Poster Note 6 43 70 Multi-channeling LC-MS/MS Forensic Methods for High-Throughput Urine Screening to Detect Buprenorphine and Ethanol Use Joseph Di Bussolo, 1 Ian White, 2 Raidiri Castillo 3 and Hashim Othman 3 1 Thermo Fisher Scientific, West Chester, PA; 2 West Chester University of Pennsylvania, West Chester, PA; 3 BioReference Laboratories, Elmwood Park, NJ

Upload: others

Post on 26-Mar-2021

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Multi-channeling LC-MS/MS Forensic Methods for High …tools.thermofisher.com/content/sfs/posters/PN-64370-LC... · 2016. 2. 3. · for EtG/EtS were diluted 1:10 with water and then

Poster Note 64370

Multi-channeling LC-MS/MS Forensic Methods for High-Throughput Urine Screening to Detect Buprenorphine and Ethanol UseJoseph Di Bussolo,1 Ian White,2 Raidiri Castillo3 and Hashim Othman3 1Thermo Fisher Scientific, West Chester, PA; 2West Chester University of Pennsylvania, West Chester, PA; 3BioReference Laboratories, Elmwood Park, NJ

Page 2: Multi-channeling LC-MS/MS Forensic Methods for High …tools.thermofisher.com/content/sfs/posters/PN-64370-LC... · 2016. 2. 3. · for EtG/EtS were diluted 1:10 with water and then

2Multi-channeling LC-MS/MS Forensic Methods for High-Throughput Urine Screening to Detect Buprenorphine and Ethanol Use

Conclusions The modified Prelude - Endura system, which utilized unique positive-

displacement pumps, produced results for Bup/Norbup and EtG/EtS methods that were consistent with those produced by the conventional Transcend - Endura system.

The maximum throughput from both 4-channel systems was 34 injections per hour when multiplexing the two forensic methods across all four channels.

The modified Prelude system reduced solvent consumption by at least 60%. Other benefits of using this system were:

Avoided pulsations of reciprocating pumps

Easier to use, purge, prepare and maintain

Smaller footprint compared to Transcend LX4

Overview Purpose: Compare the performance of running LC-MS/MS forensic methods used to detect buprenorphine and ethanol use on two different four-channel UHPLC systems - one utilizing conventional reciprocating pumps and the other using positive-displacement pumps.

Methods: Reversed-phase liquid chromatography of analytes with corresponding stable-isotope internal standards eluting from up to four UHPLC channels into a heated electro-spray ionization (HESI) source of a triple- quadrupole mass spectrometer were used to measure urine levels of the forensic compounds Buprenorphine (Bup) & Norbuprenorphine (Norbup) after hydrolysis and Ethyl-Glucuronide (EtG) and Ethyl Sulfate (EtS) after dilution.

Results: Running Bup/Norbup batches on two channels while EtG/EtS batches ran on the remaining two channels yielded a throughput of 17 injections per hour for each batch, which totaled 34 injections per hour. The four-channel system with positive-displacement pumps reduced solvent consumption by at least 60%. Results from this system were within +/- 15% of those determined on a conventional multichannel system using reciprocating pumps.

Introduction Many forensic laboratories run several different LC-MS methods in series on a single-channel LC-MS system. If the methods involve different ion sources, columns and mobile phases, the changeover is time consuming, labor intensive and increases the risk of mistakes and contamination. A four-channel UHPLC system multiplexed into one mass spectrometer permits parallel batches of up to four different methods utilizing a common ion source and unique columns and mobile phases to be completed in a fraction of the time and effort. However, conventional systems utilizing reciprocating pumps needlessly consume mobile-phase solvents between injections. The Thermo Scientific™ Prelude™ SPLC system, which utilizes positive-displacement pumps does not waste solvents between injections. A modified version of this multi-channel UHPLC system was tested by running batches for two forensic methods - Bup/Norbup and EtG/EtS - and comparing those results to those from a conventional UHPLC system that ran the same sets of calibrators, QCs and specimens.

Mass Spectrometry System Control & Data Analysis

Thermo Scientific™ TraceFinder™ with Aria™ MX software was used to control the Transcend II LX4 and Thermo Scientific™ TSQ Endura™ MS/MS systems, submit batches to desired channels as well as for analyzing data and reporting results.

The Endura triple-quadrupole mass spectrometer was used with HESI. Ion source and MS/MS conditions are described in Figure 4.

FIGURE 4. Mass spectrometry conditions.

Common source

conditions:

Bup/Norbup

acquisition

method:

EtG/EtS

acquisition

method:

FIGURE 5. Multi-channeling Bup/Norbup and EtG/EtS batches.

FIGURE 5. Calibration Plots from Modified Prelude - Endura system.

Multi-channeling LC-MS/MS Forensic Methods for High-Throughput Urine Screening to Detect Buprenorphine and Ethanol Use Joseph Di Bussolo1, Catherine Lafontaine1, Mercedes Castillo2, Thomaskutty Thomas2 & Hashim Othman2. 1Thermo Fisher Scientific, West Chester, PA, 2BioReference Laboratories, Elmwood Park, NJ

Methods Consumables

HPLC-grade solvents, reagents and most other consumables were from Thermo Fisher Scientific. Buprenorphine and norbuprenorphine and their corresponding deuterated internal standards as well as ethyl sulfate and ethyl glucuronide and their corresponding -deuterated internal standards were from Cerilliant (Round Rock, TX). Calibrators were made by mixing these standards with synthetic urine. Liquid urine controls were from Biochemical Diagnostics (Edgewood, NY). β-glucuronidase powder was purchased from Sigma-Aldrich.

Sample Preparation:

Urine specimens and corresponding calibrators and QCs to be analyzed for Bup/Norbup were hydrolyzed by incubating a mixture of 150 µL of β-glucuronidase solution (10,000 U/mL, pH 5) with 200 µL of specimen, and 50 µL of IS solution containing - buprenorphine-D3 & norbuprenorphine-D4 for 1.5 hours at 60°C. Each preparation was then mixed with 200 µL of cold methanol and refrigerated for 10 minutes before centrifugation. 20 µL injections of supernatants from each preparation were made into the UHPLC systems.

Urine specimens and corresponding calibrators and QCs to be analyzed for EtG/EtS were diluted 1:10 with water and then spiked with 50 µL of IS solution containing EtG-D5 & EtS-D5 before making 20 µL injections into the UHPLC system.

