MRI Annual Physics Tests - Beyond Just the Scans & Measurements July 21, 2005 Carl R. Keener, Ph.D. 1 MR Accreditation Program Quality Control Beyond Just the Scans and Measurements July 2005 Carl R. Keener, Ph.D., DABMP, DABR [email protected] MAR P & Medical Radiation Physics , Inc. Department of Radiology University of Texas Health Science Center at San Antonio © 2005 ACR MRI accreditation program u phantom tests u clinical images u personnel u qualifications u responsibilities u QC u daily weekly u acceptance testing u annual surveys ACR MRAP medical physicist / MR scientist* support u medical physicist / MR scientist* participation in: u initial phantom tests u technologist daily/weekly QC program u acceptance testing u annual MR testing u full-time MRI medical physicist/MR scientist * u on staff u consultant u medical physicist with some MRI responsibilities u on staff u consultant * in this talk , "medical physicist" will be used instead of "qualified medical physicist/MR scientist" MR physics tests u magnetic field homogeneity u slice position accuracy u slice thickness accuracy u RF coil stability u SNR u ghosting u uniformity u inter-slice RF interference u artifacts u softcopy displays u phantom tests u geometric accuracy u high-contrast spatial resolution u slice thickness accuracy u slice position accuracy u image intensity uniformity u percent signal ghosting u low-contrast object detectability u image artifacts MR physics tests - resources u what can the medical physicist offer to the MRI site? MR physics tests strategies u communication u have there been changes?….problems? u comparison u baselines from previous surveys not just pass/fail u measurements from similar scanners u manufacturer's specs u QC values u combine test results u creativity u additional tests and measurements

Upload: others

Post on 01-Mar-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: MRI Annual Physics Tests - Beyond the Ba · 2005. 7. 21. · Title: MRI Annual Physics Tests - Beyond the Ba Author: Carl Created Date: 7/21/2005 10:09:15 AM

MRI Annual Physics Tests - Beyond Just the Scans & Measurements July 21, 2005

Carl R. Keener, Ph.D. 1

MR Accreditation Program Quality ControlBeyond Just the Scans and Measurements

July 2005

Carl R. Keener, Ph.D., DABMP, [email protected]

MARP &Medical Radiation Physics, Inc.

Department of RadiologyUniversity of Texas Health Science Center at San Antonio

© 2005

ACR MRI accreditation program

u phantom testsu clinical imagesu personnel

u qualificationsu responsibilities

u QCu daily weekly

u acceptance testingu annual surveys

ACR MRAP medical physicist / MR scientist* support

u medical physicist / MRscientist* participationin:

u initial phantom testsu technologist

daily/weekly QCprogram

u acceptance testingu annual MR testing

u full-time MRI medicalphysicist/MR scientist *

u on staff

u consultant

u medical physicist withsome MRIresponsibilities

u on staff

u consultant

* in this talk , "medical physicist" will be used instead of "qualified medicalphysicist/MR scientist"

MR physics tests

u magnetic fieldhomogeneity

u slice position accuracyu slice thickness

accuracyu RF coil stability

u SNR

u ghostingu uniformity

u inter-slice RFinterference

u artifactsu softcopy displays

u phantom testsu geometric accuracy

u high-contrast spatialresolution

u slice thicknessaccuracy

u slice position accuracy

u image intensityuniformity

u percent signal ghosting

u low-contrast objectdetectability

u image artifacts

MR physics tests - resources

u what can the medicalphysicist offer to theMRI site?

