Int J Clin Exp Med 2015;8(11):20413-20423 www.ijcem.com /ISSN:1940-5901/IJCEM0015075 Original Article The clinical performance evaluation of novel protein chips for eleven biomarkers detection and the diagnostic model study Yuan Luo 1* , Xu Zhu 2* , Pengjun Zhang 2 , Qian Shen 3 , Zi Wang 1 , Xinyu Wen 4 , Ling Wang 4 , Jing Gao 4 , Jin Dong 4 , Caie Yang 1 , Tangming Wu 5 , Zheng Zhu 5 , Yaping Tian 1 1 Core Laboratory of Translational Medicine, State Key Laboratory of Kidney Disease, Chinese PLA General Hos- pital, Beijing, P. R. China; 2 Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/ Beijing), Interventional Therapy Department, Peking University Cancer Hospital&Institute, Beijing, P. R. China; 3 Department of Laboratory Diagnosis, Changhai Hospital, The Second Military Medical University, Shanghai, P. R. China; 4 Department of Clinical Biochemistry, Chinese PLA General Hospital, Beijing, P. R. China; 5 Jiangsu Sunlant Bioengineering Co., Ltd, Wuxi, P. R. China. * Co-first authors. Received August 25, 2015; Accepted October 25, 2015; Epub November 15, 2015; Published November 30, 2015 Abstract: We aimed to develop and validate two novel protein chips, which are based on microarray chemilumines- cence immunoassay and can simultaneously detected 11 biomarkers, and then to evaluate their clinical diagnostic value by comparing with the traditional methods. Protein chips were evaluated for limit of detection, specificity, common interferences, linearity, precision and accuracy. 11 biomarkers were simultaneously detected by traditional methods and protein chips in 3683 samples, which included 1723 cancer patients, 1798 benign diseases patients and 162 healthy controls. After assay validation, protein chips demonstrated high sensitivity, high specificity, good linearity, low imprecision and were free of common interferences. Compared with the traditional methods, protein chips have good correlation in the detection of all the 13 kinds of biomarkers (r≥0.935, P<0.001). For specific can- cer detection, there were no statistically significant differences between the traditional method and novel protein chips, except that male protein chip showed significantly better diagnostic value on NSE detection (P=0.004) but sig- nificantly worse value on pro-GRP detection (P=0.012), female chip showed significantly better diagnostic value on pro-GRP detection (P=0.005). Furthermore, both male and female multivariate diagnostic models had significantly better diagnostic value than single detection of PGI, PG II, pro-GRP, NSE and CA125 (P<0.05). In addition, male models had significantly better diagnostic value than single CA199 and free-PSA (P<0.05), while female models observed significantly better diagnostic value than single CA724 and β-HCG (P<0.05). For total disease or cancer detection, the AUC of multivariate logistic regression for the male and female disease detection was 0.981 (95% CI: 0.975-0.987) and 0.836 (95% CI: 0.798-0.874), respectively. While, that for total cancer detection was 0.691 (95% CI: 0.666-0.717) and 0.753 (95% CI: 0.731-0.775), respectively. The new designed protein chips are simple, multiplex and reliable clinical assays and the multi-parameter diagnostic models based on them could significantly improve their clinical performance. Keywords: Cancer biomarkers, protein array analysis, diagnostic tests, early detection of cancer, logistic models, ROC curve Introduction Nowadays, remarkable progresses have been made in understanding the molecular and cel- lular events that transform a normal function- ing cell into a malignant one [1, 2]. Based on these progresses, cancer detection has entered into molecular diagnostic phase and the types of biomarkers are rapidly expanding, including proteins, peptides, hormones, certain DNA, microRNAs, and so on [3, 4]. Although it’s very significant to focus on the development of improved therapies for cancer patients, it’s still important to invent multiplex early detection methods, which could scent precancerous changes with high specificity and sensitivity [5]. Finding cancer at its earliest, most treatable stage gives patients the best chance to survive. Common methods recommended by American Cancer Society (ACS) for cancer early detection

Upload: others

Post on 24-Dec-2021

5 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Original Article The clinical performance evaluation of

Int J Clin Exp Med 2015;8(11):20413-20423www.ijcem.com /ISSN:1940-5901/IJCEM0015075

Original ArticleThe clinical performance evaluation of novel protein chips for eleven biomarkers detection and the diagnostic model study

Yuan Luo1*, Xu Zhu2*, Pengjun Zhang2, Qian Shen3, Zi Wang1, Xinyu Wen4, Ling Wang4, Jing Gao4, Jin Dong4, Caie Yang1, Tangming Wu5, Zheng Zhu5, Yaping Tian1

1Core Laboratory of Translational Medicine, State Key Laboratory of Kidney Disease, Chinese PLA General Hos-pital, Beijing, P. R. China; 2Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Interventional Therapy Department, Peking University Cancer Hospital&Institute, Beijing, P. R. China; 3Department of Laboratory Diagnosis, Changhai Hospital, The Second Military Medical University, Shanghai, P. R. China; 4Department of Clinical Biochemistry, Chinese PLA General Hospital, Beijing, P. R. China; 5Jiangsu Sunlant Bioengineering Co., Ltd, Wuxi, P. R. China. *Co-first authors.

Received August 25, 2015; Accepted October 25, 2015; Epub November 15, 2015; Published November 30, 2015

Abstract: We aimed to develop and validate two novel protein chips, which are based on microarray chemilumines-cence immunoassay and can simultaneously detected 11 biomarkers, and then to evaluate their clinical diagnostic value by comparing with the traditional methods. Protein chips were evaluated for limit of detection, specificity, common interferences, linearity, precision and accuracy. 11 biomarkers were simultaneously detected by traditional methods and protein chips in 3683 samples, which included 1723 cancer patients, 1798 benign diseases patients and 162 healthy controls. After assay validation, protein chips demonstrated high sensitivity, high specificity, good linearity, low imprecision and were free of common interferences. Compared with the traditional methods, protein chips have good correlation in the detection of all the 13 kinds of biomarkers (r≥0.935, P<0.001). For specific can-cer detection, there were no statistically significant differences between the traditional method and novel protein chips, except that male protein chip showed significantly better diagnostic value on NSE detection (P=0.004) but sig-nificantly worse value on pro-GRP detection (P=0.012), female chip showed significantly better diagnostic value on pro-GRP detection (P=0.005). Furthermore, both male and female multivariate diagnostic models had significantly better diagnostic value than single detection of PGI, PG II, pro-GRP, NSE and CA125 (P<0.05). In addition, male models had significantly better diagnostic value than single CA199 and free-PSA (P<0.05), while female models observed significantly better diagnostic value than single CA724 and β-HCG (P<0.05). For total disease or cancer detection, the AUC of multivariate logistic regression for the male and female disease detection was 0.981 (95% CI: 0.975-0.987) and 0.836 (95% CI: 0.798-0.874), respectively. While, that for total cancer detection was 0.691 (95% CI: 0.666-0.717) and 0.753 (95% CI: 0.731-0.775), respectively. The new designed protein chips are simple, multiplex and reliable clinical assays and the multi-parameter diagnostic models based on them could significantly improve their clinical performance.

