EDX Marco Cantoni 1 MSE-603 2009 Introduction to EDX Energy Dispersive X-ray Microanalysis (EDS, Energy dispersive Spectroscopy) EDX Marco Cantoni 2 MSE-603 2009 Basics of EDX a) Generation of X-rays b) Detection Si(Li) Detector, SDD Detector, EDS (<-> WDS) c) Quantification EDX in SEM, Interaction volume Monte-Carlo-Simulations EDX in TEM [ d) EDX in SEM-STEM ]

Upload: duongdiep

Post on 26-Aug-2018

268 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 1MSE-603 2009

Introduction to EDXEnergy Dispersive X-ray Microanalysis

(EDS, Energy dispersive Spectroscopy)

EDX Marco Cantoni 2MSE-603 2009

Basics of EDX

a) Generation of X-rays

b) DetectionSi(Li) Detector, SDD Detector, EDS (<-> WDS)

c) QuantificationEDX in SEM, Interaction volumeMonte-Carlo-SimulationsEDX in TEM

[ d) EDX in SEM-STEM ]

Page 2: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 3MSE-603 2009

Inelastic scattering of electrons at atomsEelectron_in > Eelectron_out

• Continuum X-ray production(Bremsstrahlung, Synchrotron)

•• Continuum XContinuum X--ray ray productionproduction((BremsstrahlungBremsstrahlung, , Synchrotron)Synchrotron)

SE

SE, BSE, EELS

•Inner shell ionization

••Inner shell Inner shell ionizationionization

EDX Marco Cantoni 4MSE-603 2009

Core shell ionisation: chemical microanalysis by X-ray, Auger electrons and Electron Energy Loss

Spectrometries

e-

K

L1 L2L3

KL2L3

Emission Auger

+K

L1 L2L3

Kα2

RX

Emission X

+

K

L1 L2L3

e-

e-

Ionisation

+

Ka1 Ka2 Kb

La1 La2

KL1L2 KL1L3 KL2L3

L1M1M2

M5

M4

M3

M2

M1

L3

L2

L1

K

Rayons X Electrons Auger

1ps

Page 3: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 5MSE-603 2009

Emission of characteristic X-ray and Auger electron

EDX Marco Cantoni 6MSE-603 2009

Designation of x-ray emission lines

Page 4: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 7MSE-603 2009

EDX Marco Cantoni 8MSE-603 2009

Efficiency of X-ray generationRelative efficiency of X-ray and Auger emission vs. atomic number for K lines

Ionization cross-section vs. overvoltage U=Eo/Eedge

(electron in -> X-ray out)

To ionized the incident electron MUST have an energy larger than the core shell level U>1. To be efficient, it should have about twice the edge energy U>2.

Light element atoms return to fundamental state mainly by Auger emission. For that reason, their K-lines are weak. In addition their low energy makes them easily absorbed.

SEM TEM ->Light elements

Auger Spectroscopy

Heavy elements

EDS

Cu-K 8.1kV, HT 15kVU = 15/8.1 = 1.85

Page 5: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 9MSE-603 2009

X-ray production vs. atomic number Z

Low efficiencyfor light

elements!

EDX Marco Cantoni 10MSE-603 2009

Characteristic lines: Moseley's Law

EDS range ~ 0.3-20 keV

To assess an elementall detectables lines

MUST be present!!!

!

known ambiguities:

Al Kα = Br LlS Kα = Mo Ll

Page 6: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 11MSE-603 2009

Moseley‘s law for K-series

Frequency ν of X-rays emitted from K-level vs. atomic number

E= hν et λ=c/ν

with the Planck constant:h=6.626 068 76(52) × 10-34 J·sand 1eV = 1.6 10-19 J

( )215 1Z4810.2 −=ν

So, lets measure the X-rays emitted from my sample and determine the composition !

But how to detect it ?

Energy of characteristic XEnergy of characteristic X--rayray--> Element> ElementQualitative EDXQualitative EDX--AnalysisAnalysis

EDX Marco Cantoni 12MSE-603 2009

b) Detection of X-rays (EDX)

Page 7: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 13MSE-603 2009

EDX Marco Cantoni 14MSE-603 2009

X-Ray energy conversion to electrical charges:3.8eV / electron-hole pair in averageelectronic noise+ imperfect charge collection:130 eV resolution / Mn Ka line

• Detector acts like a diode: at room temperaturethe leak current for 1000V would be too high !

