metallic iplants

18
METALLIC IMPLANT MATERIALS ABSTRACT With increase in the average age of the population has, in turn, led to a rapidly increasing number of surgical procedures involving prosthesis implantation, because as the human body ages, the load-bearing joints become more prone to ailments. This has resulted in an urgent need for improved biomaterials and processing technologies for implants, more so for orthopaedic and dental. We are using different biomaterial like metal. Polymer, ceramic and composites for implants for implant application. Among them metallic implants is the oldest one and still the best in strength, ductility and durability.Metallic biomaterials continue to be used extensively for the fabrication of surgical implants due to high strength and resistance to fracture along with a relative ease of fabrication of both simple and complex shapes using well-established and widely available fabrication techniques. They are used in joint replacement, hip/knee replacement, cochlear implants, cardiovascular implants, dental implants etc. In this paper we are discussing about different metal implants and their advantages. 1.INTRODUCTION In surgery, a biocompatible material (sometimes shortened to biomaterial) is a synthetic material used to replace part of a living system or to function in intimate contact with living tissue. The Clemson University Advisory Board for Biomaterials has formally defined a biomaterial to be a systematically and pharmacologically inert substance designed for implantation within or incorporation with living systems. By contrast a biological material is a material such as bone matrix or tooth enamel, produced by a biological system. [1] The use of biomaterials is shown in Table 1, which include replacement of a body part that has lost function due to 1 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 38

Upload: jobin-jose

Post on 21-Jul-2016

3 views

Category:

Documents


1 download

DESCRIPTION

With increase in the average age of the population has, in turn, led to a rapidly increasing number of surgical procedures involving prosthesis implantation, because as the human body ages, the load-bearing joints become more prone to ailments. This has resulted in an urgent need for improved biomaterials and processing technologies for implants, more so for orthopaedic and dental. We are using different biomaterial like metal. Polymer, ceramic and composites for implants for implant application. Among them metallic implants is the oldest one and still the best in strength, ductility and durability.Metallic biomaterials continue to be used extensively for the fabrication of surgical implants due to high strength and resistance to fracture along with a relative ease of fabrication of both simple and complex shapes using well-established and widely available fabrication techniques. They are used in joint replacement, hip/knee replacement, cochlear implants, cardiovascular implants, dental implants etc. In this paper we are discussing about different metal implants and their advantages.

TRANSCRIPT

Page 1: Metallic Iplants

METALLIC IMPLANT MATERIALS

ABSTRACT

With increase in the average age of the population has, in turn, led to a rapidly increasing number of surgical procedures involving prosthesis implantation, because as the human body ages, the load-bearing joints become more prone to ailments. This has resulted in an urgent need for improved biomaterials and processing technologies for implants, more so for orthopaedic and dental. We are using different biomaterial like metal. Polymer, ceramic and composites for implants for implant application. Among them metallic implants is the oldest one and still the best in strength, ductility and durability.Metallic biomaterials continue to be used extensively for the fabrication of surgical implants due to high strength and resistance to fracture along with a relative ease of fabrication of both simple and complex shapes using well-established and widely available fabrication techniques. They are used in joint replacement, hip/knee replacement, cochlear implants, cardiovascular implants, dental implants etc. In this paper we are discussing about different metal implants and their advantages.

1. INTRODUCTIONIn surgery, a biocompatible material (sometimes shortened to biomaterial) is a synthetic material used to replace part of a living system or to function in intimate contact with living tissue. The Clemson University Advisory Board for Biomaterials has formally defined a biomaterial to be a systematically and pharmacologically inert substance designed for implantation within or incorporation with living systems. By contrast a biological material is a material such as bone matrix or tooth enamel, produced by a biological system. [1]The use of biomaterials is shown in Table 1, which include replacement of a body part that has lost function due to disease or trauma, to assist in healing, to improve function, and to correct abnormalities.Biomaterials can be classified from the point of view of the problem area that is to be solved (Table 1), the body on a tissue level, an organ level (Table 2), or a system level (Table 3). Also classified as metals, polymers, ceramics and composites (Table 4).