Keywords: Cancer biomarkers, protein array analysis, diagnostic tests, early detection of cancer, logistic models, ROC curve

Introduction

Nowadays, remarkable progresses have been made in understanding the molecular and cel-lular events that transform a normal function-ing cell into a malignant one [1, 2]. Based on these progresses, cancer detection has entered into molecular diagnostic phase and the types of biomarkers are rapidly expanding, including proteins, peptides, hormones, certain

DNA, microRNAs, and so on [3, 4]. Although it’s very significant to focus on the development of improved therapies for cancer patients, it’s still important to invent multiplex early detection methods, which could scent precancerous changes with high specificity and sensitivity [5]. Finding cancer at its earliest, most treatable stage gives patients the best chance to survive. Common methods recommended by American Cancer Society (ACS) for cancer early detection

Page 2: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

20414 Int J Clin Exp Med 2015;8(11):20413-20423

include many types, such as microbial detec-tion, fecal occult blood test, digital rectal exam-ination, double-contrast barium enema, com-puted tomography, tumor markers test, and so on [6]. Among these, the detection of tumor markers, which are directly produced or indi-rectly induced by the tumor cell and could reflect its growth and/or activity, is one of the most convenient means to monitor the insignifi-cant or potential cancer transformation [7].

However, tumor markers are not specific to can-cer as their name looks like and the majority of them could be synthesized and released by normal cells, too. Consequently, the specificity of tumor markers is not their presence at tumor sites but the concentrations detected there, which is why normal values are needed [8]. Moreover, present clinical proven methods, such as radioimmunoassay or chemilumines-cence, could only detect one biomarker at a time. In other words, they don’t observe high-throughput abilities, which result in low working efficiency and suppress their use in mass screening. On the contrary, protein array tech-nology possesses the greatest potential for screening thousands of recombinant-, frac-tions- or purified proteins, as well as synthetic peptides and even unknown proteins in a high throughput [9, 10]. Furthermore, since various benign pathologies could also produce tumor markers, which give rise to false positives, sin-gle tumor marker has relatively low specificity and sensitivity and might sometimes shows equivocal results. Therefore, how to improve the detection efficiency and increase the speci-ficity and sensitivity is the most impor- tant aspect of diagnostic methods [11-12]. To break through this bottleneck, using multi-

or woman, respectively. Specifically, Pepsinogen I (PGI), Pepsinogen II (PGII), pro-Gastrin-releas-ing peptide (pro-GRP), alpha-fetoprotein (AFP), Neuron-specific enolase (NSE), Carcino-embryonic antigen (CEA), Carbohydrate antigen 125 (CA125), Carbohydrate antigen 199 (CA199) and Carbohydrate antigen 724 (CA724) are common in both male and female chips. Meanwhile, total-Prostate specific antigen (t-PSA) and free-Prostate specific antigen (f-PSA) are unique for male chip, beta-Human chorionic gonadotropin (β-HCG) and Car- bohydrate antigen 153 (CA153) are unique for female chip. Then, protein chips were evaluat-ed for limit of detection, specificity, common interferences, linearity, precision and accuracy. After that, diagnostic models were established to improve the detection specificity and sensi-tivity at the same time. We aimed to evaluate the clinical performance of the protein chips and the feasibility of using multiple biomarkers to establish diagnostic models for cancer detection.

Patients and methods

Patients and healthy controls

Our study was approved by ethics committee of the Chinese PLA General Hospital, Shanghai Changzheng Hospital, Nanjing general hospital of Nanjing military command and the second affiliated hospital of Zhejiang university school of medicine and has been performed in accor-dance with the ethical standards laid down in an appropriate version of the 2000 Declaration of Helsinki as well as the Declaration of Istanbul

Table 1. Characteristics of total studied subjects

Characteristics Malignant tumor(n=1723)

Benign diseases(n=1798)

Normal control(n=162)

Age (years) Mean 57.70 57.79 39.53 Range 9-91 9-97 13-54Sex, n (%) Male 837 (48.6) 822 (45.7) 97 (59.9) Female 886 (51.4) 976 (54.3) 65 (40.1)Tissue distribution, n (%) Lung 522 (30.3) 152 (8.5) - Gastrointestinal 419 (24.3) 183 (10.2) - Liver and gall 192 (11.1) 226 (12.6) - Others 590 (34.2) 1237 (68.8) -

ple biomarkers or indica-tors to establish diagno- stic models is a common effective strategy, which could integrate the advan-tages of each index and simultaneously improve bo- th sensitivity and specificity [13-15].

In our study, we have devel-oped two protein chips, which are based on micro-array chemiluminescence immunoassay and could simultaneously detect 11 kinds of biomarkers of man

Page 3: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

20415 Int J Clin Exp Med 2015;8(11):20413-20423

Figure 1. Detection correlation between conventional method and protein chips. New- represents results detected by protein chips, while old- represents results detected by conventional method.

Page 4: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

20416 Int J Clin Exp Med 2015;8(11):20413-20423

Figure 2. Receiver operating characteristic curve analysis of biomarkers detected by conventional methods, male chip or multivariate diagnostic models for evaluat-ing the diagnostic value of specific cancer prediction. *P is the p value of z statistic compared between conventional methods and male chip, while #p is the p value of that between conventional method and multivariate diagnostic model.

