manual hplc primaide hitachi

424
INSTRUCTION MANUAL FOR PRIMAIDE SYSTEM MANAGER (DETAILED OPERATION) This system is intended for research use only. It is not to be used for reporting patient diagnostic or therapeutic results. Please read through this manual carefully and keep it in a safe place. Before using the instrument, read the safety instructions and precautions carefully. Keep this manual in a safe place nearby so it can be referred to whenever needed. Hitachi High-Technologies Corporation 24-14, Nishi-Shimbashi 1-chome, Minato-ku, Tokyo, Japan Copyright C Hitachi High-Technologies Corporation 2012. 1st Edition, 2012 All rights reserved. Printed in China. Part No. 895-9635

Upload: aline

Post on 09-Nov-2015

490 views

Category:

Documents


47 download

DESCRIPTION

Manual HPLC Primaide Hitachi

TRANSCRIPT

  • INSTRUCTION MANUAL

    FOR

    PRIMAIDE SYSTEM MANAGER (DETAILED OPERATION)

    This system is intended for research use only.

    It is not to be used for reporting patient diagnostic or therapeutic results.

    Please read through this manual carefully

    and keep it in a safe place.

    Before using the instrument, read the safety

    instructions and precautions carefully.

    Keep this manual in a safe place nearby so it can

    be referred to whenever needed.

    Hitachi High-Technologies Corporation

    24-14, Nishi-Shimbashi 1-chome, Minato-ku, Tokyo, Japan

    Copyright C Hitachi High-Technologies Corporation 2012. 1st Edition, 2012

    All rights reserved. Printed in China. Part No. 895-9635

  • Be sure to read through and understand the following points

    with regard to this manual.

    1. Information contained in this document is subject to

    change without notice for improvement.

    2. This manual is copyrighted by Hitachi High-Technologies

    Corporation with all rights reserved.

    No part of this manual may be reproduced, transmitted

    or disclosed to a third party in any form or by any means

    without the express written permission of Hitachi

    High-Technologies Corporation.

    3. Hitachi High-Technologies Corporation assumes no

    liability for any direct, indirect, or consequential damages

    arising from use not described in this manual.

    Utmost care must be exercised when using the

    instrument.

    4. This document does not provide any warranty or

    permission for industrial properties or any rights to grant

    license lawfully and without infringement.

  • PREFACE

    Thank you very much for purchasing the Primaide System Manager.

    The Primaide System Manager is the Chromato data station for high

    performance liquid chromatography (HPLC).

    This software is intended for use by persons having a basic

    knowledge of chemical analysis.

    Remember that improper use of analytical instruments,

    chemicals or samples can result not only in wrong analytical

    data but also in consequences adverse to safety. Note that only

    persons with a basic knowledge of chemical analysis procedures

    may use this instrument.

    Carefully read this instruction manual before attempting operation. For

    proper use of the software, please acquaint yourself with it.

    ABOUT THIS MANUAL

    The manual of the Primaide System Manager is consists of ;

    (1) INSTRUCTION MANUAL FOR Primaide System Manager

    (INSTALLATION)

    (2) INSTRUCTION MANUAL FOR Primaide System Manager

    (ADMINISTRATION)

    (3) INSTRUCTION MANUAL FOR Primaide System Manager

    (DETAILE OPERATION)

    (4) INSTRUCTION MANUAL FOR Primaide System Manager

    (MAINTENANCE SOFTWARE)

    Manuals (2), (3) and (4) have been saved to the Instruction Manual

    CD as PDF files.

    These instruction manuals describe how to install and use the

    Primaide System Manager. The operating procedures and points

    that should be especially noted in regard to the software are

    contained in this manual.

    First of all, read "IMPORTANT" and "SAFETY SUMMARY" at the

    beginning of this manual to ensure safe operation of the

    Primaide System Manager.

    For operation of the Primaide System Manager read the each

    instruction manual.

  • IMPORTANT - 1

    IMPORTANT

    Warranty on Product

    The Primaide System Manager is warranted to be free from

    defects in material or workmanship under normal use within the

    product specifications indicated in this manual and under

    conditions given below. This warranty is void if the software is

    not used according to the instruction manual.

    The manufacturer makes no warranties, either express or implied,

    except as provided herein, including without limitation thereof,

    warranties as to marketability or merchantability, for a particular

    purpose or use, or against infringement of any patent.

    No oral or written information or advice given by the

    manufacturer, its dealers, distributors, agents or employees shall

    create a warranty or in any way increase the scope of this

    warranty.

    (1) Scope of Warranty

    Any parts that prove to be defective in design or

    workmanship during the warranty period will be repaired,

    adjusted or replaced without charge. A substitute part

    may be used for repair, or replacement with an equivalent

    product may be made instead of repair. Such system

    components as a personal computer and printer to be

    updated frequently for improvement may not be available in

    original versions at the time of replacement.

    Note that this warranty does not apply to the instrument

    after it is discarded, or if modified by the user or resold

    without permission from the manufacturer, consumable

    parts, and any failure of lifetime-expired parts.

    The manufacturer assumes no liability for any damage to

    data or application software due to any possible fault or

    failure of this instrument.

    (2) Warranty Period

    One year from the date of initial installation.

    (In case a separate warranty document has been issued,

    the warranty period indicated in it takes precedence over

    the above period.)

  • IMPORTANT - 2

    (3) Limitations and Exclusions on Warranty

    Note that the following cases are excluded from the scope

    of this warranty, i.e., these cases are beyond the coverage

    of free-of-charge repair even during the warranty period

    indicated above.

    (a) Failure due to operation at a place not meeting the

    installation requirements specified by the

    manufacturer.

    (b) Failure due to power supply voltage/frequency other

    than specified by the manufacturer or due to

    abnormality in power supply.

    (c) Corrosion or deterioration of the piping due to

    impurities contained in gas, compressed air or

    cooling water supplied by the user.

    (d) Corrosion of the electric circuits or deterioration of

    the optical elements due to highly corrosive

    atmospheric gas.

    (e) Failure due to use of software, hardware or spare

    parts not supplied by the manufacturer.

    (f) Failure due to use not described in the manual or

    improper repair not approved by the manufacturer.

    (g) Failure due to maintenance or repair by other than

    service personnel qualified by the manufacturer.

    (h) Failure due to relocation or transport conducted not

    under the supervision of the manufacturer after the

    initial installation of the instrument.

    (i) Failure due to disassembly, modification or relocation

    not approved by the manufacturer.

    (j) Failure due to acts of God, including fire, earthquake,

    storm, flood, lightning, social disturbance, riot, crime,

    insurrection, terrorism, war (declared or undeclared),

    radioactive pollution, contamination with harmful

    substances, etc.

  • IMPORTANT - 3

    (k) Failure of the hardware, or damage to the system

    software, application software or data due to

    computer virus infection.

    (l) After disposal of this instrument, after its resale

    without prior approval from the manufacturer,

    consumable parts, and failure of any part that have

    reached the end of its service life.

    (4) Disclaimer of Warranty

    THE MANUFACTURER MAKES NO WARRANTIES,

    EITHER EXPRESS OR IMPLIED, EXCEPT AS

    PROVIDED HEREIN, INCLUDING WITHOUT LIMITATION

    THEREOF, WARRANTIES AS TO MARKETABILITY,

    MERCHANTABILITY, FOR A PARTICULAR PURPOSE

    OR USE, OR AGAINST INFRINGEMENT OF ANY

    PATENT. IN NO EVENT SHALL THE MANUFACTURER

    BE LIABLE FOR ANY DIRECT, INCIDENTAL OR

    CONSEQUENTIAL DAMAGES OF ANY NATURE, OR

    LOSSES OR EXPENSES RESULTING FROM ANY

    DEFECTIVE PRODUCT OR THE USE OF ANY

    PRODUCT.

    NO ORAL OR WRITTEN INFORMATION OR ADVICE

    GIVEN BY THE MANUFACTURER, ITS DEALERS,

    DISTRIBUTORS, AGENTS OR EMPLOYEES SHALL

    CREATE A WARRANTY OR IN ANY WAY INCREASE

    THE SCOPE OF THIS WARRANTY.

  • IMPORTANT - 4

    Installation, Relocation and After-sale Technical Service

    (1) Installation and Relocation

    (a) Installation at delivery shall not be carried out by the

    user. It shall be carried out by our sales

    representative or the engineers who have been

    trained and qualified for this purpose by us in order to

    use the instrument safely and accurately.

    (b) Before installation, the user shall make preparations

    for satisfying the installation requirements in

    accordance with this instruction manual.

    (c) If relocation becomes necessary after initial

    installation (delivery), please contact the dealer from

    whom you purchased the instrument or our sales

    representative.

    (2) After-sales Service

    (a) For after-sales service, contact our sales

    representative or service office of our sales

    representative.

