m6-030.891.01.01.02_planningguide

302
  © Si emen s A G The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model or design, are reserved. Print No.: P o l s t e r MAGNETOM System Planning Guide MR 2005 M6-030.891.01.01.02 Repl aces : n.a. English Doc. Gen. Date: 05.05 Planning Guide Trio, A Tim System 10018418 10018620

Upload: jand20

Post on 01-Nov-2015

63 views

Category:

Documents


4 download

DESCRIPTION

Magnetom

TRANSCRIPT

  • System

    Planning Guide

    A Tim SystemTrio, Planning Guide Siemens AGe reproduction, tra

    this document or irmitted without thority. Offenders mages. All rights

    eated by patent gra a utility model served.

    Print No.: M6-030.891.01.Replaces: n.a.

    nglishoc. Gen. Da

    1001841810018620

    nsmission or usets contents is notexpress written

    will be liable for, including rightsnt or registrationor design, are

    2005

    te: 05.05Thofpeaudacrofre

    01.02 EDMAGNETOM MRPolster

  • 2 Revision / Disclaimer

    1Revision / Disclaimer

    Document revision levelThe document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.Please contact your local Siemens office to order current revision levels.

    DisclaimerThe installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro-cedures.MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 2 of 302Medical Solutions

  • Table of Contents 3

    1- 0Table of Contents

    1 _______ General _______________________________________________________ 11

    Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Controlled access area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Bruising hazard sign (Valid for the Upgrade only!) . . . . . . . . . . . . . . . . . . . . . . . . . . 13Warning sign Ear Protection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

    Regulatory affairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Pressure Equipment Directive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Pressure Device Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

    Project Manager Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Project management. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Service and installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18Environmental protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19

    System component overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20System configuration with SEP. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20System configuration with dedicated KKT chiller and IFP . . . . . . . . . . . . . . . . . . . . 21System Components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

    2 _______ Room planning_________________________________________________ 24

    General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24Information for using optical stimulation equipment . . . . . . . . . . . . . . . . . . . . . . . . . 24

    Planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Magnetic field effects on peripheral devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

    Flux density Bmax . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26Magnet siting requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

    Minimum distances between Siemens Medical Solutions magnets . . . . . . . . . . . . . 27Interferences. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

    Dynamic interference . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29AC interferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30Static interferences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31

    Iron room shielding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32Building vibration specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33

    Additional information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33Site measurement decision matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

    Noise emission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Terz-band and octave spectrum . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Noise emission values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37Noise cancellation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37

    Fringe field distribution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38Graphical fringe field distribution. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Fringe field distribution in table form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43

    Co-Siting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Magnet distances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Co-siting prerequisites . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44Field decay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 3 of 302Medical Solutions

  • 4 Table of ContentsACC distance for systems with the same operating frequency . . . . . . . . . . . . . . . . . 46Magnet orientation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

    Room configurations and sizes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50Room configuration example. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51Minimum technical finished room size. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56

    Magnet floor loading . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58Seismic anchor positions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60

    RF filter plate installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62Recommended installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63Alternative installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64

    3 _______ System Components ____________________________________________ 68

    Magnet OR64 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69Magnet positioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77Seismic anchoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77Quench tube connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 77

    Electronics cabinet (GPA/ACC cabinet) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81Service area . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 82

    RF filter panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85

    MKS 35kW RF amplifier . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89Technical Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90Exclusion zone and service area. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91

    Console components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94Siemens Color Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94Host PC MRC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94Console components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96

    Alarm box . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98

    Magnet stop button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100

    SEP cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 4 of 302Medical Solutions

  • Table of Contents 5Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103

    KKT KCC 215 chiller. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 108Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 114KKT representations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115

    IFP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 122

    Helium compressor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 124Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 125Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 126

    4 _______ On-site electrical installation ____________________________________ 128

    General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128Electromagnetic Compatibility (EMV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128Equipotential bonding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 129

    On-site power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130On-site power specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130Notes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 130On-site power requirement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131Average power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132Mains connection of the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 133

    On-site power distribution panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 134Power distribution panel diagram example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 136Line power on/off; EPO. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 138

    Grounding. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140Grounding points. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 140

    Customer RF filter connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141RF filter for customer use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141RF filter 6A/25A. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 142

    Alarm contact . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143Connection details . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143

    Room lighting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144Intensity of illumination . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144Image screen workstations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144Planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144Reflections, glare . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 144Examination room. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145DC voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145AC voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145Recommendation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 5 of 302Medical Solutions

  • 6 Table of ContentsProtective Conductor Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 146

    5 _______ On-site water supply ___________________________________________ 147

    General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147Water-cooled components. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 148

    On-site water supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149Piping specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149Water quality specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 150

    Primary chilled water supply specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151Primary chilled water supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 151

    6 _______ Air-conditioning _______________________________________________ 153

    General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153Air-conditioning requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153

    Climate conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155Climate specifications - Examination room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 155Climate specifications - Equipment room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156Climate specifications - Operator room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 156Climate specifications - Evaluation room. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 157Absolute humidity - Molier graph . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 158

    7 _______ System connections ___________________________________________ 159

    System connection lengths . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159System cable connection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 159System water hose connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 172System pressurized helium hose connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 179MRC Host PC and Imager PC environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 181MRSC Host PC and Imager PC environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 182Information for using Digital and/or Laser cameras. . . . . . . . . . . . . . . . . . . . . . . . . 182

    Connectivity Concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 183

    Cable routing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184Bending radius for cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184Cable separation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 184Door switches at the RF room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 186

    8 _______ RF Room _____________________________________________________ 187

    RF room prerequisites. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 187Requirements for the RF room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189

    Specifications for the RF room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 189RF room construction (example) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 190RF door opening . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 191RF room interior. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 192Quench tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 193Air conditioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 198Medical gases . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 199Standards used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 200MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 6 of 302Medical Solutions

  • Table of Contents 7RF room . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 201RF filter panel cut-out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 202

    Cu RF room RF filter panel cut-out . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 203Cu RF room floor (example) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 204

    RF room isolation from on-site ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205Measurement devices. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205Preferred method of measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 205

    Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 206RF room manufacturers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207

    RF room manufacturer contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 207

    9 _______ Quench Tube _________________________________________________ 210

    General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210Basis for design calculations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 210

    Quench tube components checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 211Sizing procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 218

    Quench line elbows. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 219Diffuser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 220

    Maximum permissible tube lengths . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221Smooth pipe . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221Flexible convoluted tube . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 221Effective length of bends. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 222

