la neta de la goniometria
TRANSCRIPT
GONIOMETRYa. Definition - measuring the available range of motion or the position of a joint
- typically this is a measure of PASSIVE motion. If you are documenting active range of motion, document that this is so.
b. For clarity of communication – measure one direction at a time (e.g. elbow flexion = 130o, not elbow flexion/extension 130o/0o)
GONIOMETER PARTSStandard (universal) Goniometer
Gravity Goniometer (inclinometer)
GONIOMETRY PROCEDURE a. Position joint in zero position and stabilize proximal joint component b. Move joint to end of range of motion (to assess quality of movement) c. Determine end-feel at point where measurement will be taken (at end of
available range of motion) d. Identify and palpate bony landmarks e. Align goniometer with bony landmarks while holding joint at end of range f. Read the goniometer g. Record measurement (e.g. elbow flexion = 130o)
POSITIONING
a. Start with joint at zero position - This is the reference point for the measurement. If zero position can't be achieved, this must be documented.
b. permit complete range of motion
1) If you are assessing joint ROM, be sure that some other structure (eg. a tight muscle) doesn't interfere.
2) If you are assessing some other structure (eg. a tight muscle, pain limiting the motion) document exactly what is limiting the range of motion. (eg. hamstring tightness at 65o of hip flexion)
STABILIZATION a. Poor stabilization is the most frequent cause of invlaid measurements. (eg. observe a "normal" ROM of elbow extension when movement of shoulder and arm masks a limitation - actually measuring shoulder and arm movement)
Poor Stabilization for Elbow Extension
b. Usually stabilize proximal joint components
c. Promote patient relaxation so voluntary muscle contraction doesn't interfere
VALIDITY
a. Validity is a measurement concept that asks whether a measurement system actually measures what it's supposed to (i.e., joint range of motion in the case of goniometry)
b. Goniometric measurements can be invalid; usually because of poor stabilization. (See positioning and stabilization)
RELIABILITY a. Reliability is a measurement concept that asks whether successive measurements are consistent, repeatable or reproducible. Upper extremity measurements are more reliable than lower extremity measurements.
1) Intratester reliability (same tester on different occasions)- measurement error should be less than 5 degrees
2) Intertester reliability (different testers) - measurement error probably greater than 5 degrees
b. To maximize reliability always use the same: 1) Goniometer 2) Positioning 3) Procedure 4) Examiner
END FEEL
a. The quality of resistance at end of range
b. Each joint has a normal end feel at a normal point in the range of motion (ROM)
c. Incorrect end feel, or correct end feel at incorrect ROM indicate pathology
d. Terms: I suggest my terms for clarity of communication (to clearly identify the structure that the tester feels is limiting the ROM). Other authors use different terms - eg. hard, firm, soft, etc. I feel these "vague" terms lead to communication errors.
Capsular - indicates that the joint capsule is limiting the ROM. Feels like stretching a leather belt. Example - knee extension.
Ligamentous - indicates that ligament tightness is limiting the ROM. Feels like stretching a leather belt. Example - wrist radial deviation.
Bony - indicates that bone touching bone is limiting the ROM. Feels like pushing two wooden surfaces together. Example - elbow extension.>
Muscle Stretch - indicates that muscle tightness is limiting the ROM. Feels like stretching a bicycle tire innertube. Example - hip flexion while maintaining knee extension (straight leg raise) when hamstrings are tight.
Soft Tissue Approximation - indicates that subcutaneous tissues (muscle bulk, fat) are pushing against each other and limiting the ROM. Feels like squeezing two balloons together. Example - calf pressing against thigh during knee flexion.
Springy - indicates that a loose body is limiting the ROM. Feels "bouncy" like you are compressing a spring. Example - torn meniscal (knee) tissue limiting knee extension.
Empty - indicates that the examiner did not reach the end feel (usually the patient is not willing to allow motion to end of range because of anticipated pain). Feels like the joint has more range available, but the patient is purposefully preventing movement through the full ROM.
