OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM 1

Upload: others

Post on 30-Apr-2020

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

1

Page 2: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

2

This operation manual describes only the option PARTICULATE MATTER MEASUREMENT together with the OPTIMA 7/SPECTRAplus/NOVAplus.

When using the NOVAplus, connect the hose for gas to the NOVAplus basic unit and the electrical cable connection to the AUX socket of the RCU comfort remote control.

For the general handling of the OPTIMA 7/SPECTRAplus/NOVAplus including the connectors please read the operation manuals of the OPTIMA 7/SPECTRAplus/NOVAplus.

Page 3: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

3

1 CONTENT 1 CONTENT .................................................................................................................................. 3

2 INTRODUCTION ......................................................................................................................... 6

2.1 The MRU particulate matter measurement system (FSM) ....................................................... 6

2.2 The company MRU GmbH ................................................................................................... 7

2.3 Important general information VDI 4206-2 ............................................................................ 8

2.4 Important information about the users/operation manual ..................................................... 8

3 SAFETY INFORMATION ............................................................................................................... 8

3.1 Safety regulations .............................................................................................................. 8

3.2 Specific safety regulations ................................................................................................... 9

4 RETURNED GOODS .................................................................................................................... 9

4.1 Return of hazardous waste .................................................................................................. 9

4.2 Return of analyzer according to ElektroG .............................................................................. 9

5 MEASUREMENT PRINCIPAL ....................................................................................................... 10

5.1 Basic principal of a particulate matter measurement ............................................................ 10

5.1.1 In the user's office ......................................................................................................... 10

5.1.2 On site measuring ......................................................................................................... 10

5.1.3 In the user's office ......................................................................................................... 10

5.2 FSM Gas flow diagram ....................................................................................................... 11

6 DESCRIPTION........................................................................................................................... 12

6.1 Connections at FSM case ................................................................................................... 12

6.2 Hose diagram ................................................................................................................... 13

7 ACCESSORIES .......................................................................................................................... 14

7.1 Sampling gas/particulate matter probe .............................................................................. 14

8 OPERATING ............................................................................................................................. 15

8.1 LED display FSM case ........................................................................................................ 15

8.2 Display FSM scale ............................................................................................................. 15

9 FIRST USE OF THE FSM PARTICULATE MATTER MESASURING SYSTEM AND NEW FILTER CARTRIDGE DATASET ........................................................................................................................................ 16

9.1 Device ready for operation ................................................................................................ 16

9.2 Generate new filter cage data set ....................................................................................... 16

10 MAINTENANCE ..................................................................................................................... 16

10.1 FSM care program ............................................................................................................ 16

10.2 Maintenance and cleaning ................................................................................................ 16

10.3 Service and Maintenance .................................................................................................. 16

11 ADMINISTRATION OF FILTER ................................................................................................. 17

11.1 Edit filter cage status ......................................................................................................... 17

Page 4: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

4

11.2 Delete empty filter cages ................................................................................................... 17

11.3 Generate filter cage .......................................................................................................... 17

12 General description of a weighing process ............................................................................. 18

13 Initially weighed (in the user's office) ..................................................................................... 21

14 Measuring with FSM ............................................................................................................. 22

15 Finally weighed (in the user's office)....................................................................................... 26

16 TECHNICAL SPECIFICATIONS ................................................................................................. 28

16.1 General Specifications ....................................................................................................... 28

16.2 Interfaces ......................................................................................................................... 29

16.3 Probe............................................................................................................................... 29

16.4 Measuring values .............................................................................................................. 30

16.5 Calculated values .............................................................................................................. 31

17 APPENDIX ............................................................................................................................ 31

17.1 Further menu items .......................................................................................................... 31

17.1.1 Filter cages information ............................................................................................. 31

17.1.2 FSM-Info ................................................................................................................... 32

17.1.3 FSM-Scale info ........................................................................................................... 32

17.1.4 FSM Service values ..................................................................................................... 32

17.1.5 FSM bench test .......................................................................................................... 33

17.1.5.1 FSM Tightness test .............................................................................................. 33

17.1.5.2 FSM Flow rate test .............................................................................................. 33

17.2 Replacement - and Consumables ....................................................................................... 34

17.3 Declaration of conformity FSM module .............................................................................. 35

18 GENERAL INFORMATION ABOUT THE FSM SCALE .................................................................... 36

19 UNPACKING, ASSEMBLING AND STARTUP .............................................................................. 36

19.1 Unpacking ....................................................................................................................... 36

19.2 ASSEMBLING .................................................................................................................... 37

19.3 Level setting..................................................................................................................... 39

19.4 Place of installation ........................................................................................................... 40

19.5 Connect OPTIMA 7 with the FSM scale ................................................................................ 40

19.6 Bench test of the FSM scale ............................................................................................... 42

19.7 Measurement of the heating temperature of the FSM scale .................................................. 42

19.8 Error messages ................................................................................................................. 43

19.9 Packaging ........................................................................................................................ 43

20 Declaration of conformity FSM scale ..................................................................................... 44

Page 5: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

5

This operation manual describes only the option PARTICULATE MATTER MEASUREMENT together with the OPTIMA 7/SPECTRAplus/NOVAplus.

When using the NOVAplus, connect the hose for gas to the NOVAplus basic unit and the electrical cable connection to the AUX socket of the RCU comfort remote control.