    Quench tube sizing example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223Galvanic separation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 225

    Quench tube - RF room feed-through . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 227Quench tube sketch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228

    Compensator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 228Horizontal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 229Vertical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 230

    Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 231Additional information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 233

    10 ______ Cryogen supply _______________________________________________ 234

    General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 234Structural requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 235Transport methods for helium. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 236Dimensions of the dewars . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 237Delivery of magnet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 238

    The magnet is shipped cold (standard case). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 238Boil-off rate and helium level for the OR64 magnet . . . . . . . . . . . . . . . . . . . . . . . . . . . 239

    Helium content . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 239Helium content . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240Boil-off rate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 240

    11 ______ Options ______________________________________________________ 241Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 7 of 302Medical Solutions

  • 8 Table of ContentsGeneral . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 241MRSC / Leonardo workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 242

    Power requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 242In-room MRC. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244

    Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244Component . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 244For planning purposes, the following are important: . . . . . . . . . . . . . . . . . . . . . . . . 247

    Foot switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 248

    Additional monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 250Patient supervision unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 251

    Functionality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 251Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 251

    Sensor docking station . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254Drawing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 255

    Console table and container . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 257Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 257Container for the Host PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 258

    Surface coils . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 259Patient transport trolley 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 260

    Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 260Adaptation transformer for the SEP cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 261

    Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 261Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 261Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 261

    Power Distribution Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 264

    UPS System for the Host PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 265Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 266Scheme . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 266Connection box for UPS PW 9125 3000i . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 267

    UPS systems for the complete system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 269Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 269Power requirements if the on-site power is provided through a UPS system . . . . . 269UPS system configurations (examples). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 270

    Primary water filter system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 273

    12 ______ Installation and service tools ____________________________________ 274

    Service tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 274Dedicated service tools for MAGNETOM Trio, A Tim System . . . . . . . . . . . . . . . . 274MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 8 of 302Medical Solutions

  • Table of Contents 9Standard service tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 276Magnet transport device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 277

    Magnet transportation tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 279Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 279

    MPS (Magnet Power Supply) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 280Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 280Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 280

    Array Shim Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281Application . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281Transport box dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 281

    13 ______ Information regarding transport__________________________________ 282

    Transport and storage specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 282Mechanical conditions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 282Transport and storage specifications for the magnet . . . . . . . . . . . . . . . . . . . . . . . 282Transport and storage specifications for the system components . . . . . . . . . . . . . 282

    Transport route . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283Requirements for the transport route . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283Issuing the order . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 283

    Transport dimension. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284Magnet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 284GPA/ACC cabinet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285Component weight and dimension . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 285

    14 ______ Technical data ________________________________________________ 286

    Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 286General classification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 286Power requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 286Heating/air conditioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 286Primary water specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 287RF shielding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 288Noise emission . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 288

    15 ______ Trio to Trio, A Tim System Upgrade ______________________________ 289

    General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 289Project manager . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 289Application goals and key advantages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 289System overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 289

    Planning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 290On-site power requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 290On-site water requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 290On-site air-conditioning requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 291Room size requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 291Storage space requirement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 295

    System Cabeling; Upgrade. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 296Cables to be kept on-site for reuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 296Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 9 of 302Medical Solutions

  • 10 Table of ContentsParts handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 298Parts to be kept on-site for reuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 298Surface coils . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 299Parts of the TRIO system not to be used for the upgrade . . . . . . . . . . . . . . . . . . . . 299

    Service tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 301

    16 ______ Changes to the previous version _________________________________ 302MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 10 of 302Medical Solutions

  • General 11

    2- 1GeneralSafety 0

    The safety notes are supplemented by specific descriptions in the documents provided by us, e.g. product documents, ARTD (general guidelines) as well as written instructions for specific cases. All safety instructions have to be observed when performing work activities and tests. In addition, compliance with pre-scribed country-specific requirements has to be ensured (e.g. occupational safety and accident prevention regulations).

    NOTE Read these pages carefully and comply with the safety informa-tion!Read the documents General Safety and MR-specific Safety.

    Controlled access area 0

    NOTE The project manager has to inform the customer to protect thecontrolled access area (0.5mT) with the delivered warning signs.All doors leading into the 0.5mT area must be identified/protectedwith a warning sign. The project manager provides the customerwith the position of the 0.5mT line based on, e.g., a table from theplanning guide that shows the distances or the customer sitedrawings that indicate the 0.5mT line. The warning sign cardiacpacemaker is, e.g., for identifying the 0.5mT line. The warningsign strong magnetic field is, e.g., for identifying the entrancedoors into the examination room or the 0.5mT line.

    The following warning signs are part of the delivery volume.

    Fig. 1: Warning sign PacemakerSiemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 11 of 302Medical Solutions

  • 12 GeneralFig. 2: Warning sign Strong Magnetic FieldFor MR EQUIPMENT that generates a stray field exceeding 0.5 mT outside its perma-nently attached cover and/or an electromagnetic interference level that does not comply with IEC 60601-1-2, the technical description will indicate the necessity to define and permanently install a CONTROLLED

    ACCESS AREA around the MR EQUIPMENT such that outside this area a) the magnetic fringe field strength will not exceed 0.5 mT and b) the electromagnetic interference level complies with IEC 60601-1-2

    will give clear recommendations as to how this CONTROLLED ACCESS AREA will be delimited, e.g. by markings on the floor, barriers and/or other means to allow responsi-ble staff to adequately control access to this area by unauthorized persons; and

    the CONTROLLED ACCESS AREA must be labeled at all entries using the warning signs provided, including indication of the presence of magnetic fields and their force or torque on ferromagnetic materials.