Liquid Chromatography

The “conventional” multi-channel LC system was a Thermo Scientific™ Transcend™ LX4 equipped with binary-solvent pumps and a dual-arm autosampler configuration. Thermo Scientific™ Accucore™ RP-MS, 2.6 µm, 50 x 2.1 mm HPLC columns were used for the Bup/Norbup method. Mobile phase conditions for this method are described in Figure 1. Thermo Scientific™ Synchronis™ aQ, 5 µm, 50 x 3.0 mm HPLC columns were used for the EtG/EtS method. Mobile phase conditions for this method are described in Figure 2.

A Thermo Scientific™ Prelude™ SPLC system was modified to permit injections across four channels, similar to the Transcend LX4 system. The Bup/Norbup method on this system utilized the same column and mobile phase conditions as the Transcend system. However, the EtG/EtS method was modified to use the Thermo Scientific™ Synchronis™ aQ, 3 µm, 100 x 3.0 mm HPLC columns, as shown in Figure 3.

Results Throughput and Multi-channeling Bup/Norbup and EtG/EtS batches

Multi-channeling Bup/Norbup across 2 channels while EtG/EtS runs on one allowed 23 Bup/Norbup and 12 EtG/EtS injections/hour. Using 2 channels for each does not increase throughputs but ensures completion of all batches even if one channel stops because of leakage or over-pressurization. Multi-channeling two batches is illustrated in Figure 5.

FIGURE 1. Transcend Bup/norbup LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

Comparison between Transcend and Modified Prelude Systems

Batches of calibrators, QCs and at least 20 specimens were prepared and shared between the two multi-channel systems. Typical calibration plots from the modified Prelude - Endura system are shown in Figure 5. Desired quantitation ranges were consistently linear (r2 > 0.99 with 1/X weighting) whether when calibrators were injected into one channel or across all channels.

Only results that were within the analytical ranges for each method were used for comparisons. Differences in calculated amounts averaged less than +/- 5% within a maximum of +/- 15%. Scatter plots in Figure 6 showed good correlations between the two systems.

FIGURE 6. Scatter plots of results - Transcend vs. modified Prelude.

FIGURE 2. Transcend EtG/EtS LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

FIGURE 3. Modified Prelude EtG/EtS LC conditions.

Start data 0.75 min Data window: 1.5 min Total run time: 4.0 min

0

50

100

150

200

250

300

350

400

0 100 200 300 400

Tran

scen

d

Modified Prelude

Bup

0

100

200

300

400

500

600

700

0 100 200 300 400 500 600 700

Tran

scen

d

Modified Prelude

Norbup

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

0 2000 4000 6000 8000 10000

Tran

scen

d

Modified Prelude

EtG

0.00

1000.00

2000.00

3000.00

4000.00

5000.00

6000.00

0 1000 2000 3000 4000 5000 6000 7000

Tran

scen

d

Modified Prelude

EtS

For forensic toxicology use only.

©2015 Thermo Fisher Scientific. All Rights Reserved.

Cerilliant is a registered trademark of the Sigma-Aldrich Corporation. Biochemical Diagnostics is a registered trademark of Biochemical Diagnostics, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.

This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.

Conclusions The modified Prelude - Endura system, which utilized unique positive-

displacement pumps, produced results for Bup/Norbup and EtG/EtS methods that were consistent with those produced by the conventional Transcend - Endura system.

The maximum throughput from both 4-channel systems was 34 injections per hour when multiplexing the two forensic methods across all four channels.

The modified Prelude system reduced solvent consumption by at least 60%. Other benefits of using this system were:

Avoided pulsations of reciprocating pumps

Easier to use, purge, prepare and maintain

Smaller footprint compared to Transcend LX4

Overview Purpose: Compare the performance of running LC-MS/MS forensic methods used to detect buprenorphine and ethanol use on two different four-channel UHPLC systems - one utilizing conventional reciprocating pumps and the other using positive-displacement pumps.

Methods: Reversed-phase liquid chromatography of analytes with corresponding stable-isotope internal standards eluting from up to four UHPLC channels into a heated electro-spray ionization (HESI) source of a triple- quadrupole mass spectrometer were used to measure urine levels of the forensic compounds Buprenorphine (Bup) & Norbuprenorphine (Norbup) after hydrolysis and Ethyl-Glucuronide (EtG) and Ethyl Sulfate (EtS) after dilution.

Results: Running Bup/Norbup batches on two channels while EtG/EtS batches ran on the remaining two channels yielded a throughput of 17 injections per hour for each batch, which totaled 34 injections per hour. The four-channel system with positive-displacement pumps reduced solvent consumption by at least 60%. Results from this system were within +/- 15% of those determined on a conventional multichannel system using reciprocating pumps.

Introduction Many forensic laboratories run several different LC-MS methods in series on a single-channel LC-MS system. If the methods involve different ion sources, columns and mobile phases, the changeover is time consuming, labor intensive and increases the risk of mistakes and contamination. A four-channel UHPLC system multiplexed into one mass spectrometer permits parallel batches of up to four different methods utilizing a common ion source and unique columns and mobile phases to be completed in a fraction of the time and effort. However, conventional systems utilizing reciprocating pumps needlessly consume mobile-phase solvents between injections. The Thermo Scientific™ Prelude™ SPLC system, which utilizes positive-displacement pumps does not waste solvents between injections. A modified version of this multi-channel UHPLC system was tested by running batches for two forensic methods - Bup/Norbup and EtG/EtS - and comparing those results to those from a conventional UHPLC system that ran the same sets of calibrators, QCs and specimens.

Mass Spectrometry System Control & Data Analysis

Thermo Scientific™ TraceFinder™ with Aria™ MX software was used to control the Transcend II LX4 and Thermo Scientific™ TSQ Endura™ MS/MS systems, submit batches to desired channels as well as for analyzing data and reporting results.

The Endura triple-quadrupole mass spectrometer was used with HESI. Ion source and MS/MS conditions are described in Figure 4.

FIGURE 4. Mass spectrometry conditions.

Common source

conditions:

Bup/Norbup

acquisition

method:

EtG/EtS

acquisition

method:

FIGURE 5. Multi-channeling Bup/Norbup and EtG/EtS batches.

FIGURE 5. Calibration Plots from Modified Prelude - Endura system.