Page 5: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

20417 Int J Clin Exp Med 2015;8(11):20413-20423

Figure 3. Receiver operating characteristic curve analysis of biomarkers detected by old methods, female chip or multivariate diagnostic models for evaluating the diagnostic value of specific cancer prediction. *P is the p value of z statistic compared between conventional methods and female chip, while #p is the p value of that between conventional method and multivariate diagnostic model.

Page 6: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

20418 Int J Clin Exp Med 2015;8(11):20413-20423

2009. With written consent, the serum sam-ples were collected and detected in four hospi-tals mentioned above. All the four hospitals used the same batch protein chip detection reagent and machine to ensure that the results were reliable and comparable. 1723 cancer patients, 1798 benign disease patients and 162 healthy controls were enrolled in the study. Both of the cancer and benign patients’ sam-ples were collected before any treatment, such as surgery, chemotherapy or radiation therapy, and further confirmed by histopathological analysis. Healthy controls were enrolled based on their negative results on the blood biomark-er test, x-ray, ultrasound, computed tomogra-phy examination and fecal occult blood testing. The clinical characteristics of the participants were shown in Table 1.

Serum collection

A total of 5 ml of peripheral blood samples was collected in tubes containing a separating gel and clot activator. After blood collection, the tubes should be upside down ten times quickly and then rest for about 1 hour. After centrifug-ing at 2500 rpm for 5 min, the samples could be directly detected by corresponding tech-niques. If it’s not convenient to do the test at once, the supernatant could be transferred into new tubes, and the serum should be stored in aliquots at -80°C until detection.

Measurement of serum biomarker levels using novel protein chips

The microarray-based sandwich immunoassay kits (Sunlant Bioengineering, Wuxi, China), which are cooperated with protein chips, were used to detect the serum concentrations of cancer related biomarkers. Briefly, serum sam-ple and microarray that was covalently coupled to specific capture antibodies were incubated together. After washing, a cocktail of horserad-ish peroxidase-labeled antibodies were added to the array, which was shaken at room tem-perature. Then, the array was transferred to a cuvette with substrate solution and detected by an automatic biochip reader. The results were recorded by standard curves established by five different concentrations.

Measurement of serum biomarker levels using traditional detection methods

PGI, PGII and pro-GRP were detected by an Abbott i2000 instrument (Abbott Laboratories,

USA). AFP, NSE, CEA, CA125, CA199, CA724, PSA, f-PSA, β-HCG and CA153 were detected by Roche Cobas e601 instrument (Roche Diagnostics, Germany). The main principle of these methods is competitive protein binding (CPB), however they have different detection methods: electrochemiluminescence detection in Roche Cobas e601 and chemiluminescence detection in Abbott i2000. Detection was con-ducted according to the manufacturer’s reagent kit manual for specific biomarkers.

Evaluation of detection sensitivity and specific-ity of protein chips

For sensitivity evaluation, the experiment to determine the limit of detection (LOD) was per-formed on 3 separate occasions, as follows: the zero concentration calibrator was evaluat-ed for 15 replicates, respectively. The average and 2SDs (Standard deviation) of the determi-nations was used to calculate LOD. For specific-ity evaluation, 50 samples of patients with high expression of various biomarkers were detect-ed by protein chips to validate their detection specificity. In addition, 30 samples from hemo-lysis, hyperlipidemia or jaundice were used to validate the detection stability.

Evaluation of detection linearity of protein chips

For each biomarker, linearity was evaluated using Abbott or Roche calibrators in triplicate measurements. Linearity was also evaluated using a patient sample with high concentration of specific biomarker. This sample was diluted with blank whole blood at different proportions to achieve 7 dilutions. These diluted specimens were each examined in triplicate.

Evaluation of precision performance of protein chips

According to NCCLS EP5-A program [16], both intra- and inter-batch examination used the same machine and standard curve but three reagent box with different lot number. For the former, repeat detecting two internal quality controls for 20 times in one day and then calcu-lating the intra-batch precision. For the latter, repeat detecting one internal quality control for 2 times in one day and continuously monitoring for 20 days and then calculating the inter-batch precision.

Page 7: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

20419 Int J Clin Exp Med 2015;8(11):20413-20423

Statistical analyses

Statistical analyses were carried out using SPSS 22.0 (SPSS Inc., USA) and MedCalc 13.0 (MedCalc Software, Belgium). Pearson’s corre-lation coefficient was used for correlation anal-ysis of the associated quantitative data. Kappa test was used to detect the diagnostic consis-tency between methods. Receiver operating characteristic (ROC) curves were established to evaluate the diagnostic value, and the Areas under the ROC curves (AUC) were compared by z-scores test. The cutoff value, which deter-mined the sensitivity and specificity, was cho-sen by the Youden index. Binary logistic regres-sion models were built to evaluate the multivariate diagnostic value of multiple bio-markers detected by protein chips. A two-tailed p-value of less than 0.05 was considered sta-tistically significant.

Results

Evaluation of detection sensitivity and specific-ity of protein chips

The LOD for protein chips were defined here as the determination at 2SDs above the zero con-centration calibrators. Based on 3 separate analyses, the average LOD for PGI, PGII, pro-GRP, AFP, NSE, CEA, CA125, CA199, CA724, t-PSA, f-PSA, β-HCG and CA153 was 4 ng/ml, 1.5 ng/ml, 0.02 ng/ml, 2 ng/ml, 2.5 ng/ml, 0.5 ng/ml, 2.3 U/ml, 5 U/ml, 1 U/ml, 0.1 ng/ml, 0.1 ng/ml, 0.4 ng/ml and 5 U/ml. Meanwhile, pro-tein chips could specifically detect every bio-marker on the chips without cross-reactivity. Additionally, interference potential was evalu-ated by determining biomarker concentration for samples from hemolysis, hyperlipidemia or jaundice. The results showed that severe hemo-lysis samples are prone to lead false positive of NSE index, while samples from hyperlipidemia or jaundice have no impacts on the chip results (data not shown). It suggested that protein chips had comparable detection sensitivity and specificity with traditional methods and could meet the basic requirements of clinical use.