Testing End Feel for Elbow Extension
EXTREMIDAD SUPERIOR
SHOULDER FLEXION
Test Position
Subject supine Flatten lumbar spine (flex knees) Shoulder no abduction, adduction
or rotation (Note: to measure gleno-humeral
motion, stabilize scapula)
Normal Range
(For shoulder complex flexion)
167o + or - 4.7o (American Academy of Orthopaedic Surgeons)
150o (American Medical Association)
166o (mean), 4.7o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – center of humeral head near acromion process
Stationary arm – parallel mid-axillary line
Moving arm – aligned with midline of humerus (lateral epicondyle)
Normal End Feel
Muscle Stretch
EXTENSION DEL HOMBRO
Test Position
Subject prone Shoulder no abduction, adduction
or rotation (note: to measure gleno-humeral
motion, stabilize scapula)
Normal Range
(for shoulder complex flexion)
62o + or - 9.5o (American Academy of Orthopaedic Surgeons)
50o (American Medical Association)
62.3o (mean), 9.5o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – center of humeral head near acromion process
Stationary arm – parallel mid-axillary line
Moving arm – aligned with midline of humerus (lateral epicondyle)
Normal End Feel
Capsular or ligamentous
ABDUCCION DEL HOMBRO
Test Position
Subject supine Shoulder 0o flexion and extension Shoulder laterally (externally)
rotated Shoulder abducted Stabilize thorax (note: to measure
gleno-humeral motion, stabilize scapula)
Normal Range
(for shoulder complex abduction)
184o + or - 7.0o (American Academy of Orthopaedic Surgeons)
180o (American Medical Association)
184o (mean), 7.0o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – center of humeral head near acromion process
Stationary arm – parallel to sternum
Moving arm – aligned with midline of humerus
Normal End Feel
Muscle Stretch
ROTACION INTERNA DEL HOMBRO
Test Position
Subject supine Shoulder 90o abduction Forearm neutral Elbow flexed 90o Stabilize arm
Normal Range
69o + or - 4.6o (American Academy of Orthopaedic Surgeons)
90o (American Medical Association)
68.8o (mean), 4.6o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – olecranon process of ulna Stationary arm – aligned vertically
Moving arm – aligned with ulna (styloid process)
Normal End Feel
Capsular
ROTACION EXTERNA DEL HOMBRO
Test Position
Subject supine Shoulder 90o abduction Forearm neutral Elbow flexed 90o Stabilize arm
Normal Range
104o + or - 8.5o (American Academy of Orthopaedic Surgeons)
90o (American Medical Association)
103o (mean), 8.5o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – olecranon process of ulna Stationary arm – aligned vertically
Moving arm – aligned with ulna (styloid process)
Normal End Feel
Capsular
FLEXION DEL ANTEBRAZO
Test Position
Subject supine Shoulder neutral (arm at side) Forearm supinated Elbow flexed Stabilize arm
Normal Range
141.0o + or - 4.9o (American Academy of Orthopaedic Surgeons)
140.0o (American Medical Association)
142.9o (mean), 5.6o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – lateral epicondyle of humerus
Stationary arm – aligned with humerus (center of acromion process)
Moving arm – aligned with radius (styloid process)
Normal End Feel
Soft tissue approximation (capsular for thin subjects)
EXTENSION DEL CODO
Test Position
Subject supine Shoulder neutral (arm at side) Forearm supinated Elbow extended Stabilize arm
Normal Range
0.3o + or - 2.0o (American Academy of Orthopaedic Surgeons)
0.0o (American Medical Association)
0.6o (mean), 3.1o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – lateral epicondyle of humerus
Stationary arm – aligned with humerus (center of acromion process)
Moving arm – aligned with radius (styloid process)
Normal End Feel
Bone on bone
FOREARM SUPINATION
Test Position
Subject sitting Shoulder neutral (arm at side) Elbow flexed to 90o Stabilize arm
Supinate forearm
Normal Range
81o + or - 4.0o (American Academy of Orthopaedic Surgeons)
80o (American Medical Association)
82.1o (mean), 3.8o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – medial to ulnar styloid Stationary arm – parallel to
Normal End Feel
Capsular
humerus
Moving arm – aligned with ventral aspect of radius
FOREARM PRONATION
Test Position
Subject sitting Shoulder neutral (arm at side) Elbow flexed to 90o Stabilize arm
