ITC TroubleshootingWhat can go wrong?
Gamze KARAKULLUKÇU, MScApplication Engineer
[email protected] | +90 549 745 30 04
![Page 2: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/2.jpg)
ITC Troubleshooting
The Symptoms:
Baseline position
Baseline drift
Non sigmoidal binding isotherm
Stoichiometry far from 1 or integer value
Split (oscillating) peaks
![Page 3: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/3.jpg)
Heat to Sample Cell = Negative Baseline Deflection
Heat to Reference Cell = Positive Baseline
Deflection
Cooling the Sample Cell = Positive Baseline
Deflection
Cooling the Reference Cell = Negative Baseline
Deflection
0
5
DP
DP
Temperature effects
![Page 4: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/4.jpg)
Some instrumental factors
1. Lab conditions
2. Syringe
3. Stir speeds
4. Temperature setting
5. Plunger tip
![Page 5: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/5.jpg)
0
5
DP
DP
Lab condition influence
![Page 6: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/6.jpg)
0
5
DP
DP
Bent syringe
![Page 7: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/7.jpg)
0
5
DP
MicroCal iTC200 - Syringe holding nut loose
DP
Syringe height
![Page 8: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/8.jpg)
0
5
DP
Faster Stirring
Stir speeds
![Page 9: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/9.jpg)
10 20 30 40 50 60 70
9.8
10.0
10.2
10.4
10.6
10.8
Ba
se
line
Po
sitio
n W
ith
Stirr
ing
(u
ca
l/se
c.)
Experimental Temperature (Deg. C)
Cell848
Cell849
Cell855
Cell857
Cell859
Cell860
Cell862
Cell864
Cell865
Cell866
Cell868
Cell869
###
###
Reference Power = 11 ucal/sec for all data shown below!
Baseline temperature dependence
Reference power=11 ucal/sec for all data shown
below!
![Page 10: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/10.jpg)
Non instrumental factors
Baseline position is the first diagnostic for data
quality-information on:
Air bubbles
Cell cleanliness/“Sticky” proteins
Time between injections
![Page 11: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/11.jpg)
0
5
DP
DP
Bubbles
![Page 12: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/12.jpg)
0.00 20.00
9.6
9.8
10.0
Time (min)
µc
al/
se
c
De-gassing problem
0.00 20.008.28.38.48.58.68.78.88.99.09.19.29.39.49.59.69.79.89.9
Time (min)
µc
al/se
c
![Page 13: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/13.jpg)
Before experiment:
De-gas, use proper filling technique (MicroCal iTC200)
During experiment:
Reload
Gently tap the bottom of the cell with the loading syringe
Re-insert injection syringe
The Cure
![Page 14: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/14.jpg)
-8.33 0.00 8.33 16.67 25.00 33.33 41.67 50.00 58.33 66.67
2
4
6
8
10
12
14
16
Time (min)
µca
l/se
c
Not long enough between injections
Time (min)
mca
l/s
![Page 15: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/15.jpg)
0.00 33.33 66.67 100.00 133.33 166.67
14
15
16
17
18
19
Time (min)
µca
l/se
c
Not long enough between injections
Time (min)
mca
l/s
![Page 16: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/16.jpg)
Increase the time between the injections
This can be done ‘on the fly’ by going to the advanced
experimental tab and ‘update’ the change to injection
parameters > spacing (secs)
The Cure
![Page 17: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/17.jpg)
“Sticky” proteins or cleanliness
0.00 8.33 16.67 25.00 33.33 41.67 50.00
-4
-2
0
2
4
6
Time (min)
µca
l/se
c
-16.67 0.00 16.67 33.33 50.00 66.67 83.33 100.00 116.67 133.33
14
16
Time (min)
µcal/sec
-33.33 0.00 33.33 66.67 100.00 133.33 166.67 200.00 233.33
10
12
14
16
Time (min)
µcal/sec
![Page 18: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/18.jpg)
0.00 16.67
9.38
9.40
9.42
9.44
9.46
9.48
MESdegasb98RAW_CP
MESdegasb98BASE
MESdegasb113RAW_CP
MESdegasb113BASE
Time (min)
µcal/sec
Time (min)
mcal/
s
Dirty cell
![Page 19: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/19.jpg)
Clean the cells with Contrad 70 (Decon 90) and repeat
If the problem persists after cleaning then may have to
“do biochemistry”
Change solution conditions e.g. pH, salt, co-solutes
Change protein construct
The cure
![Page 20: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/20.jpg)
0.00 10.00 20.00 30.009.10
9.20
9.30
9.40
Data1RAW_cp
###
Time (min)
µcal/sec
Nice data but dirty cell
![Page 21: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/21.jpg)
Examine effect on stoichiometry and if error is not
acceptable:
Reload
Clean cell
The cure
![Page 22: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/22.jpg)
Non sigmoidal binding isotherm
No Binding
No Heat
Buffer mismatch
More than one binding event
![Page 23: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/23.jpg)
Effect of DMSO mismatch
-5 0 5 10 15 20 25 30 35 40 45 50 55
0
1
2
3
4
5
6
7
C: 2 % mismatch in DMSO: syringe: 20 ml DMSO added to 1.0 ml buffer; cell: buffer only (no DMSO).
B: Slightly mismatched solution: syringe: 20 ml DMSO added to 1.0 ml buffer;
cell: 280 ml DMSO added to 14 ml buffer.
A: Matched solution: both cell & syringe have same solution (280 ml DMSO added to 14 ml buffer).