    (b) For service after the warranty period, consult us with

    regard to a maintenance and inspection service

    contract.

    Technical Seminars and Training Courses for Users

    We offer technical seminars and training courses at either our or

    users facilities to ensure proper and safe operation of the

    analytical instrument to its full performance. For further

    information, contact our sales representative. (Applicants will

    be charged.)

  • IMPORTANT - 5

    Other Precautions

    (1) Handling of Chemicals and Samples

    (a) The user is responsible for following relevant legal

    standards and regulations in handling, storage and

    disposal of chemicals and samples used in analytical

    operations with this instrument.

    (b) Reagents, standard solutions and accuracy-control

    samples shall be handled, stored and discarded as

    instructed by the respective suppliers.

    (2) Trademark Acknowledgements

    Primaide System Manager is registered trademarks of

    Hitachi High-Technologies Corporation.

    Microsoft, MS, Microsoft Excel, Microsoft Word, and

    Windows are either registered trademarks or trademarks of

    Microsoft Corporation in the United States and other

    countries.

    Windows 7 is a trademark of Microsoft Corporation (USA).

    All other trademarks are the property of their respective

    holders and are hereby acknowledged.

  • SAFETY - 1

    SAFETY SUMMARY

    General Safety Guidelines

    Before using the Primaide System Manager, be sure to read the

    following safety instructions carefully.

    The hazard warnings which appear on the warning labels on the

    product or in the manual have one of the following alert headings

    consisting of a safety alert symbol and signal word

    DANGER, WARNING or CAUTION.

    : Safety alert symbol used for calling

    attention to a potential hazard which

    could cause personal injury.

    To avoid possible injury or death,

    observe all the safety messages

    following this symbol.

    DANGER : Indicates an imminently hazardous

    situation which, if not avoided, will result

    in death or serious injury.

    WARNING : Indicates a potentially hazardous

    situation which, if not avoided, can

    result in death or serious injury.

    CAUTION : Indicates a hazardous situation which, if

    not avoided, can result in minor or

    moderate injury.

    NOTICE : Indicates a hazardous situation which, if

    avoided, can result in damage to

    property.

    In addition to the above, the following signal word is used to

    indicate instructions for ensuring proper use of the product.

    NOTE: Indicates an instruction for ensuring correct use

    of the product and accurate analysis therewith.

  • SAFETY - 2

    SAFETY SUMMARY

    Common Safety Precautions

    Prior to Use

    Before using the product, be sure to read this instruction

    manual carefully to attain a full understanding of its

    operations.

    Keep the instruction manual handy nearby so it can be

    referred to whenever needed.

    Be sure to observe the procedures specified in the manual.

    Be sure to understand and follow all the safety instructions

    given in the manual.

    Be sure to observe all the hazard warnings attached to the

    instrument or provided in the manual. Failure to do so could

    result in personal injury or damage to the instrument.

    Be sure to follow all the methods of use instructed in the

    manual for proper application of the product.

    Absolutely avoid modifying the product, using non-specified

    parts, or removing safety devices as it could be hazardous.

    Do not perform any operation or action other than described

    in the manual.

    On occurrence of any trouble in the instrument, notify the

    nearest sales representative or service office of Hitachi High-

    Technologies Corporation.

    When using chemicals for the instrument, be sure to provide

    proper ventilation of the room. Inadequate ventilation could

    endanger human health.

    Keep in mind that the hazard warnings in the manuals or on

    the product cannot cover every possible case, as it is

    impossible to predict and evaluate all circumstances

    beforehand. Always be alert and use your common sense.

  • SAFETY - 3

    SAFETY SUMMARY

    Common Safety Precautions (Continued)

    In Use

    If an abnormality such as unusual noise, odor, fuming or gas

    leakage occurs during operation of the instrument, immediately

    disconnect power to the instrument, and take proper safety

    measures as required. Then, notify the nearest Hitachi High-

    Technologies Corporation sales representative or service office

    of Hitachi High-Technologies Corporation sales representative.

    Installation, Maintenance, and Relocation

    At the time of delivery, installation of the instrument shall be

    carried out by or under supervision of qualified service

    personnel of the manufacturer or its authorized service agent

    for ensuring safety and high accuracy in operation of the

    instrument. It is not permitted for the user to carry out

    installation.

    After completion of installation, check that all the standard

    parts are equipped. If the instrument is made active with any

    one of the standard parts not equipped, a failure could occur

    to result in a hazardous condition.

    If any item is missing or damaged or if you have any question,

    notify the installation personnel at site or the nearest Hitachi

    High-Technologies Corporation sales representative or

    service office of Hitachi High-Technologies Corporation sales

    representative.

    The maintenance and checkup procedures to be taken by the

    user are only those described in the manual. When taking

    the maintenance and checkup procedures described in the

    manual, attain a clear understanding of them.

    Do not perform other maintenance and checkup procedures

    to avoid jeopardizing safety and causing troubles in the

    instrument.

  • SAFETY - 4

    SAFETY SUMMARY

    Common Safety Precautions (Continued)

    After installation, do not relocate the instrument. If the

    instrument is relocated, vibration or impact to be applied

    during relocation could cause a malfunction in the optical

    components that have been adjusted precisely.

  • SAFETY - 5

    SAFETY SUMMARY

    Safety Instructions in This Manual

    Shown below are the safety instructions contained in this manual

    and their relevant sections in it.

    DANGER Indications

    The indication DANGER does not apply to this product.

    WARNING Indications

    The indication WARNING does not apply to this product.

    CAUTION Indications

    Fatigue due to Long-Hour Operation

    If you keep working with the display monitor and keyboard for

    long hours, your eyes and body will be fatigued to jeopardize

    your health. When working with the display monitor for a long

    time, take a break for 10 to 15 minutes per hour for health of

    your eyes and body.

    (Chapter 1)

  • SAFETY - 6

    SAFETY SUMMARY

    NOTICE

    Accuracy and Precision of Measured Values

    Carry out control sample measurements to ensure that the

    performance of the instrument is normal.

    Data Backup

    Data on the hard disk may become unusable due to a system

    failure, wrong operation, computer virus infection, etc.

    To ensure data integrity in case of accidental damage to the hard

    disk, periodically make backup copies of hard disk files onto

    floppy disks. Be sure to carry out this data backup procedure

    periodically.

    Power Interruption

    On occurrence of momentary power voltage drop due to power

    interruption or lightning, the personal computer may become

    faulty or the basic software, application software or data may be

    damaged.

    For protection against momentary power voltage drop, it is

    advisable to use an AC uninterruptible power supply unit (stated

    according to the Japanese Electronic Industry Development

    Association guidelines for protection against momentary power

    voltage drop in personal computers).

  • SAFETY - 7

    SAFETY SUMMARY

    NOTICE (Continued)

    About Personal Computer

    About handling of the personal computer, read the manual

    that is standard-furnished for personal computer and use as

    directed.

    When using this product, be sure to observe the warnings

    and cautions shown by the personal computer manufacturer.

    Do not turn off the personal computer main circuit while the

    hard disk or removable disk drive is active. If power to the

    personal computer is turned off while the hard disk or

    removable disk is being accessed, the personal computer

    may become faulty or data/software stored in it may be

    damaged. Before turning off power to the personal

    computer, quit the Primaide System Manager and then take

    the shutdown sequence of the operating system of the

    personal computer.

    Do not locate the personal computer on an unstable place or

    a narrow place. Failure to do so could result in personal

    injury or damage to the personal computer.

    Do not spill the solvent on the personal computer because

    there is a possibility that the personal computer fails. Should

    you spill the solvent on the personal computer, wipe it off

    immediately.

    If you use as it is, it cause an electric shock, release of fume

    and ignition.

    Do not use the volatile liquid near the personal computer

    because of the possibility of ignition hazard.

  • SAFETY - 8

    SAFETY SUMMARY

    NOTICE (Continued)

    Protection against Computer Viruses

    If any program/data is damaged suddenly or an unexpected

    operation/screen is encountered, the personal computer may be

    infected by a computer virus. Computer viruses are malicious

    programs that sneak into personal computers to cause

    misbehavior or damage to data. And, a program designed to

    offer protection against computer viruses is called a vaccine

    program.

    Possible causes of virus infection are:

    (a) Downloading a virus-laden program through

    communication.

    (b) Using a floppy disk or other storage medium infected by a

    virus.

    Note also that once a virus infects any personal computer, it

    may spread to other computers via communication or storage

    medium.

    Never use a program or storage medium that is suspected of

    containing a virus.

    If there is a possibility of virus infection, check for a virus using a

    vaccine program. Note, however, that some kinds of vaccine

    programs cannot eradicate particular viruses.

    In such a case, be sure to make a backup of hard disk files.