1.1. Need for implants Joints such as hips, knees, and shoulders are often prone to degenerative and inflammatory diseases that result in pain and joint stiffness.[2] Apart from the usual decay of articular cartilage due to age, there are many illnesses such as osteoarthritis (inflammation of bone), rheumatoid arthritis (inflammation of synovial membrane), and chondromalacia (softening of cartilage) which affect the joints. The bone cell density of a osteoporosis bone is substantially lower than normal bone. Such premature joint degeneration may arise mainly from three conditions: deficiencies in joint biomaterial properties, excessive loading conditions, and failure of normal repair [2]. Although minor surgical treatments are done to provide temporary relief to numerous patients, there is a consensus that the ultimate step is to replace the dysfunctional natural joints for prolonged pain relief and mobility. Arthroplasty is an operative procedure of orthopedic surgery performed, in which the arthritic or dysfunctional joint surface is replaced with something better .Currently, one of the main achieve-ments in the field of arthroplasty is Total Joint Replacement (TJR), where the entire load-bearing joint (mainly in the knee, hip, or shoul-der) is replaced surgically by ceramic,

1

1

2345

6789

101112131415161718192021222324252627282930313233343536373839404142434445464748

Page 2: Metallic Iplants

metal, or polymeric artificial materials. As stated earlier, the problem is that not all artificial materials could be used for such purposes, only the ones that fulfil certain broad specifications. similarly the human tooth, consisting of enamel, dentin, pulp, and cementum, is a highly specialized calcified structure used to break down food. It is a site where most surgical procedures in humans are performed.

1.2. Implant Properties The property requirements of a modern-dayimplant can broadly be categorized into three equally important features.

1. The human body must be compatible with the material used for the prosthesis.2. The implant should have the desired balance of mechanical and physical properties

necessary to perform as expected.3. The device under question should be relatively easy to fabricate, being reproducible.

Apart from these factors, the selection of the implant material itself is the principal criterion for proper functioning. Material should be biologically and mechanically compatible. . For example, plastic or ceramic material is used for smaller implants like cochlear and dental prostheses. However, for making total hip replacements and total knee replacements, metallic implants the best candidate due to their higher tensile load-bearing capabilities.

2. METALLIC IMPLANTS

2.1 Importance of Metallic Implants.

Metallic biomaterials continue to be used extensively for the fabrication of surgical implants due to high strength and resistance to fracture along with a relative ease of fabrication of both simple and complex shapes using well-established and widely available fabrication techniques. They are used in joint replacement, hip/knee replacement, cochlear implants, cardiovascular implants, dental implants etc. Various metal implants and their uses are noted in table 5.

2.2. Types of metallic implant materials

Most metals such as iron (Fe), chromium (Cr), cobalt (Co), nickel (Ni), titanium (Ti), tantalum (Ta), niobium (Nb), molybdenum (Mo), and tungsten (W) that were used to make alloys for manufacturing implants.[4] Sometimes those metallic elements, in naturally occurring forms, are essential in red blood cell functions (Fe) or synthesis of a vitamin B-12 (Co), but cannot be tolerated in large amounts in the body. [4] The biocompatibility of the metallic implant is of considerable concern because these implants can corrode in an in vivo environment [5]. The consequences of corrosion are the disintegration of the implant material per se, which will weaken the implant, and the harmful effect of corrosion products on the surrounding tissues and organs. [6].

The various metals and alloys used for implant fabrications are- Stainless steels- Cobalt based alloys- Titanium and titanium based alloys