Pronate forearm
Normal Range
75o + or - 5.3o (American Academy of Orthopaedic Surgeons)
80o (American Medical Association)
75.8o (mean), 5.1o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – lateral to ulnar styloid Stationary arm – parallel to
humerus
Moving arm – aligned with dorsum of radius
Normal End Feel
Capsular
WRIST FLEXION
Test Position
Subject seated Forearm stabilized on table Flex wrist (fingers relaxed)
Normal Range
75o + or - 6.6o (American Academy of Orthopaedic Surgeons)
60o (American Medical Association)
76.4o (mean), 6.3o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – lateral wrist (triquetrum)
Normal End Feel
Stationary arm – aligned with ulna
Moving arm – aligned with fifth metacarpal
Capsular
WRIST EXTENSION
Test Position
Subject seated Forearm stabilized on table Extend wrist (fingers relaxed)
Normal Range
74o + or - 6.6o (American Academy of Orthopaedic Surgeons)
60o (American Medical Association)
74.9o (mean), 6.4o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – lateral wrist (triquetrum)
Normal End Feel
Capsular
Stationary arm – aligned with ulna
Moving arm – aligned with fifth metacarpal
WRIST RADIAL DEVIATION
Test Position Subject sitting with forearm resting
on table
Stabilize forearm to prevent pronation or supination
Normal Range 21o + or - 4o (American Academy
of Orthopaedic Surgeons) 20o (American Medical
Association)
21.5o (mean), 4.0o (standard deviation), (Boon and Azen)
Goniometer Alignment Axis – capitate Stationary arm – aligned with
forearm (lateral epicondyle)
Moving arm – aligned with metacarpal of middle finger
Normal End Feel Ligamentous (ulnar collateral
ligament)
WRIST ULNAR DEVIATION
Test Position
Subject sitting with forearm resting on table
Stabilize forearm to prevent pronation or supination
Normal Range
35o + or - 3.8o (American Academy of Orthopaedic Surgeons)
30o (American Medical Association)
36.0o (mean), 3.8o (standard deviation), (Boon and Azen)
Goniometer Alignment
Axis – capitate Stationary arm – aligned with
forearm (lateral epicondyle)
Moving arm – aligned with metacarpal of middle finger
Normal End Feel
Ligamentous (radial collateral ligament)
METACARPOPHALANGEAL JOINT FLEXION
Test Position
Subject sitting with forearm resting on table
Wrist and interphalangeal joints relaxed
Forearm neutral
Stabilize metacarpal to prevent motion
Normal Range
86o (index), 91o (long), 99o (ring), 105o (little) (American Academy of Orthopaedic Surgeons - active motion)
90o (American Medical Association)
Goniometer Alignment
dorsal metacarpophalangeal joint Stationary arm – aligned with
metacarpal
Moving arm – aligned with proximal phalange
Normal End Feel
capsular
METACARPOPHALANGEAL JOINT EXTENSION
Test Position
Subject sitting with forearm resting on table
Wrist and interphalangeal joints relaxed
Forearm neutral
Stabilize metacarpal to prevent motion
Normal Range
22o (index), 18o (long), 23o (ring), 19o (little) (American Academy of Orthopaedic Surgeons - active motion)
20o (American Medical Association)
Goniometer Alignment
dorsal metacarpophalangeal joint Stationary arm – aligned with
metacarpal
Moving arm – aligned with proximal phalange
Normal End Feel
capsular
METACARPOPHALANGEAL JOINT ABDUCTION
Test Position
Subject sitting with forearm resting on table
Wrist neutral Forearm neutral
Stabilize metacarpal to prevent motion
Normal Range
???
Goniometer Alignment
dorsal metacarpophalangeal joint Stationary arm – aligned with
metacarpal
Moving arm – aligned with proximal phalange
Normal End Feel
capsular
METACARPOPHALANGEAL JOINT ADDUCTION
Test Position
Subject sitting with forearm resting on table
Wrist neutral Forearm neutral
Stabilize metacarpal to prevent motion
Normal Range
???
Goniometer Alignment
dorsal metacarpophalangeal joint Stationary arm – aligned with
metacarpal
Moving arm – aligned with proximal phalange
Normal End Feel
capsular
INTERPHALANGEAL JOINT FLEXION
Note: This page demonstrates the technique for index proximal interphalangeal joint flexion. The technique for all other interphalangeal joints is similar. Simply align the goniometer over the proximal and distal joint partners (bones) for the joint you wish to measure.