Multi-channeling LC-MS/MS Forensic Methods for High-Throughput Urine Screening to Detect Buprenorphine and Ethanol Use Joseph Di Bussolo1, Catherine Lafontaine1, Mercedes Castillo2, Thomaskutty Thomas2 & Hashim Othman2. 1Thermo Fisher Scientific, West Chester, PA, 2BioReference Laboratories, Elmwood Park, NJ

Methods Consumables

HPLC-grade solvents, reagents and most other consumables were from Thermo Fisher Scientific. Buprenorphine and norbuprenorphine and their corresponding deuterated internal standards as well as ethyl sulfate and ethyl glucuronide and their corresponding -deuterated internal standards were from Cerilliant (Round Rock, TX). Calibrators were made by mixing these standards with synthetic urine. Liquid urine controls were from Biochemical Diagnostics (Edgewood, NY). β-glucuronidase powder was purchased from Sigma-Aldrich.

Sample Preparation:

Urine specimens and corresponding calibrators and QCs to be analyzed for Bup/Norbup were hydrolyzed by incubating a mixture of 150 µL of β-glucuronidase solution (10,000 U/mL, pH 5) with 200 µL of specimen, and 50 µL of IS solution containing - buprenorphine-D3 & norbuprenorphine-D4 for 1.5 hours at 60°C. Each preparation was then mixed with 200 µL of cold methanol and refrigerated for 10 minutes before centrifugation. 20 µL injections of supernatants from each preparation were made into the UHPLC systems.

Urine specimens and corresponding calibrators and QCs to be analyzed for EtG/EtS were diluted 1:10 with water and then spiked with 50 µL of IS solution containing EtG-D5 & EtS-D5 before making 20 µL injections into the UHPLC system.

Liquid Chromatography

The “conventional” multi-channel LC system was a Thermo Scientific™ Transcend™ LX4 equipped with binary-solvent pumps and a dual-arm autosampler configuration. Thermo Scientific™ Accucore™ RP-MS, 2.6 µm, 50 x 2.1 mm HPLC columns were used for the Bup/Norbup method. Mobile phase conditions for this method are described in Figure 1. Thermo Scientific™ Synchronis™ aQ, 5 µm, 50 x 3.0 mm HPLC columns were used for the EtG/EtS method. Mobile phase conditions for this method are described in Figure 2.

A Thermo Scientific™ Prelude™ SPLC system was modified to permit injections across four channels, similar to the Transcend LX4 system. The Bup/Norbup method on this system utilized the same column and mobile phase conditions as the Transcend system. However, the EtG/EtS method was modified to use the Thermo Scientific™ Synchronis™ aQ, 3 µm, 100 x 3.0 mm HPLC columns, as shown in Figure 3.

Results Throughput and Multi-channeling Bup/Norbup and EtG/EtS batches

Multi-channeling Bup/Norbup across 2 channels while EtG/EtS runs on one allowed 23 Bup/Norbup and 12 EtG/EtS injections/hour. Using 2 channels for each does not increase throughputs but ensures completion of all batches even if one channel stops because of leakage or over-pressurization. Multi-channeling two batches is illustrated in Figure 5.

FIGURE 1. Transcend Bup/norbup LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

Comparison between Transcend and Modified Prelude Systems

Batches of calibrators, QCs and at least 20 specimens were prepared and shared between the two multi-channel systems. Typical calibration plots from the modified Prelude - Endura system are shown in Figure 5. Desired quantitation ranges were consistently linear (r2 > 0.99 with 1/X weighting) whether when calibrators were injected into one channel or across all channels.

Only results that were within the analytical ranges for each method were used for comparisons. Differences in calculated amounts averaged less than +/- 5% within a maximum of +/- 15%. Scatter plots in Figure 6 showed good correlations between the two systems.

FIGURE 6. Scatter plots of results - Transcend vs. modified Prelude.

FIGURE 2. Transcend EtG/EtS LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

FIGURE 3. Modified Prelude EtG/EtS LC conditions.

Start data 0.75 min Data window: 1.5 min Total run time: 4.0 min

0

50

100

150

200

250

300

350

400

0 100 200 300 400

Tran

scen

d

Modified Prelude

Bup

0

100

200

300

400

500

600

700

0 100 200 300 400 500 600 700

Tran

scen

d

Modified Prelude

Norbup

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

0 2000 4000 6000 8000 10000

Tran

scen

d

Modified Prelude

EtG

0.00

1000.00

2000.00

3000.00

4000.00

5000.00

6000.00

0 1000 2000 3000 4000 5000 6000 7000

Tran

scen

d

Modified Prelude

EtS

For forensic toxicology use only.

©2015 Thermo Fisher Scientific. All Rights Reserved.

Cerilliant is a registered trademark of the Sigma-Aldrich Corporation. Biochemical Diagnostics is a registered trademark of Biochemical Diagnostics, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.

This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.

Conclusions The modified Prelude - Endura system, which utilized unique positive-

displacement pumps, produced results for Bup/Norbup and EtG/EtS methods that were consistent with those produced by the conventional Transcend - Endura system.

The maximum throughput from both 4-channel systems was 34 injections per hour when multiplexing the two forensic methods across all four channels.

The modified Prelude system reduced solvent consumption by at least 60%. Other benefits of using this system were:

Avoided pulsations of reciprocating pumps

Easier to use, purge, prepare and maintain

Smaller footprint compared to Transcend LX4

Overview Purpose: Compare the performance of running LC-MS/MS forensic methods used to detect buprenorphine and ethanol use on two different four-channel UHPLC systems - one utilizing conventional reciprocating pumps and the other using positive-displacement pumps.

Methods: Reversed-phase liquid chromatography of analytes with corresponding stable-isotope internal standards eluting from up to four UHPLC channels into a heated electro-spray ionization (HESI) source of a triple- quadrupole mass spectrometer were used to measure urine levels of the forensic compounds Buprenorphine (Bup) & Norbuprenorphine (Norbup) after hydrolysis and Ethyl-Glucuronide (EtG) and Ethyl Sulfate (EtS) after dilution.

Results: Running Bup/Norbup batches on two channels while EtG/EtS batches ran on the remaining two channels yielded a throughput of 17 injections per hour for each batch, which totaled 34 injections per hour. The four-channel system with positive-displacement pumps reduced solvent consumption by at least 60%. Results from this system were within +/- 15% of those determined on a conventional multichannel system using reciprocating pumps.