• The FET produces less noise if cooled !• Li migration at room temperature !• ->Detector cooling by L-N

Page 8: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 15MSE-603 2009

Pulse detection and analysis

Pulse detection: Charge ~ energy

Shorter time constant = process time to analyze voltage-> peak broadening (lower energy resolution)

Longer time constant (higher energy resolution)

->pulse rejection (dead time)

EDX Marco Cantoni 16MSE-603 2009

Process timetime constant

Long

low count rate

Higher dead time

narrow peaks

short

high count rate

lower dead time

broad peaks~2’00050

~4’50030

~11’00015

~30’0004

Count rate, cts/sec.

(30% dead time)

Time const.

(μsec.)

resolutionpeak identification

beam currentcount rate

Page 9: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 17MSE-603 2009

Detection and artifacts

X-Ray energy conversion to electrical charges:3.8eV / electron-hole pair in averageelectronic noise+ imperfect charge collection:130 eV resolution / Mn Ka line

Take care when looking for “trace” elements (low concentrations). Don’t confuse small peaks with escape peaks !

EDX Marco Cantoni 18MSE-603 2009

Wavelength Dispersive Spectroscopy (WDS)

The specimen, the diffracting crystal and the detector stay on the Rowland circle. To scan the wavelength, this circle rotates around the specimen to satisfyingthe Bragg law

Johansson focusing spectrometer:The diffracting crystal is bent with a curvatureradius double of that of the Rowland circleThe crystal surface is cylindrically ground to the radius of the Rowland circle

Page 10: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 19MSE-603 2009

Electron Microprobe EPMA (Electron Probe MicroAnalyser)with WDS

EDX Marco Cantoni 20MSE-603 2009

5-30min.1 min.Acquisition time

1000:1100:1

Peak to background

ratio

difficult++StandardlessAnalysis

difficultEasyUse

≈5eV80-180eVEnergy Resolution

WDSEDS

EDS <-> WDS

Page 11: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 21MSE-603 2009

(K,Na)NbO3

Overvoltage,10keV

Continuum,Bremsstrahlung

Electron beam: 10keV Duane-Hunt limit

EDX Marco Cantoni 22MSE-603 2009

Bremstrahlung (background)

• when a charged particle (des-) accelarates or changes direction, it emits an electromagnetic wave.

– This is widely used to produce synchrotron radiation

• On a bulk sample of atomic number Z:• N(E) is the number of photons of energy E, E0 the energy of the incident electron and K the

Kramers constant

( ) ( )E

EEKZEN 0 −

=

Page 12: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 23MSE-603 2009

EDS in SEM• Acquisition under best conditions

– Flat surface without contamination(no Au coating, use C instead)

– Sample must be homogenous at the place of analysis (interaction volume !!)

– Horizontal orientation of the surface– High count rate (but dead time below 30%)– Overvoltage U=Eo/Ec >1.5-2

•• For acquisition times of 100sec. :For acquisition times of 100sec. :detection of ~0.5at% for almost all elementsdetection of ~0.5at% for almost all elements

EDX Marco Cantoni 24MSE-603 2009

(K,Na)NbO3

Overvoltage,10keV

Continuum,Bremsstrahlung

Not ideal !

Duane-Hunt limit

Page 13: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 25MSE-603 2009

60.4820.368.667.82Min.

61.5121.1310.259.79Max.

100.0061.1420.869.408.60Spectrum 8

100.0060.4920.3710.258.89Spectrum 7

100.0060.8420.639.079.46Spectrum 6

100.0061.0220.779.358.86Spectrum 5

100.0060.4820.369.379.79Spectrum 4

100.0061.5121.139.547.82Spectrum 3

100.0061.0020.758.669.59Spectrum 2

100.0060.9320.7010.188.19Spectrum 1

TotalONbKNaSpectrum

(K,Na)NbO3

EDX Marco Cantoni 26MSE-603 2009

c) Quantification

• First approach:compare X-ray intensity with a standard (sample with known concentration, same beam current of the electron beam)

ci: wt concentration of element i

Ii: X-ray intensity of char. Line

ki: concentration ratio

istdi

istdi

i kI

I

c

c==

Yes, but…..

Page 14: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 27MSE-603 2009

Intensity ~ Concentration…?

How manydifferent

samples…?

