med associates product manual€¦ · the rotorat™ main menu ... figure 2.4 - computer with...

27
instrumentation and software for research ROTORAT™ SOF-801 USER’S MANUAL DOC-037 Rev. 2.8 Copyright ©2018 All Rights Reserved Med Associates, Inc. P.O. Box 319 St. Albans, Vermont 05478 Phone: 802.527.2343 Fax: 802.527.5095 www.med-associates.com

Upload: others

Post on 22-Jun-2020

1 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

instrumentation and software for research

ROTORAT™ SOF-801 USER’S MANUAL

DOC-037 Rev. 2.8 Copyright ©2018 All Rights Reserved Med Associates, Inc. P.O. Box 319 St. Albans, Vermont 05478 Phone: 802.527.2343 Fax: 802.527.5095 www.med-associates.com

Page 2: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

This page intentionally left blank

Page 3: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Table of Contents

Chapter 1 | Introduction ................................................................................................ 1

Chapter 2 | Hardware ..................................................................................................... 2

Hardware Guide ............................................................................................................................. 2

Cable Guide .................................................................................................................................... 3

Wiring Instructions ........................................................................................................................ 4

Chapter 3 | Software ...................................................................................................... 6

Installing the RotoRat™ Software ............................................................................................... 6

The RotoRat™ Main Menu ............................................................................................................ 6

File Menu Options .......................................................................................................................... 7

Data Directory .............................................................................................................................. 8

Experiment Menu Options .......................................................................................................... 10

View Menu Options ...................................................................................................................... 12

Help Menu Options ...................................................................................................................... 12

Chapter 4 | Running Experiments ............................................................................. 13

Hardware Configuration .............................................................................................................. 13

Experiment Configuration ........................................................................................................... 13

Loading an Animal ....................................................................................................................... 14

Adjusting the Swivel Spring Length .......................................................................................... 15

Starting the Boxes ....................................................................................................................... 15

Chapter 5 | Data Files ................................................................................................... 16

Appendix A | Chamber Assembly ............................................................................... 18

Appendix B | Contact Information............................................................................. 23

- iii - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 4: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Notes _____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

_____________________________________________________________________________________

Diagrams

- iv - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 5: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

CHAPTER 1 | INTRODUCTION The RotoRat™ system has been designed to provide the maximum flexibility in both hardware and software for the measurement of rotational movement or activity of a rat.

Key features include:

• Graphical representation of the animal’s position at all times;

• Measure both clockwise and counterclockwise rotations; • Three levels of sensitivity;

• Records direction changes. Requirements:

• Windows XP, Vista, 7, 10, 32-bit or 64-bit • MED-RRS RotoRat™ Starter Package

Figure 1.1 - RotoRat™ Chamber

- 1 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 6: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

CHAPTER 2 | HARDWARE

Hardware Guide

Figure 2.1 - RotoRat™ Chamber

Figure 2.2 - Interface Cabinet with DIG-700G Decoder Card and DIG-713A SuperPort Card

Figure 2.3 - ENV-500A Junction Box

- 2 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 7: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Figure 2.4 - Computer with DIG-704 Interface Card

Cable Guide

Figure 2.5 - DIG-700C Ribbon Cable

Figure 2.6 - SG-218-10 Cable

DIG-704 Interface Card

- 3 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 8: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Wiring Instructions

If a complete system with computer was purchased from MED Associates, Inc, then the DIG-704 Interface Card and driver will already be installed. If using a computer purchased elsewhere, first install the DIG-704 Interface Card in any available PCI slot in the computer cabinet following the directions included with the computer. The DIG-704 Interface Card driver is included on the RotoRat™ software CD. After installing the DIG-704 Interface Card into the computer, turn on the computer and follow the Hardware Wizard prompts to select and install the DIG-704 driver.

Next, connect each RotoRat™ chamber to the ENV-500A Junction Box. Connect Chamber 1 to connector 1, Chamber 2 to connector 2, etc.