For the general handling of the OPTIMA 7/SPECTRAplus/NOVAplus including the connectors please read the operation manuals of the OPTIMA 7/SPECTRAplus/NOVAplus.

Inspect Shipment for Damage Carefully inspect the entire shipment for damage in the presence of the shipper’s agent, removing packaging material if necessary. Note any damage to packaging and/or goods on Packing List and have it signed by the shipper’s agent prior to accepting the shipment. Submit damage claim to MRU immediately.

NOTE: Damage claims not received by MRU within 3 days of receipt of shipment will not be accepted.

Save the original box and the packing material for use if the scale must be shipped in the future.

The products described in this manual are subject to continuous development and improvement and it is therefore acknowledged that this manual may contain errors or omissions. MRU encourages customer feedback and welcomes any comments or suggestions relating to the product or documentation.

Please forward all comments or suggestions to the Customer Feedback Department at the following address:

MRU GmbH, Fuchshalde 8 + 12, 74172 Neckarsulm / Obereisesheim, GERMANY

Phone (+49)7132-9962-0 (Front desk)

Phone (+49)7132-9962-61 (Service)

Fax (+49)7132-9962-20

Email: [email protected], Homepage: www.mru.eu

This manual is intended solely as a guide to the use of the product.

MRU shall not be liable for any loss or damage whatsoever arising from content errors or misinterpretation of information’s from this manual or any misuse resulting from the use of this manual.

Page 6: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

6

2 INTRODUCTION 2.1 The MRU particulate matter measurement system (FSM)

Basic principle of the MRU particulate matter measurement (FSM) is the loading of a filter with particulate matter and the weighing of this filter. The amount of particulate matter resulting from the difference in weight before and after the filter has been loaded. On site the FSM case is needed and the weighing of the filter will be done off site in the user’s office.

This separation enables a more reliable overall system, the mobile component of the FSM particulate matter collection suitcase is simple and robust. The weighing method is a reliable method to determine the particulate matter weight. Using a special scale that simultaneously dry’s and weigh’s the filter in one step, errors are excluded by moisture in the filter without having to implement a lengthy filter conditioning. The complete system FSM is designed as an accessory to the proven basic unit OPTIMA 7, enabling the complete menu guided operation of the particulate matter measurement. Both, FSM scales and FSM particulate matter collection box can only be operated via the OPTIMA 7, which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations for particulate matter content, as well as the O2 and CO - measurement and allows transfer of results to either the printer, the data memory or to a management software. The FSM module is installed fine for transporting and operating in a case.

Page 7: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

7

2.2 The company MRU GmbH

Your particulate matter measurement system is produced by the MRU GmbH in Neckarsulm Germany (founded in 1984), a medium-sized company that is specialized in developing, producing and marketing of high quality emission monitoring analyzers/systems. MRU GmbH produces a wide range of instruments, from standard analyzers up to tailor made Process Gas Analyzers and CEMS. MRU GmbH contact details are listed on one of the previous page.

MRU Plant 1 (service department, development division, administration)

MRU Plant 2 (Product division)

Page 8: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

8

2.3 Important general information VDI 4206-2

This measuring system is not designed to be used for continuous measurements.

This measurement system is designed to correspond the requirements of the first BlmSchV and VDI 4206-2.

A change of sensors at the device which is used in the context of 1.BImSchV is permitted only by the manufacturer or by qualified personnel authorized by the manufacturer. The measurement system must be checked by a competent authority under state law office announced semi-annually. Improper operation may cause an electric shock.

2.4 Important information about the users/operation manual The users/operation manual is an important part of this delivery. It will explain how to use this analyzer properly and sets forth safety and environmentally friendly procedures.

It is the responsibility of all users to read and familiarize themselves with this manual, paying particular attention to the safety instructions.

The most important safety details are listed in chapter 3 (Safety Information). Additional safety details in other chapters are clearly marked with an attention sign.

3 SAFETY INFORMATION The following safety procedures must to be followed at all times. They are a significant and essential part of this manual. Failure to follow safety procedures can result in the loss of your warranty.

3.1 Safety regulations

1. The particulate matter measurement system may only be used as indicated in this manual.

2. Our devices are checked according to the following regulations: VDE 0411 (EN61010) and DIN VDE 0701 before they leave the MRU GmbH factory.

3. MRU technical products are designed and manufactured according to DIN 31000/ VDE 1000 and UVV=VBG 4 of the professional guilds for fine mechanics and electrical engineering.

MRU GmbH assures that the analyzer complies to the essential requirements of the legal regulations of the member states of the electro-magnetic compatibility (89/336/EWG) and to the low-voltage regulations (3/23/EWG).

Page 9: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

9

3.2 Specific safety regulations

1. Use only the power supply supplied with the device for the system

2. No part of the system, including the metal probe tube and all other metal parts & accessories are to be used as electric conductors.

3. The system is not to be used in or under water.

4. The system is not to be placed near or directly exposed to open fire or heat.

5. The specified probe temperature range is not to be exceeded, as the probe, temperature sensory mechanism and sensor could be damaged or destroyed.

6. Don’t drop any components of this system.

7. Caution: Moisture, being pumped out of the condensates trap can be slightly acidic. In case of skin contact IMMEDIATELY: clean affected parts of the body. Avoid getting liquid in eyes. Please carefully clean all parts that come into contact with the condensates.