Introduction Many forensic laboratories run several different LC-MS methods in series on a single-channel LC-MS system. If the methods involve different ion sources, columns and mobile phases, the changeover is time consuming, labor intensive and increases the risk of mistakes and contamination. A four-channel UHPLC system multiplexed into one mass spectrometer permits parallel batches of up to four different methods utilizing a common ion source and unique columns and mobile phases to be completed in a fraction of the time and effort. However, conventional systems utilizing reciprocating pumps needlessly consume mobile-phase solvents between injections. The Thermo Scientific™ Prelude™ SPLC system, which utilizes positive-displacement pumps does not waste solvents between injections. A modified version of this multi-channel UHPLC system was tested by running batches for two forensic methods - Bup/Norbup and EtG/EtS - and comparing those results to those from a conventional UHPLC system that ran the same sets of calibrators, QCs and specimens.

Mass Spectrometry System Control & Data Analysis

Thermo Scientific™ TraceFinder™ with Aria™ MX software was used to control the Transcend II LX4 and Thermo Scientific™ TSQ Endura™ MS/MS systems, submit batches to desired channels as well as for analyzing data and reporting results.

The Endura triple-quadrupole mass spectrometer was used with HESI. Ion source and MS/MS conditions are described in Figure 4.

FIGURE 4. Mass spectrometry conditions.

Common source

conditions:

Bup/Norbup

acquisition

method:

EtG/EtS

acquisition

method:

FIGURE 5. Multi-channeling Bup/Norbup and EtG/EtS batches.

FIGURE 5. Calibration Plots from Modified Prelude - Endura system.

Multi-channeling LC-MS/MS Forensic Methods for High-Throughput Urine Screening to Detect Buprenorphine and Ethanol Use Joseph Di Bussolo1, Catherine Lafontaine1, Mercedes Castillo2, Thomaskutty Thomas2 & Hashim Othman2. 1Thermo Fisher Scientific, West Chester, PA, 2BioReference Laboratories, Elmwood Park, NJ

Methods Consumables

HPLC-grade solvents, reagents and most other consumables were from Thermo Fisher Scientific. Buprenorphine and norbuprenorphine and their corresponding deuterated internal standards as well as ethyl sulfate and ethyl glucuronide and their corresponding -deuterated internal standards were from Cerilliant (Round Rock, TX). Calibrators were made by mixing these standards with synthetic urine. Liquid urine controls were from Biochemical Diagnostics (Edgewood, NY). β-glucuronidase powder was purchased from Sigma-Aldrich.

Sample Preparation:

Urine specimens and corresponding calibrators and QCs to be analyzed for Bup/Norbup were hydrolyzed by incubating a mixture of 150 µL of β-glucuronidase solution (10,000 U/mL, pH 5) with 200 µL of specimen, and 50 µL of IS solution containing - buprenorphine-D3 & norbuprenorphine-D4 for 1.5 hours at 60°C. Each preparation was then mixed with 200 µL of cold methanol and refrigerated for 10 minutes before centrifugation. 20 µL injections of supernatants from each preparation were made into the UHPLC systems.

Urine specimens and corresponding calibrators and QCs to be analyzed for EtG/EtS were diluted 1:10 with water and then spiked with 50 µL of IS solution containing EtG-D5 & EtS-D5 before making 20 µL injections into the UHPLC system.

Liquid Chromatography

The “conventional” multi-channel LC system was a Thermo Scientific™ Transcend™ LX4 equipped with binary-solvent pumps and a dual-arm autosampler configuration. Thermo Scientific™ Accucore™ RP-MS, 2.6 µm, 50 x 2.1 mm HPLC columns were used for the Bup/Norbup method. Mobile phase conditions for this method are described in Figure 1. Thermo Scientific™ Synchronis™ aQ, 5 µm, 50 x 3.0 mm HPLC columns were used for the EtG/EtS method. Mobile phase conditions for this method are described in Figure 2.

A Thermo Scientific™ Prelude™ SPLC system was modified to permit injections across four channels, similar to the Transcend LX4 system. The Bup/Norbup method on this system utilized the same column and mobile phase conditions as the Transcend system. However, the EtG/EtS method was modified to use the Thermo Scientific™ Synchronis™ aQ, 3 µm, 100 x 3.0 mm HPLC columns, as shown in Figure 3.

Results Throughput and Multi-channeling Bup/Norbup and EtG/EtS batches

Multi-channeling Bup/Norbup across 2 channels while EtG/EtS runs on one allowed 23 Bup/Norbup and 12 EtG/EtS injections/hour. Using 2 channels for each does not increase throughputs but ensures completion of all batches even if one channel stops because of leakage or over-pressurization. Multi-channeling two batches is illustrated in Figure 5.

FIGURE 1. Transcend Bup/norbup LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

Comparison between Transcend and Modified Prelude Systems

Batches of calibrators, QCs and at least 20 specimens were prepared and shared between the two multi-channel systems. Typical calibration plots from the modified Prelude - Endura system are shown in Figure 5. Desired quantitation ranges were consistently linear (r2 > 0.99 with 1/X weighting) whether when calibrators were injected into one channel or across all channels.

Only results that were within the analytical ranges for each method were used for comparisons. Differences in calculated amounts averaged less than +/- 5% within a maximum of +/- 15%. Scatter plots in Figure 6 showed good correlations between the two systems.

FIGURE 6. Scatter plots of results - Transcend vs. modified Prelude.

FIGURE 2. Transcend EtG/EtS LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

FIGURE 3. Modified Prelude EtG/EtS LC conditions.

Start data 0.75 min Data window: 1.5 min Total run time: 4.0 min

0

50

100

150

200

250

300

350

400

0 100 200 300 400

Tran

scen

d

Modified Prelude

Bup

0

100

200

300

400

500

600

700

0 100 200 300 400 500 600 700

Tran

scen

d

Modified Prelude

Norbup

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

0 2000 4000 6000 8000 10000

Tran

scen

d

Modified Prelude

EtG

0.00

1000.00

2000.00

3000.00

4000.00

5000.00

6000.00

0 1000 2000 3000 4000 5000 6000 7000

Tran

scen

d

Modified Prelude

EtS

For forensic toxicology use only.

©2015 Thermo Fisher Scientific. All Rights Reserved.

Cerilliant is a registered trademark of the Sigma-Aldrich Corporation. Biochemical Diagnostics is a registered trademark of Biochemical Diagnostics, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.

This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.

