multislice perfusion imaging with arterial spin labelling · 2005. 2. 9. · multislice perfusion...

98
Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A. Giil, B. Sc. Medical Physics Unit McGill University Montreal, Quebec, Canada A thesis submitted to the Faculty of Gtaduate Studies and Research in partial fuifilIrnent of the requirements of the Degree of Master of Science in Medical Physics. @Bradford A. Gill, August 1999

Upload: others

Post on 22-Sep-2020

6 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI

Bradford A. Giil, B. Sc.

Medical Physics Unit

McGill University

Montreal, Quebec, Canada

A thesis submitted to the Faculty of Gtaduate

Studies and Research in partial fuifilIrnent of the

requirements of the Degree of

Master of Science in Medical Physics.

@Bradford A. Gill, August 1999

Page 2: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

National Librafy Bibliothèque nationafe du Canada

Acquisitions and Acquisitions et Bibîiiraphic Services services bibliographiques

The author has granted a non- exclusive licence ailowing the National Library of Caaada to reproduce, loan, distriibute or seil copies of this thesis in microform, paper or electronic formats.

The author retains ownership of the copyright in this thesis. Neither the thesis nor substantial extracts &om it may be printed or otherwise reproduced without the author's permission.

L'auteur a accordé une licence non exclusive pennettant é la Bibliothèque nationale du Canada de reproduire, prêter, distribuer ou vendre des copies de cette thèse sous la forme de microfiche/nlm, de reproduction sur papier ou sur formai électronique.

L'auteur conserve la propriété du droit d'auteur qui protège cette thèse. Ni la thèse ni des extraits substantiels de celle-ci ne doivent être imprimes -

ou autrement reproduits sans son autorisation.

Page 3: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Abstract

This thesis presents the design, implementation and testing of a perfusion-weighted

magnetic resonance imaging sequence that is capable of acquiring several slices at a

time. Various methods of image acquisition and perfiision contrast generation are con-

sidered and tested.

The results of multi-slice acquisitions are compared with those from single-slice se-

quences in a variety of expenments to elucidate the effects of various imaging parameters

on the measured perfusion signai. A seven slice version was implernented, and was

found to give good results in the tests perfonned. This sequence will be useful in

perfusion-based finctional magnetic resonance imaging studies where the region of

interest can not be covered with a single image slice.

Page 4: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Resumé

Cette thèse présente la conception, l'implantation et l'essai d'une séquence d'imagerie

par résonance magnetique de la perfusion ayan la capacité d'acquerir plusieurs tranches

à la fois. Diverses méthodes d'acquisition d'images et de génération de contraste de

perfusion y sont considérées et testées.

Le résultat des acquisitions multi-tranches sont comparées a celles issues de séquences

à tranche simple, dans une séries d'experiences dont le but est d'élucider les effets de

divers paramètres d'imagerie sur le signal de periùsion mesuré. Une version a sept

tranches fut implantée et produisit de bons résultats lors des essais. Cette séquence sera

utile pour les études d'imagerie fonctionelle par résonance magnétique basées sur la

perfusion ou une tranche simple s'avérera insuffisante pour couvrir la zone d'interêt.

Page 5: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Acknowledgements

1 would like to thank my supervisor, Dr. G. Bruce Pike, for providing guidance, support

and an open and stimulating environment in the imaging lab.

1 wouId also like to thank my fellow students and CO-workers at the Neuroimaging

Lab, and in the Medical Physics program. 1 would especially like to thank Dr. Gérard

Crelier, Dr. Rick Hoge, Patrice Munger, John Sled, Dr. Jeff Atkinson, Valentina Petre

and Andreas Lazada for their technical assistance and advice.

For invaluable help in various administrative matters I would like to thank the Med-

ical Physics Graduate Secretary, Margery Knewstubb, as well as BIC Administrative

Assistant Jemifer Quinn.

My gratitude is also extended to the Medical Research Council of Canada, who

provided the financial support that made this project possible.

Page 6: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Contents

Abstract 1

Resumé

Acknowledgements 3

Table of Contents 4

List of Figures 6

Glossary of Terms 7

1 Introduction 9

2 Background 12 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2 MRSignal 12

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.2 Imaging 19 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.3 Fast Imaging 26

. . . . . . . . . . . . . . . . . . . . . . . . . 2.4 Functional Brain Imaging 28 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2.5 BOLDContrast 33

. . . . . . . . . . . . . . . . . . . . . . . 2.6 Perfusion Weighted Imaging 35

3 Arterial Spin Labelling 36 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.1 Introduction 36

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.2 Theory 37 . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3 Pulsed ASL Sequences 42

. . . . . . . . . . . . . . . . . . . 3 .3.1 Fast Acquisition Techniques 42 3.3.2 Spin Tagging Approaches . . . . . . . . . . . . . . . . . . . . 48 3 .3.3 Considerations for Multi-Slice Acquisitions . . . . . . . . . . . 54

4 Methods 56 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.1 Sequence Design 59

4.1.1 Low-Resolution Acquisitions . . . . . . . . . . . . . . . . . . 59

Page 7: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

. . . . . . . . . . . . . . . . . . . . . . 4.2 Spin-Tagging implementation 60 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2.1 FAIR 61 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.2.2 STAR 61

. . . . . . . . . . . . . . . . . . 4.2.3 Post-Processing of Image Data 62 . . . . . . . . . . . . . . . . . . . . . . . 4.3 The Perfusion Measurement 63 . . . . . . . . . . . . . . . . . . . . . . 4.3.1 Subject Immobilization 64

. . . . . . . . . . . . . . . . . . . . 4.3.2 Functional Data Processing 65 . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.4 Experimental Design 66

. . . . . . . . . . . . . . . . . . . . . 4.4.1 Single-Slice Experiments 67

. . . . . . . . . . . . . . . . . . . . . 4.4.2 Multi-Slice Experiments 67 . . . . . . . . . . . . . . . 4.4.3 The Effect of Inversion Slab Width 68 . . . . . . . . . . . . . . 4.4.4 Slice Profile Effects on Perfusion Data 68 . . . . . . . . . . . . . . 4.4.5 The effect of different inversion times 68

. . . . . 4.4.6 A Seven-SIice Perfusion-Based Functional Experiment 69 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4.5 QUIPPS 70

. . . . . . . . . . . . . . . . . . . . . . . . 4.6 Reproducibility of Results 70

5 Results 71 . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1 Single-Slice Results 71

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1.1 FAIR 71

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.1.2 STAR 73 . . . . . . . . . . . . . . . . . . . 5.2 The Effect of Inversion Band Width 73

. . . . . . . . . . . . . . . . . . . . . . . . 5.3 The Effect of Slice Profile 74 . . . . . . . . . . . . . . . . . . . . . . . 5.4 The Effect of Inversion Time 75

. . . . . . . . . . . . . . 5.5 Comparison of Response Seen Across the ROI 77 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.6 QUIPPS 78

. . . . . . . . . . . . . . . . . . . . 5.6.1 Reproducibility of Results 78

6 Discussion 80 . . . . . . . . . . . . . . . . . . . . . . . . . . . 6.1 Experimental Results 80

. . . . . . . . . . . . . . . . 6.2 The Seven-Slice GRASE-FAIR Sequence 84 . . . . . . . . . . . . . . . . . . . . . . . . . 6.3 The Interleaved Sequence 84

7 Conclusions 86

Bibliography 87

Page 8: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

List of Figures

. . . . . . . . . . . . . . . 2.1 Nuclear effect of an extemal magnetic field 14

. . . . . . . . . . . . . . . . . . . . . 2.2 Free Induction Decay schematic 15

. . . . . . . . . . . . . . . . . . . . . . . . . . 2.3 The spin-echo sequence 19

. . . . . . . . . . . . . . . . . . . . . . . 2.4 One-dimensional localization 21

. . . . . . . . . . . . . . . . . . . . . . . . . 2 -5 Slice-selective excitation 24

. . . . . . . . . . . . . . . . . . . 2.6 Two-dimensional spin-warp imaging 25

. . . . . . . . . . . . . . . . . . 2.7 Three-dimensional spin-warp irnaging 26

. . . . . . . . . . . . . . . . . . . . . . . . . . . 2.8 Image reconstruction 27

. . . . . . . . . . . . . . . . . 2.9 Functional imaging expenmental design 32

. . . . . . . . . . . . . . . . . . . . . 2.10 BOLD contrast source schematic 34

. . . . . . . . . . . . . 3.1 Echo-planar imaging pulse sequence schematic 43

. . . . . . . . . . . . . . . . . . . . 3.2 GRASE pulse sequence schematic 45

. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3 The GRASE signal 46

. . . . . . . . . . . . . . . . . . . . . . . . 3.4 EPI-STAR image formation 50

. . . . . . . . . . . . . . . . . . . . . . . . . . 3.5 FAIR image formation 51

. . . . . . . . . . . . . . . . . . . . . 4.1 Visual stimulus delivery system 58

. . . . . . . . . . . . . . . . . . . . . . . . . . 4.2 Visual stimulus pattern 58

Page 9: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

4.3 Hyperbolic secant pulse profile . 1 0.24ms duration . . . . . . . . . . .

1.4 Hyperbolic secant inversion pulse profile . 2.56ms duration . . . . . . .

. . . . . . . . . . . . . . . 5 . 1 Cornparison of FAIR acquisition sequences

. . . . . . . . . . . . . . . . 5.2 Comparison of STAR acquisition sequences

5.3 The effect of inversion band width on measured relative perfùsion change .

. . . . . 5.4 Effect of slice separation on measured relative perfusion change

5.5 The effect of varying inversion time on the measured relative perfusion

change . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

5.6 Seven-slice acquisitions with visual area positioned at three separate

. . . . . . . . . . . . . . . . . . . . . . . . . . . . locations intheRO1

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5.7 QUIPPS Results

Page 10: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Glossary of Terms

ADC ; ANOVA : ASL : BOLD : CBF : CBV : CMRO : EEG : EPt : FAIR : FID : FFT : fMN: FSE : GRASE : MEG : MN1 : M R : Ml21 : MT: NMR : PET : PICORE : QUIPSS : RARE : RF: ROI : SE : SPECT : SNR : STAR : T : TI : T2 : T; : TE : TIC : TR:

analogue to digital converter analysis of variance arteriai spin labeling blood oxygenation level dependent cerebral blood flow cerebral blood volume cerebral metabolic rate of oxygen elec troencep halography echo planar imaging flow altemating inversion recovery fiee induction decay fast Fourier transfonn functional magnetic resonance imaging fast spin ec ho gradient and spin echo magnetoencephalography Montreal Neurological Institute magnetic resonance magnetic resonance imaging magent ization transfer nuclear magnetic resonance positron emission tomography proximal inversion with a control for off-resonance effects quantitative imaging of perfusion using a single subtraction rapid acquisition with relaxation enhancement radio fiequency region of interest spin echo single photon emission computed tomography signal to noise ratio signal targening using altemating radiofiequency Tesla spin lattice relaxation time constant spin-spin relaxation time constant transverse relaxation time constant echo time tirne-intensity curve repetition time

Page 11: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Chapter 1

Introduction

The high-resolution anatomical images produced by magnetic resonance imaging (MM)

scanners have become increasingly important in modem medicine, and the rapid growth

of MRI technology in the past decade has allowed researchers to explore new apptica-

tions of this imaging modality.

In addition to imaging simple anatomy, MR imaging methods that measure dynamic

processes in the body are being actively explored. These methods, which attempt to

image some aspect of biological function, are broadly referred to as functional MRI, or

simply fMRI. The most intensly pursued area of fMRI is the mapping of brain function.

and it is this application which motivates this thesis. In fact, in the context of this thesis,

the term fMRI will explicitly refer to the imaging of brain function.

MRi depends on stimulated signals fiom certain nuclei in the body. This endoge-

nous signal source permits great flexibility in the possible imaging strategies.

fMRI brain mapping is perfonned by acquiring a series of images in the brain re-

gion of interest while sorne stimulus is presented or task is performed. The brain areas

activated during the experiment will show changes in local neuronal and metabolic ac-

Page 12: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

tivities, and the relevant portions of the brain images involved will show signal changes

that are correlated with the stimulus or task. fMRl imaging sequences are designed to

be sensitive to some aspect of the metabolic changes associatied with neuronal activity

in order to produce and maximize these signal changes.

The most common method of producing neuronal activity-correlated signal changes

for use in fMW studies is the use of blood oxygenation ZeveZ dependent (BOLD) con-

trast technique. This contrast source relies on the different signal decay rates seen

between regions with diffenng concentrations of deoxyhaemoglobin in the local blood

supply. Activation increases the amount of blood fiowing into an area of the brain and

decreases the deoxyhaemoglobin concentration locally. Signal fiom this area will have

an altered relaxation time constant which c m be exploited to provide image contrast.

BOLD contrast-based fMEU using a single-shot echo-planar acquistion provides good

results, and is relatively simple to implement.

A more direct method of measuring neuronal activity is the measurement of cere-

bral blood flow (perfusion), and this can be performed by altering the magnetization

of blood flowing into the region of interest. Such methods are conceptually similar

to radioactive labelling techniques such as PET (positron emission tomography) and

SPECT (single photon emission computed tomography). These methods have a po-

tentially broader applicability than BOLD, and can be used for quantitative blood flow

measurernents.

The use of BOLD contrast in the vast majority of brain fMRI studies has been

partly due to the limited spatial extent of non-invasive perfusion-weighted imaging

techniques. Contrast agents can be used in MRi perfusion measurements with im-

proved spatial resolution, but problems with build-up in multiple-trial experiments, cost

and increased experimental complexity have limited their use. The goal of this thesis

Page 13: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

was to design and implement non-invasive, multi-slice, perfusion-weighted irnaging

sequences to increase the spatial coverage of these techniques, which should greatly

improve their usefilness in research applications.

The thesis consists of seven chapters. The second begins with a brief review of ba-

sic M W theory and proceeds to outline fast imaging techniques, and some applications

of them in the pursuit of fûnctional brain imaging. The signal theory and spin tagging

approaches used in pulsed arterial spin labelling (pulsed ASL) perfusion imaging are

discussed in Chapter Three, along with considerations for the extension of pulsed ASL

to mu1 tiple slice acquisitions. The experimental procedures for both functional imaging

studies and for the individual tests performed to detennine the effect of vanous imag-

ing parameters on functional results are outlined in Chapter Four. The resul ts of the

performed experiments are presented in Chapter Five. The discussion section, Chapter

Six, reviews the results in the context of the function of a seven-slice perfusion acqui-

sition. Chapter Six concludes with the discussion of results obtained at our lab fiom a

sequence used to obtain simultaneous BOLD/perfÜsion measurements in experiments

to elucidate metabolic fùnctioning of the primary visual area of the human brain.

Original contributions made in this thesis are: 1) the development and testing of

low-resolution, single-shot Gradient and Spin Echo (GRASE), half-Fourier GRASE

and echo-planar acquisition and reconstruction sequences; 2) the development and

testing of Flow-measurement using Altemating Inversion Recovery (FAIR) and Sig-

nal Targetting with Altemating Radiofiequency (STAR) pulsed ASL sequences using

the above acquisition techniques; 3)expenmental evaluation of the effects of varying

various sequence design parameters on multi-slice pulsed ASL sequences; and 4) the

design of an interleaved BOLDEAIR echo-planar sequence and its use in numerous

basic activation physiology experiments.

Page 14: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Chapter 2

Background

2.1 MR Signal

Magnetic resonance imaging ( M N ) is based on the nuclear magnetic resonance (NMR)

phenomenon which arkes from certain nuclei in the presence of external magnetic

fields.

The NMR phenomenon involves nuclei that possess an odd number of protons

andor neutrons. Such nuclei necessarily have non-zero angular momentum and rnag-

netic moment, and, as shown by Stem and Gerlach in 1922 [22], they can be made to

interact with extemal magnetic fields to yield observable results. Stem and Gerlach's

experiment involved the splitting of a beam of Ag atoms by an externally-applied, in-

homogeneous magnetic field.