Figure 2.7 - Connect Chambers to Junction Box

- 4 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 9: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Connect the ENV-500A Junction Box to the DIG-713A SuperPort Input Card (located in the interface cabinet) using the SG-218-10 Serial Port Cable, as shown in Figure 2.8.

Figure 2.8 - Connect the Junction Box to the SuperPort Input Card

Connect the DIG-700G Decoder Card (located in the interface cabinet) to the DIG-704 Interface Card (located in the computer) using the DIG-700C Ribbon Cable, as shown in Figure 2.9.

Figure 2.9 - Connect the Decoder Card to the Interface Card

- 5 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 10: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

CHAPTER 3 | SOFTWARE

Instal l ing the RotoRat™ Software

Insert the RotoRat™ CD into the CD-ROM drive. If the installation program does not automatically start, right-click on the Start button and choose “Explore” or “Open Windows Explorer.” Browse to the CD Rom Drive and double-click Autorun.exe. Follow the prompts to install the program. Register your software with Med Associates to obtain software password.

The RotoRat™ Main Menu

Double click on the RotoRat™ desktop icon, or open the Windows Start menu and select All Programs | Med Associates | RotoRat. Select RotoRat to start the program.

Upon opening the RotoRat™ software application, the Main Menu screen as shown in Figure will be displayed.

Figure 3.1 - The RotoRat™ Main Menu

- 6 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 11: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

The RotoRat™ menu bar contains four menus. These menus and options are:

Table 3.1 - Main Menu Options Defined

Option Description

File Change the Hardware Configuration parameters, import old Hardware/Software configuration files, and Exit the program.

Experiment Change the Experiment Configuration parameters and Start/Stop Experiments.

View Select which Box to display in the large graph, select which Boxes to display in the left portion of the Window, and turn on/off the large graph.

Help Get information about the program version.

Fi le Menu Options

Configuration

The Configuration menu option opens the Hardware Configuration window, shown in Figure .

Figure 3.2 - Hardware Configuration Options

- 7 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 12: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Table 3.2 - Configuration Options Defined

Option Description

Data Directory The directory where data files will be saved. Selecting the Browse button and navigating to the desired folder will change the Data Directory.

Data Reporting Mode Data may be accumulated in either Absolute (A) or Relative (R) modes. In Absolute mode the counters continue to increment in each Time Block with the last Time Block providing a total count for the entire trial time. In Relative mode the counters are reset following each Time Block. Total counts may be obtained by adding the incremental counts in each Block.

Number of Boxes The number of RotoRat chambers available to run experiments in.

Resolution Length of time between each measurement. This value is set at 2ms and cannot be changed.

Automatic File Naming If selected, then the file name will be automatically generated using the values that are entered in the Subject, Experiment, and Group fields on the Experiment Configuration form. If not selected, then the user may enter the file name in the Experiment Setup Options dialog box.

Box The chamber for which the Port, Offset, and Bit values is being configured.

Port The input Port that corresponds to the setting on the interface card. All Superport input cards have the port address of 789.

Offset The Offset address that corresponds to the setting on the interface cards. Valid values are from 0 – 254 for SuperPort cards. Refer to the Superport manual for information on configuring Superport cards.

Bit There are two Bit values for each Box that corresponds to the input values on the interface card where the inputs are connected to the control panel. Refer to the Superport manual for bit values. Enter the lower bit value in the top box and the higher bit value in the bottom box.

Auto Configure Opens the Auto Port Configuration window, which allows the user to enter the starting Port and Offset for Box 1. These values are then used to calculate the Port, Offset, and Bit values for the rest of the Boxes. It is recommended that the default values obtained using the Auto Configure button are used.

Previous Load the Port, Offset, and Bit values for the Previous Box.

Next Load the Port, Offset, and Bit values for the Next Box.

Load Load a saved Hardware configuration from the Windows Registry.

Save Save the current Hardware configuration to the Windows Registry.

Delete Delete a previously saved Hardware configuration from the Windows Registry.