Page 3: Multi-channeling LC-MS/MS Forensic Methods for High …tools.thermofisher.com/content/sfs/posters/PN-64370-LC... · 2016. 2. 3. · for EtG/EtS were diluted 1:10 with water and then

3Thermo Scientific Poster Note • MASCL • PN64370-EN 0315S

Conclusions The modified Prelude - Endura system, which utilized unique positive-

displacement pumps, produced results for Bup/Norbup and EtG/EtS methods that were consistent with those produced by the conventional Transcend - Endura system.

The maximum throughput from both 4-channel systems was 34 injections per hour when multiplexing the two forensic methods across all four channels.

The modified Prelude system reduced solvent consumption by at least 60%. Other benefits of using this system were:

Avoided pulsations of reciprocating pumps

Easier to use, purge, prepare and maintain

Smaller footprint compared to Transcend LX4

Overview Purpose: Compare the performance of running LC-MS/MS forensic methods used to detect buprenorphine and ethanol use on two different four-channel UHPLC systems - one utilizing conventional reciprocating pumps and the other using positive-displacement pumps.

Methods: Reversed-phase liquid chromatography of analytes with corresponding stable-isotope internal standards eluting from up to four UHPLC channels into a heated electro-spray ionization (HESI) source of a triple- quadrupole mass spectrometer were used to measure urine levels of the forensic compounds Buprenorphine (Bup) & Norbuprenorphine (Norbup) after hydrolysis and Ethyl-Glucuronide (EtG) and Ethyl Sulfate (EtS) after dilution.

Results: Running Bup/Norbup batches on two channels while EtG/EtS batches ran on the remaining two channels yielded a throughput of 17 injections per hour for each batch, which totaled 34 injections per hour. The four-channel system with positive-displacement pumps reduced solvent consumption by at least 60%. Results from this system were within +/- 15% of those determined on a conventional multichannel system using reciprocating pumps.

Introduction Many forensic laboratories run several different LC-MS methods in series on a single-channel LC-MS system. If the methods involve different ion sources, columns and mobile phases, the changeover is time consuming, labor intensive and increases the risk of mistakes and contamination. A four-channel UHPLC system multiplexed into one mass spectrometer permits parallel batches of up to four different methods utilizing a common ion source and unique columns and mobile phases to be completed in a fraction of the time and effort. However, conventional systems utilizing reciprocating pumps needlessly consume mobile-phase solvents between injections. The Thermo Scientific™ Prelude™ SPLC system, which utilizes positive-displacement pumps does not waste solvents between injections. A modified version of this multi-channel UHPLC system was tested by running batches for two forensic methods - Bup/Norbup and EtG/EtS - and comparing those results to those from a conventional UHPLC system that ran the same sets of calibrators, QCs and specimens.

Mass Spectrometry System Control & Data Analysis

Thermo Scientific™ TraceFinder™ with Aria™ MX software was used to control the Transcend II LX4 and Thermo Scientific™ TSQ Endura™ MS/MS systems, submit batches to desired channels as well as for analyzing data and reporting results.

The Endura triple-quadrupole mass spectrometer was used with HESI. Ion source and MS/MS conditions are described in Figure 4.

FIGURE 4. Mass spectrometry conditions.

Common source

conditions:

Bup/Norbup

acquisition

method:

EtG/EtS

acquisition

method:

FIGURE 5. Multi-channeling Bup/Norbup and EtG/EtS batches.

FIGURE 5. Calibration Plots from Modified Prelude - Endura system.

Multi-channeling LC-MS/MS Forensic Methods for High-Throughput Urine Screening to Detect Buprenorphine and Ethanol Use Joseph Di Bussolo1, Catherine Lafontaine1, Mercedes Castillo2, Thomaskutty Thomas2 & Hashim Othman2. 1Thermo Fisher Scientific, West Chester, PA, 2BioReference Laboratories, Elmwood Park, NJ

Methods Consumables

HPLC-grade solvents, reagents and most other consumables were from Thermo Fisher Scientific. Buprenorphine and norbuprenorphine and their corresponding deuterated internal standards as well as ethyl sulfate and ethyl glucuronide and their corresponding -deuterated internal standards were from Cerilliant (Round Rock, TX). Calibrators were made by mixing these standards with synthetic urine. Liquid urine controls were from Biochemical Diagnostics (Edgewood, NY). β-glucuronidase powder was purchased from Sigma-Aldrich.

Sample Preparation:

Urine specimens and corresponding calibrators and QCs to be analyzed for Bup/Norbup were hydrolyzed by incubating a mixture of 150 µL of β-glucuronidase solution (10,000 U/mL, pH 5) with 200 µL of specimen, and 50 µL of IS solution containing - buprenorphine-D3 & norbuprenorphine-D4 for 1.5 hours at 60°C. Each preparation was then mixed with 200 µL of cold methanol and refrigerated for 10 minutes before centrifugation. 20 µL injections of supernatants from each preparation were made into the UHPLC systems.

Urine specimens and corresponding calibrators and QCs to be analyzed for EtG/EtS were diluted 1:10 with water and then spiked with 50 µL of IS solution containing EtG-D5 & EtS-D5 before making 20 µL injections into the UHPLC system.

Liquid Chromatography

The “conventional” multi-channel LC system was a Thermo Scientific™ Transcend™ LX4 equipped with binary-solvent pumps and a dual-arm autosampler configuration. Thermo Scientific™ Accucore™ RP-MS, 2.6 µm, 50 x 2.1 mm HPLC columns were used for the Bup/Norbup method. Mobile phase conditions for this method are described in Figure 1. Thermo Scientific™ Synchronis™ aQ, 5 µm, 50 x 3.0 mm HPLC columns were used for the EtG/EtS method. Mobile phase conditions for this method are described in Figure 2.

A Thermo Scientific™ Prelude™ SPLC system was modified to permit injections across four channels, similar to the Transcend LX4 system. The Bup/Norbup method on this system utilized the same column and mobile phase conditions as the Transcend system. However, the EtG/EtS method was modified to use the Thermo Scientific™ Synchronis™ aQ, 3 µm, 100 x 3.0 mm HPLC columns, as shown in Figure 3.

Results Throughput and Multi-channeling Bup/Norbup and EtG/EtS batches

Multi-channeling Bup/Norbup across 2 channels while EtG/EtS runs on one allowed 23 Bup/Norbup and 12 EtG/EtS injections/hour. Using 2 channels for each does not increase throughputs but ensures completion of all batches even if one channel stops because of leakage or over-pressurization. Multi-channeling two batches is illustrated in Figure 5.

