.

5. Record the power value provided by the power-meter.

6. Switch the laser signal to another channel, while the power-meter is still measuring Ch1.

7. Record the power on the power-meter.

8. Subtract the second power from the first measure to get the crosstalk.

9. Verify that the calculated value is over the specified value of 50 dB.

10. Perform the same operation for the Chn SWT channel.

Laser sourceCom OSICS

SWT 1xnPower-meter

Ch1

Chn

Page 149: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 149

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN12. Using the OSICS SWT APC Module

This section explains how to operate the OSICS SWT APC via the SWT APC Unit Setup menu. For more details on how to install and initialize the module into the OSICS Mainframe, see section Installing and Connecting the OSICS Multifunction Platform, p. 21.

12.1 OSICS SWT APC Presentation

Features The OSICS SWT APC is an optical-switch module that comes in a 1x2 or 1x4 configuration: one common input/output channel and two or four output/input channels (Ch1 and Ch2 or Ch1, Ch2, Ch3 and Ch4) to be used sequentially. You can use the OSICS SWT APC module as a bi-directional switch, or as a tunable laser, as explained below.The OSICS SWT APC module can only be used with OSICS Mainframe v. 3.01 and higher versions.

OSICS SWT APC Used in Switch Mode

The OSICS SWT APC module is fully bi-directional: you can use the module as 1 input and 2 or 4 outputs (see Figure 106, p. 149), or 1 output and 2 or 4 inputs (see Figure 107, p. 149), depending on the needs of your measurement setup.• In a standard configuration, it allows you to direct a laser signal from the common

channel Com to any output channel Ch1 to Chn, as shown in the following figure:

Figure 106: SWT APC Standard Configuration

• In a reverse configuration, you can direct any signal one after another from Ch1 to Chn to the common channel Com as illustrated in the following figure:

Figure 107: SWT APC Reverse Configuration

The SWT APC optical-switch module can be combined with a variety of modular sources such as tunable external cavity laser (T100), distributed feedback laser diode (DFB) and tunable laser source (TLS), within the same OSICS multifunction platform.

Output

Output

OSICSSWT APC

InputCom

Ch1

Chn

Input

Input

Ch1

Chn

OSICSSWT APC

ComOutput

Page 150: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

150 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

OSICS SWT APC Used in Full-Band Mode

Using T100 and ECL modules, you can use the SWT APC as a tunable laser: from the OSICS SWT APC setup menu you can drive in power and wavelength from 1 to 4 T100 and ECL modules as illustrated in the following figure.A monitoring photodiode checks the optical output power of the module.

Figure 108: SWT APC in T100 Full-Band Configuration

The OSICS SWT APC module can drive the T100 and ECL modules only if they are located in the same OSICS Mainframe.

Examples of SWT APC Implementations

Measurement of Two Laser Sources with a Wavemeter

If you want two laser sources to share the same measurement device (such as a wavemeter), you can connect two laser sources to Ch1 and Ch2, which can then be sequentially measured during a specific time frame (if you define one):

Figure 109: SWT APC Implementation Example – Two Laser Sources and a Wavemeter

Sequential Testing of Two or Four Devices

To sequentially test two or four devices, you can use two OSICS SWT APC modules. A laser source and the input of the devices-under-test are connected to the first module. The output of the devices-under-test are connected to the second module, which sequentially directs the signal to a unique measurement device:

Figure 110: SWT APC Implementation Example – Sequential Testing of Two or Four Devices

Testing of an Optical Amplifier

To test an optical amplifier, you need to connect both the non-amplified and amplified signals to a measurement device. In this setup two OSICS SWT APC modules are needed.

T100 1

T100 n

Ch1

Chn

OSICSSWT APC

ComOutput

Laser Source

Laser Source

Ch1

Ch2

SWT APC 1x2or

SWT APC 1x4

ComWavemeter

Laser SourceCh1SWT APC

1x2

TestedDevice

TestedDevice

SWT APC1x2

Measurement Device

Ch2

Com ComCh1

Ch2

Page 151: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

OSICS User Manual 151

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 111: SWT APC Implementation Example – Testing of an Optical Amplifier

Testing of an Optical Device over a Wide Spectrum

To test an optical device over a wide spectrum, you must employ several tunable lasers with distinct wavelength coverage:

Figure 112: SWT APC Implementation Example – Testing a DUT over a Wide Spectrum

Laser SourceCh1SWT APC

1x2

EDFA SWT APC1x2

Yenista Optics OSA400 with

EDFA Characterization

FunctionCh2

Com ComCh1

Ch2

Amplified signal

Source signal

T100 1

T100 n

Ch1

Chn

OSICSSWT APC

ComDUT

Measurement Device

Page 152: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

152 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

12.1.1 OSICS SWT APC Technical Specifications

OSICS SWT APC

Spectral Range (nm) 1250-1650 nm

Insertion Loss*1*2

*1: Including connectors.*2: On the whole wavelength range.

< 1.5 dB

Polarization Dependent Loss*1*3

*3: At 1550 nm.

< 0.1 dB

Repeatability*2*4

*4: At constant temperature, over 100 successive cycles.

±0.02 dB

Return Loss*1 > 50 dB

Crosstalk*2 > 50 dB

Optical Input/Output Connectors(Module Front Panel)

FC/APC narrow key

Automatic Power Control Yes (with OSICS T100 & ECL)

Input/Output Fiber Type SMF-28 fiber

Synchronization (mainframe) BNC connector N2: 50 ms TTL pulses

Dimensions W x H x D 35 x 128 x 230 mm (1.4 x 5.0 x 9.0 in)

Weight 1 kg (2.21 lb)

Temperature Range 15-35 °C

Specifications apply for wavelengths not equal to any water absorption line.

Page 153: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

OSICS User Manual 153

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN12.1.2 OSICS SWT APC Module Overview

Figure 113: SWT APC Module – Front and Rear Panels

Front Panel Com Connector

Fiber optic connector. It is a standard input/output FC/APC type connector. For more details on the appropriate fiber type, see section OSICS SWT APC Technical Specifications, p. 152.

Chn Connector and LEDs

Fiber optic connector. It is a standard input/output FC/APC type connector. For more details on the appropriate fiber type, see section OSICS SWT APC Technical Specifications, p. 152.The green LEDs indicates which channel is currently active. If the channel is enabled, the LED is lit and CHn is displayed in the System Status screen (see section System Status Screen Description, p. 30).

The label located on the faceplate indicates the locations of the laser source.

Openingfor air output

Top Notch

Bottom Notch

96-pin VMEConnector

Com Connector

Retaining Screw

Chn Connector

LED

Do not connect several laser sources at the same time on the Com and Chn channels, or you may severely damage your optical equipment.

Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

CAUTION

WARNING

Page 154: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

154 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Retaining Screw

The captive screw enables you to lock/unlock the module inserted into the slot of the OSICS Mainframe.

Rear Panel Notches

The top and bottom notches enable you to slide the module into the OSICS Mainframe slot (see section Installing an OSICS Module Into the OSICS Mainframe, p. 23).

96-pin VME Connector

The rear panel of the OSICS modules interfaces with the OSICS Mainframe backplane by means of the 96-pin VME connector.

Opening

The opening ensures proper flow of cooled air drawn from the OSICS Mainframe chassis.

Side Panel The labels affixed on the side panel of the OSICS SWT APC are described in section OSICS Information and Warning Markings, p. 19.

12.2 Selecting the Operating Mode

Subject You can operate the SWT APC in Switch or Full-Band mode. In the Unit Setup menu, the Mode : menu displays the active mode.

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

The Mode : menu display the currently active mode:

2. Turn the rotary knob to put the cursor before the Mode : Switch or Mode : Full-Band menu.

3. Press the knob to toggle from one mode to another.

The Mode : menu displays the selected mode and the corresponding setting menus are displayed.

4. If you set the module to Full-Band mode, you must configure the modules connected to the SWT APC module as explained in section Configuring the Full-Band Mode, p. 156.

The OSICS Mainframe does not support "hot-swap" of modules; for more details, see section Handling OSICS Modules Into the OSICS Mainframe, p. 23).

CAUTION

Page 155: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

OSICS User Manual 155

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN12.3 Selecting the OSICS SWT APC Input/Output Channel (Switch Mode)

Subject This section explains how to select the channel through which the signal will be directed.You can only activate one output channel at a time.

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Set the SWT APC to Switch mode (see section Selecting the Operating Mode, p. 154).

3. Turn the rotary knob to put the cursor before the Select Channel (for SWT APC 1x2 or 1x4) or Select Output Mode (for SWT APC 2x2 or 2x(1x1)) menu and press the knob.

The Channel Selection menu appears and indicates the currently active SWT APC channel or output mode with a cross next to the selected channel.

Figure 114: SWT APC Unit Setup – Channel Selection menu (SWT APC 1x4 Example Screen)

4. Turn the rotary knob to put the cursor before the channel or output mode you want to activate and press it to validate your selection.

The corresponding channel is immediately activated and a cross appears next to the selected channel. On the SWT APC front panel, the LED corresponding to the selected channel or output mode is lit.

5. Press one of the following control buttons:

• ESCAPE: to go back to the SWT APC Unit Setup menu.

• EXIT: to display the System Status screen.

In the System Status screen, the selected channel number or output mode is displayed in the corresponding SWT APC channel Info field.

Main

EXIT

1 2 3 4 5 6 7 8SWT Unit Setup - Channel Selection

ESCAPE

Select Channel:

Ch1Ch2 xCh3Ch4

Page 156: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

156 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

12.4 Configuring the Full-Band Mode

The OSICS SWT APC optical switch module enables you to drive from 1 to 2 or to 1 to 4 (depending on your SWT APC model) OSICS T100 and OSICS ECL modules plugged in the same OSICS Mainframe: you select a wavelength and an output power in the OSICS SWT APC setup menu, and the module automatically selects the channel and drives the OSICS T100 or ECL to provide the wanted values.To use the OSICS SWT APC in Full-Band mode, the module must know where OSICS T100 and OSICS ECL modules are located in the Mainframe. For this, you can use one of the following methods:• Manual configuration: use this method if an OSICS T100 or ECL is running with

another application, or if an ECL has a software version prior to v. 2.21 (in this case the wavelength range cannot be automatically defined).

• Automatic detection: use this method if the OSICS T100 and ECL modules installed in the Mainframe are not running for other applications.

12.4.1 Automatically Detecting the T100 and ECL Modules Connected to the SWT APC

Subject The OSICS T100 and ECL modules already installed in the Mainframe (not being used by other applications) can be automatically detected by the OSICS SWT APC module.The automatic detection does not work for ECL modules with a software version prior to v. 2.21.The auto configuration takes approximately 20 seconds. During this operation, the optical outputs of the connected OSICS T100 or OSICS ECL are sequentially turned off and on.

Before Starting • Connect the wanted OSICS T100 and ECL to the input ports of the OSICS SWT APC Ch1 to Chn.

• Set the SWT APC to Full-Band mode (see section Selecting the Operating Mode, p. 154) and configure the connected modules (section Configuring the Full-Band Mode, p. 156).

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the ECL vs Channel menu enter the Auto menu.

The ECL vs Channel menu appears, prompting you to start the detection.

3. Press the knob to start the detection.

The SWT APC sequentially disables and enables the optical output of the connected OSICS T100 and ECL modules.

If two OSICS T100 or ECL modules have common wavelengths, the SWT APC selects the first module in the mainframe to achieve the range.

Once the automatic detection is complete the automatically detected values are displayed.

Page 157: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

OSICS User Manual 157

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 115: SWT APC Setup – Channel Configuration menu (1x4 Example Screen)

4. Do one of the following:

• If you want to manually change the configuration, modify the wanted wavelength range as explained in section Manually Configuring the T100 and ECL Modules Connected to the SWT APC, p. 157.

• If you don’t want to modify the displayed values press the ENTER control button.

The module configuration is validated and the SWT APC Unit Setup screen is displayed.

12.4.2 Manually Configuring the T100 and ECL Modules Connected to the SWT APC

Subject If one of your OSICS T100 or ECL is used by other applications, or if an ECL has a software version older than v. 2.21, you can manually configure the OSICS T100 and ECL modules connected to your SWT APC module.

Before Starting • Connect the wanted OSICS T100 and ECL to the input ports of the OSICS SWT APC Ch1 to Chn.

• Set the SWT APC to Full-Band mode (see section Selecting the Operating Mode, p. 154) and configure the connected modules (section Configuring the Full-Band Mode, p. 156).

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the ECL vs Channel menu enter the Edit menu.

The ECL vs Channel configuration screen appears and displays the following information:• Ch column: SWT APC input port number.• ECL column: mainframe slot number in which is installed the T100 or ECL module.• λmin/λmax columns: wavelength range of the connected module in nm.

Main

ENTER

2 3 4 5 6 7 8SWT Unit Setup - ECL vs Channel

ESCAPE

Press the knob to change the digit.

1

Ch ECL λmin (nm) λmax (nm) 1 2 1260 1360 2 6 1360 1470 3 3 1460 1580

4 8 1560 1680

Page 158: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

158 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Figure 116: SWT APC Setup – Channel Configuration menu (1x4 Example Screen)

3. Turn the rotary knob to put the cursor before the channel you want to modify and press it to validate your selection.

The corresponding ECL slot number is highlighted.

4. Modify the wanted values as follows:

If you select for two channels the same OSICS T100 or ECL, the new selection cancels the previous one.

e. Turn the knob clockwise to increase the value or anticlockwise to decrease it of the slot.Available slot numbers are the one where a T100 or ECL is located.If you have a channel without OSICS T100 or ECL connected on, select the -- symbols.In the λ column, the wavelength limits are automatically defined for OSICS ECL and T100 with version higher than 2.21.

f. Press the knob to validate the selected digit.The following column digit is highlighted

g. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

h. Perform steps a. to c. again with the other digits to modify.

i. Press the ENTER control button to validate the channel configuration.

5. Perform steps 3. and 4. again with the other channels to modify.

6. Press the ENTER control button again to validate the overall channel configuration.

The module configuration is validated and the SWT APC Unit Setup screen is displayed.

Main

ENTER

2 3 4 5 6 7 8SWT Unit Setup - ECL vs Channel

ESCAPE

Press the knob to change the digit.

1

Ch ECL λmin (nm) λmax (nm) 1 2 1260 1360 2 6 1360 1470 3 3 1460 1580

4 8 1560 1680

Page 159: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

OSICS User Manual 159

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN12.5 Enabling/Disabling the Light Source of all Connected Modules (Full-Band Mode)

Subject In Full-Band mode, you can enable/disable the laser of all modules connected to the SWT APC, as explained in the following procedure.

Before Starting • Connect an optical device to the OSICS SWT APC (see section Connecting Optical Devices to OSICS Modules, p. 26).

• Set the SWT APC to Full-Band mode (see section Selecting the Operating Mode, p. 154) and specify the connected modules (section Configuring the Full-Band Mode, p. 156).

Procedure Enabling the Light Source of All Connected Modules

1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Optical Output : DISABLED menu.

3. Press the rotary knob to enable the laser of all installed modules.

The OSICS modules beep and the laser is enabled on all connected modules simultaneously (the green Enable LED of all modules lights on).

The menu displays Optical Output : ENABLED.

Disabling the Light Source of All Connected Modules

1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Optical Output : ENABLED menu.

3. Press the rotary knob to disable the laser of all connected modules.

The laser is disabled on all modules simultaneously (the green Enable LED of all modules lights off).

• Do not enable the laser while there is no fiber connected to the optical output connector.

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 160: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

160 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

12.6 Setting the Optical Output Power of the SWT APC Module (Full-Band Mode)

Subject You can set the optical output power in two different ways:• From the OSICS Mainframe Main Setup menu, you can simultaneously set the

optical power of all modules to the same value: see section Setting the Optical Output Power of All Installed Modules, p. 38.

• From the SWT APC Unit Setup menu, you can set the output power of the module individually as explained in the following procedure.

In Full-Band mode, you can set the optical output power of the SWT APC module, which drives the T100 or ECL output power with its integrated photodiode, to get a continuous output power on the whole range of wavelength. If you change the wavelength or the output power, the photodiode in the OSICS SWT APC monitors an optical power that is not the real time optical power adjustment (this function is done by the OSICS T100 or ECL).

The power value is software-limited by the SWT APC in a range between -6.99 dBm and +3 dBm.

Before Starting • Set the optical power unit from the OSICS Mainframe Main Setup menu (see section Selecting the Power Unit for All Installed Modules, p. 38). You cannot set the power unit of an individual module.

• Set the SWT APC to Full-Band mode (see section Selecting the Operating Mode, p. 154) and specify the connected modules (section Configuring the Full-Band Mode, p. 156).

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set Power menu and press the knob.

The Set Power menu appears and displays the current power setting.

Figure 117: SWT APC Unit Setup – Set Power menu

3. Modify the power value as follows:

a. Turn the rotary knob to put the cursor under the digit of +/- sign you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Power

ESCAPE

Power :

P = +03.00 dBm ENTER

Modify the value.

Page 161: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

OSICS User Manual 161

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INd. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new power value.

The Unit Setup menu is automatically displayed and the optical power value of the module is now modified (according to the module power limits).

12.7 Setting the Optical Emission Wavelength/Frequency (Full-Band Mode)

12.7.1 Specifying the Optical Emission Wavelength/Frequency

Subject The following procedure explains how to set the emission wavelength or optical frequency of the OSICS SWT APC. The available range is defined by the ECL vs Channel configuration panel of the OSICS SWT APC (see section Configuring the Full-Band Mode, p. 156).

Before Starting • Set the spectral unit from the OSICS Mainframe Main Setup menu (see section Selecting the Spectral Unit for all Installed Modules, p. 39). You cannot set the spectral unit of an individual module.

• Set the SWT APC to Full-Band mode (see section Selecting the Operating Mode, p. 154) and specify the connected modules (section Configuring the Full-Band Mode, p. 156).

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set λ/f menu and press the knob.

The Set Wavelength or Set Frequency (depending on the selected spectral unit) menu appears and displays the emission wavelength or frequency.

Figure 118: SWT APC Unit Setup – Set Wavelength/Frequency menu

3. Modify the wavelength/frequency value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Wavelength

ESCAPE

Wavelength :

λ = 1556.02 nm ENTER

Modify the value.

Page 162: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

162 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new wavelength value.

The Unit Setup menu is automatically displayed and the wavelength/frequency value of the module is now modified (according to the module limits).