8. After measurement, vent the system with ambient air and allow the parts to cool. A hot filter cartridge or probe could cause burns or ignite flammable material.

Your quality management MRU GmbH

4 RETURNED GOODS Packing regulation of 12.07.1991

If your local waste facility does not except MRU packing materials for disposal, you may return it to MRU or our local sales representative. Packing materials returned to MRU must be returned prepaid.

4.1 Return of hazardous waste

- Waste Disposal/Returns/Warranty -

MRU GmbH is required to accept the return of hazardous waste such as electro-chemical sensors that cannot be disposed of locally. Hazardous waste must be returned to MRU prepaid.

4.2 Return of analyzer according to ElektroG MRU GmbH is required to accept the return, for proper disposal, of all analyzers delivered after 13th of August 2005.

Devices must be returned to MRU prepaid.

Page 10: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

10

5 MEASUREMENT PRINCIPAL 5.1 Basic principal of a particulate matter measurement

5.1.1 In the user's office

a) Connect FSM scale with OPTIMA 7

b) Announce the number of the initial weighing FSM filter cage the OPTIMA 7

c) Initial weighing of the FSM filter cage incl. particulate collection filter.

d) Packing of FSM filter case and particulate collection filter for transportation with the FSM case.

5.1.2 On site measuring a) Setting up the FSM particulate matter collection case

b) Insert a pre-weighed FSM filter case into the probe

c) Connect the OPTIMA 7 with the FSM particulate matter collection unit and select the pre-weight filter cage.

d) Perform a leak tight test (hold tube closed - takes about 5-7 seconds)

e) Insert probe into the stack

f) After the heating process is completed the measurement can be started

g) Once measurement is completed, disconnect the inlet hose from the FSM unit and start the purging process.

h) Remove the FSM filter cage from the probe (ATTENTION: hot) and store it in an appropriate transport container.

i) If necessary, empty and clean the condensate separator

j) Pack up the FSM particulate matter collection case

5.1.3 In the user's office

a) Connect FSM scale with the OPTIMA 7

b) Announce the number of the final weighing FSM filter cage to the OPTIMA 7

c) Take the FSM filter cartridge from the transport container and reweigh it.

d) Review result

Page 11: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

11

5.2 FSM Gas flow diagram

Position Description

1 Inlet gas sampling probe

2 Condensate separator with filter

3 Flow rate unit

4 Pump unit

5 Condensate separator

6 Outlet condensate / Connect the waste hose

7 Filter set (option)

8 OPTIMA7

A Backup-filter in condensate separator

B Silica gel in filter set

C Activated carbon in filter set

D Purafil in filter set

Page 12: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

12

6 DESCRIPTION 6.1 Connections at FSM case

Position Description

1 Connector AUX to OPTIMA 7

2 Connector flue gas measurement via filter unit

3 Connector flue gas inlet via filter

4 Connector gas outlet at condensate

5 Electrical connector to gas/particulate matter sampling probe

6 Connector gas outlet at condensate separator

7 Connector gas inlet filter

8 Connector filter at gas/particulate matter sampling probe

9 Power supply port

A Foil with LEDs

B Condensate separator

C Filter

Page 13: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

13

6.2 Hose diagram

FSM-case SPECTRAplus/ OPTIMA7

NOVAplus Basic unit

NOVAplus RCUcomfort

1 Gas inlet Gas inlet (Non-existent)

2 AUX-socket (Non-existent) AUX-socket

3 exhaust to the outside of the site

For particulate matter measurement the waste hose from the gas outlet connector has to be placed outside the site!

The filter unit is optional, if it’s not available, the hose from the gas inlet of the OPTIMA 7 is directly connected to the flue gas inlet at the FSM case.

Page 14: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

14

7 ACCESSORIES 7.1 Sampling gas/particulate matter probe

In der Kombination mit der FSM Feinstaubmessung wird folgende Sonde angeboten:

Sonde mit 300 mm Sondenrohr und 4,0 m Schlauchlänge.

Position Description

1 Probe tube 300 mm / 10 mm

2 Probe cone 10 mm

3 Gasket

4 Probe handle with filter cage, heated

5 Gas sampling hose 4 m

6 Electrical connector probe, 4 m

7 Filter cartridge

Including Pos. 8, 9 and 10

8 Particulate collection filter

9 Sealing cone of filter cage

10 Filter cage

11 additionally Aluminum transportation container

Page 15: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

15

8 OPERATING The operation of the FSM particulate matter measuring system will be performed using the Optima 7. Establish a connection between the FSM and the OPTIMA 7 using the AUX cable (or RS232 Cable adapter).

The red marking on the plug and socket must match when inserting the plug into the socket.

For the general handling of the OPTIMA 7 please read the operation manual of the OPTIMA 7 (# 62876GB).

8.1 LED display FSM case

The LED display shows the current operating status of the FSM particulate matter measuring system.