Evaluation of detection linearity of protein chips

After linearity evaluation with Abbot or Roche calibrators and serial dilutions of patient sam-ples, protein chips were found to be linear to detect PGI, PGII, pro-GRP, AFP, NSE, CEA, CA125, CA199, CA724, t-PSA, f-PSA, β-HCG

and CA153 in the range of 6-300 ng/ml, 2-200 ng/ml, 0.03-5 ng/ml, 3-1000 ng/ml, 2.6-300 ng/ml, 1-300 ng/ml, 3.5-3000 U/ml, 7-2000 U/ml, 1.35-400 U/ml, 1-300 ng/ml, 0.2-100 ng/ml, 0.6-200 ng/ml and 7-300 U/ml. For every biomarker, the correlation coefficient of linearity was higher than 0.99 (r≥0.99). It sug-gested that protein chip had good detection linearity and wide linear range.

Evaluation of precision performance of novel protein chips

In order to evaluate the precision performance, protein chips for male and female were respec-tively used to do the intra- and inter-batch examination. As shown in Supplementary Tables 1, 2, 3 and 4, coefficient of variation (%CV) of male chip in intra-batch examination ranged from 2.4% to 10.0%, %CV of male chip in inter-batch examination ranged from 5.0% to 9.1%, %CV of female chip in intra-batch exami-nation ranged from 1.89% to 9.95%, %CV of female chip in inter-batch examination ranged from 4.66% to 9.56%. It suggested that both male and female protein chips have good detection precision in intra- and inter-batch. In other words, protein chips are reliable detec-tion methods and had good repeatability and stability.

Correlation analysis between the traditional methods and novel protein chips

All kinds of biomarkers were detected by both the traditional methods and novel protein chips, and then the same biomarkers from both chips were analyzed together. As showed in Figure 1, the correlation coefficient for PGI, PGII, pro-GRP, AFP, NSE, CEA, CA125, CA199, CA724, t-PSA, f-PSA, β-HCG and CA153 was 0.935, 0.947, 0.988, 0.985, 0.969, 0.980, 0.980, 0.980, 0.978, 0.979, 0.982, 0.981 and 0.966. And all the p values for them were 0.000. It suggested that protein chips have good correlation and detection consistency with the traditional method in the detection of all the 13 kinds of biomarkers.

Diagnostic value of the 11 biomarkers detect-ed by traditional methods and novel protein chips for specific cancers detection

After the assay validation and correlation analy-sis, we examined the diagnostic ability of the

Page 8: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

20420 Int J Clin Exp Med 2015;8(11):20413-20423

biomarkers depending on their clinical utility. Generally, PGI, PGII, CEA, CA199 and CA724 were used for gastrointestinal tumors detec-tion, while pro-GRP and NSE for lung cancer, AFP for liver cancer, t-PSA and f-PSA for pros-tate cancer, β-HCG for gynecologic tumor, CA153 for female breast cancer, and CA125 for all cancers. The diagnostic value of the two methods on specific cancer screening was eval-uated by ROC curves and their AUCs were com-pared by z-score test. Benefit from the multi-plex detection ability of protein chips, we further established binary logistic regression models to evaluate their multivariate diagnostic value

in specific cancer screening. The diagnostic models and their specificity (SPC) and sensitiv-ity (SEN) were listed in Supplementary Table 5.

As shown in Figures 2, 3, there were no statisti-cally significant differences between the tradi-tional method and novel protein chips, except that male protein chip showed significantly bet-ter diagnostic value on NSE detection (P=0.004) but significantly worse value on pro-GRP detec-tion (P=0.012), female chip showed significant-ly better diagnostic value on pro-GRP detection (P=0.005). Furthermore, compared with tradi-tional methods, multivariate diagnostic models

Figure 4. Receiver operating characteristic curve analysis of multivariate diagnostic models for evaluating the diag-nostic value of disease or cancer differentiation. The sensitivity and specificity were determined by the maximum Youden index. AUC: Areas under the curve.

Page 9: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

20421 Int J Clin Exp Med 2015;8(11):20413-20423

established on chip results had significantly better diagnostic value as a whole. Specifically, both male and female models had significantly better diagnostic value than single detection of PGI, PG II, pro-GRP, NSE and CA125 (P<0.05). In addition, male models had significantly bet-ter diagnostic value than single CA199 and free-PSA (P<0.05), while female models observed significantly better diagnostic value than single CA724 and β-HCG (P<0.05). It sug-gested that protein chips are comparable with traditional methods in single biomarker detec-tion manner. Once establishing diagnostic models based on multiplex chip results, it observed significantly higher diagnostic value than single detection of 7 biomarkers in 11. It showed that protein chips could not only increase the detection efficiency, but also improve the diagnostic value a lot.

Multivariate diagnostic value of protein chips for total disease or cancer detection

Similarly, based on the multiplex detection abil-ity of protein chips, we have established binary logistic regression models to evaluate their multivariate diagnostic value in total disease or cancer detection. Specifically, in total disease identification, both the cancer and benign dis-ease patients were considered as “disease positive” group, while the normal control as “disease negative” group. Meanwhile, in the case of total cancer identification, the cancer patients were set as “cancer positive” group, while the benign disease patients as “cancer negative” group. The diagnostic models and their SPC and SEN were listed in Supplementary Table 6. As shown in Figure 4, the AUC of multi-variate logistic regression for the male and female disease detection was 0.981 (95% CI: 0.975-0.987) and 0.836 (95% CI: 0.798-0.874), respectively. While, that for cancer detection was 0.691 (95% CI: 0.666-0.717) and 0.753 (95% CI: 0.731-0.775), respectively. The results preliminarily indicated that protein chips might have a certain diagnostic value in the detection of total diseases or tumors.

Discussion

Cancer, which is one of the most common causes of disease-associated death, is a major health issue in the world [17]. Fortunately, a substantial proportion of cancers, which are caused by cigarette smoking, heavy use of alcohol or other bad habits, could be prevented

if people could reduce such harmful exposure. Moreover, some cancers could even be cured if people could scent precancerous changes and get diagnosed at an early stage, such as can-cers of the breast, cervix, colon, and rectum [18-20]. Therefore, early detection and timely diagnosis is the key point of cancer prevention and control, which also predicts good clinical outcomes. Unexpectedly, although developed countries could provide the best services for detection, diagnosis, and treatment, they have the highest incidence of cancer. The highest prevalence proportions of cancer also occur in these populations. This could be explained by the following three reasons. Firstly, developed countries observe high level of medical tech-nology, which could detect more precancerous patients at early stages, while these patients often miss their survival chances in developing countries. Secondly, most developed countries are aging society, which has a higher incidence of cancer. At last, higher incidence of cancer often means that other diseases, such as infec-tious ones, are less. So, this unexpected inci-dence level doesn’t mean that there’s no need to achieve more abilities of cancer detection, diagnosis and treatment.