Time (min)
µcal/sec
2% DMSO into
2% DMSO-same stock
2% DMSO into
2% DMSO-separate stocks
2% DMSO into
0% DMSOm
ca
l/s
![Page 24: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/24.jpg)
Buffer mismatch-no dialysis
0 20 40 60 80 100 120 140 160 180
-0.5
0.0
0.5
1.0
1.5
2.0
2.5
without dialysis
with dialysis
Time (min)
µca
l/se
c
![Page 25: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/25.jpg)
The cure
Dialyze, or use desalting columns if possible
Check for additives that are in not in both cell and syringe. Also, ask what was sample purified from e.g. was protein lyophilized in buffer and not dialyzed?
Check pH of final solutions-should differ by less than 0.1 pH units. This issue is common when working with high concentrations of ligand-e.g. 500 mM and above-weak binding
![Page 26: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/26.jpg)
-16.67 0.00 16.67 33.33 50.00 66.67 83.33 100.00 116.67 133.33 150.00
7.5
8.0
8.5
9.0
Experiment 2:
ligand into protein 2
Time (min)
µca
l/se
c
Control: ligand
into buffer
Experiment 1:
ligand into protein
1
No
saturation
or protein
not
matched
to ligand
Effect of buffer mismatch and/or low C
value
Time (min)
mca
l/s
![Page 27: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/27.jpg)
Effect of buffer mismatch and/or low C
value
-16.67 0.00 16.67 33.33 50.00 66.67 83.33 100.00 116.67 133.33
29.0
29.5
30.0
30.5
31.0
31.5
Time (min)
µca
l/se
c
Time (min)
mca
l/s
![Page 28: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/28.jpg)
See “mismatch” plus binding may be weaker than you
think
Increase sample concentrations
The cure
![Page 29: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/29.jpg)
ITC: low heat
Heats for
experiment
same as control
0.00 16.67 33.33 50.00 66.67 83.33
4.26
4.28
4.30
4.32
4.34
4.36
4.38
4.40
4.42
4.44
4.46
4.48
4.50
4.52
Time (min)
µcal/sec
Control: ligand into buffer
ExperimentL ligand into protein
![Page 30: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/30.jpg)
Change experimental temperature by at least 10ºC
and/or
Increase sample concentration
The cure
![Page 31: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/31.jpg)
Unexpected stoichiometry
0.00 33.33 66.67 100.00 133.33 166.67
18.6
18.7
18.8
18.9
19.0
19.1
19.2
19.3
Time (min)
µca
l/se
c
![Page 32: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/32.jpg)
Increase protein concentration or decrease ligand
concentration
Make sure protein is active
Also –c-value is high so we may want to change initial
[Lig]/[Prot] ratio
The cure
![Page 33: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/33.jpg)
Split (oscillating) peaks
-8.33 0.00 8.33 16.67 25.00 33.33 41.67 50.00 58.33
-4
-2
0
2
Time (min)
µca
l/se
cOscillating
signal:
Power below 0
Time (min)
mcal/s
![Page 34: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/34.jpg)
Instrument reference power too low
Clean sample and reference cell and increase ‘reference
power’ in advanced experimental tab
Note that ITC data can have exothermic and endothermic
peaks in same injection.
The cure
![Page 35: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/35.jpg)
Insoluble Ligands
‘One site’ interactions are symmetrical and as such the
ligand can be put in the cell and the protein in the syringe
Reverse titration
![Page 36: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/36.jpg)
-1.5
-1.0
-0.5
0.0
-10 0 10 20 30 40 50 60 70 80 90 100 110 120
Time (min)
µca
l/se
c
0.0 0.5 1.0 1.5 2.0 2.5
-12
-10
-8
-6
-4
-2
0
2
Molar Ratio
kca
l/m
ole
of in
jecta
nt
-0.8
-0.6
-0.4
-0.2
0.0
-10 0 10 20 30 40 50 60 70 80 90 100
Time (min)
µca
l/se
c
0.0 0.5 1.0 1.5 2.0 2.5
-14
-12
-10
-8
-6
-4
-2
0
2
Molar Ratio
kca
l/m
ole
of in
jecta
nt
Parameter Ligand in syringe Ligand in cell
N 0.99 0.97
KD 104 nM 105 nM
DG -9.4 kcal/mol -9.4 kcal/mol
DH -11.9 kcal/mol -12.4 kcal/mol
TDS -2.5 kcal/mol -3.0 kcal/mol
29.5µM Protein titrated with
1.1mM Compound
11.5µM Compound titrated with
179µM Protein
Cure-reverse the titration
![Page 37: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/37.jpg)
Stoichiometry
“N” is the average number of binding sites per mole of
protein in your solution, assuming:
that all binding sites are identical and independent
that you have pure protein (and ligand)
that you have given the correct protein and ligand
concentrations
that all your protein is correctly folded and active
![Page 38: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/38.jpg)
Stoichiometry not expected value
If stoichiometry less than expected value:
Protein concentration lower than you think
Ligand concentration higher than you think
Impure ligand and/or macromolecule
Misfolded protein (esp. for recombinants)
If 0.5, 1 ligand binding to 2 macromolecules (dimer?)
Binding does not fit simple independent model
![Page 39: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/39.jpg)
Stoichiometry not expected value
If stoichiometry greater than expected value
Multiple binding sites on protein
Protein concentration higher than you think
Ligand concentration lower than you think
Binding does not fit simple independent model
![Page 40: ITC Troubleshooting - Atomika Teknik · ITC Troubleshooting What can go wrong? Gamze KARAKULLUKÇU, MSc Application Engineer ... experimental tab and ‘update’ the change to injection](https://reader034.vdocuments.mx/reader034/viewer/2022052104/603eece2eec6cf4b4954b237/html5/thumbnails/40.jpg)
ITC general conclusions!
Take care with sample preparation
Use appropriate run parameters
Check initial baseline position
Perform water-water titration to remove instrumental issues
De-gassing
Load carefully
Clean cell thoroughly, Clean cell thoroughly, Clean cell thoroughly