LED Status off on blinking

POWER No power supply connected

Power supply connected

COMMUNIKATION No AUX connection between the FSM and OPTIMA 7

AUX connection between the FSM and OPTIMA 7 ready for measurement

MEASUREMENT ACTIVE

Measurement not active

Pump not working

Measurement active

Pump is working

PROBE HEATING Operating temperature has been reached

Slow blinking: probe in heat up process Fast blinking: Probe is not connected to the FSM

ERROR No error Error: contact MRU service department at www.mru.eu

8.2 Display FSM scale The display of the FSM scale is for information only; you will receive all necessary information on the OPTIMA 7 display.

Page 16: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

16

9 FIRST USE OF THE FSM PARTICULATE MATTER MESASURING SYSTEM AND NEW FILTER CARTRIDGE DATASET 9.1 Device ready for operation

• Unpack the device, read the complete manual

• The FSM case was shipped completely assembled, in working condition and ready for operation. It is recommended that the FSM case is thoroughly inspected for completeness and damage.

• For the general handling of the OPTIMA 7 please read the operation manual of the OPTIMA 7.

9.2 Generate new filter cage data set The FSM includes a number of filter cages. Each filter cage has an engraved number. Each number must be inserted in the OPTIMA 7 cage management under „Generate filter cage“(Chapter 11.2).

10 MAINTENANCE 10.1 FSM care program

We recommend to perform the care program daily at the end of the measurement day.

OK

We recommend the FSM-care program. 1 hour to run. Press F1 for START.

10.2 Maintenance and cleaning

The FSM unit is a low maintenance unit we do however recommend to:

• now and then: Cleaning of the probe and the probe tube

• after every measurement: remove gas sampling tube from the FSM case, so that the hose can dry

• after each measurement: Inspect condensate show glass of the FSM and empty the condensate separator if necessary before closing the case.

• Check the condensate backup filter and replace if necessary. This filter is designed to protect the pump unit if the particulate matter collection filter in the probe is not inserted.

• Check solids filter unit if necessary - renew granules (optional).

10.3 Service and Maintenance

An annual service check and if necessary adjustment of the scale at an MRU service department (www.mru.eu) are recommended for the preservation of value.

Page 17: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

17

11 ADMINISTRATION OF FILTER The administration of filter cages allows you to create, manage and delete filter cages.

OK OK ,

11.1 Edit filter cage status

Generated and empty

The menu item "Generated and empty" shows all the cages that are generated but not yet in use. I.e. an initially weighing, FSM measurement or final weighing has not occurred.

Initially weighed

The menu item "Initially weighed" displays all initially weighed cages. With this function also initially weighed cages can be deleted. After deletion this filter cage can be weighed again.

Measured with FSM

The menu item „Measured with FSM“ displays all FSM filter cages which are used for the FSM measurement. With this function, the particulate matter measurement (sampling with FSM) can be deleted. Then, a new measurement with the FSM is possible. Please note: If particulate matter has already been collected in this cage, a new initially weighing is necessary!

Finally weighed

The menu item „Finally weighed“ displays all FSM filter cages which are finally weighed. If the reweighing is deleted; a new reweighing with this FSM filter cage is then again possible.

11.2 Delete empty filter cages Not be used FSM filter cages can be deleted using the delete function. To remove a FSM filter cage, the cage must have the status "generated and empty."

11.3 Generate filter cage If you want to use a new FSM filter cage, please enter the 3-digit number. The administration of filter cages then puts that FSM filter cage with the status "Generated and empty".

Page 18: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

18

12 General description of a weighing process This general description of a weighing process applies to both the standard weighing of FSM filter cages in the scale, as well as initial- and final weighing.

Switch on both the scale and the OPTIMA 7. (The connection must be made!) Make sure that the selected number of the FSM filter cages corresponds to the FSM filter cages that you are using. Handle each FSM filter cage so that a transfer of material is avoided in the FSM filter cage.

Please follow the instructions on the display of OPTIMA 7.

Placing the FSM filter cage in the center of the weighing pan, the elevations will prevent the filter cage from rolling.

Then the drying process begins.

Page 19: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

19

The drying stops after about 10 minutes and the completion of the drying is displayed on your OPTIMA 7 also giving a beeping signal. Remove FSM filter cage within about 1 minute and then close the lid off the scale.

If the drying time was not sufficient, a re-calls the measuring system to drying process.

Please note the FSM filter sleeve is hot after drying!

Use for gripping the sleeve filter the supplied tweezers!

After that the OPTIMA 7 determines, the deviation of the zero point. Please wait a moment until the message appears.

Please confirm with OK. The display shows the result for the weighing.

The weighing process is completed.

Before you can reuse the scale again you must allow the scale to cool down for a couple minutes. This time can be shortened by opening the lid of the scale.

Page 20: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

20

The OPTIMA 7 will allow the next measurement once the temperature is below 65 °C.

Page 21: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

21

13 Initially weighed (in the user's office) Voraussetzungen für das Vorwiegen

Conditions for initially weighed

Before weighing please insert a clean particulate matter collection filter into the filter cage. After inserting the particulate matter collection filter in the filter cage, screw on the sealing cone - hand-tight.

Connect the OPTIMA 7 with the FSM scale.

Select the menu item particulate matter measurement "Initial weighing". After selecting OK, the FSM filer cages selection appears.

Select one (already in the administration) of the filter cages) and confirm with OK.

The weighing process is described in "General Description of a weighing process”

If the FSM filter cage was successfully weighed this cage number will be shifted from the "Initially weighed" to "Measured with FSM “

The transport of the filter cages takes place in an aluminum carrying case.