OK Save all changes and exit the window.

Cancel Discard all changes and exit the window.

- 8 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 13: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Figure 3.3 - Auto Port Configuration Options

The Auto Port Configuration window allows the user to enter the starting Port and Offset for Box 1. These values are then used to calculate the Port, Offset, and Bit values for the rest of the Boxes.

NOTE: In order for this to work each Box must be setup in sequential order by Bit, Offset, and then Port. This is the normal setup when using SuperPort cards.

Table 3.3 – Auto Port Configuration Options Defined

Option Description

Card Type Select SuperPort if using a SuperPort input card. Select Standard if using a Standard or SmartCtrl card.

Starting Port Number The Port number of the first Box.

Starting Offset The Offset address of the first Box.

OK Calculate the Port, Offset, and Bit values for all Boxes using the entered values.

Cancel Discard all changes and exit the window.

Import

The Import command allows the user to import configuration files from DOS versions of the RotoRat™ program. Because the DOS version saved the hardware and experiment configuration information into the same file, the Import command has two menu items:

Old Configuration

Imports just the hardware configuration from the DOS configuration files.

Old Experiment

Imports just the experiment configuration from the DOS configuration files.

Exit

The Exit command exits the program.

- 9 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 14: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Experiment Menu Options

Start Experiment

The Start Experiment command opens the Experiment Setup Options window for Box 1. This window allows the user to select the Trial Time, Block Time, etc. for each Box. A different experiment may be run in each box.

Figure 3.4 - Experiment Setup Options

Table 3.4 - Experiment Setup Options Defined

Option Description

File Name The file name that will be used to save the data. The file will be saved in the Data Directory that was selected on the Hardware Configuration window. If the Automatic File Naming option was selected, then this field will be filled in automatically using values from the Subject, Experiment, and Group fields listed below. If this option was not selected, then the user may enter in any valid file name.

Subject Can be up to 25 characters in length and may also be used to help create the File Name above.

Experiment Can be up to 30 characters in length and may also be used to help create the File Name above.

Group Can be up to 25 characters in length and may also be used to help create the File Name above.

Comments Used to help describe the Experiment that is being run. Because this field can be very large it is not saved to the Windows Registry with the rest of the fields. The comments appear in the data file.

Trial Time (min) The time in minutes that the experiment will run. Note: Data is not saved to the disk until the end of the Trial Time. This means that excessively long trial times may be impractical. At the end of the trial time the data is automatically saved to the output file specified. If more than one Box uses the same File Name, then the data will be appended to the existing file when the trial is completed.

- 10 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 15: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Option Description

Block Time (sec) The time in seconds that RotoRat will continue to increment the data counters before saving them in memory. Each Block produces a single line in the data file. At the end of each Block the data counters are reset to zero if the Relative (R) data option has been selected (See Data Reporting Mode).

Counts/Rev In addition to providing a count of complete (Full) revolutions of 360 degrees (without changing direction) and a count representing the smallest degree of movement interpreted by the optical encoder (Max), RotoRat provides a user selectable count which may be set from 2 counts per revolution (one count for a 180 degree rotation) to 32 counts per revolution (one count for a 11.25 degree rotation). Use the following table to determine the best “sensitivity.” The value of the counter controlled by this setting appears in the data file under the “Partial” heading.

Counts/Revolution Degrees Rotation/Count Data File

1 360 Full

2 180

4 90

8 45 Partial

16 22.5

32 11.25

64 5.625 Max

Retrace Determines the sensitivity of the system to direction changes. A setting of 1 is the most sensitive, requiring only 5.625 degrees of movement to register a change. This can happen from random head movements while the animal is rotating and will result in zero full rotations. Any number from 1 to 16 may be entered with each increment adding an additional 5.625 degrees. A setting of 8 requires 45 degrees of movement in the opposite direction before a direction change is registered. Until a direction change is recognized the “zero” reference point for Partial and Full rotations remains fixed. Once a direction change is recognized the “zero” reference point becomes the point at which the direction change began.