12.7.2 Step-adjusting the Wavelength/Frequency Value of the Module

Subject The following procedure explains how to modify the wavelength/frequency of the module step by step using the increment defined in the Set Step sub-menu.You can define a valid step value within the range 00.00–50.00 nm. If you use two non-adjacent T100 or ECL, check that the step is high enough.The starting wavelength is the value originally defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 161).

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Step menu and enter the Set Step sub-menu.

The Set Step sub-menu appears and displays the current value of the wavelength/frequency step in nm or GHz .

Figure 119: SWT APC Unit Setup – Set Step sub-menu

3. Modify the step value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new wavelength value.

The Unit Setup menu is automatically displayed.

4. To increase or decrease the wavelength/frequency value defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 161) step by step, expand the Step menu and put the cursor before the Step + or Step - sub-menus and press the knob to step-adjust the value.While the module is updating the wavelength/frequency, the message Moving… appears along with the new wavelength/frequency value after step-change.

5. Repeat step 4. until the wanted wavelength/frequency is reached.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Step (nm)

ESCAPE

Step:

01.00 nm ENTER

Modify the value.

Page 163: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

OSICS User Manual 163

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN12.7.3 Scanning a Predefined Wavelength/Frequency Range

Subject The Scan function enables you to scan a predefined wavelength range step-by-step, using the channel increment defined in the Step sub-menu to directly set the emission wavelength (or optical frequency) of the SWT APC tunable laser.Starting from the first wavelength (the lower wavelength limit you have set), the emission wavelength increases incrementally (the wavelength increment is stored in the Step parameter) up to the last wavelength value (the upper wavelength limit you have set). During each scan interval, the OSICS T100 or ECL laser driven by the OSICS SWT APC module scans at a selected wavelength range for the duration specified by the Delay parameter.

12.7.3.1 Setting the Scan Parameters

Subject This section explains how to configure the scan mode by setting the following parameters:• The first and last wavelength/frequency of the scan• The value of the wavelength/frequency step that is used in the scan• The time between steps• The number of cycles of the scan

Before Starting • Set the SWT APC to Full-Band mode (see section Selecting the Operating Mode, p. 154) and specify the connected modules (section Configuring the Full-Band Mode, p. 156).

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Scan menu and enter the wanted setting menus:

• λ/f start sub-menu: to select the first wavelength/frequencyof the scan.

• λ/f end sub-menu: to select the last wavelength/frequencyof the scan.

• Step sub-menu: to select the value of the wavelength/frequencystep that you want to use for the scan.

• Delay sub-menu: to select the duration of a scan step.

• Scan Number sub-menu: to select the number of scan cycles. If you select 0000, the OSICS SWT APC will scan indefinitely.

The sub-menu appears and displays the value of the corresponding scan parameter.

3. Modify the value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new value.

The SWT APC Unit Setup menu is automatically displayed and the scan parameters are set.

Page 164: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

164 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

12.7.3.2 Controlling the Scan

Subject This section explains how to start, stop and resume the scan.

Before Starting • Configure the scan parameters as explained in section Setting the Scan Parameters, p. 163.

• Set the optical output power of the SWT APC module (see section Setting the Optical Output Power of the SWT APC Module (Full-Band Mode), p. 160).

• Set the SWT APC to Full-Band mode (see section Selecting the Operating Mode, p. 154) and specify the connected modules (section Configuring the Full-Band Mode, p. 156).

Procedures Starting/Stopping a New Scan

1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Scan menu and enter the Control sub-menu.

Figure 120: SWT APC Unit Setup – Scan Control sub-menu

3. Turn the rotary knob to put the cursor before START and press it to validate your selection.

The Scan sub-menu appears and displays real-time information on the scanning process.

4. To stop the scan, press the ESCAPE control button.

Resuming the Scan

1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Scan menu and enter the Control sub-menu.

3. Turn the rotary knob to put the cursor before RESUME and press it to validate your selection.

The Scan sub-menu appears and displays real-time information on the scanning process.

Main

EXIT

1 2 3 4 5 6 7 8SWT Unit Setup - Scan Control

ESCAPE

Scan Control :

STARTRESUME

Select a new value.

Page 165: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

OSICS User Manual 165

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN12.8 Activating Coherence Control on all Connected T100 and ECL Modules (Full-Band Mode)

Subject The OSICS SWT APC module enables you to control the Coherence Control function of all the connected ECL and T100 modules.The Coherence Control function induces broadening of the OSICS T100 or ECL tunable laser linewidth. This feature is especially useful in helping reduce interferometric noise.Coherence control consists in modulating the frequency with a pseudo-noise signal that widens the OSICS T100 or ECL optical signal linewidth. The specified linewidth for the OSICS T100 or ECL module is 150 kHz (typical value) as indicated in the module specifications without coherence control. This linewidth is brought to 100 MHz when the coherence control function is on.The OSICS T100 and ECL modules feature a built-in modulation source used to implement coherence control function on-board. You do not need to connect any external modulation device to the OSICS system.The following procedure explains how to activate the coherence control function to produce the linewidth broadening required.

Before Starting • Set the SWT APC to Full-Band mode (see section Selecting the Operating Mode, p. 154) and specify the connected modules (section Configuring the Full-Band Mode, p. 156).

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Coherence menu and press the knob.

The Coherence menu appears and the cursor points to the current setting.

Figure 121: SWT APC Unit Setup – Coherence Control menu

3. Turn the rotary knob to put the cursor before the wanted setting, and press it to validate your selection.

The new state of the coherence control function is implemented immediately in all the ECL and T100 modules connected to the SWT APC. The Unit Setup menu is automatically displayed.

In the OSICS System Status screen, the message C is displayed in the OSICS module channel Info field if the coherence control function is currently enabled on this module.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Coherence

ESCAPE

Coherence :

OFF

ON

Select a new value.

Page 166: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

166 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

12.9 Activating the Auto-peak Find Function (Full-Band Mode)

Subject The OSICS SWT APC module enables you to activate the Auto-peak find function for all the connected ECL and T100 modules.The Auto peak find function generates current pulses in order to stabilize the cavity operating conditions and stabilize the linewidth.When implemented, the Auto peak find function generates a small power fluctuation during a brief instant that may disrupt certain applications.If enabled, the Auto-peak Find is performed 1 second immediately after the laser output is enabled and then every 10 minutes afterward.

Before Starting • Set the SWT APC to Full-Band mode (see section Selecting the Operating Mode, p. 154) and specify the connected modules (section Configuring the Full-Band Mode, p. 156).

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Auto peak find menu and press the knob.

The Auto peak find menu appears and the cursor points to the current setting.

Figure 122: SWT APC Unit Setup – Auto peak find menu

3. Turn the rotary knob to put the cursor before the wanted setting, and press it to validate your selection.

The new state is implemented immediatelyon all the ECL and T100 modules connected to the SWT APC. The Unit Setup menu is automatically displayed.

In the OSICS System Status screen, the message A is displayed in the OSICS module channel Info field if the Auto peak find function is currently enabled on this module.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Auto peak find

ESCAPE

Auto peak find :

OFF

ON

Select a new value.

Page 167: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

OSICS User Manual 167

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN12.10 Configuring the Modulation for All Connected T100 and ECL Modules (Full-Band Mode)

Subject The modulation is a typical function of OSICS T100 and ECL modules. For more details on the modulation of these modules, see section Using the OSICS T100 Module, p. 173 and section Using the OSICS ECL Module, p. 99.From the OSICS SWT APC, you can configure the modulation for all connected T100 and ECL modules:• To define the modulation source for all connected T100 and ECL modules, see section

Defining the Modulation Source for All Connected T100 and ECL Modules, p. 167.

• To control the modulation of all connected T100 and ECL modules, see section Controlling the Modulation of All Connected T100 and ECL Modules, p. 168.

Before Starting • Set the SWT APC to Full-Band mode (see section Selecting the Operating Mode, p. 154) and specify the connected modules (section Configuring the Full-Band Mode, p. 156).

12.10.1 Defining the Modulation Source for All Connected T100 and ECL Modules

Subject The following section explains how to configure the wanted modulation source for all T100 and ECL modules connected to the OSICS SWT APC.The SWT APC has no internal modulation source, so if you set the modulation source to INTERNAL, the internal modulation source of the corresponding T100 or ECL module will be used for modulation.

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Source sub-menu.

The Modulation Source sub-menu appears and displays the current selected modulation source.

Figure 123: SWT APC Unit Setup – Modulation Source

3. Turn the rotary knob to put the cursor before the wanted modulation source and press it to validate your selection.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

MAINFRAME

Select a new value.

Page 168: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

168 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

• INTERNAL designates the internal modulation source of the T100 or ECL modules.

• MAINFRAME designates the OSICS Mainframe modulation source, which can be internal or external.

The SWT APC Unit Setup menu is automatically displayed and the modulation source is set for all the T100 and ECL modules connected to the SWT APC.

• If you have selected INTERNAL, set the frequency of the individual T100 and ECL modules using the Set Frequency sub-menu of the module setup menus (as explained in the corresponding module sections).

• If you have selected MAINFRAME, configure the wanted mainframe modulation as explained in the T100 and ECL module sections.

12.10.2 Controlling the Modulation of All Connected T100 and ECL Modules

Subject From the OSICS SWT APC, you can control the modulation of all T100 and ECL modules connected to the SWT APC. To achieve this, the modulation source and frequency must be set for all modules.

Before Starting • Set the wanted modulation source for all T100 and ECL modules connected to the SWT APC, as explained in section Defining the Modulation Source for All Connected T100 and ECL Modules, p. 167, or directly from the setup menu of the modules.

• Set the wanted modulation frequency for each connected module using the Set Frequency sub-menu of the module Unit Setup menu or of the OSICS Mainframe Main Setup menu, depending on the selected modulation source.

Procedure 1. Access the SWT APC Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Figure 124: SWT APC Unit Setup – Modulation Control (On)

3. Turn the rotary knob to put the cursor before the wanted modulation:

• ON for standard modulation.

• ON INVERTED for reverse modulation

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation control

ESCAPE

Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Page 169: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

OSICS User Manual 169

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN4. Press the rotary knob to validate your selection.

Modulation is enabled. In the System Status screen, the letter M is displayed in the channel Info field of the corresponding modules.

5. To stop the modulation: in the Control sub-menu, select OFF.

12.11 Monitoring the Optical Output Power of the SWT APC Module

Subject The OSICS Mainframe rear panel (see Figure 2, p. 18) features eight channel ports, each of them is dedicated to a specific channel-slot and identified by its channel number (CH1 to CH8). As they are numbered from left to right, channel ports do not face across their corresponding slots.The channel port OUT 1 of the SWT APC provides an analog signal that you can use to monitor the module output power in real time. The OUT 1 voltage is electronically linked to the photodiode transimpedance amplifier (not generated by software); so the signal level depends on the photodiode spectral sensitivity and the insertion loss of internal components: at constant power, the OUT 1 signal level depends on wavelength (see Figure 125, p. 169 below).

Figure 125: BNC OUT1 signal level depending on wavelength (at 2 mW)

Before Starting Make sure you have a voltmeter.

Procedure 1. Enable the light source of all connected modules as described in section Enabling/Disabling the Light Source of all Connected Modules (Full-Band Mode), p. 159.

2. To monitor the selected signal, connect your voltmeter to the OUT 1 BNC connector corresponding to the channel slot number of the SWT APC module.

Page 170: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

170 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

12.12 Testing the Performance of the SWT APC Module

Use the procedures described in this section to validate warranted specifications of the OSICS SWT APC module, as needed.

12.12.1 Preparing Performance Tests

Before Starting Make sure you have the following required equipment for conducting performance tests on OSICS SWT APC modules:

The various types as well as the number of fiber-optic connectors needed may vary depending on the type of equipment used and the distance between the test equipment and the OSICS SWT APC modular optical source under test.

Procedure 1. In the OSICS Mainframe, install the SWT APC modules you want to test.

2. Turn-on the OSICS multifunction platform (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

3. Let the OSICS multifunction platform and other test equipments run continuously for 2 hours at room temperature (within the specified operating temperature range) to achieve proper equipment warm-up.

Equipment Minimum RequirementOptical power meter (x2) Wavelength range: 1250 nm to 1650 nm

Resolution: < 0.001 dBDynamic range: > 70 dB

Laser source Wavelength range: 1250 to 1650 nmFiber optic coupler FC/APC connectors

• Disable the laser output before connecting or disconnecting a fiber optic cable on the module (see section Enabling/Disabling the Light Source of All Modules, p. 37).

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 171: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

OSICS User Manual 171

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

12.12.2 Verifying the Repeatability

Subject This procedure verifies that the SWT APC module repeatability is within specification range over 100 measurements for each channel.The following figure illustrates the appropriate setup to perform the repeatability test:

Figure 126: SWT APC Performance Test – Repeatability

Before Starting Make sure you have the appropriate power-meter (see section Preparing Performance Tests, p. 170).

Procedure 1. Disable all laser source outputs and other modules in the OSICS Mainframe.

2. Connect the measurement equipments as indicated on Figure 126, p. 171.

3. If you are using an OSICS T100 or OSICS DFB as the external source, enable the laser output.

4. Enable the output channel under test (in our example Ch1) and record the power output value P1 and P2 provided respectively by OPM1 and OPM2 after the power reading has stabilized.

5. Select dBm as the power-meter measurement unit and hold the range on OPM2 for better accuracy.

6. Turn off and on Ch1 for 100 times, and record each time the power reading P1 and P2 (in dBm) on both power-meters OPM1 and OPM2.

7. Subtract OPM2 reading from OPM1 reading (P1-P2) for each of the 100 measurements and record the result.

8. Find the maximum (|P1-P2|max) and minimum (|P1-P2|min) values among the |P1-P2| records calculated in step 6. above.

9. Subtract the maximum and minimum values to get the module repeatability.

10. Verify that the calculated repeatability value is within the specification range of 0.02 dB.

11. Perform the same operations for the second OSICS SWT APC module channel.

Power-meterOPM1

CouplerLaser sourceIn

Out1

Out2OSICS

SWT APC 1xnPower-meter

OPM2

Ch1

Chn

Com

Page 172: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS SWT APC Module

172 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

12.12.3 Verifying the Crosstalk

Subject The following figure illustrates the appropriate setup to perform the crosstalk test:

Figure 127: SWT APC Performance Test – Crosstalk

Before Starting Make sure you have the appropriate power-meter (see section Preparing Performance Tests, p. 170).

Procedure 1. Disable all laser source outputs and other modules in the OSICS Mainframe.

2. Connect the measurement equipments as indicated on Figure 127, p. 172.

3. Select dBm as the power-meter unit.

4. Select the channel (Ch1 in our example) that you want to measure.

5. Record the power value provided by the power-meter.

6. Switch the laser signal to another channel, while the power-meter is still measuring Ch1.

7. Record the power on the power-meter.

8. Subtract the second power from the first measure to get the crosstalk.

9. Verify that the calculated value is over the specified value of 50 dB.

10. Perform the same operation for the Chn SWT APC channel.

Laser sourceCom OSICS

SWT APC 1xnPower-meter

Ch1

Chn

Page 173: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 173

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN13. Using the OSICS T100 Module

This section explains how to operate the OSICS T100 via the T100 Unit Setup menu. For more details on how to install and initialize the module into the OSICS Mainframe, see section Installing and Connecting the OSICS Multifunction Platform, p. 21.

13.1 OSICS T100 Presentation

The OSICS T100 is a tunable external cavity laser in a pluggable module utilizing Yenista Optics patented T100 cavity. The optical layout of the cavity conforms to a modified Sagnac interferometer configuration. A beam-separator is placed right outside the path of laser-diode emission. Within the cavity, a double-pass reflection on the grating provides maximum dispersion, while the very short cavity, which maximizes mode spacing, yields a spectrally pure and truly single-mode operation.

13.1.1 OSICS T100 Technical Specifications

T100 1310 T100 1415 T100 1520 T100 1550 T100 1575 T100 1620Wavelength Range (nm)

P = 3 dBm 1260–1360 1360–1470 1465–1575 1490–1610 1520–1630 1560–1680P = 6 dBm 1290–1340 1390–1445 1495–1555 1520–1590 1540–1610 1580–1660

Wavelength Setting Accuracy*5 ±0.2 nm

Wavelength Setting Resolution 0.01 nm (0.001 nm option R)

Wavelength Stability*1 ±0.01 nm/h (±0.01 nm / 24 h typical)

Wavelength Setting Repeatability ±0.01 nm typical

Tuning Speed (step-by step)*2*3 10 nm/s typical

Laser Output CharacteristicsSSSER > 90 dB / 0.1 nm typical

SMSR*4 ≥ 45 dB

Output Power Stability*5 ±0.01 dB/h (±0.01 dB / 24 h typical)

Relative Intensity Noise*6 <-140 dB/Hz

Spectral Width (FWHM) 150 kHz typical (coherence control off)> 100 MHz (coherence control on)

Laser Safety Classification Class 1MInterfacesAnalog Modulation 50 Hz to 50 MHz (external)Digital Modulation 50 Hz to 1 MHz (internal & external)Output Fiber Type SMF or PMF (option M)Output Connector FC/APC narrow key

Page 174: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

174 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Permissible Values

Absolute Maximum Ratings

Physical SpecificationsDimensions (W x H x D) 35 mm (1.4 in) x 128 mm (5.0 in) x 230 mm (9.0 in)Weight 1.07 kg (2.36 lb)All specifications are given after 60 minutes warm-up and apply for wavelengths not equal to any water absorption line.

*1: At constant temperature.*2: With the high resolution option (R) the tuning speed is 2.5 nm/s typ.*3: The kinematic chain of the laser does not allow for swept operation.*4: Measured over a 0.1 nm bandwidth ±1nm from the signal.*5: Measured at 0 dBm output power.*6: RIN within the range 100 MHz–3 GHz measured at +3 dBm output power with RBW = 30 kHz.

T100 1310 T100 1415 T100 1520 T100 1550 T100 1575 T100 1620

T100 Model Maximum Power Laser Diode Current LevelP= 3 dBm 2 mW 200 mAP= 6 dBm 4 mW 200 mA

Minimum output power > 0.2 mWMaximum output power < 15 mWMaximum current 200 mA

Page 175: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 175

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN13.1.2 OSICS T100 Module Overview

Figure 128: T100 Module – Front and Rear Panels

Front Panel On LED

This LED indicates that the module is turned on (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

Enable Button

This button enables you to turn on/off the laser inside the module. The green LED of the button is lit if the optical output is enabled, and the power value is displayed in the System Status screen (see section System Status Screen Description, p. 30).

Modulation Input Connector (Coaxial SMB type)

This connector enables you to directly modulate the laser diode chip of the module (see section Setting Up Analog Modulation Using an External Source, p. 193).