    When the MR EQUIPMENT is designed for installation in a room in which audio visual contact to the PATIENT is likely to be limited, the technical descriptions will specify provi-sions in the design of the room and in the EQUIPMENT to enable audio and visual contact with the PATIENT during the MR EXAMINATION. The audio visual contact will be suffi-cient not to hamper the possible need for MEDICAL SUPERVISION of the PATIENT.When the installation of a CONTROLLED ACCESS AREA is required for the MR EQUIP-MENT, the INSTRUCTIONS FOR USEMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 12 of 302Medical Solutions

  • General 13 will state clearly that it is the responsibility of the USER to follow local statutory require-ments with respect to access to the CONTROLLED ACCESS AREA;

    will specify, preferably accompanied by a sketch, the size and shape of the CON-TROLLED ACCESS AREA;

    will indicate the need to establish adequate rules for controlling access to the CON-TROLLED ACCESS AREA in terms of the potential risk to PATIENTS and staff within the CONTROLLED ACCESS AREA regarding the attraction of objects containing iron or other magnetically active materials or regarding the torque on such metallic materi-als and the potential risk to persons inadvertently entering the area who may be affect-ed by the possible malfunction of medical implants such as pacemakers;

    NOTE For magnetic field strengths less than 0.5 mT, no administrativecontrols are required.

    will list EQUIPMENT and tools specified or recommended by the MANUFACTURER for use in the CONTROLLED ACCESS AREA. For all EQUIPMENT, ACCESSORIES or tools listed, a description should be given of special measures that are needed, if any, for its installation as well as special precautions, if any, for their use;

    will state clearly that peripheral equipment, including PATIENT monitoring, life support-ing devices and emergency care equipment, which are not specified or recommended for use in the CONTROLLED ACCESS AREA, may be disturbed by the radio frequen-cy field or the magnetic fringe field of the MR EQUIPMENT and that this peripheral equipment may also disturb the proper function of the MR EQUIPMENT.

    Bruising hazard sign (Valid for the Upgrade only!) 0NOTE If the distance from the fully extended patient table from the the

    service end to the back wall is less than 500 mm, a label or warningsign has to be posted to identify the hazardous area. This can bedone e.g. with a color-coordinated tape attached to the floor.

    NOTE The bruising hazard sign must be posted in back of the patient ta-ble if the distance between patient table and back wall is less than500 mm.Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 13 of 302Medical Solutions

  • 14 GeneralFig. 3: Bruising hazard sign

    Fig. 4: Bruising hazardPos. 1 LabelMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 14 of 302Medical Solutions

  • General 15Fig. 5: Service end areaPos. 1 Marking on the floorPos. 2 Patient tablePos. 3 Wall of the RF room

    Warning sign Ear Protection 0

    NOTE The MAGNETOM Trio, ATim System is equipped with a Stop-Chocspring kit or Sylomer/Sylodamp pad kit for reducing structural-ly-borne noise.In addition, the covers are equipped with passive noise cancella-tion for reducing airborne noise.

    NOTE The warning sign Ear protection has to be posted on all entrancedoors to the examination room.

    Fig. 6: Hearing protection signThe physical characteristics of the MR system generate a certain amount of noise, emit-ted as contact or acoustic noise. Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 15 of 302Medical Solutions

  • 16 GeneralRegulatory affairs 0

    NOTE The UI MR 020/03/S Quench Tube Inspection has to be per-formed prior to the upgrade!All constructional measures and hardware modifications, requiredas a result of the UI, have to be performed prior to the upgrade.

    NOTE The UI MR 020/03/S Quench Tube Inspection has to be per-formed prior to the upgrade!

    Pressure Equipment Directive 0

    NOTE Local national laws, codes, regulations and procedures have to bemaintained.

    The magnet qualifies as pressure equipment. When starting up and while operating pres-sure equipment, national regulations have to be observed. In Europe, the Pressure Equip-ment Directive (97/23/EU) regulates the market introduction of pressure equipment. In Germany, the Pressure Equipment Directive as well as the "Betriebsicherheitsverord-nung" (BetrSichV)(Industrial Safety Regulations) apply to both system start-up and opera-tion. The BetrSichV stipulates that prior to start-up, the user or others must perform a final acceptance test consisting of the inspection of the documentation, safety equipment, and installation. Periodic inspections of the cryostat are not required. However, periodic inter-nal tests and periodic strength tests are mandatory, if the equipment is taken out of oper-ation for maintenance or repair work.

    Declaration of Conformity The Declaration of Conformity is delivered together with the magnet documenta-

    tion.The CSE, ISE or factory installation member performing the initial installation of the sys-tem has to add the Declaration of Conformity to the System Owner Manual, Register 6, Certificates.The Declaration of Conformity is important for the customer in several countries for acceptance inspections by e.g. the government or any technical inspection authority.

    Pressure Device Regulations 0The magnet is a pressure device. Fuer die Inbetriebnahme und den Betrieb von Druck-geraeten sind die nationalen Vorschriften zu beachten.In Europa regelt die Druckgera-etrichtlinie (97/23/EG) [Pressure Equipment Directive] die Inverkehrbringung des Druckgeraetes.In Deutschland gilt die Druckgeraeteverordnung sowie die fr die Inbetrie-bnahme und den Betrieb der Anlage relevante Betriebssicherheitsverordnung (Betr-SichV). Die BetrSichV schreibt dem Betreiber u.a. eine Abnahmepruefung vor Inbetriebnahme vor, die aus einer Ordnungspruefung, Pruefung der Sicherheitsausrues-MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 16 of 302Medical Solutions

  • General 17tung und Pruefung der Aufstellung besteht. Wiederkehrende Pruefungen des Kryostaten sind nicht vorgeschrieben. Bei Druckgeraeten muessen jedoch wiederkehrende innere Pruefungen und wiederkehrende Festigkeitspruefungen durchgefuehrt werden, wenn sie zu Instandsetzungsarbeiten ausser Betrieb genommen werden.

    Declaration of Conformity The declaration of Conformity is delivered together with the magnet documenta-

    tion.The CSE, ISE or factory installation member which is performing the initial installation of the system had to add the Declaration of Conformity to the System Owner Manual, Register 6, Certificates.The Declaration of Conformity is important for the customer in several countries for acceptance inspections by e.g. the government or any technical inspection authority.Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 17 of 302Medical Solutions

  • 18 GeneralProject Manager Information 0

    Project management 0NOTE The infrastructure for site preparation

    NOTE For each project the actual data/information and the PlanningGuides have to be used!It is the responsibility of the Project Manager to check the Intranetfor the latest version of the Planning Guide.

    Service and installation 0 Information regarding service and installation tools can be found in the following

    chapter: (Installation and service tools / p. 274)

    Installation

    NOTE The transportation, mechanical installation and cabling of the sys-tem and system components have to be performed by trained per-sonnel of our service providers (Hegele, UniData and RemeTronix)only!The system installation, e.g tune-up and start-up will be performedby the US Installation team, the Erlangen Factory installation teamor the HSC team in Erlangen.

    NOTE High priority must be given to preparing the site for a trouble-freereading!This means ensuring that the system components can be installedand connected as soon as possible to achieve a customer- orien-tated installation time and to reduce the amount of liquid heliumboil off during the installation process.