.

Page 22: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

22

14 Measuring with FSM Setup the FSM particulate matter collection unit near the measuring point. 230 Vac power supply is required for the FSM particulate matter collection unit.

Remove an initially weighed filter particulate matter collector (1) from the transport container and put this cartridge into the probe handle (2) of the gas sampling probe.

IMPORTANT! A missing gasket (3) will lead to wrong measuring results.

Insert the gas sampling hose (5) to the gas inlet port of the condensate separator and the connector (6) on the probe socket of the FSM particulate matter collecting unit. (review hose schematic (chapter 6.2)

Connect the OPTIMA 7 via the AUX connection cable to the connection "data connection" of the FSM particulate matter collecting unit.

Switch on OPTIMA 7.

Select menu item „Particulate matter measurement“, Measuring with FSM“.

OK OK

If T gas and draft have not been measured, these values can be measured using the by F1 key.

This can’t be measured with the FSM probe. You need a flue gas probe or a separate Tgas sensor for the OPTIMA7.

With F3, the initially weighed FSM filter cases are listed.

Page 23: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

23

OK

Select an FSM filter cage and start with OK the particulate matter measurement.

Before using the instrument, an optically overall visual inspection of the device must be performed to ensure error-free operation of the device.

Follow the display instructions:

Performing tightness test

(Seal probe tube - takes about 5-7 seconds)

OK

After a successful tightness test follows the fuel selection, then the selection of limits.

For more information on the fuel list, and user-defined fuels can be found in the OPTIMA 7 operation manual as needed.

Page 24: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

24

Insert probe in flue gas

After finishing of the heating process the measurement can be started

F1

Start FSM measuring

You will start the particulate matter measurement using the F3 key. This process can be ended by using the F1 key at any given time.

The current measured values for CO, O2 and T-gas are displayed during the measurement period of 15 minutes on the OPTIMA 7 display. For more information about other options for the OPTIMA 7 display, refer to the operation manual OPTIMA 7 as needed.

Note:

During the measurement, T gas can only be displayed! The stored Tgas value is always "Temperature / draft" is determined in the window!

After completion of the measurement, please press the F1 key. The measurement result is stored under the current filter cage number.

If you also want a site number be assigned, please confirm with "Yes". (see Appendix administrative user manual OPTIMA7)

Page 25: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

25

After completing the measurement remove the inlet hose from the FSM particulate matter collection case and start the "cleaning phase

"IMPORTANT: If the hose is not removed, dust will collect and thus falsify the result!”

Please note the FSM filter cage is hot after the measurement!

Remove the FSM filter cage with a pair of tweezers from the probe (Warning: hot) and store in the appropriate transport container.

If necessary, empty the condensate separator!

Caution: Moisture, being pumped out of the condensates separator can be slightly acidic. In case of skin contact IMMEDIATELY: clean affected parts of the body. Avoid getting liquid in eyes. Please carefully clean all parts that came into contact with the condensates.

Pack up the FSM particulate matter collection case

Page 26: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

26

15 Finally weighed (in the user's office) Connect the OPTIMA 7 via the AUX - RS232 Cable Adapter with the FSM scale. The red marking on the plug must match when inserting with the red dot on the socket.

Select the menu item „Particulate matter measurement“, „Final weighing

OK OK

After pressing OK, the FSM filter cage selection appears. Choose a cage and confirm with OK.

For the “general weighing procedure” please refer to the chapter "General Description of a weighing process" (chapter 12).

After successful weighing the result is displayed. The cage number will be moved from the „Finally weighed list „to the "result list".

Please note the FSM filter cage is hot after finally weighing! Please dispose of the particulate matter collection filter (household waste). If necessary, clean the cage.

Result window:

In the result window, the result of the measurement can be considered (with initial und final values) again.

cage cage number date date of measurement

site number (if site is allocated)

Split logs: fuel (with reference to O2)

O2 [%]: average value of O2-measurement [%]

Limit value: limit value of site [mg/m³]

dust[g/m³O2]: determined dust value (interium value) (referenced to O2 content )

MU[g/m³O2]: expanded measurement uncertainty (referenced to the particular limit)

dust.U [g/m³O2]: result of dust measurement [g]

(less the measurement uncertainty (MU), based on the oxygen content

Page 27: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

27

On the following page, the identification numbers of the instrument are displayed.

The result of the particle measurements can be stored in the data memory. The particle measurements can then be viewed, printed and deleted data in memory even when the corresponding cage number was deleted. Subsequent conditioning assignment in the result window is possible.

For more information on the documentation of the measurement, the print out, the data storage in the customer management and data processing can be found in the OPTIMA 7 manual as needed.

Take the filter cartridge with the included tweezers from the scale.

Please note the FSM filter cage is hot after the reweighing! Remove the dust filter from the filter cage. Please dispose of the dust filter (household waste). If necessary, clean the filter cage.

Enter the filter cartridge in the filter cage administration with the F3 key again. All filter cages data will be deleted or if the cage of a site is assigned remains the result in the data memory condition.

The administration of filter cages moves the FSM filter cage back to the initially weighed list!