In 1946, Bloch and hrcel l [7, 721, working independently, were given credit for

discovering NMR; a feat for which they were awarded a Nobel Prize in Physics in 1952.

They provided the f b t theoretical insights into the physics describing the behaviour of

the excited nuclei, and it is an extension of their basic experimental setup that is used

Page 15: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

in al1 NMR spectroscopy measurements and MRi.

Although it is a quantum-mechanical process, a classical description of the physics

behind NMR is adequate in an explanation of MN. In the simplest case - that of an ex-

ternal magnetic field interacting with the magnetic moments of spin-possessing nuclei

in a sample - the magnetic moment of each spin 1/2 nucleus behaves as a small bar

magnet, and the extemal field will attempt to align these nuclear moments with itself.

individual nuclei cannot be treated classically, and are forbidden from aligning entirely

with the external field. Instead, these nuclei show a majority alignment with the exter-

na1 field, with a defined projection of the magnetic moment along the direction of the

applied magnetic field. In addition to the aligning of the nuclear magnetic moments, the

nuclei in the external field are rapidly precessing. The basic precessional behavior of

nuclei possessing intrinsic angular momentum and a magnetic moment in an external

magnetic field is described by the Larmor Equation:

where B is the magnetic field strength, w the angular precessional frequency of the

atomic nucleus or 'Larmor' frequency and y the gyromagnetic ratio of the nucleus. The

gyromagnetic ratio is a constant unique to eacb type of particle with spin. 'H nuclei

are used in most MRI applications due to their prevalence (by far the most numerous

nucleus in biology) and their high sensitivity. The gyromagnetic ratio for LH nuclei is

42.58 MHdTesla Cl].

Conventionally a Cartesian coordinate system is used to describe orientation in

Page 16: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 2.1 : Formation of the net magnetization vector by the alignment of the majority o f the nuclear spins with the external field

MM, wherein the extemal field direction is along the z mis. ' The physics goveming the behavior of the individual nuclei are quantum mechani-

cal in nature, and a quantum mechanical restriction exists on the possible orientations

of 'H nuclei in extemal magnetic fields. More specifically, because 'H nuclei are spin

1/2 nuclei, two orientations are permitted, one with the projection of the magnetic mo-

ment of the nucleus parallel and the other anti-parallel to the extemal field [22]. The

potential energy possessed by a spin depends on its orientation in the field. The two

orientations can therefore be referred to as separate energy levels.

The population of the spins in the sample is split between the two energy levels with

the lower-energy @araIlel orientation) being favoured by a ratio given by the Boltzmann

factor [l]. At room temperature, the factor is very close to unity because the energy

level difference of the spins is small compared to the thermal energy of the molecules.

The presence of a majority of spins aligning with the extemal field has the effect of

polarizing the sample and creating a net magnetization vector (M) aligned parallel to

the extemal field (Fig 2.1 ). It is the manipulation of this vector that forms the basis of

al1 NMR-based measurements including M M [ 1 1.

'Vector quantities are indicated by bold font, while normal font will represent scalar quantities.

14

Page 17: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 2.2: Free Induction Decay (FID) schematic: the experimental setup (lefi) mea- sures the signal fiom the magnetization vector precessing in the transverse plane. The received signal is dependent on the magnetization vector in the transverse plane, and decays away exponentially with a time constant Tg. At the same time M, recovers ex- ponentially with a different time constant. The Fourier transfonn of the decaying signal yields its fiequency components (spectrum) shown in the lower right.

The magnetization of a sample can be manipdated through the application of a

correctly-tuned radio-fiequency (RF) pulse whose orientation of propagation is perpen-

dicular to the direction of the extemal magnetic field. Such an RF pulse should be tuned

to the Larmor fiequency of the sample spins in order to efficiently transfer energy to

them. The classical description of the interaction between the pulse and magnetization

vector features the magnetic field component of the RF pulse (BI) rotating M away

fiom the extemal field direction. The resonance condition in the classical description

Page 18: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

exists in that, for the efficient rotation of M to occur. BI must be synchronized with

the rotating frame of the precessing spins - a condition only met if the RF and Larmor

fiequencies match. By controlling the duration and amplitude of the RF pulse used,

any rotation angle for M can be chosen.

The rotation of the magnetization vector by BI changes its orientation with respect

to the Bo field. The result then is that a component of the magnetization vector will

now be in the plane transverse to the Bo direction - except in special cases of rotations

that are integer multiples of 180".

An experiment in NMR consists of three basic steps: 1) the alignment of the nuclei

in an external Bo field to form a magnetization vector M; 2) the rotation of the vector

via RF fields; and 3) the detection of the resultant signal by coils surrounding the sam-

ple. The component of M rotating in the transverse plane changes the flux in a receive

coi1 and induces an electro-motive force. The detected signal fiom such an event is the

basis of al1 NMR experiments including MRI. A graphical depiction is given in Fig 2.2.

The signal fiom a simple excitation and detection routine is called a free induction

decay (FID). The strongest received signal strength occurs immediately following an

RF pulse that tilts !'II by 90" so that it lies entirely in the transverse plane. Analysis

of the FID signal fiom chemicai compounds fonns the basis of a simple experiment

in NMR spectroscopy. Mucb about the structure of molecules can be deduced from

the amount of magnetic shielding they provide to their constituent nuclei. The amount

of shielding given by chemical environment results in a change in the precessional

fiequency of nuclei. This change in precessional fiequency is called a chemical sh@.

The concept of chemical shifi is important in MRI in that species with a significant

chemical shift and abundance (eg. fat) can be incorrectly

The FiD signal strength decays away exponentially

localized.

fiom its initial value foliow-

Page 19: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

ing excitation. The re-growth of M to its steady-state strength and orientation is an

exponential process as well, but this re-growth occurs with a different tirne constant.

The time constant Tl describes Al's re-growth to the steady state M0z. Tl is called

the spin-lotlice relaxation time constant - so called by early NMR researchers [25] .

The basic mechanisms of a system's renim to steady state energy level populations are

the quantum interactions between excited spins and the surrounding molecular environ-

ment. More specifically, an excited spin can give up a quantum of energy to the lattice,

and to do this there needs to exist a magnetic field generated by nearby molecules that is

flucniating at the spin's precessional (Larmor) fiequency. These environmental or lat-

tice fiequencies originate fiom the thermal motion of nearby atoms. The time constant

is a statistical manifestation of the likelihood of these energy transitions fiom the popu-

lation of excited spins. The steady-state populations of energy levels are themselves the

result of dynarnic processes of excitation and energy exchange, and the retum to steady

state afier excitation depends on these same processes. Tl time constants depend on

temperatures and the molecular reonentation fiequencies - characterized by correlation

times, as these in part determine the lattice magnetic field fluctuation fiequencies. and

Bo field strengths as this determines the size of the energy quanta required to cause

transitions [25] .

The exponential decay of the magnetization in the transverse plane (FID) follows a

different time constant, T;. This relaxation behaviour results fiom contributions from

two separate sources: nuclear interactions and the effects of field inhomogeneity. The

contribution from field inhomogeneity can be nulled by experimental techniques. If

such techniques are implemented, the signal decays with T2, the true spin-spin relax-

ation time constant. T2 results fiom nuclear interactions including the transitions that

give rise to the Tl constant, but also Uiclude interactions that degrade the magnetization

Page 20: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

vector strength in the transverse plane by de-phasing the spins of the constituent nuclei.

The effect of Bo inhomogeneity on the decay of the transverse magnetization of

an excited sample comes about due to the different precessional fiequencies of the

nuclei in the different regions of the field. These effects can be removed by altering

the RF pulse sequence to inchde a 180" pulse following the initial excitation. Such a

pulse sequence (Carr-Purcell) [12] refocuses the signal that was de-phased by the field

inhomogeneity to form a spin echo. The process of spin-echo formation is shown in

Fig 2.3.

The peak of the spin echo signal is the point at which the spins refocus, eliminating

the effect of Bo inhomogeneity. The decay of the peak value of the spin echo is an

exponential process characterized by the T2 time constant. Subsequent 180" pulses can

be inserted at regular intervals to form an echo train of spin echos. This is comrnonly

done in clinical imaging procedures (e-g. Fast Spin Echo) to lengthen the time for

which signals can be acquired following the 90" excitation pulse [30].

The general behaviour of the magnetization vector can be described phenomeno-

logically by the Bloch equafion [68]:

where M, il&, My and A, are the magnetization vector and its Cartesian components

along the unit vector directions i, j and k.

Page 21: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 2.3: Schematic of spin echo sequence. Afier a 180" excitation is applied at T E / & the spins that had dephased due to field inhomogeneity begin to re-orient them- selves to form an echo at TE.

2.2 Imaging

The extension of NMR fiom simple excitation and reception to MRI involves the use of

additional hardware to mod ie the basic experiment. More specifically, the addition of

Iinear magnetic field gradients allows spatial encoding of the received signal to occur.

The gradient fields are oriented in the sarne direction as the Bo field, but their field

strengths Vary in a linear fashion within the magnet. This gives a spatial dependence to

the magnetic field strength within the field:

or more simply:

Page 22: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

where G,, G, and Gz are the field gradients along x, y and z respectively, and B( r) is

the external magnetic field strength at r.

The effect of these gradients is to change the precessional frequencies of the spins

at different locations within the sarnple:

A single gradient field allows simple localization of the spins. If two samples in a

Bo field are spatially separated in the presence of a linear gradient (say G,). and the

gradient direction is the same as that of the samples' separation, the received signal post

excitation will contain al1 of the precessional fiequencies of both samples. The preces-

sional fiequency of the spins in the sarnples will depend on their spatial location. A

Fourier Transform of the data set will then yield a result representative of the samples'

projection (Fig 2.4).

X-ray radiography is based entirely on projection imaging, and, x-ray Computed

Tomography (CT) reconstruction techniques can be applied to MR projection data sets.

The data for such a measurement is acquired at many projection angles - corresponding

to several gradient directions, and reconstructed using cornputer algorithms to yield the

desired image. A more elegant way of acquiring and reconstmcting MR signals uses

multi-dimensional Fourier Transforms of the received MR signals to reconstruct the

image.

The MR signal can be manipulated by the linear gradient fields such that the re-

Page 23: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

f requency - Figure 2.4: 1D Localization: in the presence of a gradient field the precessional fie- quencies of the nuclei in the samples depend on their spatial location. The received sig- nal contains contributions from al1 of the nuclei in the sarnples. The Fourier Transform of the received data will quanti@ the amounts of nuclei precessing at each frequency, which is equivalent to the projection image of the sample.

ceived signal provides a Cartesian sampling of k-space (spatial fiequency Fourier space)

that will yield an image of the excited magnetization distribution after reconstruction

with a Fast Fourier Transform (FFT) algonthm. This can be illustrated by consider-

ing the signal equation for an excited two-dimensional sample of spins (given by the

solution to the Bloch Equation) in the presence of linear field gradients:

Page 24: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

where kz ( t ) and kg ( t ) are given by:

kx ( t ) = 1 GZ ( ~ ) d r . 27r O

and

or rather, the wclvenumber modulating the received signal is given by the time integral

of the field gradient as they are applied to the excited spins.

The two-dimensional Fourier Transform of the desired image is given by:

Equations 2.5 and 2.5 are identical in form, and imply that a point in the k-space

interpretation of an image can be inferred directly fiom the received signal strength

of the object if the excited spins are correctly manipulated by the gradient waveforms.

Furthemore, the k-space position of the received signal is given by the time integral of

the gradient wavefonns (Equations 2.6,and 2.7).

Slice-selective excitation utilizes short RF pulses that possess a discreet bandwidth

of frequencies - al1 pulses of finite duration must be composed of a range of frequencies

depending on the pulse shape. The pulses used are typically tnincated sinc or Gaussian

Page 25: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

in shape and centred around the central carrier fiequency which is the Larmor fiequen-

cy for the spins to be excited [68]. Sinc and Gaussian RF pulse shapes are prefemed

in most applications because they result in the most usefûl slice profiles for imaging

applications.

The process of selective excitation is based on the requirement that, for the spins

of the sample to receive energy fiom the RF field, there must exist resonance between

the precessing spins and the RF field, ie. the RF field must at least contain the Lar-

mor fiequency of the spins to be excited. If the nuclei experience a gradient field in

addition to Bo such that spins are precessing at fiequencies outside the bandwidth of

the excitation pulse, only those spins that precess within the region corresponding to

the RF bandwidth will be excited, leaving those spins precessing outside the RF range

unaffected (Fig 2.5).

After excitation, the received signal is demodulated (to remove the camer fie-

quency) and sampled by an analogue-to-digital converter (or ADC) before being stored

at the appropriate position in the raw data (k-space) matrix. The matrix is fiIled by

rapidly sarnpling data with the ADC as a gradient is applied to move the signal Iinearty

through k-space. The particular gradient used in this manner is retèrred to as the read-

out gradient. The k-space trajectory is moved in the direction perpendicular ta the

readout direction by the application of short gradient pulses. These carefully chosen

phase encode gradient pulses allow readout lines to be shifted so that a full two dimen-

sional sampiing of the desired range of k-space can be obtained. This strategy can be

directly extended to three dimensions by phase encoding in the slice selection direc-

tions. Simple two and three dimensional imaging sequences along with their k-space

trajectories are shown in Fig. 2.6 and 2.7; such sequences are exarnples of spin-waa,p

imaging.

Page 26: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

time 4 distance j +

profile

G-slice

Figure 2.5: Slice Selective Excitation Schematic: the RF pulse contains a well-defined range of fiequencies (shown by the Fourier Transform of the wavefom) which corre- sponds to a select range of precessional fiequencies in a sarnple of spins in the presence of a gradient field. The spins in this siice of the sample will be the only ones excited by the RF pulse.

The signal received by the ADC as the readout Iine is acquired (concomitant with

the passage between negative and positive k-space values) is called a gradient-wcczlled

echo. The centre position of k-space is weighted by the signal strength of the sample

spins precessing coherently. Positions off-centre are samples of the spins that have

a phase modulation in the directions of the gradients that were applied to the spins.

The echo is formed as the phase of the spins is untwisted from an edge of k-space

to the central (coherent) position and then re-modulated as the far edge of k-space is

approac hed.

Care must be taken in the acquisition of the k-space data for an image. The sample

spacing must be sufficiently close to ensure that the reconstructed image is fiee from

Page 27: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

N D I I

time - Figure 2.6: Two-dimensional spin-warp imaging: The sequence of RF and gradient pulses move the signal around in k-space. Each excitation is followed by a G, gradient pulse (the phase-encode gradient) that moves the signal along k,. and a shon negative and longer positive G, gradient pulse (the readout gradient) moving the signal left and then nght along k,. The Gy pulse is shown as a box as many repetitions of this sequence using different Gy gradient pulse magnitudes of both signs are used in the filling of k- space.

the effects of aliasing. These effects are avoided by ensuring that the image data is

sarnpled to fulfill the Nvquist criterion which dictates that the sampling frequency must

be at least double that of the highest frequency found within the received signal [13].

There is always a tradeoff in MRI between the achievable spatial resolution, acqui-

sition time and signal-to-noise ratio (SNR) possible in different scanning procedures.

The time taken by conventional s c a ~ i n g techniques is given by the product of the num-

ber of lines that need to be acquired and the interval between excitations (as one line is

acquired for each excitation in the simplest case).

One of the great strengths of MRI is the ability to provide images weighted with

several different types of contrast simply by varying some of the parameters of the

Page 28: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 2.7: Three-Dimensional Spin-Warp Imaging: The sequence in 2.6 is extended so that kz lines are now acquired following G , gradient pulses (phase-encode gradient). which move the signal along k:.

acquisition. CT scans provide image contrast based on subject x-ray attenuation co-

efficients. MRI acquisitions are commody weighted to provide image contrast based

upon proton density, Tl , and G. Newer, more intncate acquisition sequences can pro-

vide images weighted by other physiologically relevant parameters, such as difision,

perfision and flow velocity 1131.