2

49505152535455565758

59606162

636465666768

69

70

717273747576

77

7879808182838485868788899091

Page 3: Metallic Iplants

- Dental metals- Other metals

2.2.1 Stainless SteelThe family of alloy steels usually containing 10 to 30 percent chromium is called stainless steel. In conjunction with low carbon contents, chromium imparts remarkable resistance to corrosion and heat. Other elements such as nickel, molybdenum, titanium, aluminum, niobium, copper, nitrogen, sulfur, phosphorus, and selenium may be added to increase corrosion resistance to specific environments, enhance oxidation resistance, and impart special characteristics. The first metal alloy developed specifically for human use was the “vanadium steel” which was used to manufacture bone fracture plates (Sherman plates) and screws. It is no longerused in implants due to inadequate corrosion resistance. [8] The first stainless steel utilized for implant fabrication was the 18-8 (type 302 in modern classification), which is stronger and more resistant to corrosion than the vanadium steel. To increase the corrosion resistance a small percentage of Molybdenum is introduced (type 316). Carbon content of 316 stainless steel was reduced from 0.08 to a maximum amount of 0.03% and it became known as type 316L stainless steel. The austenitic stainless steels, especially types 316 and 316L are most widely used for implants. These are not hard enable by heat treatment but can be hardened by cold-working. This group of stainless steels is nonmagnetic and possesses better corrosion resistance than any others. The inclusion of molybdenum enhances resistance to pitting corrosion in salt water. The ASTM (American Society for Testing and Materials) recommends type 316L rather than 316 for implant fabrication. The specifications for 316 and 316L stainless steels are given in Table 6.

Properties of stainless steels

A wide range of properties can be obtained for 316 and 316L depending on the heat treatment (to obtain softer materials) or cold working (for greater strength and hardness). The designer must be careful when selecting materials of this type. Even the type 316L stainless steels may corrode inside the body under certain circumstances such as in a highly stressed and oxygen-depleted region. They are however, suitable to use in temporary devices such as fracture plates, screws, and hip nails, some joint replacement components [8,9] Surface modification methods such as anodization, passivation, and glow-discharge nitrogen-implantation, are widely used in order to improve corro-sion resistance, wear resistance, and fatiguestrength of 316L stainless steel [8]

2.2.2. Cobalt Based Alloys

These materials are usually referred to as cobalt-chromium alloys. There are basically two types- the castable CoCrMo - CoNiCrMo alloy, which is usually wrought by (hot) forging. The castable CoCrMo alloy has been used in making artificial joints. The wrought CoNiCrMo alloy is a newcomer now used for making the stems of prosthesis for heavily loaded joints such as the knee and hip. The two basic elements of the CoCr alloys form a solid solution of up to 65% Co. The molybdenumis added to produce finer grains which results in higher strengths after casting or

3

9293

949596979899

100101102103104105106107108109110111112113114115116117118119120121122123124125126127128129130131132133134135136137138139140

Page 4: Metallic Iplants

forging. The chromiumenhances corrosion resistance as well as solid solution strengthening of the alloy.

Types and compositions of Co-Based AlloysASTM lists four types of Co based alloys that are recommended for surgical implant applications.1. cast CoCrMo alloy (F67)2. wrought CoCrWNi alloy (F90)3. wrought CoNiCrMo alloy (F562)4. wrought CoNiCrMoWFe alloy (F563)At the present time, only two of the four alloys are used extensively in implant fabrications, the castable CoCrMo and the wrought CoNiCrMo alloy.Chemical Composition of Co-Based Alloys is given in table 7.[8]

Properties of Co-Based Alloys

The two basic elements of the Co-based alloys form a solid solution of up to 65 wt% Co and the remainder is Cr. Molybdenum is added to produce finer grains, which results in higher strengths after casting or forging.

The wrought Co-based alloy CoNiCrMo alloy has a higher degree of corrosion resistance to sea water (containing chloride ions) under stress. The cold working can increase the strength of the alloy. The abrasive properties of this alloy are similar to the cast CoCrMo alloy. However the former is not recommended for the bearing surfaces of a joint. The superior fatigue and ultimate tensile strength of the wrought CoNiCrMo alloy make it very suitable for applications requiring long service without fracture or stress failure, thus used for stems of the hip joint prostheses. This advantage is more appreciated when the implant has to be replaced with another one since it is quite difficult to remove the failed piece of implant embedded deep in the femoral medullary canal.[7] As is the case with other alloys, the increased strength is accompanied by decreased ductility. Both the cast and wrought alloys have excellent corrosion resistance. The modulus of elasticity for the cobalt based alloys does not change with the change in their ultimate tensile strength. The castable CoCrMo alloy has been used in making artificial joints. The wrought CoNiCrMo alloy is a newcomer now used for making the stems of prosthesis for heavily loaded joints such as the knee and hip.[11]. 2.2.3. Titanium And Titanium Based Alloys