    NOTE All transport locking devices e.g., the red metal beams on top ofthe ACC cabinet or the aluminium mounting blocks for the magnet,have to be stored on-site.The reason for this is the possibility of future location involvingchanging or removing/transporting the system.MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 18 of 302Medical Solutions

  • General 19Environmental protection 0 Refer to the latest version of the Disposal Instructions

    NOTE The transport packing materials, e.g. all port pallets, have to betaken back by the service provider!All other packing materials, e.g. cardboard containers , can be dis-posed of as required by local codes.Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 19 of 302Medical Solutions

  • 20 GeneralSystem component overview 0

    System configuration with SEP 0

    Fig. 7: System component configuration with SEP cabinet

    ACCGPASEP

    Alarm Box

    RF filter plate

    GPA/ACCSEP

    Magnet

    Console

    Astex RF amplifierMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 20 of 302Medical Solutions

  • General 21System configuration with dedicated KKT chiller and IFP 0

    Fig. 8: System component configuration with KKT chiller and IFP

    System Components 0

    Magnet(Function / p. 68)

    System electronics cabinet (GPA/ACC)(Function / p. 80)

    Alarm Box

    Console

    1050

    IFP KKT chillerHelium compressor

    RF filter plate

    ACCGPA

    GPA/ACC

    Magnet

    Astex RF amplifierSiemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 21 of 302Medical Solutions

  • 22 GeneralRF filter plate(Function / p. 84)

    Alarm box(Function / p. 97)

    MRC consoleThe MRC console consists of the HOST PC, monitor, keyboard and mouse used to oper-ate the MR system. In addition, an intercom is part of the console electronics.

    SEP system configuration(Function / p. 102)

    KKT chiller system configuration(Function / p. 108)

    IFP(Function / p. 121)

    Integrated components

    Whole-body gradient systemIn addition to the basic magnetic field, the whole-body gradient system generates linear dynamic magnetic field gradients in the three room orientations. The whole- body gradient system includes the gradient amplifier and the gradient coil.

    Astex RFPAThe Astex RFPA is linear and fully automatic. The Astex RFPA is capable of delivering 35kW at 123.272MHz. Typical gain at 35kW output will be 77.5dB.(Function / p. 89)

    RF systemThe RF system generates the RF fields and transmits them to the patient (transmit func-tion). It also receives the MR signals emitted by the patient and provides them in digital format for image reconstruction (receive system).

    RF coilsThe system includes a whole-body coil for emitting RF to the patient and for receiving MR signals from the patient. The tune box which adjusts the resonator to the performance amplifier as a function of frequency and load as well as the connection cables to these units are part of the body coil. The support tube of the body coil also functions as the inner cover of the magnet tunnel. RF local coils improve image quality when examining specific body regions.RF coil dimensions can be found under the following chapter: (Surface coils / p. 259)MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 22 of 302Medical Solutions

  • General 23Patient handlingPrior to the MR examination the patient is positioned on the table using the positioning accessories for additional comfort. During the examination, the patient is monitored accordingly, e.g. via video camera. The system handling function covers all requirements and includes the patient table, lighting, ventilation, and patient monitoring. Since the patient is usually alone in the examination room during the measurements, patient and physician communicate via the intercom. The intercom includes the configura-tions necessary in the examination room and the operating console as well as the cabling.Whenever required, the patient is monitored during the examination via a video camera. The setup consists of the camera, the monitor, as well as the cabling.Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 23 of 302Medical Solutions

  • 24 Room planning

    3- 2Room planningGeneral 0

    Information for using optical stimulation equipment 0One of the major applications of the MAGNETOM Trio, A Tim Systrem is fMRI (functional imaging). Due to the design, the easiest way to position the optical stimulation equipment may be from the back. This fact has to be considered during planning if the customer will be using fMRI. Additional space is required for the stimulation equipment in the back (ser-vice end) of the magnet (e.g. beamer, panel etc.).

    Due to this fact we recommend a 1.5m to 2.0m distance between the magnet and the back RF wall.

    Consult the user's manufacturer's manual for the stimulation equipment regarding details involving installation aspects as well as space and distance required to the magnet (fringe field).MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 24 of 302Medical Solutions

  • Room planning 25Planning 0

    NOTE All system-related requirements have to meet Siemens MedicalSolutions specifications identified in this document!The specifications are part of this Planning Guide.

    When planning the MR-suite, the following influence the suitability of the site: Interference generated by the magnet Interference affecting the magnet Load capacity Noise emission Minimum door height Minimum room height Cable routing On-site preparation Installation qualitySiemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 25 of 302Medical Solutions

  • 26 Room planningMagnetic field effects on peripheral devices 0The fringe field may affect the functions of devices operated in the vicinity of the magnet. To prevent interference, the magnetic flux density at the location of these devices may not exceed certain values. The flux density acceptable for these devices depends on a num-ber of different parameters. All pertinent information is available through the manufac-turer.

    NOTE The fringe field of the MR system may make its location critical asspecified in the planning documentation. Additional room shield-ing can be calculated and recommended by the ES PM PS planningdepartment in Erlangen.

    Flux density Bmax 0Reference values for acceptable magnetic flux densities Bmax

    Bmax in mT1

    1. 1mT = 10 Gauss

    distance 2

    2. The distance is measured from the magnet isocenter

    Noteradial direction (X/Y) in m

    axial direction (Z) in m

    20 1.7 2.6 Servo ventilator by Siemens

    10 1.9 3.0 RF filter plate

    5 2.2 3.4 MR electronics cabinet

    3 2.4 3.8 Small motors, watches, cameras, magnetic data carriers

    1 2.9 5.0 Processors, magnetic disk drives, oscilloscopes

    0.5 3.4 5.9 Pacemakers, B/W monitors, X-ray tubes, magnetic data carriers, insulin pumps

    0.3 4.0 6.8 Color monitor with active and passive shielding

    0.2 4.6 7.5 CT systems by Siemens

    0.15 5.1 8.2 Computer monitors (PC)0.1 5.9 9.2 Linear accelerators by Siemens

    0.05 7.6 11.2 X-ray image intensifiers, gamma cameras, linear accelerators from other manufacturersMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 26 of 302Medical Solutions

  • Room planning 27Magnet siting requirements 0 The final location selected for the magnet has to ensure that the stability and

    homogeneity of the magnetic field are not adversely affected by extraneous fields. In addition, the magnetic fringe field should not impair the safety of per-sonnel or the functions of other devices in the vicinity of the magnet.

    NOTE At present, a site survey measurement for Trio, A Tim System hasto be performed.Contact the ES PM PS department in Erlangen (Mr. MichaelSchmitz) to arrange this measurement.