2.3 Fast Imaging

Fast techniques have been developed which greatly reduce the time required to acquire

an image. Short imaging times are clinically useful in that the motion of an organ or

body part can be effectivelyfrozen if the image is taken quickly enough. There are also

physiological phenornena that would be impossible to measure with MR if the image

times were not appreciably short.

Page 29: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

magnitude raw data magnitude reconstructed image

Figure 2.8: An image is formed fiom its k-space data.

Fast imaging techniques reduce the time required to fil1 k-space through the use

of several different modifications of the basic sequence. The reduction of the repeti-

tion time between excitations (TR), the acquisition of multiple phase encode lines per

excitation and the reduction of the total number of phase encode lines are al1 used.

An effective method to reduce the time per scan is to acquire oniy partial k-space

data sets for individual images - there is a fundamental redundancy (Hermitian sym-

metry) in the k-space representation of a real image that allows reconstmction to occur

with as little as one half of the full data set.

In recent years, modifications to MR scanner hardware have enabled entire k-space

data sets to be sampled following a single excitation. Such modifications require pow-

e h 1 gradient drivers and finely-tuned coils to function comectly. So-called single-shot

techniques have permitted imaging of physiological phenornena in times that are typi-

cally in the tens of milliseconds [63].

Page 30: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

2.4 Functional Brain Imaging

The functioning of the human brain is an important topic in science, with far-reaching

implications for modern medicine. The gross anatomy of the brain has been more or

iess known for some time, but only in the past century has technology progressed to the

point where accurate fwictional importance can be ascribed to the myriad structures to

be found.

Brain function involves a complex physiological interrelationship benveen electri-

cal and chemically-mediated events. The basic unit of brain fùnction is the neuron.

which is an excitable ce11 that conducts electrical signals. or action potentials through

the use of voltage-gated ion channels in the cellular membrane.

The electrical activity of functioning neurons can be detected at short distances

by measurements of the electrical field. The technique of electroencephalography, or

EEG was originally performed directly on the surface of the brain in animals, but soon

afier the technique was extended to the placing of electrodes on the scalp. Simple

localization of the measured activity could be inferred by keeping track of the signal

amplitude measured at various locations on the scalp surface. The use of multiple

electrode arrays allowed these measurements to be done simultaneously [14].

Magnetoencephalography, or MEG, is a relatively new innovation that provides

higher-resolution activity maps than EEG (within a few millimeters as opposed to tens

of mm's in EEG) [27]. MEG is based on the detection of magnetic fields that are in-

duced by the neuronal electrical currents. MEG and EEG stand alone in methodologies

that non-invasively and directly measure electrical events in the brain, but these tech-

niques suffer the rigorous calculation routines required to do the source localizations

and the prohibitive cost of MEG machinery.

Page 31: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Auto-radiographic studies of animal brains begm in the 1950's. These techniques

were centred around the injection of a radioactive dye while the animal was being

exposed to a stimulus. Immediately following the stimulus, the animal would be sacri-

ficed, and a thin section of the animal's brain would be taken and placed in contact with

a radiographie plate [5 11. Functionaily important areas in the brain associated with the

stimulus would show preferential uptake of the radioactive tracer and expose the plate

to a larger degree. Development of the plate would allow these areas to be localized

with high resolution. Although modem scaming techniques have reduced the need for

animal autoradiograph studies in recent years, the technique still sees use due to the

high resolution of the data acquired.

The use of radioactive substances to measure brain fknction was also extended to

in vivo studies in humans. The passage of a radioactive bolus in the blood could be

quantitatively measured by the use of a pulse counter sensitive only to a small section

of the brain. The numbers of nuclear events recorded would provide an estimate of

the amount of radioactive blood in the small area of the sensor's field. Single photon

emission tomography (SPECT) and positron emission tornography (PET) were intro-

duced in the 1970's to provide images of radioactive tracers in the body [2]. PET in

particular was used to study several physiological parameters in the brain, and these

measurements could then be used as the basis for localizing brain function changes

brought about by various stimuli. PET had the advantage of localizing the radioactive

tracers within the human brain to a higher resolution than SPECT (about l5mrn) , with

an acceptably low radiation dose to the subject.

PET functional studies typically measure brain perfusion as the correlate of brain

activity in functional studies. Perfusion as a parameter quantifies the passage of blood

water into the tissue, and is an important physiological parameter, as it provides in-

Page 32: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

formation about possible brain pathology and it is the basis for calculations of other

brain physiology parameters [80, 171. For Our purposes, perfusion will be assumed to

be equivalent to measurements of cerebral blood flow (CBF)[2].

Problems with PET-based functional imaging include a limited spatial resolution

and a low, but significant radiation dose to the subject. Cost is also a factor; the isotopes

used in al1 human nuclear medicine-based imaging are very short lived, and do not store

well. PET imaging facilities require that a cyclotron or some other related equipment

be located nearby to generate the necessary tracers for the studies.

The explosion of MRI technology in the eighties provided a rneans of obtaining

high resolution anatornical images of the body without exposing the subject to a dose

of ionizing radiation. The extension of PET-based functional imaging techniques to

MRI was first studied in the late eighties [6. 5, 77, 761. In these studies, the radioactive

tracers were replaced with substances that locally disrupt the magnetic field seen by

blood water. Typically images of the region of interest would be rapidly acquired while

a bolus of contrast agent passed through. Fluctuations in quantitative measurements

of the signal fiom the region could then be correlated with the stimulus given to the

subject, and functional localization of the resultant brain activity could be obtained.

Problems with contrast agent-based functional MRI (fMRi) are in large part due

to the limited reproducibility of experiments due to the accumulation of contrast agent

in the blood of the subject. The situation changed with the discovery that endogenous

contrast sources related to correlates of neuronal activation existed and could be utilized

in functional studies by correct pulse sequence designs. Images could be weighted to

show contrast based on blood oxygcnation or perfusion of tissues - through the use

of inversion pulses to prepare in-flowing blood so that it would act as its own contrast

agent [ 1 81.

Page 33: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

MR-based fûnctional studies follow several different paradigrns, with the block-

design study being the most common, In a block study, the experiment consists of a

series of altemating t h e penods (blocks) of stimulus presentation or task performance

and control. Throughout the study, images are acquired of the region of interest in the

brain (Fig. 2.9). More recent experimental design modifications involve changing the

time course of stimulus presentation. Event-related functional studies allow researchers

to investigate brain responses to stimuli of very short durations - a more realistic repre-

sentation of many normal fünctions [78].

A functional imaging study can produce results of various forrns. The raw data set

afier reconstruction, is composed of a series of images that show the same region of

interest throughout the time course of the study. From the raw data, activation maps,

time-intensity curves and quantitative information c m potentially be calculated (Fig

2.9).

Activation maps are images that represent the statistical significance of a voxel-by-

voxel companson of the intensity values in the raw data set with the applied stimulus

in the study. A wide variety of statistical methods are used in the analysis of functional

data sets [26,90].

Time course data, or time-intensity curves, are plots of the pixel values in a region

of interest across the time scale of the raw data set acquisition. Modulations in pix-

el values averaged or summed across the region of interest show the degree of signal

change brought about by the stimulus. Regions of interest centred on areas with fiinc-

tional significance to the applied stimulus (ofien defined via an activation map) should

show changes in pixel values closely correlated with the stimulus. Other regions not

functionally involved with the stimulus should show no modulations in pixel values

outside of normal noise values.

Page 34: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 2.9: fMRI: the block design schematic (lefi) outlines a comrnon procedure used in fMRi. Many images are acquired of the same slice or slices (top) while a stimulus is applied for discrete time periods. A tirne-intensity curve (right) shows modulations of the pixel values in a region of interest across the time course of the experiment which correspond to the times of stimulus presentation in the block design.

Quantitative data fiom activation studies attempt to relate the degree of pixel val-

ue change seen between activated and baseline States through the course of the time-

intensity curve to a physiologically relevant scale. Periùsion-based functional studies

are good candidates for quantitative results, as perfusion is itself one of the most phys-

iologically important parameters used to signie the degree of brain activation.

Page 35: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

2.5 BOLD Contrast

The basis of blood oxygenation level dependent (BOLD) contrast techniques is that

the magnetic nature of the haemoglobin molecule within red blood cells is dependent

on whether or not it is carrying oxygen. Deoxyhaemoglobin, possesses an unpaired

electron and hence is paramagnetic as this electron will align its magnetic moment

with the applied field. The magnitude of the magnetic moment of an unpaired electron

is many times that of a 'H nucleus and will have a strong disruptive effect on the signal

seen From nearby spins due to its effect on the local B field . Oxyhaemoglobin is

slightly diamagnetic as paired electrons will align in opposite orientations in an extemal

field and cancel one another's effect on the local field environment. Such molecules

contain electronic currents inducing magnetic fields that weakly oppose the applied

magnetic field - hence the diamagnetism. At 1.5T, T1 of blood varies from 30ms to

25Oms for the range of oxyhaemoglobin concentration from 30% to 96% [91]. The

susceptibility difference between the two oxygenation States is [82]:

A x ~ ~ ~ ~ ~ - l x o a = +0.18 x 106

(2.9)

The phase change in spins caused by deoxyhaemoglobin's field disruptions shorten

the T,* relaxation time constant locally. BOLD-weighted images are acquired with

timing parameters chosen to maximize these T.'. differences in the generation of image

contrast.

The exact relationship between the BOLD signal change and functional activation

is complex, and still a matter of some debate [45]. What is observed is that areas or"

increased neuronal activation show an increase in cerebral blood flow and a decrease in

the concentration of deoxygenated blood. This results in a higher signal fiom these acti-

Page 36: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 2.10: (top) BOLD Baseline: The blood passing through the capillary delivers a set supply of oxygen to the tissue. A concentration of deoxyhaemoglobin is present that shortens the T,* tirne constant of the local signal. (bottom) BOLD Activation: Brain activation stimulates an increase in the local blood flow resulting in an infiux of oxyhaemoglobin which decreases the concentration of deoxyhaemoglobin in the capillary and lengthens the local T; time constant.

vated areas during image acquisition. Since the images used in BOLD studies are heavi-

ly T;-weighted, they are also susceptible to various artifacts. In particular, T;-weighted

images are degraded by the effects of magnetic field inhomogeneities brought about by

magnetic susceptibility differences. Such di fferences in susceptibility are prevalent at

Page 37: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

the interfaces between tissue and air; limiting the use of BOLD contrast in brain studies

where the region of interest lies near the air sinuses or brain stem, for example.

2.6 Perfusion Weighted Imaging

Measurements of brain perfûsion are of use in the diagnosis of disease and brain injury,

and c m also form the basis of functional brain mapping.

MN-based measurements of perfusion were first performed in the late 1980's using

exogenous contrast agents [62]. The subject in these studies would be injected with a

paramagnetic contrast agent which would be imaged as it passed through the region

of interest [77]. Ln the early 19901s, methods were developed in which the contrast

agents were replaced by magnetic labeling of blood water spins immediately proximal

to the region of interest. These methods. generally referred to as artenal spin labelling

(ASL), effectively transformed blood water into an endogenous contrast source and

allow increased repeatability of perfusion experiments in a single session.

Problems with MRI-based perfusion measurements have stemmed fiom low SNR.

the difficulty of producing quantitative results and the limited spatial extent that can be

imaged at once - early methods could only be performed in a single slice. The physical

basis of ASL techniques are expanded upon in the next chapter.

Page 38: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Chapter 3

Arterial Spin Labelling

3.1 Introduction

Arterial spin labelling (ASL) perfusion imaging involves the acquisition of two sepa-

rately prepared images. These images are then subtracted to yield a d~rerence Nnage

that will have pixel values dependent on the amount of perfusion occurring in the im-

aged region. These images are generally called tag and conrrol images - depending on

the sequence of pulses that were used to prepare them.

The original arterial spin labelling approaches used continuous saturation or inver-

sion of the spins of a defined volume proximal to the imaged slice [18,83]. Continuous-

wave experiments of cerebral perfusion often use a separate excitation coi1 placed

around the neck of the subject for the inversion of inflowing spins. The inverted spins

flow through the brain and exchange with the tissue in the imaging regions. An image

thusly acquired, the tag image, will show some contribution fiom the inverted water

spins. The effect of inverted spins on the image can be isolated by subtracting fiom

the tag image a similar image that was acquired without the spin inversion, the control

Page 39: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

image.

Although a higher signal-to-noise is theoretically possible with continuous-inversion

methods [87], the technique suffers fiom problems associated with excessive energy

deposition and rnagnetization transfer (MT) artifacts when the separate inversion coi1

is not used. Magnetization transfer contrast comes about due to interactions between

excited and non-excited ' H nuclei [84]. The effect in continuous-wave inversion comes

from the off-resonance inversion pulse which doesn't excite water spins, but will excite

' H nuclei in macromolecular environrnents. Macromolecular spins have very broad

spectral linewidths and can be excited well away fiom water proton resonance - which

is very narrow [a]. The macromolecular spins that have been excited can then transfer

their magnetization to nearby water spins and change the image contrast.

h l s e d ASL techniques provide a means of controlling subtraction errors due to

magnetization effects, while providing reasonable - al though lower - SNR wi thout the

need of additional hardware to invert the inflowing spins.

3.2 Theory

A comprehensive model of the ASL signal in tissue is necessary to interpret images

in physiological terms. Viirious models specific to individual experimental techniques

have been proposed 159, 1 1, 83, 55, 54, 73, 18,961 , but recent work by Buxton et.al.

[ 101 has unified the theories into general and standard models of ASL signal for both

continuous and pulsed techniques.

The general model consists of a signal function that depends on three separate and

independent func tions representing assumptions made about the introduction, washout

and decay of inverted spins in the image slice. The standard model is the solution to

Page 40: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

the general model with a specific set of ideal assumptions about the three independent

functions. Other solutions to the general model based on more physiologically real-

istic assumptions can be obtained, but the mathematics involved quickly become very

complicated and numencal methods ofien need to be employed to solve the equations.

The general model is based on considerations of the relative concentrations of

tagged and untagged spins in the tissue; compounded with the relaxation of the tagged

spins with the time constants of their surroundings: first with the Tl of blood - Tib - and

then with TI of tissue (after the spins have been extracted). The model represents the

magnetization signal in the voxels of the difference image and has the form:

where c ( t l ) represents the normalized concentration of magnetization amving at the

voxel at time t' (delivery function), r ( t - t l ) , the residne fincrion, gives the fraction of

the tagged spins that arrived at time t' and remain in the voxel at time t - accounting

for the effects of spin washout, and m ( t - t'), the magnetization relaxationfuncrion,

represents the fraction of the magnetization of the spins that arrived at t' and remains at

t -accounting for the reduction of magnetization by the decay process.

From the general model, the direc! relationship between pixel intensity values and

the amount of flow f in the tissue is shown. A set of ideal assumptions about the nature

of c( t ) , r ( t ) and m(t) are made to provide a specific solution to Equation 3.1 which is

applicable to most ASL applications.

Page 41: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

The idealized representations of the component parts of Equation 3.1 that comprise

the standard ASL signal model are:

1. The delivery fiinction used in the standard model cornes about fiom the assump-

tion that the tagged spins arrive in the imaging slice via uniform plug flow. The

plug is created at a distance - typically about a centimeter - fiom the imaging

slice, and therefore a short. but unknown time period bt will pass before any of

the tagged spins reach the tissue of interest. The plug is created with a defined

width and will hence flow through the slice for a defined period of time Ï. The

short, well-defined duration of the plug fiow model allows the delivery function

to be represented as:

during the time period that the plug is passing through the image slice, and c ( t ) =

O before and afler.

2. The residue function is based on assumptions about the nature of water ex-

change between blood and tissue in the image voxels. Single-compartment ki-

netics are assumed, and this implies that instantaneous exchange between sub-

compartrnents that may exist within the voxel occurs, and therefore the tissue

concentration of inverted spins is related to the vascular concentration by a con-

stant ratio equal to the tissuehlood partition coefficient of water, A. The function

Page 42: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

has the form:

r ( t - t') = exp[- f ( t - t ' ) /A ] .

