michael hartman, crna, dnp, msn, bsn, ba · 2018. 3. 31. · **leads to heart failure and require...

46
Michael Hartman, CRNA, DNP, MSN, BSN, BA

Upload: others

Post on 25-Feb-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Michael Hartman, CRNA, DNP, MSN, BSN, BA

Page 2: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize
Presenter
Presentation Notes
Already aware of anatomy and physiology- chambers-valves-vessels of the heart http://www.bem.fi/book/06/06.htm http://www.google.com/imgres?imgurl=http://home.comcast.net/~pegglestoncbsd/cardiac%2520conduction.gif&imgrefurl=http://home.comcast.net/~pegglestoncbsd/cardiovascular.htm&usg=__J4QBxqOadcbz1CKIFWv8-kGIR9c=&h=364&w=500&sz=70&hl=en&start=0&zoom=1&tbnid=9o0KbS1Z5FY07M:&tbnh=119&tbnw=164&prev=/images%3Fq%3Dcardiac%2Bconduction%2Bsystem%2Banamation%26um%3D1%26hl%3Den%26sa%3DX%26rls%3Dcom.microsoft:en-us:IE-SearchBox%26rlz%3D1I7TSNA_enUS350US350%26nfpr%3D1%26biw%3D1041%26bih%3D433%26tbs%3Disch:1&um=1&itbs=1&iact=hc&vpx=621&vpy=147&dur=1762&hovh=191&hovw=263&tx=127&ty=117&ei=7xFzTP6BJ4OglAfhypGEDw&oei=7xFzTP6BJ4OglAfhypGEDw&esq=1&page=1&ndsp=12&ved=1t:429,r:10,s:0
Page 3: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize
Presenter
Presentation Notes
Resting membrane of -70 once the anions levels reach the -70 than depolarization occurs this is why K can increase or decrease the excitability of the cardiac ventricular myocyte (threshold voltage of about -70 mV ) 0-3 absolute refractory period stimulus will cause VF or VT arrhythmia
Page 4: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Intrinsic Conduction Automaticity-cardiac cells ability to spontaneously

depolarize and initiate impulse Depolarization-cardiac cells go to a+ intracellular

and – extracellular charges from influx of Na and Ca

Repolarization –return to resting

Presenter
Presentation Notes
SA node 70-100 beats/min , AV node 40-60 bts/min, Bundle and purkingie fibers 20-40 bts/min
Page 5: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize
Page 6: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Transcutaneous External Pacing (R2 Pads with Defibrillators) Epicardial Temporary lead placement in the epicardium and utilized

with open heart surgery Transvenous Temporary lead placed within a vein (femoral, subclavian,

or internal jugular veins) Permanent Surgical placement with fluoroscopy of the lead(s) within a

vein and connected to a generator with implantation under the skin

Page 7: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Diagnostic Testing Sensing: assesses ability of device to identify intrinsic

cardiac activity within the atria and/or ventricles Impedance: assesses integrity of lead Threshold: determines the minimal amount of energy

required to potentiate depolarization of the atria or ventricle

Presenter
Presentation Notes
All help to prolong device longevity
Page 8: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Existence as of 2007:

PPM- 3Million ICD’s-300,000

Presenter
Presentation Notes
Intrinsic Conduction will be assisted either as DEPENDENT or ASSISTED
Page 9: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Founded in 1980

Page 10: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize
Presenter
Presentation Notes
FACTS:: Pulse Generator or Battery has an average of 5-10years life expectancy and weigh an average of 20-50grams (example 10 Nickels or one small sized apple) First implanted between 1980-1982 in John Hopkins Hospital and Stanford University Hospital Criteria for placement included: Pt need to survive two events of cardiac arrest not attributed to MI and documented VF arrest and on antiarrtyhmic medications
Page 11: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize
Presenter
Presentation Notes
http://www.blackwellpublishing.com/content/BPL_Images/Content_store/Sample_chapter/0727915665/Chow%20Sample%20Chap%201.pdf
Page 12: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Uni-polar Primarily utilized with

pacers Creates large pacer spike

on EKG Very sensitive to EMI

Bi-polar Common with pacers Less operating watts More resistant to EMI

Page 13: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize
Presenter
Presentation Notes
NOTE insulator on X-ray of ICD leads
Page 14: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Bipolar Single-coil Dual-coil *Active and Passive Fixation * Multiple lengths and diameter *Flexible “Floppy”

Presenter
Presentation Notes
Evolving since 1993 first with applicable lead to diameter size and consolidation Latest release of a 6.5 fr icd lead by St. Jude in Europe http://www.cardiotextpublishing.com/pdf/9780979016462-ch12-sample.pdf
Page 15: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Rhythm Bradycardia

