lesson4p3
TRANSCRIPT
-
7/28/2019 Lesson4p3
1/8
Hardness Testing
For information regarding this presentation, please contact Dan Emelander (Metallurgist) 713-939-2672
Hardness Testing Equipment
Part 3: Reading a Brinell Impression
-
7/28/2019 Lesson4p3
2/8
Reading the Impression
Brinell Scopes
Light source MUST be attached!
Computerized OpticalScanner
-
7/28/2019 Lesson4p3
3/8
Comparison of Equipment
Advantages
Relatively inexpensive More portable
More durable
Disadvantages
High variability (~10% ofBrinell value)
Can only read at 0.05mmincrements
Inspectors sometimes donot measure 2 diameters
Advantages
Low variability, only 2% ofBrinell value
Can read to 0.01mm
Always reads 8+ diameters
Disadvantages Less portable (but still very
portable)
Higher cost; ~$4000
Less durable
Brinell ScopesComputerized Optical
Scanner
-
7/28/2019 Lesson4p3
4/8
Using the Brinell Scope
Turn on the flashlight, or lighting system The flashlight should be attached to the scope! Holding the
light, or not having a light is unacceptable
Place the scope on the part No rocking should take place, it should rest flat
Read the impression Measure 2 diameters
Diameters are to be 90 apart
Average the diameters, and use the table listed inASTM E10 to convert the diameter into a HBWnumber
-
7/28/2019 Lesson4p3
5/8
Oval Indentations
Cameron requires that when the difference betweentwo diameters is greater than 0.1mm, it is invalid
Oval indentations are a key indicator that the setup ofthe tester was not acceptable
Oval indentations are most often caused by: Bottom support not directly below test location
Tester not perpendicular to the test piece
Movement during the test
Test surface not flat
-
7/28/2019 Lesson4p3
6/8
Which Indentation is Invalid?
4.15
4.20
A. B.
C. D.
4.10
4.25
4.15
4.15
4.15
4.25
Difference =0.05
Difference =0.10 Difference =0.00
Difference =0.15
-
7/28/2019 Lesson4p3
7/8
Using the Optical Scanner
Once the scanner is turned on and calibrated: Place the scanner over the indentation
The scanner should rest flat on the surface, no rocking
should occur
Make sure the entire indentation is visible on the screen, andpress the button
The software will now display the hardness value on thescreen
It will also kick out oval readings, or if the surface is too roughto get a good reading
-
7/28/2019 Lesson4p3
8/8
Reporting a Brinell Hardness
The standard Brinell hardness number shall bereported using 3 significant numbers, followed byHBW.
Example: 237 HBW Non-standard tests shall be reported using the above
notation, followed by the items below, separated by a/: The load used
The ball diameter used (mm)
The time the load was held
300 HBW 1500 / 5 / 7 A 300 HBW was obtained using a 1500kg load, a 5mm
tungsten carbide ball, and the load was applied for 7 seconds.