As reported by Chinese National Cancer Institute, lung cancer, gastric cancer, liver can-cer, esophageal cancer, colorectal cancer, female breast cancer and cervical cancer were the most common cancers in China [21]. In our study, various biomarkers most related to the aforementioned cancers were detected by clini-cal proven methods and new designed protein chips, named male chip and female chip, which are based on microarray chemiluminescent immunoassay and could simultaneously detect 11 kinds of biomarkers. Mukoubayashi C and his colleagues have found that serum PGI/II testing is useful in detecting early-stage gastric cancers arising from atrophic gastric mucosa [22]. As reviewed by Ischia J, pro-GRP is the most sensitive serum biomarker in patients with small cell lung cancer and is a significant predictor of poor survival in patients with advanced prostate cancer [22]. Interestingly, β-HCG is not only an important hormone, but also could be used as tumor marker for some gynecological cancers [24]. Although with vari-ous levels of SPC and SEN, the rest of the bio-markers targeted by protein chips are common clinical proven tumor markers [25-27].

Page 10: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

20422 Int J Clin Exp Med 2015;8(11):20413-20423

After comprehensive assay validation, protein chips demonstrated high sensitivity, high speci-ficity, good linearity, low imprecision and were free of common interferences. Compared with the traditional methods, protein chips have good correlation in the detection of all the 13 kinds of biomarkers (r≥0.935, P<0.001). As detailed in the results, we further compared the diagnostic ability among traditional meth-ods, protein chips and diagnostic models based on the multiplex chip results. The ROC curve results suggest that both traditional methods and protein chips are roughly the same in detecting single biomarker for cancer detec-tion, except NSE and pro-GRP. Specifically, female chip observed better diagnostic value in predicting lung cancer through pro-GRP detec-tion, while male chip had worse value of that. In addition, male chip had better diagnostic value in NSE detection for lung cancer discrimination. Once establishing diagnostic models based on multiplex chip results, it could significantly enhance their diagnostic ability. Thereafter, try-ing to improve the detection specificity and sen-sitivity of male or female chip, we have estab-lished 4 diagnostic formulas based on multivariate logistic regression analyses for total disease or cancer detection, respectively. Considering the SPC and SEN, it showed that protein chips might be useful in total disease detection but play limited roles in cancer dis-crimination. However, due to the small ratio of normal control in both male (97 in 1756) and female (65 in 1927) group, the high SPC of total disease differentiation might be not as real as it showed.

In conclusion, we have comprehensively assessed the new designed protein chips, which could simultaneously detect 11 kinds of biomarkers closely related to the most com-mon cancers in China. After assay validation and diagnostic value comparison, we found that the new designed protein chips have gen-erally equal diagnostic ability to the traditional methods and they observe high-throughput detection capabilities. Furthermore, combining with multivariate logistic regression, we found that diagnostic models established from the chip results might be useful in cancer detec-tion. In a word, these chips might play impor-tant roles in cancer early detection, especially for mass screening, which has urgent need for high-throughput testing methods. Although this study was conducted in four different hospitals,

the sample size was still small, especially that of the healthy control group. Besides, the sam-ple size of specific diseases or cancer was also small. Hence, multicenter studies with larger sample sizes will be needed to further validate the value of the chips in diagnosing cancer.

Acknowledgements

This study was supported by the National High Technology Research and Development Pro- gram of China (863 Program, No. 2011AA02- A111), the National Natural Science Foundation of China (No. 81371906) and the Capital Health Development Special Scientific Research Projects (2014-2-2154). And we wish to thank all of the doctors and nurses in the above four mentioned hospitals for their kindly support and assistance.

Disclosure of conflict of interest

None.

Address correspondence to: Yaping Tian, Core Laboratory of Translational Medicine, State Key Laboratory of Kidney Disease, Chinese PLA General Hospital, Beijing, P. R. China. E-mail: [email protected]

References

[1] Hanahan D, Weinberg RA. Hallmarks of can-cer: the next generation. Cell 2011; 144: 646-674.

[2] Shibue T, Brooks MW, Weinberg RA. An integ-rin-linked machinery of cytoskeletal regulation that enables experimental tumor initiation and metastatic colonization. Cancer Cell 2013; 24: 481-498.

[3] Vockley JG, Niederhuber JE. Diagnosis and treatment of cancer using genomics. BMJ 2015; 350: h1832.

[4] Cheng G. Circulating miRNAs: roles in cancer diagnosis, prognosis and therapy. Adv Drug Deliv Rev 2015; 81: 75-93.

[5] Diamandis EP. Present and future of cancer biomarkers. Clin Chem Lab Med 2014; 52: 791-794.

[6] American Cancer Society. Cancer Facts & Fig-ures 2013. Atlanta, USA: American Cancer So-ciety; 2013. pp. 1-8.

[7] Smith RA, Manassaram-Baptiste D, Brooks D, Cokkinides V, Doroshenk M, Saslow D, Wender RC, Brawley OW. Cancer screening in the Unit-ed States, 2014: a review of current American Cancer Society guidelines and current issues in cancer screening. CA Cancer J Clin 2014; 64: 30-51.

Page 11: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

20423 Int J Clin Exp Med 2015;8(11):20413-20423

[8] Sturgeon CM, Duffy MJ, Stenman UH, Lilja H, Brünner N, Chan DW, Babaian R, Bast RC Jr, Dowell B, Esteva FJ, Haglund C, Harbeck N, Hayes DF, Holten-Andersen M, Klee GG, Lamerz R, Looijenga LH, Molina R, Nielsen HJ, Rittenhouse H, Semjonow A, Shih IeM, Sibley P, Sölétormos G, Stephan C, Sokoll L, Hoffman BR, Diamandis EP. National Academy of Clini-cal Biochemistry. National Academy of Clinical Biochemistry laboratory medicine practice guidelines for use of tumor markers in testicu-lar, prostate, colorectal, breast, and ovarian cancers. Clin Chem 2008; 54: e11-79.