Attempts to use titanium for implant fabrication date to the late 1930s. it was found that titanium was tolerated in femurs as was stainless steel and Vitallium (CoCrMo alloy). Its lightness (4.5 g/cu cm compared to 7.9 g/cu cm for 316 stainless steel, 8.3 g/cu cm for cast CoCrMo, and 9.2 g/cu cm for wrought CoNiCrMo alloys) and good mechanical properties are salient features for implant application. Composition of Ti and Ti-Based Alloys

There are four grades of unalloyed titanium for surgical implant applications. The impurity contents separate them; oxygen, iron, and nitrogen should be controlled carefully. Oxygen in particular has a great influence on the ductility and strength.

4

141142143144145146147148149150151152153154155156157158159160161162163164165166167168169170171172173174175176177178179180181182183184185186187188189190

Page 5: Metallic Iplants

One Titanium alloy (Ti6Al4V) is widely used to manufacture implants. The main alloying elements of the alloy are Aluminum (5.5-6.5 %) and Vanadium (3.5-4.5 wt %).Chemical Composition of Titanium and its Alloys is shown in table 8.

Titanum Nickel Alloys

The nickel-titanium alloys show an unusual property, i.e., after the material is deformed it can snap back to its previous shape following heating the material. This phenomenon is called shape memory effect (SME. The equiatomic Ni- Ti alloy (Nitinol) exhibits an exceptional SME near room temperature

Properties of NiTi Alloys The SME can be generally due to martensite and autensite transformation At high temperatures, nitinol assumes an interpenetrating primitive cubic crystal structure referred to as austenite (also known as the parent phase). At low temperatures, nitinol spontaneously transforms to a more complicated monoclinic crystal structure known as martensite(daughter phase). This transformation is reversible.Some possible applications of shape memory alloys are orthodontic dental arch wires, intracranial aneurysm clips, a vena cava filter, and contractile artificial muscles for an artificial heart, orthopedic implants and other medical devices.Chemical Composition of Ni-Ti Alloys is shown in table 9

Structure and Properties of Ti and Ti-Based Alloys

Titanium is an allotropic material that exists as a hexagonal close packed structure (hcp, α-Ti) up to 882o C and body centered cubic structure (bcc, β-Ti) above that temperature. The addition of alloying elements to titanium enables it to have a wide range of properties.

1. Aluminum tends to stabilize the α-phase, that is increase the transformation temperature from α- to β-phase .2. Vanadium stabilizes the β-phase by lowering the temperature of the transformation from αto β.

Titanium derives its resistance to oxidation by the formation of a solid oxide layer. Under in vivo conditions the oxide (TiO2) is the only stable reaction product. The oxide layer forms a thin adherent film and passivates the material. [1] Titanium is mainly used in joint implants and dental implants.

2.2.4 Dental Metals2.2.4.1 Dental Amalgam An amalgam is an alloy in which one of the component metals is mercury. The rationale for using mercury as a tooth filling material is that since mercury is a liquid at room temperature, it can react with other metals such as silver and tin and form a plastic mass that can be packed into the cavity, and that hardens (sets) with time. To fill a cavity, the dentist mixes solid alloy, supplied in particulate form, with mercury in a mechanical triturator. The resulting material is readily deformable and is then packed into the prepared cavity. The solid alloy is composed of atleast 65% silver, and not more than 29% tin, 6% copper, 2% zinc, 3% mercury.

5

191192193194195196197198199200201202203204205206207208209210211212213214215216217218219220221222223224225226227228229230231232233234235236237238239240

Page 6: Metallic Iplants

2.2.4.2 Gold and gold alloys Gold and gold alloys are useful metals in dentistry as a result of their durability, stability and corrosion resistance. Gold fillings are introduced by two methods: casting and malleting. Cast restorations are made by taking a wax impression of the prepared cavity, making a mold from this impression in a material such as gypsum silica, which tolerates high temperature, and casting molten gold in the mold. The patient is given a temporary filling for the intervening time. Gold alloys are also used for cast restorations, since they have mechanical properties that are superior to those of pure gold. Corrosion resistance is retained in these alloys provided they contain 75% or more of gold and other noble metals. Copper, alloyed with gold, significantly increases its strength. Platinum also improves the strength, but no more than about 4% can be added, or the melting point of the alloy is elevated excessively. Silver compensates for the colour of copper.[1] A small amount of zinc may be added to lower the melting point and to scavenge oxides formed during melting. Gold alloys of different composition are available. Softer alloys containing more than 83% gold are used for inlays, which are not subjected to much stress. Harder alloys containing less gold are chosen for crowns and cusps which are more heavily stressed. Pure gold is relatively soft, so this type of restoration is limited to areas not subjected to much stress.