    Minimum distances between Siemens Medical Solutions magnets 0

    NOTE It is not possible to install electronics components side-by-side(GPA/ACC/SEP/Astex RFPA) of MR systems with the same operat-ing frequency! Refer to: (Co-Siting / p. 44)The GPA/ACC cabinets and the Astex RFPA have to be separatedby >/= 5 m. The magnetic field of one system has to be ramped to the upperfrequency limit, the other magnetic field to the lower frequencylimit!Shimming is only optimized with both magnets ramped up duringthe shimming procedure.

    The following minimum distances between magnets are valid if: Both magnets are at field. Neither magnet is ramping, while the other is running applications. If one magnet is switched off, the other magnet has to be re-shimmed. Contact the ES PM PS department if you have non-Siemens products or older system

    types, or if you have to plan, e.g., two systems with the same operating frequency close to each other.

    Min. distance of MAGNETOM Trio, A Tim System to:

    the magnet isocenter in m

    1.5T Avanto 7.0

    1.0T Harmony 7.0

    1.5T Symphony 7.0

    1.5T Espree 7.0

    1.5T Sonata 7.0

    3.0T Allegra 10.0Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 27 of 302Medical Solutions

  • 28 Room planningNOTE Shimming is only optimized with both magnets ramped up! Neither magnet is ramping, while the other is running applications.If one magnet is switched off, the other magnet has to bere-shimmed.

    3.0T Trio 10.0

    0.2T Concerto 10.0

    0.2T Jazz 10.0

    0.35T C! 10.0

    Min. distance of MAGNETOM Trio, A Tim System to:

    the magnet isocenter in mMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 28 of 302Medical Solutions

  • Room planning 29Interferences 0Effects on the magnetic field are divided into: static interferences caused by, e.g., ferromagnetic objects, especially those located

    underneath the magnet dynamic interferences caused by, e.g., moving ferromagnetic objects or low frequen-

    cy magnetic fields In critical cases, either Siemens or a third party authorized by Siemens Medical

    Solutions checks the suitability of the site. The tests are designed to detect elec-tromagnetic and mechanical sources of interference.

    Siemens is not responsible for inspecting structural prerequisites, for example, the load capacity of floors or the heating and air conditioning system of the site. Similarly, Siemens is not responsible for executing or monitoring preliminary on-site installation steps. For these reasons, Siemens will not be held liable for the customers failure to maintain specific operating conditions.

    NOTE Static effects can be compensated for by shimming.Dynamic effects are compensated for by maintaining the maxi-mum values indicated.

    NOTE A ferromagnetic iron shielding is required, e.g., if minimum dis-tances to static / moving iron masses or AC/DC sources cannot bemaintained.

    Dynamic interference 0

    Dynamic interference min. distance

    1 in m Notes

    Iron object in radial direction (X/Y)in axial direction

    (Z)< 50 kg 5.5 6.5 With additional iron shielding, it is

    possible to reduce these distances by about 1.0 m!For objects that appear under an angle of 45 to 0 from the Z-axis, the axial values must be applied.The distances are measured from the magnet isocenter to the edge of the object!

    < 200 kg 6.0 7.0

    < 900 kg 6.5 8.0

    < 4500 kg 7.0 9.5Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 29 of 302Medical Solutions

  • 30 Room planningAC interferences 0

    Wheelchair / Bed 5.5 6.5 With additional iron shielding, it is

    possible to reduce these distances by about 1 m!For objects that appear under an angle of 45 to 0 from the Z-axis, the axial values must be applied.The distances are measured from the magnet isocenter to the edge of the object!

    Car 6.5 8.0

    Truck 7.0 9.5

    Train 40.0 40.0

    Subway / Tram 40.0 40.0

    Axiom Artis dFC 30.0 30.0

    1. Distance from the magnet isocenter

    Dynamic interference min. distance

    1 in m Notes

    AC interference min. distance1 in m

    1. Distance from the magnet isocenter

    Notes

    AC Transformer radial direc-tion (X/Y)axial direction (Z)

    < 100 kVA 12.0 8.0The distances are measured from the magnet isocenter to

    the edge of the object!< 250 kVA 12.5 10.0

    < 650 kVA 13.0 12.0

    < 1600 kVA 14.0 15.0

    AC interference min. distance1 in m Notes

    AC Cables radial direc-tion (X/Y)axial direction (Z)

    < 10 A 2.0 2.0

    The distances are measured from the magnet isocenter to

    the edge of the object!

    < 25 A 2.0 2.0

    < 5 A 2.0 2.0

    < 100 A 3.0 2.0

    < 250 A 7.0 3.0

    < 1000 A 12.0 5.0MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 30 of 302Medical Solutions

  • Room planning 31Static interferences 0

    For statical iron (reinforcements) observe the following limitations:

    NOTE The mass of the bottom slab plate should be about 600 kg/m2 (cor-responding to a thickness of about min. 20 cm or 8") to achieve vi-bration and structure-borne sound isolation

    1. Distance from the magnet isocenter

    Static interfer-ences

    min. distance1 in m

    1. Distance from the magnet isocenter

    Notes

    radial direction (X/Y)

    axial direction (Z)

    Water cooling unit 4.0 4.0

    valid for MAGNETOM Trio, A Tim System

    Allowable iron mass consisting of:

    reinforcement steel in the floor

  • 32 Room planningIron room shielding 0 Additional iron shielding is required e.g. if the 0.5mT line has to be within the

    examination room walls, dynamic interferences are present, and public areas have to be protected against the 0.5mT fringe field.

    For additional ferromagnetic iron shielding calculations (if required), contact the planning department of:

    NOTE Existing iron masses will effect the magnetic field and thereforethe image quality of the system. The following information has tobe taken into account.

    NOTE If an additional iron room shielding has to be installed, inform theresponsible structural engineer regarding the additional weightload of the iron for the examination room.

    ES PM PS

    phone ++49 (0) 9131-84-7624 or 7665fax ++49 (0) 9131-84-8632

    Contact PersonsMr. Michael Schmitz; [email protected]

    Mr. Bernhard Heinrich; [email protected]

    Iron masses in the environment

    E.g. existing/required additional iron shielding underneath the magnet

    E.g. iron beams or reinforcement iron at >/= 1.25m distance from the magnet isocenter .

    Contact Mr. Bernhard Heinrich ES PM PS; E-mail: [email protected]

    Local iron mass conditions Required damping solution kit

    Standard site conditions(none of the below described circum-stances takes place)

    Sylomer pad kit(Standard)

    Additional iron shielding, calculated by ES PM PS, underneath the magnet position is required.