Set All Boxes Like This Copies all values, except for the Subject value, to the other Boxes.

Previous Displays the values from the Previous Box in the sequence.

Next Displays the values of the Next Box in the sequence.

Load Load a saved Experiment configuration from the Windows Registry.

Save Save the current Experiment configuration to the Windows Registry.

Delete Delete a previously saved Experiment configuration from the Windows Registry.

OK Save all changes and open the Start Boxes window.

Cancel Discard all changes and exit the window.

- 11 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 16: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Start Boxes

Allows the User to select which Boxes to start.

Stop Boxes

Allows the User to select which Boxes to stop (i.e. End the experiment prematurely).

View Menu Options

Graph

Turns on or off the large graph. If the large graph has been turned off, then all of the smaller graphs will be visible at one time.

Select

Allows the user to select which Box to display in the large graph. This can also be accomplished by using the shortcut keys or by clicking on the desired Box with the mouse while the large graph is displayed.

Boxes

Allows the user to select which set of Boxes (1-8, 9-16, or 17-24) to display while the large graph is also being displayed. This can also be accomplished by moving the scroll bar.

Help Menu Options

About

Displays the About RotoRat™ window with information regarding the software version and copyright information.

Figure 3.5 - About RotoRat™ Window

- 12 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 17: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

CHAPTER 4 | RUNNING EXPERIMENTS

Hardware Configuration

Open the software by double-clicking on the RotoRat™ icon on the desktop. Begin by selecting File | Configuration. The screen shown in Figure will appear. Refer to Table for definitions of each of the Configuration Menu parameters.

Select the data directory, data reporting mode, number of boxes (or chambers) in the system, resolution and whether or not Automatic File Naming should be used. These configuration parameters apply to the entire system.

Note: The settings entered in the Input Configuration section of the screen are unique to each box. These settings can be entered automatically by selecting Auto Configure. The screen shown in Figure will appear. The options on the Auto Port Configuration screen are defined in Table .

Select the Input Card Type, Starting Port Number and Starting Offset. Once the Input Card Type is selected, the default values for the Starting Port Number and Starting Offset will automatically appear. These values are acceptable for all standard RotoRat™ systems. Click OK and the Input Configuration information for each box will appear on the Configuration Screen.

Once all of the necessary information has been entered into the Configuration screen, there are several options. This configuration can be saved for later use by clicking Save. The RotoRat™ software allows the user to save multiple configurations. Select Load to open a previously saved configuration. Click OK to save all changes and exit this screen or click Exit to discard all changes and exit this screen. Selecting Delete allows the user to remove any previously saved configurations.

Experiment Configuration

Select Experiment | Start Experiment to define the experiment parameters. The screen shown in Figure will appear. Notice that the number of the box being configured, in this case Box 1, appears at the top of the screen. Refer to Table for definitions of each of the experiment parameters. These parameters can be different for each box, or they can all be the same. If all of the boxes should have the same parameters it is only necessary to enter them once, then select Set All Boxes Like This.

If Automatic File Naming was selected on the Configuration screen, a file name will already appear on the Start Experiment screen. If not, enter the desired File Name, and then enter the Subject, Experiment and Group, as well as any desired Comments. Next enter the desired Trial Time, Block Time, Counts/Rev and Retrace values.

Click Next to advance to the next Box screen, or Previous to go back. Once the experiment parameters have been entered for each box, there are several options. As with the Hardware Configuration, the RotoRat™ software allows the user to save multiple Experiment Configurations. Click Save to save this experiment, or click Load to load a previously saved experiment. Selecting Delete allows the user to remove any unwanted experiments. Click OK to save all changes and open the Start Boxes screen or click Exit to discard all changes and exit this screen.

- 13 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 18: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Figure 4.1 - Start Boxes Screen

The Start Boxes screen allows the user to select which boxes to run. Select individual boxes, or select All to start all of the boxes in the system. Clicking OK will begin the experiment; clicking Cancel will close the Start Boxes screen.