Optical Output Connector

The standard output connector is a FC/APC type connector.For more details on the appropriate fiber type, see section OSICS T100 Technical Specifications, p. 173.

The label located on the optical output connector cap indicates the location of the laser source.

Openingfor air output

Top Notch

Bottom Notch

96-pin VMEConnector

On LED

Enable Button

Modulation Connector

Optical OutputConnector

Retaining Screw

Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.WARNING

Page 176: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

176 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Retaining Screw

The captive screw enables you to lock/unlock the module inserted into the slot of the OSICS Mainframe.

Rear Panel Notches

The top and bottom notches enable you to slide the module into the OSICS Mainframe slot (see section Installing an OSICS Module Into the OSICS Mainframe, p. 23).

96-pin VME Connector

The rear panel of the OSICS modules interfaces with the OSICS Mainframe backplane by means of the 96-pin VME connector.

Opening

The opening ensures proper flow of cooled air drawn from the OSICS Mainframe chassis.

Side Panel The labels affixed on the side panel of the OSICS T100 are described in section OSICS Information and Warning Markings, p. 19.

The OSICS Mainframe does not support "hot-swap" of modules; for more details, see section Handling OSICS Modules Into the OSICS Mainframe, p. 23).

CAUTION

Page 177: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 177

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN13.2 Enabling/Disabling the Light Source of the Module

Subject You can enable/disable the light source in two different ways:• From the OSICS Mainframe Main Setup menu (see section Enabling/Disabling the

Light Source of All Modules, p. 37), you can simultaneously enable or disable the light source in all the modules installed in the OSICS Mainframe. This function overrides the enable/disable state individually set in all modules.

• From the T100 front panel, you can individually enable or disable the light source of the module by using the Enable button as explained in the following procedure. This action does not affect the enable/disable state of other modules installed in the OSICS Mainframe.

For laser safety reasons, the light source in modules are disabled after OSICS initialization.

Before Starting Connect an optical device to the OSICS module (see section Connecting Optical Devices to OSICS Modules, p. 26).

Procedures Enabling the Light Source of the Module

• On the T100 module front panel press the Enable button.

The green LED on the button is lit.In the System Status screen, the output power is displayed in the P field (in dBm or mW depending on the selected power unit) for the corresponding module.

Disabling the Light Source of the Module

• On the T100 module front panel press the Enable button.

The green LED on the button is turned off.In the System Status screen, the message Disabled is displayed for the corresponding module.

• Do not enable the laser while there is no fiber connected to the optical output connector.

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 178: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

178 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

13.3 Setting the Optical Output Power of the T100 Module

13.3.1 Specifying the Optical Output Power of the Module

Subject You can set the optical output power in two different ways:• From the OSICS Mainframe Main Setup menu, you can simultaneously set the

optical power of all modules to the same value: see section Setting the Optical Output Power of All Installed Modules, p. 38.

• From the T100 Unit Setup menu, you can set the output power of the module individually as explained in the following procedure.

The power value is software-limited to stay within the specified range.

Before Starting • Set the optical power unit from the OSICS Mainframe Main Setup menu (see section Selecting the Power Unit for All Installed Modules, p. 38). You cannot set the power unit of an individual module.

Procedure 1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set Power menu and press the knob.

The Set Power menu appears and displays the current power setting.

Figure 129: T100 Unit Setup – Set Power menu

3. Modify the power value as follows:

a. Turn the rotary knob to put the cursor under the digit of +/- sign you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new power value.

The Unit Setup menu is automatically displayed and the optical power value of the module is now modified (according to the module power limits).

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Power

ESCAPE

Power :

P = +03.00 dBm ENTER

Modify the value.

Page 179: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 179

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN13.3.2 Performing Real-time Adjustments to the Output Power

Subject The following procedure explains how to adjust the output power of the module in real-time (within the accuracy determined by the smallest decimal-digit of the power value).The starting power is the value defined in the Set Power menu.

Before Starting Enable the optical output of the module (see section Enabling/Disabling the Light Source of the Module, p. 177).

Procedure 1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Adjust Power menu and press the knob.

The Adjust Power menu appears and displays the current power setting, with the first digit highlighted.

Figure 130: T100 Unit Setup – Adjust Power menu

3. Adjust the power value as follows:

a. Turn the knob clockwise to increase the digit value or anticlockwise to decrease it and press the knob to validate the selected digit.Rotating the knob increments or decrements the digit above the cursor. Going above "9" increases the next significant digit; going below "0" decreases the next significant digit.

b. Perform steps a. and b. again with the other digits to modify.

The module output power is updated in real time when you change the value in this menu.

4. Press one of the following control buttons:

• ESCAPE: to validate the displayed value and go back to the T100 Unit Setup menu.

• EXIT: to display the System Status screen.

The optical power value of the module is now modified (according to the module power limits).

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Adjust Power

ESCAPE

Measured Power :

P = +02.10 dBm

Modify the value.

Press the knob to change the digit

Page 180: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

180 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

13.4 Setting the Optical Emission Wavelength/Frequency

13.4.1 Specifying the Optical Emission Wavelength/Frequency

Subject The following procedure explains how to set the emission wavelength or optical frequency of the OSICS T100 module according to its specifications. Even though the module lets you set wavelength values outside this range, performance of your module is guaranteed only within the wavelength range established by the module specifications.

Before Starting • Set the spectral unit from the OSICS Mainframe Main Setup menu (see section Selecting the Spectral Unit for all Installed Modules, p. 39). You cannot set the spectral unit of an individual module.

Procedure 1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set λ/f menu and press the knob.

The Set Wavelength menu appears and displays the emission wavelength or frequency.

Figure 131: T100 Unit Setup – Set Wavelength/Frequency menu

3. Modify the wavelength/frequency value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new wavelength value.

The Unit Setup menu is automatically displayed and the wavelength/frequency value of the module is now modified (according to the module limits).

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Wavelength

ESCAPE

Wavelength :

λ = 1556.02 nm ENTER

Modify the value.

Page 181: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 181

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

13.4.2 Step-adjusting the Wavelength/Frequency Value of the Module

Subject The following procedure explains how to modify the wavelength/frequency of the module step by step using the increment defined in the Set Step sub-menu.You can define a valid step value within the range 00.00–50.00 nm. The starting wavelength is the value originally defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 180).

Procedure 1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Step menu and enter the Set Step sub-menu.

The Set Step sub-menu appears and displays the current value of the wavelength/frequency step in nm or GHz .

Figure 132: T100 Unit Setup – Set Step sub-menu

3. Modify the step value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new wavelength value.

The Unit Setup menu is automatically displayed.

4. To increase or decrease the wavelength/frequency value defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 180) step by step, expand the Step menu and put the cursor before the Step + or Step - sub-menus and press the knob to step-adjust the value.While the module is updating the wavelength/frequency, the message Moving… appears along with the new wavelength/frequency value after step-change.

5. Repeat step 4. until the wanted wavelength/frequency is reached.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Step (nm)

ESCAPE

Step:

01.00 nm ENTER

Modify the value.

Page 182: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

182 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

13.5 Activating Coherence Control

Subject The Coherence Control function induces broadening of the OSICS T100 tunable laser linewidth. This feature is especially useful in helping reduce interferometric noise.Coherence control consists in modulating the frequency with a pseudo-noise signal that widens the OSICS T100 optical signal linewidth. The specified linewidth for the OSICS T100 module is 150 kHz (typical value) as indicated in the module specifications without coherence control. This linewidth is brought to 100 MHz when the coherence control function is on.The OSICS T100 module features a built-in modulation source used to implement coherence control function on-board. You do not need to connect any external modulation device to the OSICS system.The following procedure explains how to activate the coherence control function to produce the linewidth broadening required.

Procedure 1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Coherence menu and press the knob.

The Coherence menu appears and the cursor points to the current setting.

Figure 133: T100 Unit Setup – Coherence Control menu

3. Turn the rotary knob to put the cursor before the wanted setting, and press it to validate your selection.

The new state of the coherence control function is implemented immediately. The Unit Setup menu is automatically displayed.

In the OSICS System Status screen, the message C is displayed in the OSICS module channel Info field if the coherence control function is currently enabled on this module.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Coherence

ESCAPE

Coherence :

OFF

ON

Select a new value.

Page 183: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 183

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN13.6 Activating the Auto-peak Find Function

Subject The Auto peak find function generates current pulses in order to stabilize the cavity operating conditions and stabilize the linewidth.When implemented, the Auto peak find function generates a small power fluctuation during a brief instant that may disrupt certain applications.If enabled, the Auto-peak Find is performed 1 second immediately after the laser output is enabled and then every 10 minutes afterward.

Procedure 1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Auto peak find menu and press the knob.

The Auto peak find menu appears and the cursor points to the current setting.

Figure 134: T100 Unit Setup – Auto peak find menu

3. Turn the rotary knob to put the cursor before the wanted setting, and press it to validate your selection.

The new state is implemented immediately. The Unit Setup menu is automatically displayed.

In the OSICS System Status screen, the message A is displayed in the OSICS module channel Info field if the Auto peak find function is currently enabled on this module.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Auto peak find

ESCAPE

Auto peak find :

OFF

ON

Select a new value.

Page 184: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

184 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

13.7 Configuring the OSICS T100 Signal ModulationPresentation In addition to the OSICS Mainframe built-in modulation source (see section Configuring

the OSICS Mainframe Modulation, p. 40), the OSICS T100 module holds a built-in digital modulation source.The OSICS T100 module supports the following modulation schemes:• Internal modulation:

• Digital (TTL) modulation using the OSICS T100 internal source• Digital (TTL) modulation using the OSICS Mainframe internal source

• External modulation:• Digital (TTL) modulation using an external source connected to the OSICS

Mainframe• Analog modulation using an external source connected to the OSICS T100

The OUT 2 BNC connector can be used as an external modulation source for other devices.

Modulation Scheme

Figure 135: Mainframe and T100 Module Modulation Scheme

Page 185: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 185

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN13.7.1 Setting Up Digital (TTL) Modulation Using the T100 Module Internal Source

Subject Digital modulation capabilities of the T100 module is within the 50 Hz–1 MHz frequency range (see the T100 specifications).Internal digital modulation uses the T100 module's built-in TTL-level signal generator (see Figure 135, p. 184). The T100 module’s built-in TTL signal-generator can provide a TTL modulation signal in the range of 153 Hz–1 MHz, but for optimum modulation performance we recommend applying modulation frequencies within the specified range of the T100 module.The resulting modulated optical output signal follows a pattern as shown in the following figure.

Figure 136: TTL-Modulated Output Power Signal (T100 Module)

Procedure Step 1 – Set the module modulation source to INTERNAL:

1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Source sub-menu.

The Modulation source sub-menu appears and the cursor automatically points to the current modulation source setting.

Figure 137: T100 Unit Setup – Modulation Source (Internal)

3. Turn the rotary knob to put the cursor before INTERNAL and press it to validate your selection.

The T100 Unit Setup menu is automatically displayed and the modulation is set to the selected source.

Time Reference

Power Setting

2 x Power Setting

Optical OutputPower

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

MAINFRAME

Select a new value.

Page 186: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

186 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Step 2 – Set the frequency of the module INTERNAL modulation source:

1. In the T100 Unit Setup menu, expand the Modulation menu and enter the Set Frequency sub-menu.

The Set Frequency sub-menu appears and displays the current selected modulation frequency.

Figure 138: T100 Unit Setup – Set Frequency

2. Modify the frequency value as follows:

The T100 module’s built-in TTL signal-generator can provide a TTL modulation signal in the range of 153 Hz–1 MHz, but for optimum modulation performance we recommend applying modulation frequencies within the specified range of the T100 module.

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps 1. and 2. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new frequency value.

The frequency is updated to the "real frequency". This screen displays the actual modulation frequency produced by the OSICS T100 oscillator, which may slightly differ from the values you have set.

3. Press the knob to validate the displayed real frequency.

Step 3 – Set the modulation control:

1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Main

EXIT

2 3 4 5 6 7 8Unit Setup - Set Frequency

ESCAPE

Frequency :

f = 0016700 Hz

Modify the value.

ENTER

1

Page 187: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 187

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 139: T100 Unit Setup – Modulation Control (On)

3. Turn the rotary knob to put the cursor before the wanted modulation:

• ON for standard modulation.

• ON INVERTED for reverse modulation

4. Press the rotary knob to validate your selection.

Modulation is enabled. In the System Status screen, the letter M is displayed in the channel Info field of the corresponding module.

5. To stop the modulation: in the Control sub-menu, select OFF.

13.7.2 Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source

Subject This modulation setting uses the OSICS Mainframe’s built-in internal oscillator as a source of TTL-level modulated signals (see Figure 135, p. 184). The main purpose of this setting is to provide simultaneous and synchronous modulation to several modules installed in the OSICS Mainframe, or to an external device connected to the OSICS Mainframe Synchro output connector.The range of the frequency of the modulation provided by the OSICS Mainframe is defined to encompass the frequency requirements of all types of OSICS T100 modules. The frequency settings for the lower and upper bounds may therefore lie outside of the modulation capabilities of a particular module, in which case the module's own frequency range sets the modulation specifications.The OSICS Mainframe signal generator can provide a TTL signal with a repetition frequency in the range 123 Hz–1 MHz, but for optimum modulation performance, we recommend applying modulation frequencies within the specified range of the T100 module.The resulting signal-shape of the modulated output power is similar to the internally modulated pattern of the OSICS T100, as shown in Figure 136, p. 185.

Procedure Step 1 – Set the OSICS Mainframe modulation source to INTERNAL:

1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Enter the Source sub-menu as follows:

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation control

ESCAPE

Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Page 188: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

188 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

a. Turn the rotary knob to put the cursor before the Modulation menu and press the knob to expand it.

b. Turn the rotary knob to put the cursor before the Source sub-menu and press the knob to enter it.

The Modulation Source sub-menu appears and displays the current selected modulation source.

Figure 140: Main Setup – Modulation Source (Internal)

3. Turn the rotary knob to put the cursor before INTERNAL and press it to validate your selection.

The Main Setup menu is automatically displayed and the modulation is set to the selected source.

Step 2 – Set the frequency of the OSICS Mainframe internal modulated-signal:

1. In the Main Setup menu, expand the Modulation menu and select the Set frequency sub-menu.

The Modulation frequency sub-menu appears and displays the current selected modulation frequency-value.

Figure 141: Main Setup – Modulation Frequency

2. Modify the frequency value as follows:

The internal TTL-signal generator can provide a TTL-modulated signal in the range 123 Hz to 1 MHz, but for optimum modulation performance we recommend applying modulation frequencies within the specified range of the T100 module.

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

EXTERNAL

Select a new value.

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation frequency

ESCAPE

Frequency value :

f = 0014700 Hz

Modify the value.

ENTER

Page 189: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 189

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INc. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new frequency value.

The frequency is updated to the "real frequency" as shown in Figure 142, p. 189. This screen displays the current modulation frequency produced by the OSICS Mainframe oscillator, which may slightly differ from the values you have set.

Figure 142: Main Setup – Modulation Real Frequency

3. Press the knob to go back to the Main Setup menu.

Step 3 – Set the module modulation source to MAINFRAME:

1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Source sub-menu.

The Modulation source sub-menu appears and the cursor automatically points to the current modulation source setting.

Figure 143: T100 Unit Setup – Modulation Source (Mainframe)

3. Turn the rotary knob to put the cursor before MAINFRAME and press it to validate your selection.

The T100 Unit Setup menu is automatically displayed and the modulation is set to the selected source.

Step 4 – Set the modulation control:

1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation frequency

ESCAPE

The real frequency is :

14706 Hz

Please press the knob.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

MAINFRAME

Select a new value.

Page 190: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

190 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

2. Expand the Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Figure 144: T100 Unit Setup – Modulation Control (On)

3. Turn the rotary knob to put the cursor before the wanted modulation:

• ON for standard modulation.

• ON INVERTED for reverse modulation

4. Press the rotary knob to validate your selection.

Modulation is enabled. In the System Status screen, the letter M is displayed in the channel Info field of the corresponding module.

5. To stop the modulation: in the Control sub-menu, select OFF.

13.7.3 Setting Up Digital (TTL) Modulation Using an External Source

Subject External digital-modulation is achieved by applying a digital signal to the BNC Mod. In connector located on the OSICS Mainframe rear panel (see Figure 135, p. 184). This allows all the modules (or an external device connected to the OSICS Mainframe Synchro output connector) to be modulated in a synchronous way by means of an external source. The OSICS Mainframe distributes the external modulated-signal to all installed modules in order to synchronize all modules with a single modulated signal, amplitude, and frequency.For optimum stability of the modulated signals, we recommend applying a digital modulation in the specified range of the modules.

Before Starting Make sure you have a digital-signal generator.

Procedure Step 1 – Set the OSICS Mainframe modulation source to EXTERNAL:

1. Access the Main Setup menu (see section Accessing the OSICS Mainframe Main Setup Menu, p. 33).

2. Enter the Source sub-menu as follows:

a. Turn the rotary knob to put the cursor before the Modulation menu and press the knob to expand it.

b. Turn the rotary knob to put the cursor before the Source sub-menu and press the knob to enter it.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation control

ESCAPE

Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Page 191: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 191

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INThe Modulation source sub-menu appears and displays the current selected modulation source.

Figure 145: Main Setup – Modulation Source (External)

3. Turn the rotary knob to put the cursor before EXTERNAL and press it to validate your selection.

The Main Setup menu is automatically displayed and the modulation is set to the selected source.

Step 2 – Set the TTL-level signal generator:

1. On the OSICS Mainframe rear panel, connect your external digital signal-generator to the Mod. in BNC connector (see Figure 2, p. 18).

2. Apply the TTL signal not exceeding the maximum voltage of +5 V and repetition frequency between 50 Hz and 1 MHz.

The green LED located next to the Mod. In connector is lit to indicate that an external signal-generator is connected.

Step 3 – Set the module modulation source to MAINFRAME:

1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Source sub-menu.

The Modulation source sub-menu appears and the cursor automatically points to the current modulation source setting.

Figure 146: T100 Unit Setup – Modulation Source (Mainframe)

Main

EXIT

1 2 3 4 5 6 7 8Main Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

EXTERNAL

Select a new value.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation source

ESCAPE

Modulation source :

INTERNAL

MAINFRAME

Select a new value.

Page 192: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

192 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

3. Turn the rotary knob to put the cursor before MAINFRAME and press it to validate your selection.

The T100 Unit Setup menu is automatically displayed and the modulation is set to the selected source.