FIGURE 1. Transcend Bup/norbup LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

Comparison between Transcend and Modified Prelude Systems

Batches of calibrators, QCs and at least 20 specimens were prepared and shared between the two multi-channel systems. Typical calibration plots from the modified Prelude - Endura system are shown in Figure 5. Desired quantitation ranges were consistently linear (r2 > 0.99 with 1/X weighting) whether when calibrators were injected into one channel or across all channels.

Only results that were within the analytical ranges for each method were used for comparisons. Differences in calculated amounts averaged less than +/- 5% within a maximum of +/- 15%. Scatter plots in Figure 6 showed good correlations between the two systems.

FIGURE 6. Scatter plots of results - Transcend vs. modified Prelude.

FIGURE 2. Transcend EtG/EtS LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

FIGURE 3. Modified Prelude EtG/EtS LC conditions.

Start data 0.75 min Data window: 1.5 min Total run time: 4.0 min

0

50

100

150

200

250

300

350

400

0 100 200 300 400

Tran

scen

d

Modified Prelude

Bup

0

100

200

300

400

500

600

700

0 100 200 300 400 500 600 700

Tran

scen

d

Modified Prelude

Norbup

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

0 2000 4000 6000 8000 10000

Tran

scen

d

Modified Prelude

EtG

0.00

1000.00

2000.00

3000.00

4000.00

5000.00

6000.00

0 1000 2000 3000 4000 5000 6000 7000

Tran

scen

d

Modified Prelude

EtS

For forensic toxicology use only.

©2015 Thermo Fisher Scientific. All Rights Reserved.

Cerilliant is a registered trademark of the Sigma-Aldrich Corporation. Biochemical Diagnostics is a registered trademark of Biochemical Diagnostics, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.

This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.

Conclusions The modified Prelude - Endura system, which utilized unique positive-

displacement pumps, produced results for Bup/Norbup and EtG/EtS methods that were consistent with those produced by the conventional Transcend - Endura system.

The maximum throughput from both 4-channel systems was 34 injections per hour when multiplexing the two forensic methods across all four channels.

The modified Prelude system reduced solvent consumption by at least 60%. Other benefits of using this system were:

Avoided pulsations of reciprocating pumps

Easier to use, purge, prepare and maintain

Smaller footprint compared to Transcend LX4

Overview Purpose: Compare the performance of running LC-MS/MS forensic methods used to detect buprenorphine and ethanol use on two different four-channel UHPLC systems - one utilizing conventional reciprocating pumps and the other using positive-displacement pumps.

Methods: Reversed-phase liquid chromatography of analytes with corresponding stable-isotope internal standards eluting from up to four UHPLC channels into a heated electro-spray ionization (HESI) source of a triple- quadrupole mass spectrometer were used to measure urine levels of the forensic compounds Buprenorphine (Bup) & Norbuprenorphine (Norbup) after hydrolysis and Ethyl-Glucuronide (EtG) and Ethyl Sulfate (EtS) after dilution.

Results: Running Bup/Norbup batches on two channels while EtG/EtS batches ran on the remaining two channels yielded a throughput of 17 injections per hour for each batch, which totaled 34 injections per hour. The four-channel system with positive-displacement pumps reduced solvent consumption by at least 60%. Results from this system were within +/- 15% of those determined on a conventional multichannel system using reciprocating pumps.

Introduction Many forensic laboratories run several different LC-MS methods in series on a single-channel LC-MS system. If the methods involve different ion sources, columns and mobile phases, the changeover is time consuming, labor intensive and increases the risk of mistakes and contamination. A four-channel UHPLC system multiplexed into one mass spectrometer permits parallel batches of up to four different methods utilizing a common ion source and unique columns and mobile phases to be completed in a fraction of the time and effort. However, conventional systems utilizing reciprocating pumps needlessly consume mobile-phase solvents between injections. The Thermo Scientific™ Prelude™ SPLC system, which utilizes positive-displacement pumps does not waste solvents between injections. A modified version of this multi-channel UHPLC system was tested by running batches for two forensic methods - Bup/Norbup and EtG/EtS - and comparing those results to those from a conventional UHPLC system that ran the same sets of calibrators, QCs and specimens.

Mass Spectrometry System Control & Data Analysis

Thermo Scientific™ TraceFinder™ with Aria™ MX software was used to control the Transcend II LX4 and Thermo Scientific™ TSQ Endura™ MS/MS systems, submit batches to desired channels as well as for analyzing data and reporting results.

The Endura triple-quadrupole mass spectrometer was used with HESI. Ion source and MS/MS conditions are described in Figure 4.

FIGURE 4. Mass spectrometry conditions.

Common source

conditions:

Bup/Norbup

acquisition

method:

EtG/EtS

acquisition

method:

FIGURE 5. Multi-channeling Bup/Norbup and EtG/EtS batches.

FIGURE 5. Calibration Plots from Modified Prelude - Endura system.

Multi-channeling LC-MS/MS Forensic Methods for High-Throughput Urine Screening to Detect Buprenorphine and Ethanol Use Joseph Di Bussolo1, Catherine Lafontaine1, Mercedes Castillo2, Thomaskutty Thomas2 & Hashim Othman2. 1Thermo Fisher Scientific, West Chester, PA, 2BioReference Laboratories, Elmwood Park, NJ

Methods Consumables

HPLC-grade solvents, reagents and most other consumables were from Thermo Fisher Scientific. Buprenorphine and norbuprenorphine and their corresponding deuterated internal standards as well as ethyl sulfate and ethyl glucuronide and their corresponding -deuterated internal standards were from Cerilliant (Round Rock, TX). Calibrators were made by mixing these standards with synthetic urine. Liquid urine controls were from Biochemical Diagnostics (Edgewood, NY). β-glucuronidase powder was purchased from Sigma-Aldrich.

Sample Preparation:

Urine specimens and corresponding calibrators and QCs to be analyzed for Bup/Norbup were hydrolyzed by incubating a mixture of 150 µL of β-glucuronidase solution (10,000 U/mL, pH 5) with 200 µL of specimen, and 50 µL of IS solution containing - buprenorphine-D3 & norbuprenorphine-D4 for 1.5 hours at 60°C. Each preparation was then mixed with 200 µL of cold methanol and refrigerated for 10 minutes before centrifugation. 20 µL injections of supernatants from each preparation were made into the UHPLC systems.