3. The rnagnetization relaxation function assumes that the transition fiorn the vas-

cular to the tissue environment for the water spins occurs instantaneously. The

water entering the image voxel does so at a time t' and the magnetic tag decays

fiom that point forward with the relaxation time of tissue T l . This results in a

fiinction of the fonn:

The expression for the signal strength in the difference image is then given by the

solution to Equation 3.1, with the assumptions given by Equations 3.2,3.3 and 3.4:

where the expression q, is dimensionless and represents the processes of relaxation and

clearance that are occumng in each time domain; it has the fonn:

40

Page 43: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

where:

and

The standard model solution suggests that the received difference image signal de-

pends strongly on the parameters At and r. This is a probiem from an experimental

point of view as these values are typically not known beforehand (as they depend di-

rectly on the rate of blood flow), may change with activation during a functional ex-

periment, and require that measurements be made at several different inversion times

(TI) before they can be estimated. Relative quantitative perfusion measurements - as-

sessing the fractional change of pixel values between activated and baseline States - are

less sensitive overall, but changes in the transit delay At can affect these calculations

as well. Quantitation of perfusion using the standard model requires that the technique

used should be done in conjunction with measurements of At and T l , or be modified to

minimize the effect of these parameters on the acquisition.

Page 44: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

3.3 Pulsed ASL Sequences

Several single-shot image acquisition sequences are applicable to pulsed spin tagging

approaches. Signal to noise ratio, image artifacts, resolution, imaging time, and energy

deposition in tissue Vary between these sequences, and represent factors which must be

weighed when selecting an optimal sequence for a particular application.

3.3.1 Fast Acquisition Techniques

Perfusion measurements with pulsed ASL depend on the rapid measurement of the

signal from prepared spins in the slice or volume of interest. The dynamic nature of

these signals requires that the image be measured quickly.

The fastest MRI acquisition techniques are based on the acquisition of a complete

set of k-space data with a single excitation. Such techniques, called single-shot ac-

quisitions, require high-performance gradient systems to allow the necessary k-space

coverage in the short period of time that the signal exists following an excitation.

Problems exist in single-shot techniques due to several effects: chemical shift; the

phase errors that accrue due to error in the signal sampling; the signal difference be-

tween readout lines acquired from different sections of the signal decay curve; and

several others. These need to be corrected either through reconstruction algorithms or

modifications to the acquisition technique.

The conceptually simplest single-shot technique, echo-planar imaging (EPI), was

the first to be developed [63]. EPI is essentially a single-shot version of a standard

spin-warp acquisition as s h o w in Fig. 2.6 (previous chapter).

Post-excitation, the signal phase is manipulated by applied gradient waveforms so

that, in the k-space representation, it sits at a corner of the space to be filled. Raster lines

Page 45: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

of k-space data are then acquired as an oscillating hi&-intensity gradient is applied in

the readout direction. The received signal is moved in the phase encode direction by

the application of a continuous, low-intensity gradient field or the application of higher-

intensity, short-duration gradient blips between the readout lines (Fig. 3.1 ).

Figure 3.1 : (top) Gradient echo (GE) echo-planar imaging (EPI) sequence schematic: n i e sequence features a 90" slice-selective excitation pulse (applied in the presence of the slice-select gradient Gs) followed by an oscillating gradient in the readout direc- tion (GR) and short, phase-encoding gradient blips (Gp) applied between readout line acquisitions. The gradient echoes formed by the oscillating readout gradient are shown on the RF line. (bonom) The k-space trajectory of a blipped echo-planar acquisition.

The SNR of an EPI acquisition can be improved, and the Bo inhomogeneities re-

duced, by inserting a 180" pulse between the excitation pulse and readout window such

Page 46: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

that the readout window is centred on the resulting spin echo peak. The time window

over which strong signal can be detected is extended, and the signal strength at the cen-

tre of k-space - which determines the global intensity of the reconstructed image - is

significantly increased. Such a sequence is called spin-echo EPI (SE-EPI) [68].

By inserting 180" refocusing pulses between each readout iine and acquinng each

line at the centre of the resultant spin echo, the signal strength at each line of data

c m be increased although the time between hnes increases. Such a sequence is titled

RARE for t-apid acquisition with relaxation enhancetnent or FSE for fasr spin echo

[30]. RARE sequences ofien have a fürther modification in that the order in which

the readout lines of k-space is changed so that the centre lines of k-space are acquired

before the outlying ones. This modification, temed centric phase encoding, improves

SNR and decreases the blur in the reconstmcted image due to T2 decay between readout

Iines, resulting in less T2 weighting in the reconstructed image.

Because the lines of k-space in single-shot techniques have to be acquired from a

single decay curve, the signal strength o f lines read early in the acquisition will typically

be higher (excepting SE-EPI). Consecutive lines of k-space acquired with a gradient-

echo EPI technique will show signal differences based on the T; time constant; consec-

utively acquired RARE lines are different by a factor based on T2. The simple, linear

manner in which k-space lines are acquired in EPI results in a blur artifact in the recon-

structed image. This blur c m be calculated by considering the T; exponential decay

fbnction as it applies across the k-space trajectory. In addition, the central k-space line

is acquired after half of the acquisition window has passed. This results in a reduction

in SNR for the reconstructed image. The reordering of a RARE sequence places the

first - and strongest in signal - echo at the centre of k-space, and consecutive echoes

near to the origin on either side. The resultant image is of higher SNR than one ac-

Page 47: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 3.2: (top) GRASE imaging sequence schematic; The pulse sequence consists of slice-selective excitation (applied in the presence of a slice-select gradient Gs) fotlowed by a senes of 180" inversion pulses that f o m an echo train. Through the formation of each echo, oscillating readout gradients (GR) are applied that form three gradient echoes per spin echo (five and seven GREISE are also commonly used). Between each echo, phase encode pulses (along Gp) are applied as short blips, and the spin phase is re-set to zero between inversion pulses. (bottom) The centre portion of a k-banded GRASE k-space trajectory. The numbers at the left hand side outline the sequence in which the lines were acquired (e.g. the third line acquired is numbered as 3). The numbers shown outline the centre of k-space in the GRASE sequences developed here, which used 5GREISE.

quired via EPI, and suffers less fiom the effects of T2 decay as consecutive lines differ

by a smaller factor. Problems with RARE centre around the longer imaging time re-

quired and the increased energy deposition (due to the large number of RF pulses) in

the imaged tissues.

The Gradient And Spin Echo or GRASE sequence is a hybnd of the EPI and RARE

Page 48: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Phase Encoding Lim

Figure 3.3: Schematic of the signal strength and phase modulation in k-banded GRASE imaging [53]. The three bands (centre, left and right) are acquired sequentially. With- in each band, the phase-encoding pulses are arranged such that readout lines obtained at the same position relative to the spin echo are piaced together. The pattern of the readout lines is chosen to minimize coherent patterns in the signal strength modula- tions across the phase-encodes in k-space. Such coherent patterns will result in image artifacts after reconstruction. The phase Iine in the figure shows the position relative to the spin echo at which the readout Iine was sampled. A sequence with 75 readout lines is shown.

techniques [71]. In this rnethod a series of 180" pulses is used to generate an RF-

refocused echo train, and several gradient echoes are acquired over eac h spin echo. This

results in a net decrease in the acquisition time when compared to RARE imaging be-

cause the number of time-consuming 180" pulses is significantly reduced. The GRASE

acquisition typically takes longer to acquire than EPI measurements, but, due to the

refocusing pulses, the signal strength rernains high enough to measure for a longer pe-

riod of time. A GRASE sequence with only one 180" pulse is a (reordered) SE-EPI

sequence, and one with only one gradient-recalled echo per spin echo is a RARE ac-

quisition. The GRASE and RARE acquisitions' refocusing of the signal also results

Page 49: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

in the calculated image being less prone to the artifacts created by static field inhomo-

geneities due, for example, to magnetic susceptibility di fferences in the imaged object.

interfaces between air and tissue or tissue and bone do not have uniform magnetic field-

s. This is due to the change in magnetic susceptibility at the interface of the two. The

change in susceptibility tends to dismpt the magnetic field locally and can drastically

change the value of T; in this region. The refocusing pulses in the echo train result in

the tissue signal decaying with T2 instead of T;, and the resulting image will be free

from these artifacts.

The ordering of k-space readout lines in GRASE acquisitions can be done in a

number of ways including linear, 'standard', partially randomized and k-banded (kb)

GRASE [53]. The short total acquisition time of funciional brain images makes the kb-

GRASE ordering scheme the logical choice, as this method produces the least amount

of artifacts for large field of view images in a short penod of time [53].

The speed of an acquisition is greatly increased for sequences that require only a

portion of k-space for image reconstruction. Full images c m be reconstructed from ac-

quisitions where as linle as half of the required k-space data is measured. The missing

data fiom a partial acquisition can be accounted for by changing the relative weighting

the existing data prior to reconstruction. This is possible because the magnetization

distribution of the object being imaged may be considered to be pure real, and this im-

plies that the k-space representation of the object must be Hermitian. A Hennitian data

set is symmetncal about a diagonal line through the origin of k-space. For incomplete

acquisitions, the rnissing data points c m be extrapolated fiom the existing data by a

point-by-point calculation of complex conjugates about the diagonal.

Problems with the reconstruction of partial-Fowier acquisitions corne about fiom

the presence of low-fiequency-phase modulations in the magnetization distribution that

Page 50: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

cause artifacts in the resultant image - that is, the magnetization distribution being re-

constructed is not purely real [16, 28, 521. There exist several correction methods for

removing this phase error, including the use of homodyne detection [70]. In the homo-

dyne technique, k-space is filled such that the high fiequency data beyond a few lines

on one side of the origin are left blank, and the remaining k-space is read normally

(approximately 65% of the fi111 k-space coverage was acquired in the partial-Fourier

sequences developed here). Then two reconstructions are done: one on the fùll data set

with weighting of the high frequency terms to account for the missing data, and one

which reconstmcts only the low-frequency terms that exist on either side of the origin

in the partial acquisition. The low-fiequency reconstruction results in a reconstmcted

image that is of a very low spatiaI resolution, but contains the tow-frequency phase er-

rors that reduce the resolution of the image formed by the reconstruction of the entire

partial data set. By subtracting the phase of the low-resolution image from the phase

of the entire-set reconstruction, and then taking the real part of the resultant complex

matrix, an image is generated that is fiee fiom the phase error caused by the reduced

k-space acquisition.

33.2 Spin Tagging Approaches

Two spin tagging approaches, STAR [21] and FAIR [59, 541, are currently the most

widely-used pulsed ASL techniques. STAR is the older of the two, and is the most like

continous-wave ASL which preceded it. A third imaging technique, PICORE [89, 861,

is the most recent addition to the ASL family. This sequence is very similar to the STAR

technique, and differs only in the way in which the effects of magnetization transfer are

controlled.

STAR - s ipal fargeting using altemafing radiofiequency, is a spin labelling ap-

Page 51: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

proach used to produce perfusion-weighted images [2 11. This technique most-often

uses an EPI sequence to acquire the images. and the term EPI-STAR is thus used to

reference such acquisitions. The technique involves the acquisition of two separately

prepared images of the same slices in the brain. One of the images is preceded by a

slab-selective inversion pulse applied proximally to the region of interest. Image acqui-

sition follows afier a time penod has elapsed to allow inflowing spins to perfuse tissue

in the region of interest. The images thus acquired are now flow-sensitive and are re-

ferred to as tag images. The second image set is acquired in a similar fashion with the

difference being that the slab-selective inversion pulse is applied distally to the region

of interest. These images should show no flow-sensitivity, and are referred to as being

the conîrol image. A subtraction of the control fkom the tag images yields difference

images that are weighted such that pixel values are directly proportional to flow. STAR

is shown schematically in Fig. 3.4.

lnstead of offset slab-selective inversion pulses, the FAIR - Jlorv-sensitive alrer-nat-

hg inversion recover y - technique employs slab-selec tive and non-selec tive invers ion

pulses centred on the slice(s) of interest in the generation of the tag and control images.

In FAIR, the control image is acquired following a non-selective inversion of the

spins surrounding the image slice, and the tag is obtained by inverting only a small

volume of spins immediately surrounding the image slice. Flow-sensitivity only exists

in the selective-inversion images as inflowing spins have a different magnetization his-

tory than those within the image slice - providing image contrast. The non-selective

inversion is insensitive to the effects of flow as there is almost no difference in the

magnetization histories of infiowing and image slice spins [55, 591. The magnetization

transfer effects are equal in both images, and hence should not appear in the difference

image. FAR is s b schematically in Fig. 3.5.

Page 52: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 3.4: (top) EPI-STAR tag image formation. The tag image is forrned by inverting the spins in the tag region with a slab-selective inversion pulse. The inverted spins then flow through the region of interest and perfùse the tissue there, where they provide contrast to the acquired images at Tl . (bottom) EPI-STAR control image formation. The control image is formed with the slab-selective spin inversion pulse applied distally to the region of interest so no tagged spins flow through the image slices pnor to imaging.

Inc luded for completeness, the PICORE sequence -proximal inversion with a con-

trol for ofiresonance effects - is identical to the STAR technique, with the exception

that the control image is acquired with the distal slab-selective inversion pulse replaced

with a repetition of the proximal inversion pulse, but now in the absence of the selection

gradient. The off-resonançe pulse will then invert no water spins, but the macromolec-

ular spins will be excited identically in both images (resulting in no magnetization

transfer contrast in the difference image). The PICORE modification has the advantage

that only spins proximal to the image slice are inverted, and therefore blood flow into

Page 53: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 3.5: (top) FAIR selective inversion pulse (tag image formation). The inversion band is centred around the region of interest and only a small volume of spins is invert- ed. The inflowing spins have not been inverted and hence behave as the tag spins in EPI-STAR. (bottom) FAIR non-selective inversion pulse (control image): here the spins are non- selectively inverted. Tagged blood flowing into the region of interest wiIl now be in- verted and act as the spins in the control image in EPI-STAR.

the slice fiom areas distal to the image slice contributes nothing to the difference image.

The dependence of the difference image signal strength on the parameters At and

r (Equation 3.5) imply that, for absolute or relative quantitative measurements of brain

perfusion, these parameters need to be measured or ehminated fiom the signal equation

by experimental design.

The development of simple pulse sequences that allow quantitative, perfusion-weighted

difference images to be generated with just two image acquisitions was performed by

Wong et a1.1871. Two modifications to the basic pulsed ASL methods were created and

Page 54: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

titled QUIPSS, for quantifutive imaging of perfusion using a single subtrac fion, and

QUIPSS 11. These modifications are applicable to al1 of the pulsed ASL techniques,

and allow quantitative results to be directly obtained by eliminating the unknown pa-

rameters fiom the signal equation.

The original modification - QUIPSS - eliminates the dependence of the signal on

the transit delay At by saturating the image slice after a time delay TI1 chosen such

that:

T I L > At.

The image is then acquired at TJ2, where:

The tagged blood then enters the slice for a time period AT1 = T l 2 - TlI, and the

standard mode! solution of the signal equation for the difference image becomes:

Under these exact circumstances, the signal equation has lost al1 of its dependence

on the transit delay At. By accurately estimating the physiological constants in the

equation, a quantitative difference image can be produced. The importance of this

modification to perfusion-based functional imaging is that any variability of the transit

delay due to activation-induced changes in blood flow is eliminated. This enables the

direct cornparison of difference image pixel values to obtain quantitative relative blood

flow changes between activation and baseline conditions without further measurement.

Page 55: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

The QUIPSS method has the limitation that only a single slice at a time can be

measured. The saturation of the imaging slice after inverted spins have begun perfusing

the imaged tissue means that slices distal to the first slice to be perfised will experience

variable amounts of transit delay that will affect the quantitative nature of the di fference

image in a manner similar to the transit delay existing between the inversion band and

the imaging of the slices.