    Check with ES PM PS because of magnet isocenter distance to the shielding iron.

    Additional iron shielding underneath the magnet position is already present e.g. from a previous system.

    Check with ES PM PS because of magnet isocenter distance to the shielding iron.MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 32 of 302Medical Solutions

  • Room planning 33Building vibration specifications 0 External vibrations or shocks affecting the magnet may degrade image quality.

    The building must not exceed the following vibrational specifications in the three spatial orientations:

    Tab. 1 Building vibration specifications

    The requirement for amax is -83 dB (g) is measured as a maximum rms value per frequency component > 0.5Hz in the Fourier Transformation of the recorded signal (spectrum).

    Additional information 0 The reference value is the acceleration of gravity (g=9.81m/s2). -83 dB correspond to a factor of 1/14125 amax = 1 mm/s2

    NOTE According to our experience, the mass of the floor plate should beabout 600 kg/m2 (corresponding to a thickness of about min. 20cmor 8") to achieve good vibration and structure-borne sound isola-tion

    NOTE The fringe field of the MR system may make its location critical asspecified in the planning documentation. Additional room shield-ing can be calculated and recommended by the ES PM PS planningdepartment in Erlangen.Refer also to: (Iron room shielding / p. 32)

    AccelerationBuilding vibration

    Specificationsin the frequency range

    amax -83 dB (g) from 0 to 70 HzSiemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 33 of 302Medical Solutions

  • 34 Room planningSite measurement decision matrix 0

    Fig. 9: Site measurement decision matrixMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 34 of 302Medical Solutions

  • Room planning 35Noise emission 0

    NOTE The MAGNETOM Trio, A Tim System is equipped with Sylomer/Sy-lodamp pad kit for reducing structurally-borne noise.In addition, the covers are equipped with passive noise cancella-tion for reducing airborne noise.

    The physical characteristics of the MR system generate a certain amount of noise, emit-ted as contact or acoustic noise. The table below provides you with the information to install the noise attenuation as required, e.g., by law. The attenuation usually incurs additional structural expenditures. The noise level for the various systems averaged across different sequences is shown in the table below.

    Terz-band and octave spectrum 0

    NOTE These spectras are showing the noise behaviour of the MAGNE-TOM Trio, A Tim System for the eight hour average value of 77.9 dB(A) at the In-Room MRC position.

    Fig. 10: Terz-spectrum

    1/3 Octave Synthesis(SPL1)

    Working : Input : Input : FFT Analyzer

    125 250 500 1k 2k

    0

    10

    20

    30

    40

    50

    60

    70

    80

    [Hz]

    [dB(A)/20,0u Pa]

    1/3 Octave Synthesis(SPL1)

    Working : Input : Input : FFT Analyzer

    125 250 500 1000 2000

    0

    10

    20

    30

    40

    50

    60

    70

    80

    [Hz]

    [dB(A)/20,0u Pa]Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 35 of 302Medical Solutions

  • 36 Room planningFig. 11: Octave spectrum

    Regulations 0

    Local regulations (e.g. VGB 121)

    NOTE Refer to local regulations for noise protection.

    1/1 Octave Synthesis(SPL1)

    Working : Input : Input : FFT Analyzer

    31,5 63 125 250 500 1k 2k

    0

    10

    20

    30

    40

    50

    60

    70

    80

    [Hz]

    [dB(A)/20,0u Pa]

    1/1 Octave Synthesis(SPL1)

    Working : Input : Input : FFT Analyzer

    31,5 63 125 250 500 1000 2000

    0

    10

    20

    30

    40

    50

    60

    70

    80

    [Hz]

    [dB(A)/20,0u Pa]

    Allowable noise level for operating personnel1

    1. The noise level is valid for the operating personnel working in the operating room based on VGB121.

    Room Level

    Limiting value related to 20 Pa acoustic pressure

    Operating room

  • Room planning 37Noise emission values 0Tab. 2 Noise value

    Noise cancellation 0

    NOTE If you expect noise problems, the RF room manufacturer you areusing may be able to deliver noise reduction components with theRF room (e.g., interior; RF window; RF door, etc.)

    Allowable noise level for neighbors Level

    Limiting value related to 20 Pa acoustic pressure

  • 38 Room planningFringe field distribution 0

    NOTE The field stability over time is < 0.1 ppm/h.MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 38 of 302Medical Solutions

  • Room planning 39Graphical fringe field distribution 0

    Fig. 12: Fringe field 1.0 mT to 200.0 mTSiemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 39 of 302Medical Solutions

  • 40 Room planningFig. 13: Fringe field 0.1mT to 1.0mTMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 40 of 302Medical Solutions

  • Room planning 41Fig. 14: Fringe field Z-direction; Y-directionSiemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 41 of 302Medical Solutions

  • 42 Room planningFig. 15: Fringe field Z-direction; Y-directionMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 42 of 302Medical Solutions

  • Room planning 43Fringe field distribution in table form 0Tab. 3 Fringe field distribution MAGNETOM Trio, A Tim System

    The fringe field extends spatially in three dimensions around the magnet iso-center and can be reduced by additional iron shielding. The typical fringe field lines represent the ideal flux density distribution in air which may be distorted by steel reinforcements or other iron masses in buildings.

    Fringe fieldDistance from the magnet isocenter in the direction of

    the X-axis in m the Y-axis in m the Z-axis in m

    200 mT x x x

    100 mT 1.25 1.25 1.9

    80 mT 1.3 1.3 1.95

    50 mT 1.4 1.4 2.15

    40 mT 1.45 1.45 2.25

    30 mT 1.55 1.55 2.35

    20 mT 1.65 1.65 2.6

    15 mT 1.75 1.75 2.75

    10 mT 1.9 1.9 3.0

    5 mT 2.15 2.15 3.4

    3 mT 2.35 2.35 3.8

    2 mT x x x

    1 mT 2.9 2.9 5.0

    0.5 mT 3.40 3.40 5.90

    0.3 mT 4.0 4.0 6.75

    0.2 mT 4.6 4.6 7.5

    0.15 mT 5.1 5.1 8.2

    0.1 mT 5.9 5.9 9.2

    0.05 mT 7.6 7.6 11.2Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 43 of 302Medical Solutions

  • 44 Room planningCo-Siting 0

    General 0 Co-siting describes the prerequisites for planning/installing two MR systems

    side-by-side at the same time and with the same operating frequency (field strength). One other scenario could be: A customer already has a 1.5T system in operation and is getting a second 1.5T system to be located in close vicinity of the first one. In this case, the new sys-tem has to meet all requirements specified in this section. The already installed system should be modified whenever possible.

