G3C

G3 Patient Monitor

Service Manual

GENERAL MEDITECH, INC.RevisionRevision Date Changes

1.0 5/2009 First Issue

All Rights ReservedThis publication is protected by copyright and all rights are reserved. No part of this manual may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any purpose, without the express written permission of GENERAL MEDITECH, INC. (hereinafter called GMI).Restrictions and Liabilities

Information in this document is subject to change and does not represent a commitment by GMI. Changes made to the information in this document will be incorporated in new editions of the publication. No responsibility is assumed by GMI for the use or reliability of software or equipment that is not supplied by GMI or its affiliated dealers.

WarrantyG3 Multi-Parameter Patient Monitor is warranted by GMI for 1.5 years of free charge from the date of the purchase, while the SpO2 sensor, blood pressure cuff, ECG cable and body temperature sensor is warranted for 1 year. The consumables are excluded from the scope of this warranty. The following situations are not included in the scope of the warranty: The serial number of the monitor was ripped off or unreadable; or

The monitor was damaged due to improper connection with other equipment; or

The monitor was damaged due to accident; or

The user modified the monitor without GMIs written authorization.

ContactManufacturer: GENERAL MEDITECH, INC.Add: South Office 4/F, Kezhi Rd.W.No. 1, Science Park, Nanshan District, Shenzhen, Guangdong P.R.China