A new measurement with this cage is now possible again.

cage cage number

reweighing. [g]: weight reweighing [g]

initial weighing. [g]: weight initial weighing [g]

St. [mg/m³]: Intermediate value from the intake gas amount and the difference between the reweighing and initial weighing

period: period of dust measurement

Scale Serial number FSM scale

FSM Serial number FSM module

cage cage number

Zug [Pa]: measuring value of draft [Pa]

T-Gas [°C]: measuring value T-Gas [°C]

CO [ppm]: Average value of CO-measurement [ppm]

measured CO [g/m³ O2]: measuring value of CO [g/m³] referenced to O2 content

MU [g/m³ O2]: determined measurement uncertainty of CO measurement

CO.U[g/m³O2]: result of CO average measurement in [g/m³O2]

(less the measurement uncertainty CO measurement based on the oxygen content)

Page 28: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

28

16 TECHNICAL SPECIFICATIONS 16.1 General Specifications

Deutsch OPTIMA7 FSM –Staub-sammelkoffer

FSM-Waage English

Umgebungs-temperatur bei Betrieb

->BA OPTIMA7 +5°C ... +45 °C +15 °C ... +30 °C Ambient temperature

during operation

Lager und –Trans-porttemperatur

->BA OPTIMA7 - 20 °C ... +50 °C -20 °C ... +50 °C Storage and transportation

temperature

Luftfeuchtigkeit 10 % ... 70 % 10 % ... 70 % 10 % ... 70 % Humidity

Anzahl und Typ Batterien

->BA OPTIMA7 0 0 Qty and size of batteries

Gehäusematerial ->BA OPTIMA7 Aluminium ABS Housing material

Gewicht ->BA OPTIMA7 9 kg 4,8 kg Mass

Maße/ mm ->BA OPTIMA7 470*230*315 335*210*158 Size/ mm

Gasfluss Sonde Sample Gas Flow

Durchflussmenge (trocken, 0°C)

->BA OPTIMA7

1l/min 13 ln/min -

Flow (0 % RH, 0 °C)

Durchflussregelung (Differenzdruck-messung)

- ♦ -

Flow regulation (based on diff.

pressure)

Genauigkeit der Durchflussregelung

- ± 5% - Accuracy of Flow

Regulation

Max. Abgastemperatur abhängig vom Feuchtegehalt;

Typische maximale Temperatur beiScheitholz

350 °C

Max flue gas temperature depends on humidity content.

Typical max flue gas temperature with split logs

Page 29: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

29

16.2 Interfaces

Deutsch OPTIMA7 FSM –Staub-sammelkoffer

FSM-Waage English

Gasanschluss Sonde - ♦ - sample gas

connector

Gasanschluss Abgasanalyse ♦ ♦ -

Temperatur-anschluss Thermoelement, T-Gas

->BA OPTIMA7 - -

temperature connector

thermocouple, T-Gas

SD - Karte ♦ - - SD card

USB - Anschluss ♦ - - USB interface

Serielle Schnittstelle AUX AUX ♦ serial interface

Netzspannung zur Versorgung

-> OPTIMA7 100 - 240 V / 50 Hz / 1,6A

24V DC / 5A

100 - 240 V 50 Hz AC Power Supply

16.3 Probe

Deutsch Angaben English

Länge des Sondenrohres 300 mm Length of Sample Gas Probe tube

Durchmesser des Sondenrohrs

10 mm Probe Tube Diameter

Länge des Sondenschlauchs 4 m Length of Probe Tube Hose

Filtertemperatur in der Sonde, geregelt 70 °C Temperature of filter in the

probe, regulated

Sondenmaterial Edelstahl / stainless steel

PTFE

Material of Probe

Filtertype MRU #64389 Filter Type

Filtermaterial Borosilikat Material of Filter

Page 30: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

30

16.4 Measuring values

Deutsch Gesamtsystem English

Staubgehalt im Abgas * Sample Gas Dust Content

Messprinzip: Gravimetrisch ♦ Measuring Principle:

Gravimetry

Messbereich 0 ... 1000 mg/Nm3 Measuring Range

Mittelungszeit 15 min Average Time Interval

Genauigkeit der Filterwiegung

entsprechend der Massekonzentration

< 1 mg

< 5 mg/m3

Accuracy of filter weighing

corresponding to dust mass concentration

Erweiterte Messunsicherheit des Gesamtsystems

% v. MW @ Staubgrenzwert

gemäß TÜV Prüfbericht M-BI 1164-00/13

43% @ 20mg/m3

13% @ 60 mg/m3

25% @ 90 mg/m3

23% @ 100 mg/m3

27% @ 150 mg/m3

Expanded Uncertainty of Complete System

% reading @ Dust Emission Limit

acc. to TUV Report M-BI 1164-00/13

Elektrochemischer Sensor O2 Electrochemical Sensor

alle technischen Daten gemäß -> OPTIMA7 all technical data acc to

Elektrochemischer Sensor CO Electrochemical Sensor

alle technischen Daten gemäß ->BA OPTIMA7 all technical data acc to

Abgastemperatur TA Flue Gas Temperature

alle technischen Daten gemäß ->BA OPTIMA7 all technical data acc to

Kaminzug

Draft

alle technischen Daten gemäß ->BA OPTIMA7 all technical data acc to

* (A limit is now set for coal, the MU is always 40%; except 150mg limit, then the MU = 0%)