2.2.5 Other MetalsSeveral other metals have been used for a variety of specialized implant applications.Tantalum has been subjected to animal implant studies and has been shown very biocompatible. Due to its poor mechanical properties and its high density (16.6 g/cm3), it is restricted to few applications such as wire sutures for plastic surgeons and neurosurgeons and a radioisotope for bladder tumors.[8]Platinum and other noble metals in the platinum group are extremely corrosion resistant but have poor mechanical properties. They are mainly used as alloys for electrodes such as pacemaker tips and cochlear implants[10] because of their high resistance to corrosion and low threshold potentials.

3. CORROSION OF METALLIC IMPLANTS One of the main problem of metallic implants in human body is corrosion. Corrosion is the unwanted chemical reaction of a metal with its environment, resulting in its continued degradation to oxides, hydroxides, or other compounds.Tissue fluid in the human body contains water, dissolved oxygen, proteins, and various ions such as chloride and hydroxide. As a result, the human body presents a very aggressive environment to metals used for implantation. Corrosion resistance of a metallic implant material is consequently an important aspect of its biocompatibility.

3.1. Corrosion of Available Metals Choice of a metal for implantation should take into account the corrosion properties Metals that are in current use as biomaterials include gold, cobalt-chromium alloys, type 316 stainless steel, titanium, nickel-titanium alloys, and silver-mercury amalgam. The noble metals are immune to corrosion and would be ideal materials if corrosion resistance were the only concern. Gold is widely used in dental restorations and in that setting it offers superior performance and longevity. Gold is not, however, used in orthopaedic applications as a result of its high density, insufficient strength, and high cost. Titanium is a base metal in the context of the electrochemical series; however, it forms a robust passivation layer and, it remains passive under physiological conditions. Corrosion

6

241242243244245246247248249250251252253254255256257258259260261262263264265266267268269270271272273274275276277278279280281282283284285286287288289290

Page 7: Metallic Iplants

currents in normal saline are very low: 10-8 A/cm2. Titanium implants remain virtually unchanged in appearance. Ti offers superior corrosion resistance but is not as stiff or strong as steel. Cobalt-chromium alloys, like titanium, are passive in the human body. They are widely in use in orthopaedic applications. They do not exhibit pitting. Stainless steels contain enough chromium to confer corrosion resistance by passivity. The passive layer is not as robust as in the case of titanium or the cobalt-chromium alloys. Only the most corrosion resistant of the stainless steels are suitable for implants. These are the austenitic types 316, 316L, and 317, which contain molybdenum. Even these types of stainless steel are vulnerable to pitting and to crevice corrosion around screws.

3.2. Minimization of Corrosion

Although laboratory investigations are essential in the choice of a metal, clinical evaluation in follow-up is also essential. Corrosion of an implant in the clinical setting can result in symptoms such as local pain and swelling in the region of the implant, with no evidence of infection; cracking or flaking of the implant as seen on X-ray films, and excretion of excess metal ions. At surgery, grey or black discoloration of the surrounding tissue may be seen, and flakes of metal may be found in the tissue. Corrosion also plays a role in the mechanical failures of orthopaedic implants. Most of these failures are due to fatigue, and the presence of a saline environment certainly exacerbates fatigue.

Experience in the orthopaedic setting suggests that corrosion is minimized by the following:

l. Use appropriate metals.2. Avoid implantation of different types of metal in the same region. In the manufacturing process, provide matched parts from the same batch of the same variant of a given alloy.3. Design the implant to minimize pits and crevices.4. In surgery, avoid transfer of metal from tools to the implant or tissue. A avoid contact between metal tools and the implant, unless special care is taken.5. Recognize that a metal that resists corrosion in one body environment may corrode in another part of the body.