NOTE: Do not start the boxes until the animals are loaded. Refer to the “Loading an Animal” section of this manual.

Loading an Animal

The animals should be loaded into the chambers prior to starting the boxes. Begin by removing the swivel from the chamber by loosening the upper set screw, shown in Figure 0.12. Next, select the appropriate rodent harness from Table . Place the harness on the animal and attach the harness to the Velcro at the end of the swivel. Secure the harness by looping a length of the included string through the hooks on the harness, as shown in Figure 4.2. Place the animal in the chamber and reattach the swivel to the chamber. The boxes may now be started.

Table 4.1 - Rodent Harness Sizes and Part Numbers

Animal Weight (grams) Rodent Harness Part Number

100 – 150 ENV-500JXS

150 – 250 ENV-500JS

250 – 350 ENV-500JM

300 – 500 ENV-500J

Figure 4.2 - Rodent Harness Side and RearView

- 14 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 19: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Adjusting the Swivel Spring Length

It may be necessary to adjust the length of the swivel spring. Too much slack in the swivel spring may result in erroneous data. Adjust the length of the spring by loosening the lower setscrew (Figure 0.12) and removing the spring from the coupler. Trim the spring to the desired length and install back into the coupler, being sure to tighten the setscrew adequately.

Figure 4.3 - Swivel Spring with Too Much Slack

Figure 4.4 - Properly Adjusted Swivel Spring

Starting the Boxes

Once the boxes are started, the screen shown in Figure will appear. All of the boxes that are running will appear down the left side of the screen. The rotational movement will be displayed in real time. Refer to View Menu Options for detailed information regarding different ways of viewing the boxes. At the end of the Trial, the boxes will automatically be stopped.

- 15 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 20: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

CHAPTER 5 | DATA FILES Definition of Rotation

The optical encoder used in RotoRat™ divides a 360-degree circle into 64 equal segments of 5.625 degrees each. Movement is defined from wherever the encoder is positioned when a Box is “Started.” The direction of movement is sensed when a rotation of 5.625 degrees occurs and the “Max” counter is incremented for clockwise or counterclockwise movement as appropriate. As long as rotation continues in the same direction, the same “Max” counter will continue to increment every 5.625 degrees, the corresponding “Partial” counter will increment at its set sensitivity (Counts/Revolution), and the “Full” counter will increment every 360 degrees. As soon as the direction is reversed, a new starting point is established, the “Change Direction” counter is incremented, and the “Max,” Partial,” and “Full” counters corresponding to the new direction.

In the Relative mode, all data counters are zeroed at the end of each Time Block, however, internal counters are not. This means an animal could rotate 350 degrees when a time slice ends and the “Max,” “Partial,” and “Full” counters would be set to zero. In the new time slice an additional rotation in the same direction of 10 degrees would increment 1 on the “Max” counter and would also be sufficient to complete a revolution for the “Full” counter to increment to 1, however, the “Max” counter would have to reach 65 before the “Full” counter could reach 2.

Relative Data File

Start Date: 05/17/2004 End Date: 05/17/2004 Start Time: 14:52:17 End Time: 15:02:17 Subject: LEO1 Experiment: AM10 Group: EXP Comment: Relative Data File Box #: 1 Block Time: 120 Seconds Counts/Rev: 4 (or) 90.00 Degrees/Count Retrace: 8 Data Mode: Relative Time | CW | CC | ====================================================================== Block | Max Partial Full | Max Partial Full | Change 120 213 8 0 96 2 0 19 120 177 6 0 96 2 0 15 120 221 11 2 58 2 0 6 120 322 19 4 103 4 1 5 120 129 6 1 20 1 0 5

- 16 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 21: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

Absolute Data File

Taking the identical input that created the previous data file, except for changing the comment and running RotoRat™ in the Absolute mode, produces the following data file:

Start Date: 05/17/2004 End Date: 05/17/2004 Start Time: 14:52:17 End Time: 15:02:17 Subject: LEO1 Experiment: AM10 Group: EXP Comment: Absolute Data File Box #: 1 Block Time: 120 Seconds Counts/Rev: 4 (or) 90.00 Degrees/Count Retrace: 8 Data Mode: Absolute Time | CW | CC | ====================================================================== Block | Max Partial Full | Max Partial Full | Change 120 213 8 0 96 2 0 19 240 390 14 0 192 4 0 34 360 611 25 2 250 6 0 40 480 933 44 6 353 10 1 45

600 1092 50 7 373 11 1 50

- 17 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 22: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

APPENDIX A | CHAMBER ASSEMBLY

1. Attach the enclosure braces to the base using the predrilled holes and eight of the included 6-32 x ½” screws, as shown in Figure 0.1.

Figure 0.1 - Base with Enclosure Braces Installed

2. Attach the supports to the base using the predrilled holes and the included #8 x ½” screws, as shown in Figure 0.2.

Figure 0.2 - Attach Supports to Base

- 18 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 23: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

3. Place the enclosure (metal bowl) on the enclosure braces and fix in place using the included brackets, the remaining four 6-32 x ½” screws and the four #6 nylon flat washers. The washers go between the bracket and the brace.

Figure 0.3 - Install Brackets to Secure Enclosure

4. Attach the latches to the three shield supports using the six 4-40 x ¼” screws, as shown in Figure 0.4.

Figure 0.4 - Attach Latches to Shield Supports

5. Attach two shield supports to the shield using the predrilled holes and eight of the included 6-32 x

3/8” screws, as shown in Figure 0.5. Be sure that the latches are oriented on the same end of the shield.

Figure 0.5 - Attach Shield Supports

- 19 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 24: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

6. Using the remaining shield support, form the shield into a cylinder and close using four of the 6-32 x 3/8” screws. Do not tighten the screws down, as the size of the cylinder may need to be adjusted once it is mounted on the enclosure.

Figure 0.6 - Complete the Shield Assembly

7. Place the shield over the enclosure and latch in place, as shown in Figure 0.7.

Figure 0.7 - Latch Shield in Place

- 20 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 25: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

8. Adjust the size of the shield and tighten the screws to keep in place, as shown in Figure 0.8.

Figure 0.8 - Tighten Shield Adjustment Screws

9. Mount the encoder on the crossbar using the two #2 lock washers and the two 2-56 x ¼” screws, as

shown in Figure 0.9.

Figure 0.9 - Mount Encoder

10. Secure the encoder cable with included cable clamp. Place the cable clamp around the encoder cable and fix it in place in the predrilled hole using the 6-32 x 3/8” screw through the #6 lock washer and the #6 flat washer.

Figure 0.10 - Secure Encoder Cable

- 21 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 26: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

11. Secure the crossbar to the supports using the four 8-32 x 5/16” screws through #8 lock washers, as shown in Figure 0.11.

Figure 0.11 - Install Crossbar

12. Attach the swivel coupler to the encoder and tighten the top setscrew using the included hex key, as shown in Figure 0.12.

Figure 0.12 - Attach Swivel Coupler to Encoder

13. The chamber is now completely assembled and should appear as shown in Figure 1.1.

- 22 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.

Page 27: Med Associates Product Manual€¦ · The RotoRat™ Main Menu ... Figure 2.4 - Computer with DIG-704 Interface Card Cable Guide Figure 2.5 - DIG-700C Ribbon Cable Figure 2.6 - SG-218-

MED A S SOCI ATES, I NC. SO F-8 01 ROTOR AT ™

APPENDIX B | CONTACT INFORMATION Please contact Med Associates, Inc. for information regarding any of our products.

For Technical questions, email [email protected].

For Sales questions, email [email protected].

Visit our website at www.med-associates.com.

- 23 - DOC-037 Rev. 2.8 Copyright © 2018 Med Associates, Inc.