A second acquisition sequence was developed that provides multi-slice quantitative

difference images based on simple subtraction of tag and control images. The second

modification, called QUIPSS II, is similar to its precursor, but applies the saturation

pulse differently. In QUIPSS II, the saturation pulse is applied to the region of the

inversion band at time T I I . If TII is shorter than the time width of the inversion band

r, the application of the saturation pulse truncates the inversion band so that it wi Il be

of well-defined time width T I I . So if:

Ti, < r

and

then the tagged bolus of blood will have passed entirely through the region of interest

before the slices are imaged at TI2. This results in the signal equation for the difference

image being given by:

Page 56: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Again, by choosing the saturation T I 1 and imaging TI:, times correctly, the subtncted

images will be independent of the effects of transit delay and hence will reflect quanti-

tative perfusion values.

3.3.3 Considerations for Multi-Slice Acquisitions

The extension of pulsed ASL techniques fiom single to rnulti-slice acquisitions intro-

duces potential sources of error into the difference images.

Long imaging times are a cause of some concem in multi-slice perfusion imaging.

The distal-to-proximal excitation order seems most logical as the spins saturated dunng

the image acquisition of each slice would not flow into the next slice to be imaged and

affect its quality. The problem with this excitation order is that the time over which

perfùsed spins have had to disperse within the image slice varies greatly between the

most distal and proximal slices. Theoretical calculations have shown that a proximal-

to-distal excitation order helps to alleviate the problem of differential perfusion times

and does not degrade slice quality as long as the individual image times are sufficiently

short [87]. For the flow values seen in the brain, the images will not be degraded if the

image times are kept below approximately 80ms, as this implies that the progress of the

slice excitations is faster than the fastest of the flowing spins in the brain.

Careful attention must also be paid to the profiles of the slab-selective inversion

bands in a multi-slice sequence. The hyperbolic secant pulses [S 11 often used to invert

the spins in pulsed ASL show well-defined inversion margins in inhomogeneous Bo

environrnents or flowing spins, but show some rounding of the edges of the inversion

slabs. This rounding effect can alter the slice intensities depending on their proximity to

the inversion band and these modulations will differ between the tag and control images

depending on the particular tagging approach used. Expenmental approaches using

Page 57: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

shorter (less than 20ms), lower intensity inversion pulses improve the slice profiles -

by reducing the effects of relaxation, and a saturation pulse applied to the image area

prior to the inversion pulse to nul1 the static tissue signal help co alleviate this problem

CW- Limitations exist on the spatial extent that can be irnaged during a single pulsed

ASL measurement. Using EPI-STAR as an example, the inverted tag spins that flow

through the imaging area decay with Tl tirne constants - Tib and then Tl of tissue afier

they are extracted fiom the vascular space [IO] - fiom the moment of their inversion

onwards. This limited time window for the signal sets limits on the spatial extent of the

measurernent in that a significant time penod must pass while the spins are in transit

before they can begin perfusing distal tissue slices. The proximal slices in a perfusion

measurement have a longer period of time for perfusion of extracted spins than do

distal slices. Rapid imaging sequences help to alleviate some of these problems in that

the maximum amount of time post-inversion can be allotted for the perfusion of distal

slices, but differences in the signals seen between slices due to the different amval

times of the inverted spins can affect perfusion measurement.

Page 58: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Chapter 4

Methods

fMRI can be performed on subjects using a wide variety of stimuli and tasks. The mag-

nitude and location of the response seen in a fMRI experiment can Vary significantly

depending upon stimulus parameters or task details.

For al1 of the experiments performed here we used a robust visual stimulus. Visual

stimuli are well-suited to fMRI methodology development due to the relatively large

signal response seen and the welI known location and fbnctional organization of the

visual areas [75].

The stimulus used here consisted of a yellow/blue radial checkerboard pattern with

30% luminance contrast (luminance variation divided by average luminance) that alter-

nated at a fiequency of 8Hz and covered 27" of the visual field of view. The pattern

was composed of 30 spokes and 4.5 rings. This was projected onto a screen at one

end of the scanner's bore, and was visible to the subject through the use of an angled

mirror added to the head imrnobilization setup. A schematic of the stimulus presenta-

tion equipment configuration is shown in Fig. 4.1. Visual stimuli were presented with a

NECMT820LCD projector operating in 640x480 mode at 60Hz, and generated using

Page 59: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

OpenGL-based software on a Silicon Graphics O2 cornputer [32].

As illustrated in Fig. 4.2, a small triangle was projected at the centre of the stimulus

pattern. This tnangle was also presented during the control (baseline) presentation -

consisting of a unifonn grey background at the mean stimulus luminance. To ensure

that the subject's attention was focused on the centre of the presented pattern, this trian-

gle was used to prompt feedback fiom the subject between acquisitions. The subjects

were asked to report the orientation (pointing right or lefi) of this triangle by pressing

the relevant bunon on a MRI-compatible mouse.

The visual area of the human brain is highly-organized and exists at a well defined

location along the calcarine fissure in the back of the occipital cortex. Signals generated

in the retina are processed in the lateral geniculate nucleus and then passed along nerve

fibres terrninating in the primary visual area (V1 ) or srriate corfer. The organization of

the primary visual area is such that a direct correspondence (contralateral mirror image)

exists between locations in the visual field and points in VI. Other visual areas (V2 and

up) exist lateral to V1 and perform higher-order processing of visual information [75].

The selective activation of various regions of the visual area can be achieved by the

application of specific visual stimuli. Elucidation of the relationship between stimulus

location in the visual field and the location of the evoked response is titled retinotopic

mapping and has been the focus of several experiments in PET, autoradiographic studies

and fMR1 [79, 751. The stimulus used here to test the perfusion-weighted imaging

techniques was chosen because of the large and consistant response that it evokes.

Page 60: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

NEC UT

scanner hardware

* visual field - 220

Figure 4.1 : Visual stimulus delivery schematic. The stimulus of Fig. 4.2 was delivered to the subject lying in the scanner bore. The stimulus pattern was generated by a SGI O2 computer and displayed on a screen through the use of an LCD projector. The prompt to change the stimulus during an acquisition came fiom trigger signals from the scanner hardware.

Figure 4.2: The radial checkerboard visual stimulus pattern. Experiments were per- fonned with a yellowiblue colour scheme altemating at 8Hz. Control images were acquired with a uniforni, grey screen.

Page 61: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

4.1 Sequence Design

The sequences used were generated using PARGEN 5.0 (Parameter Generator) for the

Siemens Magnetom Vision l.5T MRI system using the NUMARIS /3 V3B3 1 A oper-

ating system (SIEMENS Medical Systems, Erlangen, GDR.). The GRASE sequences

and the echo-planar sequence were designed to have the same readout bandwidth and

sequence timings.

4.1.1 Low-Resolution Acquisitions

Previous work at our lab [15] showed that a 128 x 128 matrix GRASE sequence provid-

ed good results in the acquisition of single-slice perfusion-weighted imaging using the

FAIR technique. Images from this sequence had a rectangular field of view in which

75 phase encode lines were acquired in order'to reduce the time required per image.

The sequence was still relatively long (160ms), and we hypothesized that a reduction

in acquisition time would facilitate the extension of the technique to multi-slice.

To this end, a lower-resolution 64x64 matnx GRASE sequence was implemented

that featured the acquisition of 45 phase-encode lines in a rectangular field of view in-

stead of the original 75. The reduction of the matrix size also meant that the acquisition

time for each readout line could be reduced as well. The end result was the reduction

of the total imaging tirne for each acquisition to Zms.

A M e r time savings was gained through the use of a half-Fourier sequence. In

this acquisition, entire brain images were reconstructed fiom raw data sets that featured

incomplete (two-thirds) coverage of k-space . A homodyne detection-based reconstruc-

tion technique was implemented to reduce the phase errors associated with half-Fourier

acquisitions [70]. This implementation reduced the tirne per acquisition to 5Oms with

Page 62: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

little reduction in image quaiity.

The GRASE sequence developed here acquired five readout lines (five gradient-

recalled echoes) for each 180" refocusing pulse. The k-space data was separated into

three regions that were filled according to the k-banding technique proposed by Fein-

berg et.al. as discussed in the previous chapter [23]. The initial five lines of data were

recorded without phase encoding and were used to correct phase error in subsequently

acquired lines. This was done by a point-by-point phase correction algorithm wherein

the phase of these original reference lines was subtracted from the corresponding data

lines.

The half-Fourier (HF) GRASE sequence reconstntcted images from incomplete k-

space data sets (two thirds of the data required by full-Fourier GRASE). The k-space

lines are organized into two bands [23] which are reversed in order with respect to one

another. Selective weighting - doubling of the intensities of the high-fiequency k-space

data, and a homodyne detection technique - to reduce the effects of low-frequency

modulations of the incomplete data set - were used to construct the full images from

the reduced data sets [69].

4.2 Spin-Tagging Implementation

Acquisitions featuring odd numbers of image slices were used in the generation of al1

perfusion-weighted images. The inversion pulses used were positioned at the central

slice location, resulting in a symmetrical positioning of the image slices within the

inversion slabs. This was done as a convenience as the slab centre is positioned along

an image slice.

Page 63: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

4.2.1 FAIR

The FAIR technique was implemented using a hyperbolic secant inversion pulse to

invert blood water spins prior to image acquisition [8 11. An inversion slab wide enough

to encompass the image slices plus an extra margin of 25% on each side, to minimize

profile effects, was used in the generation of the selective inversion images. The sarne

inversion pulse applied in the absence of a selection gradient was employed to yield the

non-selective inversion images. The inversion pulse profile was measured by acquiting

a magnitude image in a gel phantom in a plane perpendicular to the orientation of the

inversion pulse at an inversion time of 30ms. A plot of pixel values from a vector

passing through the inversion region is shown in Fig. 4.3.

Figure 4.3: Inversion profile for hyperbolic secant inversion pulse of 10.24ms duration.

4.2.2 STAR

The STAR implementation involved the use of regional inversion bands to generate the

tag and control images needed. These bands were positioned to be parallel to the image

slice(s) with a lem separation between the closer edge of the band to the most proximal

or distal slice. The positioning of the bands was done at the scanner's interface terminal

61

Page 64: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

in a marner similar to that of the slice positioning. Hyperbolic secant pulses were also

used to invert the spins in this case, however PARGEN lirnited the pulse duration to

approximately 3ms. A puise profile was generated for the shortened pulse and the

results are plotted in Fig. 4.4.

?au, 1

mD ' 1 P 1 9 ~ 1 1 P n Y 4 4 n

Am I*AI

Figure 4.4: Inversion profile for hyperbolic secant inversion pulse of 2.56ms duration.

The inversion profile of the shortened pulse is not noticably different than that of

the longer version.

4.2.3 Post-Processing of Image Data

Images generated on the MRI scanner were transferred to the MRI Research Lab com-

puters via ethemet. Activation maps, time-intensity curves (TIC'S) and relative per-

fusion change values were calculated fiom the data sets using routines witten in the

PERL scripting language and MATLAB ( V5.2 The Math Works Inc., Natick MA).

PERL scripts were written to convert the native data format of the scanner to MiNC

- a file format created at the Montreal Neurological Institute to facilitate easy post-

processing of images [66]. Once in MINC format, the individual image files for each

functional run were concatenated to foxm selective and non-selective inversion image

data files - or tag and control sets in the case of STAR imaging sequences. From these a

Page 65: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

difference image set was created. Quantitative processing was then perfomed on these

files.

4.3 The Perfusion Measurement

The data sets generated in each FAIR-based functional snidy consisted of a selective

image set acquisition followed by the imaging of the same slice(s) following the non-

selective inversion pulse. The STAR sequences featured proximal followed by distal

inversions. The repetition time (TR) of the slice measurements was set to three seconds.

This choice of TR allows for fresh (steady-state) spins to flow into the inversion slab

area following the acquisition of images. TR's as low as 2.0 seconds can be used in

perfusion weighted studies as spins flowing into the inversion slab areas can be assumed

to have no prior spin history [87]. Inversion times are typically on the order of 1.0

second (time between inversion and the excitation of the first slice to be imaged). This

time period allows inverted spins to perfUse tissue in the ROI and exit the macrovascu1ar

space, while maintaining acceptable signal strength (the effects of varying the inversion

time are studied in section 5.4). The functional experirnents were conducted for four or

six minute durations. The six minute duration experiments were used initially to test the

single-slice acquisitions. Based on the results of these studies, it was decided that four-

minute scan durations were sufficient, as the time savings allowed additional studies

to be perfonned while a single subject was in the scanner (each functional expenment

takes significantly longer than its nominal duration because of sequence load delays

between runs due to the processing and storage of images fiom the previous run).

The visuai stimulus presentation was synchronized with acquisition of the function-

al data with the baseline (uniform grey background) image being presented for the fvst

Page 66: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

minute. The activation pattern was then presented to the subject for the second minute.

This pattem was continued for the duration of the functional nin. An output trigger

pulse from the scanner hardware was used to synchronize the stimulus to the acquisi-

tion, and it was also used to change the direction of the triangular indicator at the centre

of the visual field in both activation and control images in a random fashion. The sub-

ject was asked to report the orientation of this indicator by pressing the correct button

on an MM-compatible mouse. A record of the subject's response was displayed in the

interface room to ensure that he/she was awake and paying anention to the presented

stimulus during the acquisition.

4.3.1 Subject Immobilization

Dwing the acquisition of the functional data, the subject was required to remain as still

as possible. The extremely rapid nature of the image acquisition ensures the 'freezing'

of the subject's motion during the measurement of a slice (Le. intrascan motion), but

movement between slice acquisitions (interscan motion) can greatly affect the results

of a functional expenment.

Subject immobilization during the study was facilitated by the use of an MM-

compatible headholder consisting of a foam forrn into which the subject's head was

fitted, a custom dental imprinted bite-bar that was placed into the subject's mouth, a

saddle-shaped nose-bridge piece that was pressed lightly onto the bridge of the nose,

and two ear cups that could be cinched over the the subject's ears - with the added

benefit to the subject of reducing the scanner noise dunng the acquisition.

Page 67: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

4.3.2 Functional Data Processing

The various image sets produced by the PERL routines could be analyzed in a number

of different ways in the generation of the study results.

MATLAB routines were developed for the calculation of t-maps [3]. These were

based on the pixel values of the difference image set referenced to a vector describing

the stimulus presented to the subject during the acquisition- Spearman Rank Correlation

was used as the statistical test in t-map generation [20]. The t-map of each slice was

then supenmposed on the anatomical data set to relate the function and anatomy.

Time-lntensity Curves (TIC'S) of averaged pixel values in a region of interest (ROI)

were plotted using a separate MATLAB routine. The ROI for these studies could

be chosen based on a t-map (wherein only pixels corresponding to significantly ac-

tive brain areas would be included by eliminating those pixels below a critical t-value

threshold), or by comparison with a previously-generated retinotopic map. Retinotopic

mapping was performed on some subjects in prior experiments performed at our lab

[45]. The mapping procedure consisted of using graded stimuli in a series of functional

experiments to generate activation maps (delineating V 1 ) that excluded contributions

from large vessels. The procedure used was developed earlier by Sereno et. al. at

UCSD [79]. The retinotopic maps were used as masks in determining the ROI'S used

in the functional experiments.

Changes behveen activated and control states are demonstrated graphically by plot-

ting the averaged pixel value in the ROI (either within a slice or a volume given by con-

sidering several slices at a time) throughout the time course of the selective-inversion

images. The same averaging was applied to the non-selective images, and the average

pixel value across the time course is shown on the TIC. The average pixel values in the

ROI for activation and baseline stimulus conditions are shown on the TIC as well, and

Page 68: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

the relative perfùsion value was calculated according to:

-4ctivation Sel. Image - BaseEine Sel. Image Relatiue Perfusion =

Baseline Sel. Image - ,Vonsel. Image (4.1)

for the FAIR images, and:

-4ctiuation Tag Image - B a d i n e T u g Image Relative Per f vs ion = -

,4ctiuation Co.ntro1 Image - Bnseline C'ontrol Image

(4.2)

for the STAR images.