EDX Marco Cantoni 28MSE-603 2009

Page 15: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 29MSE-603 2009

Electron Flight Simulator

EDX Marco Cantoni 30MSE-603 2009

Casino

Page 16: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 31MSE-603 2009

EDX Marco Cantoni 32MSE-603 2009

QuantificationWhen the going gets tough…..

[ ] istdi

istdi

i kI

I

c

cFAZ ==××

• "Z" describe how the electron beampenetrates in the sample (Z-dependant and density dependant) and loose energy

• "A" takes in account the absorption of the X-rays photons along the path to samplesurface

• "F" adds some photons when (secondary) fluorescence occurs

Correction matrix

Page 17: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 33MSE-603 2009

Flow chart of quantificationMeasure the intensities

and calculate the concentrationswithout ZAF corrections

Calculate the ZAF correctionsand the density of the sample

Calculate the concentrations with the corrections

Is the differencebetween the new and the old

concentrations smaller than the calculation error?

no Yes !stop

EDX Marco Cantoni 34MSE-603 2009

Correction methods:

• ZAF (purely theoretical)

• PROZA Phi-Rho-Z

• PaP (Pouchou and Pichoir)

• XPP (extended Puchou/Pichoir)

• with standards (same HT, current, detector settings)

• Standardless: theoretical calculation of Istd

• Standardless optimized: « hidden » standards, user definedpeak profiles

Page 18: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 35MSE-603 2009

Quantitative EDX in SEM• Acquisition under best conditions

– Flat surface without contamination, horizontal orientation of the surface (no Au coating, use C instead)

– Sample must be homogenous at the place of analysis (interaction volume !!)

– High count rate (but dead time below 30%)

– Overvoltage U=Eo/Ec >1.5-2

• For acquisition times of 100sec. :detection of ~0.5at% possible for almost all elements

•• StandardlessStandardless acquisitionacquisition possible with high accuracy (intensities of references under the given conditions can be calculated for a great range of elements), test with samples of known composition, light elements (like O) are critical…

• Spatial resolution depends strongly on HT and the density of the sample

EDX Marco Cantoni 36MSE-603 2009

Demo NSS/INCA

• Peak finding, synthetic spectrum

• Spectrum imaging (extraction of elements)

Page 19: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 37MSE-603 2009

Modern EDX systems:

• User friendly interfaces• New and more powerful

electronics (stability of calibrations, higher count rate)

• Drift compensation for long acquisition times (element mapping on CM300 at high mag, “sitelock”)

• Synthesized spectra(spectrum overlay)easier identification

• Advanced element mapping: Spectral imaging (data cube), selection of elements and regions post-acquisition

• Powerfull reporting and Export tools (Word, Powerpoint, html, tifetc.)

EDX Marco Cantoni 38MSE-603 2009

Synthesized spectrum

Spectrum imaging

Data cube

Extraction of element maps

Page 20: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 39MSE-603 2009

EDS in TEM

PZT bulk

20nm thick PZTHigh spatial resolution !

EDX Marco Cantoni 40MSE-603 2009

EDS in TEM• Thin samples -> correction

factors weak (A and F can be neglected)

• Very weak beam broadening -> high spatial resolution ~ beam diameter (~nm)

High energy: artifacts !

If only there wasn’t this specimen preparation……

Page 21: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 41MSE-603 2009

STEM point analysisSTEM point analysisPbMg1/3Nb2/3O3 (bulk)

Processing option : Oxygen by stoichiometry (Normalised)

Spectrum Mg Si Nb Pb O Total Spectrum 1 30.02 13.32 56.66 100.00 Spectrum 2 19.15 7.96 4.11 11.72 57.06 100.00 Spectrum 3 6.01 12.49 22.13 59.37 100.00 Spectrum 4 5.65 12.39 22.67 59.29 100.00 Spectrum 5 5.63 12.48 22.52 59.36 100.00 Spectrum 6 5.98 13.66 20.11 60.25 100.00 Spectrum 7 5.55 12.45 22.66 59.34 100.00 Spectrum 8 5.49 12.96 21.84 59.72 100.00 Spectrum 9 5.63 12.19 23.04 59.14 100.00 Max. 30.02 13.32 13.66 23.04 60.25 Min. 5.49 7.96 4.11 11.72 56.66

All results in Atomic Percent

EDX Marco Cantoni 42MSE-603 2009

STEM STEM linescanlinescanPb(Zr,Ti)O3 (thick film), slight Pb excess

Page 22: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 43MSE-603 2009

STEM Element MappingSTEM Element MappingPMN/PT 90/10 (bulk)

