monoclonal antibody characterization on q...

31
Monoclonal Antibody Characterization on Q Exactive and Oribtrap Elite Yi Zhang, Ph.D Senior Proteomic Marketing Specialist Oct. 12, 2011

Upload: others

Post on 31-Dec-2019

8 views

Category:

Documents


0 download

TRANSCRIPT

Monoclonal Antibody Characterization on Q Exactive and Oribtrap Elite

Yi Zhang, Ph.D Senior Proteomic Marketing Specialist Oct. 12, 2011

2

Outline

• Orbitrap Mass Spectrometer in mAb Characterization

• Intact Protein Analysis on Q Exactive Using Protein Deconvolution

• Top-Down Analysis of mAb Light Chain on Orbitrap Elite

• Conclusion

3

Mass Spectrometry in mAb Characterization

• Monoclonal antibodies have become a major drug class. • Highly specific • Fewer side effects

• Quality control is challenging

• Big molecules (150kD) More things can go wrong.

• The role of mass spectrometry • Mass measurement of intact mAb • Sequence confirmation • Disulfide mapping • PTM identification • Glycan analysis • ……

4

Challenges of Protein Detection in Orbitrap Analyzer

0 1000 2000 3000 t, ms

Myoglobin (16.9 kD) z: +16

0 1000 2000 3000 t, ms

0 1000 2000 3000 t, ms 0

Transferrin (78kD) z: +40

Enolase (46kD) z: +24

Solutions: • Higher Vacuum • Hardware Innovation • Software Innovation

Fast Signal Decay: • Collision of ions • Fragmentation of ions • De-phasing of ion packets

5

Hardware Inovations in New Orbitrap Analyzer

0 1 2 3 4 5 6 7 8 9 10m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rela

tive I

nten

sity

6.56290.3585

6.4464

6.40610.5010

6.34466.3241

6.2629

6.24326.10036.0396

5.83635.77555.63274.96222.14822.0607 4.51544.08912.18082.0204

3.68333.52083.2571 9.90649.55888.95501.1884 8.84477.86337.2506

0 1 2 3 4 5 6 7 8 9 10m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Relat

ive In

tensit

y

0.1758

0.3435

0.4006

0.5435 4.7990 7.63040.6147 4.32664.0796 7.36875.5774 10.39.58811.0611 8.87952.9427 6.8835 7.7499

ms ms

8.9 ms < 0.6 ms

Exactive Q Exactive & Orbitrap Elite

Improved Detector Hardware Reduces Delay in Transient Acquisition

See aslo ASMS Poster MP093 Oliver Lange et al.

0 1000 2000 3000 t, ms 0

Transferrin (78kD) z: +40

<10ms Efficient Capture of Initial Transient

Stronger Signal Better Mass Accuracy

6

Software Innovation – Advanced Signal Processing

• Increases resolution by 2 fold for the same transient time

0

20

40

60

80

100

993.97701 R=110221

993.89200 R=108377 994.08274

R=107785

993.82838 R=113310

994.23048 R=103203

Maximum Resolution:

Q Exactive: 140K (m/z 200) 100K (m/z 400)

Orbitrap Elite: 240K (m/z 400)