(Symptomatic) Brady/Tachy Syndrome Tachycardia

Conduction Complete AV Block Second Degree AV Block Intraventricular

Conduction Defects

Presenter
Presentation Notes
Atrial Fib have utilized with success bi-atrial pacing
Page 16: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Anatomical Hypertrophic Cardiomyopathy Dilated Cardiomyopathy **Leads to heart failure and require Bi-ventricular pacing or cardiac

resynchronization therapy (CRT)s

Presenter
Presentation Notes
Chest xray of a 38 year old pregnant female with cardiomyopathy and lupus (SLE)
Page 17: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Types: Single Chamber- one lead implanted within the appropriate chamber atrial or

ventricular Dual Chamber- two leads implanted with one in the right atrium and one in the right

ventricle Bi-ventricular- right atrial and ventricular leads with a lead passed through the

coronary sinus to pace the left ventricle (Goal is to resynchronize the heart chambers thus optimizing cardiac output)

Presenter
Presentation Notes
http://missinglink.ucsf.edu/lm/pacemaker_module/index.htm
Page 18: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

North American Society of Pacing and Electrophysiology (NASPE ) and British Pacing and Electrophysiology Group (BPEG)

Programming Codes :

I II III IV V

Chamber(s) Paced

Chamber(s) Sensed

Response to Sensing

Rate Modulation

Multisite Pacing

O=None O=None O=None O=None O=None

A=Atrium A=Atrium T=Triggered R=Rate Modulation

A=Atrium

V=Ventricle V=Ventricle I=Inhibited V=Ventricle

D=Dual (A+V)

D=Dual (A+V)

D=Dual (T+I)

D=Dual (A+V)

Presenter
Presentation Notes
Example explain of a DDDR device Multisite pacing is stimulation along a lead to specified areas within the atrial or ventricular chamber theorized to decrease stress of on area of the heart chamber Multisite pacing for prevention of atrial tachyarrhythmias: potential mechanisms
Page 19: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

North American Society of Pacing and Electrophysiology (NASPE) and British Pacing and Electrophysiology Group (BPEG)

Programming Codes:

I II III IV

Shock Chamber Antitachycardia Pacing Chamber

Tachycardia Detection

Antibradycardia Pacing Chamber

O = None O = None E=Electrogram O=None

A= Atrium A= Atrium H=Hemodynamic A= Atrium

V= Ventricle V=Ventricle V=Ventricle

D=Dual (A+V) D=Dual (A+V) D=Dual (A+V)

Page 20: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize
Page 21: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Class I: 1. Cardiac arrest due to ventricular fibrillation (VF) or ventricular tachycardia

(VT) not due to a transient or reversible cause. 2. Spontaneous sustained VT in association with structural heart disease. 3. Syncope of undetermined origin with clinically relevant, hemodynamically

significant sustained VT or VF induced at electrophysiologic study when drug therapy is ineffective, not tolerated, or not preferred.

4. Nonsustained VT in patients with coronary artery disease, prior myocardial

infarction (MI), left ventricular (LV) dysfunction, and inducible VF or sustained VT at electrophysiologicstudy that is not suppressible by a Class I antiarrhythmic agent.

5. Spontaneous sustained VT in patients who do not have structural heart

disease that is not amenable to other treatments.

Page 22: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Class IIa • Patients with LV ejection fraction of 30% or less, at

least 1 month post-MI and 3 months post coronary artery revascularization surgery.

Page 23: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Class IIb: 1. Cardiac arrest presumed to be due to VF when electrophysiologic testing is

precluded by other medical conditions. 2. Severe symptoms (eg, syncope) attributable to sustained ventricular

tachyarrhythmias in patients awaiting cardiac transplantation. 3. Familial or inherited conditions with a high risk for life threatening

ventricular tachyarrhythmias, such as long QT syndrome or hypertrophic cardiomyopathy.

4. Nonsustained VT with coronary artery disease, prior MI, LV dysfunction, and inducible sustained VT or VF at electrophysiologic Study.

5. Recurrent syncope of undetermined etiology in the presence of ventricular dysfunction and inducible ventricular arrhythmias at electrophysiologic study when other

causes of syncope have been excluded. 6. Syncope of unexplained etiology or family history of unexplained sudden

cardiac death in association with typical or atypical right bundle-branch block and ST-segment elevations (Brugada syndrome).

7. Syncope in patients with advanced structural heart disease in which thorough invasive and noninvasive investigation has failed to define a cause.