Step 4 – Set the modulation control:

1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Figure 147: T100 Unit Setup – Modulation Control (On)

3. Turn the rotary knob to put the cursor before the wanted modulation:

• ON for standard modulation.

• ON INVERTED for reverse modulation

4. Press the rotary knob to validate your selection.

Modulation is enabled. In the System Status screen, the letter M is displayed in the channel Info field of the corresponding module.

5. To stop the modulation: in the Control sub-menu, select OFF.

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Modulation control

ESCAPE

Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Page 193: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 193

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

13.7.4 Setting Up Analog Modulation Using an External Source

Subject The SMB connector located on the front panel of the T100 module is directly connected to the laser chip and allows you to couple an external modulating-current to the system bias current (see Figure 135, p. 184).By modulating this external current, you can modulate the optical power output in a frequency range from 50 Hz to 50 MHz.

Figure 148: External Analog Modulation Setup

The relationship between the voltage applied at the SMB input and the current added to the diode is given by: Iin = Vin/Rinternal

where:• Iin is the added current• Vin is the voltage applied to the SMB input• Rinternal is the 50 Ω internal impedance of your T100 module

TTL modulation can also be applied directly on the module's front panel using the input SMB connector generally used for the analog modulation. The modulation is applied directly on the laser chip. In this case, simultaneous modulation for all the modules is not possible.Laser diode current level is available in section OSICS T100 Technical Specifications, p. 173.

Before Starting Make sure you have a 50 Ω coaxial cable to connect your HF-generator to the 50 Ω input SMB connector.

Procedure 1. Set the module modulation control to off as follows:

OSICS Module Laser Diode Source

Rinternal

50 Ω

Waveform Generator

Ibias

Polarization Current

Iin

Idiode

Vin

Coaxial cable 50Ω

Mod.SMB Connector

50Ω

Before applying an external analog-modulation, make sure the modulation control is set to off.

CAUTION

Page 194: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

194 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

a. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

b. Expand the Modulation menu and enter the Control sub-menu.

The Modulation control sub-menu appears and displays the current state of the modulation control.

Figure 149: T100 Unit Setup – Modulation Control (Off)

c. Turn the rotary knob to put the cursor before OFF and press it to validate your selection.

Modulation control is disabled and the T100 Unit Setup menu is displayed.

2. Connect your HF-generator to the 50 Ω input SMB connector using a 50 Ω coaxial cable.

3. Since the SMB input is directly connected to the laser diode chip, make sure that the input voltage is such that the following conditions are always met:

• The diode laser is forward-biased.

• The current flowing through the diode is lower than the maximum allowed current (Imax) indicated in the system acceptance test report.

The best way to fulfill the above conditions is to monitor the T100 module bias-current (Ibias) via the rated output voltage measured at the OUT 1 port located on the rear panel of the OSICS Mainframe (see Figure 2, p. 18), and to compute the total current using the following formula:

Idiode = Ibias + Vin/Rinternal

Where:

• Idiode is the total current through the diode• Ibias is the bias current• Vin is the voltage applied to the SMB input• Rinternal is the 50 Ω internal impedance of the T100 moduleYou must monitor Vin to ensure that the total current remains positive and lower than the maximum allowed current (Imax), as shown in the following figure:

0 < Idiode < Imax

The SMB input is directly connected to the laser diode chip via a 50 Ω resistor. It is essential to make sure that at all times the diode is forward-biased to avoid damage or possible destruction of the diode.

Main

EXIT

1 2 3 4 5 6 7 8

ESCAPE

Modulation control :

OFF

ON

Select a new value.

ON INVERTED

Unit Setup - Modulation control

CAUTION

Page 195: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 195

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 150: Resulting Diode Current with External Analog Modulation Source

If the voltage applied to the SMB input is an AC signal with no DC component and with peak-to-peak amplitude is Vpp, the above condition becomes:Vpp < 2 x Rinternal (Imax - Ibias)Vpp < 2 x Rinternal (Ibias)

Ibias

tIbias

tVin

t

Idiode

Idiode Ibias

Diode modulated current

Current set on OSICSwithout modulation

Input modulated current

Imax

IA

IB

IA < Imax

IB > 0

0

Page 196: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

196 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

13.7.5 Using the T100 Module as a Modulation Source for External Devices (OUT 2 Output)

Subject The OSICS Mainframe rear panel (see Figure 2, p. 18) features eight channel ports, each dedicated to a specific channel-slot and identified by its channel number (CH1 to CH8).As they are numbered from left to right, channel ports do not face across their corresponding slots.Each channel can produce two types of output signals on BNC connectors:• OUT 1: analog signal for constant monitoring of the laser diode current. For more

details, see section Monitoring the OSICS T100 Module Parameters Using the Analog OUT 1 Output, p. 197.

• OUT 2: analog signal from the T100 module (see Figure 135, p. 184) that you can use as an external modulation source for other devices. The following procedure explains how to set the OUT 2 analog signal.

You can also use the OSICS Mainframe as a modulation source for external devices with the Synchro BNC output connector, as explained in section Using the Mainframe as a Modulation Source for External Devices (Synchro Output), p. 44.

Procedure 1. Set up the wanted digital modulation to the T100 module using one of the following methods:

• T100 module internal modulation source, as explained in section Setting Up Digital (TTL) Modulation Using the T100 Module Internal Source, p. 185.

• OSICS Mainframe internal modulation source, as explained in section Setting Up Digital (TTL) Modulation Using the Mainframe Internal Source, p. 187.

• OSICS Mainframe external modulation source, as explained in section Setting Up Digital (TTL) Modulation Using an External Source, p. 190.

2. Connect the wanted appropriate device to the OUT 2 BNC connector corresponding to the channel slot number of the module.

The selected modulation frequency is applied to the connected device.

Page 197: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 197

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN13.8 Monitoring the OSICS T100 Module Parameters Using the Analog OUT 1 Output

Subject The OSICS Mainframe rear panel (see Figure 2, p. 18) features eight channel ports, each of them is dedicated to a specific channel-slot and identified by its channel number (CH1 to CH8).As they are numbered from left to right, channel ports do not face across their corresponding slots.Each channel can produce two types of output signals on BNC connectors:• OUT 2: analog signal from the T100 module that you can use as an external

modulation source for other devices. For more details, see section Using the T100 Module as a Modulation Source for External Devices (OUT 2 Output), p. 196.

• OUT 1: analog signal for constant monitoring of the T100 laser diode current (ILD) or output power (POUT).

The scale-factors for the OUT 1 signal are the following (FPGA v.3.00):

• Laser diode current (ILD): 10 V / 260.4 mA

• Optical output power (POUT): 10 V / 15 mW

The following procedure explains how to set the OUT 1 analog signal.

Before Starting Make sure you have an oscilloscope.

Procedure 1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Analog out menu and press the knob.

The Analog Out menu appears and displays the monitoring signal currently active on the OUT1 connector.

Figure 151: T100 Unit Setup – Analog Out setting

3. Turn the rotary knob to put the cursor before the parameter you want to monitor and press it to validate your selection.

The selected parameter is validated and the T100 Unit Setup menu is displayed.

4. To monitor the selected signal, connect your oscilloscope to the OUT 1 BNC connector corresponding to the channel slot number of the module.

Main

EXIT

2 3 4 5 6 7 8ECL Unit Setup - Analog Out

ESCAPE

Analog Out :

Select a new signal.

1

CURRENT

POWER

Page 198: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

198 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

13.9 Calibrating the T100 Module

13.9.1 Performing a Wavelength Reference Calibration

Subject The Wavelength Reference function allows the OSICS T100 to perform its own internal calibration based on a built-in mechanical reference.

Procedure 1. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Calibration menu and enter the Wavelength Reference sub-menu.

The Wavelength reference sub-menu appears prompting you to confirm whether you want to run the wavelength reference calibration.

3. Press the rotary knob to run the wavelength reference calibration.

The message Calibrating… appears. The referencing process may take up to two minutes. Once the calibration is complete, the T100 Unit Setup menu is displayed.

13.9.2 Performing a Wavelength/Frequency (λ/f) Calibration

Subject In the λ/f calibration sub-menu, first the system internally calibrates the T100 module wavelength or frequency (depending on the unit setting) with a mechanical reference (as explained in section Performing a Wavelength Reference Calibration, p. 198), then the system enters the two-point wavelength/frequency calibration sequence.The purpose of this function is to correct the discrepancies between the wavelength/frequency displayed by OSICS and the wavelength/frequency reading of your reference wavemeter. This calibration is made at two different wavelengths/frequencies using the T100 Unit Setup menu.The new calibration wavelengths/frequencies sets the new system calibration default.

Before Starting Make sure you have a wavemeter.

Procedure 1. Connect the optical output of the OSICS T100 module to the wavemeter.

2. Enable the module optical output (see section Enabling/Disabling the Light Source of All Modules, p. 37).

3. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

4. Expand the Calibration menu and enter the λ/f calibration sub-menu.

The system internally calibrates the T100 module wavelength or frequency with a mechanical reference (as explained in section Performing a Wavelength Reference Calibration, p. 198).The Wavelength/Frequency calibration sub-menu appears, showing the first wavelength/frequency value.

Page 199: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 199

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 152: T100 Unit Setup – λ/f calibration, first wavelength

This is the first step of the two-point wavelength/frequency calibration method.The first wavelength/frequency is the first factory calibration wavelength/frequency. The value you enter in this field is the wavemeter reading.

5. Enter the value measured at the wavemeter as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the value.

The message Moving... appears while the first wavelength/frequency is being set.

The Wavelength/Frequency calibration window appears, showing the second wavelength value.

Figure 153: T100 Unit Setup – λ/f calibration, second wavelength

This is the second step of the two-point wavelength/frequency calibration method.

6. Measure the second wavelength/frequency using the wavemeter.

7. Enter the wavelength/frequency reading in this field and select ENTER.

A confirmation message appears.

8. Press the knob to confirm.

While updating the emission wavelength/frequency, the system displays the Moving… message until the wavelength/frequency is set.

Main

EXIT

2 3 4 5 6 7 8ECL Unit Setup - Wavelength calibration

ESCAPE

First wavelength value :

Correct the value.

ENTER

1

λ = 1540.461 nm

Main

EXIT

2 3 4 5 6 7 8ECL Unit Setup - Wavelength calibration

ESCAPE

Second wavelength value :

Correct the value.

ENTER

1

λ = 1589.694 nm

Page 200: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

200 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Once both calibration points are entered, the new calibration values are implemented immediately and stored in the system's flash-memory, and the current wavelength/frequency setting of the T100 module is updated to match the new calibration data. The new calibration wavelengths/frequencies sets the new system calibration default.

The T100 Unit Setup menu is displayed.

13.9.3 Performing a Power Calibration

Subject This section explains how to perform a two-point power calibration sequence using the Power calibration function. The purpose of this calibration is to correct the discrepancies between the power displayed by the OSICS T100 module and the power measured at your reference power meter.Also, calibration is made at two different wavelengths to account for small wavelength-dependence of the power meter.The new calibration data sets the new system calibration default.

Before Starting Make sure you have a power meter.

Procedure 1. Set the T100 output power to 1 mW.

2. Connect the optical output of the OSICS T100 module to a power meter.

3. Enable the module optical output (see section Enabling/Disabling the Light Source of the Module, p. 177).

4. Access the T100 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

5. Expand the Calibration menu and enter the Power calibration sub-menu.

The message Moving... appears, and the Power calibration sub-menu appears, prompting you to set the power at the first wavelength value.

Figure 154: T100 Unit Setup – Power calibration, power value at first wavelength

6. Measure the power level at the first wavelength on the power meter.

A valid power value can range from 0.6 mW to 1.2 mW.

7. Enter the power value measured by the power meter and select ENTER.

The message Moving... appears while the power value is set. The Power calibration window appears, showing the second window. The T100 tunable laser

Main

EXIT

2 3 4 5 6 7 8ECL Unit Setup - Power calibration

ESCAPE

First power value :

Correct the value.

ENTER

1

P = +09.160 dBm

First wavelength : 1540.000 nm.

Page 201: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 201

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INmoves to a second wavelength in the tuning range and displays the power level corresponding to the calibration data.

Figure 155: T100 Unit Setup – Power calibration, power value at second wavelength

8. Measure the power level at the second wavelength on the power meter.

A valid power value can range from 0.6 mW to 1.2 mW.

9. Enter the power value measured by the power meter in this field and select ENTER.

A confirmation message appears.

10. Press the knob to confirm.

The message Moving... appears while the power value is set. The power calibration is now complete and the values are stored in the module's non-volatile memory.

The T100 Unit Setup menu is displayed.

The new calibration data sets the new system calibration default.

13.9.4 Asking for a Factory Recalibration

Subject To externally verify the performance of the OSICS, refer to the procedure described in section Performing a Wavelength Reference Calibration, p. 198 and section Performing a Power Calibration, p. 200. If the performance does not meet warranted specifications, a complete calibration of the instrument may be required.We recommend performing factory calibration of OSICS T100 modules over one-year intervals.

Procedure Contact the Yenista Optics customer support service (see section Contact Information, p. 4) for a complete calibration, as some calibration steps require the adjustment of internal components including the laser head.

Main

EXIT

2 3 4 5 6 7 8ECL Unit Setup - Power calibration

ESCAPE

Second power value :

Correct the value.

ENTER

1

P = +09.160 dBm

Second wavelength : 1590.000 nm.

Page 202: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

202 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

13.10 Testing the Performance of the T100 Module

Use the procedures described in this section to validate warranted specifications of the OSICS T100 module, as needed.During the performance verification, the settings for wavelength and power may not agree with the measured values. If this occurs, you may calibrate wavelength and power settings as explained in section Performing a Wavelength/Frequency (λ/f) Calibration, p. 198 and section Performing a Power Calibration, p. 200.Notwithstanding performance tests described in this section, we recommend performing calibration of OSICS T100 modules over one-year intervals (see section Asking for a Factory Recalibration, p. 201).

13.10.1 Preparing Performance Tests

Before Starting Make sure you have the following required equipment for conducting performance tests on OSICS T100 modules:

The various types as well as the number of fiber-optic connectors needed may vary depending on the type of equipment used and the distance between the test equipment and the OSICS T100 modular optical source under test.

Procedure 1. In the OSICS Mainframe, install the T100 modules you want to test.

2. Turn-on the OSICS multifunction platform (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

3. Let the OSICS multifunction platform and other test equipments run continuously for 2 hours at room temperature (within the specified operating temperature range) to achieve proper equipment warm-up.

Equipment Minimum RequirementOptical power meter Wavelength range: 1260 nm to 1680 nmOptical spectrum analyzer (OSA) Wavelength range: 1260 nm to 1680 nmWavemeter Wavelength range: 1260 nm to 1680 nmJumper (x2) FC/APC to FC/PC or to FC/APCIn-line optical adapter FC to FC

• Disable the laser output before connecting or disconnecting a fiber optic cable on the module (see section Enabling/Disabling the Light Source of All Modules, p. 37).

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 203: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

OSICS User Manual 203

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

13.10.2 Measuring the Absolute Wavelength Accuracy

Subject This test procedure enables you to verify that the actual measured wavelength stays within acceptable deviation range from the wavelength setting.

Before Starting • Make sure you have a wavemeter with ±1 pm resolution and a NIST-traceable calibration (see section Preparing Performance Tests, p. 202) and jumper.

Procedure 1. Connect the OSICS T100 module to the wavemeter.

2. Set the OSICS T100 output power to 0 dBm.

3. Set the wavelength of the OSICS T100 tunable laser to the minimum value of the wavelength range. The minimum wavelength depends on the T100 model you are using (see section Specifying the Optical Emission Wavelength/Frequency, p. 180).

4. Set the wavelength step to +5 nm.

5. Enable the laser output by pressing the Enable button on the module front panel.6. Compute the wavelength deviation of the T100 module and check that it is not

greater than ±0.2 nm compared with the wavelength reading of the wavemeter.

The absolute wavelength accuracy is the deviation between the wavelength setting and the wavelength reading provided by the wavemeter at each measurement step.

7. Increment the T100 wavelength setting one wavelength step (+5 nm) and repeat measurement step 6. to check that the wavelength deviation is not greater than ±0.2 nm.

8. Stop when you reach the T100 maximum wavelength limit.

If the minimum and maximum wavelength limits do not meet the specifications, you can calibrate these values as explained in section Performing a Wavelength/Frequency (λ/f) Calibration, p. 198.

13.10.3 Checking the Wavelength Range and Output Power

Subject This procedure enables you to verify that the T100 module can achieve the rated output power at both ends of the T100 wavelength range.

Before Starting Make sure you have a powermeter with a NIST-traceable calibration (see section Preparing Performance Tests, p. 202) and jumper.

Procedure 1. Disable the T100 laser output.

2. Connect the T100 laser output to the optical power meter.

3. From the T100 Unit Setup menu, set the wavelength of the T100 tunable laser to the minimum wavelength that corresponds to the type and power option of the T100 module you are using (see OSICS T100 Technical Specifications).

4. Enable the T100 module laser output by pressing the Enable button.

Page 204: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS T100 Module

204 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

5. Set the output power value to the rated output power, as given in section OSICS T100 Technical Specifications and check that the T100 module can deliver the desired output power at the given wavelength.

6. Set the wavelength of the T100 tunable laser to the maximum wavelength that corresponds to the type and power option of the T100 module you are using (see section OSICS T100 Technical Specifications) and check that the T100 module can deliver the desired output power at the given wavelength.

7. Do one of the following, depending on the measured values:

• If the T100 power settings show small discrepancies compared to the measured powers, perform a two-point manual calibration of the T100 output power (see section Performing a Power Calibration, p. 200).

• If either of the rated output power at the wavelength range ends cannot be achieved, the T100 module requires servicing. Please contact your Yenista Optics customer support service (see section Contact Information, p. 4) immediately.

13.10.4 Validating the Side Mode Suppression

Subject This procedure enables you to verify the suppression of any wave emissions that are more than 1 nm away from the selected wavelength.

Before Starting Make sure you have an appropriate optical spectrum analyzer (see section Preparing Performance Tests, p. 202) and jumper.

Procedure 1. Disable the T100 laser output.

2. Connect the T100 laser output to the optical spectrum analyzer.

3. Set the wavelength of the T100 module to a value at mid-range of the wavelength range as set in the T100 module specifications (see OSICS T100 Technical Specifications).