The error was estimated by using the standard error of the means calculated for

the ROI pixel values in the selective image activation and control sets and in the non-

selective image data set. Such an error estimation is given by:

a Standard Error of the Mean = -

fi (4.3

where O is the standard deviation and n is the number of samples used in the calculation.

Experimental Design

Perfusion-weighted sequences were wrïtten and tested for STAR and FAIR spin-tagging

methods using both single and multi-slice acquisitions. Activation maps based on t-

values of Spearman Rank Correlation tests performed on the data sets were generated

along with time-intensity curves showing the relative perfusion values between activat-

ed and baseline stimulus presentation. The results of the single-slice studies provide a

Page 69: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

basis of cornpanson that enabled the selection of an imaging technique as being most

suitable for multi-slice applications. Once this was accomplished, irnaging parameters

were studied to find their effects on multi-slice sequences and predict a limit for the

possible spatial extent to be covered.

4.4.1 Single-Slice Experiments

Single-slice FAIR acquisitions used a 7.5 mm inversion band centred on the 5mm image

slice while STAR sequences used a 90 mm inversion band offset fiom the image slice

by a distance of 10 mm.

Typical relative pefision changes between activation and baseline States were found

to be on the order of 20 - 60%, as expected, and were found to Vary between subjects.

The results shown are fiom multiple subjects.

4.4.2 Multi-Slice Experiments

The GRASE FAIR sequence was chosen as the method that was most suitable for ex-

tension to a multi-slice acquisition. This decision was made because the GRASE-FAIR

sequence provided good functional data in preliminary studies, and had the advantage

of eliminating much of the image distortion seen in EPI.

The quality of perfusion data acquired as part of a multi-slice acquisition can suffer

from errors arising from slight changes in the acquisition parameters. Images in a multi-

slice sequence will have flow weighting based on a much wider inversion pulse, may

show effects fiom slice profile effects due to the proximity of adjacent slices, and may

also suffer due to slight differences in inversion times that exist between slices (as each

slice acquisition takes 80 ms).

Page 70: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

4.4-3 The Effect of Inversion Slab Width

The inversion band width's effect on perfusion data was studied using a series of

single-slice GRASE-FAIR measurements acquired at the centre of several inversion

slab widths. Single-slice acquisitions were chosen for this set of experiments as they

will not have confounding effects fiom nearby slice acquisitions. Inversion band widths

fiom 55.5m.m to 106.5mm were tested. Testing of this range of slab widths provides

an estimate of this parameter's effect on the results of functional experiments with slab

widths wide enough to incorporate up to seven 5mm slices with lmm separation - the

centre of an inversion slab of 106mm will correspond with the most distal slice of a

seven-slice acquisition.

4-4-4 Slice Profile Effects on Perfusion Data

To evaluate the effect that imaging nearby slices has on the perfusion data within a

slice, a seven slice sequence with contiguous 5 mm slices and an inversion band of

52.5 mm was used. Functional studies were performed using this sequence centred

on the visual area of the brain. Studies difEered in the distance separating the imaged

slices. Cornparisons of the perfusion values seen in the centre slice - slice 4 - were used

to assess the separation at which the perfusion values were the sarne as those seen in

single-slice acquisitions.

4-4.5 The effect of different inversion times

The inverted magnetization recovery during the acquisition of a single slice is negli-

gable, but becomes appreciable when multiple slices are acquired. The inversion time

difference between the imaging of separate slices is a possible source of error in a perfi-

Page 71: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

sion measurement. To find the range of possible inversion tirnes over which acceptable

perfusion data can be obtained, a series of functional data sets were acquired in which

the inversion time of the image set was varied between runs. The inversion time was

defined relative to the centre slice. The ROI used to generate the TIC included the entire

masked volume defined by a retinotopic map. The different slices contributing to this

result al1 diEer slightly in inversion time across the ROI.

4.4.6 A Seven-Slice Perfusion-Based Functional Experiment

The results of the tests performed in the previous sections suggested that a multi-slice

sequence would be successful in acquiring up to seven slices at a tirne. A seven-slice

sequence was assembled, with 5 mm slice width, I mm slice separation and an inversion

band width of 55.5 mm centred on slice four in the imaged area. The spatial coverage

of the imaging sequence was now greater than the size of the primary visual cortex.

To compare the response measured by the sequence across all of the slices, functional

studies were performed with the centre of the primary visual cortex - the Calcarine

fissure - lying along slices at different locations in the excitation order. A study with

one of the more proximal slices aligned with the calcarine will have flow weighting due

to spins that have flowed a short distance from the edge of the inversion band and will

have seen a shorter inversion time than slices in more distal locations. By comparing the

activations seen fkom the visual area as it is centred on the various slice locations while

keeping al1 of the other acquisition parameters constant, a measure of the uniformity

of perfusion sensitivity across slices was obtained. The TIC'S calculated here were

averaged across the entire ROI volume covered by the slices.

Page 72: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

4.5 QUIPPS

A single-slice QUIPPS GRASE FAIR sequence was developed. An experiment com-

paring the response seen with that fiom GRASE FAIR was performed to test the ef-

fect of the QUIPSS modification on the fiinctional results. The QUIPSS modification

(according to the Standard ASL Signal Model) eliminates the effects of the unknown

parameters r and At fiom the perfusion-weighted images.

This cornparison will yield an estimate of the accuracy of the non-QUIPPS perfu-

sion results obtained in Our studies.

4.6 Reproducibility of Results

To test the reliability of the methods used. An experiment was designed where finc-

tional data was acquired fiom a singie subject on two occasions a week apart. The

results fiom these tests were compared to ensure consistency.

Page 73: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Chapter 5

Results

Results of the various iÛnctional data experirnents outlined in the previous chapter are

presented here. The t-map and TIC of runs are presented. as well as the results of an

analysis of variance (ANOVA) test performed on the data sets. The ANOVA test was

used to test the probability that the difference in means of the functional results are due

to chance alone. In cases where a significant difference was found, Tukey's Honestly

Significant Difference (HSD) Test was used for post hoc-pairwise comparison of the

individual data sets [95].

5.1 Single-Slice Results

5.11 FAIR

The FAIR single-slice results are shown in Fig.5.1. The experiments were of six minute

duration, and used a retinotopic map to mask the ROI. The studies were performed

sequentially on a single subject.

The measured relative pefision change was measured as 30 5 3% for EPI-FAIR,

Page 74: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

23 dz 3% for GRASE FAIR and 24 k 3% for Partial-Fourier GRASE FAIR. The ANOVA

test was performed on the data sets to detennine if it was reasonable to assume that the

differences in the three relative pefision values seen was actually significant, and not

due to chance alone. The results of this statistical test indicate that the means of the data

sets do not differ significantly enough to mie out chance as the cause of the differences

seen (F=2.33, dW9, p>O.OS).

The localization of function seen in the three aqusition modalities is displayed in

the upper row of Fig. 5.1. The three modalities show activation in the same area of

the brain. The EPI-FAIR sequence results showed the largest area of activation, the

GRASE FAIR results showed activation in a smaller area located within the activation

area of the EPI-FAIR results, and the Partial-Fourier GRASE FAIR results were smaller

still, but were located at the same position within the brain as the GRASE FAIR results.

Figure 5.1 : EPI-FAIR (lefi), GRASE FAIR (centre) and Partial-Fourier GRASE FAIR (right) t-maps and TIC'S.

Page 75: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

5.1.2 STAR

The results of the single-slice STAR techniques are shown in Fig. 5.2. Each experiment

was of a six minute duration, and used retinotopic maps to mask the ROI.

The STAR results were significantly noisier than their single-slice FAIR counter-

parts - the standard deviations of the data sets were approxirnately 25% of the calcu-

lated value, as compared to 10% for the FAIR results. The ANOVA analysis found no

significant difference in the relative perfusion change values of the data sets outside of

that expected fiom chance alone (F=0.45, deS9, p>0.05).

The relative perfusion change values calculated fiom the STAR data sets were 24 * 4% for EPI-STAR, 24 -t5% for GFWSE STAR, and 28 3 ~ 6 % for Partial-Fourier GRASE

STAR.

Activation t-maps are displayed in the upper portion of Fig. 5.2. The t-maps show

activation in the same areas of the primary visual cortex with similarly-shaped areas of

activation.

5.2 The Effect of Inversion Band Width

The results of the study investigating the effects of inversion band width are s h o w in

Fig. 5.3. Al1 studies were performed sequentially on a single subject, and were four

minutes in duration.

The various inversion bands produced results of regular shape and consistant rela-

tive perfusion change values (404%).

The ANOVA test performed on the data sets showed there to be no significant dif-

ference in the calculated relative perfusion change values (F= 1 S9, df=79, p>0.05).

Page 76: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 5.2: EPI-STAR (left), GRASE STAR (centre) and Partial-Fourier GRASE STAR (right) t-maps and TIC'S.

5.3 The Effect of Süce Profile

The results of the expenments exploring the effect of varying the slice separation are

shown in Fig. 5.4. These four-minute studies were perfomed sequentially on a single

subject. A retinotopic rnap was used to mask the ROI.

The centre slice of seven-slice acquisitions at various slice separations was used

to determine the effects of imperfect slice profiles on the measured relative perfusion

change. Increased relative perfûsion values (relative to that measured in a single-slice

experiment) were seen at slice separations below Imm.

Page 77: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

q ml. - - . - - . - - - - . - - - . g 1 - A- ROI RXN Vau.

lm- kavabon Avg Sm I m g . . , - - - - - - - - - - - - - Casnn ~ u g 5r lrnig.

- Ncn-Sm b n ~ g e R O I A v g --

m. ; .9 .; A P 1 4 1 - - 6 - m u

4 a. 2 2 ,, --- - 4 / - ~ v g ROI mi Vatus

.a- Ccovamn Aug S.( lm- . - - I Nan-SH bnage ROI Avg

-0 ; O O n . > r-". r "

Figure 5.3: The TIC and relative perfusion change value measured in a single slice centred in inversion bands of various widths: 55.5mm (top left), ïl.5mm (top right), 90mm (bottom lefi) and 106.5mm (bottom right).

5.4 The Effect of Inversion Time

The results of measurements of the effects of differing inversion times are shown in Fig.

5.5. The four-minute experiments used here were perfomed sequentialty on a single

subject.

Seven slice acquisitions with one millimeter separation between slices were used

to test for the effects of inversion tirne on the measured relative perhision change. A

retinotopic map was used to mask the ROI in al1 slices. The response fiom al1 slices are

included in the results. inversion times stated are to the first image slice excitation.

Page 78: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 5.4: Calculated relative perfusion changes at various slice separations. The measured relative perfusion change measured in a single slice aqusition is included as the dotted line-

The earliest inversion time study (2OOms) shows the pixel averages for the setective

image to be lower than those of the non-selective inversion. The caiculated relative

perfusion change value does not differ in sign from those of longer inversion times, but

is significantly different in the appearance of the TIC. The calculated relative perfusion

change was 24 zt: 5% for this inversion time.

The remaining inversion time experiments were of a more regular form. The ANO-

VA test perfonned on the data sets showed that a significant difference existed be-

tween the relative perfusion change values of the data sets in the study (F=7.58 d e l 19

p<0.000 1). Tukey's HSD test showed that results fiom the study with an inversion time

of 1400ms were significantly different from every other data set.

The measured relative perfusion change has a maximum value at an inversion time

of 1000ms. The response seen fkom the 1200ms inversion time study is slightly lower,

but still within the error ranges of the values.

Page 79: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 5.5: Graph of relative perfusion change values measured with a seven slice GRASE FAIR sequence as the inversion time was varied between 600 and 1100ms.

5.5 Cornparison of Response Seen Across the ROI

The results of the measurements of the primary visual area response with different

positionings of the ROI are shown in Fig. 5.6. The three four-minute studies were

performed sequentially on a single subject. A retinotopic map was used to mask the

ROI. Only the results from the slice centred on the calcarine fissure are included. Al1

acquisitions were performed using an inversion time of 1000ms.

The relative perfusion values measured in the tùnctional runs were 30 f 6% for

the acquisition centred on the calcarine fissure; 30 & 3% for the acquisition with the

calcarine aligned along the second-most distal slice; and 33 & 4% for the acquisition

with the calcarine aligned along the second-most proximal slice. The centre slice was

acquired with an effective inversion time of 12lOms; the proximally positioned acqui-

sition had an inversion time of 1 OiOms, and the distally-positioned acquisition had an

effective inversion time of 1420ms. The TIC'S are of a regular form, with the centred

acquisition having the highest level of noise in the data set. The ANOVA test showed no

significant difference in the relative perfusion change values calculated from the studies

(F=0.29 df=59 p>0.05).

Page 80: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 5.6: Relative perfusion changes seen in one slice (centred on the calcarine fis- sure) of seven-slice GRASE-FAIR acquisitions with calcarine fissure aligned along the second-most proximal slice (lefi side data point), the centre slice (centre data point), and the second-most distal slice (right side data point).

5.6 QUIPPS

The effect of the QUIPSS modification was studied using the GRASE FAIR sequence,

the results are dispiayed in Fig. 5.7. The four-minute runs were performed sequentially

on a single subject. A retinotopic map was used to mask the ROI.

The two TIC'S are similar in appearance with a higher level of noise seen in the

GRASE FAlR acquisition. The ANOVA test perforrned showed no significant dif-

ference in the calculated relative perfusion values of the two studies (F=0.84 d e 3 9

p>0.05).

5.6.1 Reproducibility of Results

To test the reproducibility of pefision measurements, a single subject was tested at two

separate times a week apart. Each expenment consisted of five functional runs during

each of which a single slice aligned with the calcarine fissure was imaged in four minute

runs with the usuaI stimulus presentation. The results of these measurements are shown

Page 81: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Figure 5.7: Cornparison of TIC'S for GRASE FAIR single-slice acquisitions. The con- ventional results are displayed at lefi, and those fiom a sequence with the QUIPPS modification are shown at right.

Table 5.1 : Table of experiinental results testing the reproducibility of perfusion results. A single subject was tested with an identical set of experiments a week apart.

below.

A two-tailed, paired t-test was applied to the results to predict the likelihood that the

difference seen between the two means was due to something other than chance [74].

The result of the test indicated that the results between the two tests are not significantly

different t ( 4 ) = - 1.16, p > 0.05.

Page 82: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Chapter 6

Discussion

6.1 Experimental Results

The single-slice results for the FAIR and STAR perfusion-weighting approaches show

similar results for the calculated relative perfusion change value but differ in the degree

of noise seen in the image. The results fiom both approaches are largely independent

of the imaging method used, but the FAIR techniques show cleaner TIC data (better

Sm). This improvement in signal may result fiom the decreased distance that the

leading edge of the tag bolus has to travel to get to the imaging area (a few mm in

FAIR as opposed to slightly more than a cm in STAR), and may also be in part due to

the limited size of the inversion band used in the STAR technique (a 9 cm inversion

slab was used). The selection of GRASE-FAIR as the method best-suited to extension

to multiple slices was based on the quality of the results fiom the single-slice sfudy.

The results were essentially the same as those fkom the EPI-FAIR sequence, but the

refocusing of the signal in the GRASE acquisition results in less spatial distortion than

that seen in EPI sequences [7 11.