MR physics tests strategies

u communicationu have there been changes?….problems?

u comparisonu baselines from previous surveys

• not just pass/fail

u measurements from similar scannersu manufacturer's specsu QC values

u combine test resultsu creativity

u additional tests and measurements

Page 2: MRI Annual Physics Tests - Beyond the Ba · 2005. 7. 21. · Title: MRI Annual Physics Tests - Beyond the Ba Author: Carl Created Date: 7/21/2005 10:09:15 AM

MRI Annual Physics Tests - Beyond Just the Scans & Measurements July 21, 2005

Carl R. Keener, Ph.D. 2

MR annual testing tools

u ACR phantomu site available phantoms

u manufacturer-specific

u specialty phantomsu custom made

u non MR-specific tools

MR annual testing – outline

u homogeneityu center frequency / QCu ACR phantom testsu coil testsu non-phantom tests

homogeneity

u concernsu geometric distortion?

u problems with fatsaturation?

u methodsu spectral

u phase-difference

u visual distortion

homogeneity – spectral analysis

u measures frequency ofsignal-producingphantom in magnet

u DSV• ↑↑ homogeneity at center

& with ↓↓ FOV• ? FOV… ? phantom size

u spectrum displayed onscreen (Hz)

u homogeneitymeasured in ppm

u compare to specs orbaseline values

homogeneity – spectral analysis

u manual prescanu GE

u EPI tuningu Elscint

u FWHM (Hz)

u volume ~ phantom

0.3526DSV20030.3826DSV2004

ppmcmorientationdate

ppmHz

MHzFWHM

LarmorFrequency

=

magnetic field homogeneity – phase difference

u 2 FE (GRE) scansu slightly different TEs

u subtracted phase imagesshow phase differences

• inhomogeneities

u change (ppm) depends onchange in TE in protocol

• specified in manufacturer'sdocumentation

u allows you to viewinhomogenieties

u multiple slices

u three planes

Page 3: MRI Annual Physics Tests - Beyond the Ba · 2005. 7. 21. · Title: MRI Annual Physics Tests - Beyond the Ba Author: Carl Created Date: 7/21/2005 10:09:15 AM

MRI Annual Physics Tests - Beyond Just the Scans & Measurements July 21, 2005

Carl R. Keener, Ph.D. 3

phase-difference (Siemens)

u fast-field mappingu selectable resolution

based on TE

u B →→W →→B = 1/TE/B0

B→→ W →→B = 0.039ppmB→→ W →→B = 0.39ppm

fast field mapping (Siemens)

<0.210transverse

<0.210sagittal

<0.210coronal

1.1724sagittal

0.3924transverse

0.3924coronal

ppmcmorientation

phase-difference (Philips)

u shim_checku option under phantom

studies

u FFE, fixed technique &FOV

u 40 cm cylindricalphantom

u holder to centerphantom

u 3 orientations

u view real imagesu count B →→W transitions

shim_check (Philips)

u example:u 1.5T, FFE, TR=400, TE=16,

30º, 450 mm FOV

u B →→W = 1 ppm

u example:u coin taped to phantom

shim_check (Philips)

ppm

2

2

1

12

6

8

5/6/03

1.020transverse

0.2520sagittal

0.2520coronal

3.540sagittal

640transverse

540coronal

5/11/04cmorientation

spectral – three planes

u Picker / Marconi Eclipseu "Shim and CF Tune"

u graph of spectrum

u 30 cm x 10 cm cylinderu holder centers phantom

in 3 orientations

Page 4: MRI Annual Physics Tests - Beyond the Ba · 2005. 7. 21. · Title: MRI Annual Physics Tests - Beyond the Ba Author: Carl Created Date: 7/21/2005 10:09:15 AM

MRI Annual Physics Tests - Beyond Just the Scans & Measurements July 21, 2005

Carl R. Keener, Ph.D. 4

shim & CF tune (Picker /Marconi)

u sagittal u transverse

shim & CF tune (Picker /Marconi)