4. DISCUSSION

The strength and the modulus of elasticity of the metallic implants vary from metals to metals and type of metals used in alloy. Entire properties of the metal vary by slight variation in composition. The modulous of elasticity properties of commercially pure and 6Al4V alloys is about 110GPa, which is half the value of Co-based alloys. The higher impurity content leads to higher strength and reduced ductility. The strength of the material varies much lower than that of 316 stainless steel or the Co-based alloys to a value about equal to that of annealed 316 stainless steel of the cast CoCrMo alloy. However, when compared by the specific strength (strength per density) the titanium alloy excels over any other implant material. Titanium, nevertheless, has poor shear strength, making it less desirable for bone screws, plates, and similar applications. It also tends to gall or seize when in sliding contact with itself or another metal. The yield strength or yield point of a material is defined stress at which a material begins to deform plastically. Yield strength of various metals are shown in fig.1.

7

291292293294295296297298299300301302303304305306307308309310311312313314315316317318319320321322323324

325326327328329330331332333334335336337338

Page 8: Metallic Iplants

Metallic biomaterials widely for the fabrication of surgical implants due to high strength and resistance to fracture along with a relative ease of fabrication. They are used in joint replacement, hip/knee replacement, cochlear implants, cardiovascular implants, dental implants etc. the main dis advantage of metallic implants is the chance of corrosion.

REFERENCES1. Joon B. Park, Roderic S. Lakes. Bio Materials, An Introduction, second edition.2. M.J. Long and H.J. Rack, Titanium Alloysin Total Joint Replacement-A Material Science

Perspective, Biomaterials, Vol 19,1998, p 1621–16393. Nag Soumya and Banerjee Rajarshi. “Fundamentals of Medical Implant Materials “.

ASM Handbook,Volume 23,Materials for Medical Devices. Pages 6-174. Patitapabana Parida, Behera Ajit, Mishra Subash.” Classification of Biomaterials used in

Medicine”. International Journal of Advances in Applied Sciences. Vol. 1, No. 3, September 2012, pages. 125-129

5. Y. R. Yoo, H. H. Cho, S. G. Jang, K. Y. Lee, H. Y. Son, J. G. Kim, Y. S. Kim, Effect Of Co Content On The Corrosion Of High Performance Stainless Steels In Simulated Biosolutions, Key Engineering Materials, Vol. 342-343, 585-588, 2007.

6. U Kamachi Mudali, T M Sridhar and Baldev Raj, Corrosion of bio implants, Sadhana, Vol. 28, Parts 3 & 4, 601-637

7. Annual Book of ASTM Standards, Part 46, American Society for Testing and Materials, Philadelphia, 1980, p. 578.

8. Joseph D. Bronzino.” The Biomedical Engineering HandBook, Second Edition.” Chapter 37

9. Roger Narayan. “Biomedical Materials” springer publication, Academic edition 2009. Page 42-81

10. Tykocinski M, Duan Y, Tabor B, Cowan RS.” Chronic electrical stimulation of the auditory nerve using high surface area (HiQ) platinum electrodes.”. Hear Res. 2001 Sep. pp53-68.

11. T. M. Devine and J. Wulff, "Cast vs. Wrought Cobalt-Chromium Surgical Implant Alloys," J. Biomed. Mater. Res., 9, 151-167, 1975.

12. C.J. E. Smith and A. N. Hughes, ''The Corrosion Fatigue Behavior of a Titanium-6 w/o Aluminum-4 w/o Vanadium Alloy," Eng, Med., 7, 158-171, 1966.