    The user is requested to prepare a vaccine program and carry

    out virus removal on his or her responsibility.

  • - i -

    CONTENTS

    PREFACE

    ABOUT THIS MANUAL

    IMPORTANT ....................................................................................................... IMPORTANT-1

    Warranty on Product ................................................ IMPORTANT-1

    Installation, Relocation and After-sale

    Technical Service ..................................................... IMPORTANT-4

    Technical Seminars and Training Courses for

    Users ....................................................................... IMPORTANT-4

    Other Precautions .................................................... IMPORTANT-5

    SAFETY SUMMARY .................................................................................................. SAFETY-1

    General Safety Guidelines ....................................... SAFETY-1

    Common Safety Precautions ................................... SAFETY-2

    Safety Instructions in This Manual ........................... SAFETY-5

    DANGER Indications ..................................... SAFETY-5

    WARNING Indications................................... SAFETY-5

    CAUTION Indications .................................... SAFETY-5

    Fatigue due to Long-Hour Operation ............ SAFETY-5

    NOTICE .......................................................................... SAFETY-6

    1. OVERVIEW ..................................................................................................................... 1-1

    1.1 The Primaide System Manager ...................................... 1-1

    1.2 The Main Function .......................................................... 1-2

    2. STARTING/EXITING THE PRIMAIDE SYSTEM MANAGER ........................................... 2-1

    2.1 Introduction .................................................................... 2-1

    2.2 Starting/Exiting the Primaide System Manager ........... 2-2

    2.3 Menu ................................................................................ 2-3

    2.3.1 File Menu ............................................................. 2-3

    2.3.2 Edit Menu ............................................................. 2-7

    2.3.3 Window Menu ...................................................... 2-8

    2.3.4 Help Menu ............................................................ 2-8

    3. PRIMAIDE SYSTEM MANAGER MAIN WINDOW .......................................................... 3-1

    3.1 Menu Command Function of Main Window ................. 3-3

    3.1.1 File Menu ............................................................. 3-3

    3.1.2 Help Menu ............................................................ 3-4

    3.2 Main Tool Bar Icons ....................................................... 3-5

    3.2.1 Change Application Icon ...................................... 3-6

    3.2.2 Method Setup Icon ............................................... 3-6

  • - ii -

    3.2.3 Sampler Setup Icon .............................................. 3-8

    3.2.4 Acquire Data Icon ................................................. 3-8

    3.2.5 Reprocess Data Icon ............................................ 3-9

    3.2.6 Report Icon ........................................................ 3-10

    3.2.7 System Status Icon ............................................ 3-11

    3.2.8 Pump(A) ON/OFF and Pump(B) On/Off Icon ..... 3-13

    3.2.9 Module Detailed Information Icon ....................... 3-13

    3.2.10 Quick Analysis Start Icon ................................... 3-13

    4. FUNCTION AND OPERATION OF METHOD WINDOW ................................................. 4-1

    4.1 Menu Command Function of Main Window ................. 4-1

    4.1.1 File Menu ............................................................. 4-2

    4.1.2 Edit Menu ............................................................. 4-2

    4.1.3 Module Setup Menu ............................................. 4-3

    4.1.4 Data Process Setup Menu ................................... 4-4

    4.1.5 Option Menu......................................................... 4-5

    4.1.6 Window Menu ...................................................... 4-7

    4.1.7 Help Menu ............................................................ 4-8

    4.2 Using a Method ............................................................... 4-9

    4.2.1 Opening an Existing Method .............................. 4-10

    4.2.2 Creating a New Method ...................................... 4-11

    4.2.3 Adding or Modifying Method Parameters ........... 4-12

    4.2.4 Using Reference Methods .................................. 4-12

    4.2.5 Saving a Method ................................................ 4-14

    4.3 Module Setup Menu ...................................................... 4-15

    4.3.1 Method Information ............................................ 4-15

    4.3.2 Method Configuration ......................................... 4-16

    4.3.3 Pump Setup ....................................................... 4-18

    4.3.4 Autosampler Setup ............................................. 4-20

    4.3.5 Column Oven Setup ........................................... 4-22

    4.3.6 Channel 1 or 2 Detector Setup ........................... 4-23

    4.3.7 1410 UV Detector Setup .................................... 4-23

    4.3.8 USB-AID Setup (AID: Analog Input Device) ....... 4-26

    4.3.9 1430 DAD Setup ................................................ 4-28

    4.3.10 Time Summary ................................................... 4-30

    4.4 Data Processing Setup Menu ...................................... 4-31

    4.4.1 Channel 1 or Channel 2 ..................................... 4-31

    4.4.2 Calculation Method ............................................. 4-31

    4.4.3 Component Table ............................................... 4-36

    4.4.4 Concentration Table ........................................... 4-40

    4.4.5 Coefficient Table ................................................ 4-42

    4.4.6 Integration Time Table ....................................... 4-44

    4.4.7 DAD Data Processing ........................................ 4-47

    4.4.8 Chromatogram Display Format .......................... 4-49

    4.4.9 DAD Data Display .............................................. 4-52

  • - iii -

    4.4.10 Confidence Report ............................................. 4-54

    4.4.11 Report Format .................................................... 4-56

    5. SAMPLE TABLE ............................................................................................................. 5-1

    5.1 Sample Table setup ........................................................ 5-1

    5.1.1 File Menu ............................................................. 5-2

    5.1.2 Edit Menu ............................................................. 5-3

    5.1.3 SampleSetup Menu and Sample Table Tool Bar .. 5-4

    5.1.4 Window Menu ...................................................... 5-4

    5.1.5 Help Menu ............................................................ 5-4

    5.2 Using Sample Table ....................................................... 5-5

    5.2.1 Opening an Existing Sample Table ...................... 5-5

    5.2.2 Creating a Single-Method Sample Table .............. 5-5

    5.2.3 Creating a Multi-Method Sample Table ................ 5-7

    5.2.4 Editing a Sample Table During Data Acquisition .. 5-8

    5.2.5 Saving a Sample Table ........................................ 5-9

    5.3 Sample Table Window .................................................. 5-10

    5.3.1 Setup Information Screen................................... 5-10

    5.3.2 Edit Table Screen ............................................... 5-14

    6. ACQUIRING DATA .......................................................................................................... 6-1