Page 83: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

The inversion slab width was shown to have little effect on the quality of the func-

tional data and the measured relative perfiision changes over the range of widths stud-

ied. Only the inversion widths that were relevant to the numbers of slices studied in

other investigations were explored. The limit of this slab width was not realized dur-

ing this investigation, but it must necessarily exist as extremely wide inversion bands

would be indistinguishable fiom the non-selective inversion pulse, and not be flow-

weighted at all. The results fiom the widest inversion pulse studied suggest that an

inversion pulse wide enough to incorporate seven slices would not significantly alter

the perfusion-weighting of the imaged slices, i.e. the centre slice from an inversion

band wide enough to include thirteen slices would roughly correspond to the most dis-

ta1 slice of 3 seven slice acquisition.

lmaging slice profile has a large effect on the calculated value of relative pef i s ion

change seen in the expenments. The changes seen in the contiguous slice and half-

millimeter separation experiments show relative changes that are approximately 25%

higher than that of a single-slice experiment perfomed on the same subject. Separa-

tions of one millimeter or larger show perfusion values are within experimental error

of the single-slice results. The inflated perfusion change values result from the dif-

ferent effects that imperfect slice profiles have on inflowing spins following selective

and non-selective inversion pulse. Signal from inflowing spins in selec tive-inversion

images have steady-state magnetization, but non-selective inversion spins are only par-

tially recovered at the time of image acquisition. Imperfect slice profiles result in slight

rotations of spins in nearby slices, and this will attenuate the signal from steady-state

inflowing spins to a larger degree than their non-selective inversion counterparts. The

relative perfusion change calculation relies on the change seen in activation and base-

line pixel averages in the selective inversion images. These will be attenuated in a

Page 84: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

similar fashion by the imperfect slice profiles of the adjoining slices, not greatly af-

fecting the difference. The pixel average in the non-selective images, however, will

show less attenuation fiom slice profile effects due to the saturation of the inflowing

spins. The overestimation in relative perfusion therefore likely results from the reduced

difference between the selective-inversion baseline pixel average and the non-selective

pixel average (a reduction in the value of the denorninator).

Inversion time was found to have little efTect on the calculated relative perfusion

change and activation t-maps

Relative perfusion change measurements were found to be robust to changes in in-

version time between 600ms and 1200ms. Studies using inversion times outside of this

range showed noisier TIC'S, poorer-quality t-maps and changes in the value of the rela-

tive perfbsion change calculation. These results suggest that inversion time differences

between slices acquired as part of a seven-slice sequence would have minimal effect on

the quality of fiinctional image data.

Within the inversion band of a seven-slice GRASE-FAIR acquisition a similar re-

sponse was seen after positioning the same responding structure, the primary visual

cortex, at proximal, central and distal locations. A noisier functional data set was ob-

tained fiom the central positioning experiment, but the calculated relative perfusion

change value seen was very similar to the two more-proximal studies. The centred

slice results having an increased level of noise is most likely due to imperfect position-

ing of slices along the calcarine fissure between experiments. Slight diflerences in the

alignment of the slice of interest along the calcarine may affect the signal seen in an

experiment .

The QUIPPS modification to the single-slice GRASE FAIR sequence showed no

significant difference in the resultant data sets. Similar TIC'S and calculated relative

Page 85: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

perfusion changes were seen. The increased level of noise seen in the non-QUIPSS

acquistion may be due to imperfect subtraction of the tissue signal between the selective

and non-selective inversion images. The signal from static tissue in the QUIPSS images

will be more efficiently nulled by the saturation pulsed applied to the image slices,

making this sequence less sensitive to this source of noise.

The QUIPSS modification is important to absolute quantitative measurements of

perfusion level. as this method eliminates the confounding factor of At fiom the signal

equation of the standard model. in FAIR measurements, the spatial gap between the

Ieading edge of the inversion slab and the imaged slice is typically a few mm, and Our

use of the body coi1 in the spin inversions results in a large bolus width (large T ) as al1

of the spins in the sensitive volume of the scanner would be inverted in the generation

of the non-selective image. The implications of this in reference to Buxton's Standard

model (outlined in Section 3.2 of this thesis) are that the second condition of equation

3.5 would apply for the signal function. The small spatial gap between the leading edge

of the inversion bolus and the first imaged slice would result in a small value of At in

equation 3.5. If we assume this At contribution to be negligable, and the value of T to

be large enough so that the second condition of equation 3.5 is valid for the inversion

time used, the relative changes in measured signal during an activation experiment will

be directly proportional to the relative flow (and hence perfusion) changes in the brain.

Questions remain about the validity of the above suggestions. The actual value of

At and T for a typical FAIR experiment have not been studied at Our lab in reference to

the standard model. The result of the single-slice QUIPPS cornparison performed here,

however, suggests that our assumptions are valid.

Page 86: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

6.2 The Seven-Slice GRASE-FAIR Sequence

The results of the various tests performed on sequence parameters suggest that a seven-

slice GRASE-FAIR sequence with lmm slice separation is effective in extending the

spatial coverage of pefision-weighted imaging sequences fiom single slice acquisi-

tions. An increase in spatial coverage fiom 5mm to 41mm was obtained (although the

increased coverage necessarily contains 6 lmm slice gaps that do not contribute to the

functional data).

6.3 The Interleaved Sequence

Currently, BOLD contrast-based fMRI is predominmt over perfusion-weighted imag-

ing techniques. This is due to limitations in SNR and the extent of the spatial coverage

possible in the perfusion methods. Questions exist, however. pertaining to the relation-

ship between BOLD and perfusion-based contrast sources, and the changes in cerebral

metabolic rate of oxygen consumption (CMR02) (assumed to be coupled to changes in

cerebral blood flow - CBF). Several models describing this relationship have been pro-

posed, and experiments performed at Our lab have tested this relationship and compared

it to the relationship proposed by the Deoxyhaemoglobin Dilution Mode1 [45].

The interleaved sequence that was the means of acquiring this functional data fea-

tured an echo-planar readout and acquired a BOLD-weighted image after the selective

inversion image acquisition of a FAIR sequence and another afier the non-selective im-

age acquisition. The use of an interleaved sequence (developed by the author) provided

the direct cornparison of BOLD and perfusion-weighted measurements dunng a sin-

gle functional run (the two data sets are temporally interleaved and have exact spatial

correspondence). The expenments done at Our lab were performed with graded visual

Page 87: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

stimuli, and with different levels of hypercapnia - which allowed the manipulation of

CBF independent of CMR03.

The results of these experiments supported the Deoxyhaemoglobin Dilution Model,

suggested that ABOLD signal (relative BOLD signal change) is a linear measure of tis-

sue workload in visuai activation studies, and related ACBF to 3CMRO2 in a constant

2 : 1 ratio. Furthemore, a PET-based determination of ACMRO? with a radial

checkerboard stimulus (which had been performed in a separate experiment) produced

results which were the same as those found at in out MN experiment (2.5% change).

This work formed the bulk of the PhD thesis of Rick Hoge, and was the source of three

papers [45] - [47] and numerous abstracts [34] - [43] and [48] - [SOI.

The extension of the interleaved perfusion-BOLD teclmique to multiple slices us-

ing an echo-planar sequence would allow the activation physiology experiments to be

extended to other brain structures. As well, increased spatial coverage in further visual

studies would allow the entire primary visual cortex to be studied in a single experi-

ment.

Page 88: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Chapter 7

Conclusions

The pursuit of a perfusion imaging technique that, at least partially, alleviates the prob-

lem of limited spatial extent has been the focus of this thesis. Candidate acquisition

sequences were designed and tested with FAIR and STAR spin-labelling modalities in

single-slice acquisitions. The GRASE FAIR sequence was chosen for extension to a

multiple-slice acquisition, and was found that up to seven slices could be effectively

imaged at a time.

The imaging of perfusion in multiple slices introduces potential problems. Vari-

ous possible sources of error in multi-slice acquisitions were addressed and tested to

determine their effect on perfusion results. The results of the study suggested that the

perfusion measures seen were robust to most of the parameters tested over the ranges

of interest.

An increase in the spatial coverage of perfusion-weighted sequences will allow the

acquisition of data fiom activation foci less easily identified than the primary visu-

al area, whose location along the calcarine fissure makes it particularly convenient to

study. Also, tùnctional structures in areas of high magnetic susceptibility change are not

Page 89: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

good candidates for BOLD imaging methods, as the TG-weighting of the images leads

to an unacceptable level of artifacts. Perfusion-weighted GRASE imaging is especially

well-suited to these tasks as the constant refocusing of the signal dunng acquisition

greatly reduces artifacts fiom this source.

Issues remain in regard to the quantitative nature of the blood flow signal measured

in MR-based perfusion imaging. The goal of much of the current research is to provide

absolute quantitative measurements of cerebral perfusion. The pursuit of this has gen-

erated much of the theory that was used in this thesis (the Standard Model. QUIPSS,

and QUIPSS II).

Questions remain about the validity of the perfusion change results reported here.

The assumption that relative blood flow changes are directly proportional to the perfision-

weighted signal changes seen in Our images is as yet unverified. Encouraging results

such as the constancy of calculated values between single and multi-slice acquisition-

s, the agreement that was seen between Our methods and PET for the same stimulus

between different subjects, and the constancy of results with and without the QUIPSS

modification have been reported here. A more exhaustive study comparing the multi-

slice sequence reported here with a QUIPSS II version would be useful in answering

the questions that remain.

Additionally, quantitative comparisons with PET studies performed on the same

subjects would be the most reliable assay of the effectiveness of ASL perfusion mea-

surements. PET results remain the most reliable assays of brain perfusion, and the

reproduction of these using ASL methods would be of great importance to fMRI.

Page 90: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Bibliography

[ 1 ] A. Abragam. Principles of nuclear rnugnetisrn. Oxford Science Publications, 1961.

[2] C. J. Aine. A conceptual overview and critique of functional neuroimaging tech- niques in humans: 1. MRVfMRI and PET. Critical Reviews in Netrrobiology, 9(2-3):229-309, 1995.

(31 P.A. Bandettini and A. Jesmanowicz. Processing strategies for time-course data sets in functional MRI of the human brain. Magnetic Resonance in Medicine, 30(2): 16 1-73, Aug. 1993.

[4] J. C. Baron, R. S. Frackowiak, K. Herholz, T. Jones, A. A. Lammertsma, B. Ma- zoyer, and K. Wienhard. Use of pet methods for measurernent of cerebral energy metabolism and hemodynamics in cerebrovascular disease. Jownal of Cere61-al Blood Flow and Metabolism, 9(6): 72-2, Dec 1989.

[SI J. W. Belliveau, D. N. Kennedy Jr., R. C. McKinstry, B. R. Buchbinder, R. M. Weisskoff, M. S. Cohen, J. M. Vevea, T. J. Brady, and B. R. Rosen. Functional mapping of the human visual cortex by magnetic resonance imaging. Science, 254(5032):7 16-9, Nov 1 199 1.

[6] J. W. Belliveau, B. R. Rosen, H. L. Kantor, R. R. Rzedzian, D. N. Kennedy, R. C. McKinstry, J. M. Vevea, M. S. Cohen, 1. L. Pykett, and T. J. Brady. Func- tional cerebral imaging by susceptibility-contrast NMR. Magnetic Resonance in Medicine, 14(3):538-46, Jun 1990.

[7] F. Bloch. Nuclear induction. Physical Reviav, 70:46W73, 1946.

[SI M. J. Bronskill and P. Sprawls, editors. n e physics of MRi 1992 surnmer school proceedings. American Association of Pysicists in Medicine, 1992.

[9] R. B. Buxton and L. R. Frank. A mode1 for the coupling between cerebral blood flow and oxygen metabolism during neural stimulation. Journal of Cerebral Blood Flow and Metabolism, 17(1):6472, Jan 1997-

Page 91: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

[ i O] R.B. Buxton, L.R. Frank, E.C. Wong, B. Siewert, S. Wanch, and R.R. Edelman. A general kinetic model for quantitative perfusion imaging with artenal spin la- belling. Magnetic Resonance in Medicine, 40:383-396, 1998.

[ I l ] F. Calamante, S. R. Williams, N. van Bruggen, K. K. Kwong, and R. Turner. A model for quantification of perfusion in pulsed labelling techniques Cpublished erratum appears in NMR biomed 1996 sep;9(6):277]. NMR in Biomedicine. 9(2):79-83, Apr 1996.

[12] H.Y. Cam and E.M. Purcell. Effects of diffusion on fkee precession in nuclear magnetic resonance experirnents. Ph-vsical Review, 94(3):63M3 8, 1 954.

[ 131 Z.H. Cho, J. P. Jones, and M. Singh. Foundations of Medical Imuging. John Wiley and Sons, hc., 1993.

[14] E. Clarke and K Dewhurst. An illusmted HrSrory of Bvain Funcrion. Norman Publishing, second edition, 1996.

[ 151 G.R. Crelier, R.D. Hoge, P. Munger, and G.B. Pike. Perfusion-based functional magnetic resonance imaging wi th single-shot RARE and GRASE acquisitions. Magnetic Resonance in Medicine, 4 1 ( 1 ): 1 32-1 36, January 1 999.

[16] J. Cuppen and Andre van Est. Reducing MRimaging time by one-sided recon- struction. In Topical Conference on Fast MM Techniques, pages 1 5-1 8, May 1987.

[17] J. A. Detre, D. C. Alsop, L. R. Vives, L. Maccotta, J. W. Teener, and E. C. Raps. Noninvasive MW evaluation of cerebral blood flow in cerebrovascular disease. Neurologv, 50(3):633-4 1, Mar 1998.

[18] J. A. Detre, J. S. Leigh, D. S. Williams, and A. P. Koretsky. Pefision imaging. Magnetic Resonance in Medicine, 23 ( 1 ) : 3 7 4 5 , Jan 1 992.

[19] J. A. Detre, W. Zhang, D. A. Roberts, A. C. Silva, D. S. Williams, D. J. Grandis, A. P. Koretsky, and J. S. Leigh. Tissue specific perfusion imaging using arteriat spin labeling. NMR in Biomedicine, 7( 1 -2):75-82, Mar 1994.

[20] J.L Devore. Probabiiiîy and Statistics for Engineering and the Sciences. Duxbury Press, fourth edition, 1995.

[2 l ] R. R. Edelman, B. Siewert, D. G. Darby, V. Thangaraj, A. C. Nobre, M. M. Mesu- larn, and S. Warach. Qualitative mapping of cerebral blood flow and functional localization with echo-planar mr imaging and signal targeting with alternating ra- dio fiequency. Radiology, 192(2):5 1 3-20, Aug 1994.

[22] R.M. Eisberg. Fundamentals of modem physics. John Wiley and Sons, Inc., 196 1.

Page 92: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

[23] D. A. Feinberg, B. Kiefer, and G. Johnson. Grase improves spatial resolution in single shot imaging. Magnetic Resonance in Medicine, 33(4):52%3 3, Apr 1995.

[24] L. R. Frank, E. C. Wong, and R. B. Buxton. Slice profile effects in adiabatic inversion: application to multislice perfusion imaging. Magnetic Resonaiice in Medicine, 38(4):55844, Oct 1997.

[25] R. Freeman. A handbook of nuclear magnetic resonance. Longman Scientific and Technical, 1988.

[26] K.J Friston, A.P. Holmes, K.J. Worsley, J.P. Poline, C.D. Frith, and R.S.J. Frack- owiak. Statistical parametric maps in functional imaging: a gened iinear ap- proach. Human Brain Mapping, 2: 189-2 10, 1995.

[27] J.S. George, C.J. Aine, J-C. Mosher, D.M. Schmidt, D.M. Ranken. H.A. Schlirt, C.C. Wood, J.D. Lewine, J.A. Sanders, and J.W. Belliveau. Mapping function in the human brain with magnetoencephalography, anatomical magnetic resonance imaging, and functional magnetic resonance imaging. Journal of Chical Newo- ph-vsiology, 12(5):40&3 1, Sep 1995.

[28] P. M. Glover, P. F. Tokarczuk, and R. W- Bowtell. A robust single-shot partial sampling scheme. Magnetic Resonance in Medicine, 34( 1 ): 74-9, Jul 1 995.

1291 D. F. Gochberg, R. P. K e ~ a n , and J. C. Gore. Quantitative studies of magne- tization transfer by selective excitation and t l recovery. Magnetic Resonance in Medicine, 38(2):224-3 1, Aug 1997.

1301 J. Hennig, A. Nauerth, and H. Friedburg. Rare imaging: a fast imaging method for clinical IN. Magnetic Resonance in Medicine, 3(6): 823-3 3, Dec 1 986.

[3 11 P. Herscovitch and M. E. Raichle. What is the correct value for the brain-blood partition coefficient for water? Journal of Cerebral Blood Flow and Metabolism. 5(1):6S9, Mar 1985.

