mechanical aspects important for determining the final wire position accuracy

29
Mechanical aspects important for determining the final wire position accuracy Review of the New CERN Fast Wirescanner Juan Herranz

Upload: elaine

Post on 11-Feb-2016

41 views

Category:

Documents


0 download

DESCRIPTION

Review of the New CERN Fast Wirescanner. Mechanical aspects important for determining the final wire position accuracy . Juan Herranz . Wire Scanner general considerations Sources of uncertainty Strategy to minimize uncertainties Summary Conclusions. General considerations. HOME position. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Mechanical aspects important for determining the final wire position accuracy

Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast Wirescanner

Juan Herranz

Page 2: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

• Wire Scanner general considerations

• Sources of uncertainty

• Strategy to minimize uncertainties • Summary

• Conclusions

Page 3: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

General considerations

Peak lin. speed 20 m/sFork length 156 mmPeak ang. speed 128.20 rad/s Peak ang. acceleration 6700 - 10000 rad/s^2Tang. acceleration 95 – 159 gNormal acceleration 260 g

Fork

Shaft

R =

156

mm

Deceleration

Acceleration

Cons. Speed

SPS configuration

Beam

HOME position

IN positionSame speed and position but different pick acceleration

Page 4: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

disk angular origin

𝜃𝑑

𝐿 𝑓

Motor side

In this model we can determine the position of the point P as function ofthe angular position of the disk andthe length of the fork

𝑃Arm 1

Arm 2

Optical disk

Shaft

Wire

𝑅

General considerations

Page 5: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

MisalignmentDisk – forkForks-arms

Sources of uncertainty

Page 6: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Shaft twist reference

Shaft twist 𝑻= 𝑰 𝜶 Motor Torque

Sources of uncertainty

MisalignmentDisk – forkForks-arms

Acceleration and inertia produce a torque which produce a twist on the shaft

Page 7: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Shaft twist reference

Acceleration and inertia produce a torque which produce a twist on the shaft

Shaft twist

Shaft Twist 𝑻= 𝑰 𝜶 Motor Torque

Sources of uncertainty

MisalignmentDisk – forkForks-arms

Page 8: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Shaft twist reference

Because tangential acceleration the fork bends

Shaft twist

Shaft Twist

𝑭=𝒎𝒂

Motor Torque𝑻= 𝑰 𝜶

Sources of uncertainty

MisalignmentDisk – forkForks-arms

Page 9: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Shaft twist reference

Because tangential acceleration the wire have a deflection

Shaft twist

Shaft Twist

𝑭=𝒎𝒂

𝑻= 𝑰 𝜶 Motor Torque

𝑭=𝒎𝒂

Sources of uncertainty

MisalignmentDisk – forkForks-arms

Page 10: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Measured position

𝛼1 𝛼2

𝛼3

𝛼𝑑

𝛼𝑒

𝛿2

𝛿3“Deterministic” displacement due to the peak acceleration and theinertia of the system

𝛿𝑑

Ø 312Ø 140

𝛿1

α δTwist of the shaft (fork) 50 7.8Fork bending 256 40Wire deflection 128 20Twist of the shaft (disk) 112.0 17.5 Total deformation 547 85.3

𝛿𝑒

Shaft

DiskWire Path

Real position

𝛼𝑒=𝛼1+𝛼2+𝛼3−𝛼𝑑

𝛿𝑒=𝛿1+𝛿2+𝛿3−𝛿𝑑

FEA estimation

Analytical estimationFEA estimation

FEA estimation

Pick accel. 6700 rad/s^2Optical disk, 3 mm thicknessWire, 12 strands of 7um

Sources of uncertainty

Page 11: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Measured position

𝛼1 𝛼2

𝛼3

𝛼𝑑

𝛼𝑒

𝛿2

𝛿3

𝛿𝑑𝛿1

𝛿𝑒

Shaft Disk Wire Path

Real position

α δTwist of the shaft (fork) ± 50 ± 7.8Fork bending ± 256 ± 40Wire deflection ± 128 ± 20Twist of the shaft (disk) ± 112.0 ± 17.5 Vibration amplitude ± 547 ± 85.3

Sources of uncertainty

Due to the variation of the acceleration we have a vibrational behaviour which amplitude is + / - the value of the deformation

Page 12: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Beam-wire relative positionuncertainty +/- 10 mm

BeamWire

Fork

Shaft twist between arms not represented

Motion

𝛿𝑤𝑖𝑟𝑒

𝛿𝑤−𝑏

For a wire deflection () of 20 μm and a beam-wire relative position of +/- 10 mm

Sources of uncertainty

Systematic Statistical Total Misalignment ± 160 ± 25 ± 160 ± 25Pure deformations Vibrations ± 547 ± 85.3 ± 547 ± 85.3Rel. Beam-Wire pos. ± 32 ± 5 ± 32 ± 5