    6.1 Data Acquisition Window............................................... 6-2

    6.1.1 File Menu ............................................................. 6-3

    6.1.2 Acquire Menu ....................................................... 6-4

    6.1.3 Data Display Menu ............................................... 6-9

    6.2 Running Data Acquisition ............................................ 6-12

    6.2.1 Initiating the Idle Monitor .................................... 6-13

    6.2.2 Starting a Run .................................................... 6-15

    6.2.3 Starting Series ................................................... 6-16

    6.2.4 Performing a Manual and Auto Noise Test ......... 6-20

    6.2.5 Editing a Sample Table During Data Acquisition 6-22

    6.2.6 Data Acquisition from arbitrary line of

    Sample Table ..................................................... 6-23

    6.2.7 Data Reprocess of Data Series of incomplete

    Starting series .................................................... 6-24

    6.3 Performing On-Line Data Processing ......................... 6-25

    6.3.1 Printing of a Report ............................................ 6-25

    6.3.2 Statistical Calculation ......................................... 6-25

    6.3.3 Calibration .......................................................... 6-26

    6.3.4 Blank Subtraction ............................................... 6-26

    6.3.5 DAD Data Processing ........................................ 6-27

    6.3.6 Data Processing for Two Detectors .................... 6-27

    6.4 Quick Analysis Start ..................................................... 6-28

    6.4.1 Setting Parameter .............................................. 6-28

  • - iv -

    7. FUCTION AND OPERATION OF DATA-PROCESSING CONTROL WINDOW .............. 7-1

    7.1 Configuration of Data-Processing Control Window .... 7-1

    7.1.1 File Menu ............................................................. 7-2

    7.1.2 Edit Menu ............................................................. 7-2

    7.1.3 Option Menu......................................................... 7-3

    7.1.4 Window Menu ...................................................... 7-4

    7.1.5 Help Menu ............................................................ 7-4

    7.2 Operation of Data-Processing Control Window ........... 7-5

    7.2.1 Overview .............................................................. 7-5

    7.2.2 Injection Table ...................................................... 7-5

    7.2.3 Function Buttons .................................................. 7-6

    7.3 Calibration Curve Window ........................................... 7-10

    7.3.1 Overview ............................................................ 7-10

    7.3.2 Options Menu ..................................................... 7-11

    7.3.3 View Menu ......................................................... 7-13

    7.3.4 Using Calibration Curves .................................... 7-14

    7.4 Modify Report screen ................................................... 7-18

    7.4.1 File Menu ........................................................... 7-18

    7.4.2 Options Menu ..................................................... 7-19

    7.4.3 Help Menu .......................................................... 7-20

    7.5 Injection Table .............................................................. 7-21

    7.5.1 Injection Table .................................................... 7-21

    7.5.2 Recalculating...................................................... 7-22

    8. DATA-PROCESSING CONTROL WINDOW -CHROMATOGRAM- ................................ 8-1

    8.1 Configuration of Data-Processing control Window ..... 8-2

    8.1.1 File Menu ............................................................. 8-3

    8.1.2 Edit Menu ............................................................. 8-4

    8.1.3 Data control Menu ................................................ 8-5

    8.1.4 Options Menu ....................................................... 8-7

    8.1.5 View Menu ........................................................... 8-9

    8.1.6 Window Menu .................................................... 8-10

    8.1.7 Manual Baseline Icons ....................................... 8-10

    8.2 Setting up/Modifying Data Processing Parameters ... 8-11

    8.2.1 Setting Up a Component Table with a

    Chromatogram ................................................... 8-11

    8.2.2 Setting Up Integration Time Table Using the

    Displayed Chromatogram ................................... 8-12

    8.2.3 Other Data Control Parameters .......................... 8-13

    8.2.4 Manually Correcting Baselines ........................... 8-14

    8.3 Using Multi-Display Mode ............................................ 8-16

    8.3.1 Multi-Display window .......................................... 8-16

    8.3.2 File Menu ........................................................... 8-17

    8.3.3 Edit Menu ........................................................... 8-18

    8.3.4 Option Menu....................................................... 8-19

  • - v -

    8.3.5 Window Menu .................................................... 8-26

    8.3.6 Help Menu .......................................................... 8-26

    8.4 Manipulating Display Features .................................... 8-27

    8.4.1 Manipulating Display Features ........................... 8-27

    8.4.2 Panning Graphs ................................................. 8-28

    8.4.3 Changing the Upper Y-Axis Value ...................... 8-28

    8.4.4 Moving the Line Cursor ...................................... 8-29

    8.5 Include Statistics Report for QC or Unk Vials ............ 8-30

    9. MODIFY REPORT ........................................................................................................... 9-1

    9.1 Print Preview Window .................................................... 9-2

    9.2 Opening the Report Preview Window ........................... 9-3

    9.3 Functional Buttons ......................................................... 9-3

    10. REPORT LAYOUT EDITOR .......................................................................................... 10-1

    10.1 Outline for Creating Report ......................................... 10-1

    10.2 Report Layout Editor .................................................... 10-3

    10.2.1 Report Layout Editor Window ............................. 10-3

    10.2.2 Template Display Area ....................................... 10-4

    10.2.3 Separator ........................................................... 10-4

    10.2.4 Flame ................................................................. 10-5

    10.3 Menu Bar/Toolbar ......................................................... 10-6

    10.3.1 File Menu ........................................................... 10-6

    10.3.2 Edit Menu ........................................................... 10-7

    10.3.3 Option Menu....................................................... 10-8

    10.3.4 Window Menu .................................................... 10-8

    10.3.5 Help Menu .......................................................... 10-9

    10.4 Selection Buttons ....................................................... 10-10

    10.4.1 Selection Buttons ............................................. 10-10

    10.4.2 Report Items List .............................................. 10-12

    10.4.3 Framed Report Items ....................................... 10-13

    10.5 Report Layout Editor .................................................. 10-14

    10.5.1 Generating a Report ......................................... 10-14

    10.5.2 Mismatched Layout Items Dialog ..................... 10-15

    10.6 Method Layout Template ........................................... 10-16

    10.6.1 Creating a New Master Layout ......................... 10-16

    10.6.2 Load Master Layout Dialog............................... 10-16

    10.6.3 Basic Steps in Formatting a Layout .................. 10-17

    10.7 Dependency Function of Report Items ..................... 10-34

    10.7.1 Modifying Dependencies .................................. 10-34

    10.7.2 Modify Dependencies Dialog ............................ 10-35

    10.8 Master Layout ............................................................. 10-37

    10.8.1 Modify Master Layout Dialog ............................ 10-37

    10.8.2 Save Master Layout Dialog .............................. 10-38

    10.9 Master Layout File ...................................................... 10-38

  • - vi -

    10.10 Special Report Items .................................................. 10-39

    10.10.1 Multi-Chrom Overlay Report ............................. 10-39

    10.10.2 Resolution Report ............................................ 10-39

    11. FUNCTIONS AND OPERATIONS OF MODULE

    DETAILED INFORMATION WINDOW .......................................................................... 11-1

    11.1 Status Display on Module Detailed Information

    Window ......................................................................... 11-1

    11.1.1 Common (system status) ................................... 11-1

    11.1.2 1110 Pump ......................................................... 11-2

    11.1.3 1210 autosampler .............................................. 11-2

    11.1.4 1310 column oven .............................................. 11-3

    11.1.5 1410 UV detector ............................................... 11-3

    11.1.6 USB-AID ............................................................ 11-3

    11.1.7 1430 DAD .......................................................... 11-4

    11.2 Control Commands on Module Detailed Information

    Window ......................................................................... 11-5

    11.3 Status Display ............................................................... 11-8

    11.3.1 1110 Pump Status Display ................................. 11-8

    11.3.2 1210 Autosampler Status Display .................... 11-12

    11.3.3 1310 Column Oven Status Display ................... 11-16

    11.3.4 1410 UV Detector Status Display ..................... 11-17

    11.3.5 USB-AID Status Display ................................... 11-20

    11.3.6 1430 DAD Status Display ................................. 11-21

    12. DATA RE-PROCESSING - 3D DATA - .......................................................................... 12-1

    12.1 Composition of a data re-processing window ........... 12-2

    12.1.1 File Menu ........................................................... 12-3

    12.1.2 Edit Menu ........................................................... 12-4

    12.1.3 Process Data Menu ............................................ 12-5

    12.1.4 Options Menu ..................................................... 12-7

    12.1.5 Library Menu .................................................... 12-11

    12.1.6 View Menu ....................................................... 12-12

    12.1.7 Window Menu .................................................. 12-15

    12.2 Set and Modify Extracted Parameters of

    Chromatogram ............................................................ 12-16

    12.2.1 Display Chromatogram with

    Arbitrary Wavelength ........................................ 12-16

    12.2.2 Selecting/Displaying Chromatograms with

    Fixed Wavelengths ........................................... 12-17

    12.2.3 Displaying Integrated Chromatogram ............... 12-18

    12.2.4 Displaying/Editing the Best Chromatogram ...... 12-19

    12.2.5 Other Data Processing Parameter ................... 12-22

  • - vii -

    12.2.6 Data Reprocessing of Extracted

    Chromatogram ................................................. 12-23

    12.3 Set and Modify Extracted Parameter of Spectrum ... 12-24

    12.3.1 Spectrum Display of Arbitrary Time .................. 12-24

    12.3.2 Overlapping Display of Spectrum ..................... 12-25

    12.3.3 Selecting a Background-Spectrum/Displaying

    a Background-Subtracted Spectrum ................ 12-26

    12.3.4 Specify Integrated Spectrum ............................ 12-27

    12.4 Performing/Displaying Peak-Purity Checks ............. 12-28

    12.4.1 Perform Peak Purity Check .............................. 12-28

    12.4.2 Check purity Check Spectrum .......................... 12-29

    12.4.3 Set and modify Peak Purity Check Processing

    Parameter ........................................................ 12-30

    12.5 Other Display Window ............................................... 12-31

    12.5.1 Chromatogram Display Window ....................... 12-31

    12.5.2 Spectrum Display Window ............................... 12-32

    12.5.3 Using 3-D Mesh View ....................................... 12-34

    12.6 Change Operation of Graph View ............................. 12-36

    12.6.1 Zooming and Un-zooming of Graph ................. 12-36

    12.6.2 Moving the Chromatogram ............................... 12-37

    12.6.3 changing Signal Intensity Scale ........................ 12-37

    12.6.4 Moving Wavelength Cursor and Time Cursor ... 12-37

    13. Spectrum Library Window .......................................................................................... 13-1

    13.1 Configuration of Spectrum library Window ................ 13-1

    13.1.1 Process Menu .................................................... 13-2

    13.1.2 Options Menu ..................................................... 13-3

    13.2 Managing Spectrum Library ........................................ 13-5

    13.2.1 Modifying Library Spectrum Information ............. 13-5

    13.2.2 Filtering of Spectrum List ................................... 13-8

    13.2.3 Delete Spectrum ................................................ 13-9

    13.3 Using Spectrum Library ............................................. 13-10

    13.3.1 Library Menu .................................................... 13-12

    13.3.2 Searching and Saving Spectrum Library .......... 13-14

    13.3.3 Reverse Search library ..................................... 13-19

    13.3.4 Search library under Data Acquisition .............. 13-21

    14. HPLC APPLICATION INFORMATION .......................................................................... 14-1

    14.1 Introduction .................................................................. 14-1

    14.2 Example of a Solvent Time Table for

    Gradient Elution ........................................................... 14-1