Test Position
Subject sitting with forearm resting on table
Wrist, metacarpal, and non-tested interphalangeal joints relaxed
Forearm neutral
Stabilize proximal bone to prevent motion
Normal Range
American Academy of Orthopaedic Surgeons
PIP fingers - 102o (index), 105o (long), 108o (ring), 106o (little) ( active motion)
DIP fingers - 72o (index), 71o (long), 63o (ring), 65o (little) ( active motion)
IP thumb - 73o
American Medical Association
100o (PIP fingers), 70o (DIP fingers), 80o (IP thumb)
Normal End Feel
Goniometer Alignment
dorsal proximal interphalangeal joint
Stationary arm – aligned with proximal phalange
Moving arm – aligned with middle phalange
Proximal Interphalangeal Finger Joints
bone on bone (if tissues overlying palmar aspect of bones is thin)
soft tissue approximation (if tissues overlying palmar aspect of bones is thick)
Distal Interphalangeal Finger Joints and Thumb Interphalangeal Joint
capsular
THUMB CARPOMETACARPAL JOINT FLEXION
Test Position
Subject sitting with forearm supinated and resting on table
Wrist neutral
Stabilize carpals to prevent wrist motion
Normal Range
???
Goniometer Alignment Normal End Feel
Axis – carpometacarpal joint Stationary arm – aligned with
radius
Moving arm – aligned with metacarpal of thumb
Capsular
THUMB CARPOMETACARPAL JOINT EXTENSION
Test Position
Subject sitting with forearm supinated and resting on table
Wrist neutral
Stabilize carpals to prevent wrist motion
Normal Range
???
Goniometer Alignment
Axis – carpometacarpal joint Stationary arm – aligned with
radius
Normal End Feel
Capsular
Moving arm – aligned with metacarpal of thumb
THUMB CARPOMETACARPAL JOINT ABDUCTION
Test Position
Subject sitting with forearm resting on table
Wrist neutral Forearm neutral
Stabilize carpals to prevent wrist motion
Normal Range
70o (American Academy of Orthopaedic Surgeons)
Goniometer Alignment
Axis – radial styloid Stationary arm – aligned with
metacarpal of index finger
Moving arm – aligned with metacarpal of thumb
Normal End Feel
Muscle stretch (adductor pollicus, skin, fascia)
THUMB CARPOMETACARPAL JOINT ADDUCTION
Note: Thumb adduction is the return to neutral from thumb abduction. Thumb adduction is rarely measured, probably because it is rarely limited.
Test Position Subject sitting with forearm resting
on table Wrist neutral Forearm neutral
Stabilize carpals to prevent wrist motion
Normal Range 0o ???
Goniometer Alignment Axis – radial styloid Stationary arm – aligned with
metacarpal of index finger
Moving arm – aligned with metacarpal of thumb
Normal End Feel Soft tissue approximation
THUMB CARPOMETACARPAL JOINT OPPOSITION
Note: Opposition of the thumb causes the pad of the thumb to face (oppose) the pads of the fingers. Opposition cannot be measured with a goniometer. The American Academy of Orthopaedic Surgeons suggests that opposition range is normal when the tip of the thumb can touch the base of the fifth finger. When range is not adequate, a ruler can be used to measure the distance between the tip of the thumb and the base of the fifth finger.
Test Position
Subject sitting with forearm supinated and resting on table
Wrist neutral
Stabilize fifth metacarpal
Normal Range
Able to touch tip of thumb to base of fifth finger (American Academy of Orthopaedic Surgeons)
Goniometer Alignment
Goniometer cannot be used
Use a ruler to measure distance between tip of thumb and base of fifth finger
Normal End Feel
Capsular or soft tissue approximation
EXTREMIDAD INFERIOR
HIP FLEXION
Test Position
Subject supine Allow knee to flex (to avoid
limitation by tight hamstrings) Stabilize pelvis to prevent rotation
Flex hip
Normal Range
121.0o + or - 6.4o (American Academy of Orthopaedic Surgeons)
100.0o (American Medical Association)
122.3o (mean), 6.1o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – greater trochanter Stationary arm – aligned with
midline of plevis
Moving arm – aligned with femur (lateral epicondyle)