EDX Marco Cantoni 44MSE-603 2009

Artifactshow to recognize/minimize them

Page 23: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 45MSE-603 2009

EDSdetector

upper obj. lens polepiece

lower obj. lens polepiece

X-Ray collection in TEM EDS

collimator

sampleholder collection solid angle

EDX Marco Cantoni 46MSE-603 2009

EDSdetector

upper obj. lens polepiece

lower obj. lens polepiece

X-Ray collection in TEM EDS

collimator

sampleholder collection solid angle

Page 24: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 47MSE-603 2009

EDSdetector

upper obj. lens polepiece

lower obj. lens polepiece

Stray X-Rays in TEM EDS

collimator

sampleholder collection solid angle

X-rays shower

hole count

EDX Marco Cantoni 48MSE-603 2009

EDSdetector

upper obj. lens polepiece

lower obj. lens polepiece

Stray X-Rays in TEM EDS

collimator

back-scattered e-

to thick sample

Page 25: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 49MSE-603 2009

EDSdetector

upper obj. lens polepiece

lower obj. lens polepiece

Stray X-Rays in TEM EDS

collimator

scattered e- to polepieces + thick sample

EDX Marco Cantoni 50MSE-603 2009

EDSdetector

upper obj. lens polepiece

lower obj. lens polepiece

Stray X-Rays in TEM EDS: characteristic x-rays

collimator

Page 26: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 51MSE-603 2009

EDSdetector

upper obj. lens polepiece

lower obj. lens polepiece

Stray X-Rays in TEM EDS: continuum (Bremsstrahlung)

collimator

EDX Marco Cantoni 52MSE-603 2009

Ideal samples:

• FIB samples: almost uniform thickness, smallsample size(less bulk material around)

Page 27: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 53MSE-603 2009

EDS in TEM

• Thin samples -> correction factors weak (A and F can be neglected), quantification “easy”

• Very weak beam broadening -> high spatial resolution ~ beam diameter (~nm)

• High energy -> artifacts

• Sample preparation, sample geometry……

EDX Marco Cantoni 54MSE-603 2009

Bonus

Page 28: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 55MSE-603 2009

SEM

TEMholey Carbon film

KNbO3

Nano-rods

EDX ofPowders

nano particlesSEM or TEM ???

EDX ofEDX ofPowdersPowders

nanonano particlesparticlesSEM or TEM ???SEM or TEM ???

Optical microscopereflected light

EDX Marco Cantoni 56MSE-603 2009

Simulation Electron Flight Simulator EFS

Sample holder

C film

x

x

Page 29: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 57MSE-603 2009

SEM 10kVTEM 300kV

x

x

EDX Marco Cantoni 58MSE-603 2009

SEM 10kVSEM-STEM 30kV

x

x

Page 30: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 59MSE-603 2009

SEM 10kV STEM 30kV

• Easy interpretation“no” contribution from substrate (C)

• MBTF corrections for quantification

• Interaction volume >> particle size• Sample holder analyzed• No deconvolution possible• Low HT = limited energy range for

ionization energies

EDX Marco Cantoni 60MSE-603 2009

SEM/STEM

TEM sample

BSE-Detector (upside-down)

XL30XL30--SFEGSFEG

Annular dark Annular dark fieldfield

Bright fieldBright field

““Poor manPoor man’’s TEMs TEM””

Page 31: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 61MSE-603 2009

SEMSEM--STEM BFSTEM BF SEMSEM--STEM ADFSTEM ADF

KNbO3

EDX Marco Cantoni 62MSE-603 2009

SEM-STEM BF SE detector

Page 32: Introduction to EDX - CIME | EPFL · Introduction to EDX ... Spectrum 1 8.19 10.18 20.70 60.93 100.00 Spectrum Na K Nb O Total (K,Na)NbO3 MSE-603 2009 EDX Marco Cantoni 26 c) Quantification

EDX Marco Cantoni 63MSE-603 2009

SEMSEM--STEM ADF 30kV, FIB lamellaSTEM ADF 30kV, FIB lamella

SEMSEM--SE 30kVSE 30kV

EDX Marco Cantoni 64MSE-603 2009

FIBSEM SE of TEM sampleSEM SE of TEM sample

SEM STEM ADF, 30kVSEM STEM ADF, 30kV

“true” STEM, CM300, 300kV. ADF