Page 1: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Update on Cardiac Risk

Managing a Patient with a Cardiac Device

Jutta Novalija, MD, PhD

Medical College of Wisconsin

Page 2: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Disclosure

Nothing to disclose

Page 3: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Relevant?

1993 to 2006 nationwide:

2.4 million patients with

primary PM

0.80 million patients with

primary ICD

369,000 PM replacement

74,000 ICD revision

Broadening indications for PM

Resynchronization

Heartfailure

Pacing Clin Electrophysiol. 2010;33(6):705-711

Page 4: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

PM/ICD

History and Classification

ICD Therapy

Perioperative Management

Electromagnetic Interference (EMI)

Page 5: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

History

First battery operated PM developed 1958 and

first implanted PM 1960

First ICD placed 1980

Page 6: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

PM - NASPE/BPEG Generic Code

I II III IV V

Chamber(s)

paced

Chamber(s)

sensed

Response to

sensing

Programmability rate

modulation

Multisite Pacing

O = None

A = Atrium

V = Ventricle

D = Dual (A+V)

O = None

A = Atrium

V = Ventricle

D = Dual (A+V)

O = None

T = Triggered

I = Inhibited

D = Dual (T+I)

O = None

R = Rate Response

O = None

A = Atrium

V = Ventricle

D = Dual (A+V)

NASPE - North American Society of Pacing and Electrophasiology

BPEG – British Pacing and Electrophysiology Group

Page 7: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

ICD - NASPE/BPEG Generic Code

I II III IV

Shock Chamber(s) Antitachycardia PacingChamber(s)

Tachycardia Detection Antibradycardia PacingChambers(s)

O = NoneA = AtriumV = VentricleD = Dual (A+V)

O = NoneA = AtriumV = VentricleD = Dual (A+V)

E = Electrogram

H = Hemodynamic

O = NoneA = AtriumV = VentricleD = Dual (A+V)

NASPE - North American Society of Pacing and Electrophasiology

BPEG – British Pacing and Electrophysiology Group

Page 8: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

ICD/PM

History and Classification

ICD Therapy

Perioperative Management

Electromagnetic Interference (EMI)

Page 9: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

How does an ICD work?

Early ICD’s - shock only

Now

Low energy shock for VT, high energy shock

for VF

Antitachycardic Pacing (ATP) - saves shocks

and battery

Backup pacing (post shock)

Page 10: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

ICD- Detection Algorithm

VT-1 Zone VT-Zone VF-Zone

Rate

Duration

RedetectionDuration

Post-ShockDuration

90-200 bpm

1-60 sec

1-15 sec

1-60 sec

110-220 bpm

1-30 sec

1-15 sec

1-30 sec

130-250 bpm

1-15 sec

1 sec

1 sec

Page 11: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

ICD - ATP

Redetection

Attempt 1 with 3 pulses Attempt 2 with 4 pulses

300

ms

300

ms

300

ms

300

ms

300

ms

300

ms

292

ms

284

ms

Ramp Burst

Redetection

300

ms

292

ms

284

ms

276

ms

268

ms

284

ms

Ramp burst Ramp/Scan burst

Burst

Ramp Ramp/Scan

Redetection

300

ms

300

ms

284

ms

284

ms

284

ms

Scan burst

Scan

Adapted from www.guidant.com

Page 12: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

PM/ICD

History and Classification

ICD Therapy

Perioperative Management

Electromagnetic Interference (EMI)

Page 13: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Heart Rhythm, Vol 8, No 7, July 2011

Page 14: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Perioperative Management

Preop

Past medical history

Interrogation, reprogramming

Intraop

Monitoring

Magnet

Postop

Interrogation

Page 15: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Preoperative Evaluation

Patients History

EKG

CXR

Electrolytes

Interrogation

Page 16: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Patients History

Indication, time and symptoms before

implantation

Manufacturer, model and program

Last evaluation

underlying rhythm

battery status

recent shocks (if ICD)

New symptoms (syncope, dizziness, fatigue,

CP, SOB, palpitations, confusion)

Page 17: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

EKG

Underlying rhythm vs. paced rhythm

(vagal maneuver discouraged)

PM dependency?

- higher risk if CHB or severe bradycardia

- lower risk if SSS or carotid sinus

hypersensitivity

Signs of malfunction?