Page 31: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

31

16.5 Calculated values

Deutsch English

Gesamtes Absaugvolumen

Total Extracted Sample

Gas Volume

Absaugzeit 15 min Extraction Interval

Umrechnung auf Normzustand 0 °C, 1013 hPa Conversion to reference

Genauigkeit ± 5% Accuracy

Mittlerer Sauerstoffgehalt Average O2 value

Mittelungszeit 15 min Average Interval

Mittlerer CO Gehalt Average CO value

Mittelungszeit 15 min Average Interval

Umrechnung auf Bezugssauerstoffgehalt ♦ Conversion to reference O2

Staubgehalt im Abgas Sample Gas Dust Content

Umrechnung auf Bezugssauerstoff ♦ Conversion to reference O2

♦ = availiable

17 APPENDIX 17.1 Further menu items

17.1.1 Filter cages information

OK

Page 32: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

32

17.1.2 FSM-Info

OK

Changing the instrument identification number (MIN) for the FSM-fine dust measuring module Please change the menu, FSM-info '. Connect the FSM-module. Showing the MIN is stored in the FSM module fine. To change the MIN, press the F3 key.

17.1.3 FSM-Scale info

To change the MIN, press the F3 key. With the OK button the value can be changed. Connect the FSM-scale and transfer the serial number automatically from the scale.

17.1.4 FSM Service values In case of a device failure, the error can usually be detected in the service value window. When contacting our service department be ready to provide the information shown below: FSM Service values window.

OK

Page 33: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

33

17.1.5 FSM bench test

OK

17.1.5.1 FSM Tightness test

OK

17.1.5.2 FSM Flow rate test

OK

Page 34: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

34

17.2 Replacement - and Consumables

Particulate matter collecting filter

Package unit 10 filter cartridges

(1 filter cartridge for every measurement required) 64872

Filter cage

stainless steel perforated sheet, complete with sealing cone 64390

Alu – transport container, particulate matter-proof, screwed

for the safe transport of FSM filter cage and filter cartridge 64510

Filter for condensate separator

Backup - filter to protect the internal pump (in case the Particulate matter collecting filter is not connected ) 64382

Sight glass for condensate trap at the FSM case 64493

Gasket for probe tube,

from temperature-resistant Viton 64559

Options:

Filter set

Complete with 2 m hose 64828

Nachfüllgranulate:

Purafil

100 g incl. Filtertab 60825

Silicagel

100 g incl. Filtertab 60826

Activated carbone

100 g incl. Filtertab 60827

Page 35: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

35

17.3 Declaration of conformity FSM module

Page 36: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

36

18 GENERAL INFORMATION ABOUT THE FSM SCALE 1. The FSM scale is not designed as a mobile element of the measurement of particulate mattern,

the scale remains at a site in the user's office

2. The FSM scale can be shipped in the manufacturer's packaging. For packing and unpacking the scale the notes in chapter 4.1 & 4.8 are to be observed. Retain for future use on the transport packaging!

3. The FSM scale is a precision instrument that can be damaged by improper handling. Note especially the assembling instructions for mounting the pan btacket and the maximum load of the balance of 50 g

4. The FSM scale as a precision instrument requires a suitable place. Please follow the appropriate instructions in chapter 6.5. Check it after the assembling the scale precision reached by the scale test program in OPTIMA 7, as described in chapter 6.6.

5. The FSM - scale compensated fluctuating environmental conditions independently. To avoid an erroneous determination of the dust content by changing the scale properties over a long period of time, to pre-and reweighing are performed each FSM filter cartridge within a period of one week.

6. The FSM scale must remain in the same place between pre-and reweighing a FSM filter cartridge to avoid measurement errors caused by positional changes.

7. Leave the FSM scale on power and turn it only with the On / Off - key on the FSM scale off and on. This will give you a quick measurement readiness of the FSM scale.

1. The FSM scale can only be operated via the OPTIMA 7 and may be used only for the weighing of FSM filter cartridges within the MRU dust measurement system.

19 UNPACKING, ASSEMBLING AND STARTUP 19.1 Unpacking Carefully remove the FSM scale from its packaging Max. load of the FSM scale is 50 g!!! (Risk of destruction)

At www.youtube.com you can find a short video film "FSM scale" (search for "MRU GmbH") for correct commissioning of the FSM level.

Remove the transport protection from the drying

Page 37: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

37

19.2 ASSEMBLING

Insert chamber cover Insert pan bracket.

Pay attention to the mandrel! Drawing on page 35 below!

Insert pan handle

Insert disponable pan

Page 38: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

38

Setting drying pan’s supporter (bracket).

1 – disposable pan

2 – pan bracket

3 – pan handle

4 – chamber cover

5 – chamber basis

When inserting the pan bracket caution!

Max. load 50 g !

When assembling the pan supporter (bracket) pay special attention to the correct positioning of its mandrel. The mandrel features a cut for its unique positioning against the pan handle, thus preventing their contact and friction. Setting pan’s bracket: On assembling the pan bracket onto the mandrel, turn the bracket slightly, so that the cut on the mandrels are located in their unique and correct position,

Page 39: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

39

19.3 Level setting

Important: An accurate weighing is only possible with a properly aligned scales.

Level the FSM scale using adjustable feet. Turn the feet in a way that the air bubble of the level is located centrally.