    NOTE It is not possible to install electronic components (GPA/ACC/SEP)side-by-side for systems with the same operating frequency. Addi-tional prerequisites are mandatory!The ACC cabinets have to be separated by 5.0 m (center to center).The magnetic field of one system has to be ramped, according tothe ramping procedure, to the upper frequency limit , the othermagnet to the lower frequency limit!Shimming is only optimized with both magnets ramped up duringthe shimming procedure.

    Magnet distances 0

    Co-siting prerequisites 0

    NOTE The following prerequisites for Co-siting are mandatory for bothMR systems.

    Min. distance between magnets 1

    1. Contact VR P PS for details if the on-site situation requires, e.g., additional iron shielding

    1.0T 1.5T 0.2T 2

    2. With EFI unit, no ramping during measurement; additional iron shielding can reduce distance

    3.0T

    0.2T 5.0 m 6.0 m 10.0 m 10.0 m

    0.35T 5.0 m 6.0 m 10.0 m 10.0 m

    1.0T 4.5 m 5.0 m 5.0 m 7.0 m

    1.5T 5.0 m 5.0 m 6.0 m 7.0 m

    3.0T 7.0 m 7.0 m 10.0 m 10.0 mMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 44 of 302Medical Solutions

  • Room planning 45If two systems with the same operating frequency have to be planned (or one is already installed), take the following into account: The attenuation of the RF room must be 100 dB at the required operating frequency. For additional/optional wave guides, the following calculation applies:Lmin = 4 x diLmin = wave guide length; di = inner tube diameter (di = < 300mm) The two RF rooms cannot be connected electrically. The openings of the RF rooms (e.g., door, window, wave guide) have to be turned

    away from each other. Avoid having the openings (e.g., door, window, wave guide) facing each other.

    The ACC cabinet must be installed at a min. distance of 5.0 m from each other. The cable ducts (ACC side) of the two systems have to be installed separated by 5 m. The short cable set (1.5 m) is not permitted for use with systems in Co-siting, even if

    the ACC cabinet is installed directly underneath the RF filter plate. There must be separate on-site power distribution panels for each system . They

    have to be installed as far away as possible from each other. The RF filter plates have to be positioned as far away as possible from each other

    The mininum distance is 5.0 m. The connection of the RF filters (gradient and power) has to be performed as requested

    in the installation manual (e.g., torque). According to the ramp procedure, one of the MR systems has to be ramped to the up-

    per frequency limit which is 123.265 MHz (3.0T) , the other MR system has to be ramped to approx. 100 kHz lower than the first one.

    NOTE If two 3 T systems are installed side by side:One of the MR system has to be ramped up to the upper frequencylimit of e.g. 123.265 MHz,the other MR system has to be ramped to app. 100kHz less than thefirst one.

    NOTE Shimming is only optimized with both magnets ramped up duringthe shimming procedure.

    Field decay 0 The magnet marked "magnet 2" in the example has to be re-ramped first,

    because the normal magnetic field decay causes the lower operating frequency tolerance limit to be reached earlier.Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 45 of 302Medical Solutions

  • 46 Room planningNOTE The special ramping procedure could require a different re-ramp-ing time interval (to the operating frequency after the normal fielddecay) of the two systems!

    Fig. 16: Co-Siting frequency settings

    NOTE If two 3 T systems are installed side by side:One of the MR system has to be ramped up to the upper frequencylimit of e.g. 123.265 MHz,the other MR system has to be ramped to app. 100kHz less than thefirst one.

    ACC distance for systems with the same operating frequency 0

    NOTE The distance from the isocenter of the electronics cabinet to eachother must be min. 5.0m.MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 46 of 302Medical Solutions

  • Room planning 47Fig. 17: ACC installation with two systems at the same operating frequency

    Magnet orientation 0

    NOTE Two magnets with the same operating frequency positioned in "Z"direction should be avoided due to the required distances and in-creased RF coupling between the systems! Contact ES PM PS inthis case.Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 47 of 302Medical Solutions

  • 48 Room planningFig. 18: 2 magnets line up in Z axis

    NOTE It is permitted to position two magnets with the same operating fre-quency in the "X" direction.MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 48 of 302Medical Solutions

  • Room planning 49Fig. 19: Magnet parallel installation

    NOTE It is permitted to position two magnets with the same operating fre-quency at an angle to each other.

    Fig. 20: 2 magnets 90 to each otherSiemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 49 of 302Medical Solutions

  • 50 Room planningRoom configurations and sizes 0

    Fig. 21: Example of room layoutMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 50 of 302Medical Solutions

  • Room planning 51Room configuration example 0

    Fig. 22: Minimum finished room dimensions

    Building wall RF-enclosure Finished room wall

    min. 3800

    min.6380

    min. 500 safety distance

    min. 800 trolley distance

    min.1000

    min.2070

    min.600

    service

    cleara

    nce

    min.600

    service

    cleara

    nce

    min. 1970Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 51 of 302Medical Solutions

  • 52 Room planningFig. 23: Example of room layout

    140 above finished floorAlarm Box

    180 above finished floorMagnetstop

    RF door opens to the outside

    Examination Room

    Customer Storage

    Preparation

    Equipment Room

    Operator Room

    RFPA

    GPA

    MRC ConsoleMR Host Computer

    ACC

    Carry-in opening for the magnet

    Domesticelectricalinstalla

    tion

    RF-filter panel

    Main Distribution SwitchboardMR System

    SEP

    Preferredlocationofthe

    RFfilterpanel

    Block off fence for the 0.5mT line(if required)

    min. 1970

    min. 600

    min.800

    min.2070MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 52 of 302Medical Solutions

  • Room planning 53Fig. 24: Example of room layout

    MR-Examination Room

    EDV

    TEL

    HZ

    EL

    Preparation

    Equipment Room

    Operating Room

    GPAACC SEP

    1.01

    1.02

    315

    320

    232.5

    Astex

    294

    Alarm-Box

    Operator console

    RF filter panel

    Remote Magnet Stop buttonSiemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 53 of 302Medical Solutions

  • 54 Room planningNOTE If the patient cannot be observed (visually) through an RF windowbecause of restrictions in the room layout, the Patient Video optionhas to be included in the room planning. Audible contact with thepatient is provided via the built-in intercom system (e.g. loud-speaker, pneumatic headphone).

    NOTE For installation and service (ramping the magnet), an area of ap-prox. 1.0m2 has to be provided for the MPS and cables in the equip-ment room.