4. Set the T100 laser output power to 0 dBm.

5. Configure the optical spectrum analyzer as follows:

• Center wavelength: same as T100 laser wavelength setting• Span: 20 nm• Resolution: 0.1 nm• Reference level: 0 dBm• Level scale: 10 dB/div• Scan: Repeat

6. Enable the T100 module laser output by pressing the Enable button.

7. Check that, at wavelength deviations larger than 1 nm away from the center wavelength of the laser peak, the power level of any residual emission is lower than-45 dB down from the peak level.

Page 205: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 205

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN14. Using the OSICS TLS-50 Module

This section explains how to operate the OSICS TLS-50 via the TLS-50 Unit Setup menu. For more details on how to install and initialize the module into the OSICS Mainframe, see section Installing and Connecting the OSICS Multifunction Platform, p. 21.

14.1 OSICS TLS-50 PresentationThe OSICS TLS-50 modules are tunable laser sources with high output power and very high accuracy based on the ITU-T grid. The wavelength can be switched over about 90 channels of the ITU-T grid by 50 GHz steps, covering around 35 nm in C or L band. The TLS-50 module is particularly adapted to testing of WDM routing.

14.1.1 OSICS TLS-50 Technical Specifications

C-band L-bandNumber of ITU Channels 89 (50 GHz spacing) 93 (50 GHz spacing)

Wavelength Range*1 196.1 to 191.7 THz 1528.77 to 1563.86 nm

191.1 to 186.5 THz 1568.77 to 1607.47 nm

Output Power 20 mW (+13 dBm) 10 mW (+10 dBm)Power Range (typ.) +8 to +14 dBm +5 to +11 dBmWavelength Accuracy ±1.8 GHzWavelength Setting Resolution 50 GHzSwitching Speed < 20 ms (typ. between two channels)

Power Stability*1*2 ±0.05 dB

Absolute Power Deviation Across Tuning Range 0.5 dB

Linewidth (FWHM) < 5 MHz (1 MHz typ.)

SBS Suppression*3 Yes

Frequency Modulation*4 10 kHZ to 100 kHz

Trace Tone*5

(Amplitude Modulation)10 kHZ to 500 kHz

Side Mode Suppression Ratio*1 > 40 dB (45 dB typ.)

RIN*1*2*6 < -150 dB/Hz

Operating Temperature Range +15 to +35 °C (+59 to +95 °F)Optical Interface FC/APC connector on PM15 PER > 20 dBDimensions (WxHxD) & Weight 35 x 130 x 250 mm (single slot), 1 kgLaser Safety Classification Class 1 M

Page 206: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

206 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

14.1.2 OSICS TLS-50 Module Overview

Figure 156: TLS-50 Module – Front and Rear Panels

Front Panel On LED

This LED indicates that the module is turned on (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

Enable Button

This button enables you to turn on/off the laser inside the module. The green LED of the button is lit if the optical output is enabled and the power value is displayed in the System Status screen (see section System Status Screen Description, p. 30).

Optical Output Connector

The standard output connector is a FC/APC type connector.For more details on the appropriate fiber type, see section OSICS TLS-50 Technical Specifications, p. 205.

*1: After warm-up of 60 s.*2: Over 2 h at a constant temperature.*3: Enable: on/off. Linewidth from 250 to 1000 MHz depending on the selected frequency mo-

dulation.*4: Waveshape selection: sinusoidal or triangular.*5: Tone depth: up to 10 %*6: RIN within the range 100 MHz–20 GHz measured at 10 dBm output power with RBW = 30 kHz

Openingfor air output

Top Notch

Bottom Notch

96-pin VMEConnector

On LED

Enable Button

Optical OutputConnector

Retaining Screw

Page 207: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

OSICS User Manual 207

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INThe label located on the optical output connector cap indicates the location of the laser source.

Retaining Screw

The captive screw enables you to lock/unlock the module inserted into the slot of the OSICS Mainframe.

Rear Panel Notches

The top and bottom notches enable you to slide the module into the OSICS Mainframe slot (see section Installing an OSICS Module Into the OSICS Mainframe, p. 23).

96-pin VME Connector

The rear panel of the OSICS modules interfaces with the OSICS Mainframe backplane by means of the 96-pin VME connector.

Opening

The opening ensures proper flow of cooled air drawn from the OSICS Mainframe chassis.

Side Panel The labels affixed on the side panel of the OSICS TLS-50 are described in section OSICS Information and Warning Markings, p. 19.

Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

The OSICS Mainframe does not support "hot-swap" of modules; for more details, see section Handling OSICS Modules Into the OSICS Mainframe, p. 23).

WARNING

CAUTION

Page 208: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

208 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

14.2 Enabling/Disabling the Light Source of the Module

Subject You can enable/disable the light source in two different ways:• From the OSICS Mainframe Main Setup menu (see section Enabling/Disabling the

Light Source of All Modules, p. 37), you can simultaneously enable or disable the light source in all the modules installed in the OSICS Mainframe. This function overrides the enable/disable state individually set in all modules.

• From the TLS-50 front panel, you can individually enable or disable the light source of the module by using the Enable button as explained in the following procedure. This action does not affect the enable/disable state of other modules installed in the OSICS Mainframe.

For laser safety reasons, the light source in modules are disabled after OSICS initialization.

Before Starting Connect an optical device to the OSICS module (see section Connecting Optical Devices to OSICS Modules, p. 26).

Procedures Enabling the Light Source of the Module

• On the TLS-50 module front panel press the Enable button.

The green LED on the button is lit.In the System Status screen, the output power is displayed in the P field (in dBm or mW depending on the selected power unit) for the corresponding module.

Disabling the Light Source of the Module

• On the TLS-50 module front panel press the Enable button.

The green LED on the button is turned off.In the System Status screen, the message Disabled is displayed for the corresponding module.

• Do not enable the laser while there is no fiber connected to the optical output connector.

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 209: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

OSICS User Manual 209

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN14.3 Setting the Optical Output Power of the TLS-50 Module

Subject You can set the optical output power in two different ways:• From the OSICS Mainframe Main Setup menu, you can simultaneously set the

optical power of all modules to the same value: see section Setting the Optical Output Power of All Installed Modules, p. 38.

• From the TLS-50 Unit Setup menu, you can set the output power of the module individually as explained in the following procedure.

Before Starting • Set the optical power unit from the OSICS Mainframe Main Setup menu (see section Selecting the Power Unit for All Installed Modules, p. 38). You cannot set the power unit of an individual module.

Procedure 1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set Power menu and press the knob.

The Set Power menu appears and displays the current power setting.

Figure 157: TLS-50 Unit Setup – Set Power menu

3. Modify the power value as follows:

a. Turn the rotary knob to put the cursor under the digit of +/- sign you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new power value.

The Unit Setup menu is automatically displayed and the optical power value of the module is now modified (according to the module power limits).

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Power

ESCAPE

Power :

P = +03.00 dBm ENTER

Modify the value.

Page 210: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

210 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

14.4 Setting the Optical Emission Wavelength/Frequency

14.4.1 Specifying the Optical Emission Wavelength/Frequency

Subject The following procedure explains how to set the emission wavelength or optical frequency of the OSICS TLS-50 module according to its specifications.

Before Starting • Set the spectral unit from the OSICS Mainframe Main Setup menu (see section Selecting the Spectral Unit for all Installed Modules, p. 39). You cannot set the spectral unit of an individual module.

Procedure 1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set λ/f menu and press the knob.

The Set λ/f menu appears and displays the current channel and the corresponding emission wavelength or frequency.

Figure 158: TLS-50 Unit Setup – Set Wavelength/Frequency menu

3. Modify the wavelength/frequency value channel value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Press the ENTER control button to validate the new wavelength and to go back to the Unit Setup menu.

The Unit Setup menu is automatically displayed and the wavelength/frequency value of the module is now modified (according to the module limits).

Main

ENTER

1 2 3 4 5 6 7 8TLS Unit Setup - Set λ/F

ESCAPE

Channel:

CH = 01 (1565.701 nm)

Modify the value.

Page 211: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

OSICS User Manual 211

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

14.4.2 Step-adjusting the Wavelength/Frequency Value of the Module

Subject The following procedure explains how to modify the wavelength/frequency of the module step by step using the increment defined in the Set Step sub-menu.

The starting wavelength is the value originally defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 210).

Procedure 1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Step menu and enter the Set Step sub-menu.

The Set Step sub-menu appears and displays the current value of the channel step and the corresponding step in GHz.

Figure 159: TLS-50 Unit Setup – Set Step sub-menu

3. Modify the step value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Press the ENTER control button to go back to the Unit Setup menu.

The Unit Setup menu is automatically displayed.

4. To increase or decrease the wavelength/frequency value defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 210) step by step, expand the Step menu and put the cursor before the Step + or Step - sub-menus and press the knob to step-adjust the value.While the module is updating the wavelength/frequency, the message Moving… appears along with the new wavelength/frequency value after step-change.

5. Repeat step 4. until the wanted wavelength/frequency is reached.

Main

ENTER

1 2 3 4 5 6 7 8TLS Unit Setup - Set Step

ESCAPE

Step in Channel(s):

dCH = 02 (df = 50 GHz)

Modify the value.

Page 212: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

212 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

14.4.3 Scanning a Predefined Wavelength/Frequency Range

Subject The Scan function enables you to scan a predefined wavelength range step-by-step, using the channel increment defined in the Step sub-menu to directly set the emission wavelength (or optical frequency) of the TLS-50 tunable laser.Starting from the first wavelength (the lower wavelength limit you have set), the emission wavelength increases incrementally (the wavelength increment is stored in the Step parameter) up to the last wavelength value (the upper wavelength limit you have set). During each scan interval, the OSICS TLS-50 module scans at a selected wavelength range for the duration specified by the Delay parameter.

14.4.3.1 Setting the Scan Parameters

Subject This section explains how to configure the scan mode by setting the following parameters:• The first and last channel of the scan• The value of the channel step that is used in the scan• The time between steps• The number of cycles of the scanThe laser stabilization is in the millisecond range. You can immediately use it for accurate measurements.

Procedure 1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Scan menu and enter the wanted setting menus:

• Start Channel sub-menu: to select the first channelof the scan.

• End Channel sub-menu: to select the last channelof the scan.

• Step sub-menu: to select the value of the channelstep that you want to use for the scan.

• Delay sub-menu: to select the duration of a scan step.

• Scan Number sub-menu: to select the number of scan cycles. If you select 0000, the OSICS TLS-50 will scan indefinitely.

The sub-menu appears and displays the value of the corresponding scan parameter.

3. Modify the value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new value.

The TLS-50 Unit Setup menu is automatically displayed and the scan parameters are set.

14.4.3.2 Controlling the Scan

Subject This section explains how to start, stop and resume the scan.

Page 213: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

OSICS User Manual 213

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INBefore Starting • Configure the scan parameters as explained in section Setting the Scan Parameters, p. 212.

The laser stabilization is in the millisecond range. You can immediately use it for accurate measurements.

Procedures Starting a New Scan

1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Scan menu and enter the Control sub-menu.

Figure 160: TLS-50 Unit Setup – Scan Control sub-menu

3. Turn the rotary knob to put the cursor before START and press it to validate your selection.

The Unit Setup menu is displayed and the scan starts. Real-time information on the scanning process is displayed in the System Status screen.

Stopping the Scan

1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Scan menu and enter the Control sub-menu.

3. Turn the rotary knob to put the cursor before STOP and press it to validate your selection.

The scan stops at the wavelength/frequency reached when the Stop function was activated; the Unit Setup menu is displayed.

Resuming the Scan

1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Scan menu and enter the Control sub-menu.

3. Turn the rotary knob to put the cursor before RESUME and press it to validate your selection.

The Unit Setup menu is displayed and the scan starts at the last channel selected in the previous scan. Real-time information on the scanning process is displayed in the System Status screen.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - Scan Control

ESCAPE

Scan Control:

STARTRESUMESTOP

ENTER

Select a new value.

Page 214: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

214 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

14.5 Setting the SBS Suppression Mode

The SBS suppression mode allows you to add internal modulation to the source, to broaden the laser line width. You can select one of the following (both modulation types cannot be selected at the same time):• Amplitude Modulation (AM)

The amplitude modulation (trace tone) operates at a frequency that can be detected by an optical receiver. The operational frequency band is within the opto RX detectable bandwidth > 50 kHz.

AM signaling is typically used during channel commissioning to identify a specific transmitter wavelength detected by an optical receiver.

Figure 161: TLS-50 – Amplitude Modulation

You can configure this modulation mode with the following parameters:

• The frequency of this modulation from 10 to 500 kHz (SBS Rate, see section Selecting the SBS Rate, p. 216)

• The shape: sinusoidal or triangular (SBS Waveform, see section Selecting the SBS Waveform, p. 216)

• The amplitude deviation is the amplitude modulation depth in % = X/Y*100% between 0.1 and 10% (SBS AM, see section Selecting the Amplitude Deviation, p. 218)

OIF standard recommendation: Sinusoidal, 100 to 200 kHz (within the optical receiver unit detectable bandwidth).

• Frequency Modulation (FM)

Frequency modulation requires the application of a dither signal to modulate the frequency of the laser, to broaden the effective linewidth of the laser and reduce the effects of Stimulated Brillouin Scattering in the optical fiber, the laser is modulated with a single frequency dither signal (triangular or sinusoidal). Ideally the frequency (repetition rate) of the SBS dither signal must be kept within the lower kHz range.

You can configure this modulation mode with the following parameters:

• The frequency of this modulation from 10 to 100 kHz (SBS Rate, see section Selecting the SBS Rate, p. 216)

• The shape: sinusoidal or triangular (SBS Waveform, see section Selecting the SBS Waveform, p. 216)

X Mean CWPower Level

Y

ZeroPower Level

Page 215: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

OSICS User Manual 215

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN• The frequency deviation from 0.3 to 1 GHz (SBS FM, see section Selecting the Frequency Deviation, p. 217)

OIF standard recommendation: Triangular, 10 to 50 kHz (generally below the optical receiver unit detectable bandwidth)

If the SBS function is activated, a S flag appears on the right side of the System Status screen (see Figure 9, p. 30).

14.5.1 Selecting the Modulation Mode

Subject The following procedure explains how to select the wanted modulation mode: amplitude modulation (AM) or frequency modulation (FM).

Procedure 1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the SBS Suppression menu and enter the SBS Modulation Mode sub-menu.

The sub-menu appears and displays the current state of the SBS Suppression.

Figure 162: TLS-50 Unit Setup – SBS Modulation Mode sub-menu

3. Turn the rotary knob to put the cursor before the wanted mode:

• AM for amplitude modulation

• FM for frequency modulation

4. Press the knob to validate your selection.

The SBS Suppression is set to the selected mode and the Unit Setup menu is automatically displayed.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - SBS Modulation Mode

ESCAPE

AMFM

SBS modulation mode :

Select a new value.

Page 216: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

216 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

14.5.2 Selecting the SBS Rate

Subject The OIF recommended values for SBS rate (for amplitude and frequency modulation) are given in section Setting the SBS Suppression Mode, p. 214.

Procedure 1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the SBS Suppression menu and enter the SBS Rate sub-menu.

The sub-menu appears and displays the current frequency deviation.

Figure 163: TLS-50 Unit Setup – SBS FM sub-menu

3. Modify the value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digit to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new value.

The SBS rate is set and the TLS-50 Unit Setup menu is automatically displayed.

14.5.3 Selecting the SBS Waveform

Subject You can choose between triangular or sinusoidal shape, whatever the selected modulation mode. The OIF recommended values for waveform (for amplitude and frequency modulation) are given in section Setting the SBS Suppression Mode, p. 214.

Procedure 1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the SBS Suppression menu and enter the SBS Waveform sub-menu.

The sub-menu appears and displays the current state of the SBS Suppression.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - SBS Rate

ESCAPE

010 kHz

SBS Rate :

Select a new value.

ENTER

Page 217: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

OSICS User Manual 217

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 164: TLS-50 Unit Setup – SBS Waveform sub-menu

3. Turn the rotary knob to put the cursor before the wanted shape and press it to validate your selection.

The SBS Suppression is set to the selected waveform and the Unit Setup menu is automatically displayed.

14.5.4 Selecting the Frequency Deviation

Subject Frequency modulation requires the application of a dither signal to modulate the frequency of the laser; to broaden the effective linewidth of the laser and reduce the effects of Stimulated Brillouin Scattering in the optical fiber.The OIF recommended values for SBS rate (for amplitude and frequency modulation) are given in section Setting the SBS Suppression Mode, p. 214.

Procedure 1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the SBS Suppression menu and enter the SBS FM sub-menu.

The sub-menu appears and displays the current optical frequency deviation.

Figure 165: TLS-50 Unit Setup – SBS FM sub-menu

3. Modify the value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - SBS Waveform

ESCAPE

SinusoidalTriangular

SBS waveform :

Select a new value.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - SBS FM

ESCAPE

1.0 GHz ENTER

Frequency Deviation :

Select a new value.

Page 218: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

218 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

c. Perform steps a. and b. again with the other digit to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new value.

The TLS-50 Unit Setup menu is automatically displayed and the frequency deviation is set.

14.5.5 Selecting the Amplitude Deviation

Subject The OIF recommended values for SBS rate (for amplitude and frequency modulation) are given in section Setting the SBS Suppression Mode, p. 214.

Before Starting Make sure the selected modulation mode is AM.

Procedure 1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the SBS Suppression menu and enter the SBS AM sub-menu.

The sub-menu appears and displays the current amplitude deviation.

Figure 166: TLS-50 Unit Setup – SBS AM sub-menu

3. Modify the value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digit to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new value.

The TLS-50 Unit Setup menu is automatically displayed and the amplitude deviation is set.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - SBS AM

ESCAPE

00.0 %

Amplitude Deviation :

Select a new value.

ENTER

Page 219: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

OSICS User Manual 219

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

14.5.6 Enabling/Disabling the SBS Suppression

Subject If the SBS function is activated, a S flag appears in the Info column of the System Status screen (see Figure 9, p. 30).

Procedure 1. Access the TLS-50 Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the SBS Suppression menu and enter the SBS On/Off sub-menu.

The sub-menu appears and displays the current state of the SBS Suppression.

Figure 167: TLS-50 Unit Setup – SBS Suppression sub-menu

3. Turn the rotary knob to put the cursor before the On setting, and press it to validate your selection.

The SBS Suppression is set to On and the Unit Setup menu is automatically displayed.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - SBS On/Off

ESCAPE

OffOn

SBS suppression :

Select a new value.

Page 220: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

220 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

14.6 Testing the Performance of the TLS-50 Module

Use the procedures described in this section to validate warranted specifications of the OSICS TLS-50 module, as needed.Notwithstanding performance tests described in this section, we recommend performing calibration of OSICS TLS-50 modules over one-year intervals (see section Asking for a Factory Recalibration, p. 222).