    14.3 Determining a Sampling Period .................................. 14-2

    14.3.1 Automatic Sampling Period Mode ...................... 14-2

    14.3.2 Manual Sampling Period Mode........................... 14-2

    14.4 Detector WL Table and Best WL Chromatogram ....... 14-3

  • - viii -

    14.5 Quantification Methods ................................................ 14-3

    14.5.1 Area% / Height% Method ................................... 14-3

    14.5.2 Calibration Curve ............................................... 14-4

    14.5.3 External Standard Method .................................. 14-5

    14.5.4 Normalized-% Method ........................................ 14-7

    14.5.5 Internal Standard Method ................................... 14-8

    14.6 Peak Identification ...................................................... 14-10

    14.6.1 Example of Percentage Time Window (%TIME)

    ......................................................................... 14-10

    14.6.2 Example of Absolute Time Window

    (Absolute Time) ................................................ 14-10

    14.6.3 Peak Identification Rules .................................. 14-11

    14.7 Blank Subtraction ....................................................... 14-12

    14.8 Grouping Individual Components ............................. 14-14

    14.9 Retention Time ........................................................... 14-15

    14.9.1 Relative Retention Time (RRT) ........................ 14-15

    14.9.2 Corrected Retention Time (CRT)...................... 14-16

    14.10 Integration and Baseline Correction ......................... 14-18

    14.10.1 Peak Detection and Integration ........................ 14-18

    14.10.2 Noise ................................................................ 14-21

    14.10.3 Bunching .......................................................... 14-22

    14.10.4 Smoothing ........................................................ 14-22

    14.10.5 Sensitivity ......................................................... 14-23

    14.10.6 Peak Detection ................................................. 14-23

    14.10.7 N-Method ......................................................... 14-23

    14.10.8 Integration-Inhibit ............................................. 14-24

    14.10.9 Tail ................................................................... 14-24

    14.10.10 Vertical .......................................................... 14-24

    14.10.11 Forward- and Backward-Horizontal ................ 14-24

    14.10.12 Negative Peak ............................................... 14-25

    14.10.13 Group ............................................................ 14-25

    14.10.14 Precedence of Baseline Functions ................. 14-26

    14.10.15 Baseline Code ............................................... 14-26

    14.11 Peak Purity Check Using Spectra ............................. 14-28

    14.12 Spectral Correlation in Spectral Peak ID, Peak Purity,

    and Spectrum Library Search .................................... 14-30

    14.13 Statistical Calculations Used by Primaide System

    Manager ...................................................................... 14-31

    14.13.1 Average ............................................................ 14-31

    14.13.2 Dispersion (Variance) ....................................... 14-31

    14.13.3 Standard Deviation ........................................... 14-31

    14.13.4 Relative Standard Deviation ............................. 14-31

    14.13.5 Coefficient of Determination ............................. 14-32

    14.14 System Suitability Test (SST) .................................... 14-33

    14.14.1 Capacity Factor, (k') ......................................... 14-34

  • - ix -

    14.14.2 Selectivity, () .................................................. 14-34

    14.14.3 Number of Theoretical Plates, N ...................... 14-35

    14.14.4 Resolution, R.................................................... 14-37

    14.14.5 Asymmetry, Asym ............................................ 14-39

    14.14.6 Signal -to- Noise Ratio ..................................... 15-40

    14.15 Data Diagnosis ........................................................... 14-42

    14.15.1 Designating a Diagnosis Peak ......................... 14-42

    14.15.2 Designating Expected Concentration (E-Conc)

    Peaks ............................................................... 14-43

    14.16 Decomposed Substance Report ................................ 14-44

    14.16.1 Rule on Component Peak Search .................... 14-44

    14.16.2 Calculation Method for Decomposed Substance

    Report .............................................................. 14-45

    14.17 Troubleshooting ......................................................... 14-46

    14.17.1 Peak Determination Problems .......................... 14-46

    14.17.2 Qualitative and Quantitative Analysis Problems14-49

    14.17.3 Error Clearing Method ...................................... 14-50

    APPENDIX .......................................................................................................... APPENDIX 1-1

    APPENDIX 1. FUNCTION AND OPERATION OF

    ONLINE DDE .................................... APPENDIX 1-1

    AP.1.1 Operating Environment of Attached Online DDE

    Program ............................................ APPENDIX 1-1

    AP.1.2 Setting of Method .............................. APPENDIX 1-2

    AP.1.3 Details of Attached Online DDE

    Program ............................................ APPENDIX 1-3

    AP.1.4 Setup of Online DDE ......................... APPENDIX 1-5

    AP.1.5 Example of Online DDE Result ......... APPENDIX 1-6

    APPENDIX 2. APPLICATION OF PRIMAIDE SYSTEM MANAGER

    .......................................................... APPENDIX 2-1

    AP.2.1 Application in Internal Standard

    Method .............................................. APPENDIX 2-1

    AP.2.2 Application of Report Output ............. APPENDIX 2-3

    APPENDIX. 3 AIA FILE CONVERSIONS ................ APPENDIX 3-1

    AP3.1 To Convert from Raw Data to

    AIA Format ........................................ APPENDIX 3-1

    APPENDIX. 4 CAUTION ON SETTING OF

    SAMPLING PERIOD ......................... APPENDIX 4-1

    APPENDIX. 5 CAUTION ON SETTING OF

    RESPONSE ...................................... APPENDIX 5-1

    APPENDIX. 6 ABOUT ERROR MESSAGE ............. APPENDIX 6-1

    AP6.1 1110Pump A/B Message ................... APPENDIX 6-1

    AP6.2 1210Autosampler .............................. APPENDIX 6-2

    AP6.3 1310Column Oven ............................ APPENDIX 6-4

    AP6.4 1410UV Detector ............................... APPENDIX 6-6

  • - x -

    AP6.5 USB-AID ........................................... APPENDIX 6-7

    AP6.6 1430 DAD ......................................... APPENDIX 6-8

    TERMINOLOGY .............................................. TERMINOLOGY - 1

    INDEX .............................................................................. INDEX - 1

  • 1.1

    1-1

    1. OVERVIEW

    CAUTION

    Fatigue due to Long-Hour Operation

    If you keep working with the display monitor and

    keyboard for long hours, your eyes and body will be

    fatigued to jeopardize your health. When working

    with the display monitor for a long time, take a break

    for 10 to 15 minutes per hour for health of your eyes

    and body.

    1.1 The Primaide System Manager The Primaide System Manager is supported the control function, the data

    acquisition and the data processing function to a High Performance Liquid

    Chromatography (HPLC) with the personal computer (PC).

    The Primaide System Manager is linked through a USB interface board (IFB) to

    control Primaide 1000 Series Pumps, autosampler, column oven, and UV

    detector using Hitachi e-Line (elite Line).

    3 D data acquisition and 3 D data processing from PC by the Primaide System

    Manager become possible.

    Primaide System Manager

    HPLC System

    Control

  • 1.2 The Main Function

    1-2

    1.2 The Main Function 1) Method Development

    Instrument-control parameters

    Data-processing parameters

    Report Layout Editor

    2) Data Acquisition

    On-line monitoring of HPLC units

    On-line display of Chromatogram data

    Multi-Method data acquisition

    On-line monitor of 3 D contour map

    Spectrum search library

    3) Chromatogram Data Processing

    Quantitation

    System suitability testing

    Identification of components by time window and spectra

    Identifying the component by retention time and spectrum

    4) 3 D Data Processing

    Spectrum data processing

    Contour map display and 3 D map display

    Creating and searching the spectrum library

    Comparing spectrum

    Blank subtraction

    Peak purity check

    Extracting the chromatogram from 3 D data (DAD data)

    5) Report Generation

    On-line quantitation report

    On-line confidence report

    Reprocess-mode report generation

    Export report files to Microsoft Excel and Microsoft Word

    Data report of on-line DAD

    6) File Management

    Opening a Method, sample table, data, or report file

    Saving a Method, sample table, data or report file

    Copying files from one application to another

    Deleting a Method, sample table, data, or report file

  • 1.2

    1-3

    7) Module

    Detailed information

    Module control

    Maintenance information/reset

  • 2.

    2-1

    2. STARTING/EXITING THE PRIMAIDE SYSTEM MANAGER

    2.1 Introduction The Primaide System Manager program uses the basic terminology and

    techniques established for Windows operating system with regard to screen,

    menu, mouse, keyboard, and Help functions.

    Refer to the Microsoft Windows operating system User's Guide for a review of

    these fundamentals.