Enolase (46.64 kDa), 47+, Orbitrap Elite

• Isotopically Resolved LC & HC • Top-Down Analysis on LC Time Scale

7

Outline

• Orbitrap Mass Spectrometer in mAb Characterization

• Intact Protein Analysis on Q Exactive Using Protein Deconvolution

• Top Down Analysis of mAb Light Chain on Orbitrap Elite

• Conclusion

8

Intact Protein Mass Measurement on Q Exactive

1600 1800 2000 2200 2400 2600 2800 3000 3200 3400 3600 3800 4000 m/z

0 5

10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95

100

Rel

ativ

e Ab

unda

nce

2745.7720

2851.3544

2907.2595 2556.4844

2965.3764 2471.3405

3025.8632 2391.6288

1300 1400 1500 1600 1700 1800 1900 2000 2100 2200 2300 2400 2500 m/z

0 5

10 15

20

25 30

35

40 45

50

55 60

65

70 75

80

85 90

95

100

Rel

ativ

e A

bund

ance

1805.6858 z=10

2006.2045 z=9

2099.6662 z=9 1641.6240

z=11

1889.9027 z=10

1504.8222 z=12

2362.1277 z=8

2256.9799 z=8 1717.9145

z=11 1389.2991 z=13

2399.6221 z=8

2135.3361 z=9

1575.0062 z=12

1805.0 1805.5 1806.0 1806.5 1807.0 1807.5 1808.0 m/z

0 5

10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95

100

Rel

ativ

e A

bund

ance

1805.6858 z=10

1805.8863 z=10 1805.4848

z=10

1806.0875 z=10

1805.2842 z=10

1807.2867 z=10

1806.2882 z=10 1807.5870

z=10 1806.9859

z=10

1805.0834 z=10 1806.4884

z=10 1807.9868 z=10

Protein: ~20kD R: 140K

mAb:~150kD R: 17.5K

Isotopically Resolved

Isotopically Unresolved

9

Protein Deconvolution 1.0

• Workflow software for intact protein mass determination

• Includes 2 deconvolution algorithms: • Xtract for isotopically resolved proteins • ReSpectTM for isotopically unresolved proteins (e.g. IgG)

• As of today, the software is feature complete and undergoing testing

• Target release date: Late October, early November

10

Protein Deconvolution 1.0 – A Workflow Design

Select Algorithm Load File Create/Select Method

11

Protein Deconvolution Method Parameters

Straight-forward Parameter Setting

Automatic Resolution Detection

Estimated Target Mass

12

Protein Deconvolution – Chromatogram Tab

Averaged spectrum

13

Protein Deconvolution - ReSpectTM Deconvolution

Fast Data Processing:1-2s

Deconvoluted Spectrum

Deconvolution Data

14

Protein Deconvolution - Report

Comprehensive, Exportable Report

15

Isotopic Resolution for Accurate Mass Detection

1300 1400 1500 1600 1700 1800 1900 2000 2100 2200 2300 2400 2500 m/z

0 5

10 15

20

25 30

35

40 45

50

55 60

65

70 75

80

85 90

95

100

Rel

ativ

e A

bund

ance

1805.6858 z=10

2006.2045 z=9

2099.6662 z=9 1641.6240

z=11

1889.9027 z=10

1504.8222 z=12

2362.1277 z=8

2256.9799 z=8 1717.9145

z=11 1389.2991 z=13

2399.6221 z=8

2135.3361 z=9

1575.0062 z=12

1805.0 1805.5 1806.0 1806.5 1807.0 1807.5 1808.0 m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e A

bund

ance

1805.6858 z=10

1805.8863 z=10 1805.4848

z=10

1806.0875 z=10

1805.2842 z=10

1807.2867 z=10

1806.2882 z=10 1807.5870

z=10 1806.9859

z=10

1805.0834 z=10 1806.4884

z=10 1807.9868 z=10

Protein:~20kD R: 140K

Xtract

1 Oxidation Monoisotopic Mass

(<5ppm) 15.96 amu

16

1600 1800 2000 2200 2400 2600 2800 3000 3200 3400 3600 3800 4000 m/z

0 5

10 15 20 25 30 35 40 45 50 55 60 65 70 75 80 85 90 95

100

Rel

ativ

e Ab

unda

nce

2745.7720

2851.3544

2907.2595 2556.4844

2965.3764 2471.3405

3025.8632 2391.6288

-7 ppm G0F+G1F

G0F+G0F G0F+G2F (or 2G1F)

