multifunctional gas analyzer of multi-component mixtures ГАММА · hydrogen – oxygen carbon...

3
Temperature range of ambient and analyzable mediums Relative ambient humidity Pressure range of analyzable gas medium Flow rate of analyzable gas medium Humidity of analyzable gas medium Mass concentration of dust in analyzable medium Industrial vibration: - frequency - amplitude +5 / +45 up to 80 % from 630 to 800 mm Hg (0,9 ±0,1) l/min 3 5 g/m at most 3 10 mg/m at most from 10 to 55 Hz 0,35 mm at most Parameters Values Functional capabilities > Digital indication of measured component concentration on each measuring channel; > Generation of unified output current signal proportional to analyte concentration; > Generation of green-light indication upon gas analyzer powering on; > Generation of text and light alarm indicating that analyte concentration exceeds or drops below threshold values; > Self-diagnostics with output of its results onto display; Multifunctional gas analyzer of multi-component mixtures ГАММА-100 Designed for process and ecological monitoring by means of continuous automatic measurement of 1 or up to 3 components. Application In automated process control systems in ferrous and non-ferrous metallurgy, cement and glass industries, at brick and ceramics manufacturing plants. Gas analyzer type - fixed. Sampling – forced (flow rate pump is not a component of the device). > Commutation of external circuits with two groups of switching relay (250 V, 2,8 A) contacts (“dry contacts”) for automatic actuating device switching-on (-off) upon alarm actuation indicating that the measured component has reached the set threshold values (excess or drop); > Output if data on measured component concentration in a digital form via RS 232, RS 485 and Ethernet (depending on version), in addition to the option of concentration reading on main channels, gas analyzer also has an option of control via RS 232 (RS 485) communication channel, i.e.: - adjustment of scale initial value; - adjustment of scale finite value; - thresholds setting. Operating conditions Interval between verifications for thermoconductometric and infrared sensors is extended up to 30 days, for thermomagnetic – up to 6 months. Instruments for ecological and process monitoring

Upload: others

Post on 22-Mar-2020

24 views

Category:

Documents


0 download

TRANSCRIPT

Temperature range of ambient and analyzable mediums

Relative ambient humidity

Pressure range of analyzable gas medium

Flow rate of analyzable gas medium

Humidity of analyzable gas medium

Mass concentration of dust in analyzable medium

Industrial vibration:- frequency- amplitude

+5 / +45

up to 80 %

from 630 to 800 mm Hg

(0,9 ±0,1) l/min

35 g/m at most

310 mg/m at most

from 10 to 55 Hz0,35 mm at most

Parameters Values

Functional capabilities> Digital indication of measured component concentration on each measuring channel;> Generation of unified output current signal proportional to analyte concentration;> Generation of green-light indication upon gas analyzer powering on;> Generation of text and light alarm indicating that analyte concentration exceeds or drops below threshold values;> Self-diagnostics with output of its results onto display;

Multifunctional gas analyzer of multi-component mixtures ГАММА-100

Designed for process and ecological monitoring by means of continuous automatic measurement of 1 or up to 3 components.

ApplicationIn automated process control systems in ferrous and non-ferrous metallurgy, cement and glass industries, at brick and ceramics manufacturing plants.

Gas analyzer type - fixed.Sampling – forced (flow rate pump is not a component of the device).

> Commutation of external circuits with two groups of switching relay (250 V, 2,8 A) contacts (“dry contacts”) for automatic actuating device switching-on (-off) upon alarm actuation indicating that the measured component has reached the set threshold values (excess or drop);> Output if data on measured component concentration in a digital form via RS 232, RS 485 and Ethernet (depending on version), in addition to the option of concentration reading on main channels, gas analyzer also has an option of control via RS 232 (RS 485) communication channel, i.e.: - adjustment of scale initial value; - adjustment of scale finite value; - thresholds setting.

Operating conditions

Interval between verifications for thermoconductometric and infrared sensors is extended up to 30 days, for thermomagnetic – up to 6 months.

Instruments for ecological and process monitoring

Hydrogen Н2

Oxygen О2

Oxygen О2

Helium Не

Oxygen О2

Oxygen О2

Oxygen О2

vol. fraction %

vol. fraction %

vol. fraction %

vol. fraction %

vol. fraction %

vol. fraction %

vol. fraction %

0-0,5; 0-1; 0-2;0-5; 0-10; 0-20;

0-30; 0-50; 0-70; 0-100

50-100; 70-100

0-1; 0-2

0-1; 0-2; 0-3

10-100

0-1; 0-2; 0-5;0-10; 0-21; 0-30;

0-50; 0-80; 0-100;15-30; 50-80;