    NOTE: NOT RESPONDING message will be indicated on the screen of

    Primaide System Manager when you execute a low-speed operation

    such as a communication connection to an instrument. However,

    software will not be affected. Please wait for a completion of

    processing.

  • 2.2 Starting/Exiting the Primaide System Manager

    2-2

    2.2 Starting/Exiting the Primaide System Manager Click [Primaide] icon on desktop.

    The Primaide System Manager is started and the Main Window appears.

    The Primaide System Manager clicks [All Program][Primaide][Primaide] of

    the Start menu under the left of the window, Or, the Main Window is displayed

    by starting when the [Primaide] icon of desktop is double-clicked.

    Windows 7 Start Menu

    (Primaide Icon)

    Primaide System Manager is operated from the Main Window.

    Chose [Exit] in File Menu of the Main Window for exiting the Primaide System

    Manager. Or, please click button on the right of the Main Window.

  • 2.3

    2-3

    2.3 Menu The main menu item is as follows.

    2.3.1 File Menu

    The File menu on the Primaide System Manager program contains the

    following commands.

    New Ctrl+N

    Opens the New dialog box and creates a new Method or Sample Table.

    Choose the New command from the File menu to open the New dialog box

    shown below.

    Highlight the desired file type (Method or Sample Table) and choose OK.

    Either the Method Setup or Sample Setup starts with parameters set to

    program defaults or blank settings.

    You can modify parameters and choose Save As from the File menu to save the

    file with a new name.

    Open... Ctrl+O

    Opens the Open File dialog box so you can open an existing Method, Sample

    Table, Data Series, or Report.

    Choose Open from the File menu to display the Open File dialog box shown

    below:

  • 2.3.1 File Menu

    2-4

    Refer to Table 2.1 for the functions of the Open File command buttons.

    Table 2-1 Open File Commands

    Button Function

    Copy Opens the Copy dialog box to allow you to copy Methods,

    Sample Tables, Data Series, and Reports.

    Delete

    The Delete button opens the Delete dialog box to delete files.

    The Delete function is only available to certain users whom the

    System Administrator has granted file deletion rights using the

    Primaide Administration program.

    Sort

    Opens the Sort dialog box to sort Methods, Sample Tables,

    Data, and Reports. The type of file being sorted is reflected in

    the name visible in the Title Bar of the dialog box (e. g. , Sort

    Sample). Files cannot be sorted by name or by date/time. The

    sort order can be either ascending or descending.

    Filter Initiates the Data Filter dialog box, but only when the chosen

    File Type is Data.

    Rename Opens the Rename dialog when the File Type is either Data or

    Report. Use to change the file name or report name.

    OK

    Opens the highlighted file. If the File Type is Method, the

    Method Information screen is opened. If it is Sample Tables,

    the Setup Information screen is opened. If the File Type is

    Data, the associated Injection Table is displayed. If the File

    type is Report, the selected report is displayed in the Report

    Preview window.

    Cancel Cancels the Open File dialog box.

    The Open File dialog box is accessible via the Method, Sample Table,

    Data-Processing Control, and Preview/Print Report icons on the Main Tool Bar

    in particular situations (See Section 3.2, Main Tool Bar Icons) Regardless of how

    the dialog box is opened, you may select any file type.

  • 2.3.1

    2-5

    Close

    In general, choose Close from the File menu to close the window that is active.

    The window can be for a Method, Sample Table, Injection Table, or data display.

    In the following cases, however, the Close command is disabled by the Primaide

    System Manager program:

    (1) If one Method and one or more Injection Tables are open, the Primaide

    System Manager program disables the Close command so that the Method

    cannot be closed before the Injection Table.

    (2) If one Sample Table is open and active, and the Data acquisition window is

    also open, the Primaide System Manager program disables the Close

    command so that the Sample Table that is currently in use for data

    acquisition cannot be closed until the Data Acquisition window is closed.

    When closing an Injection Table with the Close command, the Primaide System

    Manager program automatically closes any associated data display windows

    that are open.

    Close All

    Closes all open functions in all open windows regardless of focus. If any window

    has been modified, the program offers the opportunity to save it.

    When Data Acquisition is in a mode other than Idle Monitor, the Primaide

    System Manager program closes all other types of windows except the Data

    Acquisition window and the Sample Table that currently is in use.

    If data acquisition is in Idle Monitor mode, the Primaide System Manager

    program closes all windows.

    Print Preview

    The preview screen of the contents of the display can be displayed by clicking

    menu bar. The print output can be performed from this screen.

    Print Setup...

    Lets you pick a printer, paper orientation (Portrait or Landscape), paper size

    (letter, legal, A4, B5) and the paper source (tray or manual).

    Print... Ctrl+P

    The contents of the display can be printed.

    Save Method

    Saves the current Method on display in the Window. The previous version on

    the hard disk is over-written after you respond to all Method validation

    messages, if any.

  • 2.3.1 File Menu

    2-6

    Save Method As...

    The Save As dialog boxes allow you to specify the name of the current Method or

    Sample Table and the name of the Application under which the Method or Table

    is to be saved.

    The Save As dialog box offers the following options:

    a)Name

    Enter a new name in the text-entry box or highlight an existing name displayed

    in the list box.

    b)Application

    Highlight the application name in the list box to indicate where the Method or

    Sample Table is to be saved.

    c)Comment

    Enter appropriate comments (99 characters, maximum) regarding the file saved

    using the Save As function.

    d)OK

    Choose the OK button to save the Method or Sample Table with the specified

    name to the specified application. If the name already exists in the application,

    the Primaide System Manager program requests an OK to overwrite the

    existing file.

    e)Cancel

    Choose the Cancel function to cancel the Save As dialog box.

  • 2.3.2

    2-7

    Save Sample

    Saves the Sample Table in focus.

    When you select this command, the Primaide System Manager program starts

    to validate the Sample Table and then initiate the Save As dialog box. The

    previous file version on the hard disk is overwritten.

    Save Sample As...

    Initiates Sample Table validation and opens the Save As dialog box so that you

    can specify the Title of the Sample Table and the name of the Application.

    Exit

    Exits the Primaide System Manager program.

    Exit is inactive while you are acquiring data using Data Acquisition.

    Prompts to save documents.

    2.3.2 Edit Menu

    The Edit menu on the Primaide System Manager program contains the

    following commands.

    Copy Ctrl+C

    Copies a selection to the Windows clipboard. The original from which the

    selection was copied remains unchanged. To transfer the selected (highlighted)

    Method, Sample Table, Data or Report files from the current application to

    another application, select (highlight) the application on the Copy dialog box.

    Cut Ctrl+X

    If it is a screen of the table, copies a highlighted item to the Windows clip-board,

    and deletes the highlighted item.

    Delete Ctrl+Del (table) , Del (Report edit)

    If it is a screen of the table, deletes the highlighted item.

    Paste Ctrl+V

    Used in conjunction with the Cut and Copy commands.

    Inserts a copy of the selection stored in the Windows clipboard at the insertion

    point determined by the mouse pointer.

    Edit Colors...

    Opens the Color Selection dialog box to allow you to change screen colors. Items

    associated with the current window are listed in the upper left box.

    The Color Selection dialog box appears as shown below. Select an item and

    change colors using the score bars. Click on the Preview button to preview the

    new color; click on the Save button to save the color.

  • 2.3.3 Window Menu

    2-8

    2.3.3 Window Menu

    The File menu on the Primaide System Manager program contains the

    following commands.

    Arrange Icons

    Arrange icons at the bottom of the window.

    Cascade

    Arrange windows so they overlap.

    Tile

    Arrange windows as non-overlapping tiles.

    2.3.4 Help Menu

    The File menu on the Primaide System Manager program contains the

    following commands.

    Index

    List Help topics.

    Using Help

    Display instructions about how to use help.

    About Primaide...

    Displays the About the Primaide System Manager Program dialog box. This box

    provides specific information such as the program name, part number, version

    number, serial number, licensee name, and a copyright notice.

  • 3.

    3-1

    3. PRIMAIDE SYSTEM MANAGER MAIN WINDOW Basic items on the Main window are shown below:

    1) Title Bar

    The title bar is the horizontal bar located along the top of the Main window. It

    displays the program title (Primaide System Manager) and the names of the

    current application (e. g. , samples, in the above illustration) and the current

    Method.

    2) Menu Bar

    The menu bar is the horizontal bar located directly below the title bar. It lists

    program menus.

    The menus that are available change according to the task that is active. Each

    menu, when opened, provides a list of command options. See a more detailed

    description of the menus available to the Primaide System Manager.

    3) Display Area

    The display area is the large open area of the Primaide System Manager

    program Main window. Initially, when the program boots up, the display area

    shows the Primaide System Manager program logo. Subsequently, dependent

    on the task selected, it displays one of a variety of task windows.