Page 40: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

40

19.4 Place of installation

The FSM scale requires a precision instrument to reach a suitable place for the required measurement accuracy. One such suitable measuring situation can be realized by simple means in each office and the achieved accuracy of measurement to be tested.

• Since the FSM scale is heating up the filter cartridge of the installation site should provide good air circulation (space about 20 cm each side and 1 m above the FSM scale).

• The FSM scale is to be free of vibrations on a stable surface. For this, a separate table in office quality is sufficient.

• Drafts as it occurs near open windows or doors should be avoided during each measurement.

• Direct sunlight should be avoided during the weighing.

• A typical suitable location for the FSM scale is therefore a separate office table on one wall of the office to which any other work to be carried out during the weighing.

19.5 Connect OPTIMA 7 with the FSM scale

Connect FSM scale using RS232 cable adapter to the OPTIMA 7 AUX connector. Connect the power cord to the FSM scale.

The keyboard of the FSM scale is locked after power on. An operation is necessary only on the OPTIMA 7. Similarly, the display of the FSM scale is for information only, you will receive all the necessary information about the OPTIMA 7.

Before switching on the FSM scale, i.e. BEFORE a zeroing the scale tray must be used. If a zeroing after about 30 sec. is not completed, must be examined - whether weighing pan WITHOUT an additional weight such as one sleeve is in the balance.

For a re-zeroing the FSM scale must be disconnected from the power supply first.

Die The FSM scale is at any time via the On / Off - key switched off and on.

Page 41: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

41

Checking the scale

After the FSM scales unpacked and check was placed in a suitable place and aligned please use the function of the scale.

The re-test is used to determine the repeatability of the FSM scale.

OK

The aim is to examine the repeatability of weighing of the filter cartridge with the FSM scale is in the appropriate range. For this, a standard filter cartidge is measured several times in succession. A standard filter cartidge is any of your existing filter cartidges with new filter inserted.

Simply connect the balance to the power supply, turn the scale on and allow the system to stabilize for 2 hours.

Connect the FSM-scale with the OPTIMA7.

1. Select one of your FSM filter cartidges for this process as "Standard filter cartridge" and insert the number of this cartridge in OPTIMA 7 under the menu item "Create standard fiter cartidge".

2. Select the menu item "standard filter cartridge weigh" and perform a complete weighing procedure with this standard filter cartridge.

3. Repeat after the other later in the day or the weighing of the standard filter cartridge as often, at least 3 times

4. In the menu item "View Statistics" you can see the individual results of the standard filter cartidge weighings. In the menu item "Evaluation filter cartidge" you see summary of the stored standard filter cartridge weighings (number of weighings, minimum and maximum values, as well as mean and dispersion (standard deviation). The FSM scale is at your installation site be fully functional if the achieved dispersion is less than 1 mg

Page 42: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

42

19.6 Bench test of the FSM scale

The bench test is used to check the FSM weighing accuracy for official measurements.

OK OK

Review of Weighing accuracy

a. Setting up the FSM scale, leveling, plug in, switch and for 2 h to acclimate.

b. Connection of a OPTIMA7 with the FSM scale.

c. Selection of test program and preparation of an empty FSM filter cage.

d. Weighing the empty FSM filter cage. Weight about 29 g.

e. Removing the FSM filter cage and cooling the FSM scale.

f. Reweighing the empty FSM filter cage plus a reference weight of 10 mg.

g. The reported difference in weight from pre and reweighing must be 10 mg +/- 1mg.

19.7 Measurement of the heating temperature of the FSM scale

The drying temperature of the scale is checked by a temperature gauge. The temperature gauge is to be inserted in a cage without filter (must be unlocked by cone, see photo). Then a 10-minute drying is carried out with this filter. Then remove the gauge: it should show a uniform density to 110 ° C.

Use this one-time-use temperature measuring strips, eg Article no. 2077.1 from Carl Roth.

Page 43: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

43

19.8 Error messages

Er 1 Hi Start mass exceeds permissible range (it refers to start and adjustment – start mass)

FSM scale is not unloaded Pan bracket is not inserted correctly?

Er 4 Fu Overload scale

Er 5 Ad A/D converter error Ad Tc L Error of temperature sensor

Ad Tc H Error of temperature sensor

If at an error message a restart the FSM scale, switch off - and on again (including pulling the plug) does not help, please contact a service MRU center (www.mru.eu).

19.9 Packaging

To transport the FSM scale (eg. Test facility) it's ABSOLUTELY necessary, the keep the FSM scale by the original packaging (including the foam inserts) to protect!

When removing the pan bracket exercise caution and thereby exert pressure on the pan bracket!

Page 44: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

44

20 Declaration of conformity FSM scale

Page 45: OPERATION MANUAL PARTICULATE MATTER MEASUREMENT€¦ · , which manages the filters used and their weight and displays the final results. The OPTIMA 7 performs all the necessary calculations

OPERATION MANUAL PARTICULATE MATTER MEASUREMENT FSM

45

Änderungen vorbehalten! Ausgabe: 20131203 Ausgabe: 20131216 Ausgabe: 20131217 Ausgabe: 20140120 Ausgabe: 20140121 Ausgabe: 20140711 Ausgabe: 20141009 Ausgabe: 20141117 Ausgabe: 20181220 Ausgabe: 20190405