    NOTE The possibility of refilling the magnet with liquid helium has to beguaranteed. As a result, space has to be provided for the heliumdewars.The min. distance between patient table and RF cabin wall has tobe suitable for He- dewars.

    NOTE The RF room door has to be opened to the outside, if this is e.g.not possible due to building restrictions, refer to the register onthe RF-room for details (Alternatives for a pressure equalizationopening / p. 192).

    NOTE Ensure that no cable ducts, light fittings, air conditioning ducts orany other component is installed above the cold head and the he-lium syphon port.Keep those areas clear of components.

    Distance of the magnet to the rear wall

    Maximum patient table travel distance in the longitudinal direction is 2680mm

    Maximum patient table excess length behind the magnet cover is 500mm

    A distance from the back magnet cover to the back wall of 1000mm is recom-mended in any case and mandatory for the whole-body suite option.Minumum required distance from the back magnet cover to the back wall for service purposes (e.g. changing the patient tabletop) is xxxxmm.The distance from the magnet isocenter to the back wall of 2065mm is recom-mended in any case and mandatory for the whole-body suite option.MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 54 of 302Medical Solutions

  • Room planning 55Fig. 25: Required minimum finished room length

    Patient table travel

    Vertical patient table movement 570mm to 1000mm

    Longitudinal table movement 2680mm

    Table excess length behind the magnet for whole body suite 500mm

    Max. scan range with whole body suite 1810mm 1

    1. 1400mm without the option Tim Whole Body Suite

    Minimum required service distance from the back magnet cover to the wall

    1000mm

    Minimum required service distance from the magnet isocenter to the wall

    2065mm

    Minimum required distance from the back magnet cover to the wall including a 500 mm safety distance.

    1000mm

    Minimum required distance from the magnet isocenter to the wall including 500 mm safety distance.

    2065mm

    Minimum required distance in front of the magnet from the fully extended patient table to the wall for using the Patient table trolley option.

    800mm

    800500500

    4730

    6380

    2065

    4580Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 55 of 302Medical Solutions

  • 56 Room planningMinimum technical finished room size 0

    Minimum distancesRoom planning for MR systems must be based on the minimum distances specified for safety and installation-related reasons. In the RF room, the subfloor together with the floor cannot exceed the tolerance values specified.

    NOTE If the system is designated for using the 2nd tabletop with trolley,a distance of 800 mm, in front of the patient table fully extended tothe wall, has to be maintained

    NOTE The minimum distance from the back magnet cover to the backwall has to be 1000 mm in any case. This is needed because of ser-vice requirements.

    Minimum technical room height

    NOTE The minimum dimensions provided present the minimum techni-cal requirements.It is the decision of the customer to increase these dimensions.

    NOTE The minimum finished room height measures from the finishedfloor to the finished (e.g., suspended) ceiling.To service the magnet (e.g., cold head, current probe, etc.), theminimum finished room height must be guaranteed.

    NOTE The system is delivered with a built-in set of Sylomer/Sylodamppad elements for structurally-borne sound reduction.

    NOTE If, e.g., additional iron room shielding or noise damping has to bebuilt into the RF room, take the size of the noise damping into con-sideration to achieve the min. finished room sizes!

    Minimum technical finished room height 1 Minimum technical finished room size 2 in l x w

    Examination room 2700 mm 3800 mm/6380 mm

    Control room 2200 mm 2000 mm/2500 mm

    Equipment room 2200 mm 2810 mm/1650 mmMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 56 of 302Medical Solutions

  • Room planning 571. The finished room height is the distance from the finished floor (e.g., anti-static linoleum floor) to thefinished ceiling (e.g., suspended ceiling). These values are technical measures.

    2. The minimum finished room size is measured from the finished wall to finished wall.Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 57 of 302Medical Solutions

  • 58 Room planningMagnet floor loading 0

    Fig. 26: Magnet footprint

    Magnet floor loading

    Total magnet weight ~ 13.000kg

    Sylomer/Sylodamp pad kit4 support feet each 200 mm X 600 mm(each 1200cm2)3250kg/1200cm2

    Total floor loading ~ 127.5kN

    Foot position 1 31.88 kN

    Foot position 2 31.88 kN

    Foot position 3 31.88 kN

    Foot position 4 31.88 kN

    Patientend

    flush

    1042

    600

    200

    Serviceend

    1360

    flush

    2010

    1725

    1276

    877

    105 180

    80 x 80MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 58 of 302Medical Solutions

  • Room planning 59Fig. 27: Configuration under the box sectionPos. 1 Shim plates for leveling the magnetPos. 2 Sylodamp platePos. 3 Sylomer plate

    Fig. 28: Configuration under the box section

    Pos. 1 GRP plate and shim plates for leveling the magnetPos. 2 Sylodamp platePos. 3 Sylomer platePos. 4 Box section

    NOTE The floor in the vicinity of the magnet (2.6 m x 4.8 m) has to be leveled within max. +/-2.0 mm.Siemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 59 of 302Medical Solutions

  • 60 Room planningSeismic anchor positions 0

    Fig. 29: Seismic bracket position for the magnetMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 60 of 302Medical Solutions

  • Room planning 61Fig. 30: Seismic bracket position for the magnetSiemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 61 of 302Medical Solutions

  • 62 Room planningRF filter plate installation 0

    NOTE The position of the RF filter panel should be located in the area be-hind the magnet isocenter.MAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 62 of 302Medical Solutions

  • Room planning 63Recommended installation 0

    Fig. 31: ACC/GPA position in relation to the RF filter platePos. 1 ACC/GPAPos. 3 RF roomSiemens AG M6-030.891.01.01.02 MAGNETOM05.05 CS SD 23

    Page 63 of 302Medical Solutions

  • 64 Room planningAlternative installation 0Due to building requirements or other circumstances, it may be necessary to install the RF filter plate in other positions.

    Fig. 32: RF filter plate at ACC/GPA heightMAGNETOM M6-030.891.01.01.02 Siemens AG05.05 CS SD 23

    Page 64 of 302Medical Solutions

  • Room planning 65Pos. 1 ACC/GPAPos. 5 Minimum distance of ACC/GPA to RF room if (6) is 1,870 mm or lowerPos. 6 RF filter plate height examplePos. 7 The 1.5 m cable set will cover this configuration. If the distance (5) is more than 850 mm, the next longer cable

    set has to be used.Pos. 8 The minimum distance of 200 mm to the ceiling is to achieve the bending radius of the gradient cables.

    NOTE The electronic cabinet could also be positioned either to the left orto right of the filter plate. A longer cable set h