u ↓↓ homogeneity intransverse plane

u poor fat saturationu ½ bright / ½ dark

u metal objects removedfrom bore

u ↑↑ homogeneityu ↑↑ fat sat

4/26/04

4/26/04

4/26/04

3/29/04

3/29/04

3/29/04

date

1.2730transverse

0.2430sagittal

0.3030coronal

0.3130sagittal

2.730transverse

0.3130coronal

ppmcmorientation

homogeneity – geometric distortion

u most low-field open magnetsdo not have software to checkhomogeneity

u observe effects on geometricdistortion

• lower bandwidth

u warped images

7.4 kHz 3.6 kHz

center frequency – high field

u Action limits: 63,854,754 +/- 100 Hzu horizontal gridlines = 100 Hz

Center Frequency1.5 T

63854554

63854654

63854754

63854854

63854954

5/1/03 7/1/03 9/1/03 11/1/03 1/2/04 3/3/04 5/4/04

center frequency – low-field open

u Action limits: 15000700 +/- 500 Hzu horizontal gridlines = 500 Hz

Center Frequency0.35T

14999700

15000200

15000700

15001200

15001700

15002200

12/1/03 12/31/03 1/31/04 3/1/04 4/1/04 5/1/04 6/1/04 7/1/04

center frequency – high field (high drift)

u Action limits: > 63,825,500 MHzu change/week < 100 Hz

u horizontal gridlines = 1000 Hz

Center Frequency1.5 T

63821500

63822500

63823500

63824500

63825500

63826500

63827500

63828500

63829500

8/1/03 9/1/03 10/2/03 11/2/03 12/3/03 1/3/04 2/3/04 3/5/04

Page 5: MRI Annual Physics Tests - Beyond the Ba · 2005. 7. 21. · Title: MRI Annual Physics Tests - Beyond the Ba Author: Carl Created Date: 7/21/2005 10:09:15 AM

MRI Annual Physics Tests - Beyond Just the Scans & Measurements July 21, 2005

Carl R. Keener, Ph.D. 5

ACR phantom

u testsu geometric accuracyu high-contrast spatial

resolutionu slice thickness

accuracyu slice position accuracyu image intensity

uniformityu percent signal ghostingu low-contrast object

detectabilityu image artifactsu inter-slice gap

u standard orientationu axialu head coil

ACR phantom – other orientations

u head coilu tests

u geometric accuracyu high-contrast spatial

resolutionu slice thickness

accuracyu slice position accuracyu image intensity

uniformityu percent signal ghostingu low-contrast object

detectabilityu image artifacts

coronal

sagittal

geometric accuracy

u other orientations offeradditional measurements inz plane

u standard ACR phantomorientation gives moreinformation in x & yplanes than in z plane

geometric accuracy

sequencephantom

orientationslice

locationmeasurement

orientationmeasured

(mm)actual (mm)

difference (mm)

SE coronal slice 5 top-bottom 186.2 190 3.8

SE coronal slice 5 left-right 192.2 190 -2.2

SE coronal slice 5 diagonal (/) 187.5 190 2.5

SE coronal slice 5 diagonal (\) 187.3 190 2.7

SE coronal slice 1 top-bottom 189.3 190 0.7

SE coronal slice 1 left-right 192.5 190 -2.5

SE coronal slice 11 top-bottom 186.5 190 3.5

SE coronal slice 11 left-right 191.7 190 -1.7SE sagittal slice 5 top-bottom 186.7 190 3.3

SE sagittal slice 5 left-right 192.7 190 -2.7

SE sagittal slice 5 diagonal (/) 188.6 190 1.4

SE sagittal slice 5 diagonal (\) 189.5 190 0.5

SE sagittal slice 1 top-bottom 188.3 190 1.7

SE sagittal slice 1 left-right 193.4 190 -3.4

SE sagittal slice 11 top-bottom 189.6 190 0.4

SE sagittal slice 11 left-right 192.7 190 -2.7

slice thickness

u slice thickness based onRF pulse & gradient

u do changes in slicethickness correlate withgradient nonlinearities(based on geometricmeasurements)?

zGz

⋅∆

=∆γ

ω

sequence orientationtop length

(mm)

bottom

length (mm)

measured

thickness (mm)

actual

thickness (mm)

difference (mm)

% difference

ACR T1 axial 57.7 58.3 5.8 5.0 0.8 16%ACR T1 sagittal 51.4 60.5 5.6 5.0 0.6 11%ACR T1 coronal 50.5 69.0 5.8 5.0 0.8 17%

slice positioning

u slice positioningmeasurements shouldnot change duringsurvey if phantom hasnot been moved

u changes in sliceposition may be due toa temporal magneticfield or gradient drift