Page 24: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Class III: 1. Syncope of undetermined cause in a patient without inducible

ventricular tachyarrhythmias and without structural heart disease. 2. Incessant VT or VF. 3. VT or VF resulting from arrhythmias amenable to surgical or catheter

ablation; for example, atrial arrhythmias associated with the Wolff-Parkinson-White syndrome, right ventricular outflow tract VT, idiopathic LV tachycardia, or fascicular VT.

4. Ventricular tachyarrhythmias due to a transient or reversible disorder (eg, acute MI, electrolyte imbalance, drugs, or trauma) when correction of the disorder is considered feasible and likely to substantially reduce the risk of recurrent arrhythmia.

5. Significant psychiatric illnesses that may be aggravated by device implantation or may preclude systematic follow- up.

6. Terminal illnesses with projected life expectancy of 6 months or less. 7. Patients with coronary artery disease, LV dysfunction, and prolonged

QRS duration in the absence of spontaneous or inducible sustained or nonsustained VT who are undergoing coronary bypass surgery.

8. NYHA Class 4 drug-refractory congestive heart failure in patients who are not candidates for cardiac transplantation.

Page 25: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Explanations: 1. Literature - published conflicting information 2. Literature – current advances and care practices

not published (New Anesthesia Machines- Battery and MH

preparation)

Information Confusion

Presenter
Presentation Notes
PPM and ICD: Dr, Rozner, MDA, PhD
Page 26: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize
Page 27: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Anesthesia Evaluation for PPM and ICD : Chest x-ray Serum K+, H&H, and coagulation Device Card Device Type Device Last Interrogation Determine use of EMI (Cautery) **Prevention: Delays Cancellation Sentinel Complications

Presenter
Presentation Notes
High K+ can cause VT Low K+ can cause loss of pacer capture Information of device is vital on preadmission testing
Page 28: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Best practices are to have the device evaluated or contact company representative for specific care instructions.

Questions:

PPM -Is the patient dependent? ICD- Is it magnet resistant?

Presenter
Presentation Notes
We always call the company but make sure they understand what you are asking as I will discuss later in a case study
Page 29: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Medtronic and Biotronik with magnet on and off will disable shock therapy with ICD –indicating no need to contact representative unless the surgery is within 6-inches of the device.

Exceptions: If PPM/ICD is within the sterile field (i.e. Rotator cuff repair

or carotid endarectomy) than device representative needs to program the device into asynchronous mode/no shock therapy

If patient positioning (i.e. prone) not optimal to maintain magnet than device representative should program device to asynchronous mode/ no shock therapy

Page 30: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Pacer Dependent

Cautery with 6”

Prone Positioning

Device in Sterile Field

MRI Lithotripsy (ESWL)

Boston Scientific / Guidant

Reprogramming pre and post-operatively

Reprogramming pre and post-operatively

Reprogramming pre and post-operatively

Reprogramming pre and post-operatively

Contraindicated Reprogramming pre, intra, and post-operatively

Call representative to determine magnet response

Page 31: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Medtronic Guidant/Boston Scientific Biotronik St. Jude / Pacesetter

Implantable Device Companies:

Presenter
Presentation Notes
Could also reference you EP or Cardiac Catherization Lab or unit for Representative information or Device management
Page 32: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Heart Rate and rhythm monitoring Place magnet on pacer if EMI is interfering or if patient is pacer dependent Remove magnet from ICD to activate detection and therapy settings for a

patient displaying an aberrant arrhythmia External defibrillation equipment readily available (i.e. defibrillator, zoll

pacer, transvenous pacing cather, transespohgeal pacing) Minimize use of SA and AV node sensitive medications such as Precedex

and opiates with known underlying rhythm of SSS or asystole

Presenter
Presentation Notes
External equipment in OR arena, ACLS
Page 33: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Pacer and ICD: Monitor heart rate and rhythm Contact device representative Follow care guidelines of the facility Available external pacing and defibrillator devices

Page 34: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Pacemakers Designed to evaluate generator life Pacing threshold safety factors

Page 35: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Medtronic (PPM) Paced at a heart rate of

85 indicates generator OK

Paced at a heart below 65 indicates that generator requires replacement

Paced at a heart rate of 98 indicates generator OK

Paced at a heart below 87 indicates that generator requires replacement

St. Jude Medical (PPM)

Presenter
Presentation Notes
Responses with magnet ON!!! ELEVATED HEART RATE PERMITS TELEPHONE TRANSMISSION AND INTERROGATION*****
Page 36: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