13. J. H. Dumbleton and J. Black, An Introduction to Orthopaedic Materials, Charles C. Thomas, Springfield, III., 1975.

APPENDIX:

List of tables and figures

Tables

8

339340341342343344345346347348349350351352353354355356357358359360361362363

364365366367368369370371372373

374

375

376

377

378

379

380

Page 9: Metallic Iplants

Table 1 Uses of BiomaterialsProblem Area ExamplesReplacement of diseased or damaged part

Artificial hip joint, kidney dialysis machine

Assist in healing Sutures, bone plates and screws

Improve function Cardiac pacemaker, contact lens

Correct functional abnormality

Harrington spinal rod

Correct cosmetic problem

Augmentation mammoplasty, chin augmentation

Aid to diagnosis Probes and catheters

Aid to treatment Catheters, drains

Table 2 Biomaterials in OrgansOrgan ExamplesHeart Cardiac pacemaker,

artificial heart valveLung Oxygenator machineEye Contact lens, eye lens

replacementEar Artificial stapes,

cosmetic reconstruction of outer ear

Bone Bone plateKidney Kidney dialysis

machineBladder Catheter

Table 3 Biomaterials in Body SystemsSystem ExamplesSkeletal Bone plate, total joint

replacementsMuscular SuturesDigestive SuturesCirculatory Artificial heart valves, blood

vesselsRespiratory Oxygenator machineIntegumentary Sutures, burn dressings,

artificial skin

9

381

382383

384385386

Page 10: Metallic Iplants

Urinary Catheters, kidney dialysis machine

Nervous Hydrocephalus drain, cardiac pacemaker

Endocrine Microencapsulated pancreatic islet cells

Reproductive Augmentation mammoplasty, other cosmetic replacements

Table 4 Materials for use in body

type example Advantages Disadvantages

use

Polymers Nylon, Silicones, Teflon, Dacron

No additional surgery required for implant removal.

Not strong

May degrade

Deform with time

Sutures,blood vessels, hip socket, ear, nose, other soft tissues

No permanent implant in the body.Safe and biocompatible material, no risk of metal allergic reactionsReduced traumaNo long-term implant palpabilityCompatible with Magnetic Resonance Imaging (MRI Reduced radiographic scatter/obstruction

Metals Titanium, Ti alloys,Stainless steel,Co-Cr alloys,Gold

Easy to fabricate May corrode ,Dense, Brittle , Difficult to make ,Not resilient, difficulty in MRI imaging

Joint replacement, bone plates & screws, dental root implants

Strong, toughDuctile

Ceramics Aluminum oxide,Hydroxyapatite

Strong in compression Brittle, Difficult to make, Not resilient

Dental, hip socketStrong, tailor made

Not strongDeform with time

Composites carbon-carbon, wire or fiber reinforced bone cement

Strong Difficult to make

Joint implants, heart valves

tailor-made

10

387388389

Page 11: Metallic Iplants

Table 5. metal implants and applications

Application Implant materialCochlear implant Platinum electrodesCardiovascular implants (heart valve) Stainless steelJoint replacement Ti, Ti Alloys, Stainless steelBone plate for fracture fixation Stainless steel. Cobalt alloysDental implant titanium

Table 6. Compositions of 316 and 316L Stainless Steels

Element compositionCarbon 0.03 maxManganese 2.00 maxPhosphorus 0.03 maxSulphur 0.03 maxSilicon 0.75 maxChromium 17.00–20.00Nickel 12.00–14.00Molybdenum 2.00–4.00 Table 7. Chemical Composition of CoCr Alloys

Table 8. Chemical Composition of Titanium and its Alloys.

Element Grade 1 Grade 2 Grade 3 Grade 4 Ti6Al4V*

Nitrogen 0.03 0.03 0.05 0.05 0.05Carbon 0.10 0.10 0.10 0.10 0.08Hydrogen 0.015 0.015 0.015 0.015 0.0125Iron 0.20 0.30 0.30 0.50 0.25Oxygen 0.18 0.25 0.35 0.40 0.13

11

390

391

392

393394

395396

397398399400

Page 12: Metallic Iplants

Titanium balance

* Aluminum 6.00% (5.50~6.50), vanadium 4.00% (3.50~4.50), andother elements 0.1% maximum or 0.4% total.

Table 9. chemical composition of NitinolElement Composition (wt%) Composition %Ni 54.01Co 0.64Cr 0.76Mn 0.64Fe 0.66Ti Balance

Figures

Fig.1. Yield strength of various metallic implants.

12

401402

403

404

405

406

407408409410