Page 6: MRI Annual Physics Tests - Beyond the Ba · 2005. 7. 21. · Title: MRI Annual Physics Tests - Beyond the Ba Author: Carl Created Date: 7/21/2005 10:09:15 AM

MRI Annual Physics Tests - Beyond Just the Scans & Measurements July 21, 2005

Carl R. Keener, Ph.D. 6

rf inter-slice interference

u how close can slices be before adjacent slicesinterfere with each other?

z

SNR

u single image of uniformslice

u signal is mean of uniform areau noise is SD of area outside

phantom

u tests• interslice gap• coils

u easy to use, but…….u sensitive to techniqueu different than NEMA method

or manufacturer's specsu sensitive to fluctuations in

noise

SNRSignalNoise

=

hardware setup, phantom,sequence, pulse

)(

⋅∆∝

SNRNNBW

NEXFOVFOVzMSNRyxreceiver

yxTExy

SNR / field strength

Field Strength vs. SNR

0

100

200

300

400

0 0.5 1 1.5 2 2.5 3 3.5

Field Strength (Tesla)

SN

R

u field strength & SNR correlate closelyu 20 scanners (0.2 – 3.0T), 42 surveys

u ACR T1 series on ACR phantom

u 1 operator (physicist)

u @ 1.5T:

u mean = 251, SD = 47, n = 31

rf inter-slice interference

u measure SNR of uniform slices with different gaps

91.2%

94.6%

95.7%

97.7%

100.0%

%mean

93.3%

111.7%

101.5%

107.5%

100.0%

%SNR

0.0

0.5

1.0

2.0

5.0

gap (mm)

2444.931202.58

2924.271247.20

2814.581288.08

2664.751261.38

2625.041318.23

SNRSDmean

Interslice Gap (Signal)

70%

80%

90%

100%

110%

120%

0.0 1.0 2.0 3.0 4.0 5.0gap (mm)

% S

igna

l

Interslice Gap (SNR)

70%

80%

90%

100%

110%

120%

0.0 1.0 2.0 3.0 4.0 5.0

gap (mm)

% S

NR

low-contrast detectability

u slices 8-11 on ACRphantom

u 4 series / 4 slices forphantom evaluation

u 1 series / 1 slice for daily /weekly QC

u pass/failu can also be used to track

performanceu used in place of SNR

u correlates w/ SNR but….u subjective

u sensitive to unrelatedartifacts

low contrast detectability / SNR

u LCD & SNR correlate closelyu 17 scanners (0.2 – 3.0T)

u ACR T1 series on ACR phantomu 1 operator (physicist)

SNR vs MRAP LCD

0

10

20

30

40

0 50 100 150 200 250 300 350 400

SNR

LC

D

u @ 1.5T:

u mean = 35.4, SD = 2.7, n = 35

Page 7: MRI Annual Physics Tests - Beyond the Ba · 2005. 7. 21. · Title: MRI Annual Physics Tests - Beyond the Ba Author: Carl Created Date: 7/21/2005 10:09:15 AM

MRI Annual Physics Tests - Beyond Just the Scans & Measurements July 21, 2005

Carl R. Keener, Ph.D. 7

low contrast detectability / SNR

u LCD & SNR correlate closelyu 1 scanner, 1 surveyu tracked within survey

7828site T2

21139ACR T2

18238site T1

29739ACR T1

SNRLCDseries

low contrast detectability / SNR

u LCD & SNRu tracked within surveyu tracked between surveys

u why did LCD & SNR drop in 2002?u QC scores for slice 8 = 10,10, 10, 10, 10, 10, etc.u 2002 physicist score for slice 8 = 4

SNRLCD

80

30

68

111

2002

76

220

333

2001

GRE

815site T2

2727ACR T2

site T1

2936ACR T1

20022001series

low contrast detectability / daily QC

u SNR was 50-60% lower on phantomu SNR had not dropped on other coils

u why did LCD & SNR drop in 2002?u technologist was scoring slice 11 instead of slice 8