Normal End Feel
Capsular
HIP EXTENSION
Test Position
Subject prone Stabilize pelvis to prevent rotation Extend hip
Normal Range
12.0o + or - 5.4o (American Academy of Orthopaedic Surgeons)
30.0o (American Medical Association)
9.8o (mean), 6.8o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – greater trochanter
Normal End Feel
Stationary arm – aligned with midline of plevis
Moving arm – aligned with femur (lateral epicondyle)
Capsular or ligamentous
HIP ABDUCTION
Test Position
Subject supine Stabilize pelvis to prevent pelvic
list Abduct hip
Normal Range
41.0o + or - 6.0o (American Academy of Orthopaedic Surgeons)
40.0o (American Medical Association)
45.9o (mean), 9.3o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – anterior superior iliac spine
Normal End Feel
(ASIS) Stationary arm – aligned with
opposite ASIS
Moving arm – aligned with femur (center of patella)
Capsular or ligamentous
HIP ADDUCTION
Test Position
Subject supine Stabilize pelvis to prevent pelvic
list Abduct opposite hip (to allow
room for tested limb to adduct)
Adduct hip
Normal Range
27.0o + or - 3.6o (American Academy of Orthopaedic Surgeons)
20.0o (American Medical Association)
26.9o (mean), 4.1o (standard deviation), (Boone and Azen)
Goniometer Alignment
Axis – anterior superior iliac spine
Normal End Feel
(ASIS) Stationary arm – aligned with
opposite ASIS
Moving arm – aligned with femur (center of patella)
Capsular or ligamentous
HIP MEDIAL (INTERNAL) ROTATION
Test Position Subject sitting on table knee flexed Stabilize distal thigh
medially (internally) rotate hip
Normal Range 44.0o + or - 4.3o (American
Academy of Orthopaedic Surgeons)
40.0o (American Medical Association)
47.3o (mean), 6.0o (standard deviation), (Boone and Azen)
Goniometer Alignment Normal End Feel
Axis – center of patella Stationary arm – aligned vertically
Moving arm – aligned with leg (crest of tibia)
Capsular
HIP LATERAL (EXTERNAL) ROTATION
Test Position Subject sitting on table knee flexed Stabilize distal thigh hip laterally (externally) rotated
Normal Range 44.0o + or - 4.8o (American
Academy of Orthopaedic Surgeons)
50.0o (American Medical Association)
47.2o (mean), 6.3o (standard deviation), (Boone and Azen)
Goniometer Alignment Normal End Feel
Axis – center of patella Stationary arm – aligned vertically
Moving arm – aligned with leg (crest of tibia)
Capsular
KNEE FLEXION
Knee flexion should be measured with the subject supine. This position allows assessment of the joint range of motion without interference from tightness in the rectus femoris muscle. If the examiner wishes to assess length of the rectus femoris, have the patient lie prone (see 2nd illustration).
Test Position
Subject supine Allow hip to flex
Flex knee
Normal Range
141o + or - 5.3o (American Academy of Orthopaedic Surgeons)
150o (American Medical Association)
142.5o (mean), 5.4o, (standard deviation) (Boone and Azen)
Goniometer Alignment Normal End Feel
Axis – lateral epicondyle of femur Stationary arm – aligned with
greater trochanter
Moving arm – aligned with lateral malleolus
Soft tissue approximation
Supine Position (Rectus Femoris Limiting)
KNEE EXTENSION
Test Position
Subject prone Stabilize femur
Extend Knee
Normal Range
minus 2.0o + or - 3.0o (American Academy of Orthopaedic Surgeons)
Goniometer Alignment
Axis – lateral epicondyle of femur Stationary arm – aligned with
greater trochanter
Moving arm – aligned with lateral malleolus
Normal End Feel
Capsular
ANKLE DORSIFLEXION
Pronation of the sub-talar joint can compensate for a loss of ankle joint dorsiflexion range of motion. To avoid measurement error (by accidentally including sub-talar pronation), the sub-talar joint must be stabilized in its neutral position. To assess the range of JOINT motion, flex the knee (first illustration). To assess tightness of the gastrocnemius muscle, extend the knee (second illustration).