Page 18: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

PM in the EKG

• DDD

Page 19: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

PM in the EKG

VVI

VDD

Page 20: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

PM in the EKG

AAI

VOO

I I I

Page 21: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

PM in the EKG

DDDR

rate response

various sensors:

muscle activity, transthoracic electrical impedance, temperature, minute ventilation, QT-Intervall, pH

Page 22: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

CXR

Pacemaker vs. ICD

Location of device and leads (transvenous vs.

epicardial, ventricular and/or atrial)

Manufacturer

Lead integrity (insertion into generator, entry into

subclavian vein, passage between clavicle and first

rib)

Signs of CHF

Page 23: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

CXR

RA lead

RV lead RV lead

RA lead

Page 24: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

CXR

RV lead

RV lead

Page 25: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

CXR

RV lead RV lead

Page 26: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

CXR

RV lead

Page 27: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

CXR

RV lead

Page 28: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Interrogation

Need to know manufacturer to bring correct

programmer

Underlying rhythm

Programmed settings

Battery status

Check threshold if PM dependency

Call manufacturer

Page 29: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Manufacturers

Miller's Anesthesia, 8th edition, p1486

Page 30: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Reprogramming options

ICD

Turn off anti-tachycardic pacing/shock

Change pacemaker program

Page 31: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Reprogramming options

PM

Asynchronous

Turn off rate response

Consider increase in pacing rate to optimize

oxygen-delivery to tissues for major surgery

Some manufacturers recommend increase of

amplitude and duration based on threshold

test. Changes in threshold by surgery are

described Levine PA: Ann Thorac Surg 1986,41:313

Page 32: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Rate Response

63yo after laser stomach surgery under sedation

developed HR 130/Min with Hypotension in PACU

when connected to Monitor - disappeared when

respiratory rate monitor disabled Walden J et al: Anesth Analg, 1998, 86:1334-9

59yo for TURP under spinal developed gradually

increase in HR to 130/Min every time the

electrocautery (unipolar) was used Wong DT: Anesthesiology 2001, 94:710-1

Page 33: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Rate Response

70yo for craniotomy under GA developed

periodically increase in HR to 120/Min during

drilling - disappeared after reprogramming to

DDD instead of DDDR Schwartzenburg CF: Anesthesiology V87, 5, Nov 1997

Similar incidents described in helicopter

transport, elevators, coughing, sneezing Gordon RS: Annals of Emergency Medicine 19:11,1990,1260

Page 34: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Intraoperative Precautions

Alternate pacing modality available

Have external defibrillator available (ICD)

Magnet

Chronotropes (Atropine, Isoproterenol,

Ephedrine)

Electrosurgery cautery precautions

Anesthetic specials

Page 35: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Alternate pacing modality

Time to initiate Disadvantages

Transcutaneous 1-2 Min Variable capture, patient discomfort

Transesophageal Minutes Unreliable ventricular capture, requires

special generator

(if rhythm responsive to atrial pacing)

Transvenous 3-20 Minutes Invasive, question of stability

Pacing PA catheter Minutes (if PA catheter

in place)

Requires specific PA catheter

Epicardial <1 Min Postoperative only, early lead failure

Trankina MF, In Faust RS 2002

Page 36: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Magnet

Page 37: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

PM - Magnet

Generally asynchronous pacing, but

Magnet response manufacturer depended

Possibly increases risk for

reprogramming in presence of EMI)

Page 38: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Europace 2011;13(9):1222-1230

Page 39: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

ICD - Magnet

Generally

‘Monitor only’ (suspend anti-tachycardia

therapy) and

no effect on pacing

but response unpredictable

Page 40: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG
Page 41: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Reprogramming

Preferred over Magnet if

PM dependent patients

Disabling of rate response

Non supine surgery

Page 42: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

M. E. Stone et al. Br. J. Anaesth. 2011;107:i16-i26

Page 43: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Intraoperative Monitoring

Electrical and mechanical evidence of heart

function

EKG, disable Filter

Blood pressure related tracing: pulse oximeter,

arterial line (consider in Biventricular PM, HOCM)

Avoid unipolar electrocautery Rozner MA, Anesthesiology 2002, 96, 773-4

Page 44: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Electrocautery

Direct current interference

Possible inhibition of pacing,

asynchronous pacing, increase

in pacing rate (rate response)

Possible device reset or

permanent device damage

Risk of inappropriate shock

Possible myocardial tissue

damage at lead tip Adapted from www.guidant.com

Page 45: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Electrocautery

If surgical site is close to the device

Consider asynchronous pacing mode

Short, intermittent, irregular bursts

Confirm normal device function and programmed

parameters postop

Bipolar electrocautery recommended, if unipolar

necessary, current path should be as far from

device/leads as possible (eg. place ‘bovie pad’ on

forearm)

Page 46: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Anesthetic specials

Concerns

Strain on device and cables (patient

movement, position)

Shivering or fasciculations (Succhinylcholine)

Ischemia and electrolyte disturbances (treat

loss of capture as ischemia until proven

otherwise)

Toxic doses of Local Anesthetics can

potentially lead to loss of capture Bourke ME: Can J Anesth 1996, 43:5,R24-32

Page 47: Managing a Patient with a Cardiac Device · 2015-09-14 · Managing a Patient with a Cardiac Device Jutta Novalija, ... Underlying rhythm vs. paced rhythm ... VDD . PM in the EKG

Summary

It is not the device, but the patients

disease!

Get as much information as possible

Vigilance if patient is PM dependent

Have backup pacing/defibrillating device

available

Magnet can be useful