    Menu Bar

    Main Tool Bar

    Status Bar

    Display Area Title Bar

  • 3. PRIMAIDE SYSTEM MANAGER MAIN WINDOW

    3-2

    4) Main Tool Bar

    The Main Tool Bar is the vertical bar located along the left side of the main

    window. It contains the following icons.

    5) Status Bar

    The status bar is the horizontal bar located along the bottom of the main

    window. It displays program messages, current date and time, and item

    information.

  • 3.1

    3-3

    3.1 Menu Command Function of Main Window The menu of the main window is as follows.

    3.1.1 File Menu

    The following items are found in the File menu of the Main Window:

    Please refer to [2.3.1 File Menu] for a general menu.

    New Ctrl+N

    Create a new document.

    Open... Ctrl+O

    Open an existing document.

    Close All

    Close all open documents.

    Print Setup...

    Change the printer and printing options.

    Spectrum Library

    The Spectrum Library window is opened. This command can be used in the

    Main Window or the Reprocess Data window (See Chapter13 for library

    window).

  • 3.1.2 Help Menu

    3-4

    Exit

    Quit the application; prompts to save documents.

    3.1.2 Help Menu

    The Help Menu contains the following commands. Index Opens a Help window

    and displays a list of main topics.

    Please refer to [2.3.4 Help Menu] for a general menu.

    Index

    List Help topics.

    Using Help

    Display instructions about how to use help.

    About Primaide...

    Display program information, version number and copyright.

  • 3.2

    3-5

    3.2 Main Tool Bar Icons The Main Tool Bar Icons contains the followings.

    Table 3-1 Main Tool Bar Icons

    Icon Icon Name and Function

    Change Application Select to open the Application dialog to switch to a different

    application.

    Method Setup Select to begin Method Setup functions.

    Sampler Setup Select to begin Sample Table Setup functions.

    Acquire Data Select to begin Data Acquisition functions.

    Reprocess Data Select to begin Data Processing Control functions.

    Report Select to open the Open File dialog box to review report

    files.

    System Status Select to open the Hardware Status dialog box to display

    the current hardware status, before and after downloading

    and during idle monitoring or data processing.

    Pump(A/B) ON/OFF Select upper icon to toggle Pump A on/off; select lower

    icon to toggle Pump B on/off.

    Module Detailed

    Information

    Detailed information on the module (status) and the control

    setting .

    Quick Analysis Start It is possible to shift to the data collection monitor

    promptly by the input of the analysis file name and the

    rack name, etc.

  • 3.2.1 Change Application Icon

    3-6

    3.2.1 Change Application Icon

    Application feature to change from one system application to another.

    When you click the Change Application icon, the Icon Primaide System

    Manager program displays the Application dialog box.

    In the list box, highlight the application name and choose Select.

    (1) If one or more types of windows other than a Data Acquisition window are

    open, the Primaide System Manager closes all the windows (i. e. , executes

    Close All) and changes to the new application.

    (2) If one or more types of windows including the Data Acquisition window are

    open and data acquisition is in progress, the Primaide System Manager

    closes all windows except the Data Acquisition window and the Sample

    Table used by Data Acquisition, and changes to the new application.

    (3) If one or more types of windows including a Data Acquisition window are

    open and the Data acquisition window is in the Idle Monitor mode, the

    Primaide System Manager closes all the windows (including the Data

    Acquisition window and the Sample Table being used) and changes to the

    new application.

    (4) If nothing is open in the Main window, the Primaide System Manager

    changes to the new application.

    3.2.2 Method Setup Icon

    Clicking on the Set Up Method icon initiates the Method Setup function.

    The Primaide System Manager program requires that only one Method be open

    at any time (except for reference Methods).

    You cannot close a Method being used for data processing.

    Depending on the situation, the Primaide System Manager program takes the

    following actions when you choose the Method Setup icon:

    (1) If no Method window is open (first-time log-in), the Primaide System

    Manager program displays the Open File dialog box with Method selected

    as the default File Type (i. e. , performs the Open command from the File

    menu). The last-used Method for the current application is highlighted.

  • 3.2.2

    3-7

    (2) If one Method window is open but not active (minimized to an icon), the

    Primaide System Manager program minimizes all other windows to icons

    and activates (maximizes) the open Method window.

    (3) If one Method window is open and active and no Injection Table is open, the

    Primaide System Manager program displays the Open File dialog box with

    Method as a default File Type (i. e. , performs the Open command from the

    File menu). The last-used Method for the current application is highlighted.

    When you select a Method, a message appears for you to confirm Close

    current Method(s)?. If Cancel is chosen, the Primaide System Manager

    cancels the entire open file function. If OK is chosen, the Primaide System

    Manager closes and saves the cur-rent Method and opens the new Method.

    All other windows are minimized.

    (4) If one Method is open and active, and one or more Injection Tables are open

    as well, the Primaide System Manager displays the Open File Dialog with

    Method as a default Type.

    The last used Method for the current application is highlighted.

    When you select a Method, the Primaide System Manager first verifies the new

    Method configuration. (See Injection Table or Data Processing Tool)

    If it does not match, an error dialog is displayed with detector types listed for

    the new Method and the open Injection Tables and the Primaide System

    Manager fails to open the new Method.

    If the succeeds, a message appears for you to confirm Close current Method(s).

    If Cancel is pressed, the Primaide System Manager cancels the entire open file

    function.

    If OK is pressed, the Primaide System Manager closes the current Method,

    opens the new Method, and minimizes all other windows.

    The Primaide System Manager then updates all open Injection Tables and their

    graphical displays according to the newly opened Method.

  • 3.2.3 Sampler Setup Icon

    3-8

    3.2.3 Sampler Setup Icon

    You can open multiple Sample Tables via the Sample Table Setup Table Icon

    function. However, it requires that one Sample Table window must be selected

    when the Data Acquisition window is open.

    Depending on the situation, the Primaide System Manager program takes one

    of the following actions when you choose the Sampler Setup icon.

    (1) If no Sample Table window is open, the Primaide System Manager program

    displays the Open File dialog box with Sample Table chosen as the File

    Type (i. e. , performs the Open command from the File menu).

    The last-used Sample Table for the current application is highlighted.

    When you select a Sample Table and choose OK, the Primaide System

    Manager program opens the Sample Table and sets all other windows to

    background.

    (2) If one or more Sample Tables are open and all of them are minimized, the

    Primaide System Manager program minimizes all other windows (e. g., a

    Method window and restores (maximizes) all Sample Tables. 3. If one or

    more Sample Tables are open and one is selected (active), the Primaide

    System Manager program displays the Open File dialog box with Sample

    Table selected as a default File Type (i. e., performs the Open command).

    The last-used Sample Table for the current application is highlighted.

    When you select a Sample Table and choose OK, the Primaide System

    Manager program opens the Sample Table, maximizes all other Sample

    Tables, and minimizes all other types of windows (e. g., a Method window).

    3.2.4 Acquire Data Icon

    The Primaide System Manager allows you to open one Data Acquisition

    Window.

    The following rules are followed when you click on the Acquire Data icon:

    (1) If no Data Acquisition Window is open, the Primaide System Manager

    displays the Open Sample Table for Data Acquisition Dialog where all

    sample tables in the current application are listed.

    Upon your selection of a Sample Table, the Primaide System Manager

    opens the Sample Table and starts the Data Acquisition window by

    validating the first Method in the Sample Table for starting Idle Monitor.

    (2) If no Data Acquisition Window is open and no Sample Tables can be found

    in the current application, the Primaide System Manager displays the

    message: No sample tables available.

  • 3.2.5

    3-9

    (3) If the Data Acquisition Window is already open but not active, the Primaide

    System Manager restores (maximize) the Data Acquisition Window.

    3.2.5 Reprocess Data Icon

    A Method must be open before injection data can be opened via the

    Data-Processing Control function.

    The Primaide System Manager program takes the following actions when you

    choose the Reprocess Data icon.

    (1) If no Method is open and no Injection Table is open, the Primaide System

    Manager program displays the Open File dialog box with Data as the

    default File Type (i. e. , performs the Open command from the File menu).

    The last-used or last-acquired Injection Table (whichever is more recent) for

    the current application is highlighted.

    When you select an Injection Table and choose OK, the Primaide System

    Manager program automatically opens the Method used to acquire the

    stored data.

    It then proceeds to the Method verification.

    (2) If one Method is open and one or more Injection Tables (including their

    graphical display windows) are also open but not active, the Primaide

    System Manager program restores all Injection Tables (including graphical

    displays) with the last-opened one set to foreground and sets the Method to

    background.

    (3) If one Method is open, and one or more Injection Tables (including their

    data display windows) are also open with one of them selected (active), the

    Primaide System Manager program displays the Open File dialog box with

    Data as the default File Type (i. e. , performs the Open command from the

    File menu).