ICD Medtronic: Anesthesia Provider’s Awareness: Magnet remains secured over device Nonprogrammable for magnet resistance Audible confirmation only if patient alert or care alert is

activated (30seconds of constant tone) NOTE: if sustained constant or pulsing tone audible- stop and contact device representative

Disables tachyarrhythmia detection and therapy from EMI Remove magnet to reactivate detection and therapy

settings for tachyarrhythmia Have device evaluate postoperatively by representative

Page 37: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

ICD Guidant and Boston Scientific: Anesthesia Provider’s Awareness: Magnet remains secured over device Programmable for magnet resistance Audible confirmation –synchronous tones with heart beat to continuous

after 30seconds NOTE: Tones present prior to magnet placement or no tones aucultated

with magnet placement requires device interrogation from company representative

Disables Tachyarrhythmia detection and therapy from EMI Remove magnet greater than 2’ to reactivate detection and therapy

settings for tachyarrhythmia Have device evaluate postoperatively by representative since electrical

reset can occur

Page 38: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

ICD St. Jude Medical: Anesthesia Provider’s Awareness: Magnet remains secured over device Note: Magnet positioning must be secured over the top or bottom

edge of the device in order to achieve magnet reversion Programmable for magnet resistance No Audible Disables Tachyarrhythmia detection and therapy from EMI Remove magnet to reactivate detection and therapy settings for

tachyarrhythmia Have device evaluate postoperatively by representative

Presenter
Presentation Notes
From St. Jude’s Cardiac rhythm management division
Page 39: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Magnet placement is important since poorly positioned may not produce the desired effect.

Magnets placed directly on top of the Medtronic, Boston

Scientific, and Biotronik ICD’s. *Exception: St. Jude ICD’s, magnet is placed off-centre

with the curve of the magnet over the bottom or top end of the ICD

Page 40: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Awareness: 1. Difficult to differentiate a PPM from an ICD on an EKG 2. Underlying rhythm 3. Magnet on ICD does not convert pacer function to

asynchronous mode (VOO) 4. ICD’s translate electromagnetic interference (EMI)

differently 5. Chest x-ray to identify lead type and Bi-Ventricular left

ventricle lead location in relation to the coronary sinus especially with central line placement

6. Reprogramming to VOO requires an increased heart rate than intrinsic rate in order to ensure tissue perfusion and deter under sensing and over sensing from EMI (cautery)

Presenter
Presentation Notes
Central line placement precaution since dislodgement can occur and prior documentation to lead location for comparison post line placement CYA
Page 41: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Development of Practice Guidelines: Surgeon notifying the scheduling department indicates devices

existence Magnets available on all anesthesia carts Access to an alternative pacing or defibrillator source No EMI within six inches of the device Utilize Bipolar EMI or harmonic scalpel Minimize EMI Joules Grounding Pad away from device but as close to the surgical field

to minimize aberrant current absorption by device Utilize burst EMI with unipolar with five second pauses to

minimize inhibition of the device Device in reprogrammed preoperatively it should be evaluate

postoperatively by a device representative

Page 42: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Colonoscopy 78yr old with prior history of colon polyps with ICD Company representative present to inform that the

device was a magnet resistant device and required reprogramming prior to start of case

As of 2007: 45,000 Magnet Resistant

Devices Exist

Presenter
Presentation Notes
by Guidant and Boston Scientific
Page 43: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Colonoscopy 70 yr old with prior history of colon polyps and ICD ICD representative present and informed that

available for reprogramming if cautery is to be utilized.

Presenter
Presentation Notes
Pt had multiple polypectomies and device reprogrammed
Page 44: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Need for Anesthesia departments to develop pacer and ICD clinical practice guidelines

Need for Anesthesia providers to be aware of such variances in order to optimize care

Question company representatives Educate fellow practitioners and students of anesthesia Ensure safe surgical experience for the patient through

vigilance

Presenter
Presentation Notes
Remember the Company upholds the “Secret” magnet resistant list
Page 45: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

The variations of models, magnet responses, and evolution of new technology requires anesthesia providers to heighten their awareness and vigilance to pre, intra, and postoperative care measures of PPM and ICD patients in order to optimize surgical outcomes.

If uncertain contact the devices representative and inquire about care measures, device settings, and patient details (i.e. magnet resistant or pacer dependent)

Page 46: Michael Hartman, CRNA, DNP, MSN, BSN, BA · 2018. 3. 31. · **Leads to heart failure and require Bi -ventricular pacing or cardiac resynchronization therapy ... (Goal is to resynchronize

Prepared without any financial incentives or obligations from any PPM and ICD device manufactures or any company in relation to interventional cardiac therapy.

Questions or Comments…