80-100; 90-100;95-100; 98-100

0-1; 0-2; 0-5;0-10; 0-21;0-30; 0-50

thermoconductometric

thermoconductometric

thermoconductometric

thermoconductometric

thermomagnetic

thermomagnetic

thermomagnetic

hydrogen – methane

oxygen – helium

oxygen – hydrogen

helium – nitrogen

oxygen - nitrogen

oxygen – argon

oxygen – combustion gas

Carbon monoxide СО

Carbon monoxide СО

Carbon monoxide СО

Carbon dioxide СО2

Carbon dioxide СО2

Carbon dioxide СО2

Methane СН4

Methane СН4

Sulfur dioxide SO2

Nitrogen N2

Hydrogen Н2

Nitrogen oxide NO

Hydrogen Н2

Hydrogen Н2

Hydrogen Н2

vol. fraction %

3г/м

vol. fraction %

vol. fraction %

vol. fraction %

3g/m

vol. fraction %

vol. fraction %

3g/m

vol. fraction %

vol. fraction %

vol. fraction %

0-200; 0-500;0-1 000; 0-2 000

0-0,5; 0-1; 0-2;0-5; 0-10; 0-20;

0-30; 0-50; 0-70; 0-100

0-15

0-100; 0-200; 0-500;0-1 000; 0-2 000

0-0,5; 0-1; 0-2;0-5; 0-10; 0-20;

0-30; 0-50; 0-70; 0-100

0-30; 0-50;40-100; 90-100

0-500; 0-1 000; 0-2 000

0-0,5; 0-1; 0-2;0-5; 0-10; 0-20;

0-30; 0-50; 0-70; 0-100

0-2; 0-5; 0-10;0-20; 0-60

0-20; 80-100;0-40; 0-60; 60-100

0-0,5; 0-1

0-1; 0-2

0-1; 0-2; 0-3; 0-5;0-10; 0-20; 0-30;0-40; 0-50; 0-60;

0-80; 0-100; 50-100;60-100; 80-100;90-100; 95-100

0-1; 0-2;0-3; 90-100

0-1; 0-2; 0-3

optical-acoustic

optical-acoustic

optical-acoustic

optical-acoustic

optical-acoustic

thermoconductometric

optical-acoustic

optical-acoustic

optical-acoustic

thermoconductometric

thermoconductometric

optical-acoustic

thermoconductometric

thermoconductometric

thermoconductometric

carbon dioxide-nitrogen

nitrogen – helium

hydrogen – nitrogen

hydrogen – air

hydrogen – oxygen

carbon dioxide-nitrogen

Multifunctional gas analyzer of multi-component mixtures ГАММА-100Monitored gases and measuring ranges

Instruments for ecological and process monitoring

Units of measurementMeasured component Measuring range Principle of measurement

Composition of analyzable medium

vol. fraction ppm

vol. fraction ppm

vol. fraction ppm

hydrogen – carbon dioxide (10%)

– oxygen (2%)– nitrogen the rest

Multifunctional gas analyzer of multi-component mixtures ГАММА-100

ИБЯЛ.413251.001

ИБЯЛ.413251.001-01

ИБЯЛ.413251.001-02

ИБЯЛ.413251.001-03

ИБЯЛ.413251.001-04

ИБЯЛ.413251.001-05

ИБЯЛ.413251.001-06

ИБЯЛ.413251.001-07

ИБЯЛ.413251.001-08

ИБЯЛ.413251.001-09

Designation

w l h

Weight, kg

3

3

2

2

2

2

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

2

2

2

2

none

none

1

1

none

none

available

none

available

none

available

none

available

none

available

none

485

485

485

485

485

485

485

485

300

300

280

280

280

280

146

146

146

146

146

146

435

435

435

435

435

435

435

435

435

435

20

20

20

18

16

16

15

15

14

14

Gas analyzer versions

Design featuresGas analyzer is a single-unit device with measuring sensors built in its enclosure. Front panel has: LCD that displays measurement data, control keyboard, flow rate indication of monitored sample. Gas analyzer is furnished with RS 232, RS 485, Ethernet interfaces, via which data can be transferred to PC. Gas analyzer also has unified current output (0 - 5 mA or 4 - 20 mA, can be switched) and relay contacts for commutation of external circuits. Gas analyzer is powered with 220 V from AC mains.

Standard equipment> Gas analyzer ГАММА-100;> SPTA set;> CD with software;> Documentation.

Optional items can be ordered> TGM-SSS cylinders;> Flow rate indicator ИБЯЛ.418622.001-07 (or analogous one) to ensure sample flow rate

3 of (0,9 ± 0,1) dm /min;> Flow rate indicator ИБЯЛ.418622.001-10 (or analogous one) to ensure gas flow rate

3 of (0,2 ± 0,05) dm /min;

In case parameters of analyzable medium at gas analyzer inlet do not meet technical requirements, a suitable sample preparation system can be supplied on separate order. This system can comprise cooler, flow rate pump, sample dehydrator and other devices.

> Fine adjustment valve ИБЯЛ.306577.002-03 (of stainless steel);> Fine adjustment valve ИБЯЛ.306577.002 (of brass);> Sample preparation elements in accordance with as analyzer application.

Instruments for ecological and process monitoring

Quantity of measuring channels

Principle of measurement and maximum quantity of measuring channels

Overall dimensions, mm

Thremo-magnetic

Thermo-conducto-

metric

Optical-acoustic

Ethernet interface