G1F+G2F G2F+G2F G0+G0F

G0+G0 2xMan5 G1F+G2F+SA

5.0 ppm

-0.7 ppm -8.5 ppm

-0.9 ppm

2680 2700 2720 2740 2760 2780 2800 2820 m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e A

bund

ance

2745.7720

2797.5697 2695.8919

Precise Mass Measurement of Intact Monoclonal Antibody

mAb: ~150kD R: 17.5K

ReSpect

Shiaw-Lin Wu, Barry Karger, Barnett Institute, Northeastern University

17

Accurate Mass Measurement from a Single Charge State

2835 2840 2845 2850 2855 2860 2865 2870 m/z

G0F/G1F Δ= 7.0 ppm

G0F/G0F Δ= 0.1 ppm

G0/G0F Δ= 10 ppm

G1F/G1F or G0F/G2F Δ = 10 ppm

G1F/G2F Δ = 8 ppm

52+

18

Reproducibility of Accurate MW Determination

ppm mass measurement errors RAW file Q Exactive G0+G0F G0F+G0F G0F+G1F G0F+G2F G1F+G2F

1 1 -10.5 0.7 -10.5 -13.8 -18.0

2 1 -11.6 -1.1 -8.8 -11.2 -12.0

3 1 5.1 -5.0 -2.6 5.1 5.6

4 2 -14.3 3.0 -6.9 -5.4 -5.9

5 2 -8.6 -2.2 -12.2 -12.5 -12.9

6 2 -14.3 -6.6 -12.3 -14.8 -10.1

-6.9 +/- 6.4 ppm mass tolerance for multiple measurements

ReSpectTM Deconvolution of Q Exactive IgG Data

19

Reproducibility of Relative Abundance Measurement

ReSpectTM Deconvolution of Q Exactive IgG Data

Relative Abundances

RAW file Q Exactive G0+G0F G0F+G0F G0F+G1F G0F+G2F G1F+G2F

1 1 12.9 74.1 100.0 67.0 23.4

2 1 12.0 72.8 100.0 66.2 22.0

3 1 12.2 75.0 100.0 67.0 23.6

4 2 12.7 75.7 100.0 63.6 21.6

5 2 13.2 75.4 100.0 64.8 21.0

6 2 12.9 76.6 100.0 64.7 21.6

CV 3.6% 1.7% N.A. 2.2% 4.9%

G0F+G1F G0F+G0F G0F+G2F (or 2G1F)

G1F+G2F G0+G0F

20

Summary of Q Exactive and Protein Deconvolution 1.0

• Q Exactive accurately and reproducibly measures the mass of intact antibody to ~7ppm.

• The new Protein Deconvolution software enables fast and accurate calculation of the intact mass of mAbs.

• Q Exactive is also a powerful HR/AM mass spectrometer for routine peptide identification and peptide mapping.

• This LC/MS-based workflow solution provides a robust QC solution for biopharmaceutical products.

21

Outline

• Orbitrap Mass Spectrometer in mAb Characterization

• Intact Protein Analysis on Q Exactive Using Protein Deconvolution

• Top-Down Analysis of mAb Light Chain on Orbitrap Elite

• Conclusion

22

Orbitrap Elite: Velos Pro Iontrap + High Field Orbitrap

Key Features for Intact Protein Analysis:

• Higher Resolution (240,000 at 1Hz)

• Faster Scan Speed • 4X more scans at same resolution

• Multiple Fragmentation Mechanism • CID • HCD • ETD

Case Study: Top-Down Analysis of mAb Light Chain

23

Step 1 – Accurate MW Determination of mAb Light Chain

1200 1400 1600 1800 2000 2200 2400 m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e A

bund

ance

1303.0917 z=18

1465.9155 z=16

1234.6137 z=19 1563.5760

z=15

1675.1174 z=14

1172.8326 z=20

1803.8945 z=13

1954.1323 z=12 2131.7785

z=11 1117.1264

z=21

1302.6 1302.8 1303.0 1303.2 1303.4 1303.6 1303.8 1304.0 m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e A

bund

ance

1303.0917 z=18

1303.2024 z=18

1303.0364 z=18

1303.3139 z=18

1302.9805 z=18

1302.9253 z=18

1303.4251 z=18

1302.8691 z=18

1303.5360 z=18 1302.7578

z=18

Xtract Measured Mass = 23424.4845 Target Mass = 23428.416 4 Dalton Mass Deviation 2 S-S?