14.6.1 Preparing Performance Tests

Before Starting Make sure you have the following required equipment for conducting performance tests on OSICS TLS-50 modules:

The various types as well as the number of fiber-optic connectors needed may vary depending on the type of equipment used and the distance between the test equipment and the OSICS TLS-50 modular optical source under test.

Procedure 1. In the OSICS Mainframe, install the TLS-50 modules you want to test.

2. Turn-on the OSICS multifunction platform (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

3. Let the OSICS multifunction platform and other test equipments run continuously for 2 hours at room temperature (within the specified operating temperature range) to achieve proper equipment warm-up.

Equipment Minimum RequirementOptical power meter Wavelength range: 1500 nm to 1630 nmWavemeter Wavelength range: 1500 nm to 1630 nmOptical attenuator -Jumper (x2) FC/APC to FC/PC or to FC/APC

• Disable the laser output before connecting or disconnecting a fiber optic cable on the module (see section Enabling/Disabling the Light Source of All Modules, p. 37).

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 221: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

OSICS User Manual 221

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

14.6.2 Measuring the Absolute Wavelength Accuracy

Subject This test procedure enables you to verify that the actual measured wavelength stays within acceptable deviation range from the wavelength setting.

Before Starting • Make sure you have a wavemeter with ±1 pm resolution and a NIST-traceable calibration (see section Preparing Performance Tests, p. 220) and jumper.

• An optical attenuator is required depending on the maximum input power accepted by your wavemeter.

Procedure 1. Connect the OSICS TLS-50 module to the wavemeter.

2. Set the OSICS TLS-50 output power to 13 dBm (C-band model) or 10 dBm (L-band model).

3. Set the wavelength of the OSICS TLS-50 tunable laser to the minimum value of the wavelength range. The minimum wavelength depends on the TLS-50 model you are using (see section Specifying the Optical Emission Wavelength/Frequency, p. 210).

4. Set the frequency step to +100 GHz.

5. Enable the laser output by pressing the Enable button on the module front panel.6. Compute the wavelength deviation of the TLS-50 module and check that it is not

greater than ±1 GHz compared with the wavelength reading of the wavemeter.

The absolute wavelength accuracy is the deviation between the wavelength setting and the wavelength reading provided by the wavemeter at each measurement step.

7. Increment the TLS-50 wavelength setting one wavelength step (+100 GHz) and repeat measurement step 6. to check that the wavelength deviation is not greater than ±1.8 GHz.

8. Stop when you reach the TLS-50 maximum wavelength limit.

9. If either of the rated output power at the wavelength range ends cannot be achieved (within specification limits), the TLS-50 module requires servicing. Please contact your Yenista Optics customer support service (see section Contact Information, p. 4) immediately.

14.6.3 Checking the Wavelength Range and Output Power

Subject This procedure enables you to verify that the TLS-50 module can achieve the rated output power at both ends of the TLS-50 wavelength range.

Before Starting Make sure you have a powermeter with a NIST-traceable calibration (see section Preparing Performance Tests, p. 220) and jumper.

Procedure 1. Disable the TLS-50 laser output.

2. Connect the TLS-50 laser output to the optical power meter.

Page 222: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-50 Module

222 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

3. From the TLS-50 Unit Setup menu, set the wavelength of the TLS-50 tunable laser to the minimum wavelength that corresponds to the type and power option of the TLS-50 module you are using (see OSICS TLS-50 Technical Specifications).

4. Enable the TLS-50 module laser output by pressing the Enable button.

5. Set the output power value to the rated output power, as given in section OSICS TLS-50 Technical Specifications and check that the TLS-50 module can deliver the desired output power at the given wavelength.

6. Set the wavelength of the TLS-50 tunable laser to the maximum wavelength that corresponds to the type and power option of the TLS-50 module you are using (see section OSICS TLS-50 Technical Specifications) and check that the TLS-50 module can deliver the desired output power at the given wavelength.

7. If either of the rated output power at the wavelength range ends cannot be achieved (within specification limits), the TLS-50 module requires servicing. Please contact your Yenista Optics customer support service (see section Contact Information, p. 4) immediately.

14.6.4 Asking for a Factory Recalibration

Subject To externally verify the performance of the OSICS, refer to the procedure described in section Testing the Performance of the TLS-50 Module, p. 220. If the performance does not meet warranted specifications, a complete calibration of the instrument may be required.We recommend performing factory calibration of OSICS TLS-50 modules over one-year intervals.

Procedure Contact the Yenista Optics customer support service (see section Contact Information, p. 4) for a complete calibration, as some calibration steps require the adjustment of internal components including the laser head.

Page 223: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 223

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN15. Using the OSICS TLS-AG Module

This section explains how to operate the OSICS TLS-AG via the TLS-AG Unit Setup menu. For more details on how to install and initialize the module into the OSICS Mainframe, see section Installing and Connecting the OSICS Multifunction Platform, p. 21.

15.1 OSICS TLS-AG Presentation

The OSICS TLS-AG modules are tunable laser sources with high output power and low linewidth thanks to external cavity design. The wavelength can be tuned to any wavelength in C or L band. The TLS-AG module is particularly adapted to WDM transmission testing.

15.1.1 OSICS TLS-AG Technical Specifications

C Band L Band Wide C Band Wide L BandGrid Selection Adjustable to any wavelength grid (ITU-T 100 GHz, 50 GHz and arbitrary grid)

Frequency/Wavelength Range

196.25–191.5 THz1527.60–1565.5 nm

190.95–186.35 THz1570.01–1608.76 nm

196.275–191.125 THz1527.41–1568.57 nm

191.100–186.200 THz1568.77–1610.06 nm

Output Power 20 mW (+13 dBm)

Power Range (typ.) +6 to +13.5 dBm

Relative Frequency(Wavelength) Accuracy (typ.)

±0.5 GHz (±4 pm)

Absolute Frequency(Wavelength) Accuracy (typ.)*1

±1.5 GHz (±12 pm)

Frequency Setting Resolution 1 GHz

Frequency Fine Tuning Setting Resolution*2 1 MHz

Tuning Speed*3 20 s (typ. between two channels)

Power Stability*2 ±0.03 dB

Absolute Output Power Deviation Across Tuning Range

±0.2 dB

Instantaneous Linewidth (FWHM) < 100 kHz

SBS Suppression Ability*4 Yes

Side Mode Suppression Ratio 50 dB (typ.)

Page 224: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

224 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

15.1.2 OSICS TLS-AG Module Overview

Figure 168: TLS-AG Module – Front and Rear Panels

Front Panel On LED

This LED indicates that the module is turned on (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

Enable Button

This button enables you to turn on/off the laser inside the module. The green LED of the button is lit if the optical output is enabled, and the power value is displayed in the System Status screen (see section System Status Screen Description, p. 30).

SSER (typ.)*5 60 dB

RIN*6 < -150 dB/Hz

Operating Temperature Range +15 to +35 °C (+59 to +95 °F)

Optical Interface FC/APC connector on PMF PER > 20 dB

Control interface Status LED, laser on/off button

Dimensions (WxHxD) & Weight 35 x 130 x 250 mm (single slot), 1 kg

Laser Safety Class 1 M

*1: ±2.5 GHz (±22 pm) end of life.*2: At a constant temperature.*3: 30 s max, including power stabilization*4: At a dither frequency of 20.8 kHz and a FM p-p modulation width from 0 to 1000 MHz.*5: Integrated over a 0.1 nm band.*6: RIN within the range 100 MHz–20 GHz measured at 10 dBm output power with RBW = 30 kHz

C Band L Band Wide C Band Wide L Band

Openingfor air output

Top Notch

Bottom Notch

96-pin VMEConnector

On LED

Enable Button

Optical OutputConnector

Retaining Screw

Page 225: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

OSICS User Manual 225

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INOptical Output Connector

The standard output connector is a FC/APC type connector.For more details on the appropriate fiber type, see section OSICS TLS-AG Technical Specifications, p. 223.

The label located on the optical output connector cap indicates the location of the laser source.

Retaining Screw

The captive screw enables you to lock/unlock the module inserted into the slot of the OSICS Mainframe.

Rear Panel Notches

The top and bottom notches enable you to slide the module into the OSICS Mainframe slot (see section Installing an OSICS Module Into the OSICS Mainframe, p. 23).

96-pin VME Connector

The rear panel of the OSICS modules interfaces with the OSICS Mainframe backplane by means of the 96-pin VME connector.

Opening

The opening ensures proper flow of cooled air drawn from the OSICS Mainframe chassis.

Side Panel The labels affixed on the side panel of the OSICS TLS-AG are described in section OSICS Information and Warning Markings, p. 19.

Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

The OSICS Mainframe does not support "hot-swap" of modules; for more details, see section Handling OSICS Modules Into the OSICS Mainframe, p. 23).

WARNING

CAUTION

Page 226: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

226 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

15.2 Enabling/Disabling the Light Source of the Module

Subject You can enable/disable the light source in two different ways:• From the OSICS Mainframe Main Setup menu (see section Enabling/Disabling the

Light Source of All Modules, p. 37), you can simultaneously enable or disable the light source in all the modules installed in the OSICS Mainframe. This function overrides the enable/disable state individually set in all modules.

• From the TLS-AG front panel, you can individually enable or disable the light source of the module by using the Enable button as explained in the following procedure. This action does not affect the enable/disable state of other modules installed in the OSICS Mainframe.

For laser safety reasons, the light source in modules are disabled after OSICS initialization.

Before Starting Connect an optical device to the OSICS module (see section Connecting Optical Devices to OSICS Modules, p. 26).

Procedures Enabling the Light Source of the Module

• On the TLS-AG module front panel press the Enable button.

The green LED on the button is lit.In the System Status screen, the output power is displayed in the P field (in dBm or mW depending on the selected power unit) for the corresponding module.

Disabling the Light Source of the Module

• On the TLS-AG module front panel press the Enable button.

The green LED on the button is turned off.In the System Status screen, the message Disabled is displayed for the corresponding module.

• Do not enable the laser while there is no fiber connected to the optical output connector.

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 227: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

OSICS User Manual 227

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN15.3 Setting the Operating Mode

15.3.1 Selecting the Operating Mode

Subject The TLS-AG module can operate in three modes:• Grid Mode

In this mode, you can select a start frequency and spacing, and then select any channel of the defined grid.

• Minimum spacing for grid is 0.1 GHz.

• Detuning from any position on the grid is possible on a ±6 GHz range with 1 MHz resolution.

• Manual Mode

In this mode, you can select wavelength/frequency directly as in any tunable laser source.

• Resolution is 0.1 MHz / 1 pm.

• Detuning from any selected wavelength/frequency is possible on a ±6 GHz range with 1 MHz resolution.

• High Resolution Mode

In this mode, you can directly enter the requested wavelength/frequency.

• Resolution is 1 MHz / 0.01 pm

Procedure 1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Mode: menu and press the knob.

The mode sub-menu appears and displays the available modes.

Figure 169: TLS-AG Unit Setup – Set Switch Mode sub-menu

3. Turn the rotary knob to put the cursor before the wanted mode and press it to validate your selection.

4. The selected mode is activated and the corresponding TLS-AG Unit Setup menu (adapted to the selected mode) appears.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - Switch Mode

ESCAPE

Mode:

GridManualHigh Resolution

Select a new value.

Page 228: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

228 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

15.3.2 Setting the Grid

Subject If you have selected the Grid operating mode, you can set the frequency of the first channel and the spacing of channels as explained in the following procedure.

Procedure Setting the Grid First Channel

1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Grid menu and enter the First Channel sub-menu.

Figure 170: TLS-AG Unit Setup – Grid First Channel sub-menu

3. Modify the channel value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the value.

The TLS-AG Unit Setup menu is automatically displayed and the first channel of the grid is set.

Setting the Channel Spacing

1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Grid menu and enter the Spacing sub-menu.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - First channel

ESCAPE

191500.0 GHz ENTER

First channel :

Modify the value.

Page 229: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

OSICS User Manual 229

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

Figure 171: TLS-AG Unit Setup – Grid Spacing sub-menu

3. Modify the channel value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the value.

The TLS-AG Unit Setup menu is automatically displayed and the spacing of the grid is set.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - Spacing

ESCAPE

0012.5 GHz ENTER

Spacing :

Modify the value.

Page 230: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

230 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

15.4 Setting the Optical Output Power of the TLS-AG Module

Subject You can set the optical output power in two different ways:• From the OSICS Mainframe Main Setup menu, you can simultaneously set the

optical power of all modules to the same value: see section Setting the Optical Output Power of All Installed Modules, p. 38.

• From the TLS-AG Unit Setup menu, you can set the output power of the module individually as explained in the following procedure.

Before Starting • Set the optical power unit from the OSICS Mainframe Main Setup menu (see section Selecting the Power Unit for All Installed Modules, p. 38). You cannot set the power unit of an individual module.

Procedure 1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set Power menu and press the knob.

The Set Power menu appears and displays the current power setting.

Figure 172: TLS-AG Unit Setup – Set Power menu

3. Modify the power value as follows:

a. Turn the rotary knob to put the cursor under the digit of +/- sign you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new power value.

The Unit Setup menu is automatically displayed and the optical power value of the module is now modified (according to the module power limits).

Main

EXIT

1 2 3 4 5 6 7 8Unit Setup - Set Power

ESCAPE

Power :

P = +03.00 dBm ENTER

Modify the value.

Page 231: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

OSICS User Manual 231

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN15.5 Setting the Optical Emission Wavelength/Frequency

15.5.1 Specifying the Optical Emission Wavelength/Frequency

Subject The following procedure explains how to set the emission wavelength or optical frequency of the OSICS TLS-AG module according to its specifications. Even though the module lets you set wavelength values outside this range, performance of your module is guaranteed only within the wavelength range established by the module specifications.• In Manual mode, the resolution setting is 1 pm or 100 MHz. If needed, it is possible

to fine-tune the laser frequency with a 1 MHz resolution (see section Fine Tuning the Wavelength/Frequency, p. 235).

• In High Resolution mode, you can directly enter the wavelength/frequency with 1 MHz/0.01 pm resolution.

• In Grid mode, you can set the wanted grid channels (see section Setting the Grid, p. 228). The channel number is indicated in the Info column of the System Status screen (see Figure 9, p. 30).

Before Starting • Set the spectral unit from the OSICS Mainframe Main Setup menu (see section Selecting the Spectral Unit for all Installed Modules, p. 39). You cannot set the spectral unit of an individual module.

• If the OSICS TLS-AG module is set to Grid mode, set the wanted grid values, as explained in section Setting the Grid, p. 228.

Procedure 1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Turn the rotary knob to put the cursor before the Set λ/f menu and press the knob.

The Set λ/f menu appears and displays the emission wavelength or frequency.If the TLS-AG module is set to Grid mode, it also displays the corresponding current channel.

Figure 173: TLS-AG Unit Setup – Set Wavelength/Frequency menu (Example in Grid mode)

3. Modify the wavelength/frequency value (or channel value if the TLS-AG module is used in Grid mode) as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

Main

ENTER

1 2 3 4 5 6 7 8TLS Unit Setup - Set λ/F

ESCAPE

Channel:

CH = 01 (1565.701 nm)

Modify the value.

Page 232: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

232 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new wavelength value; or if the TLS-AG module is used in Grid mode, press the ENTER control button to go back to the Unit Setup menu.

The Unit Setup menu is automatically displayed and the wavelength/frequency value of the module is now modified (according to the module limits).

15.5.2 Step-adjusting the Wavelength/Frequency Value of the Module

Subject The following procedure explains how to modify the wavelength/frequency of the module step by step using the increment defined in the Set Step sub-menu.

The starting wavelength is the value originally defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 231).This section only applies to OSICS TLS-AG set to Grid or Manual mode (see section Setting the Operating Mode, p. 227).• In Grid mode, the step is a number of times the channel spacing.• In Manual mode, the step is a value going down to 1 pm or 100 MHz.

Procedure 1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Step menu and enter the Set Step sub-menu.

The Set Step sub-menu appears and displays the current value of the wavelength/frequency step in nm or GHz (or if the TLS-AG module is used in Grid mode, the channel step and the corresponding step in GHz).

Figure 174: TLS-AG Unit Setup – Set Step sub-menu

3. Modify the step value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new wavelength value; or if the TLS-AG module is used in Grid mode, press the ENTER control button to go back to the Unit Setup menu.

The Unit Setup menu is automatically displayed.

4. To increase or decrease the wavelength/frequency value defined in the Set λ/f menu (see section Specifying the Optical Emission Wavelength/Frequency, p. 231) step by

Main

ENTER

1 2 3 4 5 6 7 8TLS Unit Setup - Set Step

ESCAPE

Step in Channel(s):

dCH = 02 (df = 50 GHz)

Modify the value.

Page 233: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

OSICS User Manual 233

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INstep, expand the Step menu and put the cursor before the Step + or Step - sub-menus and press the knob to step-adjust the value.While the module is updating the wavelength/frequency, the message Moving… appears along with the new wavelength/frequency value (or channel if the TLS-AG module is set to Grid mode) after step-change. The laser takes around 20 s to stabilize, we recommend to wait at least 20 seconds to get an accurate measurement.

5. Repeat step 4. until the wanted wavelength/frequency (or channel if the TLS-AG module is set to Grid mode) is reached.

15.5.3 Scanning a Predefined Wavelength/Frequency Range

Subject This section only applies to OSICS TLS-AG modules set to Grid or Manual mode (see section Selecting the Operating Mode, p. 227).The Scan function enables you to scan a predefined wavelength range step-by-step, using the channel increment defined in the Step sub-menu to directly set the emission wavelength (or optical frequency) of the TLS-AG tunable laser.Starting from the first wavelength (the lower wavelength limit you have set), the emission wavelength increases incrementally (the wavelength increment is stored in the Step parameter) up to the last wavelength value (the upper wavelength limit you have set). During each scan interval, the OSICS TLS-AG module scans at a selected wavelength range for the duration specified by the Delay parameter.

15.5.3.1 Setting the Scan Parameters

Subject This section explains how to configure the scan mode by setting the following parameters:• The first and last channel (Grid mode) or Wavelength/Frequency (Manual mode) of

the scan• The value of the channel (Grid mode) or Wavelength/Frequency (Manual mode) step

that is used in the scan• The time between steps• The number of cycles of the scanThe laser takes around 20 s to stabilize, we recommend to wait at least 20 s to get an accurate measurement.