    The last-used or last-acquired Injection Table (whichever is more recent) for

    the current application is highlighted.

    Upon your selection of an Injection Table, the Primaide System Manager

    program verifies the Method configuration with the new Injection Table.

    If it does not match, the Primaide System Manager displays the error

    dialog indicating the detector types of the Method and the Injection Table

    and fails to open the data.

    If the Method verification is successful, it opens the Injection Table.

  • 3.2.6 Report Icon

    3-10

    3.2.6 Report Icon

    Choose the Preview/Print Report icon to have the Primaide System Manager

    program perform the Open command and display the Open File dialog box with

    Report chosen as a default.

    The Open File dialog box contains a list of report files available from the data

    files open on display.

    If no file is open, the list appears blank.

    NOTE: Before using Print Preview, ensure that the proper printer driver is

    installed.

  • 3.2.7

    3-11

    3.2.7 System Status Icon

    When you choose the System Status icon, the Module Status dialog box appears

    showing the current system configuration.

    After connection for communications with each module, the following icons can

    be accessed.

    Data Acquisition

    Pump on/off

    Module Detailed Information

    Quick Analysis Start

    It changes by the Primaide Administration program in the module composition.

    You can change the configuration via the Primaide Administration program.

    NOTE: The hardware components in the HPLC system must be set up using

    the Primaide Administration program. If the program detects a

    component that has not been set up in Primaide Administration, a

    warning message appears, the component does not appear in the

    screen display of system components, and the unconfigured

    component is not considered a hardware component of the system.

  • 3.2.7 System Status Icon

    3-12

    It explains the item of the Module Detailed Information dialog box as follows.

    1) Status

    Displays system status information that is also viewable in the Data

    Acquisition window. When you first start the program, displays the message

    "Downloading" while it downloads information to the Primaide System

    Manager. Upon completion of the downloading process, the current system

    status displays on screen.

    2) Key Lock

    The Lock/Unlock buttons are only activated for use by the System

    Administrator.

    Click on the respective button to Lock/Unlock instrument key. The Lock/Unlock

    function over rides the setting on the GLP Options in the Primaide

    Administration program.

    3) Module

    Interface

    Displayed are the type of communication interface (namely, IFB), program No.

    and serial No. that are recognized by the Primaide System Manager program.

    When AID is connected, program No. will not be displayed.

    Pump A

    Pump A type of system, program No. and serial No. are displayed.

    Pump B

    Pump B type of system, program No. and serial No. are displayed.

    Autosampler

    The type of autosampler, program No. and serial No. are displayed.

    Oven

    The type of column oven, program No. and serial No. are displayed.

    Detector Ch1/Ch2

    The form, program No. , and cereal No. of the detector are shown.

    Accessory

    Connected attached device name or form (for instance, cooling unit) and

    program No. and serial No. are shown.

    4) OK

    Click to the OK button to close the dialog box.

    5) Initialize

    Click to connect each module with Primaide System Manager. This allows the

    interface module to recognize a new component that has been powered up after

    the entire system was already operational.

    6) Disconnection

    Click to disable and disconnect from all data acquisition functions.

  • 3.2.8

    3-13

    7) Error release

    When the error occurs in connected module, the Error release button is

    displayed.

    NOTE: The added component must have already been configured as a part of

    the system at some time in the past, using the Primaide

    Administration program.

    3.2.8 Pump(A) ON/OFF and Pump(B) On/Off Icon

    When you choose the Pump(A/B) On/Off icon, the pump is turned on and off.

    3.2.9 Module Detailed Information Icon

    When you choose the Module Detailed Information icon, Open the Module

    Detailed Information window.

    The Module Detailed Information window is usable for checking the detailed

    information about modules and controlling them.

    3.2.10 Quick Analysis Start Icon

    When quick analysis start icon of main tool bar is clicked, quick analysis

    starting dialog is displayed.

  • 4.

    4-1

    4. FUNCTION AND OPERATION OF METHOD WINDOW Method window contains the module operation parameters and the data

    processing parameters. You can not set/edit a file by opening two or more

    method files at the same time in a method window, but you can open a method

    file that want to refer as the reference method window. You can copy all

    parameters or a part of parameters of the reference method to the current

    method. And the current method can be compared with the reference method.

    4.1 Menu Command Function of Main Window Icons of the Method window are displayed two steps in the top of window.

    1st step : Module operation parameters

    2nd step: Data processor parameters (CH1, CH2)

  • 4.1.1 File Menu

    4-2

    4.1.1 File Menu

    The following items are found in the File menu of the Main Window:

    New... Ctrl+N

    Create a new document.

    Open... Ctrl+O

    Open an existing document.

    Close

    Close the active document.

    Close All

    Close all open documents.

    Save Method Ctrl+S

    Save the current Method.

    Save Method As...

    Save the current Method with a new name.

    Print... Ctrl+P

    Print the active document.

    Print Preview

    Display full pages.

    Print Setup...

    Change the printer and printing options.

    Exit

    Quit the application; prompts to save documents.

    4.1.2 Edit Menu

    The Edit menu allows you to perform a variety of editorial functions.

    Please refer to [2.3.2 Edit Menu] for a general menu.

    Cut Ctrl+X

    Cut the selection and put it on the Clipboard.

    Copy Ctrl+C

    Copy the selection and put it on the Clipboard.

    Paste Ctrl+V

    Insert Clipboard contents.

    Delete Ctrl+Del

    Erase the selection.

    Add New

    Adds a new row after the last row on either the Component Table, the

    Integration Time Table, and all other tables with corresponding default values.

  • 4.1.3

    4-3

    Edit Colors

    Opens the Color Selection dialog box to allow you to change screen colors. Items

    associated with the current window are listed in the upper left box.

    The Color Selection dialog box appears as shown below.

    Select an item and change colors using the score bars. Click on the Preview

    button to preview the new color; click on the Save button to save the color.

    4.1.3 Module Setup Menu

    The instrument condition menu on the Main screen of the Primaide System

    Manager program contains the following commands.

    Method Configuration

    You can review or modify Method Configuration parameters from the Method

    Configuration screen.

    Method Information

    You can review or modify general Method information such as the Method

    History, Column Type, Method Name, and Solvent Names.

    History is available only when the Method Log Book is selected in GLP options

    when running the Primaide Administration program.

    Pump Setup

    The Interface Module, the Gradient Mode, and the number of solvents specified

    for each Pump on the Method Configuration screen must be the same in both

    Methods.

    Autosampler Setup

    Autosampler setup parameters window is displayed.

  • 4.1.4 Data Process Setup Menu

    4-4

    Column Oven Setup

    The oven selected on the Method Configuration must be the same in both

    Methods.

    Channel 1/2 Detector Setup

    Channel 1/2 detector setup parameters window is displayed.

    Time Summary

    Displays the Pump (solvent) table and event table graphically.

    4.1.4 Data Process Setup Menu

    You can access the following capabilities of the Primaide System Manager

    program using the Data Processing Setup menu:

    Channel 1 (CH. 1) or Channel 2 (CH. 2)

    Use the Channel 1/2 command to review or modify the data-processing method

    (on all Data-Processing Control windows) used exclusively for analyzing data

    acquired with the channel 1/2 detector. If a single detector is configured on the

    Method Configuration screen, only the one command corresponding to the

    configured detector is enabled (the other is disabled). For example, if Channel 1

    is selected, Channel 2 is disabled and the Primaide System Manager program

    displays data-processing parameters only for the Channel 1 detector.

    Calculation Method

    This command reviews or modifies the calculation method in the Calculation

    Method screen.

    Component Table

    Sets up, reviews, or modifies the Component Table for data processing.

    Concentration Table

    Select this command or icon to review or set up the Concentration Table.

    Coefficient Table

    Reviews or modifies the coefficients of the calibration curve for each component.

  • 4.1.5

    4-5

    Integration Table

    Use this command to review or set up the Integration Time Table.

    DAD Data Processing

    Select to review or modify DAD Data Processing parameters.

    Chromatogram Display Format

    Use this command to open the Chromatogram Display Format Screen.

    DAD Display Format

    Use this command to open the DAD Display Format Screen.

    Confidence Report

    Use this command to set report parameters for System Suitability Tests,

    Module Performance Tests, and Data Diagnosis in the Confidence Report

    screen.

    Report Format

    Displays the Report Format dialog box.

    Update Method

    Click on this icon to apply the changes you have made to the current method to

    open Data Display windows.

    This icon is also available in the Component Table, and Integration Time Table

    in the Data-Processing Control window.

    4.1.5 Option Menu

    The commands and features described in this section are available in the

    Option menu.

    Copy DP Channels...

    Copies all data-processing parameters from channel 1 to channel of the current

    Method.

    You can enter an optional time offset that is added to