Urine specimens and corresponding calibrators and QCs to be analyzed for EtG/EtS were diluted 1:10 with water and then spiked with 50 µL of IS solution containing EtG-D5 & EtS-D5 before making 20 µL injections into the UHPLC system.

Liquid Chromatography

The “conventional” multi-channel LC system was a Thermo Scientific™ Transcend™ LX4 equipped with binary-solvent pumps and a dual-arm autosampler configuration. Thermo Scientific™ Accucore™ RP-MS, 2.6 µm, 50 x 2.1 mm HPLC columns were used for the Bup/Norbup method. Mobile phase conditions for this method are described in Figure 1. Thermo Scientific™ Synchronis™ aQ, 5 µm, 50 x 3.0 mm HPLC columns were used for the EtG/EtS method. Mobile phase conditions for this method are described in Figure 2.

A Thermo Scientific™ Prelude™ SPLC system was modified to permit injections across four channels, similar to the Transcend LX4 system. The Bup/Norbup method on this system utilized the same column and mobile phase conditions as the Transcend system. However, the EtG/EtS method was modified to use the Thermo Scientific™ Synchronis™ aQ, 3 µm, 100 x 3.0 mm HPLC columns, as shown in Figure 3.

Results Throughput and Multi-channeling Bup/Norbup and EtG/EtS batches

Multi-channeling Bup/Norbup across 2 channels while EtG/EtS runs on one allowed 23 Bup/Norbup and 12 EtG/EtS injections/hour. Using 2 channels for each does not increase throughputs but ensures completion of all batches even if one channel stops because of leakage or over-pressurization. Multi-channeling two batches is illustrated in Figure 5.

FIGURE 1. Transcend Bup/norbup LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

Comparison between Transcend and Modified Prelude Systems

Batches of calibrators, QCs and at least 20 specimens were prepared and shared between the two multi-channel systems. Typical calibration plots from the modified Prelude - Endura system are shown in Figure 5. Desired quantitation ranges were consistently linear (r2 > 0.99 with 1/X weighting) whether when calibrators were injected into one channel or across all channels.

Only results that were within the analytical ranges for each method were used for comparisons. Differences in calculated amounts averaged less than +/- 5% within a maximum of +/- 15%. Scatter plots in Figure 6 showed good correlations between the two systems.

FIGURE 6. Scatter plots of results - Transcend vs. modified Prelude.

FIGURE 2. Transcend EtG/EtS LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

FIGURE 3. Modified Prelude EtG/EtS LC conditions.

Start data 0.75 min Data window: 1.5 min Total run time: 4.0 min

0

50

100

150

200

250

300

350

400

0 100 200 300 400

Tran

scen

d

Modified Prelude

Bup

0

100

200

300

400

500

600

700

0 100 200 300 400 500 600 700

Tran

scen

d

Modified Prelude

Norbup

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

0 2000 4000 6000 8000 10000

Tran

scen

d

Modified Prelude

EtG

0.00

1000.00

2000.00

3000.00

4000.00

5000.00

6000.00

0 1000 2000 3000 4000 5000 6000 7000

Tran

scen

d

Modified Prelude

EtS

For forensic toxicology use only.

©2015 Thermo Fisher Scientific. All Rights Reserved.

Cerilliant is a registered trademark of the Sigma-Aldrich Corporation. Biochemical Diagnostics is a registered trademark of Biochemical Diagnostics, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries.

This information is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others.

Page 4: Multi-channeling LC-MS/MS Forensic Methods for High …tools.thermofisher.com/content/sfs/posters/PN-64370-LC... · 2016. 2. 3. · for EtG/EtS were diluted 1:10 with water and then

PN64370-EN 0315S

Africa +43 1 333 50 34 0Australia +61 3 9757 4300Austria +43 810 282 206Belgium +32 53 73 42 41Brazil +55 11 3731 5140Canada +1 800 530 8447China 800 810 5118 (free call domestic)

400 650 5118

Denmark +45 70 23 62 60Europe-Other +43 1 333 50 34 0Finland +358 9 3291 0200France +33 1 60 92 48 00Germany +49 6103 408 1014India +91 22 6742 9494Italy +39 02 950 591

Japan +81 6 6885 1213Korea +82 2 3420 8600Latin America +1 561 688 8700Middle East +43 1 333 50 34 0Netherlands +31 76 579 55 55 New Zealand +64 9 980 6700 Norway +46 8 556 468 00

Russia/CIS +43 1 333 50 34 0Singapore +65 6289 1190Sweden +46 8 556 468 00 Switzerland +41 61 716 77 00Taiwan +886 2 8751 6655UK/Ireland +44 1442 233555USA +1 800 532 4752

www.thermoscientific.com©2015 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. Cerilliant is a registered trademark of the Sigma-Aldrich Corporation. UTAK is a registered trademark of UTAK Laboratories, Inc. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details.

Thermo Fisher Scientific, San Jose, CA USA is ISO 9001 Certified.

For forensic toxicology use only.

Conclusions The modified Prelude - Endura system, which utilized unique positive-

displacement pumps, produced results for Bup/Norbup and EtG/EtS methods that were consistent with those produced by the conventional Transcend - Endura system.

The maximum throughput from both 4-channel systems was 34 injections per hour when multiplexing the two forensic methods across all four channels.

The modified Prelude system reduced solvent consumption by at least 60%. Other benefits of using this system were:

Avoided pulsations of reciprocating pumps

Easier to use, purge, prepare and maintain

Smaller footprint compared to Transcend LX4

Overview Purpose: Compare the performance of running LC-MS/MS forensic methods used to detect buprenorphine and ethanol use on two different four-channel UHPLC systems - one utilizing conventional reciprocating pumps and the other using positive-displacement pumps.

Methods: Reversed-phase liquid chromatography of analytes with corresponding stable-isotope internal standards eluting from up to four UHPLC channels into a heated electro-spray ionization (HESI) source of a triple- quadrupole mass spectrometer were used to measure urine levels of the forensic compounds Buprenorphine (Bup) & Norbuprenorphine (Norbup) after hydrolysis and Ethyl-Glucuronide (EtG) and Ethyl Sulfate (EtS) after dilution.

Results: Running Bup/Norbup batches on two channels while EtG/EtS batches ran on the remaining two channels yielded a throughput of 17 injections per hour for each batch, which totaled 34 injections per hour. The four-channel system with positive-displacement pumps reduced solvent consumption by at least 60%. Results from this system were within +/- 15% of those determined on a conventional multichannel system using reciprocating pumps.