Test Position
Subject prone Flex knee Stabilize sub-talar in neutral
Dorsiflex ankle by pushing through 5th metatarsal head
Normal Range
13o + or - 4.4o (American Academy of Orthopaedic Surgeons)
20o (American Medical Association)
12.6o (mean), 4.4o, (standard deviation) (Boone and Azen)
Goniometer Alignment
Axis – lateral malleolus
Normal End Feel
Capsular
Stationary arm – aligned with fibular head
Moving arm – aligned with fifth metatarsal
Assessing Gastrocnemius Tightness (muscle stretch end-feel)
ANKLE PLANTARFLEXION
Test Position
Subject supine Extend knee Stabilize leg
Plantarflex ankle
Normal Range
56o + or - 6.1o (American Academy of Orthopaedic Surgeons)
40o (American Medical Association)
56.2o (mean), 6.1o, (standard deviation) (Boone and Azen)
Goniometer Alignment
Axis – lateral malleolus Stationary arm – aligned with
fibular head
Moving arm – aligned with fifth metatarsal
Normal End Feel
Capsular
CALCANEAL INVERSION
Test Position
Subject prone Stabilize tibia in sagittal plane
(rotate hip or pelvis to align tibia)
Invert calcaneus
Normal Range
2/3 of total range from extreme of inversion to extreme of eversion should be inversion. About 20o inversion (and 10o eversion) on average (Seibel MO: Foot Function: A Programmed Text,p. 72, Baltimore, Williams & Wilkins, 1988)
37.0o + or - 4.5o (American Academy of Orthopaedic Surgeons)
Goniometer Alignment
Axis – automatically positioned by
Normal End Feel
alignment of goniometer arms Stationary arm – aligned with
midline of leg
Moving arm – aligned with midline of calcaneus
Capsular
CALCANEAL EVERSION
Test Position
Subject prone Stabilize tibia in sagittal plane
(rotate hip or pelvis to align tibia)
Evert calcaneus
Normal Range
1/3 of total range from extreme of inversion to extreme of eversion should be eversion. About 10o eversion (and 20o inversion) on average (Seibel MO: Foot Function: A Programmed Text,p. 72, Baltimore, Williams & Wilkins, 1988)
21.0o + or - 5.0o (American Academy of Orthopaedic Surgeons)
Goniometer Alignment
Axis – automatically positioned by alignment of goniometer arms
Stationary arm – aligned with midline of leg
Moving arm – aligned with midline of calcaneus
Normal End Feel
Capsular
MIDTARSAL INVERSION
Test Position
Subject supine Stabilize calcaneus and talus
Invert forefoot
Normal Range
???
Goniometer Alignment Normal End Feel
Axis – automatically positioned by alignment of goniometer arms
Stationary arm – aligned with midline of leg
Moving arm – aligned with plantar aspect of metatarsal heads
Capsular
MIDTARSAL EVERSION
Test Position
Subject supine Stabilize calcaneus and talus
Evert forefoot
Normal Range
???
Goniometer Alignment
Axis – automatically positioned by alignment of goniometer arms
Stationary arm – aligned with midline of leg
Moving arm – aligned with plantar
Normal End Feel
Capsular
aspect of metatarsal heads
METATARSOPHALANGEAL JOINT DORSIFLEXION(Extension)
Range of first metatarsophalangeal (MTP) joint dorsiflexion is functionally important for gait. The available range of 1st MTP joint dorsiflexion depends on the position of the 1st ray. A plantarflexed 1st ray allows greater range of 1st MTP dorsiflexion. I recommend stabilizing the 1st ray in plantarflexion to measure maximum range of 1st MTP dorsiflexion. The first photo demonstrates a good method for measuring 1st or 5th MTP joint dorsiflexion by placing the goniometer alongside the bones. This technique cannot be used for the 2nd, 3rd, or 4th MTP joints. The second photo shows a technique for measuring these joints.
Test Position Subject supine Stabilize 1st metatarsal in
plantarflexion
Dorsiflex MTP
Normal Range 1st - 65o to 75o (slightly less at
lesser MTPs) is the minimum required for normal gait (Root, Orien, Weed. Normal and Abnormal Function of the Foot, pp. 60-61, Clinical Biomechanics Corp., Los Angeles, 1977.)
1st - 50o , 2nd - 40o , 3rd - 30o , 4th - 20o , 5th - 10o (American Medical Association)
Goniometer Alignment Axis – medial to center of
metararsal head Stationary arm – aligned
metatarsal
Moving arm – aligned with proximal phalange
Normal End Feel Capsular
Assessing MTP Dorsiflexion by Placing Goniometer on Dorsum of Bones
(requires modified goniometer)
METATARSOPHALANGEAL JOINT PLANTARFLEXION
(Flexion)
The first photo demonstrates a good method for measuring 1st or 5th MTP joint plantarflexion by placing the goniometer alongside the bones. This technique cannot be used for the 2nd, 3rd, or 4th MTP joints. The second photo shows a technique for measuring these joints.
