liquid waste analysis - bca...garry smith, xrf application specialist scimed xrf, a division of...
TRANSCRIPT
-
Liquid Waste Analysis
Garry Smith, XRF Application Specialist
SciMed XRF, a division of Scientific and
Medical Products Ltd
-
About Us
• SciMed represent Rigaku (RESE and
ART) and Seiko (SIINT) XRF ranges
• Over 20 years history of working with XRF
• A wide range of UK installations across
many types of applications
-
Summary
• What are the difficulties with liquid waste
samples and how does the sample
introduce error into the test result?
• Examples of errors induced on analysis:
– Suspended solids and settling
– Immiscible liquids and separation
• What are the practical solutions?
-
What are the difficulties?
• Liquid waste samples are rarely “ideal” as
they are rarely homogeneous.
• May contain solids which settle over time.
• May be mixtures of immiscible liquids
which separate over time.
-
An Example
Example photos of multiphase sample before,
during, and after separation and settling.
-
Why is this a Problem?
• For any analysis, the measured sample
must be representative of the bulk material
being tested.
• With XRF we are not always measuring
everything in the sample cell.
• If components separate during analysis
they may not be measured accurately.
-
Critical Depth
• Fluorescent photons from analytes are re-absorbed by the sample.
• Absorption follows the Beer-Lambert law: I = I0e
-μρx
where: μ = mass absorption coefficient ρ = matrix density x = path length
-
Critical Depth
• Critical Depth is defined as the depth of sample from which 99% of the fluorescent photons are re-absorbed.
• To calculate we can re-arrange Beer-Lambert to express path length, x, where:
Io = 100 I
-
Critical Depth
• The equation simplifies to:
Critical depth, x(mm) = 46.605 μρ
• Critical depth increases with increasing photon energy.
• Critical depth decreases with increasing average atomic number and density of the sample.
-
Critical Depth
Liquid Sample in Cell
Incident X-rays
Cl
Cl
Fluorescent photon to detector
Fluorescent photon re-
absorbed by sample
Critical depth of Cl Kα
Critical depth of Zn Kα
Zn
-
Critical Depth
Line Energy (keV) Typical Critical Depth (mm)
Oil Water
S Kα 2.31 0.36 0.12
Cl Kα 2.62 0.50 0.17
Cr Kα 5.41 4.33 1.44
Zn Kα 8.64 16.3 5.42
Pb Lα 10.55 27.9 9.74
Typical critical depths of selected XRF lines in oil and water
-
Example 1 – Effect on Peak Intensity of Analyte by
Settling of Sediment (Analyte in Solution)
Overlay view of spectra for S in heavily sedimented oil measured at 2
minute intervals, showing attenuation of peak as sediment settles. The
analyte is concentrated in the solution.
-
Example 1 – Effect on Peak Intensity of Analyte by
Settling of Sediment (Analyte in Solution)(cont.)
Overlay view of corresponding spectra for Zn in in the same sample.
However there is virtually no change in intensity with time. The higher
energy of Zn Kα has greater critical depth in the sample. Intensity is
less variable with distribution throughout sample.
-
Example 1 – Effect on Calculated Concentration of
Analytes by Settling of Sediment (Analyte in
Solution)
150
160
170
180
190
200
210
220
230
240
250
3000
3500
4000
4500
5000
5500
6000
0 2 4 6 8 10 12 14 16 18
Me
as
ure
d Z
n (
pp
m)
Me
as
ure
d S
(p
pm
)
Time after mixing (min)
Measured S and Zn Concentration vs. Time for Waste Oil
-
Example 2 – Effect on Peak Intensity of Analytes
by Partitioning of Liquids (Aqueous Analyte)
Overlay view of spectra for Cl measured at 2 minute intervals after
mixing, showing enhancement of measured peak as aqueous
component containing chloride separates to bottom of liquid cell.
-
Example 2 – Effect on Peak Intensity of Analytes
by Partitioning of Liquids (Oil Analyte)
Overlay view of corresponding spectra for Cr measurement. Cr is
concentrated into the oil component and is attenuated as this layer
separates to the top of the liquid cell.
-
Example 2 – Effect on Calculated Concentration of
Analytes by Partitioning of Liquids (Analytes in Both
Phases)
150
170
190
210
230
250
45000
50000
55000
60000
65000
70000
75000
80000
85000
90000
0 2 4 6 8 10 12 14
Me
as
ure
d C
r (p
pm
)
Me
as
ure
d C
l (p
pm
)
Time after mixing (min)
Measured Cl (aqueous), and Cr (oil) vs. Time in Oil / Aqueous Mixture
Measured Cl (ppm) Measured Cr (ppm)
-
Summary
• Effect on measured intensity may be attenuation or enhancement depending on circumstances.
• Rate is dependent on the specific sample, e.g. particle size, viscosity, etc.
• Amount of error varies with analyte – higher energy lines (usually heavier elements) affected less than lower energy lines (usually lighter elements).
-
In the case of real samples, these effects
are in practice impossible to predict.
-
What are the Solutions?
• Ensure analysis is completed before
phases separate (approximate analysis)
• Separate phases to remove unwanted
components or analyse separately
• Stop phases from separating
– Use solid binders to immobilise components
-
Aqueous Samples
• Aqueous samples containing solids can be dealt with quite easily:
– Allow sample to settle or centrifuge sample and pipette supernatant liquid (dissolved components only).
– Filter sample and analyse separate components as required.
Note – can also be applied to organic liquids.
-
Immiscible Liquids
• Oil / hydrocarbon samples may contain
water.
• While it may be possible to use emulsifier
to stabilise, no universal solution for all
sample types.
• More robust solutions:
– Separation
– Immobilisation using binder
-
Separation of Liquids - Centrifuging
• Centrifuging can provide a simple approach to separating aqueous component from oils and hydrocarbons
• Also removes solids
• Component required for analysis can simply be pipetted after centrifuging
Reference: ASTM D4294 – Sulphur in Petroleum and Petroleum Products by Energy Dispersive X-Ray Fluorescence Spectrometry.
-
Immobilisation Using Binder
• Separation of liquids or solids not always a
practical solution (especially for waste oils)
• Both immiscible liquids and solids can be
immobilised by mixing the sample with a
powder binder material
– Graphite powder
– Activated alumina
-
Graphite Powder
• Fine graphite powder provides an inert and spectrally pure carrier material for oil / hydrocarbon samples
• Typically mix 10g sample with 4-6g graphite to produce a slurry
• Transfer to liquid sample cell and measure
Reference: ASTM D5839 – Trace Element Analysis of Hazardous Waste Fuel by Energy Dispersive X-Ray Fluorescence Spectrometry.
-
Activated Alumina
• Activated alumina provides a stable carrier
material for aqueous / organic solvent samples
• Typically mix 5g sample with 15g activated
alumina to produce a slurry
• Transfer to liquid sample cell and measure
Reference: ASTM D6052-97 – Preparation and
Elemental Analysis of Liquid Hazardous Waste by
Energy Dispersive X-Ray Fluorescence.
-
In Closing…
• Multiphase liquid samples can introduce significant errors into analysis results if not treated appropriately
• These errors are a result of the potential for phases to separate during measurement and the varying critical depth of different analytes.
• A number of potential solutions to remove or neutralise sources of error. Standard methodology often incorporates these precautions.