R: 240K

Shiaw-Lin Wu, Barry Karger, Barnett Institute, Northeastern University

24

Step 2 - Top-Down Analysis with ETD at 120K Resolution

1000 1200 1400 1600 1800 2000 2200 2400 2600 2800 3000 m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e A

bund

ance

1802.8888 z=13

1954.5505 z=12

1675.4727 z=14

1545.0427 z=7

2132.2379 z=11

2345.6667 z=10

1416.4512 z=8

ETD R: 120K

1820 1830 1840 1850 1860 1870 1880 1890 1900 1910 m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e A

bund

ance

1888.0963 z=6

1833.3963 z=6 1819.8239

z=13

1871.9230 z=6 1824.0591

z=6 1909.9393 z=6

1850.4055 z=6 1828.3897

z=6

1904.9250 z=5 1843.0693

z=6 1859.5754

z=6 1898.0973

z=?

25

p @ [ ]

500 1000 1500 2000 2500 3000m/z

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Rel

ativ

e A

bund

ance

1954.22z=12

2131.78z=111763.19

z=6

1511.45z=7

1697.00z=7

589.26z=1

2644.27z=4

804.36z=1397.17

z=12197.07

z=5812.34

z=1571.25z=1

654.27z=1

1237.60z=1 2620.01

z=41041.48z=1

2420.42z=4

895.45z=1

2681.29z=4

1411.66z=?

2344.76z=?

2573.89z=?

x50 x5 x50

Step 3 - Top-Down Analysis with HCD at 120K Resolution

HCD R: 120K

26

4 Dalton deviation is probably due to internal disulfides.

Step 4 – ProSightPC Data Analysis

27

Summary of Top-Down Analysis on Orbitrap Elite

• Orbitrap Elite enables mAb subunit analysis at isotopic resolution, allows accurate mass measurement (<5ppm), and quick identification of minor modifications.

• The combination of multiple fragmentation mechanisms in top-down analysis generates comprehensive sequence coverage, enables fast localization of modification with minimum sample preparation.

• Orbitrap Elite enables comprehensive characterization of intact antibody with its superior performance and versatility.

28

HR/AM Intact Mass Measurement Workflow

On-line Desalting: HPLC Separation

Data Analysis: Protein Deconvolution 1.0 with ReSpector Xtract to zero charge accurate mass

Sample Analysis: Orbitrap Elite or Q Exactive High RP for resolution of proteins ~50k low RP for larger proteins

1311.0 1311.5 1312.0 1312.5m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

z=?

1311.87212R=70792

z=?1311.54737

R=68130z=?

1311.98174R=69867

z=?

1311.43967R=58977

z=? 1312.09147R=67084

z=?

1312.20087R=56981

z=?1311.31799

R=88597z=?

1310.98824R=45346

z=?

1312.42258R=47666

z=?

1312.64646R=43558

z=?

23565 23570 23575 23580 23585 23590m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

23578.58636

23580.66451

2000 2500 3000 3500m/z

0

10

20

30

40

50

60

70

80

90

100

Rel

ativ

e A

bund

ance

2997.31777

2920.44812

2664.49830

2178.40685

3535.704273619.47065

Solutions RSLC nano/cap ProSwift RP columns

Solutions Orbitrap hybrid family Q Exactive

Solutions ReSpect(unresolved spectra) Xtract (resolved spectra)

Exact Mass Determination Solutions

29

Conclusion

Thermo Fisher provides

complete LC/MS-based workflow solutions

for intact protein analysis

and comprehensive protein characterization.

30

Acknowledgement

• Q Exactive R&D Team • Orbitrap Elite R&D Team • San Jose Proteomics Marketing Team • San Jose Demo Lab

• Barnett Institute, Northeastern University

• Shiaw-Lin Wu • Barry Karger

31

Thank You!