Tel: 86-755-26500832

Fax: 86-755-26546285

Zip: 518057

Internet: http://www.szmedtech.com

E-mail: office@szmedtech.com

EC Authorized Representative: Peretti Service S.A.S

Add: P.Le Tecchio 49/F 80125 Napoli Italy P.Iva 04552071211

Tel: 081 195 73 824Fax: 081 195 73 827

Contents1CHAPTER 1: OVERVIEW

1BRIEF INTRODUCTION

1SAFETY INFORMATION

1CLASSIFICATION

2PROHIBITION

2CAUTIONS

3CONTRAINDICATIONS

3FUNCTION

4CHAPTER 2: INTRODUCTION

4EXTERNAL APPEARANCE

4G3C

7G3D

10G3F

13G3G

16DISPLAY

18BATTERIES

18BATTERY MAINTENANCE

19BATTERY RECYCLING

20CHAPTER 3: INSTALLATION

20UNPACKING AND CHECKING

21ENVIRONMENTAL REQUIREMENTS

21INSTALLATION METHOD

21CONNECTING TO AC POWER SUPPLY

21INSTALLING THE BATTERY

22EQUIPOTENTIAL GROUNDING

22CONNECTING PATIENT SENSORS AND PROBES

22CONNECTING THE NETWORK CABLE

22CONNECTING TO VGA MONITOR

23REPLACE THE FUSE

23POWERING ON THE MONITOR

23POWERING OFF THE MONITOR

24CHAPTER 4: MENU

24MENU OPERATION

24POPUP MENU

24BROWSE MENU

24EXIT MENU

24DISPLAY

25FIXED FORMAT

26USER FORMAT

26FORMAT SETUP

27OXYCRG

27ALARM LIMIT

27WAVEFORM SPEED

28OTHER SETTINGS

29TOOLS

29REVIEW

30PATIENT

30SYSTEM SETUP

30SET TIME

31REGION SETTINGS

32NETWORK SETTINGS

33MAINTENANCE

35DEFAULT SETTINGS

36RECORDER SETUP

37CHAPTER 5: ALARMS

37ALARM CATEGORIES

37ALARM MODES

37ALARM SETTINGS

37SETTING ALARM VIA ALARM KEY

39SETTING ALARM VIA PARAMETER LABEL

40WHEN AN ALARM OCCURS

41Chapter 6: WAVEFORM FREEZING AND RECALLING

41OVERVIEW

41FREEZING AND UNFREEZING

41WAVEFORM FREEZING

41WAVEFORM UNFREEZING

41WAVEFORM RECALLING

43CHAPTER 7: PRINTING

43OVERVIEW

43PRINTING TYPES

43REAL-TIME PRINTING

43TIMER PRINTING

43ALARM PRINTING

44WAVEFORM AND PARAMETER PRINTING

44INSTALLING PRINTER PAPER

45CHAPTER 8: TRENDS

45OVERVIEW

45TREND CHANNELS

46TREND GRAPH CHANNEL

47TREND TABLE CHANNEL

48TREND WINDOWS

48TREND GRAPH WINDOW

50TREND TABLE WINDOW

51CHAPTER 9: EVENTS

51RECORD EVENTS

51BROWSE EVENTS

52CHAPTER 10: DRUG CALCULATOR

52ENTER DRUG CALCULATOR

52UNIT

53TERMS

53DRUG CALCULATOR

53NUMERIC INPUT BLOCK

53CALCULATION FORMULA

54KNOWN ITEM AND CALCULATION RESULTS

54CALCULATION FOR ANY DRUG

54CALCULATION FOR EXACT DRUG

55DRIP TABLE

56TITRATION TABLE

56RESET

57CHAPTER 11: OTHER PATIENT VIEWING

57OVERVIEW

57SELECT MONITOR

58OTHER PATIENT

59CHAPTER 12: ECG MONITORING

59OVERVIEW

59ECG MONITORING

59PREPARATION

60ELECTRODE PLACEMENT

62ECG CHANNEL

62MAIN LEAD CHANNEL

64OTHER LEAD CHANNELS

64ECG PARAMETER

64HEART RATE

66ST SEGMENT

67MAINTENANCE AND CLEANING

67TROUBLE SHOOTING

68CHAPTER 13: RESP MONITORING

68OVERVIEW

68ELECTRODE PLACEMENT

69MONITORING INTERFACE

69RESP CHANNEL

70RESPIRATION RATE

70PROMPT MESSAGES

71CHAPTER 14: T-RESP MONITORING

71T-RESP CHANNEL

71T-RESP MENU

71MAINTENANCE AND CLEANING

72CHAPTER 15: NIBP MONITORING

72OVERVIEW

72MONITORING PROCEDURE

72CUFF SELECTION AND PLACEMENT

74MEASUREMENT LIMITATIONS

74MONITORING INTERFACE

76FUNCTIONS

76NIBP MEASUREMENT

77VENIPUNCTURE

77MAINTENANCE AND CLEANING

78TROUBLE SHOOTING

79CHAPTER 16: TEMP MONITORING

79TEMPERATURE PROBE INSTALLATION

79TEMP PARAMETER

80TEMP MENU

81MAINTENANCE AND CLEANING

82CHAPTER 17: SpO2 MONITORING

82OVERVIEW

83PRECAUTIONS

84MONITORING PROCEDURE

84MEASUREMENT LIMITATIONS

84MENU

84SpO2 MENU

85PLETH MENU

85MAINTENANCE AND CLEANING

86 CHAPTER 18: IBP MONITORING

86IBP CHANNELS

87IBP PARAMETER

88PRECAUTIONS

88MONITORING PROCEDURE

89IBP MENU

89IBP PARAMETER MENU

90IBP CHANNEL MENU

90MAINTENANCE AND CLEANING

91CLEANING OF IBP TRANSDUCER

91STERILIZATION

92CHAPTER 19: CO2 MONITORING

92INTRODUCTION

92MONITORING PROCEDURE

93MONITORING INTERFACE

93CO2 CHANNEL

94CO2 PARAMETER

95MAINTENANCE AND CLEANING

96CHAPTER 20: MAINTENANCE

96INSPECTION

96CLEANING

97DISINFECTION AND STERILIZATION

98CHAPTER 21: LABELS, PACKAGING, TRANSPORT AND STORING

98LABELS

99PACKAGING

99TRANSPORT

99STORING

100APPENDIX A: PPRODUCT SPECIFICATIONS

100ENVIRONMENTAL SPECIFICATIONS

100POWER SOURCE SPECIFICATIONS

101HARDWARE SPECIFICATIONS

101DATA STORAGE

101ECG SPECIFICATIONS

102RESP SPECIFICATIONS

102NIBP SPECIFICATIONS

103TEMP SPECIFICATIONS

103SpO2 SPECIFICATIONS

103PR SPECIFICATIONS

103IBP SPECIFICATIONS

103CO2 SPECIFICATIONS

104APPENDIX B: EMC

CHAPTER 1: OVERVIEWWelcome to use G3 Multi-Parameter Patient MonitorThe main purpose of this manual is to provide operating guide, instrument repair and maintenance information to users. The detailed performance indexes, routine installation, operation and maintenance method as well as safety information are illustrated in this manual. Before use, the serviceman shall carefully read this manual in order to properly and correctly operate this monitor, so that it can reach specified safety standards and performance indexes.

BRIEF INTRODUCTIONThe G3 Multi-Parameter Patient Monitor is a multifunctional overall physiologic index monitor, which is applicable to monitor the ECG, NIBP, SpO2, PR, RESP, TEMP, IBP and CO2 in hospitals.

This product has the multi-parameter function which can be selected, combined and configured by the user according to his requirements (but the G3 Multi-Parameter Patient Monitor which you are using probably only had part parameters and corresponding accessories selected by you when you bought it).

This product is composed of main machine and corresponding functional accessories such as ECG cable, blood pressure cuff, SpO2 sensor, body temperature sensor, BP hose, IBP sensor, dehydration vase and sampling line.

This product has three input/output connectors for printer, network communication and external VGA monitor.

SAFETY INFORMATION

CLASSIFICATIONAccording to the type of protection against electric shock:----- CLASS I, INTERNALLY POWERED EQUIPMENT.

According to the degree of protection against electric shock: ----- TYPE BF APPLIED PART or TYPE CF APPLIED PART. The connector with a symbol beside is TYPE BF APPLIED PART, the connector with a symbol beside is TYPE CF APPLIED PART.

According to the degree of protection against ingress of water:

----- Not protected (ordinary).According to the method(s) of sterilisation or disinfection recommended by the manufacturer.----- Equipment with method(s) of sterilisation or disinfection recommended by the manufacturer.According to the degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR or WITH OXYGEN OR NITROUS OXIDE:

----- EQUIPMENT not suitable for use in the presence of a FLAMMABLE ANAESTHETIC MIXTURE.

According to the mode of operation:----- CONTINUOUS OPERATION.

PROHIBITION

The marked symbol on the Monitor means: Attention, consult accompanying documents (this manual).Anywhere marked with