[9] Zhang Y, Yang D, Weng L, Wang L. Early lung cancer diagnosis by biosensors. Int J Mol Sci 2013; 14: 15479-509.

[10] Berrade L, Garcia AE, Camarero JA. Protein mi-croarrays: novel developments and applica-tions. Pharm Res 2011; 28: 1480-99.

[11] Chen H, Werner S, Tao S, Zörnig I, Brenner H. Blood autoantibodies against tumor-associat-ed antigens as biomarkers in early detection of colorectal cancer. Cancer Lett 2014; 346: 178-187.

[12] Paoletti C, Hayes DF. Molecular testing in breast cancer. Annu Rev Med 2014; 65: 95-110.

[13] Montagnana M, Danese E, Ruzzenente O, Bresciani V, Nuzzo T, Gelati M, Salvagno GL, Franchi M, Lippi G, Guidi GC. The ROMA (Risk of Ovarian Malignancy Algorithm) for estimat-ing the risk of epithelial ovarian cancer in wom-en presenting with pelvic mass: is it really use-ful? Clin Chem Lab Med 2011; 49: 521-525.

[14] Zhang P, Zou M, Wen X, Gu F, Li J, Liu G, Dong J, Deng X, Gao J, Li X, Jia X, Dong Z, Chen L, Wang Y, Tian Y. Development of serum param-eters panels for the early detection of pancre-atic cancer. Int J Cancer 2014; 134: 2646-2655.

[15] Pengjun Z, Xinyu W, Feng G, Xinxin D, Yulan L, Juan L, Xingwang J, Zhennan D, Yaping T. Mul-tiplexed cytokine profiling of serum for detec-tion of colorectal cancer. Future Oncol 2013; 9: 1017-1027.

[16] NCCLS. Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guide-line. NCCLS document EP5-A [ISBN 1-56238-368-X]. NCCLS, 940 West Valley Road, Suite 1400, Wayne, PA 19087-1898 USA, 1999.

[17] Siegel R, Ma J, Zou Z, Jemal A. Cancer statis-tics, 2014. CA Cancer J Clin 2014; 64: 9-29.

[18] Kwon H, Lok AS. Does antiviral therapy prevent hepatocellular carcinoma? Antivir Ther 2011; 16: 787-795.

[19] Aggarwal P. Cervical cancer: Can it be prevent-ed? World J Clin Oncol 2014; 5: 775-780.

[20] Borchmann P, Eichenauer DA, Engert A. State of the art in the treatment of Hodgkin lympho-ma. Nat Rev Clin Oncol 2012; 9: 450-549.

[21] Chen W, Zheng R, Zhang S, Zhao P, Zeng H, Zou X. Report of cancer incidence and mortal-ity in China, 2010. Ann Transl Med 2014; 2: 61.

[22] Mukoubayashi C, Yanaoka K, Ohata H, Arii K, Tamai H, Oka M, Ichinose M. Serum pepsino-gen and gastric cancer screening. Intern Med 2007; 46: 261-266.

[23] Ischia J, Patel O, Shulkes A, Baldwin GS. Gas-trin-releasing peptide: different forms, differ-ent functions. Biofactors 2009; 35: 69-75.

[24] Stenman UH, Alfthan H. Determination of hu-man chorionic gonadotropin. Best Pract Res Clin Endocrinol Metab 2013; 27: 783-793.

[25] Juan L, Tong HL, Zhang PJ, Wen XY, Gao YH, Nan JZ, Tian YP. Clinical utility of serum tumor markers and cytokines in cervical cancer and neoplasia. Clin Chem Lab Med 2013; 51: e183-186.

[26] Wang XF, Wu YH, Wang MS, Wang YS. CEA, AFP, CA125, CA153 and CA199 in malignant pleu-ral effusions predict the cause. Asian Pac J Cancer Prev 2014; 15: 363-368.

[27] Verma A, St Onge J, Dhillon K, Chorneyko A. PSA density improves prediction of prostate cancer. Can J Urol 2014; 21: 7312-7321.

Page 12: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

1

Supplementary Table 1. Intra-batch evaluation of precision performance of male chip

Reagent Sample Index PGIng/ml

PGIIng/ml

CA199U/ml

AFPng/ml

PSAng/ml

F-PSAng/ml

NSE ng/ml

CA125U/ml

CEAng/ml

CA724U/ml

Pro-GRPng/ml

Lot 1 Internal quality control 1 Mean 66.3 6.0 113.3 328.6 5.5 3.1 13.8 141.2 23.6 19.9 0.4 SD 2.6 0.2 9. 8 24.2 0.3 0. 3 1.3 5.4 1.9 1.8 0.02

CV, % 3.9 4.1 8.6 7.4 5.0 9.4 9.1 3.8 8.2 8.9 4.0 Internal quality control 2 Mean 329.3 22.9 526.9 931.9 25.1 16.4 62.6 746.2 91.3 99.3 1.8

SD 7.8 1.6 50.5 30.8 2.2 1.5 6.1 34.8 7.9 7.5 0.1 CV, % 2.4 6.9 9.6 3.3 8.8 8.9 9.7 4.7 8.7 7.6 6.0

Lot 2 Internal quality control 1 Mean 70.5 7.5 130.3 361.2 5.3 2.9 13.8 136.3 22.6 20.5 0.4 SD 2.2 0.3 10.7 19.9 0.2 0.2 1.3 5.7 2.1 1.3 0.02

CV, % 3.1 3.8 8.2 5.5 3.9 8.1 9.1 4.2 9.3 6.4 5.3 Internal quality control 2 Mean 371.1 2.2 553.4 1021.9 26.7 16.4 64.9 684.5 97.7 93.8 1.8

SD 27.0 0.1 49.5 56.7 1.9 1.5 6.0 29.3 6.7 7.0 0.1 CV, % 7.3 6.5 8.9 5.6 7.1 9.2 9.3 4.3 6.9 7.5 5.4

Lot 3 Internal quality control 1 Mean 72.3 7.6 113.3 318.1 5.2 2.7 13.9 137.4 24.2 19.0 0.4 SD 2.9 0.3 9.8 21.6 0.2 0.3 1.4 5.1 2.3 1.7 0.02