Procedure 1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Scan menu and enter the wanted setting menus:

• Start Channel sub-menu: to select the first channel (Grid mode) or Wavelength/Frequency (Manual mode) of the scan.

• End Channel sub-menu: to select the last channel (Grid mode) or Wavelength/Frequency (Manual mode) of the scan.

• Step sub-menu: to select the value of the channel (Grid mode) or Wavelength/Frequency (Manual mode) step that you want to use for the scan.

• Delay sub-menu: to select the duration of a scan step.

• Scan Number sub-menu: to select the number of scan cycles. If you select 0000, the OSICS TLS-AG will scan indefinitely.

Page 234: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

234 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

The sub-menu appears and displays the value of the corresponding scan parameter.

3. Modify the value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new value.

The TLS-AG Unit Setup menu is automatically displayed and the scan parameters are set.

15.5.3.2 Controlling the Scan

Subject This section explains how to start, stop and resume the scan.

Before Starting • Configure the scan parameters as explained in section Setting the Scan Parameters, p. 233.

The laser takes around 20 s to stabilize, we recommend to wait at least 20 s to get an accurate measurement.

Procedures Starting a New Scan

1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Scan menu and enter the Control sub-menu.

Figure 175: TLS-AG Unit Setup – Scan Control sub-menu

3. Turn the rotary knob to put the cursor before START and press it to validate your selection.

The Unit Setup menu is displayed and the scan starts. Real-time information on the scanning process is displayed in the System Status screen.

Stopping the Scan

1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - Scan Control

ESCAPE

Scan Control:

STARTRESUMESTOP

ENTER

Select a new value.

Page 235: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

OSICS User Manual 235

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN2. Expand the Scan menu and enter the Control sub-menu.

3. Turn the rotary knob to put the cursor before STOP and press it to validate your selection.

The scan stops at the wavelength/frequency reached when the Stop function was activated; the Unit Setup menu is displayed.

Resuming the Scan

1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Scan menu and enter the Control sub-menu.

3. Turn the rotary knob to put the cursor before RESUME and press it to validate your selection.

The Unit Setup menu is displayed and the scan starts at the last channel selected in the previous scan. Real-time information on the scanning process is displayed in the System Status screen.

15.5.4 Fine Tuning the Wavelength/Frequency

Subject This section only applies to OSICS TLS-AG modules set to Grid or Manual mode (see section Setting the Operating Mode, p. 227). In High Resolution mode, the Fine Tune function is automatically activated.You can use the Fine Tune function to directly adjust the emission wavelength (or optical frequency) of the TLS-AG tunable laser with high sensitivity. This is useful in some applications, for instance in coherent applications, where you need to fine-tune the wavelength with a resolution better than 1 pm or 100 MHz. The OSICS TLS-AG allows a fine tuning with a 1 MHz resolution on a ±6 GHz range.The fine tuning value acts as an offset, it will stay constant when you use the Set λ/f function. The display value on the overall screen is given by the sum of Set λ/f + Fine Tune value.Example:• Set wavelength: 1565.485 nm• Fine Tune: +3. 19 pm• Displayed value on overall screen: 1565.488 nm

If the Fine Tune function is activated, a F flag appears on the right side of the System Status screen (see Figure 9, p. 30).If the SBS function is also activated (see section Setting the SBS Suppression Mode, p. 237), the flag becomes B. If the Frequency Control Dither function is disabled (seesection Disabling the Frequency Control Dither Function, p. 238), the flag becomes D.

Procedure Setting the Fine Tune

1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Fine Tune menu and enter the Set Fine Tune sub-menu.

The Fine Tune sub-menu appears and displays the current de-tuning of the TLS-AG module.

Page 236: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

236 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

Figure 176: TLS-AG Unit Setup – Fine Tune sub-menu

3. Modify the de-tuning value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digits to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new step value.

The Unit Setup menu is automatically displayed.

Resetting the Fine Tune to 0

1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the Fine Tune menu and select the Reset Fine Tune sub-menu.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - Fine Tune

ESCAPE

dλ = +01.12 pm ENTER

Fine Tune :

Page 237: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

OSICS User Manual 237

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN15.6 Setting the SBS Suppression Mode

Subject The Stimulated Brillouin Scattering suppression mode allows you to add internal modulation to the source. You can select frequency deviation from 0 to 1 GHz.Frequency modulation requires the application of a dither signal to modulate the frequency of the laser. To broaden the effective linewidth of the laser and reduce the effects of Stimulated Brillouin Scattering in the optical fiber, the laser is modulated with a single frequency dither signal of 20.8 kHz.

15.6.1 Selecting the Optical Frequency Deviation

Subject Frequency modulation requires the application of a dither signal to modulate the frequency of the laser; to broaden the effective linewidth of the laser and reduce the effects of Stimulated Brillouin Scattering in the optical fiber.On OSICS TLS-AG, the laser is modulated with a single frequency dither signal of 20.8 kHz.You can select the optical frequency deviation from 0 to 1 GHz.

Procedure 1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the SBS Suppression menu and enter the SBS FM sub-menu.

The sub-menu appears and displays the current optical frequency deviation.

Figure 177: TLS-AG Unit Setup – SBS FM sub-menu

3. Modify the value as follows:

a. Turn the rotary knob to put the cursor under the digit you want to modify and press the knob to highlight it.

b. Turn the knob clockwise to increase the value or anticlockwise to decrease it and press the knob to validate the selected digit.

c. Perform steps a. and b. again with the other digit to modify.

d. Turn the rotary knob clockwise to put the cursor under ENTER and press the knob to validate the new value.

The TLS-AG Unit Setup menu is automatically displayed and the frequency deviation is set.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - SBS FM

ESCAPE

1.0 GHz ENTER

Frequency Deviation :

Select a new value.

Page 238: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

238 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

15.6.2 Enabling/Disabling the SBS Suppression

Subject The SBS Suppression mode allows you to add internal frequency modulation to the source.If the SBS function is activated, a S flag appears in the Info column of the System Status screen (see Figure 9, p. 30). If the Fine Tune function is also used (value not equal to zero), the flag is B.

Procedure 1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the SBS Suppression menu and enter the SBS On/Off sub-menu.

The sub-menu appears and displays the current state of the SBS Suppression.

Figure 178: TLS-AG Unit Setup – SBS Suppression sub-menu

3. Turn the rotary knob to put the cursor before the On setting, and press it to validate your selection.

The SBS Suppression is set to On and the Unit Setup menu is automatically displayed.

15.6.3 Disabling the Frequency Control Dither Function

Subject The OSICS TLS-AG tunable laser uses a dither tone at 888 Hz used to lock the emission wavelength onto its targeted value in order to get exceptional wavelength and power stability over time.For experiments requiring very low linewidth and no jitter on the frequency, you can disable this modulation as explained in the following procedure. If the Frequency Dither Control function is disabled, a D flag appears in the Info column of the System Status screen (see Figure 9, p. 30). Long-term wavelength stability could be slightly affected.In the Info column of the System Status screen, flags record the last action performed: if the flag was S or B, it becomes D; if the flag is D and SBS suppression is activated, then it becomes B.

Procedure 1. Access the TLS-AG Unit Setup menu (see section Accessing an OSICS Module Setup Menu, p. 33).

2. Expand the SBS Suppression menu and enter the Freq. Dither Control sub-menu.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - SBS On/Off

ESCAPE

OffOn

SBS suppression :

Select a new value.

Page 239: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

OSICS User Manual 239

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

INThe sub-menu appears and displays the current state of the frequency control dither function.

Figure 179: TLS-AG Unit Setup – Frequency Dither Control sub-menu

3. Turn the rotary knob to put the cursor before the wanted setting, and press it to validate your selection.

The function is deactivated and the Unit Setup menu is automatically displayed.

After warm-up, it takes around 10 seconds before the disable function is activated. Changing wavelength or power will automatically re-activate the function for around 10 seconds and then disable it again.

During warm-up time, the 10 s could go up to one minute.

Main

EXIT

1 2 3 4 5 6 7 8TLS Unit Setup - Freq. Control Dither

ESCAPE

OffOn

Freq. control dither :

Select a new value.

Page 240: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

240 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

15.7 Testing the Performance of the TLS-AG Module

Use the procedures described in this section to validate warranted specifications of the OSICS TLS-AG module, as needed.Notwithstanding performance tests described in this section, we recommend performing calibration of OSICS TLS-AG modules over one-year intervals (see section Asking for a Factory Recalibration, p. 242).

15.7.1 Preparing Performance Tests

Before Starting Make sure you have the following required equipment for conducting performance tests on OSICS TLS-AG modules:

The various types as well as the number of fiber-optic connectors needed may vary depending on the type of equipment used and the distance between the test equipment and the OSICS TLS-AG modular optical source under test.

Procedure 1. In the OSICS Mainframe, install the TLS-AG modules you want to test.

2. Turn-on the OSICS multifunction platform (see section Turning on the OSICS Multifunction Platform and Accessing the System Status Screen, p. 29).

3. Let the OSICS multifunction platform and other test equipments run continuously for 2 hours at room temperature (within the specified operating temperature range) to achieve proper equipment warm-up.

Equipment Minimum RequirementOptical power meter Wavelength range: 1500 nm to 1630 nmWavemeter Wavelength range: 1500 nm to 1630 nmOptical attenuator -Jumper (x2) FC/APC to FC/PC or to FC/APC

• Disable the laser output before connecting or disconnecting a fiber optic cable on the module (see section Enabling/Disabling the Light Source of All Modules, p. 37).

• Do not look directly into the fiber end of an optical cable connected to any optical output while the device is in use. This may cause permanent eye damage.

WARNING

Page 241: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

OSICS User Manual 241

BKRTLS-50

ATND

FBECL

T100SLD

SWT

TLS-AG

SWT

APC

MA

IN

15.7.2 Measuring the Absolute Wavelength Accuracy

Subject This test procedure enables you to verify that the actual measured wavelength stays within acceptable deviation range from the wavelength setting.

Before Starting • Make sure you have a wavemeter with ±1 pm resolution and a NIST-traceable calibration (see section Preparing Performance Tests, p. 240) and jumper.

• An optical attenuator is required depending on the maximum input power accepted by your wavemeter.

Procedure 1. Connect the OSICS TLS-AG module to the wavemeter.

2. Set the OSICS TLS-AG output power to 13 dBm.

3. Set the wavelength of the OSICS TLS-AG tunable laser to the minimum value of the wavelength range. The minimum wavelength depends on the TLS-AG model you are using (see section Specifying the Optical Emission Wavelength/Frequency, p. 231).

4. Set the frequency step to +100 GHz.

5. Enable the laser output by pressing the Enable button on the module front panel.Keep in mind that the wavelength/power tuning takes up to 20 seconds to be stabilized.

6. Compute the wavelength deviation of the TLS-AG module and check that it is not greater than ±1 GHz compared with the wavelength reading of the wavemeter.

The absolute wavelength accuracy is the deviation between the wavelength setting and the wavelength reading provided by the wavemeter at each measurement step.

7. Increment the TLS-AG wavelength setting one wavelength step (+100 GHz) and repeat measurement step 6. to check that the wavelength deviation is not greater than ±2.5 GHz.

8. Stop when you reach the TLS-AG maximum wavelength limit.

9. If either of the rated output power at the wavelength range ends cannot be achieved (within specification limits), the TLS-AG module requires servicing. Please contact your Yenista Optics customer support service (see section Contact Information, p. 4) immediately.

15.7.3 Checking the Wavelength Range and Output Power

Subject This procedure enables you to verify that the TLS-AG module can achieve the rated output power at both ends of the TLS-AG wavelength range.

Before Starting Make sure you have a powermeter with a NIST-traceable calibration (see section Preparing Performance Tests, p. 240) and jumper.

Procedure 1. Disable the TLS-AG laser output.

2. Connect the TLS-AG laser output to the optical power meter.

Page 242: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Using the OSICS TLS-AG Module

242 OSICS User Manual

BKR

TLS-

50AT

ND

FBEC

LT1

00SL

DSW

TTL

S-A

GSW

T A

PCM

AIN

3. From the TLS-AG Unit Setup menu, set the wavelength of the TLS-AG tunable laser to the minimum wavelength that corresponds to the type and power option of the TLS-AG module you are using (see OSICS TLS-AG Technical Specifications).

4. Enable the TLS-AG module laser output by pressing the Enable button.

5. Set the output power value to the rated output power, as given in section OSICS TLS-AG Technical Specifications and check that the TLS-AG module can deliver the desired output power at the given wavelength.

6. Set the wavelength of the TLS-AG tunable laser to the maximum wavelength that corresponds to the type and power option of the TLS-AG module you are using (see section OSICS TLS-AG Technical Specifications) and check that the TLS-AG module can deliver the desired output power at the given wavelength.

7. If either of the rated output power at the wavelength range ends cannot be achieved (within specification limits), the TLS-AG module requires servicing. Please contact your Yenista Optics customer support service (see section Contact Information, p. 4) immediately.

15.7.4 Asking for a Factory Recalibration

Subject To externally verify the performance of the OSICS, refer to the procedure described in section Testing the Performance of the TLS-AG Module, p. 240. If the performance does not meet warranted specifications, a complete calibration of the instrument may be required.We recommend performing factory calibration of OSICS TLS-AG modules over one-year intervals.

Procedure Contact the Yenista Optics customer support service (see section Contact Information, p. 4) for a complete calibration, as some calibration steps require the adjustment of internal components including the laser head.

Page 243: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 243

16. Performing Basic Maintenance Operations

User maintenance of the OSICS multifunction platform is limited to basic maintenance tasks that do not require removing the instrument case-cover or accessing any internal component of the instrument.

16.1 Cleaning the OSICS Multifunction Platform

Subject If the external cover of the OSICS multifunction platform becomes dirty or dusty, clean it by following the instruction below.

Before Starting Material needed: cleaning cloth slightly damped with an isopropyl alcohol liquid.

Procedure 1. Turn the OSICS Mainframe off (see section Turning Off the OSICS Multifunction Platform, p. 36) and unplug the power supply cable from the wall socket.

2. Gently swipe dirt and dust on the external cover of the OSICS multifunction platform, without applying excessive force onto it.

16.2 Cleaning Optical ConnectorsSubject To ensure measurement accuracy and prevent loss of optical power, you must verify that

optical connectors are clean every time you connect a fiber.Handle optical fiber with appropriate care and preserve the integrity of optical connectors by keeping them free of contamination.

Before Starting Make sure you have the following material: • Optical grade cleaning cotton swabs• Clean compressed air• Isopropyl alcohol

To avoid personal injury, never remove the protective cover of the chassis to perform servicing or maintenance operations. You must refer to your Yenista Optics service representative.WARNING

Do not use chemically active or abrasive materials to clean the OSICS multifunction platform.

CAUTION

IMPORTANT To reduce the need for cleaning, immediately replace protective caps on the optical connectors when not in use.

IMPORTANT Use only high quality cleaning supplies that are non-abrasive and leave no residue.

Page 244: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Performing Basic Maintenance Operations

244 OSICS User Manual

Procedure 1. Turn the OSICS Mainframe off (see section Turning Off the OSICS Multifunction Platform, p. 36) and unplug the power supply cable from the wall socket.

2. Gently clean the connector end, with the following instructions:

a. Hold the can of compressed air upright and spray the can into the air to purge any propellant.

b. Spray the clean compressed air on the connector to remove any loose particles or moisture.

c. Moisten a clean optical swab with isopropyl alcohol and lightly wipe the surfaces of the connector with gentle circular motion.

d. Spray the clean compressed air on the connector again to remove any loose particles or isopropyl alcohol.

e. Check that the connector is clean with a fiberscope (or similar).

16.3 Replacing the Line Fuse

Before Starting • Unplug the power cable.

• Make sure you have the following material:

• Appropriate power fuse (see section OSICS Mainframe Technical Specifications, p. 16).

• Small flat-blade screwdriver

Procedure 1. On the OSICS Mainframe rear panel (see Figure 2, p. 18), insert a small flat-blade screwdriver into the notch just below the fuse holder.

2. Use the notch to pull the fuse holder straight out to remove the fuse.

3. Replace the fuse in the fuse holder and snap the fuse holder back into its housing.

16.4 Packaging for Shipment

If you need to return the OSICS Mainframe or an OSICS module to Yenista Optics for servicing or factory calibration, use the original packaging.For instructions on returning the OSICS, please contact Yenista Optics (see section Contact Information, p. 4).

To avoid fire hazard, only use the correct fuse type, voltage and current ratings (see section OSICS Mainframe Technical Specifications, p. 16).WARNING

MAINS

Io Fuse holder

Release notch

Page 245: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 245

17. Troubleshooting

Error messages displayed on screen can be generated by the OSICS Mainframe and by any module installed in the OSICS Mainframe.

17.1 OSICS Mainframe Error and Warning Messages

This section describes error and warning messages that can be generated by the OSICS Mainframe.

Message Possible Cause Possible ResolutionInfinite Wait error(10)

The module is constantly busy. The OSICS Mainframe can no longer communicate with modules.

Reset the system as follows:1. Turn off the OSICS Mainframe

power by pressing the on/off button located on the front panel.

2. Turn on the OSICS Mainframe by pressing the on/off button again and wait until the Initializing… sequence is complete.

3. Enter the password and check for any error message.

If this error message appears again, please contact the Yenista Optics customer support service (see section Contact Information, p. 4).

Infinite Wait error(20)

The OSICS Mainframe cannot send data to modules.

Infinite Wait error(30)

The OSICS Mainframe can no longer get any response from modules.

Page 246: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Troubleshooting

246 OSICS User Manual

17.2 OSICS Modules Error and Warning Messages

This section describes error and warning messages that can be generated by the OSICS modules.

Module Message Possible Cause Possible ResolutionT100ECLDFBTLS-AGTLS-50

User error User-defined stored configuration data is corrupted or cannot be retrieved.

Turn off and on the OSICS multifunction platform to reset the OSICS module.If this error message appears again, please contact the Yenista Optics customer support service (see section Contact Information, p. 4).

T100ECLDFBTLS-AGTLS-50

Fact. error The factory-set calibration data is corrupted or cannot be retrieved.

Turn off and on the OSICS multifunction platform to reset the OSICS module.If this error message appears again, please contact the Yenista Optics customer support service (see section Contact Information, p. 4).

T100ECL

Moving timeout

The maximum time for the system to try moving the module internal mechanical components has been reached.

Turn off and on the OSICS multifunction platform to reset the module.

T100ECL

Switch error Mechanical problem.The protection actuator has been triggered.

Perform an internal referencing of the module as described in section Performing a Wavelength Reference Calibration, p. 198.If this error message appears again, please contact the Yenista Optics customer support service (see section Contact Information, p. 4).