• slice 11 score = 10; slice 8 score = 4

slice 8 slice 9 slice 10 slice 11

1.5T

low contrast detectability / SNR

u head coil was defective

u ↑↑ numbers after coil correctedu site QC program correctedu LCD & SNR stable following corrections

80

147

223

306

2004

21

32

39

40

2004

SNRLCD

80

30

68

111

2002

78

182

211

297

2003

28

38

39

39

2003

76

220

333

2001

GRE

815site T2

2727ACR T2

site T1

2936ACR T1

20022001series

coils

u same coilu consistent from year to

year• SNR• ghosting• uniformity

u different coilsu are they comparable?

u ghostingu from coil or gradient?

coils – year to year

u is coil performingsimilar to previousyear?

u same setup,phantom,techniques

• photos 23594.3%0.11%1/30/2004

23993.9%0.08%2/1/2005

25194.9%0.02%12/20/2002

12993.6%0.58%9/27/2001

12793.9%0.02%8/7/2000

18993.6%1.03%10/21/1999

SNRuniformityghostingdate

Page 8: MRI Annual Physics Tests - Beyond the Ba · 2005. 7. 21. · Title: MRI Annual Physics Tests - Beyond the Ba Author: Carl Created Date: 7/21/2005 10:09:15 AM

MRI Annual Physics Tests - Beyond Just the Scans & Measurements July 21, 2005

Carl R. Keener, Ph.D. 8

comparing coils

u similar coils on samescanner

u same phantom• head SNR sphere

u same techniques• SE, TR=500, TE=20, 20 cm FOV,

256x128, 15.6 kHz

u measured max signal &noise

old shouldercoil

new shouldercoil

coil

771.88144

10402.312402

maxSNRnoisemax

signal

ghosting – comparing coils

"When troubleshooting the ghosting, concentrate on the head coil as itmay be the source of the problem. There is also noticeable ghosting onthe CTL spine coil. All other coils have minimal ghosting or none atall."

noticeableCTL

noneShoulder

noneSmall joint

0.04 %NVA

0.35 %Wrist

0.06 %Volume neck

0.11 %Quad knee

1.03 %Head

0.20 %Body

GhostingCoil

???

ghosting – most coils

u ghosting on most coilsu problem in scanner, not

coils

u alternate phase-encoding direction todeterminemalfunctioning gradient

u documentation

RF shielding

u image quality problemsu random lines from RF

noise

u increased image noise

u simple check withportable radio

u check doors

u check ceiling

RF shielding

u new scanner with grainy imagesu ACR T1 LCD = 30

• vs. 35.4 +/- 2.7 average

u ACR T1 SNR = 163• vs. 251 +/- 47 average

u radio signal strong throughoutroom

• slightly audible with door closed

u conclusion:u RF shielding was compromised

during install

magnetic fringe field

u 5 gauss line must beposted

u pacemakers

u actively shieldedmagnets

u 5 gauss line inside scanroom or very close

u unshielded magnetsu 5 gauss line ~ 30 feet

away• outside scan room• outside building

Page 9: MRI Annual Physics Tests - Beyond the Ba · 2005. 7. 21. · Title: MRI Annual Physics Tests - Beyond the Ba Author: Carl Created Date: 7/21/2005 10:09:15 AM

MRI Annual Physics Tests - Beyond Just the Scans & Measurements July 21, 2005

Carl R. Keener, Ph.D. 9

magnetic fringe field

u are signs appropriate?u has scanner been changed?u gauss meter

conclusions

u communicationu have there been

changes?….problems?

u comparisonu baselines from

previous surveysu measurements from

similar scannersu manufacturer's specsu QC values

u combine test resultsu creativity

u additional tests andmeasurements

u homogeneityu center frequency / QCu ACR phantom testsu coil testsu non-phantom tests

MR Accreditation Program Quality ControlBeyond Just the Scans and Measurements

July 2005

Carl R. Keener, Ph.D., DABMP, [email protected]

MARP &Medical Radiation Physics, Inc.

Department of RadiologyUniversity of Texas Health Science Center at San Antonio

© 2003