Test Position
Subject supine Stabilize 1st metatarsal
Plantarflex MTP
Normal Range
1st - 30o , 2nd - 30o , 3rd - 20o , 4th - 10o , 5th - 10o (American Medical Association)
Goniometer Alignment
Axis – medial to center of metararsal head
Stationary arm – aligned metatarsal
Moving arm – aligned with proximal phalange
Normal End Feel
Capsular
Assessing MTP Plantarflexion by Placing Goniometer on Dorsum of Bones
(requires modified goniometer)
METATARSOPHALANGEAL JOINT ABDUCTION
Test Position
Foot flat on table Stabilize metatarsal
Abduct MTP
Normal Range
???
Goniometer Alignment
Axis – dorsum of center of metararsal head
Stationary arm – aligned with metatarsal
Moving arm – aligned with proximal phalange
Normal End Feel
Capsular
METATARSOPHALANGEAL JOINT ADDUCTION
Test Position
Foot flat on table Stabilize metatarsal
Adduct MTP
Normal Range
???
Goniometer Alignment
Axis – dorsum of center of metararsal head
Stationary arm – aligned with metatarsal
Moving arm – aligned with proximal phalange
Normal End Feel
Capsular
COLUMNA VERTEBRAL
CERVICAL SPINE FORWARD BENDING (flexion)
Test Position
Subject sitting with lumbar and thoracic spines supported
Stabilize lumbar and thoracic spines
Flex cervical spine
Normal Range
75.5o + or - 8.5o (20 - 29 yrs.), 70.5o + or - 17.5o (30 - 49 yrs.), 64.5o + or - 7o (>50 yrs.) (American Academy of Orthopaedic Surgeons)
60o (American Medical Association)
Goniometer Alignment
Axis – external auditory meatus Stationary arm – vertical
Moving arm – aligned with nostrils
Normal End Feel
Capsular or ligamentous
CERVICAL SPINE BACKWARD BENDING (extension)
NOTE: The position of the mouth influences the available range of cervical backward bending. With the mouth closed, thghtness of the infrahyoid and
suprahyoid muscles can limit range of cervical backward bending. If you wish to assess the range of the cervical spine, the mouth should be relaxed and slightly
open.
Test Position Subject sitting with lumbar and
thoracic spines supported Stabilize lumbar and thoracic
spines Mouth relaxed and slightly open
Extend cervical spine
Normal Range 75.5o + or - 8.5o (20 - 29 yrs.),
70.5o + or - 17.5o (30 - 49 yrs.), 64.5o + or - 7o (>50 yrs.) (American Academy of Orthopaedic Surgeons)
75o (American Medical Association)
Goniometer Alignment Axis – external auditory meatus Stationary arm – vertical
Moving arm – aligned with nostrils
Normal End Feel Bony or Capsular
CERVICAL SPINE SIDEBENDING
Test Position
Subject sitting with lumbar and thoracic spines supported
Stabilize lumbar and thoracic spines
Sidebend cervical spine
Normal Range (unilateral)
50.5o + or - 5.5o (20 - 29 yrs.), 46.5o + or - 6.5o (30 - 49 yrs.), 40o + or - 8.5o (>50 yrs.) (American Academy of Orthopaedic Surgeons)
45o (American Medical Association)
Goniometer Alignment
Axis – spinous process of C7 Stationary arm – spinous
processes of thoracic spine
Moving arm – posterior midline of head at occipital protuberance
Normal End Feel
Capsular or ligamentous
CERVICAL SPINE ROTATION
Test Position
Subject sitting with lumbar and thoracic spines supported
Stabilize lumbar and thoracic spines
Rotate cervical spine
Normal Range (unilateral)
91.5o + or - 5.5o (20 - 29 yrs.), 81o + or - 6.5o (30 - 49 yrs.), 77.5o + or - 7.5o (>50 yrs.) (American Academy of Orthopaedic Surgeons)
80o (American Medical Association)
Goniometer Alignment
Axis – center of superior aspect of head
Stationary arm – aligned with acromion processes
Moving arm – aligned with tip of nose
Normal End Feel
Capsular or ligamentous
THORACO-LUMBAR SPINE FORWARD BENDING (flexion)
TEST DE SHOBER
NOTE: There are several methods for measuring the range of motion of the lumbar and thoracic spines. Each method has its own advantages and disadvantages (no method is completely valid or reliable, and normal values are not well established for any method). The method illustrated here is a good compromise. Take a baseline measurement with the patient standing upright, then take a second measurement with the subject in the forward bending position. Note the difference.