Total ± 160 ± 25 ± 579 ± 90.3 ± 739 ± 115.3

APROXIMATIVE

VALUES

Page 13: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Beam axisReal

𝜽

Sources of uncertainty

𝑷𝒆𝒓𝒓𝒐𝒓

𝑷 =|𝑬𝜽𝒘

𝜽𝒘 |+|𝐭𝐚𝐧 (𝜽 )𝑬𝜽|𝜽𝒘

𝑬 𝜽

Measured

𝜽𝒘𝒎

Relative error in the projection:

Error in position: Error in width:

At 20m/s the wire needs 15μs to travel300 μm

In 15μs an oscillation of 170 μm amplitudeand 1kHz can travel 8.5 μm

That represents a relative error 2.8%

Width error

|tan (30 ° ) 739𝑥10−6|=4.7 𝑥10− 3

Position error

Page 14: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Minimization strategies - Design

Optimization of stiffness andinertia to minimize deformations

See Sebastian’s presentation

Digital mock-up N. Chritin

Digital mock-up N. Chritin

Page 15: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

PhotodiodeLaser DisplacerBeam

expender DisplacerScanner

+/- 50 mm Calibration table

Minimization strategies - Calibration

time

diod

e sig

nal

IN OUT

Thanks to Jonathan Emery

Page 16: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Minimization strategies – Motion pattern

Displacement 25 μm 30 μm 35 μm Vibration amplitude 50 μm 8 μm 3.5 μmFactor 1 0.16 0.07

Wire fork simplified model

Pattern optimization can provide large improvements in the residual vibrations minimization

Page 17: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Minimization strategies – Vibration measurements

Wire resistance variation

Strain gauge

Wire

Gauge

Wire

Gauge

Page 18: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Carlos Daniel Morais Pereira

G = 1000 BS = ± 5V 3 kHz LP filter Fork speed: 10 m/s G = 1000 BS = ± 1V 3 kHz LP filterFork speed: 10 m/s R64 and R65 = 1kΩ

Strain Gauge (GF = 2.1)

WB

IA Amplification

Page 19: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Minimization strategies - Wire dynamic model

Strain gauge

Wire resistance

Wire analytical model

Wire deformations

ScannerFork ang. position

Mea

sure

men

ts

Excitation

Inpu

tRe

sults

ModelVerification

Dynamic model intended to provide quantitative experimental information about Residual vibration of the scanner

Page 20: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Summary

Uncertainty Minimization strategy Design Calibration Motion patternMisalignment No play in the system Correction by calibration

tables

Pure deformations Optimization stiffness and inertia

Correction by calibration tables

Smother profiles have more deterministic behaviour Residual Vibrations Optimization stiffness and

inertia Minimizing residual

vibrationsRel. Beam-Wire pos. Smother -> deterministic behaviour

Misalignment ± 32 ± 5Pure deformations ± 32 ± 5Vibrations ± 12 ± 2Rel. Beam-Wire pos. ± 32 ± 5Total ± 109 ± 17

Expected approximate values after uncertainties minimization:(the expectation is to attend values of the same order of magnitude)

Page 21: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

• The main sources of uncertainty for determining the wire position has been identified

• Actions to minimize this uncertainties has been take and integrated in the design

• Studies to minimize the influence of the motion pattern are ongoing

• Vibration measurement and quantification system are in phase of test and development using the exiting calibration test bench for PS and PSB scanners and this systems are going to be integrated in the first prototype

Conclusions

Many thanks !

Page 22: Mechanical aspects important for determining the final wire position accuracy
Page 23: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Angular speed: 90 rad/sAngular acceleration: 7000 rad/s^2

Tip deformation: 1.2 mm

Page 24: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

1st 133Hz 2nd 193Hz

2th 626Hz

Page 25: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Uncertainty Minimization strategy Design Calibration Motion patternMisalignment No play in the system Correction by calibration

tables

Pure deformations Optimization stiffness and inertia

Correction by calibration tables

Smother profiles have more deterministic behaviour Residual Vibrations Optimization stiffness and

inertia Minimizing residual

vibrationsRel. Beam-Wire pos. = minimize deformations

Total Misalignment ± 32 ± 5Pure deformations ± 32 ± 5Vibrations ± 12 ± 2Rel. Beam-Wire pos. ± 32 ± 5Total ± 109 ± 17

𝑃𝑒𝑟𝑟𝑜𝑟

𝑃 =|𝐸 𝜃𝑤

𝜃𝑤|+|tan (𝜃)𝐸 𝜃|

tan (30 ° )17 𝑥10−6=9.8 𝑥10−6

𝐸𝜃𝑤

𝜃𝑤=6.28𝑥10−4

𝑃𝑒𝑟𝑟𝑜𝑟=𝟑𝟎𝟎 𝑥6.38 𝑥10−4=𝟎 .𝟐

Page 26: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

15 m/sec check

Jonathan Emery

Typical results after calibration for PS scanner

Page 27: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Jonathan Emery

Page 28: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

Plots under discussion

Page 29: Mechanical aspects important for determining the final wire position accuracy

Review of the New CERN Fast WirescannerJuan Herranz 18.04.20139th DITANET Topical Workshop on Non-Invasive Beam Size Measurements for High Brightness Proton and Heavy Ion Accelerators

coils

wire

ferrite