T100ECL

Reference error

The module internal wavelength referencing operation has failed.

Turn off and on the OSICS multifunction platform to reset the module.

T100ECLDFB

LIM The maximum diode current has been reached. The desired power setting cannot be achieved and the system no longer operates at constant power.

Try to change the power setting and/or the operating wavelength to a value that yields a permissible diode current level.

DFB ERROR T Temperature control error Try to change the operating wavelength.Turn off and on the OSICS multifunction platform.

Page 247: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 247

Certification and Compliance

Specifications The OSICS Mainframe complies with the following specifications and standards:• CE Marking Directive

• EMC Directive 2004/108/EC: directive on the approximation of the laws of the member states relating to electromagnetic compatibility, by reference to the following standards: EN 61326-1 ed. 2006

• Low Voltage Directive (as amended by directive 2006/95/EC: Council Directive on the harmonization of the laws of Member States relating to electrical equipment designed for use within certain voltage limits).

• EN 61010-1 ed. 2001 Safety requirements for electrical equipment for measurement, control and laboratory use.

• EN 60825-1 ed. 2007 Safety of laser products.

• China RoHS

Environment Yenista Optics complies with the Waste Electrical and Electronic Equipment (WEEE) Directive 2002/96/EC and honors its obligations under this directive.

The product is marked with the WEEE symbol for recycling.This means that you must not discard the product with domestic household waste.

At the end of the product's useful life, please contact your local Yenista Optics representative to arrange disposal in accordance with your initial contract and the local law.Additionally, Yenista Optics arranges take back and disposal of historical waste as defined by the directives where Yenista Optics is supplying a new product or new equivalent product that fulfills the same function and replaces an old product (that is one placed on the market prior to August 2005).In all cases, should you require Yenista Optics to arrange such take back and disposal, please contact your local Yenista Optics representative.This obligation will be met by Yenista Optics for any equipment that is sold by Yenista Optics in the EU area irrespective of how many times it has been sold on.Yenista Optics will not be obligated for equipment that was originally sold outside the EU area and subsequently imported unless special conditions have been agreed previously between Yenista Optics and the user.

Page 248: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

Certification and Compliance

248 OSICS User Manual

Page 249: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

OSICS User Manual 249

List of Figures

ATN Unit Setup – Adjust Attenuation menu | Figure 25: ....................................................................................................................... 52ATN Unit Setup – Set Attenuation menu | Figure 24: .............................................................................................................................. 52ATN Unit Setup – Set Offset menu | Figure 28: ........................................................................................................................................... 55ATN Unit Setup – Set Step sub-menu | Figure 26:..................................................................................................................................... 53ATN Unit Setup – Set Wavelength menu | Figure 27: .............................................................................................................................. 54ATN-HP Configuration | Figure 22: ..................................................................................................................................................................... 49ATN-HP Module – Front and Rear Panels | Figure 23:............................................................................................................................. 50BKR Unit Setup – Adjust Attenuation menu | Figure 32: ........................................................................................................................ 60BKR Unit Setup – Set Attenuation menu | Figure 31: ............................................................................................................................... 59BKR Unit Setup – Set Offset menu | Figure 35: ............................................................................................................................................ 63BKR Unit Setup – Set Step sub-menu | Figure 33:...................................................................................................................................... 61BKR Unit Setup – Set Wavelength menu | Figure 34: ............................................................................................................................... 62BNC OUT1 signal level depending on wavelength (at 2 mW) | Figure 125:..............................................................................169DFB Module – Front and Rear Panels | Figure 36: ...................................................................................................................................... 67DFB Unit Setup – Adjust Power menu | Figure 38:..................................................................................................................................... 71DFB Unit Setup – Analog Out setting | Figure 61: ...................................................................................................................................... 91DFB Unit Setup – Modulation Control (Off) | Figure 55:......................................................................................................................... 84DFB Unit Setup – Modulation Control (On) | Figure 45: ......................................................................................................................... 77DFB Unit Setup – Modulation Control (On) | Figure 50: ......................................................................................................................... 80DFB Unit Setup – Modulation Control (On) | Figure 53: ......................................................................................................................... 82DFB Unit Setup – Modulation Source (Internal) | Figure 43: ................................................................................................................ 75DFB Unit Setup – Modulation Source (Mainframe) | Figure 49:......................................................................................................... 79DFB Unit Setup – Modulation Source (Mainframe) | Figure 52:......................................................................................................... 81DFB Unit Setup – Power calibration | Figure 63:.......................................................................................................................................... 93DFB Unit Setup – Set Frequency | Figure 44:................................................................................................................................................. 76DFB Unit Setup – Set Power menu | Figure 37: ............................................................................................................................................ 70DFB Unit Setup – Set Step sub-menu | Figure 40:...................................................................................................................................... 73DFB Unit Setup – Set Wavelength/Frequency menu | Figure 39:...................................................................................................... 72DFB Unit Setup – Sinus Modulation (Set Frequency) | Figure 58: ..................................................................................................... 87DFB Unit Setup – Sinus Modulation (Set Rate) | Figure 59: .................................................................................................................. 88DFB Unit Setup – Sinus Modulation Control | Figure 60: ....................................................................................................................... 89DFB Unit Setup – Wavelength calibration | Figure 62:............................................................................................................................. 92Diode current resulting from sinusoidal modulation | Figure 57:..................................................................................................... 86ECL Module – Front and Rear Panels | Figure 64:.....................................................................................................................................101ECL Unit Setup – Adjust Power menu | Figure 66:....................................................................................................................................105ECL Unit Setup – Analog Out setting | Figure 87:.....................................................................................................................................123ECL Unit Setup – Auto peak find menu | Figure 70: ................................................................................................................................109ECL Unit Setup – Coherence Control menu | Figure 69: .......................................................................................................................108ECL Unit Setup – Modulation Control (Off) | Figure 85:........................................................................................................................120ECL Unit Setup – Modulation Control (On) | Figure 75: ........................................................................................................................113ECL Unit Setup – Modulation Control (On) | Figure 80: ........................................................................................................................116ECL Unit Setup – Modulation Control (On) | Figure 83: ........................................................................................................................118ECL Unit Setup – Modulation Source (Internal) | Figure 73: ...............................................................................................................111ECL Unit Setup – Modulation Source (Mainframe) | Figure 79:........................................................................................................115ECL Unit Setup – Modulation Source (Mainframe) | Figure 82:........................................................................................................117ECL Unit Setup – Power calibration, power value at first wavelength | Figure 90:................................................................126ECL Unit Setup – Power calibration, power value at second wavelength | Figure 91: ........................................................127

Page 250: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

List of Figures

250 OSICS User Manual

ECL Unit Setup – Set Frequency | Figure 74:................................................................................................................................................112ECL Unit Setup – Set Power menu | Figure 65:...........................................................................................................................................104ECL Unit Setup – Set Step sub-menu | Figure 68:.....................................................................................................................................107ECL Unit Setup – Set Wavelength/Frequency menu | Figure 67:.....................................................................................................106ECL Unit Setup – λ/f calibration, first wavelength | Figure 88: ..........................................................................................................125ECL Unit Setup – λ/f calibration, second wavelength | Figure 89:...................................................................................................125External Analog Modulation Setup | Figure 148: ......................................................................................................................................193External Analog Modulation Setup | Figure 54: ........................................................................................................................................... 83External Analog Modulation Setup | Figure 84: .........................................................................................................................................119Main Setup – Modulation Frequency | Figure 141:..................................................................................................................................188Main Setup – Modulation Frequency | Figure 17: ...................................................................................................................................... 42Main Setup – Modulation Frequency | Figure 47: ...................................................................................................................................... 78Main Setup – Modulation Frequency | Figure 77: ....................................................................................................................................114Main Setup – Modulation Real Frequency | Figure 142:.......................................................................................................................189Main Setup – Modulation Real Frequency | Figure 18: ........................................................................................................................... 42Main Setup – Modulation Real Frequency | Figure 48: ........................................................................................................................... 79Main Setup – Modulation Real Frequency | Figure 78: .........................................................................................................................115Main Setup – Modulation Source (External) | Figure 145: ...................................................................................................................191Main Setup – Modulation Source (External) | Figure 19:........................................................................................................................ 43Main Setup – Modulation Source (External) | Figure 51:........................................................................................................................ 81Main Setup – Modulation Source (External) | Figure 81:......................................................................................................................117Main Setup – Modulation Source (Internal) | Figure 140: ....................................................................................................................188Main Setup – Modulation Source (Internal) | Figure 16:......................................................................................................................... 41Main Setup – Modulation Source (Internal) | Figure 46:......................................................................................................................... 78Main Setup – Modulation Source (Internal) | Figure 76:.......................................................................................................................114Main Setup – New Password | Figure 11: ........................................................................................................................................................ 35Main Setup – Power reference sub-menu | Figure 13: ............................................................................................................................ 39Main Setup – Save configuration | Figure 20: ............................................................................................................................................... 47Main Setup – Save configuration | Figure 21: ............................................................................................................................................... 48Main Setup – Set unit dBm/mW sub-menu | Figure 12:......................................................................................................................... 38Main Setup – Set unit GHz/nm sub-menu | Figure 14: ........................................................................................................................... 39Mainframe – Front Panel | Figure 1: ................................................................................................................................................................... 17Mainframe – Rear Panel | Figure 2: ..................................................................................................................................................................... 18Mainframe and DFB Module Modulation Scheme | Figure 41: .......................................................................................................... 74Mainframe and ECL Module Modulation Scheme | Figure 71: .........................................................................................................110Mainframe and T100 Module Modulation Scheme | Figure 135:....................................................................................................184Mainframe Modulation Scheme | Figure 15: ................................................................................................................................................. 40Module Installation – Module Insertion | Figure 5: ................................................................................................................................... 24Module Installation – Module Removal | Figure 6:.................................................................................................................................... 25Module Installation – Plate Removal | Figure 4: .......................................................................................................................................... 23OSICS BKR Module – Front and Rear Panels | Figure 30:....................................................................................................................... 58OSICS BKR Principle | Figure 29: ........................................................................................................................................................................... 57Password Screen | Figure 8:..................................................................................................................................................................................... 29Remote Interlock pin assignment | Figure 7: ................................................................................................................................................ 27Resulting Diode Current with External Analog Modulation Source | Figure 150:..................................................................195Resulting Diode Current with External Analog Modulation Source | Figure 56: ...................................................................... 85Resulting Diode Current with External Analog Modulation Source | Figure 86: ....................................................................121Retractable Legs | Figure 3:...................................................................................................................................................................................... 21Setup Menu – Examples | Figure 10: .................................................................................................................................................................. 33SLD Module – Front and Rear Panels | Figure 92: ....................................................................................................................................132SLD Unit Setup – Set Power menu | Figure 93: ..........................................................................................................................................135SWT APC Implementation Example – Sequential Testing of Two or Four Devices | Figure 110:..................................150SWT APC Implementation Example – Testing a DUT over a Wide Spectrum | Figure 112: .............................................151SWT APC Implementation Example – Testing of an Optical Amplifier | Figure 111:............................................................151

Page 251: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

List of Figures

OSICS User Manual 251

SWT APC Implementation Example – Two Laser Sources and a Wavemeter | Figure 109:..............................................150SWT APC in T100 Full-Band Configuration | Figure 108:......................................................................................................................150SWT APC Module – Front and Rear Panels | Figure 113:......................................................................................................................153SWT APC Performance Test – Crosstalk | Figure 127: ............................................................................................................................172SWT APC Performance Test – Repeatability | Figure 126:....................................................................................................................171SWT APC Reverse Configuration | Figure 107:............................................................................................................................................149SWT APC Setup – Channel Configuration menu (1x4 Example Screen) | Figure 115: .........................................................157SWT APC Setup – Channel Configuration menu (1x4 Example Screen) | Figure 116: .........................................................158SWT APC Standard Configuration | Figure 106:.........................................................................................................................................149SWT APC Unit Setup – Auto peak find menu | Figure 122:.................................................................................................................166SWT APC Unit Setup – Channel Selection menu (SWT APC 1x4 Example Screen) | Figure 114:...................................155SWT APC Unit Setup – Coherence Control menu | Figure 121:........................................................................................................165SWT APC Unit Setup – Modulation Control (On) | Figure 124:.........................................................................................................168SWT APC Unit Setup – Modulation Source | Figure 123: .....................................................................................................................167SWT APC Unit Setup – Scan Control sub-menu | Figure 120: ...........................................................................................................164SWT APC Unit Setup – Set Power menu | Figure 117:............................................................................................................................160SWT APC Unit Setup – Set Step sub-menu | Figure 119: .....................................................................................................................162SWT APC Unit Setup – Set Wavelength/Frequency menu | Figure 118: .....................................................................................161SWT Implementation Example – Sequential Testing of Two or Four Devices | Figure 97:...............................................138SWT Implementation Example – Testing of an Optical Amplifier | Figure 98:.........................................................................138SWT Implementation Example – Two Laser Sources and a Wavemeter | Figure 96: ...........................................................138SWT Module – Front and Rear Panels | Figure 99:...................................................................................................................................140SWT Performance Test – Crosstalk | Figure 105: .......................................................................................................................................148SWT Performance Test – Repeatability | Figure 104: ..............................................................................................................................147SWT Reverse Configuration | Figure 95:.........................................................................................................................................................137SWT Standard Configuration | Figure 94:......................................................................................................................................................137SWT Unit Setup – Channel Selection menu (SWT 1x4 Example Screen) | Figure 100: ........................................................142SWT Unit Setup – Sequence Switching - Control | Figure 103: ........................................................................................................145SWT Unit Setup – Sequence Switching - Number of Scans | Figure 101:...................................................................................143SWT Unit Setup – Sequence Switching - Set delay (SWT 1x4 Example Screen) | Figure 102: ........................................144System Status Screen | Figure 9:........................................................................................................................................................................... 30T100 Module – Front and Rear Panels | Figure 128:................................................................................................................................175T100 Unit Setup – Adjust Power menu | Figure 130: ..............................................................................................................................179T100 Unit Setup – Analog Out setting | Figure 151:................................................................................................................................197T100 Unit Setup – Auto peak find menu | Figure 134:...........................................................................................................................183T100 Unit Setup – Coherence Control menu | Figure 133:..................................................................................................................182T100 Unit Setup – Modulation Control (Off) | Figure 149: ..................................................................................................................194T100 Unit Setup – Modulation Control (On) | Figure 139:...................................................................................................................187T100 Unit Setup – Modulation Control (On) | Figure 144:...................................................................................................................190T100 Unit Setup – Modulation Control (On) | Figure 147:...................................................................................................................192T100 Unit Setup – Modulation Source (Internal) | Figure 137:..........................................................................................................185T100 Unit Setup – Modulation Source (Mainframe) | Figure 143: ..................................................................................................189T100 Unit Setup – Modulation Source (Mainframe) | Figure 146: ..................................................................................................191T100 Unit Setup – Power calibration, power value at first wavelength | Figure 154:...........................................................200T100 Unit Setup – Power calibration, power value at second wavelength | Figure 155: ...................................................201T100 Unit Setup – Set Frequency | Figure 138: ..........................................................................................................................................186T100 Unit Setup – Set Power menu | Figure 129: .....................................................................................................................................178T100 Unit Setup – Set Step sub-menu | Figure 132: ...............................................................................................................................181T100 Unit Setup – Set Wavelength/Frequency menu | Figure 131: ...............................................................................................180T100 Unit Setup – λ/f calibration, first wavelength | Figure 152:.....................................................................................................199T100 Unit Setup – λ/f calibration, second wavelength | Figure 153: .............................................................................................199TLS-50 – Amplitude Modulation | Figure 161: ............................................................................................................................................214TLS-50 Module – Front and Rear Panels | Figure 156: ...........................................................................................................................206TLS-50 Unit Setup – SBS AM sub-menu | Figure 166:............................................................................................................................218

Page 252: OSICS Multifunction Platform - Yonsei Universitytera.yonsei.ac.kr/PHP/measure_documents/OSICS_UM_3v2.2.pdf · 2017. 7. 25. · Manual Reference OSICS_UM_3v2.2 Typographical Conventions

List of Figures

252 OSICS User Manual

TLS-50 Unit Setup – SBS FM sub-menu | Figure 163:.............................................................................................................................216TLS-50 Unit Setup – SBS FM sub-menu | Figure 165:.............................................................................................................................217TLS-50 Unit Setup – SBS Modulation Mode sub-menu | Figure 162: ...........................................................................................215TLS-50 Unit Setup – SBS Suppression sub-menu | Figure 167: ........................................................................................................219TLS-50 Unit Setup – SBS Waveform sub-menu | Figure 164:.............................................................................................................217TLS-50 Unit Setup – Scan Control sub-menu | Figure 160:.................................................................................................................213TLS-50 Unit Setup – Set Power menu | Figure 157: .................................................................................................................................209TLS-50 Unit Setup – Set Step sub-menu | Figure 159:...........................................................................................................................211TLS-50 Unit Setup – Set Wavelength/Frequency menu | Figure 158: ...........................................................................................210TLS-AG Module – Front and Rear Panels | Figure 168:..........................................................................................................................224TLS-AG Unit Setup – Fine Tune sub-menu | Figure 176: ......................................................................................................................236TLS-AG Unit Setup – Frequency Dither Control sub-menu | Figure 179:....................................................................................239TLS-AG Unit Setup – Grid First Channel sub-menu | Figure 170: ....................................................................................................228TLS-AG Unit Setup – Grid Spacing sub-menu | Figure 171: ...............................................................................................................229TLS-AG Unit Setup – SBS FM sub-menu | Figure 177: ...........................................................................................................................237TLS-AG Unit Setup – SBS Suppression sub-menu | Figure 178: .......................................................................................................238TLS-AG Unit Setup – Scan Control sub-menu | Figure 175: ...............................................................................................................234TLS-AG Unit Setup – Set Power menu | Figure 172:................................................................................................................................230TLS-AG Unit Setup – Set Step sub-menu | Figure 174:..........................................................................................................................232TLS-AG Unit Setup – Set Switch Mode sub-menu | Figure 169: ......................................................................................................227TLS-AG Unit Setup – Set Wavelength/Frequency menu (Example in Grid mode) | Figure 173:....................................231TTL-Modulated Output Power Signal (DFB Module) | Figure 42: ..................................................................................................... 75TTL-Modulated Output Power Signal (ECL Module) | Figure 72: ....................................................................................................111TTL-Modulated Output Power Signal (T100 Module) | Figure 136: ...............................................................................................185