[32] R. Hoge. Glstim: An openGL based stimulus presentation program for functional MN. http://www.bic.mni.mcgil l .ca~users/rhoge/GLstim/GLstim. html.

[33] R.D. Hoge. Enhancement of volume and temporal resolution for functional mag- netic resonance brain imaging. Master's thesis, McGill University, 1995.

1343 R.D. Hoge, J. Atkinson, B. Gi11, G.R. Crelier, S. Marrett, and G.B. Pike. Additive combination of perfusion responses to bypercapnia and visual stimulation. In Pro- ceeding of The Eighth International Society of Magnetic Resonance in Medicine Con ference.

Page 93: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

[35] R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. Additive combination of perfusion responses to hypercapnia and visual stimulation. In Pro- ceeding of The Eighth international Society of Magnetic Resonance in Medicine Con ference.

[36] R.D. Hoge, J. Atkinson, B. Gi11, G.R. Crelier, S. Marrett, and G.B. Pike. Compar- ison of bulk CBF/CMRO coupling in human Vl during monocular and binocular stimulation. In Proceeding of The Eighth International Society of Magnetic Res- onance in Medicine Conference.

[3 71 R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. Compar- ison of bulk CBFKMRO coupling in human VI during monocular and binocular stimulation. in f roceeding of the Fïfth International Conference on Functional Mapping of the Hurnan Brain.

[38] R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. investi- gation of BOLDsignal dependence on CBF and CMRO?: The deoxyhemoglobin dilution model. In Proceedingof the Fifrh International Conference on Functionai Mapping of the Human Brain.

[39] R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. Linear coupling between cerebral blood flow and oxygen. in Proceeding of the Fi/ih International Conference on Functional Mapping of the Human Brain.

[40] R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. Metabol- ic BOLD signal attenuation demonstrated using hypercapnically matched perfu- sion levels. In Proceeding of the Fqth International Conference on Funetional Mapping of the Human Brain.

[4 11 R.D. Hoge, J. Atkinson, B. Gi11, G.R. Crelier, S. Marrett, and G.B. Pike. Non- linear BOLD and perfusion dynamics in human V 1 . in Proceeding of The Eighth International Socieiy of Magnetic Resonance in Medicine Confererzce.

(421 R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. Non- linear BOLD and perfusion dynarnics in human VI. In Proceeding of the F13h International Con ference on Functional Mapping of the Hurnan Brain.

1431 R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. Stimulus- dependent response waveforms in human V1: Detection of BOLD and perfusion overs hoo t. Ln Proceeding of the F'@h International Con ference on Functional Màpping of the Humun h i n .

[44] R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. Stimulus- dependent response waveforms in human V1: Detection of BOLD and perfusion overs hoo t. In Proceeding of The Eighth International Society of Magnetic Reso- nance in Medicine Conference.

Page 94: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

[45] R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. hves- tigation of BOLD signal dependence on CBF and CMRo,: the deoxyhemoglobin dilution model. Submitted to Magnetic Resonance in Medicine, Oct, 1 998.

1463 R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. Linear coupling between cerebral blood flow and and oxygen consumption in activated human cortex. Submitted to Proceedings of the National Acaderny of Science, Oct. 1998.

[47] R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. Non- linear BOLD and perfusion dynamics in human VI. Szibmitted to Netiroirnage, Oct. 1998.

[48] R.D. Hoge, J. Atkinson, B. Gill, G.R. Crelier, S. Marrett, and G.B. Pike. Investi- gation of BOLD signal dependence on CBF and CMRo, : The deoxyhemoglobin model. In Proceeding of The Eighth International Society of Magnetic Resonance in Medicine Conference, May 1999.

[49] R.D. Hoge, B. Gill, J. Atkinson, G.R. Crelier, S. Marrett, and G.B. Pike. Inves- tigation of CMRo,/CBF coupling in human VI using fMRi. In Proceedings of the Fourth international Conference on Functional Mapping of the Human Brain, June 1998.

[SOI R.D. Hoge, B. Gill, G.R. Crelier, and G.B. Pike. Cornparison of perîusion and (bold) responses in human visual cortex to blob-selective stimuli, inter-blob- selective stimuli, and hypercapnia. In Proceedings of the Seventh Inter-national Society of Magnetic Resonance in Medicine Con ference, A p d 1 998.

[5 11 D.H. Hubel and T.N. Wiesel. Brain mechanisms of vision. Scient@ her ican, Sept. 1979.

1521 P. M. Jakob, M. A. Griswold, K. O. Lovblad, Q. Chen, and R. R. Edelman. Half- fourier burst imaging on a clinical scanner. Magnetic Resonance in Medicine, 3 8(4):534-40, Oct 1997.

CS33 G. Johnson, D. A. Feinberg, and V. Venkataraman. A cornparison of phase en- coding ordering schemes in t2-weighted grase imaging. Magnetic Resonance in Medicine, 36(3):427-35, Sep 1996.

[54] S. G. Kim. Quantification of relative cerebral blood flow change by flow-sensitive alternating inversion recovery (fair) technique: appiication to functional mapping. Mugnetic Resonance in Medicine, 34(3):293-30 1, Sep 1995.

[55] S. G. Kim and N. V. Tsekos. Perfusion imaging by a flow-sensitive altemat- ing inversion recovery (fair) technique: application to functional brain imaging

Page 95: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

Cpublished erratum appears in magn reson med 1997 may;37(5):675]. Magnetic Resonance in Medicine, 3 7(3):42%3 5, Mar 1 997.

[56] S.G. Kim and N.V. Tsekos. Multi-slice perfusion-based fùnctional M M using the FAIR technique. In Proceedings of the Yh Meeting of the Iniernational Sociep of Magnetic Resonance in Medicine, page 3 75, 1 997.

[57] K. K. Kwong. Functional magnetic resonance imaging with echo planar imaging. Magnetic Resonance Quarteriy, 1 1 ( 1 ): 1-20, Mar 1995.

[58] K. K. Kwong, J. W. Belliveau, D. A. Chesler, 1. E. Goldberg, R. M. Weisskoff, B. P. Poncelet, D. N. Kemedy, B. E. Hoppel, M. S. Cohen, R. Turner, and a. 1. et. Dynamic magnetic resonance imaging of human brain activity during pnmary sensory stimulation. Proceedings of the National Academ-v of Sciences of the United States of America, 89(12):5675-9, Jun 15 1992.

[59] K. K. Kwong, D. A. Chesler, R. M. Weisskoff, K. M. Donahue, T. L. Davis, L. Ostergaard, T. A. Campbell, and B. R. Rosen. Mr perfusion studies with t l- weighted echo planar imaging. Magnetic Resonance in Medicine, 34(6):878-87, Dec 1995.

[60] K.K. Kwong. Current issues in fûnctional MN. NMR in Biomedicine, 1 0: 1 57- 159, 1997.

[61] S. Lai and G. H. Glover. Three-dimensional spiral fMRI technique: a comparison with 26 spiral acquisition. Magnetic Resonance in Medicine, 39( 1):68-78, Jan 1998.

[62] D. Le Bihan, editor. Dt#iusion and Perfiusion Magnetic Resonance Inzaging, Ap- plications to fMRI. Raven Press, 1 995.

[63] P. Mansfield and I.L. Pykett. Biological and medical imaging by NMR. Jorn-na/ of Magnetic Resonance, 29:3 55-3 73, 1978.

[64] A. C. McLaughlin, F. Q. Ye, J. J. Pekar, A. K. Santha, and J. A. Frank. Ef- fect of magnetization transfer on the measurement of cerebral blood flow using steady-state arterial spin tagging approaches: a theoretical investigation. Magnet- ic Resonance in Medicine, 37(4):50 1-1 0, Apr 1997.

1651 S. Meiboom and D. Gill. Modified spin-echo method for measuring nuclear re- laxation times. Review of Scientijïc Instruments, 29(8):68849 1, 1958.

[66] P. Neelin. introduction to MINC. http://www.bic.mni.mcgill.ca/so~are/minc/ minc. htm 1.

Page 96: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

[67] D. G. Nishimura, P. Irarrazabal, and C. H. Meyer. A velocity k-space analysis of flow effects in echo-planar and spiral imaging. Magnetic Resonance in Medicine, 33(4): 54-6, Apr 1 995.

[68] D.G. Nishimura. Principles of magnetic resonance imaging. Stanford University Course Notes, 1996.

[69] Douglas C Noll, Craig H Meyer, John M Pauly, Dwight G Nishimura, and Albert Macovski. A homogeneity correction method for magnetic resonance irnaging with time-varying gradients. IEEE Transaciions on Medical Imuging, 10(4):629- 637, Dec 199 1.

[70] Douglas C Noll, Dwight G Nishimura, and Albert Macovski. Homodyne detec- tion in magnetic resonance irnaging. IEEE Transactions on Medical irnaging, 1 O(2): 154-1 63, Jun 199 1.

1711 K. Oshio and D. A. Feinberg. Grase (gradient- and spin-echo) imaging: a novel fast M M technique. Magnetic Resonance in Medicine, 20(2):344-9, Aug 199 1.

[72] E.M. Purcell, H.C. Torrey, and R.V. Pound. Resonance absorbtion by nuclear magnetic moments in a solid. Physicul Review, 69:37-38, 1946.

1731 D. A. Roberts, J. A. Detre, L. Bolinger, E. K. Insko, and J. S. Leigh Jr. Quantitative magnetic resonance imaging of human brain perfusion at 1.5 t using steady-state inversion of artenal water. Proceedings of the National Actrdemy of Sciences of ihe UniiedStates ofAmerica, 91(1):33-7, Jan 4 1994.

[74] J. Rochford. Staiistics for the neurosciences. Department of Psychiatry McGill University, 1998.

(751 1. Rock, editor. The Perceptual World, chapter 1. pages 3-24. W. H. Freeman and Company, 1990.

[76] B. R. Rosen, J. W. Belliveau, B. R. Buchbinder, R. C. McKinstry, L. M. Porkka, D. N. Kennedy, M. S. Neuder, C. R. Fisel, H. J. Aronen, K. K. Kwong, and a. 1. et. Contrast agents and cerebral hemodynarnics. Magnetic Resonance in Medicine, l9(2):28%92, Jun 199 1.

[77] B. R. Rosen, J. W. Belliveau, J. M. Vevea, and T. J. Brady. Perfusion imaging with NMR contrast agents. Magnetic Resonance in Medicine, 14(2):249-65, May 1990.

[78] B. R. Rosen, R. L. Buckner, and A. M. Dale. Event-related functional MM: past, present, and future- Proceedings of the National Academy of Sciences of the United States of America, 95(3):773-80, Feb 3 1998.

Page 97: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

[79] M.I. Sereno, A.M. Dale, J.B. Reppas, K.K. Kwong, J.W. Belliveau, T.J. Brady, B.R. Rosen, and R.B.H. Tootell. Borders o f multiple visual areas in humans re- vealed by functional magnetic resonance imaging. Science, 268:889-93, May 1995.

[80] T. Siegal, R. Rubinstein, T. Tzuk-Shina, and J. M. Gomori. Utility o f relative cere- brai blood volume mapping derived fiom perfùsion magnetic resonance imaging in the routine follow up of brain tumors. Journal of Neurosurgew, 86(1):22-7. Jan 1997.

[81] M.S. Silver, R.I. Joseph, C.N. Chen, VJ. Sank, and D.I. Hoult. Highly selective n/2 and T pulse generation. Journal of Magnefic Resoriance, 34:52%33, 1 984.

[823 R.M. Weisskoff and S. Kiihne. MRI susceptometry: image-based measurement of absolute susceptibility of MR contrast agents and hwnan blood. Magneric Resonance in Medicine, 24(2):3 75-83, April 1 992.

[83] D. S. Williams, J. A. Detre, J. S. Leigh, and A. P. Koretsky. Magoetic resonance imaging of perfusion using spin inversion of arterial water [published erratum appears in prac natl acad sci u s a 1992 may 1;89(9):4220]. Proceedings of the National Academy of Sciences of the United States of Arnerica, 89( 1 ):2 1 2 4 , Jan 1 1992.

1841 S D Wolff and R S Balaban. Magnetization transfer contrast (MTC) and tissue water proton relaxation in vivo. Magnetic Resonance in Medicine, 1 O( I ): 1 3 5- 1 44, 1989.

185 ] S. D. Wolff and R. S. Balaban. Magnetization transfer imaging: practical aspects and clinical applications. Radiohgy, 1 92(3): 593-9, Sep 1 994.

[86] E. C. Wong, R. B. Buxton, and L. R. Frank. Implementation of quantitative per- fusion imaging techniques for functiona1 brain mapping using pulsed artenal spin labeling [see comments] . NMR in Biomedicine, 1 O(4-5):237+9, Jun-Aug 1 997.

[87] E. C. Wong, R. B. Buxton, and L. R. Frank. Quantitative imaging of per-fusion us- ing a single subtrac tion (quipss and quipss II). Magnetic Resonance in Medicine, 39(5):702-û, May 1998.

[88] E.C. Wong, R.B. Buxton, and L.R Frank. A theoretical and experimental compar- ison of continuous and pulsed arterial spin labelling techniques for quantitative perfusion irnaging. Magnetic Resonance in Medicine, 4O:MfM 55, 1 998.

[89] E.C. Wong, L.R. Frank, and R.B. Buxton. Quantitative multislice perfusion imag- ing using QUIPSS II, EPISTAR, FAIR and PICORE. In Proceedings of the jLh

Meeting of the Intemational Socieiy of Magnetic Resonance in Medicine, page 85, 1997.

Page 98: Multislice Perfusion Imaging With Arterial Spin Labelling · 2005. 2. 9. · Multislice Perfusion Imaging With Arterial Spin Labelling: Applications to Functional MRI Bradford A

1901 K.J. Worsley, S. Marrett, P. Neelin, A.C. Vandal, K.J. Friston, and A.C. Evans. A united statistical approach for determining significant signals in images of cerebral activation. Human Bmin Mapping, 458-73, 1996-

[91] G.A. Wright, B.S. Hu, and A. Macovski. Estimating oxygen saturation of blood in vivo with MR imaging at 1 3. Journal of Magnetic Resonance Imaging, 1 (3 ):27% 84, May 199 1.

1921 Y. Yang, J. A. Frank, L. Hou, F. Q. Ye, A. C. McLaughlin, and J. H- Duyn. Multi- slice imaging of quantitative cerebral perhision with pulsed arterial spin labeling. Magnetic Resonance in Medicine, 39(5): 825-3 2, May 1 998.

1931 Y. Yang, G. H. Glover, P- van Gelderen, A. C. Patel, V. S. Mattay, J. A. Frank, and J. H. Duyn. A comparison of fast mr scan techniques for cerebral acti- vation studies at 1.5 tesla Cpublished erratum appears in magn reson med 1998 mar;3 9(3):following 505 j. Magnetic Resonance in Medicine, 3 9( 1 ):6 1-7, Jan 1998.

[94] F. Q. Ye, V. S. Mattay, P. Jezzard, J. A. Frank, D. R. Weinberger, and A. C. M- claughlin. Correction for vascular artifacts in cerebral blood flow values mea- sured by using artenal spin tagging techniques. Magnetic Resonance in Medicine, 3 7(2):226-35, Feb 1 997.

[95] J.H. Zar. Biostatistical Analysk Prentice-Hall, foürth edition, 1 999.

[96] W. Zhang, A. C. Silva, D. S. Wiiliams, and A. P. Koretsky. Nmr measurement of pefision using arterial spin labeling without saturation of macromolecuIar spins. Magnetic Resonance in Medicine, 33(3):370-6, Mar 1995.

[97] W. Zhang, D. S. Williams, J. A. Detre, and A. P. Koretsky. Measurement of brain perfusion by volume-localized nmr spectroscopy using inversion of artenal water spins: accounting for transit tirne and cross-relaxation. Magnetic Resonance in Medicine, 25(2):362-7 1, Jun 1992.