Motion-Compensated Neuroanatomical Imaging Dylan Tisdall April 2013

Upload: others

Post on 04-Jul-2020

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

Motion-CompensatedNeuroanatomical Imaging

Dylan TisdallApril 2013

Page 2: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

Motion-compensated MRI sequences allow you to image subjects even if they move, without discarding scans and rescanning.

Page 3: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

There are two basic types of motion-compensation:

Retrospective

Prospective

Post-process to estimate data that would have been measured if the subject hadn’t moved.Examples: PROPELLER, SNAILS

Track the subject and alter the acquisition “on-the-fly” to account for subject motion.Examples: PACE, vNavs, PROMO

Page 4: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

without prospective moco with prospective moco

MPRAGE of subject prompted to change position every 45 seconds

Page 5: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction
Page 6: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction
Page 7: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

• Our vNav sequences are available now onSiemens scanners (WIP 711).

• Other groups are developing similar techniques on GE scanners (e.g., PROMO).

Who should use these sequences? Everyone!

Page 8: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

Overview

• Following the subject:EPI-navigated prospective motion correction

• More motion-resistance:automatic retrospective reacqusition

• Using FreeSurfer for validation:longitudinal, cross-contrast analysis

Page 9: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

A single TR

TI gap

Inv. pulse

Readoutblock

TR gapMPRAGE &T2SPACE

FLAIR

previous TR gap Readoutblock T2SPACE

Page 10: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

A single TR

Inv. pulse

Readoutblock

TR gapMPRAGE &T2SPACE

FLAIR

Readoutblock T2SPACE

+ EPI Navigator

EPI Nav275 ms

EPI Nav275 ms

Page 11: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

A single TR

Inv. pulse

Readoutblock

TR gapMPRAGE &T2SPACE

FLAIR

Readoutblock T2SPACE

+ EPI Navigator

EPI Nav275 ms

Reg.80 ms

+ Registration and Feedback

EPI Nav275 ms

Reg.80 ms

Page 12: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

A single TR

Inv. pulse

Readoutblock

TR gapMPRAGE &T2SPACE

FLAIR

Readoutblock T2SPACE

+ EPI Navigator

EPI Nav275 ms

Reg.80 ms

+ Registration and Feedback= updated imaging coordinates

EPI Nav275 ms

Reg.80 ms

Page 13: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

The Navigator

• 323 EPI• 8 mm iso• 256 mm FOV• 25 shots• TE 5.2 ms,

TR 11 ms• ~ 275 ms

Page 14: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

A single TR

Inv. pulse

Readoutblock

TR gapMPRAGE &T2SPACE

FLAIR

Readoutblock T2SPACE

+ EPI Navigator

EPI Nav275 ms

Reg.80 ms

+ Registration and Feedback= updated imaging coordinates

EPI Nav275 ms

Reg.80 ms

Page 15: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

...

TR nTR 1

Register each EPI nav volume back to first TR using Siemens’ PACE registration algorithm.

At 3T, observed variance of 50 microns with stationary subject (a pineapple).Accuracy estimated to be better than 300 microns in real-world examples.

Page 16: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

Unsedated pediatric multi-echo MPRAGE

without moco or navs with navs and mocoImages courtesy of Ellen Grant, Children’s Hospital Boston

Page 17: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

Overview

• Following the subject:EPI-navigated prospective motion correction

• More motion-resistance:automatic retrospective reacqusition

• Using FreeSurfer for validation:longitudinal, cross-contrast analysis

Page 18: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

w/ mocow/o reacquisiton

w/ mocow/ 10 TRs reacquired

T2SPACE corrupted by 20 seconds of free motion during acquisition of center of k-space

Page 19: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

TR 2TR 1

Register each EPI nav volume back to first TR using Siemens’ PACE registration algorithm.

TRs are in alignment now

Page 20: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

TR 2TR 1

Motion in here?

Page 21: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

Users configure the number of TRs toreacquire as part of their protocol.

Page 22: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

Users configure the number of TRs toreacquire as part of their protocol.

Page 23: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction
Page 24: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

Overview

• Following the subject:EPI-navigated prospective motion correction

• More motion-resistance:automatic retrospective reacqusition

• Using FreeSurfer for validation:longitudinal, cross-contrast analysis

Page 25: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

1. “Longitudinal” analysis of same-subject, same-day, motion-free T1 scans without navigators, with navigators but without motion-correction, and with navigators and motion-correction

2. Registration of same-subject, same-day, with-motion T1 scans to a fully segmented same-subject, same-day, without-motion T1 scan.

3. Cross-contrast registration of same-subject, same-day, with- and without-motion T2 scans to a fully segmented same-subject, same day without motion T1 scan.

Three non-standard FreeSurfer uses“Longitudinal” analysis of same-subject, same-day, motion-free T1 scans without navigators, with navigators but without motion-correction, and with navigators and motion-correction.

Page 26: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

vNav-moco-

vNav+moco-

vNav+moco+

base

“Longitudinal” analysis of same-subject, same-day, motion-free T1 scans without navigators, with navigators but without motion-correction, and with navigators and motion-correction.

longitudinal stream

Page 27: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

vNav-moco-

vNav+moco-

vNav+moco+

base

“Longitudinal” analysis of same-subject, same-day, motion-free T1 scans without navigators, with navigators but without motion-correction, and with navigators and motion-correction.

seg

longitudinal stream

Page 28: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

vNav-moco-

vNav+moco-

vNav+moco+

“Longitudinal” analysis of same-subject, same-day, motion-free T1 scans without navigators, with navigators but without motion-correction, and with navigators and motion-correction.

seg

now we have voxel-wise equivalence

Page 29: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

cross-sectional stream

Registration of same-subject, same-day, with-motion T1 scans to a fully segmented same-subject, same-day, without-motion T1 scan.

motion-vNav-

motion+vNav+

motion+vNav-

seg

Page 30: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

now we have voxel-wise equivalence

Registration of same-subject, same-day, with-motion T1 scans to a fully segmented same-subject, same-day, without-motion T1 scan.

motion-vNav-

motion+vNav+

motion+vNav-

seg

Page 31: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

We can use mri_robust_register to extrapolate a segmentation to a subsequent acquisition.

Page 32: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

cross-sectional stream

Cross-contrast registration of same-subject, same-day, with- and without-motion T2 scans to a fully segmented same-subject, same day without motion T1 scan.

motion-T1

motion+vNav+

motion+vNav-

seg

Page 33: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

now we have voxel-wise equivalence

Cross-contrast registration of same-subject, same-day, with- and without-motion T2 scans to a fully segmented same-subject, same day without motion T1 scan.

motion-T1

motion+vNav+

motion+vNav-

seg

Page 34: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

We can use bbregister to extrapolate a segmentation to a subsequent acquisition with a different contrast.

Page 35: Motion-Compensated Neuroanatomical Imaging · At 3T, observed variance of 50 microns with stationary subject (a pineapple). Accuracy estimated to be better than 300 microns ... FreeSurfer_Course_Apr_2013_Motion_Correction

• NIH grants R21EB008547, R21AA017410, R21DA026104, R01NS055754, P41RR014075

• The Ellison Medical Foundation

Acknowledgements:

• Aaron Hess• André van der Kouwe• Martin Reuter• Himanshu Bhat• Oliver Hinds• Thomas Benner

• Michael Hamm