Introduction Many forensic laboratories run several different LC-MS methods in series on a single-channel LC-MS system. If the methods involve different ion sources, columns and mobile phases, the changeover is time consuming, labor intensive and increases the risk of mistakes and contamination. A four-channel UHPLC system multiplexed into one mass spectrometer permits parallel batches of up to four different methods utilizing a common ion source and unique columns and mobile phases to be completed in a fraction of the time and effort. However, conventional systems utilizing reciprocating pumps needlessly consume mobile-phase solvents between injections. The Thermo Scientific™ Prelude™ SPLC system, which utilizes positive-displacement pumps does not waste solvents between injections. A modified version of this multi-channel UHPLC system was tested by running batches for two forensic methods - Bup/Norbup and EtG/EtS - and comparing those results to those from a conventional UHPLC system that ran the same sets of calibrators, QCs and specimens.

Mass Spectrometry System Control & Data Analysis

Thermo Scientific™ TraceFinder™ with Aria™ MX software was used to control the Transcend II LX4 and Thermo Scientific™ TSQ Endura™ MS/MS systems, submit batches to desired channels as well as for analyzing data and reporting results.

The Endura triple-quadrupole mass spectrometer was used with HESI. Ion source and MS/MS conditions are described in Figure 4.

FIGURE 4. Mass spectrometry conditions.

Common source

conditions:

Bup/Norbup

acquisition

method:

EtG/EtS

acquisition

method:

FIGURE 5. Multi-channeling Bup/Norbup and EtG/EtS batches.

FIGURE 5. Calibration Plots from Modified Prelude - Endura system.

Multi-channeling LC-MS/MS Forensic Methods for High-Throughput Urine Screening to Detect Buprenorphine and Ethanol Use Joseph Di Bussolo1, Catherine Lafontaine1, Mercedes Castillo2, Thomaskutty Thomas2 & Hashim Othman2. 1Thermo Fisher Scientific, West Chester, PA, 2BioReference Laboratories, Elmwood Park, NJ

Methods Consumables

HPLC-grade solvents, reagents and most other consumables were from Thermo Fisher Scientific. Buprenorphine and norbuprenorphine and their corresponding deuterated internal standards as well as ethyl sulfate and ethyl glucuronide and their corresponding -deuterated internal standards were from Cerilliant (Round Rock, TX). Calibrators were made by mixing these standards with synthetic urine. Liquid urine controls were from Biochemical Diagnostics (Edgewood, NY). β-glucuronidase powder was purchased from Sigma-Aldrich.

Sample Preparation:

Urine specimens and corresponding calibrators and QCs to be analyzed for Bup/Norbup were hydrolyzed by incubating a mixture of 150 µL of β-glucuronidase solution (10,000 U/mL, pH 5) with 200 µL of specimen, and 50 µL of IS solution containing - buprenorphine-D3 & norbuprenorphine-D4 for 1.5 hours at 60°C. Each preparation was then mixed with 200 µL of cold methanol and refrigerated for 10 minutes before centrifugation. 20 µL injections of supernatants from each preparation were made into the UHPLC systems.

Urine specimens and corresponding calibrators and QCs to be analyzed for EtG/EtS were diluted 1:10 with water and then spiked with 50 µL of IS solution containing EtG-D5 & EtS-D5 before making 20 µL injections into the UHPLC system.

Liquid Chromatography

The “conventional” multi-channel LC system was a Thermo Scientific™ Transcend™ LX4 equipped with binary-solvent pumps and a dual-arm autosampler configuration. Thermo Scientific™ Accucore™ RP-MS, 2.6 µm, 50 x 2.1 mm HPLC columns were used for the Bup/Norbup method. Mobile phase conditions for this method are described in Figure 1. Thermo Scientific™ Synchronis™ aQ, 5 µm, 50 x 3.0 mm HPLC columns were used for the EtG/EtS method. Mobile phase conditions for this method are described in Figure 2.

A Thermo Scientific™ Prelude™ SPLC system was modified to permit injections across four channels, similar to the Transcend LX4 system. The Bup/Norbup method on this system utilized the same column and mobile phase conditions as the Transcend system. However, the EtG/EtS method was modified to use the Thermo Scientific™ Synchronis™ aQ, 3 µm, 100 x 3.0 mm HPLC columns, as shown in Figure 3.

Results Throughput and Multi-channeling Bup/Norbup and EtG/EtS batches

Multi-channeling Bup/Norbup across 2 channels while EtG/EtS runs on one allowed 23 Bup/Norbup and 12 EtG/EtS injections/hour. Using 2 channels for each does not increase throughputs but ensures completion of all batches even if one channel stops because of leakage or over-pressurization. Multi-channeling two batches is illustrated in Figure 5.

FIGURE 1. Transcend Bup/norbup LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

Comparison between Transcend and Modified Prelude Systems

Batches of calibrators, QCs and at least 20 specimens were prepared and shared between the two multi-channel systems. Typical calibration plots from the modified Prelude - Endura system are shown in Figure 5. Desired quantitation ranges were consistently linear (r2 > 0.99 with 1/X weighting) whether when calibrators were injected into one channel or across all channels.

Only results that were within the analytical ranges for each method were used for comparisons. Differences in calculated amounts averaged less than +/- 5% within a maximum of +/- 15%. Scatter plots in Figure 6 showed good correlations between the two systems.

FIGURE 6. Scatter plots of results - Transcend vs. modified Prelude.

FIGURE 2. Transcend EtG/EtS LC conditions.

Start data 0.5 min Data window: 1.5 min Total run time: 4.0 min

FIGURE 3. Modified Prelude EtG/EtS LC conditions.

Start data 0.75 min Data window: 1.5 min Total run time: 4.0 min

0

50

100

150

200

250

300

350

400

0 100 200 300 400

Tran

scen

d

Modified Prelude

Bup

0

100

200

300

400

500

600

700

0 100 200 300 400 500 600 700

Tran

scen

d

Modified Prelude

Norbup

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

0 2000 4000 6000 8000 10000

Tran

scen

d

Modified Prelude

EtG

0.00

1000.00

2000.00

3000.00

4000.00

5000.00

6000.00

0 1000 2000 3000 4000 5000 6000 7000

Tran

scen

d

Modified Prelude

EtS

For forensic toxicology use only.