CV, % 4.0 3.6 8.6 6.8 3.3 9.6 10.0 3.7 9.5 8.7 4.4 Internal quality control 2 Mean 395.5 2.1 537.4 960.1 27.7 13.9 67.3 681.5 99.7 91.6 1.9

SD 21.6 0.2 38.4 80.8 1.3 1.3 6.3 21.1 9.8 6.7 0.1 CV, % 5.5 8.7 7.2 8.4 4.8 9.0 9.4 3.1 9.9 7.3 6.7

Page 13: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

2

Supplementary Table 2. Inter-batch evaluation of precision performance of male chip

Biomarker ReagentPrecision among days Inter-batch precision

Mean SD CV, % Mean SD CV, %CA199 Lot 1 111.3 9.1 8.1 111.1 9.5 8.5

Lot 2 111.7 10.1 9.1 Lot 3 110.3 9.3 8.5

AFP Lot 1 161.5 8.1 5.0 162.8 8.9 5.5Lot 2 162.7 8.8 5.4 Lot 3 164.2 9.7 5.9

PGI Lot 1 66.2 4.2 6.4 67.3 4.7 7.0Lot 2 67.4 5.1 7.5 Lot 3 68.4 4.7 6.9

PGII Lot 1 4.8 0.3 6.1 4.8 0.3 6.9Lot 2 4.9 0.3 6.7 Lot 3 4.9 0.4 7.6

CEA Lot 1 23.0 1.6 7.0 22.8 1.5 6.8Lot 2 22.8 1.7 7.3 Lot 3 22.6 1.4 6.0

CA125 Lot 1 146.7 10.2 7.0 145.7 11.3 7.8Lot 2 144.9 10.8 7.4 Lot 3 145.4 13.0 9.0

NSE Lot 1 13.5 0.941 7.0 13.3 1.0 7.4Lot 2 13.2 0.992 7.5 Lot 3 13.2 0.996 7.6

Pro-GRP Lot 1 0.4 0.03 6.9 0.4 0.3 6.5Lot 2 0.4 0.02 5.9 Lot 3 0.4 0.03 6.6

T-PSA Lot 1 5.7 0.4 6.5 5.7 0.4 6.5Lot 2 5.7 0.4 6.4 Lot 3 5.8 0.4 6.6

F-PSA Lot 1 3.0 0.2 5.6 2.9 0.2 5.5Lot 2 2.9 0.2 5.9 Lot 3 2.9 0.2 5.2

CA724 Lot 1 19.3 1.5 7.9 19.4 1.5 7.5Lot 2 19.5 1.5 7.8 Lot 3 19.5 1.4 7.0

Page 14: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

3

Supplementary Table 3. Inter-batch evaluation of precision performance of female chip

Reagent Sample Index PG1 ng/ml

PG2 ng/ml

CA199 U/ml

AFP ng/ml

CA153 ng/ml

β-HCG ng/ml

NSE ng/ml

CA125 U/ml

CEA ng/ml

CA724 U/ml

Pro-GRP ng/ml

Lot 1 Internal quality control 1 Mean 70.7 6.91 120.26 313.86 9.69 3.88 1.29 142.8 23.57 20.39 0.38SD 2.26 0.23 9.41 18.50 0.79 0.35 0.12 7.26 2.10 1.8 0.02

CV, % 3.19 3.34 7.82 5.89 8.11 8.99 9.00 5.08 8.92 8.83 5.69Internal quality control 2 Mean 359.3 21.18 280.36 414.55 41.26 28.75 13.88 741 86.03 102.3 1.79

SD 17.76 1.195 15.87 15.38 1.58 2. 7 1.07 29.85 3.81 6.85 0.09CV, % 4.94 5.64 5.66 3.71 3.83 9.38 7.74 4.03 4.43 6.70 4.97

Lot 2 Internal quality control 1 Mean 66.88 7.35 108.75 314.29 8.72 3.36 13.01 132.7 22.95 18.84 0.37SD 2.155 0.27 8.07 18.14 0.77 0.31 1.16 6.96 2.28 1.41 0.01

CV, % 3.22 3.64 7.42 5.77 8.77 9.10 8.92 5.25 9.95 7.48 3.43Internal quality control 2 Mean 236 16.74 235.45 415.21 41.37 26.13 116.4 714.1 79.41 95.75 1.78

SD 4.46 0.54 15.7 21.53 1.96 2.41 6.35 38.1 4.31 7.29 0.07CV, % 1.89 3.24 6.67 5.19 4.74 9.22 5.45 5.34 5.43 7.61 4.09

Lot 3 Internal quality control 1 Mean 68.54 6.97 118.85 328.6 9.32 3.56 13.58 131.3 22.98 19.35 0.37SD 2.65 0.21 10.46 24.22 0.79 0.32 1.24 4.98 2.18 1.79 0.02

CV, % 3.87 2.98 8.80 7.37 8.50 8.83 9.14 3.80 9.47 9.25 4.21Internal quality control 2 Mean 231.7 15.08 243.73 424.51 42.64 24.97 123.8 718.9 90.59 97.07 1.91

SD 12 0.55 19.24 22.54 1.62 1.81 7.12 30.99 2.0 7.29 0.11CV, % 5.18 3.67 7.89 5.31 3.81 7.26 5.75 4.31 2.21 7.51 5.51

Page 15: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

4

Supplementary Table 4. Inter-batch evaluation of precision performance of female chip

Biomarker ReagentPrecision among days Inter-batch precision

Mean SD CV, % Mean SD CV, %CA199 Lot 1 115.337 5.373 4.66 115.14 5.83 5.06

Lot 2 114.343 5.327 4.66Lot 3 115.738 6.736 5.82

AFP Lot 1 168.59 10.535 6.25 166.53 11.27 6.77Lot 2 166.09 11.975 7.21Lot 3 164.9 11.228 6.81

PGI Lot 1 68.298 5.520 8.08 67.421 5.052 7.49Lot 2 66.648 4.691 7.04Lot 3 67.32 4.90 7.28

PGII Lot 1 5.018 0.377 7.51 4.978 0.38 7.57Lot 2 4.853 0.308 6.35Lot 3 5.062 0.414 8.17