Test Position
Subject standing
Flex thoracic and lumbar spines
Normal Range
10 cm (Norkin and White)
Tape Measure Alignment
Spinous processes of C7 and S1
Normal End Feel
Capsular or ligamentous
THORACO-LUMBAR SPINE BACKWARD BENDING (extenion)
NOTE: There are several methods for measuring the range of motion of the lumbar and thoracic spines. Each method has its own advantages and disadvantages (no method is completely valid or reliable, and normal values are not well established
for any method). The method illustrated here is a good compromise. Take a baseline measurement with the patient standing upright, then take a second measurement with the subject in the backward bending position. Note the
difference.
Test Position
Subject standing
Extend thoracic and lumbar spines
Normal Range
???
Tape Measure Alignment
Spinous processes of C7 and S1
Normal End Feel
Capsular or ligamentous (sometimes bony)
THORACO-LUMBAR SPINE SIDEBENDING
NOTE: There are several methods for measuring the range of motion of the lumbar and thoracic spines. Each method has its own advantages and disadvantages (no method is completely valid or reliable, and normal values are not well established for any method). The method illustrated here is a good compromise.
Test Position Subject standing Stabilize pelvis
Sidebend thoracic and lumbar spines
Normal Range (unilateral) RIGHT :
o 20 - 29 yrs 37.6o + or - 5.8o o 30 - 39 yrs 35.3o + or - 6.5o o 40 - 49 yrs 27.1o + or - 6.5o o 50 - 59 yrs 25.3o + or - 6.2o o 60 - 69 yrs 20.2o + or - 4.8o o 70 - 79 yrs 18.0o + or - 4.7o o (Fitzgerald, Wynveen,
Rheault et al) LEFT:
o 20 - 29 yrs 38.7o + or - 5.7o o 30 - 39 yrs 36.5o + or - 6.0o o 40 - 49 yrs 28.5o + or - 5.2o
o 50 - 59 yrs 26.8o + or - 6.4o o 60 - 69 yrs 20.3o + or - 5.3o o 70 - 79 yrs 18.9o + or - 6.0o o (Fitzgerald, Wynveen,
Rheault et al)
25o (American Medical Association)
Goniometer Alignment Axis - S1 spinous process Stationary arm - vertical
Moving arm - C7 spinous process
Normal End Feel Capsular or ligamentous
ROTACION DEL TRONCONOTE: There are several methods for measuring the range of motion of the lumbar and thoracic spines. Each method has its own advantages and disadvantages (no method is completely valid or reliable, and normal values are not well established for any method). The method illustrated here is a good compromise.
Test Position Subject sitting Stabilize pelvis Do not allow sidebending, forward
Normal Range (unilateral) 45o (American Medical
Association)
bending or backward bending
Rotate thoracic and lumbar spines
Goniometer Alignment Axis - center of superior aspect of
head Stationary arm - aligned with
anterior superior iliac spines
Moving arm - aligned with acromion processes
Normal End Feel Capsular or ligamentous
TEMPEROMANDIBULAR JOINT OPENING
Ruler Method
Alternate Method
Test Position
Subject sitting Stabilize cervical spine
Open Mouth
Normal Range
35 to 50 mm (Magee)
two and 1/2 flexed PIPs (Friedman and Weisberg)
Ruler Alignment
Use a ruler to measure the distance between the upper and lower incisors
Alternate method - have the subject flex the proximal interphalangeal joints (PIPs) of the fingers and assess how many PIPs can fit between the teeth
Normal End Feel
Capsular or ligamentous
TEMPEROMANDIBULAR JOINT PROTRUSION
Test Position
Subject sitting Stabilize cervical spine
Protrude mandible forward
Normal Range
3 to 5 mm (Magee)
Ruler Alignment
Use a ruler to measure the distance between the upper and lower incisors
Normal End Feel
Capsular or ligamentous
TEMPEROMANDIBULAR JOINT LATERAL DEVIATION
Test Position
Subject sitting Stabilize cervical spine
Deviate mandible laterally
Normal Range
10 to 15mm (Magee)
Ruler Alignment
1 - Identify points on the upper and lower teeth that are aligned when the mouth is in resting position (upper and lower incisors in this illustration)
2 - Deviate the mandible laterally and use a ruler to measure the distance between the two points (upper and lower incisors in this illustration)
Normal End Feel
Capsular or ligamentous