CEA Lot 1 21.976 2.101 9.56 21.736 1.930 8.88Lot 2 21.673 1.947 8.98Lot 3 21.559 1.751 8.12

CA125 Lot 1 146.985 13.438 9.14 146.965 12.805 8.71Lot 2 147.287 13.050 8.86Lot 3 146.624 12.214 8.33

NSE Lot 1 12.580 1.027 8.17 12.479 1.016 8.14Lot 2 12.363 1.083 8.76Lot 3 12.494 0.946 7.57

Pro-GRP Lot 1 0.409 0.026 6.34 0.405 0.025 6.23Lot 2 0.405 0.026 6.43Lot 3 0.401 0.024 5.90

CA153 Lot 1 10.271 0.897 8.74 10.213 0.853 8.35Lot 2 10.219 0.767 7.51Lot 3 10.149 0.904 8.91

βHCG Lot 1 8.519 0.749 8.79 8.625 0.702 8.14Lot 2 8.608 0.715 8.31Lot 3 8.747 0.636 7.27

CA724 Lot 1 19.369 1.686 8.71 19.469 1.622 8.33Lot 2 19.691 1.776 9.02Lot 3 19.346 1.398 7.23

Page 16: Original Article The clinical performance evaluation of

Development, evaluation and model study of novel protein chips

5

Supplementary Table 5. Diagnostic models established from multiple biomarkers for various cancer predictionGender Biomarker* Diagnostic model AUC Sensitivity (%) Specificity (%)Male PGI Y=logit(P)=-1.518-0.007XPGI+0.014XPGII+0.001XCA199-0.002XCA125+0.005XCEA-0.396Xpro-GRP+0.005XCA724 0.683 40.9 76.7

PGII Y=logit(P)=-1.237-0.01XPGI+0.017XPGII+0.001XCA199-0.02XNSE+0.005XCEA-0.277Xpro-GRP+0.005XCA724 0.690 39.5 71.4pro-GRP Y=logit(P)=-1.993-0.004XAFP+0.013XNSE+0.001XCA125+0.544Xpro-GRP 0.719 21.0 93.7

AFP Y=logit(P)=-3.224+0.003XAFP 0.856 68.2 92.0NSE Y=logit(P)=-2.149+0.01XNSE+0.007XCEA+0.625Xpro-GRP 0.785 62.4 56.4CEA Y=logit(P)=-1.68-0.005XPGI+0.001XCA199-0.023XNSE+0.005XCEA+0.003XCA724 0.706 67.9 67.2

CA125 Y=logit(P)=-0.567-0.007XPGI+0.013XPGII+0.002XAFP+0.018XNSE+0.012XCEA+0.436Xpro-GRP+0.005XCA724 0.718 55.5 63.7CA199 Y=logit(P)=-1.924-0.007XPGI+0.014XPGII+0.001XCA199-0.002XCA125+0.007XCEA+0.006XCA724 0.693 42.6 74.9CA724 Y=logit(P)=-1.816-0.005XPGI+0.001XCA199-0.371Xpro-GRP+0.009XCA724 0.674 61.8 67.2

total-PSA Y=logit(P)=-4.29+0.067Xtotal-PSA 0.867 81.0 88.7free-PSA Y=logit(P)=-4.145+0.045Xtotal-PSA 0.909 84.0 79.3

Female PGI Y=logit(P)=-1.385-0.005XPGI+0.001XCA199-0.057XNSE+0.004XCEA-0.626Xpro-GRP+0.007XCA724 0.704 22.2 89.8PGII Y=logit(P)=-1.359-0.005XPGI+0.001XCA199-0.057XNSE+0.004XCEA-0.622Xpro-GRP+0.007XCA724 0.702 55.1 57.8

pro-GRP Y=logit(P)=-2.392+0.009XCEA+0.863Xpro-GRP+0.004XCA153 0.804 28.2 99.7AFP Y=logit(P)=-3.201+0.003XAFP 0.780 55.9 87.2NSE Y=logit(P)=-1.938-0.005XPGI+0.004XCEA+1.531Xpro-GRP+0.006XCA153+0.005XCA724 0.721 44.6 71.6CEA Y=logit(P)=-2.005-0.005XPGI+0.002XCA199 0.660 67.3 69.0

CA125 Y=logit(P)=-0.629-0.005XPGI+0.001XCA199+0.004XAFP+0.032XCEA+0.961Xpro-GRP+0.008XCA153+0.007XCA724 0.744 58.4 52.4CA199 Y=logit(P)=-2.234+0.001XCA199+0.005XCA724 0.681 51.0 79.6

CA724 Y=logit(P)=-1.861-0.005XPGI+0.002XCA199-0.002XCA125+0.004XCEA-0.529Xpro-GRP-0.013XCA153+0.015XCA724 0.764 47.4 80.1βHCG Y=logit(P)=-2.971-0.009XCA199+0.001XCA125+0.004XβHCG 0.701 33.3 81.1CA153 Y=logit(P)=-3.184+0.009XCA153 0.649 47.9 86.2

Note: *Biomarker was separately detected by old method and was part of the protein chip results. AUC: Areas under the curve.

Supplementary Table 6. Diagnostic models established from multiple biomarkers for total disease or cancer prediction

Gender Function Diagnostic model AUC Sensitivity (%)

Specificity (%)

Male Disease prediction Y=logit(P)=-8.325+0.028XPG2+0.057XAFP+0.455Xtotal-PSA+2.641Xfree-PSA+0.882XNSE+0.144XCA125+8.028Xpro-GRP+0.057XCA724 0.981 91.2 100

Cancer prediction Y=logit(P)=-0.197-0.006XPG1+0.01XPG2+0.001XAFP+0.01XNSE+0.008XCEA+0.5Xpro-GRP+0.003XCA724 0.691 46.1 83.1

Female Disease prediction Y=logit(P)=0.384+0.374XNSE 0.836 71.8 95.2

Cancer prediction Y=logit(P)=-0.614-0.006XPG1+0.002XCA199+0.003XAFP+0.03XCEA+1.307Xpro-GRP-0.002XβHCG+0.016XCA153+0.009XCA724 0.753 51.6 89.2

AUC: Areas under the curve.